Implants for surgery - Active implantable medical devices - Part 6: Particular requirements for active implantable medical devices intended to treat tachyarrhythmia (including implantable defibrillators) (ISO 14708-6:2019)

This document specifies requirements that are applicable to implantable cardioverter defibrillators and
CRT-Ds and the functions of active implantable medical devices intended to treat tachyarrhythmia.
The tests that are specified in ISO 14708 are type tests and are to be carried out on samples of a device
to show compliance.
This document was designed for tachyarrhythmia pulse generators used with either endocardial leads
or epicardial leads. At the time of this edition, the authors recognized the emergence of technologies
that do not use endocardial or epicardial leads for which adaptations of this part will be required. Such
adaptations are left to the discretion of manufacturers incorporating these technologies.
This document is also applicable to some non-implantable parts and accessories of the devices (see
Note 1).
The characteristics of the implantable pulse generator or lead shall be determined by either the
appropriate method detailed in this document or by any other method demonstrated to have accuracy
equal to, or better than, the method specified. In the case of dispute, the method detailed in this
document shall apply.
Any aspect of an active implantable medical device intended to treat bradyarrhythmias or cardiac
resynchronization is covered by ISO 14708-2.
NOTE 1 The device that is commonly referred to as an active implantable medical device can in fact be
a single device, a combination of devices, or a combination of a device or devices and one or more accessories.
Not all of these parts are required to be either partially or totally implantable, but there is a need to specify
some requirements of non-implantable parts and accessories if they could affect the safety or performance of the
implantable device.
NOTE 2 In this document, terms printed in italics are used as defined in Clause 3. Where a defined term is
used as a qualifier in another term, it is not printed in italics unless the concept thus qualified is also defined.

Chirurgische Implantate - Aktive implantierbare medizinische Geräte - Teil 6: Besondere Festlegungen für aktive implantierbare medizinische Produkte zur Behandlung von Tachyarrhythmie (einschließlich implantierbaren Defibrillatoren) (ISO 14708-6:2019)

Das vorliegende Dokument legt die für implantierbare Kardioverter-Defibrillatoren und CRT-D geltenden Anforderungen und die Funktionen von aktiven implantierbaren medizinischen Geräten fest, die zur Behandlung von Tachyarrhythmie vorgesehen sind.
Die in ISO 14708 festgelegten Prüfungen sind Typprüfungen und sind an Prüflingen eines Geräts durchzuführen, um die Anforderungen zu erfüllen.
Das vorliegende Dokument wurde für Tachyarrhythmie-Impulsgeneratoren ausgelegt, die entweder mit endokardialen Elektroden oder epikardialen Elektroden angewendet werden. Bei der Erarbeitung dieser Ausgabe wurde den Verfassern das Aufkommen von Technologien bewusst, bei denen keine endokardialen oder epikardialen Elektroden angewendet werden, und für welche sich die Notwendigkeit von Anpassungen des vorliegenden Teils ergeben wird. Solche Anpassungen liegen im Ermessen von Herstellern, die diese Technologien integrieren.
Das vorliegende Dokument gilt auch für einige nicht implantierbare Teile und Zubehörteile der Geräte (siehe Anmerkung 1).
Die Eigenschaften des implantierbaren Impulsgenerators oder der Elektrode müssen entweder nach dem jeweilig zutreffenden der im vorliegenden Dokument beschriebenen Verfahren oder nach einem anderen Verfahren bestimmt werden, für das nachgewiesen ist, dass dessen Präzision gleich dem festgelegten Verfahren oder besser ist. Im Zweifelsfall muss das im vorliegenden Dokument beschriebene Verfahren angewendet werden.
Jegliche Aspekte bezüglich aktiver implantierbarer medizinischer Geräte, die für die Behandlung von Bradyarrhythmien oder für kardiale Resynchronisation vorgesehen sind, werden in ISO 14708-2 behandelt.
ANMERKUNG 1 Ein Gerät, das üblicherweise als aktives implantierbares medizinisches Gerät bezeichnet wird, kann tatsächlich ein einzelnes Gerät, eine Kombination von Geräten oder eine Kombination von einem Gerät oder Geräten und einem einzelnen Zubehörteil oder mehreren Zubehörteilen sein. Nicht alle dieser Teile müssen entweder teilweise oder vollständig implantierbar sein, jedoch besteht die Notwendigkeit, einige Anforderungen an nicht implantierbare Teile und Zubehör festzulegen, wenn sie die Sicherheit oder Leistung des implantierbaren Geräts beeinflussen können.
ANMERKUNG 2 In diesem Dokument werden Begriffe, die in Abschnitt 3 definiert werden, in Kursivschrift gedruckt. Wo definierte Begriffe als Merkmal eines anderen Begriffes dienen, wird dieser nicht in Kursivschrift gedruckt, es sei denn, der fragliche Begriff ist auch definiert.

