Controlled environment storage cabinet for processed thermolabile endoscopes

The proposed standard specifies the performance requirements applying to cabinets designed to store and dry heat-sensitive endoscopes (SCHE) following automated or manual reprocessing. The storage cabinets are designed to provide a controlled environment for storage of endoscope(s) and when necessary drying of the endoscope(s), including the endoscope(s) channels. The controlled environment is provided to ensure that during storage there is no deterioration of the microbial quality of the endoscope. The drying stage is intended to supplement, if necessary, any drying provided as part of the automated or manual reprocessing cycle. The cabinet is not intended to provide any cleaning or disinfection function. Note 1 The use of a storage cabinet may allow the safe use of the endoscope for an extended period from the time of reprocessing and improve availability for emergency use. Note 2 Thorough drying of an endoscope in a washer-disinfector may require a prolonged cycle time; the use of a drying-storage cabinet may enhance throughput of the endoscopes.

Lagerungsschrank mit geregelten Umgebungsbedingungen für aufbereitete, thermolabile Endoskope

Diese Europäische Norm legt die Leistungsanforderungen fest, die für Schränke zum Lagern und Trocknen wärmeempfindlicher Endoskope (SCHE) nach der automatischen oder manuellen Wiederaufbereitung gelten. Die Lagerschränke sind dafür ausgelegt, geregelte Umgebungsbedingungen für die Endoskoplagerung und, erforderlichenfalls, die Trocknung von Endoskopen einschließlich der Endoskopkanäle herzustellen. Die geregelte Umgebung dient der Sicherstellung, dass während der Lagerung keine Verschlechterung der mikrobiellen Beschaffenheit des Endoskops auftritt. Die Trocknungsprozessphase des SCHE ist dafür vorgesehen, erforderlichenfalls jedes weitere Trocknen als Teil des automatischen oder manuellen Wieder-aufbereitungszyklus zu übernehmen. Der Schrank ist nicht dafür vorgesehen, eine Reinigungs- und/oder Desinfektionsfunktion bereitzustellen. ANMERKUNG 1 Die Verwendung eines Lagerschranks kann den sicheren Gebrauch des Endoskops über einen längeren Zeitraum nach der Wiederaufbereitung erlauben und die Verfügbarkeit für Notfälle verbessern. ANMERKUNG 2 Das gründliche Trocknen eines Endoskops in einem Reinigungs-Desinfektionsgerät kann eine verlängerte Zyklusdauer erfordern; die Verwendung eines Lagerschranks kann den Endoskop-Durchsatz erhöhen.

Enceinte de stockage à atmosphère contrôlée pour endoscopes thermosensibles traités

La présente Norme européenne définit les exigences de performance qui s’appliquent aux enceintes destinées au stockage, ou au stockage et au séchage d’endoscopes thermosensibles, après le traitement manuel ou automatique de ces derniers.
Les enceintes de stockage sont destinées à fournir une atmosphère contrôlée pour le stockage d'un ou de plusieurs endoscopes (avec ou sans canaux) et, quand cela est nécessaire, pour leur séchage (en incluant leurs canaux).
L’atmosphère contrôlée fournie par l’enceinte de stockage permet de prévenir toute dégradation de la qualité microbiologique de l’endoscope pendant la période de stockage. La fonction de séchage est prévue pour compléter, le cas échéant, tout processus de séchage faisant partie du cycle de traitement manuel ou automatique des endoscopes.
La présente Norme européenne définit les enceintes de stockage qui soufflent les canaux et les surfaces extérieures des endoscopes avec de l'air.
NOTE 1   L'enceinte de stockage est l’un des moyens permettant d’utiliser l’endoscope en toute sécurité pendant une période prolongée à partir de son traitement et d’améliorer sa disponibilité en cas d’urgence.
NOTE 2   Le séchage complet d’un endoscope dans un laveur désinfecteur peut nécessiter un temps de cycle prolongé ; l’utilisation d’une enceinte de stockage dotée d’une fonction de séchage peut améliorer le taux d’utilisation des endoscopes.
L’enceinte n’est pas prévue pour assurer une fonction de lavage ou de désinfection.
La présente Norme européenne n'inclut pas l'utilisation d'autres produits chimiques pour le séchage et le maintien de la qualité des endoscopes pendant le stockage.

