Abstract

This document specifies requirements and gives recommendations for next generation sequencing (NGS) workflows. This document covers the pre-examination processes, human RNA isolation, sequencing library preparation, sequencing, sequence analysis and reporting of the examination of sequences for diagnostic purposes from isolated RNA from, e.g. formalin-fixed and paraffin embedded tissues, fresh frozen tissues, fine needle aspirates (FNA), whole blood, circulating tumour cells (CTCs), exosomes and other extracellular vesicles, and circulating cell free RNA from plasma. NOTE 1 Typical applications include, but are not limited to, NGS for oncology and clinical genetics, certain single-cell analyses. This document is addressing in vitro diagnostic examinations including laboratory developed tests and is applicable to medical laboratories, molecular pathology laboratories and molecular genetic laboratories, in vitro diagnostic developers and manufacturers, biobanks as well as institutions and organizations performing biomedical research or sequencing services. This document is not applicable for in situ sequencing, forensic sequencing, sequencing of pathogens or microorganisms and microbiome analysis. NOTE 2 International, national or regional regulations or requirements or multiples of them can also apply to specific topics covered in this document.

Status
Not Published
Public Enquiry End Date
29-Nov-2026
Technical Committee
VAZ - Healthcare
Current Stage
4020 - Public enquire (PE) (Adopted Project)
Start Date
24-Sep-2026
Due Date
11-Feb-2027

Buy Documents

Draft

oSIST prEN ISO 25379-2:2026

English language (67 pages)
Preview
Preview
e-Library read for
×1 day

oSIST prEN ISO 25379-2:2026 is SIST’s adoption of the draft ISO document for in vitro diagnostic next generation sequencing (NGS) workflows for human RNA examination. It specifies requirements and recommendations for the full workflow - pre-examination, RNA isolation, sequencing library preparation, sequencing, sequence analysis, and reporting - for laboratories, developers, manufacturers, biobanks, and organizations that provide biomedical research or sequencing services.

What does oSIST prEN ISO 25379-2:2026 specify?

oSIST prEN ISO 25379-2:2026 specifies how an in vitro diagnostic NGS workflow for human RNA should be designed, developed, verified, validated, implemented, reported, and quality-controlled. It covers isolated RNA from formalin-fixed and paraffin embedded (FFPE) tissue, fresh frozen tissue, fine needle aspirates (FNA), whole blood, circulating tumour cells (CTCs), exosomes and other extracellular vesicles (EVs), and circulating cell free RNA (ccfRNA) from plasma.

The standard is written for in vitro diagnostic examinations, including laboratory developed examinations, and it is also relevant to biomedical research and sequencing services. It does not apply to in situ sequencing, forensic sequencing, sequencing of pathogens or microorganisms, or microbiome analysis.

The document is organized into general requirements, pre-examination processes, examination processes, reporting, implementation into routine practice, reporting and interpretation of results, and quality assurance procedures.

AnnexWhat it covers
Annex ANormative workflow for single-cell analyses
Annex BNormative exemplary workflow for spatial transcriptomics
Annex CInformative workflow scheme for RNA examination

What are the key requirements of oSIST prEN ISO 25379-2:2026?

oSIST prEN ISO 25379-2:2026 treats the RNA NGS workflow as one connected system, from specimen handling to the final report. In practice, that means the laboratory or manufacturer has to control the pre-analytical, analytical, and post-analytical steps together, not as separate activities.

Quality system, design, and intended use

  • Clause 4.1 requires a medical laboratory quality management system based on ISO 15189, or an equivalent system such as ISO/IEC 17020 or ISO/IEC 17025. For IVD manufacturers, ISO 13485 is the recommended quality system. This matters because the workflow has to fit inside a managed, documented process.
  • Clauses 4.2 and 4.3 require scientific validity, documented customer and stakeholder needs, a defined intended use, design input specifications, and development aligned with Clauses 5 and 6. The standard also points to risk assessment, Design of Experiments (DoE), and guard band studies. In practice, the assay must be designed for the clinical question it is meant to answer.
  • Clause 4.4 requires documented verification and validation plans, and it covers the whole workflow, not just one instrument or reagent. Products used in the workflow need verification or validation for their intended use. That means the claimed performance has to be demonstrated before routine use.
  • Clause 4.5 requires technical performance characteristics such as analytical accuracy, measurement uncertainty, analytical sensitivity or positive percentage agreement (PPA) when no reference method is available, analytical specificity or negative percentage agreement (NPA), positive predictive value (PPV), precision, limit of detection (LOD), limit of detection for platforms (LODP) where applicable, reportable range, reference range, interferences, sequencing base quality, and clinical performance characteristics. This is the evidence base for deciding whether the assay is fit for its purpose.

