Medical electrical equipment -- Part 1-6: General requirements for safety - Collateral standard: Usability

Medical practice is increasingly using medical electrical equipment for observation and treatment of patients. use errors caused by inadequate medical electrical equipment usability have become an increasing cause for concern. The usability engineering process is intended to achieve reasonable usability, which in turn is intended to minimise use errors and to minimise use associated risks. Some, but not all, forms of incorrect use are amenable to control by the manufacturer. The usability engineering process is part of the process of risk control. This Collateral Standard describes a usability engineering process, and provides guidance on how to implement and execute the process to provide medical electrical equipment safety. It addresses normal use and use errors but excludes abnormal use.

Medizinische elektrische Geräte -- Teil 1-6: Allgemeine Festlegungen für die Sicherheit - Ergänzungsnorm: Gebrauchstauglichkeit

Appareils électromédicaux -- Partie 1-6: Règles générales de sécurité - Norme collatérale: Aptitude à l'utilisation

Les appareils électromédicaux sont de plus en plus utilisés en pratique médicale pour l'observation et le traitement des patients. Les erreurs d'utilisation dues à une aptitude à l'utilisation inadéquate des appareils électromédicaux sont devenues une préoccupation majeure. Le processus d'ingénierie de l'aptitude à l'utilisation est destiné à obtenir une aptitude à l'utilisation raisonnable qui, a son tour, est destinée à minimiser les erreurs d'utilisation et à minimiser les risques associés à l'utilisation. Certaines formes d'utilisation incorrecte, mais pas toutes, sont maîtrisables par le constructeur. Le processus d'ingénierie de l'aptitude à l'utilisation fait partie du processus de maîtrise du risque. La présente norme collatérale décrit un processus d'ingénierie de l'aptitude à l'utilisation et fournit des lignes directrices pour la mise en oeuvre et l'exécution de ce processus pour contribuer à la sécurité des appareils électromédicaux. Elle traite de l'utilisation normale et des erreurs d'utilisation mais exclut l'utilisation anormale.

Medicinska električna oprema – 1-6. del: Splošne varnostne zahteve – Kolateralni standard: Uporabnost

General Information

Status
Withdrawn
Publication Date
31-Dec-2004
Withdrawal Date
24-Oct-2007
Current Stage
9900 - Withdrawal (Adopted Project)
Start Date
24-Oct-2007
Due Date
16-Nov-2007
Completion Date
25-Oct-2007

Relations

Buy Standard

Standard
EN 60601-1-6:2005
English language
78 pages
sale 10% off
Preview
sale 10% off
Preview
e-Library read for
1 day

Standards Content (Sample)

EUROPEAN STANDARD EN 60601-1-6
NORME EUROPÉENNE
EUROPÄISCHE NORM November 2004

ICS 11.040


English version


Medical electrical equipment
Part 1-6: General requirements for safety –
Collateral standard: Usability
(IEC 60601-1-6:2004)


Appareils électromédicaux Medizinische elektrische Geräte
Partie 1-6: Règles générales de sécurité - Teil 1-6: Allgemeine Festlegungen
Norme collatérale: Aptitude à l'utilisation für die Sicherheit –
(CEI 60601-1-6:2004) Ergänzungsnorm: Gebrauchstauglichkeit
(IEC 60601-1-6:2004)



This European Standard was approved by CENELEC on 2004-09-01. CENELEC members are bound to
comply with the CEN/CENELEC Internal Regulations which stipulate the conditions for giving this European
Standard the status of a national standard without any alteration.

Up-to-date lists and bibliographical references concerning such national standards may be obtained on
application to the Central Secretariat or to any CENELEC member.

This European Standard exists in three official versions (English, French, German). A version in any other
language made by translation under the responsibility of a CENELEC member into its own language and
notified to the Central Secretariat has the same status as the official versions.

CENELEC members are the national electrotechnical committees of Austria, Belgium, Cyprus, Czech
Republic, Denmark, Estonia, Finland, France, Germany, Greece, Hungary, Iceland, Ireland, Italy, Latvia,
Lithuania, Luxembourg, Malta, Netherlands, Norway, Poland, Portugal, Slovakia, Slovenia, Spain, Sweden,
Switzerland and United Kingdom.