Shranjevalni prostor z nadzorovanim okoljem za pripravljene termolabilne endoskope

Predlagani standard določa zahteve glede zmogljivosti prostorov za shranjevanje in sušenje občutljivih endoskopov (SCHE) po samodejni ali ročni pripravi za ponovno uporabo. Shranjevalni prostori zagotavljajo nadzorovano okolje za shranjevanje endoskopov in, kadar je to potrebno, njihovo sušenje, vključno z endoskopskimi kanali. Nadzorovano okolje zagotavlja, da se med shranjevanjem ne poslabša mikrobna kakovost endoskopa. Faza sušenja je namenjena kot dopolnilo, če je to potrebno, vsakemu sušenju, ki je del samodejnega ali ročnega cikla priprave na ponovno uporabo. Prostor ni namenjen za nobeno funkcijo čiščenja oziroma dezinfekcije. OPOMBA 1: Uporaba shranjevalnega prostora lahko dovoljuje varno uporabo endoskopa dlje časa po pripravi za ponovno uporabo in lahko izboljša razpoložljivost za uporabo v sili. OPOMBA 2: Temeljito sušenje endoskopa v pralno-dezinfekcijskem pripomočku lahko zahteva daljši čas cikla; uporaba shranjevalnega prostora pa lahko poveča pretok endoskopov.

General Information

Status
Published
Public Enquiry End Date
14-Oct-2012
Publication Date
19-Apr-2015
Technical Committee
Current Stage
6060 - National Implementation/Publication (Adopted Project)
Start Date
08-Jan-2015
Due Date
15-Mar-2015
Completion Date
20-Apr-2015

Buy Standard

Standard
EN 16442:2015
English language
59 pages
sale 10% off
Preview
sale 10% off
Preview
e-Library read for
1 day

Standards Content (Sample)

2003-01.Slovenski inštitut za standardizacijo. Razmnoževanje celote ali delov tega standarda ni dovoljeno.Shranjevalni prostor z nadzorovanim okoljem za pripravljene termolabilne endoskopeLagerungsschrank mit geregelten Umgebungsbedingungen für aufbereitete, thermolabile EndoskopeEnceinte de stockage à atmosphère contrôlée pour endoscopes thermosensibles traitésControlled environment storage cabinet for processed thermolabile endoscopes11.140Oprema bolnišnicHospital equipmentICS:Ta slovenski standard je istoveten z:EN 16442:2015SIST EN 16442:2015en,fr,de01-maj-2015SIST EN 16442:2015SLOVENSKI
STANDARD



SIST EN 16442:2015



EUROPEAN STANDARD NORME EUROPÉENNE EUROPÄISCHE NORM
EN 16442
March 2015 ICS 11.140 English Version
Controlled environment storage cabinet for processed thermolabile endoscopes
Enceinte de stockage à atmosphère contrôlée pour endoscopes thermosensibles traités
Lagerungsschrank mit geregelten Umgebungsbedingungen für aufbereitete, thermolabile Endoskope This European Standard was approved by CEN on 19 December 2014.
CEN members are bound to comply with the CEN/CENELEC Internal Regulations which stipulate the conditions for giving this European Standard the status of a national standard without any alteration. Up-to-date lists and bibliographical references concerning such national standards may be obtained on application to the CEN-CENELEC Management Centre or to any CEN member.
This European Standard exists in three official versions (English, French, German). A version in any other language made by translation under the responsibility of a CEN member into its own language and notified to the CEN-CENELEC Management Centre has the same status as the official versions.
CEN members are the national standards bodies of Austria, Belgium, Bulgaria, Croatia, Cyprus, Czech Republic, Denmark, Estonia, Finland, Former Yugoslav Republic of Macedonia, France, Germany, Greece, Hungary, Iceland, Ireland, Italy, Latvia, Lithuania, Luxembourg, Malta, Netherlands, Norway, Poland, Portugal, Romania, Slovakia, Slovenia, Spain, Sweden, Switzerland, Turkey and United Kingdom.
EUROPEAN COMMITTEE FOR STANDARDIZATION
COMITÉ EUROPÉEN DE NORMALISATION EUROPÄISCHES KOMITEE FÜR NORMUNG
CEN-CENELEC Management Centre:
Avenue Marnix 17,
B-1000 Brussels © 2015 CEN All rights of exploitation in any form and by any means reserved worldwide for CEN national Members. Ref. No. EN 16442:2015 ESIST EN 16442:2015