Implants chirurgicaux - Dispositifs médicaux implantables actifs - Partie 6: Exigences particulières pour les dispositifs médicaux implantables actifs destinés à traiter la tachyarythmie (y compris les défibrillateurs implantables) (ISO 14708-6:2019)

Les essais qui sont spécifiés dans l'ISO 14708 sont des essais de type et doivent être effectués sur des échantillons d'un dispositif pour vérifier la conformité.
Le présent document a été conçu pour les générateurs d'impulsions traitant la tachyarythmie utilisés avec des sondes endocardiques ou épicardiques. Au moment de l'élaboration de cette édition, les auteurs ont reconnu l'émergence de nouvelles technologies qui n'utilisent ni de sondes endocardiques ni de sondes épicardiques et pour lesquelles des adaptations de cette partie seront nécessaires. Ces adaptations sont laissées à la discrétion des fabricants qui intègrent ces technologies.
Ce document est également applicable à certaines parties non implantables et aux accessoires des dispositifs (voir Note 1).
Les caractéristiques du générateur d'impulsions implantable ou de la sonde doivent être déterminées soit par la méthode appropriée expliquée dans le présent document, soit par une autre méthode pour laquelle il a été démontré qu'elle possède un degré d'exactitude au moins égal à la méthode indiquée. En cas de désaccord, la méthode expliquée dans le présent document doit être appliquée.
Tout aspect d'un dispositif médical implantable actif destiné au traitement de la bradyarythmie ou de la resynchronisation cardiaque est couvert par l'ISO 14708-2.
NOTE 1    Le dispositif qui est généralement identifié comme un dispositif médical implantable actif peut être en fait un dispositif unique, une combinaison de dispositifs ou une combinaison d'un ou de plusieurs dispositifs et d'un nombre variable d'accessoires. Ces éléments ne sont pas nécessairement tous implantables, en partie ou en totalité, mais il est nécessaire de spécifier certaines exigences pour les parties et accessoires non implantables s'ils peuvent affecter la sécurité ou la performance du dispositif implantable.
NOTE 2    Dans le présent document, les termes écrits en italique sont utilisés conformément aux définitions données à l'Article 3. Lorsqu'un terme défini est utilisé comme qualificatif d'un autre terme, il n'est pas écrit en italique, sauf si le concept ainsi qualifié est également défini.

Vsadki (implantati) za kirurgijo - Aktivni medicinski pripomočki za vsaditev - 6. del: Posebne zahteve za aktivne vsadljive medicinske pripomočke, namenjene za zdravljenje tahiaritmije (vključuje vsadljive defibrilatorje) (ISO 14708-6:2019)

Ta dokument določa zahteve za vsadljive kardioverterje-defibrilatorje (ICD) in
defibrilatorje za srčno resinhronizacijsko terapijo (CRT-D) ter funkcije aktivnih vsadljivih medicinskih pripomočkov, namenjenih zdravljenju tahiaritmije.
Preskusi, ki so določeni v standardu ISO 14708, so tipski preskusi in jih je treba opraviti na vzorcih pripomočka,
da se dokaže skladnost.
Ta dokument je zasnovan za pulzne generatorje za zdravljenje tahiaritmije, ki se uporabljajo z endokardialnimi
ali epikardialnimi elektrodami. Avtorji so v času te izdaje upoštevali razvoj brezžičnih tehnologij, pri katerih se ne uporabljajo endokardialne oziroma epikardialne elektrode, zaradi česar bodo potrebne spremembe tega dela. Za te
spremembe so odgovorni proizvajalci, ki v svojih izdelkih uporabljajo omenjene tehnologije.
Ta dokument se uporablja tudi za nekatere dele in dodatno opremo pripomočkov, ki niso namenjeni vsaditvi (glej
opombo 1).
Karakteristike pulznega generatorja ali elektrod za vsaditev je treba določiti z ustrezno metodo, opisano v tem standardu, ali katero koli drugo metodo, ki je dokazano
enako ali bolj natančna od opisane metode. V primeru spora je treba uporabiti metodo, opisano v tem standardu.
Vsi vidiki aktivnih medicinskih pripomočkov za vsaditev, namenjenih zdravljenju bradiaritmije ali srčni resinhronizaciji, so zajeti v standardu ISO 14708-2.
OPOMBA 1: Pripomoček, ki se običajno imenuje aktivni medicinski pripomoček za vsaditev, je dejansko lahko
posamezen pripomoček, skupek pripomočkov ali kombinacija pripomočka ali pripomočkov in enega ali več kosov dodatne opreme.
Za nobenega od teh delov ni zahtevano, da jih je mogoče delno ali povsem vsaditi, vendar obstaja potreba po določanju nekaterih zahtev za dele in dodatno opremo, ki ni namenjena vsaditvi, če bi lahko vplivala na varnost ali delovanje pripomočka za vsaditev.
OPOMBA 2: V tem dokumentu so izrazi, natisnjeni v poševnem tisku, uporabljeni, kot je določeno v 3. točki. Če je določen izraz uporabljen kot oznaka v drugem izrazu, ni natisnjen v poševnem tisku, razen če je določen tudi tako kvalificiran koncept.

General Information

Status
Published
Public Enquiry End Date
26-Jul-2018
Publication Date
01-Sep-2022
Technical Committee
Current Stage
6060 - National Implementation/Publication (Adopted Project)
Start Date
04-Aug-2022
Due Date
09-Oct-2022
Completion Date
02-Sep-2022

Buy Standard

Standard
EN ISO 14708-6:2022 - BARVE
English language
75 pages
sale 10% off
Preview
sale 10% off
Preview
e-Library read for
1 day
Draft
prEN ISO 14708-6:2018 - BARVE
English language
82 pages
sale 10% off
Preview
sale 10% off
Preview
e-Library read for
1 day

Standards Content (Sample)