CENELEC
European Committee for Electrotechnical Standardization
Comité Européen de Normalisation Electrotechnique
Europäisches Komitee für Elektrotechnische Normung

Central Secretariat: rue de Stassart 35, B - 1050 Brussels


© 2004 CENELEC - All rights of exploitation in any form and by any means reserved worldwide for CENELEC members.

Ref. No. EN 60601-1-6:2004 E

---------------------- Page: 1 ----------------------

EN 60601-1-6:2004 - 2 -
Foreword
The text of document 62A/452/FDIS, future edition 1 of IEC 60601-1-6, prepared by SC 62A, Common
aspects of electrical equipment used in medical practice, of IEC TC 62, Electrical equipment in
medical practice, was submitted to the IEC-CENELEC parallel vote and was approved by CENELEC
as EN 60601-1-6 on 2004-09-01.
The following dates were fixed:
– latest date by which the EN has to be implemented
at national level by publication of an identical
national standard or by endorsement (dop) 2005-06-01
– latest date by which the national standards conflicting
with the EN have to be withdrawn (dow) 2007-09-01
This European Standard is a collateral standard to EN 60601-1:1990, hereinafter referred to as the
general standard.
In the 60601 series of publications, collateral standards specify general requirements for safety
applicable to
– a group of MEDICAL ELECTRICAL EQUIPMENT (e.g. radiological equipment); or
– a specific characteristic of all MEDICAL ELECTRICAL EQUIPMENT, not fully addressed in the general
standard (e.g. electromagnetic compatibility).
The numbering of sections, clauses and subclauses of this collateral standard corresponds with that of
the general standard.
Clauses, subclauses, tables and figures which are additional to those of the general standard are
numbered starting from 201; additional annexes are lettered AAA, BBB, etc., and additional items
aaa), bbb), etc.
In this collateral standard, the following print types are used:
– requirements and definitions: roman type;
– notes, examples, explanations, advice, introductions, general statements and references: smaller roman type;
– test specifications: italic type;
– TERMS DEFINED IN CLAUSE 2 OF THE GENERAL STANDARD, IN THIS COLLATERAL STANDARD OR AS NOTED
IN THE INDEX OF DEFINED TERMS: SMALL CAPITALS.
The requirements are followed by specifications for the relevant tests.
Clauses and subclauses for which a rationale is provided in the informative Annex AAA are marked
with an asterisk (*).
Annex ZA has been added by CENELEC.
__________
Endorsement notice
The text of the International Standard IEC 60601-1-6:2004 was approved by CENELEC as a
European Standard without any modification.

---------------------- Page: 2 ----------------------

- 3 - EN 60601-1-6:2004

In the official version, for Bibliography, the following notes have to be added for the standard indicated:
ISO 9000 NOTE Harmonized as EN ISO 9000:2000 (not modified)
ISO 9001 NOTE Harmonized as EN ISO 9001:2000 (not modified)
ISO 9241-11 NOTE Harmonized as EN ISO 9241-11:1998 (not modified)
ISO 13485 NOTE Harmonized as EN ISO 13485:2003 (not modified)
__________

---------------------- Page: 3 ----------------------

EN 60601-1-6:2004 - 4 -
Annex ZA
(normative)

Normative references to international publications
with their corresponding European publications
The following referenced documents are indispensable for the application of this document. For dated
references, only the edition cited applies. For undated references, the latest edition of the referenced
document (including any amendments) applies.
NOTE Where an international publication has been modified by common modifications, indicated by (mod), the relevant
EN/HD applies.
Publication Year Title EN/HD Year
IEC 60601-1 1988 Medical electrical equipment EN 60601-1 1990
Part 1: General requirements for safety + Corr. Juli 1994
A1 1991 A1 1993
+ Corr. Juli 1994
A2 1995
+ corr. June 1995 A2 1995
A13 1996

IEC 60601-1-8 2003 Medical electrical equipment EN 60601-1-8 2004
Part 1-8: General requirements for safety
- Collateral standard: General
requirements, tests and guidance for
alarm systems in medical electrical
equipment and medical electrical systems

ISO 14971 2000 Medical devices - Application of risk EN ISO 14971 2000
management to medical devices + corr. February 2002
A1 2003 A1 2003