EN 16442:2015 (E) 2 Contents
Page
Foreword . 4 Introduction . 5 1 Scope . 6 2 Normative references . 6 3 Terms and definitions . 6 4 Performance requirements . 7 4.1 General . 7 4.2 Storage . 8 4.3 Drying . 8 4.4 Endoscope storage cabinet connectors (ESC connectors) . 9 5 Mechanical and procedure requirements . 9 5.1 Materials – design, manufacture and assembly . 9 5.2 Air quality . 10 5.3 Contamination of the storage cabinet chamber surfaces . 11 5.4 Drying process control . 11 5.5 Endoscope channel aeration system . 12 5.6 Automatic temperature control . 13 5.7 Fault indication/monitoring . 13 5.8 Cycle indicators . 14 5.9 Instruments and control devices . 14 5.10 Temperature indicators . 15 5.11 Relative humidity indicator . 15 5.12 Pressure indicators . 15 5.13 Traceability . 16 5.14 Double-ended storage cabinets . 16 6 Testing for conformity . 17 6.1 General . 17 6.2 Air changes . 17 6.3 Overpressure . 17 6.4 Drying . 18 6.5 Contamination of the inside surfaces of the storage cabinet . 19 6.6 Air quality . 19 6.7 Channel aeration test . 21 6.8 Thermometric testing 1 – chamber and load temperature testing. 21 6.9 Thermometric test 2- chamber and load temperature testing . 22 6.10 Readability . 22 6.11 Tests for air filtration . 22 7 Documentation . 22 8 Information to be supplied with the storage cabinet . 22 8.1 General . 22 8.2 Information to be supplied before delivery . 23 8.3 Marking and labelling . 25 8.4 Packaging . 25 9 Information to be requested from the purchaser by the manufacturer . 25 Annex A (informative) Summary of test programmes . 26 Annex B (informative) Cross-contamination between endoscopes . 28 SIST EN 16442:2015



EN 16442:2015 (E) 3 Annex C (normative) Methods for evaluation of airborne microbial contamination
in the storage cabinet . 31 Annex D (informative) Procedure for parametric performance qualification . 32 Annex E (normative) Internal residual contamination of endoscopes after storage . 38 Annex F (normative) Establishing endoscope type test groups . 46 Annex G (normative) Establishing endoscope product families . 55 Bibliography . 59
SIST EN 16442:2015



EN 16442:2015 (E) 4 Foreword This document (EN 16442:2015) has been prepared by Technical Committee CEN/TC 102 “Sterilizers for medical purposes”, the secretariat of which is held by DIN. This European Standard shall be given the status of a national standard, either by publication of an identical text or by endorsement, at the latest by September 2015 and conflicting national standards shall be withdrawn at the latest by September 2015. Attention is drawn to the possibility that some of the elements of this document may be the subject of patent rights. CEN [and/or CENELEC] shall not be held responsible for identifying any or all such patent rights. According to the CEN-CENELEC Internal Regulations, the national standards organizations of the following countries are bound to implement this European Standard: Austria, Belgium, Bulgaria, Croatia, Cyprus, Czech Republic, Denmark, Estonia, Finland, Former Yugoslav Republic of Macedonia, France, Germany, Greece, Hungary, Iceland, Ireland, Italy, Latvia, Lithuania, Luxembourg, Malta, Netherlands, Norway, Poland, Portugal, Romania, Slovakia, Slovenia, Spain, Sweden, Switzerland, Turkey and the United Kingdom. SIST EN 16442:2015