SLOVENSKI STANDARD
SIST EN ISO 14708-6:2022
01-oktober-2022
Vsadki (implantati) za kirurgijo - Aktivni medicinski pripomočki za vsaditev - 6. del:
Posebne zahteve za aktivne vsadljive medicinske pripomočke, namenjene za
zdravljenje tahiaritmije (vključuje vsadljive defibrilatorje) (ISO 14708-6:2019)
Implants for surgery - Active implantable medical devices - Part 6: Particular
requirements for active implantable medical devices intended to treat tachyarrhythmia
(including implantable defibrillators) (ISO 14708-6:2019)
Chirurgische Implantate - Aktive implantierbare medizinische Geräte - Teil 6: Besondere
Festlegungen für aktive implantierbare medizinische Produkte zur Behandlung von
Tachyarrhythmie (einschließlich implantierbaren Defibrillatoren) (ISO 14708-6:2019)
Implants chirurgicaux - Dispositifs médicaux implantables actifs - Partie 6: Exigences
particulières pour les dispositifs médicaux implantables actifs destinés à traiter la
tachyarythmie (y compris les défibrillateurs implantables) (ISO 14708-6:2019)
Ta slovenski standard je istoveten z: EN ISO 14708-6:2022
ICS:
11.040.40 Implantanti za kirurgijo, Implants for surgery,
protetiko in ortetiko prosthetics and orthotics
SIST EN ISO 14708-6:2022 en,fr,de
2003-01.Slovenski inštitut za standardizacijo. Razmnoževanje celote ali delov tega standarda ni dovoljeno.

---------------------- Page: 1 ----------------------
SIST EN ISO 14708-6:2022

---------------------- Page: 2 ----------------------
SIST EN ISO 14708-6:2022


EUROPEAN STANDARD EN ISO 14708-6

NORME EUROPÉENNE

EUROPÄISCHE NORM
July 2022
ICS 11.040.40

English version

Implants for surgery - Active implantable medical devices -
Part 6: Particular requirements for active implantable
medical devices intended to treat tachyarrhythmia
(including implantable defibrillators) (ISO 14708-6:2019)
Implants chirurgicaux - Dispositifs médicaux Chirurgische Implantate - Aktive implantierbare
implantables actifs - Partie 6: Exigences particulières medizinische Geräte - Teil 6: Besondere Festlegungen
pour les dispositifs médicaux implantables actifs für aktive implantierbare medizinische Produkte zur
conçus pour traiter la tachyarythmie (y compris les Behandlung von Tachyarrhythmie (einschließlich
défibrillateurs implantables) (ISO 14708-6:2019) implantierbaren Defibrillatoren) (ISO 14708-6:2019)
This European Standard was approved by CEN on 6 July 2022.

CEN and CENELEC members are bound to comply with the CEN/CENELEC Internal Regulations which stipulate the conditions for
giving this European Standard the status of a national standard without any alteration. Up-to-date lists and bibliographical
references concerning such national standards may be obtained on application to the CEN-CENELEC Management Centre or to
any CEN and CENELEC member.

This European Standard exists in three official versions (English, French, German). A version in any other language made by
translation under the responsibility of a CEN and CENELEC member into its own language and notified to the CEN-CENELEC
Management Centre has the same status as the official versions.

CEN and CENELEC members are the national standards bodies and national electrotechnical committees of Austria, Belgium,
Bulgaria, Croatia, Cyprus, Czech Republic, Denmark, Estonia, Finland, France, Germany, Greece, Hungary, Iceland, Ireland, Italy,
Latvia, Lithuania, Luxembourg, Malta, Netherlands, Norway, Poland, Portugal, Republic of North Macedonia, Romania, Serbia,
Slovakia, Slovenia, Spain, Sweden, Switzerland, Türkiye and United Kingdom.

















CEN-CENELEC Management Centre:
Rue de la Science 23, B-1040 Brussels
© 2022 CEN/CENELEC All rights of exploitation in any form and by any means
Ref. No. EN ISO 14708-6:2022 E
reserved worldwide for CEN national Members and for
CENELEC Members.

---------------------- Page: 3 ----------------------
SIST EN ISO 14708-6:2022
EN ISO 14708-6:2022 (E)
Contents Page
European foreword . 3

2

---------------------- Page: 4 ----------------------
SIST EN ISO 14708-6:2022
EN ISO 14708-6:2022 (E)
European foreword
This document (EN ISO 14708-6:2022) has been prepared by Technical Committee ISO/TC 150
"Implants for surgery" in collaboration with Technical Committee CEN-CENELEC/ JTC 16 “Active
Implantable Medical Devices” the secretariat of which is held by DKE.
This European Standard shall be given the status of a national standard, either by publication of an
identical text or by endorsement, at the latest by January 2023, and conflicting national standards shall
be withdrawn at the latest by January 2023.
Attention is drawn to the possibility that some of the elements of this document may be the subject of
patent rights. CEN-CENELEC shall not be held responsible for identifying any or all such patent rights.
Any feedback and questions on this document should be directed to the users’ national standards
body/national committee. A complete listing of these bodies can be found on the CEN and CENELEC
websites.
According to the CEN-CENELEC Internal Regulations, the national standards organizations of the
following countries are bound to implement this European Standard: Austria, Belgium, Bulgaria,
Croatia, Cyprus, Czech Republic, Denmark, Estonia, Finland, France, Germany, Greece, Hungary, Iceland,
Ireland, Italy, Latvia, Lithuania, Luxembourg, Malta, Netherlands, Norway, Poland, Portugal, Republic of
North Macedonia, Romania, Serbia, Slovakia, Slovenia, Spain, Sweden, Switzerland, Türkiye and the
United Kingdom.
Endorsement notice
The text of ISO 14708-6:2019 has been approved by CEN-CENELEC as EN ISO 14708-6:2022 without
any modification.