Pre-examination and examination controls

  • Clause 5 requires source-specific RNA isolation and quality evaluation for the specimen types covered by the standard, including FFPE tissue, fresh frozen tissue, FNA, whole blood, CTCs, EVs, and plasma ccfRNA. The practical point is that extraction and quality checks must match the specimen type and the intended sequencing strategy.
  • Clause 6 covers sequencing library preparation, sequencing equipment requirements, sequencing quality control, and data analysis. Data analysis has to follow ISO 20397-2, and the bioinformatics pipeline has to be fit for the intended purpose. In practice, the laboratory has to control both wet-lab and software steps.

Reporting, implementation, and quality assurance

  • Clause 7 requires reports that are accurate, clear, unambiguous, and interoperable with the bioinformatics pipeline and, where relevant, the medical information system. It also requires report attributes such as sample-quality comments, acceptance or rejection comments, critical results, and interpretive comments. For mutations, HGVS convention or equivalents should be used.
  • Clause 7 also requires that a whole transcriptome result be handled through a virtual reporting system so that only disease-relevant findings are disclosed in the clinical report.
  • Clause 8 requires the laboratory to follow the manufacturer’s instructions, to control cross-contamination, to monitor quality metrics for each run, to use controls, to have safety procedures, and to keep a written standard operating procedure. Where a laboratory uses a laboratory developed examination, it takes responsibility for specification, verification, validation, documentation, use, and performance.
  • Clause 9 requires timely reporting, documented interpretation procedures, standardized analysis tools and parameters, defined critical results, and clear result wording such as “Detected” or “Not detected” instead of “Positive” or “Negative”.
  • Clause 10 requires ongoing quality assurance, regular performance monitoring, and proficiency testing where appropriate. This matters because NGS performance can drift over time as reagents, operators, databases, and analysis rules change.

What terms does oSIST prEN ISO 25379-2:2026 define?

  • next generation sequencing (NGS) - advanced sequencing technologies developed after dideoxy sequencing, including massively parallel sequencing (MPS), short-read sequencing, and long-read sequencing.
  • massively parallel sequencing (MPS) - nucleic acid sequencing that reads millions or billions of molecules in parallel.
  • RNA sequencing (RNA-Seq) - a technique that determines the complete or partial sequence of RNA.
  • sequencing library - DNA, cDNA, or RNA prepared for NGS with adaptors and a suitable size range.
  • sequencing library preparation - the procedures used to add tags and sequencing primer binding regions to RNA or DNA fragments for NGS.
  • laboratory developed examination - an examination manufactured or modified and used inside a health institution or medical laboratory for needs that an equivalent commercial IVD examination cannot meet.
  • bioinformatics pipeline - the set of bioinformatics tools that processes NGS data.
  • limit of detection for platforms (LODP) - the lowest relative amount of reference material for which positive identification can be achieved with defined confidence.

Who uses oSIST prEN ISO 25379-2:2026?

The standard is used by medical laboratories, molecular pathology laboratories, and molecular genetic laboratories that perform RNA-based NGS examinations. It is also relevant to IVD developers and manufacturers, biobanks, and institutions or organizations that provide biomedical research or sequencing services.

In practice, they use it to design assays, choose specimen types, validate performance, control quality, write reports, and decide how to implement or monitor RNA NGS workflows in routine use.

Which standards are used with oSIST prEN ISO 25379-2:2026?