---------------------- Page: 4 ----------------------

NORME CEI
INTERNATIONALE
IEC



60601-1-6
INTERNATIONAL


Première édition
STANDARD

First edition

2004-06


Appareils électromédicaux –
Partie 1-6:
Règles générales de sécurité –
Norme collatérale: Aptitude à l'utilisation

Medical electrical equipment –
Part 1-6:
General requirements for safety –
Collateral Standard: Usability

 IEC 2004 Droits de reproduction réservés  Copyright - all rights reserved
Aucune partie de cette publication ne peut être reproduite ni No part of this publication may be reproduced or utilized in any
utilisée sous quelque forme que ce soit et par aucun procédé, form or by any means, electronic or mechanical, including
électronique ou mécanique, y compris la photocopie et les photocopying and microfilm, without permission in writing from
microfilms, sans l'accord écrit de l'éditeur. the publisher.
International Electrotechnical Commission, 3, rue de Varembé, PO Box 131, CH-1211 Geneva 20, Switzerland
Telephone: +41 22 919 02 11 Telefax: +41 22 919 03 00 E-mail: inmail@iec.ch Web: www.iec.ch
CODE PRIX
XB
PRICE CODE
Commission Electrotechnique Internationale
International Electrotechnical Commission
МеждународнаяЭлектротехническаяКомиссия
Pour prix, voir catalogue en vigueur
For price, see current catalogue

---------------------- Page: 5 ----------------------

60601-1-6  IEC:2004 – 3 –
CONTENTS
FOREWORD.7
INTRODUCTION.11

SECTION ONE – GENERAL

1 Scope and object.13
1.201 Scope.13
1.202 Relationship to other standards.13
1.202.1 IEC 60601-1 .13
1.202.2 Particular Standards .13
1.202.3 Normative references.13
2 Terminology and definitions.15
3 General requirements .19
6 Identification, markings and documents .19
6.8 ACCOMPANYING DOCUMENTS .19
6.8.1 General .19
6.8.2 Instructions for use .19
6.8.201 TRAINING and materials for TRAINING .19

SECTIONS TWO TO SIX – NOT USED

SECTION SEVEN – PROTECTION AGAINST EXCESSIVE TEMPERATURES AND
OTHER SAFETY HAZARDS

46* USE ERROR and USABILITY.21
46.201 SAFETY for the PATIENT, OPERATOR and other persons .21
46.202 USABILITY ENGINEERING PROCESS.21
46.202.1 General.21
46.202.2 Input for the USABILITY ENGINEERING PROCESS.23
46.202.3 USABILITY SPECIFICATION .25
46.202.4 USABILITY VERIFICATION .27
46.202.5 USABILITY VALIDATION plan .27
46.202.6 USABILITY VALIDATION.27

SECTION EIGHT TO TEN – NOT USED

Annex AAA (informative) General guidance and rationale .29
Annex BBB (informative) A Taxonomy of OPERATOR action.41
Annex CCC (informative) Examples of USE ERRORS, ABNORMAL USE and design flaws
potentially leading to USE ERRORS .43
Annex DDD (informative) Guidance on the USABILITY ENGINEERING PROCESS .49
Annex EEE (informative) Sample USABILITY SPECIFICATION.107
Annex FFF (informative) Reference documents .125

---------------------- Page: 6 ----------------------

60601-1-6  IEC:2004 – 5 –
Bibliography.143
Terminology – Index of defined terms .147
Figure BBB.1 – Summary of the taxonomy of OPERATOR action .41
Figure DDD.1 – An OPERATOR-EQUIPMENT INTERFACE design cycle.55
Figure DDD.2 – Bubble diagram of the conceptual model of a physiological monitor .85
Figure EEE.1 – Example of a USABILITY SPECIFICATION for a hypothetical device .107
Table DDD.1 – Sample of design flaws and associated USE ERRORS .53
Table DDD.2 – Mapping of Figure DDD.1 to the subclauses of this standard .55
Table DDD.3 – Examples of OPERATOR-EQUIPMENT INTERFACE requirements .63
Table DDD.4 – Typical deliverables .75
Table DDD.5 – Examples of objective and subjective USABILITY goals .83
Table DDD.6 – Examples of OPERATOR-EQUIPMENT INTERFACE modelling techniques .87
Table DDD.7 – Characteristics of a typical USABILITY testing effort .87