3

---------------------- Page: 5 ----------------------
SIST EN ISO 14708-6:2022

---------------------- Page: 6 ----------------------
SIST EN ISO 14708-6:2022
INTERNATIONAL ISO
STANDARD 14708-6
Second edition
2019-09
Implants for surgery — Active
implantable medical devices —
Part 6:
Particular requirements for active
implantable medical devices intended
to treat tachyarrhythmia (including
implantable defibrillators)
Implants chirurgicaux — Dispositifs médicaux implantables actifs —
Partie 6: Exigences particulières pour les dispositifs médicaux
implantables actifs conçus pour traiter la tachyarythmie (y compris
les défibrillateurs implantables)
Reference number
ISO 14708-6:2019(E)
©
ISO 2019

---------------------- Page: 7 ----------------------
SIST EN ISO 14708-6:2022
ISO 14708-6:2019(E)

COPYRIGHT PROTECTED DOCUMENT
© ISO 2019
All rights reserved. Unless otherwise specified, or required in the context of its implementation, no part of this publication may
be reproduced or utilized otherwise in any form or by any means, electronic or mechanical, including photocopying, or posting
on the internet or an intranet, without prior written permission. Permission can be requested from either ISO at the address
below or ISO’s member body in the country of the requester.
ISO copyright office
CP 401 • Ch. de Blandonnet 8
CH-1214 Vernier, Geneva
Phone: +41 22 749 01 11
Fax: +41 22 749 09 47
Email: copyright@iso.org
Website: www.iso.org
Published in Switzerland
ii © ISO 2019 – All rights reserved

---------------------- Page: 8 ----------------------
SIST EN ISO 14708-6:2022
ISO 14708-6:2019(E)

Contents Page
Foreword .v
Introduction .vi
1 Scope . 1
2 Normative references . 1
3 Terms and definitions . 2
4 Symbols and abbreviated terms . 6
5 General requirements for non-implantable parts . 6
5.1 General requirements for non-implantable parts . 6
5.2 General requirements for software . 6
5.3 Usability of non-implantable parts . 6
5.4 Data security and protection from harm caused by unauthorized information tampering 6
5.5 General requirements for risk management . 6
5.6 Misconnection of parts of the active implantable medical device . 6
6 Measurement of implantable pulse generator and lead characteristics .6
6.1 Measurement of implantable pulse generator characteristics . 6
6.1.1 General considerations . 6
6.1.2 Measurement of the bradyarrhythmia characteristics . 7
6.1.3 Measurement of ICD output voltage . 7
6.1.4 Measurement of delivered CD pulse energy . 8
6.1.5 Measurement of the antitachyarrhythmia pacing pulse amplitude . 8
6.1.6 Measurement of the sensitivity of an implantable pulse generator with
automatic sensitivity control. 8
6.1.7 Charge time. 9
6.1.8 Capacitor formation (capacitor maintenance) . 9
6.2 Measurement of the electrical characteristic of a sensing/pacing lead . 9
7 General arrangement of the packaging . 9
8 General markings for active implantable medical devices . 9
9 Markings on the sales packaging . 9
10 Construction of the sales packaging .11
11 Markings on the sterile pack.11
12 Construction of the non-reusable pack .13
13 Markings on the active implantable medical device .13
14 Protection from unintended biological effects being caused by the active
implantable medical device .13
15 Protection from harm to the patient or user caused by external physical features of
the active implantable medical device .14
16 Protection from harm to the patient caused by electricity .14
17 Protection from harm to the patient caused by heat .18
17.1 Protection from harm to the patient caused by heat .18
17.2 Active implantable medical device intended to supply heat .19
18 Protection from ionizing radiation released or emitted from the active implantable
medical device.19
19 Protection from unintended effects caused by the device .19
20 Protection of the device from damage caused by external defibrillators .20
© ISO 2019 – All rights reserved iii

---------------------- Page: 9 ----------------------
SIST EN ISO 14708-6:2022
ISO 14708-6:2019(E)

21 Protection of the device from changes caused by high-power electrical fields
applied directly to the patient .21
22 Protection of the active implantable medical device from changes caused by
miscellaneous medical treatments .21
23 Protection of the active implantable medical device from mechanical forces .21
24 Protection of the active implantable medical device from damage caused by
electrostatic discharge .25
25 Protection of the active implantable medical device from damage caused by
atmospheric pressure changes .26
26 Protection of the active implantable medical device from damage caused by
temperature changes .26
27 Protection of the active implantable medical device from electromagnetic non-
ionizing radiation .26
28 Accompanying documentation .26
Annex A (informative) Relationship between the fundamental principles in ISO/TR 14283
and the clauses of this document .32
Annex B (informative) Notes on ISO 14708-6.53
Annex C (informative) Code for describing modes of implantable pulse generators .64
Bibliography .66
iv © ISO 2019 – All rights reserved

---------------------- Page: 10 ----------------------
SIST EN ISO 14708-6:2022
ISO 14708-6:2019(E)