StandardContribution
ISO 15189:2022Core medical laboratory quality and competence requirements; also used for collection, reporting, and quality assurance
ISO/IEC 17020:2012 and ISO/IEC 17025:2017Alternative quality and competence frameworks for inspection and testing/calibration bodies
ISO 13485Quality management requirements for IVD manufacturers
ISO 20397-1:2022Library preparation for massively parallel sequencing
ISO 20397-2Quality evaluation of sequencing data and data analysis
ISO/TS 20658Sample collection, transport, receipt, and handling
ISO 20166-1, ISO 20184-1, ISO 20186-1, ISO 7552-1, ISO 18702, ISO 18703Pre-examination processes for FFPE tissue, frozen tissue, venous whole blood, CTCs, exosomes/EVs, and plasma ccfRNA
ISO 15190Safety requirements for medical laboratories
ISO 8601-1Standardized date and time format in reports
ISO 5649:2024Concepts and specifications for laboratory developed tests

What does the oSIST prEN ISO 25379-2:2026 document contain?

The document contains requirements and recommendations for the complete RNA NGS workflow, including pre-examination handling, RNA isolation, library preparation, sequencing, data analysis, reporting, implementation, and quality assurance. It also includes detailed clauses on verification, validation, technical performance characteristics, and report content.

Annex A gives a normative workflow for single-cell analyses, including the special handling needed for rare or very small RNA inputs. Annex B gives a normative exemplary workflow for spatial transcriptomics on fresh frozen tissue, including fixation, staining, imaging, permeabilization, cDNA synthesis, sequencing, and image-based data visualization. Annex C provides an informative workflow scheme that shows the RNA examination process and its optional or reversible steps.

Relations

Effective Date
19-Feb-2025

Buy Documents

Draft

oSIST prEN ISO 25379-2:2026

English language (67 pages)
Preview
Preview
e-Library read for
×1 day

Get Certified

Connect with accredited certification bodies for this standard

BSI Group

BSI (British Standards Institution) is the business standards company that helps organizations make excellence a habit.

UKAS United Kingdom Verified

TÜV Rheinland

TÜV Rheinland is a leading international provider of technical services.

DAKKS Germany Verified

TÜV SÜD

TÜV SÜD is a trusted partner of choice for safety, security and sustainability solutions.

DAKKS Germany Verified

Sponsored listings

Frequently Asked Questions

oSIST prEN ISO 25379-2:2026 is a draft published by the Slovenian Institute for Standardization (SIST). Its full title is "In vitro diagnostic Next Generation Sequencing (NGS) workflows - Part 2: Human RNA examination (ISO/DIS 25379-2:2026)". This standard covers: This document specifies requirements and gives recommendations for next generation sequencing (NGS) workflows. This document covers the pre-examination processes, human RNA isolation, sequencing library preparation, sequencing, sequence analysis and reporting of the examination of sequences for diagnostic purposes from isolated RNA from, e.g. formalin-fixed and paraffin embedded tissues, fresh frozen tissues, fine needle aspirates (FNA), whole blood, circulating tumour cells (CTCs), exosomes and other extracellular vesicles, and circulating cell free RNA from plasma. NOTE 1 Typical applications include, but are not limited to, NGS for oncology and clinical genetics, certain single-cell analyses. This document is addressing in vitro diagnostic examinations including laboratory developed tests and is applicable to medical laboratories, molecular pathology laboratories and molecular genetic laboratories, in vitro diagnostic developers and manufacturers, biobanks as well as institutions and organizations performing biomedical research or sequencing services. This document is not applicable for in situ sequencing, forensic sequencing, sequencing of pathogens or microorganisms and microbiome analysis. NOTE 2 International, national or regional regulations or requirements or multiples of them can also apply to specific topics covered in this document.

This document specifies requirements and gives recommendations for next generation sequencing (NGS) workflows. This document covers the pre-examination processes, human RNA isolation, sequencing library preparation, sequencing, sequence analysis and reporting of the examination of sequences for diagnostic purposes from isolated RNA from, e.g. formalin-fixed and paraffin embedded tissues, fresh frozen tissues, fine needle aspirates (FNA), whole blood, circulating tumour cells (CTCs), exosomes and other extracellular vesicles, and circulating cell free RNA from plasma. NOTE 1 Typical applications include, but are not limited to, NGS for oncology and clinical genetics, certain single-cell analyses. This document is addressing in vitro diagnostic examinations including laboratory developed tests and is applicable to medical laboratories, molecular pathology laboratories and molecular genetic laboratories, in vitro diagnostic developers and manufacturers, biobanks as well as institutions and organizations performing biomedical research or sequencing services. This document is not applicable for in situ sequencing, forensic sequencing, sequencing of pathogens or microorganisms and microbiome analysis. NOTE 2 International, national or regional regulations or requirements or multiples of them can also apply to specific topics covered in this document.