EN 16442:2015 (E) 5 Introduction Endoscope storage cabinets are designed to provide a controlled environment for the storage of endoscope(s) (with or without channels) and, if necessary, drying of the endoscope(s) including the endoscope(s) channels. The controlled environment provided by the storage cabinet ensures that during storage there is no deterioration of the microbiological quality of the endoscope. The drying function is intended to supplement, if necessary, any drying conducted during automated or manual processing of the endoscope. The storage cabinet is designed to allow for the safe use of endoscopes at an extended period from the time of processing improving availability for use. NOTE 1 Drying of an endoscope in a washer-disinfector can require a prolonged cycle time. The use of a storage cabinet including a drying function can increase the number of endoscopes that can be processed in the washer-disinfector for a defined time period. NOTE 2 It is strongly recommended to verify the microbiological quality of the endoscopes intended to be stored in the cabinet before installation of the storage cabinet. NOTE 3 The storage cabinet is not designed to clean and/or disinfect endoscopes and any contaminated endoscope stored in the cabinet can still be contaminated after the storage period. NOTE 4 Storage cabinets for processed thermolabile endoscopes are not considered as medical devices. SIST EN 16442:2015



EN 16442:2015 (E) 6 1 Scope This European Standard specifies the performance requirements applying to cabinets designed to store, or store and dry, thermolabile endoscopes following automated or manual processing. The storage cabinets are designed to provide a controlled environment for storage of endoscope(s), with or without channels, and when necessary drying of the endoscope(s), including the endoscope(s) channels. The controlled environment provided by the storage cabinet ensures that during storage there is no deterioration of the microbiological quality of the endoscope. The drying function is intended to supplement, if necessary, any drying provided as part of the automated or manual processing cycle. This European Standard specifies storage cabinets which flush the channels and the external surfaces of endoscopes with air. NOTE 1 The storage cabinet is one of the means that can allow the safe use of the endoscope for an extended period from the time of processing and improve availability for emergency use. NOTE 2 Thorough drying of an endoscope in a washer-disinfector can require a prolonged cycle time; the use of a storage cabinet including a drying function can enhance throughput of the endoscopes. The cabinet is not intended to provide any cleaning or disinfection function. This European Standard does not include the use of other chemicals for drying and maintaining the quality of endoscopes during storage 2 Normative references The following documents, in whole or in part, are normatively referenced in this document and are indispensable for its application. For dated references, only the edition cited applies. For undated references, the latest edition of the referenced document (including any amendments) applies. EN 60584-1:2013, Thermocouples — Part 1: EMF specifications and tolerances (IEC 60584-1:2013) EN 60751:2008, Industrial platinum resistance thermometers and platinum temperature sensors (IEC 60751:2008) EN ISO 14644-3:2005, Cleanrooms and associated controlled environments — Part 3: Test methods (ISO 14644-3:2005) 3 Terms and definitions For the purposes of this document the following terms and definitions apply. 3.1 drying function additional feature of a storage cabinet carried out in the sequence as regulated by the automatic controller to remove moisture 3.2 drying phase part of the storage cycle that is dedicated to the drying of the endoscope 3.3 drying temperature band range of temperatures expressed as the minimum and the maximum controlled temperatures, which may prevail throughout the load during drying SIST EN 16442:2015