Foreword
ISO (the International Organization for Standardization) is a worldwide federation of national standards
bodies (ISO member bodies). The work of preparing International Standards is normally carried out
through ISO technical committees. Each member body interested in a subject for which a technical
committee has been established has the right to be represented on that committee. International
organizations, governmental and non-governmental, in liaison with ISO, also take part in the work.
ISO collaborates closely with the International Electrotechnical Commission (IEC) on all matters of
electrotechnical standardization.
The procedures used to develop this document and those intended for its further maintenance are
described in the ISO/IEC Directives, Part 1. In particular, the different approval criteria needed for the
different types of ISO documents should be noted. This document was drafted in accordance with the
editorial rules of the ISO/IEC Directives, Part 2 (see www .iso .org/directives).
Attention is drawn to the possibility that some of the elements of this document may be the subject of
patent rights. ISO shall not be held responsible for identifying any or all such patent rights. Details of
any patent rights identified during the development of the document will be in the Introduction and/or
on the ISO list of patent declarations received (see www .iso .org/patents).
Any trade name used in this document is information given for the convenience of users and does not
constitute an endorsement.
For an explanation of the voluntary nature of standards, the meaning of ISO specific terms and
expressions related to conformity assessment, as well as information about ISO's adherence to the
World Trade Organization (WTO) principles in the Technical Barriers to Trade (TBT) see www .iso
.org/iso/foreword .html.
ISO 14708-6 was prepared by Technical Committee ISO/TC 150, Implants for surgery, Subcommittee
SC 6, Active implants.
This second edition cancels and replaces the first edition (ISO 14708-6:2010), which has been technically
revised.
The main changes compared to the previous edition are as follows.
— addition of requirements for congestive heart failure devices;
— introduction of nomenclature for devices having more than two channels of pacing/sensing/
defibrillation as shown in ISO 14117:2019, Annex N;
— inclusion of new temporary exposure criteria in 17.1 for outer surface temperatures exceeding 39 °C;
— revision of atmospheric pressure test requirements in Clause 25 to align with requirements of
ISO 14708-2;
— replacement of detailed requirements in Clause 27 by reference to ISO 14117.
Other changes include updates to selected definitions and incorporation of new measurement
equipment accuracy requirements.
A list of all parts in the ISO 14708 series can be found on the ISO website.
Any feedback or questions on this document should be directed to the user’s national standards body. A
complete listing of these bodies can be found at www .iso .org/members .html.
© ISO 2019 – All rights reserved v

---------------------- Page: 11 ----------------------
SIST EN ISO 14708-6:2022
ISO 14708-6:2019(E)

Introduction
This document specifies particular requirements for implantable cardioverter defibrillators and the
functions of active implantable medical devices intended to treat tachyarrhythmia, to provide basic
assurance of safety for both patients and users.
An external defibrillator is a medical device used, in the emergency setting, to deliver a high-energy
shock to the heart, by means of electrodes applied to the external chest wall, in patients suffering
ventricular fibrillation (a rapid, disorganized and potentially lethal heart rhythm abnormality),
to restore normal heart action. External defibrillators can also be used, in emergency or elective
settings, to terminate other ventricular or atrial tachyarrhythmias by delivery of a high-energy shock,
synchronized to the intrinsic cardiac rhythm, a procedure known as cardioversion. In patients known
to be at risk of such arrhythmias, due to the occurrence of previous episodes or the presence of specific
predisposing cardiac conditions, an implantable cardioverter defibrillator might be implanted to perform
similar functions. The implantable device, which is much smaller than an external defibrillator, is
contained within a sealed, encapsulating enclosure. It generates high voltage pulses from an enclosed,
miniature, electrical battery. The pulses are transmitted to the heart by means of implanted, insulated
conductors with electrodes (leads). The implantable cardioverter defibrillator can also incorporate other
sensing and pacing functions, such as rate support for bradycardia and antitachycardia pacing (ATP) to
terminate certain tachyarrhythmias without the need of a high-energy shock. The defibrillator can be
adjusted non-invasively by means of an electronic device, known as a programmer.
In recent years, other active implantable cardiovascular devices have emerged, most notably devices
that perform the function of improving cardiac output by optimizing ventricular synchrony, in addition
to performing ICD functions.
Although these devices can deliver an additional therapy with respect to ICD devices, most of their
requirements are similar so that, in most cases, the concepts that apply to ICDs also apply to CRT-D
devices, and the appropriate way to test a CRT-D device is similar to the way ICDs are tested.
This document is relevant to all parts of active implantable medical devices intended to treat
tachyarrhythmia other than pacing functions to control bradyarrhythmia or provide cardiac
resynchronization. Typical examples are implantable pulse generators, leads, adaptors, accessories,
programmers and the related software (bradyarrhythmia and cardiac resynchronization pacing
functions are dealt with in ISO 14708-2).
The requirements of this document supplement or modify those of ISO 14708-1, Implants for surgery —
Active implantable medical devices — Part 1: General requirements for safety, marking and for information
to be provided by the manufacturer, hereinafter referred to as ISO 14708-1. The requirements of this
document take priority over those of ISO 14708-1.
In this document, terms printed in italic letters are used as defined in Clause 3. Where a defined term is
used as a qualifier in another term, it is not printed in italic letters unless the concept thus qualified is
also defined.
Information is also provided in Annex A that explains the relationship between ISO/TR 14283, Implants
for surgery — Essential principles of safety and performance, ISO 14708-1 and this document.
Notes on this document are provided in Annex B for information.
Annex C describes a coding system that may be used to designate tachyarrhythmia therapy modes. All
annexes are informative.
vi © ISO 2019 – All rights reserved