oSIST prEN ISO 25379-2:2026 is classified under the following ICS (International Classification for Standards) categories: 11.100.10 - In vitro diagnostic test systems. The ICS classification helps identify the subject area and facilitates finding related standards.

oSIST prEN ISO 25379-2:2026 has the following relationships with other standards: It is inter standard links to SIST-TS CEN/TS 17981-2:2024. Understanding these relationships helps ensure you are using the most current and applicable version of the standard.

oSIST prEN ISO 25379-2:2026 is available in PDF format for immediate download after purchase. The document can be added to your cart and obtained through the secure checkout process. Digital delivery ensures instant access to the complete standard document.

Standards Content (Sample)

Contents

  • Introduction
  • 1 Scope
  • 2 Normative references
  • 3 Terms and definitions
  • 3.1 aliquot
  • 3.2 amplicon
  • 3.3 analyte
  • 3.4 analytical accuracy
  • 3.5 analytical sensitivity
  • 3.6 analytical specificity
  • 3.7 bioinformatics pipeline
  • 3.8 biomarker
  • 3.9 circulating cell free RNA
  • 3.10 clinical accuracy
  • 3.11 clinical performance
  • 3.12 clinical sensitivity
  • 3.13 clinical specificity
  • 3.14 clinical utility
  • 3.15 clinical validity
  • 3.16 closed system
  • 3.17 Design of Experiments
  • 3.18 deoxyribonuclease
  • 3.19 deoxyribonucleic acid
  • 3.20 diagnosis
  • 3.21 examination
  • 3.22 examination manufacturer
  • 3.23 examination performance
  • 3.24 extracellular vesicle
  • 3.25 fine needle aspirate
  • 3.26 formalin
  • 3.27 fraser
  • 3.28 guard band
  • 3.29 in vitro diagnostic NGS workflow
  • 3.30 intended use
  • 3.31 intermediate precision
  • 3.32 intermediate precision condition of measurement
  • 3.33 isoform
  • 3.34 laboratory developed examination
  • 3.35 limit of detection
  • 3.36 limit of detection for platforms
  • 3.37 long-read sequencing
  • 3.38 massively parallel sequencing
  • 3.39 measurand
  • 3.40 measurement
  • 3.41 measurement procedure
  • 3.42 next generation sequencing
  • 3.43 no template control
  • 3.44 OUTRIDER
  • 3.45 panel
  • 3.46 performance characteristic
  • 3.47 performance specification
  • 3.48 pharmacogenomics
  • 3.49 post-examination processes
  • 3.50 pre-examination processes
  • 3.51 proficiency testing
  • 3.52 protein
  • 3.53 proteinase
  • 3.54 reference material
  • 3.55 reference method
  • 3.56 reference range
  • 3.57 repeatability
  • 3.58 repeatability condition of measurement
  • 3.59 reportable range
  • 3.60 reproducibility
  • 3.61 reproducibility condition of measurement
  • 3.62 reverse transcription
  • 3.63 reverse transcription dPCR
  • 3.64 reverse transcription qPCR
  • 3.65 ribonuclease
  • 3.66 ribonucleic acid
  • 3.67 RNA sequencing
  • 3.68 sample
  • 3.69 scientific validity
  • 3.70 sequencing library
  • 3.71 sequencing library preparation
  • 3.72 short-read sequencing
  • 3.73 specimen
  • 3.74 stability
  • 3.75 validation
  • 3.76 venous whole blood

Introduction

Molecular in vitro diagnostics has enabled significant progress in medicine. Further progress is expected by new technologies analysing profiles of nucleic acids, proteins, and metabolites in human tissues and body fluids. Next Generation Sequencing (NGS) takes a prominent place in the series of molecular techniques used for diagnostics. It facilitates sequence analysis of nucleic acids that can result in precise information for diagnosis and progression of diseases.