EN 16442:2015 (E) 7 3.4 endoscope storage cabinet connector
ESC connector device used to connect endoscope channels inside the cabinet to the flushing system 3.5 endoscope surrogate device item designed to represent construction elements of endoscope specific characteristics affecting the flow conditions in an endoscope Note 1 to entry: Construction elements can include channel length and diameter, connectors, channel separators, port closures, return valves, etc. 3.6 processing activity including cleaning, disinfection and sterilization (if necessary and applicable), to prepare a new or used medical device for its intended use 3.7 storage cabinet equipment controlled by an automatic control system that maintains the microbiological quality of processed thermolabile endoscope 3.8 storage cycle time between connecting and disconnecting the endoscope(s) inside the storage cabinet Note 1 to entry:
A storage cycle can include a drying phase. 3.9 storage temperature band range of temperatures expressed as the minimum and the maximum controlled temperatures, which may prevail throughout the load during storage 3.10 thermolabile damaged by exposure to temperatures within the range used for thermal disinfection Note 1 to entry: The minimum temperature for thermal disinfection specified in ISO 15883-1 is 65°C. 4 Performance requirements 4.1 General 4.1.1 Storage cabinets are designed to provide a controlled environment for storage of endoscopes (with or without channels). The controlled environment provided by the cabinet shall ensure that during storage there is no deterioration of the microbiological quality of the endoscope. An optional drying function is intended to supplement, if necessary, any drying provided as part of the automated or manual processing cycle The cabinet is not intended to provide any cleaning or disinfection function. NOTE 1 Thorough drying of an endoscope in a washer-disinfector can require a prolonged cycle time; the use of a storage cabinet including a drying function can enhance throughput of the endoscopes. NOTE 2 Table A.1 gives a summary of the tests and Clause 6 on the test methods that can be used to check that the storage cabinets meet the specified requirements. 4.1.2 Detailed requirements for information to be provided by the manufacturer are specified in Clause 8. SIST EN 16442:2015



EN 16442:2015 (E) 8 4.1.3 The value of any process variable shall be pre-set and adjustment shall require the use of a key, code or tool. 4.1.4 Throughout the drying phase and/or storage the values and rate of change in temperature, pressure or any other process variable shall be within limits which will not cause damage to the device(s) stored in the storage cabinet. 4.2 Storage 4.2.1 The storage cabinet shall maintain the microbiological quality of the endoscopes during storage. Tests shall be performed according to Annex E. 4.2.2 A risk analysis with consideration of the different parameters on the storage cabinet performance shall be performed and the means used to minimize the identified risks shall be verified (see 6.1). NOTE 1 According to the design of the storage cabinet those parameters can include: − potential for contamination between different endoscopes stored simultaneously (see Annex B); − ingress of contamination during loading and/or unloading; − potential for contamination caused by accessories and connectors/connections (see Annex B); − potential for contamination caused by endoscopes accessories; − environmental conditions (e.g. temperature, humidity, etc.) where the storage cabinet is installed; − potential for contamination caused by improper air quality in the storage compartment; − potential for contamination caused by inefficient drying procedure prior to storage; − potential for growth of the initial contamination of a contaminated endoscope accidentally introduced in the storage cabinet. NOTE 2 EN ISO 14971 establishes the requirements for risk management to determine the safety of a medical device by the manufacturer during the product life cycle and may be used. 4.2.3 If the storage cabinet does not have a drying phase, it has to be specified that the endoscope (outside surfaces and internal channels) has to be dried before storage [see 8.2 j)]. 4.2.4 Any instructions given by the manufacturer for drying the inside and outside of the endoscope shall conform to the endoscope manufacturer's instructions (on pressure issues, temperature issues, etc.). 4.2.5
Any requirements regarding the quality of the air supplied to the storage cabinet (see 5.2) shall be specified [see 8.2 f) 1)]. 4.2.6 The maximum storage time in the storage cabinet shall be determined [see 8.2 c)] NOTE The maximum storage time can be limited by national or regional recommendations or regulations. 4.3 Drying For storage cabinets that provide a drying function [see 8.2 d) 1)] the following requirements apply: a) The time required to dry the endoscopes [see 8.2 d) 4)] shall be specified when tested according to 6.4 and shall not exceed 3 hours. NOTE 1 Drying times can be different according to the type of endoscope involved. NOTE 2 3 hours is an acceptable drying time to reduce the potential risk of growth of microorganisms present in endoscope channels. SIST EN 16442:2015