---------------------- Page: 12 ----------------------
SIST EN ISO 14708-6:2022
INTERNATIONAL STANDARD ISO 14708-6:2019(E)
Implants for surgery — Active implantable medical
devices —
Part 6:
Particular requirements for active implantable medical
devices intended to treat tachyarrhythmia (including
implantable defibrillators)
1 Scope
This document specifies requirements that are applicable to implantable cardioverter defibrillators and
CRT-Ds and the functions of active implantable medical devices intended to treat tachyarrhythmia.
The tests that are specified in ISO 14708 are type tests and are to be carried out on samples of a device
to show compliance.
This document was designed for tachyarrhythmia pulse generators used with either endocardial leads
or epicardial leads. At the time of this edition, the authors recognized the emergence of technologies
that do not use endocardial or epicardial leads for which adaptations of this part will be required. Such
adaptations are left to the discretion of manufacturers incorporating these technologies.
This document is also applicable to some non-implantable parts and accessories of the devices (see
Note 1).
The characteristics of the implantable pulse generator or lead shall be determined by either the
appropriate method detailed in this document or by any other method demonstrated to have accuracy
equal to, or better than, the method specified. In the case of dispute, the method detailed in this
document shall apply.
Any aspect of an active implantable medical device intended to treat bradyarrhythmias or cardiac
resynchronization is covered by ISO 14708-2.
NOTE 1 The device that is commonly referred to as an active implantable medical device can in fact be
a single device, a combination of devices, or a combination of a device or devices and one or more accessories.
Not all of these parts are required to be either partially or totally implantable, but there is a need to specify
some requirements of non-implantable parts and accessories if they could affect the safety or performance of the
implantable device.
NOTE 2 In this document, terms printed in italics are used as defined in Clause 3. Where a defined term is
used as a qualifier in another term, it is not printed in italics unless the concept thus qualified is also defined.
2 Normative references
The following documents are referred to in the text in such a way that some or all of their content
constitutes requirements of this document. For dated references, only the edition cited applies. For
undated references, the latest edition of the referenced document (including any amendments) applies.
ISO 5841-3:2013, Implants for surgery — Cardiac pacemakers — Part 3: Low-profile connectors (IS-1) for
implantable pacemakers
ISO 11318:2002, Cardiac defibrillators — Connector assembly DF-1 for implantable defibrillators —
Dimensions and test requirements
© ISO 2019 – All rights reserved 1

---------------------- Page: 13 ----------------------
SIST EN ISO 14708-6:2022
ISO 14708-6:2019(E)

ISO 14117:2019, Active implantable medical devices — Electromagnetic compatibility — EMC test protocols
for implantable cardiac pacemakers and implantable cardioverter defibrillators
ISO 14708-1:2014, Implants for surgery — Active implantable medical devices — Part 1: General
requirements for safety, marking and for information to be provided by the manufacturer
ISO 14708-2:2019, Implants for surgery — Active implantable medical devices — Part 2: Cardiac
pacemakers
IEC/TR 60878:2015, Graphical symbols for electrical equipment in medical practice
3 Terms and definitions
For the purposes of this document, the terms and definitions given in ISO 14708-1 and the following apply.
— ISO Online browsing platform: available at http: //www .iso .org/obp
— IEC Electropedia: available at http: //www .electropedia .org/
3.1
adaptor
special connector used between an otherwise incompatible implantable pulse generator and a lead
[SOURCE: ISO 14708-2:2019, 3.2]
3.2
implantable cardioverter defibrillator
ICD
active implantable medical device comprising an implantable pulse generator and lead(s) that is intended
to detect and correct tachycardias and fibrillation by application of cardioversion/defibrillation pulse(s)
to the heart
3.3
implantable pulse generator
IPG
part of the active implantable medical device, including the power supply and electronic circuit that
produces an electrical output
Note 1 to entry: For purposes of this document, the term implantable pulse generator describes any active
implantable medical device that incorporates functions intended to treat tachyarrhythmias.
[SOURCE: ISO 14708-2:2019, 3.4, modified – “active implantable medical device” substituted for
“pacemaker”, NOTE 1 to entry added]
3.4
sensitivity
minimum signal required to control consistently the function of the implantable pulse generator
[SOURCE: ISO 14708-2:2019, 3.8]
3.5
electrode
electrically conducting part (usually the termination of a lead), which is designed to form an interface
with body tissue or body fluid
[SOURCE: ISO 14708-2:2019, 3.9]
3.6
endocardial lead
lead with an electrode designed to make contact with the endocardium, or inner surface of the heart
[SOURCE: ISO 14708-2:2019, 3.12]
2 © ISO 2019 – All rights reserved

---------------------- Page: 14 ----------------------
SIST EN ISO 14708-6:2022
ISO 14708-6:2019(E)

3.7
epicardial lead
lead with an electrode designed to ma
...

SLOVENSKI STANDARD
oSIST prEN ISO 14708-6:2018
01-julij-2018
9VDGNL LPSODQWDWL ]DNLUXUJLMR$NWLYQLPHGLFLQVNLSULSRPRþNL]DYVDGLWHYGHO
3RVHEQH]DKWHYH]DDNWLYQHYVDGOMLYHPHGLFLQVNHSULSRPRþNHQDPHQMHQH]D
]GUDYOMHQMHWDKLDULWPLMH YNOMXþXMHYVDGOMLYHGHILEULODWRUMH  ,62',6
Implants for surgery - Active implantable medical devices - Part 6: Particular
requirements for active implantable medical devices intended to treat tachyarrhythmia
(including implantable defibrillators) (ISO/DIS 14708-6:2018)
Implants chirurgicaux - Dispositifs médicaux implantables actifs - Partie 6: Exigences
particulières pour les dispositifs médicaux implantables actifs destinés à traiter la
tachyarythmie (y compris les défibrillateurs implantables) (ISO/DIS 14708-6:2018)
Ta slovenski standard je istoveten z: prEN ISO 14708-6
ICS:
11.040.40 Implantanti za kirurgijo, Implants for surgery,
protetiko in ortetiko prosthetics and orthotics
oSIST prEN ISO 14708-6:2018 en,fr,de
2003-01.Slovenski inštitut za standardizacijo. Razmnoževanje celote ali delov tega standarda ni dovoljeno.