---------------------- Page: 7 ----------------------

60601-1-6  IEC:2004 – 7 –
INTERNATIONAL ELECTROTECHNICAL COMMISSION
___________

MEDICAL ELECTRICAL EQUIPMENT –

Part 1-6: General requirements for safety –
Collateral Standard: Usability


FOREWORD
1) The International Electrotechnical Commission (IEC) is a worldwide organization for standardization comprising
all national electrotechnical committees (IEC National Committees). The object of IEC is to promote
international co-operation on all questions concerning standardization in the electrical and electronic fields. To
this end and in addition to other activities, IEC publishes International Standards, Technical Specifications,
Technical Reports, Publicly Available Specifications (PAS) and Guides (hereafter referred to as “IEC
Publication(s)”). Their preparation is entrusted to technical committees; any IEC National Committee interested
in the subject dealt with may participate in this preparatory work. International, governmental and non-
governmental organizations liaising with the IEC also participate in this preparation. IEC collaborates closely
with the International Organization for Standardization (ISO) in accordance with conditions determined by
agreement between the two organizations.
2) The formal decisions or agreements of IEC on technical matters express, as nearly as possible, an international
consensus of opinion on the relevant subjects since each technical committee has representation from all
interested IEC National Committees.
3) IEC Publications have the form of recommendations for international use and are accepted by IEC National
Committees in that sense. While all reasonable efforts are made to ensure that the technical content of IEC
Publications is accurate, IEC cannot be held responsible for the way in which they are used or for any
misinterpretation by any end user.
4) In order to promote international uniformity, IEC National Committees undertake to apply IEC Publications
transparently to the maximum extent possible in their national and regional publications. Any divergence
between any IEC Publication and the corresponding national or regional publication shall be clearly indicated in
the latter.
5) IEC provides no marking procedure to indicate its approval and cannot be rendered responsible for any
equipment declared to be in conformity with an IEC Publication.
6) All users should ensure that they have the latest edition of this publication.
7) No liability shall attach to IEC or its directors, employees, servants or agents including individual experts and
members of its technical committees and IEC National Committees for any personal injury, property damage or
other damage of any nature whatsoever, whether direct or indirect, or for costs (including legal fees) and
expenses arising out of the publication, use of, or reliance upon, this IEC Publication or any other IEC
Publications.
8) Attention is drawn to the Normative references cited in this publication. Use of the referenced publications is
indispensable for the correct application of this publication.
9) Attention is drawn to the possibility that some of the elements of this IEC Publication may be the subject of
patent rights. IEC shall not be held responsible for identifying any or all such patent rights.
International Standard IEC 60601-1-6 has been prepared by subcommittee 62A: Common
aspects of electrical equipment used in medical practice, of IEC technical committee 62:
Electrical equipment in medical practice.
This first edition constitutes a Collateral Standard to IEC 60601-1:1998, Medical electrical
equipment – Part 1: General requirements for safety, hereinafter referred to as the General
Standard.

---------------------- Page: 8 ----------------------

60601-1-6  IEC:2004 – 9 –
The text of this Collateral Standard is based on the following documents:
FDIS Report of voting
62A/452/FDIS 62A/458/RVD
Full information on the voting for the approval of this Collateral Standard can be found in the
report on voting indicated in the above table.
In the 60601 series of publications, Collateral Standards specify general requirements for
safety applicable to:
– a subgroup of MEDICAL ELECTRICAL EQUIPMENT (e.g. radiological equipment); or
– a specific characteristic of all MEDICAL ELECTRICAL EQUIPMENT, not fully addressed in the
General Standard (e.g. electromagnetic compatibility).
The numbering of sections, clauses and subclauses of this Collateral Standard corresponds
with that of the General Standard.
Clauses, subclauses, tables and figures which are additional to those of the General Standard
are numbered starting from 201; additional annexes are lettered AAA, BBB, etc., and
additional items aaa), bbb), etc.
In this Collateral Standard, the following print types are used:
– requirements and definitions: roman type;
– notes, examples, explanations, advice, general statements and references: smaller roman type;
– test specifications: italic type; and
– TERMS DEFINED IN CLAUSE 2 OF THE GENERAL STANDARD OR IN THIS COLLATERAL STANDARD OR
AS NOTED IN THE INDEX OF DEFINED TERMS: SMALL CAPITALS.
The requirements are followed by specifications for the relevant tests.
Clause and subclauses for which a rationale is provided in informative Annex AAA are marked
with an asterisk (*).
The committee has decided that the contents of this publication will remain unchanged until
the maintenance result date indicated on the IEC web site under "http://webstore.iec.ch" in
the data related to the specific publication. At this date, the publication will be
• reconfirmed;
• withdrawn;
• replaced by a revised edition, or
• amended.