EN 16442:2015 (E) 9 b) The efficacy of the drying function shall be deemed to be satisfactory if, when tested according to 6.4, there are no visible moisture droplets on the test paper. 4.4 Endoscope storage cabinet connectors (ESC connectors) 4.4.1 General Two types of ESC connectors exist: a) ESC connectors providing independent air flow to each endoscope channel; and b) ESC connectors providing air flow to a group of endoscope channels via a manifold (between the storage cabinet and the endoscope). NOTE In the absence of a system that allows control of air flow in each tube of the ESC connector, flow of air in endoscope channels depends on: — air connection tubing to the storage cabinet, — the design of the manifold that insures air separation to supply each channels of the endoscope, and — the internal design of the endoscope. 4.4.2 ESC connector qualification Each ESC connector type shall be qualified during type testing. A test protocol shall be provided to enable the user to verify the compliance of the ESC connectors with the specifications during routine testing [see 8.2 i)]. 5 Mechanical and procedure requirements 5.1 Materials – design, manufacture and assembly 5.1.1 Load carriers, trays or holding systems intended to accommodate the device(s) to be stored shall be designed and constructed to avoid the possibility of damage to the device(s) at the time of loading, during storage and during the course of unloading. 5.1.2 The procedures required to minimize microbial contamination on the internal surfaces of the storage cabinet shall be described. These procedures shall not adversely affect the quality of the load under normal conditions of use [see 8.2 r)]. 5.1.3 For storage cabinets in which the endoscopes are stored in a vertical hanging position: a) the endoscope hanger shall be designed to ensure that all endoscopes specified for storage and/or drying in the storage cabinet [see 8.2 a) and d) 2)] will not touch the bottom of the storage cabinet, or b) a means shall be provided to position the distal end of the endoscope to prevent contact with the bottom of the storage cabinet. SIST EN 16442:2015



EN 16442:2015 (E) 10 5.2 Air quality 5.2.1 Air supplied to the storage cabinet 5.2.1.1 General 5.2.1.1.1 Air supplied to the storage cabinet shall be of a quality which shall not impair the cleanliness of, nor introduce microbial contamination to, the load. 5.2.1.1.2 The quality of the air supplying the storage cabinet shall be specified [see 8.2 f)] and may include specifications for maximum relative humidity, pressure, oil content, particulate count and flow rate. 5.2.1.1.3 The material used in the pipework of the air distribution system shall be compatible with the intended use of the storage cabinet. 5.2.1.1.4 The specifications of the air quality shall be measured on installation and at specified intervals [see 8.2 f) 3)]. 5.2.1.2 Compressed air 5.2.1.2.1 Where the storage cabinet is supplied with compressed air, the compressor shall be fitted with a filter and dryer system to meet the required specifications (see 5.2.1.1.2) and minimise the risk of contamination of the storage cabinet and stored endoscopes by microorganisms. 5.2.1.2.2 Where the storage cabinet is provided with a compressor, the frequency for changing pre- and post-compression filters on the air compressor shall be specified [see 8.2 h) 2)]. 5.2.1.2.3 When compressed air is used the oil content shall not exceed 0,1 mg/ m3. NOTE 1 This quality is equivalent to Class 2 as defined in ISO 8573-1. NOTE 2 Compressed air coming from an oil free compressor is deemed to meet this requirement. 5.2.2 Environmental conditions inside the storage cabinet 5.2.2.1 General Air inside the storage cabinet and flowing through the channels of the endoscope shall be of a microbiological quality which will not impair the quality of the load. Tests shall be done according to Annex C. NOTE 1 This can be achieved by filtration of the air using filters having not less than 99,95 % retention to particles of 0,3 µm. NOTE 2 Filters conforming to Class H 13 as specified in EN 1822-1:2009 can be regarded as suitable. Recommendations on the relevant alert and action limits to be set for the results of particulate (if claimed) and microbiological monitoring [see 8.2 g) 3)] shall be specified, including the action to be taken when specified limits are exceeded [see 8.2 g) 5)]. 5.2.2.2 Overpressure When tested according to 6.3, the air pressure in the storage cabinet chamber shall be higher than the ambient pressure where the storage cabinet is located. Measurements shall be made when the doors of the storage cabinet are closed and after the defined stabilization time [see 8.2 v)]. SIST EN 16442:2015