---------------------- Page: 1 ----------------------
oSIST prEN ISO 14708-6:2018

---------------------- Page: 2 ----------------------
oSIST prEN ISO 14708-6:2018
DRAFT INTERNATIONAL STANDARD
ISO/DIS 14708-6
ISO/TC 150/SC 6 Secretariat: ANSI
Voting begins on: Voting terminates on:
2018-05-04 2018-07-27
Implants for surgery — Active implantable medical
devices —
Part 6:
Particular requirements for active implantable medical
devices intended to treat tachyarrhythmia (including
implantable defibrillators)
Implants chirurgicaux — Dispositifs médicaux implantables actifs —
Partie 6: Exigences particulières pour les dispositifs médicaux implantables actifs destinés à traiter la
tachyarythmie (y compris les défibrillateurs implantables)
ICS: 11.040.40
THIS DOCUMENT IS A DRAFT CIRCULATED
This document is circulated as received from the committee secretariat.
FOR COMMENT AND APPROVAL. IT IS
THEREFORE SUBJECT TO CHANGE AND MAY
NOT BE REFERRED TO AS AN INTERNATIONAL
STANDARD UNTIL PUBLISHED AS SUCH.
IN ADDITION TO THEIR EVALUATION AS
ISO/CEN PARALLEL PROCESSING
BEING ACCEPTABLE FOR INDUSTRIAL,
TECHNOLOGICAL, COMMERCIAL AND
USER PURPOSES, DRAFT INTERNATIONAL
STANDARDS MAY ON OCCASION HAVE TO
BE CONSIDERED IN THE LIGHT OF THEIR
POTENTIAL TO BECOME STANDARDS TO
WHICH REFERENCE MAY BE MADE IN
Reference number
NATIONAL REGULATIONS.
ISO/DIS 14708-6:2018(E)
RECIPIENTS OF THIS DRAFT ARE INVITED
TO SUBMIT, WITH THEIR COMMENTS,
NOTIFICATION OF ANY RELEVANT PATENT
RIGHTS OF WHICH THEY ARE AWARE AND TO
©
PROVIDE SUPPORTING DOCUMENTATION. ISO 2018

---------------------- Page: 3 ----------------------
oSIST prEN ISO 14708-6:2018
ISO/DIS 14708-6:2018(E)
ISO 14708-6:201x

24 Contents
25 Foreword . 5
26 Introduction. 6
27 1 Scope . 7
28 2 Normative references . 7
29 3 Terms and definitions . 8
30 4 Symbols and abbreviated terms . 12
31 5 General requirements for non-implantable parts . 12
32 6 Measurement of IMPLANTABLE PULSE GENERATOR and LEAD characteristics . 12
33 7 General arrangement of the packaging . 15
34 8 General MARKINGS for ACTIVE IMPLANTABLE MEDICAL DEVICES . 15
35 9 MARKINGS on the SALES PACKAGING . 16
36 10 Construction of the SALES PACKAGING . 17
37 11 MARKINGS on the STERILE PACK . 17
38 12 Construction of the NON-REUSABLE PACK . 18
39 13 MARKINGS on the ACTIVE IMPLANTABLE MEDICAL DEVICE . 18
40 14 Protection from unintended biological effects being caused by the ACTIVE IMPLANTABLE MEDICAL
41 DEVICE . 19
42 15 Protection from HARM to the patient or user caused by external physical features of the ACTIVE
43 IMPLANTABLE MEDICAL DEVICE . 19
COPYRIGHT PROTECTED DOCUMENT 44 16 Protection from HARM to the patient caused by electricity . 19
© ISO 2018
45 17 Protection from HARM to the patient caused by heat . 25
All rights reserved. Unless otherwise specified, or required in the context of its implementation, no part of this publication may
be reproduced or utilized otherwise in any form or by any means, electronic or mechanical, including photocopying, or posting
on the internet or an intranet, without prior written permission. Permission can be requested from either ISO at the address 46 18 Protection from ionizing radiation released or emitted from the ACTIVE IMPLANTABLE MEDICAL
below or ISO’s member body in the country of the requester.
47 DEVICE . 26
ISO copyright office
CP 401 • Ch. de Blandonnet 8
CH-1214 Vernier, Geneva 48 19 Protection from unintended effects caused by the device. 26
Phone: +41 22 749 01 11
Fax: +41 22 749 09 47
49 20 Protection of the device from damage caused by external defibrillators . 28
Email: copyright@iso.org
Website: www.iso.org
Published in Switzerland
ii © ISO 2018 – All rights reserved © ISO 2018 – All rights reserved 3