The NGS technique, however, has a very complex workflow that contains many steps. The target nucleic acids can originate from different sources, e.g. tissues, blood, and body fluids. The profiles of the isolated RNA can change during specimen collection, transport, storage and processing (e.g. formalin fixation) making the outcome from diagnostics or research unreliable or even impossible because the subsequent analytical assay will not determine the situation in the patient but an artificial profile generated during the pre-examination process. The available material can be small, the cells in a tissue can be dispersed heterogeneously (e.g. ratio of tumour to normal), the target nucleic acids can be circulating in blood or body fluids free of cells or in circulating cells (e.g. circulating tumour cells (CTCs)). For a successful and reliable sequencing result, a suitable strategy needs to be chosen for every case depending on the available material and disease conditions. Therefore, the NGS workflow can differ from case to case, and the NGS workflow steps need to be carefully considered and chosen to get a sound and reliable result to determine the best available treatment for the patient. In addition, sequencing platforms can differ in their technique (e.g. detection of a change in a current or fluorescence) and approach (e.g. whole transcriptome, panels, short-read sequencing, longread sequencing) for sequence assessment. The bioinformatics analysis can differ in approach and ability to detect non-conformities and unreliable sequencing results. To enable such capabilities, NGS metadata needs to be collected during all workflow steps from the patient to the reporting. In addition, controls and added controls need to be analysed properly. This way non-conformities or detected unreliabilities can be reported to the patient and the treating physician. The reporting of diagnostic NGS results can differ in clarity and depth, which can lead to different interpretations.

Standardization of the entire NGS workflow from specimen collection to the reporting of the results to the patient and the treating physician is needed for the development of reliable NGS examinations. This document draws upon previous work to standardize the steps for NGS examinations from tissues, blood and body fluids in what is referred to as the pre-examination phase (sample collection), the examination phase (library preparation, sequencing), and the post-examination phase (analysis and reporting). In this document, the following verbal forms are used:

  • “shall” indicates a requirement;
  • “should” indicates a recommendation;
  • “may” indicates a permission;
  • “can” indicates a possibility or a capability.

oSIST prEN ISO 25379-2:2026

oSIST prEN ISO 25379-2:2026 DRAFT International Standard ISO/DIS 25379-2:2026(en) In vitro diagnostic Next Generation Sequencing (NGS) workflows — Part 2:

Human RNA examination

1 Scope

This document specifies requirements and gives recommendations for next generation sequencing (NGS) workflows for in vitro diagnostics and biomedical research. This document covers the pre-examination processes including human RNA isolation, sequencing library preparation, sequencing, sequence analysis and reporting of the examined sequences for diagnostic purposes from isolated RNA from, e.g. formalin-fixed and paraffin embedded tissues, fresh frozen tissues, fine needle aspirates (FNA), whole blood, circulating tumour cells (CTCs), exosomes and other extracellular vesicles, and circulating cell free RNA from plasma. NOTE 1 Typical applications include, but are not limited to, NGS for oncology and clinical genetics, certain singlecell analyses. This document addresses in vitro diagnostic examinations including laboratory developed examinations and is applicable to medical laboratories, molecular pathology laboratories and molecular genetic laboratories, in vitro diagnostic developers and manufacturers, biobanks as well as institutions and organizations performing biomedical research or sequencing services.

This document is not applicable for in situ sequencing, forensic sequencing, sequencing of pathogens or microorganisms and microbiome analysis.

NOTE 2 International, national or regional regulations or requirements or multiples of them can also apply to specific topics covered in this document.