---------------------- Page: 9 ----------------------

60601-1-6  IEC:2004 – 11 –
INTRODUCTION
Medical practice is increasingly using MEDICAL ELECTRICAL EQUIPMENT for observation and
treatment of PATIENTS. USE ERRORS caused by inadequate MEDICAL ELECTRICAL EQUIPMENT
USABILITY have become an increasing cause for concern. The USABILITY ENGINEERING PROCESS
is intended to achieve reasonable USABILITY, which in turn is intended to minimise USE ERRORS
and to minimise use associated RISKS. Some, but not all, forms of incorrect use are amenable
to control by the manufacturer. The USABILITY ENGINEERING PROCESS is part of the PROCESS of
RISK CONTROL.
This Collateral Standard describes a USABILITY ENGINEERING PROCESS, and provides guidance
on how to implement and execute the PROCESS to provide MEDICAL ELECTRICAL EQUIPMENT
SAFETY. It is intended to be useful not only for manufacturers of MEDICAL ELECTRICAL
EQUIPMENT, but also for technical committees responsible for the preparation of particular
standards.

---------------------- Page: 10 ----------------------

60601-1-6  IEC:2004 – 13 –
MEDICAL ELECTRICAL EQUIPMENT –

Part 1-6: General requirements for safety –
Collateral Standard: Usability


SECTION ONE – GENERAL
1 Scope and object
1.201 Scope
This Collateral Standard specifies requirements for a PROCESS to analyse, design, verify and
validate the USABILITY, as it relates to SAFETY of MEDICAL ELECTRICAL EQUIPMENT, hereinafter
referred to as EQUIPMENT. This standard addresses NORMAL USE and USE ERRORS but excludes
ABNORMAL USE.
1.202 Relationship to other standards
1.202.1 IEC 60601-1
This Collateral Standard complements IEC 60601-1.
When referring to IEC 60601-1 or to this Collateral Standard, either individually or in
combination, the following conventions are used:
– “the General Standard” designates IEC 60601-1 alone;
– “this Collateral Standard” designates IEC 60601-1-6 alone;
– “this Standard” designates the combination of the General Standard and this Collateral
Standard.
1.202.2 Particular Standards
A requirement in a Particular Standard takes priority over the corresponding requirement in
this Collateral Standard.
1.202.3 Normative references
The following referenced documents are indispensable for the application of this document.
For dated references, only the edition cited applies. For undated references, the latest edition
of the referenced document (including any amendments) applies.
IEC 60601-1:1988, Medical electrical equipment – Part 1: General requirements for safety
Amendment 1 (1991)
Amendment 2 (1995)
IEC 60601-1-8:2003, Medical electrical equipment – Part 1-8: General requirements for safety
– Collateral standard: General requirements, tests and guidance for alarm systems in medical
electrical equipment and medical electrical systems
ISO 14971:2000, Medical devices – Application of risk management to medical devices
Amendment 1 (2003)

---------------------- Page: 11 ----------------------

60601-1-6  IEC:2004 – 15 –
2 Terminology and definitions
For the purpose of this collateral standard, the terms and definitions given in Clause 2 of
IEC 60601-1:1988, as amended by the other collateral standards, Clause 3 of
ISO 14971:2000 and the following apply.