EN 16442:2015 (E) 11 5.2.2.3 Air changes The number of air changes per hour inside the storage cabinet chamber using the method described in 6.2 shall be specified [see 8.2 w)]. NOTE An air change of at least ten times the volume of the storage compartment per hour is an acceptable value to reduce the risk of contamination from the environment following, for example, a door opening and to reduce the moisture content during drying. 5.2.2.4 Particulate contamination 5.2.2.4.1 If a specific cleanliness level is claimed and when tested according to 6.6.1 the particulate contamination within the storage cabinet shall be consistent with the claims [see 8.2 g) 1) and 2)]. NOTE Classifications of air cleanliness are defined in EN ISO 14644-1. 5.2.2.4.2 Where the air in the storage cabinet is filtered, means shall be provided to enable the filtration system to be tested (see 6.11). This shall include means of access upstream of the filter where a controlled particulate aerosol can be injected and means of access downstream of the filter where an iso-kinetic sampling probe can be placed. 5.2.2.5 Temperature control 5.2.2.5.1 If the storage cabinet operates at temperatures different from ambient, the temperature inside the storage cabinet shall be specified and controlled within the temperature limits [see 8.2 b) and 8.2 d) 3)]. The temperature limits of the endoscopes have to be considered. 5.2.2.5.2 When tested according to 6.8 and 6.9 the rate and extent of any change in temperature throughout the operating cycle shall be within specified limits, and will not cause damage to the endoscope(s) stored in the storage cabinet [see 8.2 b) and 8.2 d) 3)]. 5.2.2.6 Pressure Throughout the operating cycle the rate and extent of any change in pressure in endoscope channels shall be within specified limits, which will not cause damage to the endoscope(s) stored in the storage cabinet [see 8.2 e) 4)]. 5.3 Contamination of the storage cabinet chamber surfaces 5.3.1 The internal surfaces of the storage cabinet chamber shall be capable of withstanding routine cleaning and disinfection [see 8.2 r)], without deterioration. NOTE Removable trays of a cabinet, designed to store endoscopes, are considered as chambers. 5.3.2 A cleaning-disinfection procedure shall be provided, including any requirements regarding the frequency of the use, to ensure that surface contamination levels inside the storage cabinet that might contact endoscopes remain below 25 Colony Forming Units (CFU)/25 cm2 when tested as described in 6.5 [see 8.2 r)]. NOTE This can include the use of liquid/gas, cleaning-disinfection procedures or any other validated methods giving the same result. 5.4 Drying process control Drying shall be achieved by evaporation of residual moisture from the endoscope. The rate of drying shall be enhanced by air flow. NOTE The air can be heated and/or dry. SIST EN 16442:2015



EN 16442:2015 (E) 12 5.5 Endoscope channel aeration system 5.5.1 Throughout the storage, air has to flow through each of the internal channels and/or cavities of the device. The air circulation may be either continuous or intermittent. Instructions shall be provided on the verification of air circulation and can include: a) verifying that all channels allow the passage of air before the endoscope is loaded into the storage cabinet [see 8.2 j)]; NOTE 1 If the endoscope is cleaned and disinfected using a validated processing procedure this verification is included (e.g. EN ISO 15883-4:2009, 5.2.2). NOTE 2 Some washer-disinfectors and manual processing procedures do not monitor flow through the endoscope channels. b) confirming that all necessary connections were made before, and were still in place at the end of, the storage cycle [see 8.2 i)]; c) verifying the air circulation in each tubing of the ESC connector using specified means [see 8.2 e) 1)]; For single channel ESC connectors, means shall be provided to allow the verification of air flow in each tubing. When a manifold is used means provided shall allow the verification at least in the tube connected to the storage cabinet. NOTE 3 The attention of the user is drawn on the fact that means provided to verify the free passage of air can be either continuous or intermittent, automatic and under the control of the automatic controller of the storage cabinet (i.e. control of the air flow through each endoscope channel) or require a verification by the user (e.g. visual indication of the air circulation). d) Confirming by
...

Questions, Comments and Discussion

Ask us and Technical Secretary will try to provide an answer. You can facilitate discussion about the standard in here.