2 Normative references

The following documents are referred to in the text in such a way that some or all of their content constitutes requirements of this document. For dated references, only the edition cited applies. For undated references, the latest edition of the referenced document (including any amendments) applies. ISO 7552-1, Molecular in vitro diagnostic examinations — Specifications for pre-examination processes for circulating tumor cells (CTCs) in venous whole blood — Part 1: Isolated RNA ISO 18702, Molecular in vitro diagnostic examinations — Specifications for pre-examination processes for exosomes and other extracellular vesicles in venous whole blood — DNA, RNA and proteins ISO 18703, Molecular in vitro diagnostic examinations — Specifications for pre-examination processes for venous whole blood — Isolated circulating cell free RNA from plasma ISO 15189:2022, Medical laboratories — Requirements for quality and competence ISO/IEC 17020:2012, Conformity assessment — Requirements for the operation of various types of bodies performing inspection ISO/IEC 17025:2017, General requirements for the competence of testing and calibration laboratories ISO 20166-1, Molecular in vitro diagnostic examinations — Specifications for pre-examination processes for formalin-fixed and paraffin-embedded (FFPE) tissue — Part 1: Isolated RNA

oSIST prEN ISO 25379-2:2026 ISO/DIS 25379-2:2026(en) ISO 20184-1:2018, Molecular in vitro diagnostic examinations — Specifications for pre-examination processes for frozen tissue — Part 1: Isolated RNA ISO 20186-1, Molecular in vitro diagnostic examinations — Specifications for pre-examination processes for venous whole blood — Part 1: Isolated cellular RNA ISO 8601-1, Date and time — Representations for information interchange — Part 1: Basic rules ISO 20397-1:2022, Biotechnology — Massively parallel sequencing — Part 1: Nucleic acid and library preparation ISO 20397-2, Biotechnology — Massively parallel sequencing — Part 2: Quality evaluation of sequencing data

3 Terms and definitions

For the purposes of this document, the terms and definitions given in ISO 15189 and the following apply. ISO and IEC maintain terminology databases for use in standardization at the following addresses:

  • ISO Online browsing platform: available at https:// www .iso .org/ obp
  • IEC Electropedia: available at https:// www .electropedia .org/

3.1 aliquot

portion of a larger amount of homogenous material, assumed to be taken with negligible sampling error Note 1 to entry: The term is usually applied to fluids. Tissues are heterogeneous and therefore cannot be aliquoted. [1] [2] [3] Note 2 to entry: The definition is derived from , , .

[SOURCE: ISO 20184-3:2021, 3.1, modified ─ Note 2 to entry was added.]

3.2 amplicon

specific DNA fragment produced by a DNA-amplification technology, such as the polymerase chain reaction (PCR) [SOURCE: ISO 13495:2013, 3.3.1]

3.3 analyte

component represented in the name of a measurable quantity Note 1 to entry: The term “analyte”, or the name of a substance or compound, is not to be confused with the term “measurand”, because analytes are not quantities.

EXAMPLE Amount of substance of glucose in plasma is a measurand (3.39), while glucose is the analyte. [SOURCE: ISO 17511:2021, 3.1, modified — Note 1 and example were added]

3.4 analytical accuracy

closeness of the agreement between the result of an examination (3.21) and a true property value Note 1 to entry: For NGS-based examinations, accuracy represents the degree of concordance (or agreement) of results between a sequence obtained from the examination and the same sequence determined by a valid comparator method, or between a reference sample run on an NGS-based examination and the high confidence sequence of the reference. [SOURCE: ISO/IEC Guide 99:2007, 2.13, modified – Original notes were deleted, Note 1 to entry was added; "analytical accuracy" replaces all old terms, "a measured quantity value" was replaced by "the result of an examination" and "quantity value of a measurand" was replaced by "property value".]

oSIST prEN ISO 25379-2:2026 ISO/DIS 25379-2:2026(en)

3.5 analytical sensitivity

sensitivity of a measurement procedure quotient of the change in a measurement indication and the corresponding change in a value of a quantity being measured Note 1 to entry: The sensitivity of a measurement procedure (3.41) can depend on the value of the quantity being measured.

Note 2 to entry: The change considered in the value of the quantity being measured only becomes useful compared with the resolution.

Note 3 to entry: The analytical sensitivity of a measuring system is the slope of the calibration curve. Note 4 to entry: Analytical sensitivity is often confused with positive percentage agreement (PPA), because similar calculations are used for both the analytical sensitivity and the PPA. However, the term PPA instead of analytical sensitivity is only used, if a reference method is not available or not used. The PPA represents merely an estimation of the analytical sensitivity. PPA is applicable only to measurements of quantities that have the nature of a count. [SOURCE: ISO/IEC Guide 99:2007, 4.12, modified — terms changed from "sensitivity of a measuring system /sensitivity" changed to "analytical sensitivity/sensitivity of a measurement procedure"; "an indication of a measuring system" replaced by "a measurement indication", "measuring system" replaced by "measurement procedure" in Note 1; new Note 3 and Note 4 were added.]