NOTE An Index of all defined terms used in this collateral standard is found at the end of the document.
2.201
ABNORMAL USE
intended act or intended omission of an act by the USER or OPERATOR of EQUIPMENT as a result
of conduct that is beyond any reasonable means of RISK CONTROL by the manufacturer
NOTE 1 See also Annex BBB. Examples are given in Annex CCC.
NOTE 2 It is possible for the PATIENT to be the OPERATOR, e.g. when EQUIPMENT is used in the PATIENT’S home.
NOTE 3 ABNORMAL USE is not considered REASONABLY FORESEABLE MISUSE.
2.202
EFFECTIVENESS
accuracy and completeness with which OPERATORS achieve specified goals
[ISO 9241-11:1998, definition 3.2, modified]
2.203
EFFICIENCY
resources expended in relation to the accuracy and completeness with which OPERATORS
achieve goals
[ISO 9241-11:1998, definition 3.3 modified]
2.204
* OPERATOR-EQUIPMENT INTERFACE
means by which the OPERATOR and the EQUIPMENT communicate
[ANSI/AAMI/HE 74:2001, definition 3.24 modified]
NOTE The ACCOMPANYING DOCUMENTS are considered part of the EQUIPMENT and the OPERATOR-EQUIPMENT
INTERFACE.
2.205
OPERATOR PROFILE
summary of the mental, physical and demographic traits of the intended OPERATOR population,
as well as any special characteristics that can have a bearing on design decisions, such as
occupational skills and job requirements
2.206
* PRIMARY OPERATING FUNCTION
function that involves OPERATOR interaction that is either frequently used or related to the
SAFETY of the EQUIPMENT in NORMAL USE
2.207
* REASONABLY FORESEEABLE MISUSE
use by the OPERATOR in a way not intended by the manufacturer but which can result from
readily predictable human behaviour
[ISO/IEC Guide 51:1999, definition 3.14, modified]
NOTE 1  REASONABLY FORESEEABLE MISUSE is an intended action.

---------------------- Page: 12 ----------------------

60601-1-6  IEC:2004 – 17 –
NOTE 2  Use refers to a product, PROCESS or service.
NOTE 3  Slips, lapses, mistakes and ABNORMAL USE can also be reasonably foreseeable, but are not considered
REASONABLY FORESEEABLE MISUSE.
NOTE 4  See also Annex BBB.
2.208
TRAINING
application-specific OPERATOR-oriented instruction or exercises required for the safe and
effective use of the EQUIPMENT
2.209
USE ERROR
act or omission of an act that has a different EQUIPMENT response than intended by the
manufacturer or expected by the OPERATOR
NOTE 1 USE ERROR includes slips, lapses, mistakes, and REASONABLY FORESEEABLE MISUSE.
NOTE 2 See also Annex BBB and DDD.1.3.
NOTE 3 The physiological response of the PATIENT is not considered part of USE ERROR.
2.210
USE SCENARIO
sequence of events and tasks used to specify and test the USABILITY of the EQUIPMENT
2.211
* USABILITY
Characteristic that establishes EFFECTIVENESS, EFFICIENCY and OPERATOR learnability and
satisfaction
2.212
USABILITY ENGINEERING
application of knowledge about human behaviour, abilities, limitations, and other
characteristics to the design of tools, machines, EQUIPMENT, devices, systems, tasks, jobs,
and environments to achieve adequate USABILITY
2.213
USABILITY ENGINEERING FILE
*
set of RECORDS and other documents that are produced by USABILITY ENGINEERING activities
2.214
USABILITY SPECIFICATION
documentation defining the OPERATOR-EQUIPMENT INTERFACE requirements related to USABILITY
2.215
VALIDATION
confirmation, through the provision of objective evidence, that the requirements for a specific
intended use or application have been fulfilled
NOTE 1 The term “validated” is used to designate the corresponding status.
NOTE 2 The use conditions for VALIDATION can be real or simulated.
[ISO 9000:2000, definition 3.8.5]