3.6 analytical specificity

selectivity of a measurement procedure capability of a measuring system, using a specified measurement procedure (3.41), to provide measurement results for one or more measurands (3.39) which do not depend on each other nor on any other quantity in the system undergoing measurement (3.40) Note 1 to entry: Lack of analytical specificity is called analytical interference (see ISO 18113-1:2022, 3.2.2). Note 2 to entry: Specificity of a measurement procedure (3.40) should not be confused with diagnostic specificity (see ISO 18113-1:2022, 3.2.18).

Note 3 to entry: ISO/IEC Guide 99:2007 uses the term selectivity for this concept instead of specificity. Note 4 to entry: Adapted from ISO/IEC Guide 99:2007, 4.13.

Note 5 to entry: Selectivity as used in metrology is a concept close to specificity as it is sometimes used in chemistry. “Specificity” of measurements has some risk of ambiguity with “clinical/diagnostic specificity”. Note 6 to entry: Analytical specificity is often confused with negative percentage agreement (NPA), because the same calculations are used for both the analytical specificity and the NPA. However, the term NPA instead of analytical specificity is only used, if a reference method is not available or is not used. The NPA represents merely an estimation of the analytical specificity.

[SOURCE: ISO 18113-1:2022, 3.2.5, modified —Note 2 was deleted and 5 and 6 were added.]

3.7 bioinformatics pipeline

suite of different bioinformatics tools that process the NGS data

3.8 biomarker

molecular marker detectable and/or quantifiable molecule or group of molecules used to indicate a biological condition, state, identity or characteristic or an organism [SOURCE: ISO 16577:2022, 3.4.28]

oSIST prEN ISO 25379-2:2026 ISO/DIS 25379-2:2026(en)

3.9 circulating cell free RNA

ccfRNA extracellular human RNA (3.65) present in blood and plasma Note 1 to entry: ccfRNA includes RNA present in vesicles such as exosomes.

3.10 clinical accuracy

diagnostic accuracy extent of agreement between the outcome of the examination (3.21) and the target clinical condition

3.11 clinical performance

ability of an examination (3.21) to yield results that are correlated with a particular clinical condition, physiological or pathological state in accordance with the target population and intended use(r) Note 1 to entry: Although sometimes referred to as diagnostic performance or clinical validity; clinical performance is the harmonized term endorsed by the Global Harmonization Task Force (GHTF) and its successor, the International Medical Devices Regulators Forum (IMDRF).

Note 2 to entry: Evaluation of clinical performance often relies on the outcome of other types of clinical examinations to define “true positive or true negative” results.

3.12 clinical sensitivity

diagnostic sensitivity ability of an in vitro diagnostic examination procedure to have positive results associated with a particular disease or condition Note 1 to entry: Also defined as percent positivity in samples where the target marker is known to be present. Note 2 to entry: Diagnostic sensitivity is expressed as a percentage (number fraction multiplied by 100), calculated as 100 × the number of true positive values (TP) divided by the sum of the number of true positive values (TP) plus the number of false negative values (FP), or 100 × TP/(TP + FN). This calculation is based on a study design where only one sample is taken from each subject.

Note 3 to entry: The target condition is defined by criteria independent of the examination procedure under consideration.

Note 4 to entry: Clinical sensitivity is often confused with positive percentage agreement (PPA), because the same calculations are used for both the clinical sensitivity and the PPA. However, the term PPA instead of clinical sensitivity is only used, if a reference method is not available or is not used. The PPA represents merely an estimation of the clinical sensitivity.

[SOURCE: ISO 18113-1:2022, 3.2.17, modified — The term "clinical sensitivity" was added; second sentence in Note 1 was deleted; Note 4 was added.]

3.13 clinical specificity

diagnostic specificity ability of an in vitro diagnostic examination procedure to have negative results associated with an absence of particular disease or condition Note 1 to entry: Also defined as percent negativity in samples where the target marker is known to be absent. For information regarding description of the diagnostic performance characteristics of an IVD medical device, see Reference [4].

Note 2 to entry: Clinical specificity is expressed as a percentage (number fraction multiplied by 100), calculated as 100 × the number of true negative values (TN) divided by the sum of the number of true negative plus the number of false positive values (FP), or 100 × TN/(TN + FP). This calculation is based on a study design where only one sample is taken from each subject.

oSIST prEN ISO 25379-2:2026 ISO/DIS 25379-2:2026(en) Note 3 to entry: The target condition is defined by criteria independent of the examination procedure under consideration.

Note 4 to entry: Clinical specificity is often confused with negative percentage agreement (NPA), because the same calculations are used for both the clinical specificity and the NPA. However, the term NPA instead of clinical specificity is only used, if a reference method is not available or is not used. The NPA represents merely an estimation of the clinical specificity.

[SOURCE: ISO 18113-1:2022, 3.2.18, modified — The term "clinical specificity" was added; Note 4 was added.]

3.14 clinical utility

ability of a screening or diagnostic examination to prevent or ameliorate adverse health outcomes such as mortality, morbidity, or disability through the adoption of efficacious treatments conditioned on examination results [5] [6] Note 1 to entry: Clinical utility can be part of scientific validity and clinical performance (3.11) , . [6] [SOURCE: ]

3.15 clinical validity

predictive value of an examination (3.21) for a given clinical outcome Note 1 to entry: Clinical validity is primarily determined by the clinical sensitivity (3.12) and clinical specificity (3.13) with which an examination identifies individuals with a defined clinical condition within a given population. The clinical validity of a genetic examination is the likelihood that, e.g. cancer will develop in someone with a positive examination result.

3.16 closed system

non-modifiable system (by the user) provided by the vendor including all necessary components for the examination Note 1 to entry: Necessary components are hardware, software, procedures and reagents.

3.17 Design of Experiments

DoE <NGS> structured, statistical approach for planning sequencing runs and experimental conditions in order to systematically evaluate the effects of key variables (e.g. library preparation methods, input nucleic acid quantity, sequencing depth, and indexing strategies) on sequencing quality, coverage, and downstream data interpretation

3.18 deoxyribonuclease

DNase enzyme that catalyzes the degradation of DNA (3.19) into smaller components [SOURCE: ISO 20186-3:2019, 3.9]

3.19 deoxyribonucleic acid

DNA polymer of deoxyribonucleotides occurring in a double-stranded (dsDNA) or single-stranded (ssDNA) form

3.20 diagnosis

identification of a disease from its signs and symptoms, where the diagnostic process can involve examinations (3.21) and tests for classification of an individual's condition into separate and distinct categories or subclasses that allow medical decisions about treatment and prognosis to be made

oSIST prEN ISO 25379-2:2026 ISO/DIS 25379-2:2026(en)

3.21 examination

analytical test set of operations having the objective of determining the numerical value or characteristics of a property Note 1 to entry: An examination can be the total of a number of activities, observations or measurements required to determine a value or characteristic.

Note 2 to entry: Laboratory examinations that determine a numerical value of a property are called "quantitative examinations"; those that determine the characteristics of a property are called "qualitative examinations". Note 3 to entry: Laboratory examinations are also called "assays" or "tests". Note 4 to entry: A “measurement procedure” or “measurement system” becomes an “examination” (also in complex terms like “examination procedure”), where nominal properties (e.g. DNA bases like G, A, T, C, or a set of terms for the types of leukocytes found in human blood) or ordinal quantities (e.g. a 5-value scale (0, 1, 2, 3, 4) for dip-stick reading (assessed using 1 = not elevated, 2 = doubtfully elevated, 3 = slightly elevated, 4 = elevated, 5 = strongly elevated)) are the results of a “detection” instead of a “measurement” [derived from ISO/DIS 15193] . [SOURCE: ISO 15189:2022, 3.8, modified — The term "analytical test" was added; note 1 was added.]

3.22 examination manufacturer

analytical test manufacturer entity that manufactures and/or produces measurement or detection systems, instruments, and reagents for a specific examination (3.21)

...