---------------------- Page: 13 ----------------------

60601-1-6  IEC:2004 – 19 –
3 General requirements
3.1 Addition:
* The EQUIPMENT shall provide adequate USABILITY such that the RISKS resulting from NORMAL
USE and USE ERRORS are acceptable.
6 Identification, markings and documents
6.8 ACCOMPANYING DOCUMENTS
6.8.1 * General
Addition:
A brief description of the EQUIPMENT, its physical operating principles and significant physical
and performance characteristics relevant to its USABILITY, shall be included in the instructions
for use. The same information shall also be included in the technical description, if this is
provided as a separate document.
ACCOMPANYING DOCUMENTS for EQUIPMENT may be provided electronically, e.g. by electronic
file format or CD-ROM. If the ACCOMPANYING DOCUMENTS are provided electronically, the
USABILITY ENGINEERING PROCESS shall include consideration of which information also needs to
be provided as hard copy or as markings on the EQUIPMENT, e.g. to cover emergency
operation.
The ACCOMPANYING DOCUMENTS shall include a description of the OPERATOR PROFILE. The
ACCOMPANYING DOCUMENTS shall be written at a level consistent with the intended OPERATOR
PROFILE.
Compliance is checked by inspection of the ACCOMPANYING DOCUMENTS and the USABILITY
ENGINEERING FILE.
6.8.2 Instructions for use
Addition:
6.8.201 * TRAINING and materials for TRAINING
If EQUIPMENT specific TRAINING is required for the PRIMARY OPERATING FUNCTIONS of the
EQUIPMENT, the manufacturer shall:
– provide the necessary materials for TRAINING;
– ensure that these materials are available; or
– provide the TRAINING.
NOTE 1  EQUIPMENT-specific TRAINING provides the knowledge and skills required for safe and effective use of
EQUIPMENT in addition to the OPERATOR PROFILE.
The INTENDED USE/INTENDED PURPOSE shall be the basis for TRAINING and TRAINING material.
The instructions for use shall indicate whether specific TRAINING for this EQUIPMENT is required
and shall indicate the available TRAINING options.
1)
NOTE 2  See IEC 61258 [1] .
———————
1
) Figures in brackets refer to the Bibliography.

---------------------- Page: 14 ----------------------

60601-1-6  IEC:2004 – 21 –
SECTIONS TWO TO SIX – NOT USED
SECTION SEVEN – PROTECTION AGAINST EXCESSIVE TEMPERATURES
AND OTHER SAFETY HAZARDS
Replacement (title):
46 * USE ERROR and USABILITY
46.201 * SAFETY for the PATIENT, OPERATOR and other persons
A USABILITY ENGINEERING PROCESS shall be conducted to provide SAFETY for the PATIENT,
OPERATOR and other persons related to USABILITY of the OPERATOR-EQUIPMENT INTERFACE.
NOTE 1  To guide the application of USABILITY ENGINEERING principles, the HAZARDS to PATIENTS, OPERATORS and
other persons as listed in ISO 14971 should be considered.
NOTE 2  The following are examples of HAZARDS for the PATIENT:
– unintentional setting of the diagnostic or therapeutic EQUIPMENT, e.g. inappropriate X-ray exposure setting
requiring an additional exposure;
– unintentional interruption of delivery of therapy;
– misinterpretation of displayed values followed by an inappropriate treatment;
– confusing data presentation contributing to mental fatigue resulting in increased USE ERROR.
NOTE 3  The following are examples of HAZARDS for the OPERATOR:
– poor anthropometric design leading to musculoskeletal injury;
– repetitive-motion resulting in nerve/tendon injuries;
– poor display contrast resulting in eye fatigue;
– loud noise emanating from the EQUIPMENT resulting in hearing impairment.
Compliance with this subclause is considered to exist when compliance with the other clauses
and subclauses is demonstrated.
46.202 USABILITY ENGINEERING PROCESS
46.202.1 General
The results of the USABILITY ENGINEERING PROCESS shall be recorded in the USABILITY
ENGINEERING FILE. The USABILITY ENGINEERING PROCESS may vary in form and extent based on
the nature of the EQUIPMENT, its intended OPERATOR and its INTENDED USE/INTENDED PURPOSE
(see DDD.3.2). The records and other documents that make up the USABILITY ENGINEERING
FILE may form part of other documents and files, e.g. a manufacturer’s product file or RISK
MANAGEMENT FILE.
In the case of the modification of existing EQUIPMENT design, the USABILITY ENGINEERING
PROCESS may be scaled based on the significance of the modification depending on the
results of the RISK ANALYSIS (see DDD.3.2.2).
NOTE 1  The manufacturer
...

Questions, Comments and Discussion

Ask us and Technical Secretary will try to provide an answer. You can facilitate discussion about the standard in here.