---------------------- Page: 4 ----------------------
oSIST prEN ISO 14708-6:2018
ISO/DIS 14708-6:2018(E)
ISO 14708-6:201x
24 Contents
25 Foreword . 5
26 Introduction. 6
27 1 Scope . 7
28 2 Normative references . 7
29 3 Terms and definitions . 8
30 4 Symbols and abbreviated terms . 12
31 5 General requirements for non-implantable parts . 12
32 6 Measurement of IMPLANTABLE PULSE GENERATOR and LEAD characteristics . 12
33 7 General arrangement of the packaging . 15
34 8 General MARKINGS for ACTIVE IMPLANTABLE MEDICAL DEVICES . 15
35 9 MARKINGS on the SALES PACKAGING . 16
36 10 Construction of the SALES PACKAGING . 17
37 11 MARKINGS on the STERILE PACK . 17
38 12 Construction of the NON-REUSABLE PACK . 18
39 13 MARKINGS on the ACTIVE IMPLANTABLE MEDICAL DEVICE . 18
40 14 Protection from unintended biological effects being caused by the ACTIVE IMPLANTABLE MEDICAL
41 DEVICE . 19
42 15 Protection from HARM to the patient or user caused by external physical features of the ACTIVE
43 IMPLANTABLE MEDICAL DEVICE . 19
44 16 Protection from HARM to the patient caused by electricity . 19
45 17 Protection from HARM to the patient caused by heat . 25
46 18 Protection from ionizing radiation released or emitted from the ACTIVE IMPLANTABLE MEDICAL
47 DEVICE . 26
48 19 Protection from unintended effects caused by the device. 26
49 20 Protection of the device from damage caused by external defibrillators . 28
© ISO 2018 – All rights reserved 3

---------------------- Page: 5 ----------------------
oSIST prEN ISO 14708-6:2018
ISO/DIS 14708-6:2018(E)
ISO 14708-6:201x
50 21 Protection of the device from changes caused by high-power electrical fields applied directly
51 to the patient . 28
52 22 Protection of the ACTIVE IMPLANTABLE MEDICAL DEVICE from changes caused by miscellaneous
53 medical instruments . 28
54 23 Protection of the ACTIVE IMPLANTABLE MEDICAL DEVICE from mechanical forces . 28
55 24 Protection of the ACTIVE IMPLANTABLE MEDICAL DEVICE from damage caused by electrostatic
56 discharge . 34
57 25 Protection of the ACTIVE IMPLANTABLE MEDICAL DEVICE from damage caused by atmospheric
58 pressure changes . 34
59 26 Protection of the ACTIVE IMPLANTABLE MEDICAL DEVICE from changes caused by temperature
60 changes . 34
61 27 Protection of the ACTIVE IMPLANTABLE MEDICAL DEVICE from electromagnetic non-ionizing
62 radiation . 34
63 28 Accompanying documentation . 34
64 Annex A (informative) Relationship between the fundamental principles in ISO/TR 14283 and
65 the clauses of this standard. 40
66 Annex B (informative) Relationship between the clauses of this standard and the fundamental
67 principles listed in Annex A . 61
68 Annex C (informative) Notes on ISO 14708-6 . 63
69 Annex D (informative) Code for describing modes of IMPLANTABLE PULSE GENERATORS . 76
70 Annex E (informative) Defined terms . 79
71 Bibliography . 81
72
4 © ISO 2018 – All rights reserved

---------------------- Page: 6 ----------------------
oSIST prEN ISO 14708-6:2018
ISO/DIS 14708-6:2018(E)
ISO 14708-6:201x
73 Foreword
74 ISO (the International Organization for Standardization) is a worldwide federation of national
75 standards bodies (ISO member bodies). The work of preparing International Standards is normally
76 carried out through ISO technical committees. Each member body interested in a subject for which a
77 technical committee has been established has the right to be represented on that committee.
78 International organizations, governmental and non-governmental, in liaison with ISO, also take part in
79 the work. ISO collaborates closely with the International Electrotechnical Commission (IEC) on all
80 matters of electrotechnical standardization.
81 The procedures used to develop this document and those intended for its further maintenance are
82 described in the ISO/IEC Directives, Part 1. In particular, the different approval criteria needed for the
83 different types of ISO documents should be noted. This document was drafted in accordance with the
84 editorial rules of the ISO/IEC Directives, Part 2. www.iso.org/directives
85 Attention is drawn to the possibility that some of the elements of this document may be the subject of
86 patent rights. ISO shall not be held responsible for identifying any or all such patent rights. Details of
87 any patent rights identified during the development of the document will be in the Introduction and/or
88 on the ISO list of patent declarations received. www.iso.org/patents
89 Any trade name used in this document is information given for the convenience of users and does not
90 constitute an endorsement.
91 ISO 14708-6 was prepared by Technical Committee ISO/TC 150, Implants for surgery, Subcommittee
92 SC 6, Active implants.
93 A list of all parts in the ISO 14708 series can be found on the ISO website.
94 NOTE The attention of Member Bodies is drawn to the fact that equipment MANUFACTURERS and testing
95 organizations might need a transitional period following publication of a new, amended, or revised ISO
96 publication in which to make products in accordance with the new requirements and to equip them for
97 conducting new or revised tests. It is the recommendation of the committee that the content of this publication
98 not be adopted for mandatory implementation nationally earlier than three years from the date of publication.
© ISO 2018 – All rights reserved 5

---------------------- Page: 7 ----------------------
oSIST prEN ISO 14708-6:2018
ISO/DIS 14708-6:2018(E)
ISO 14708-6:201x
99 Introduction
100 This document specifies particular requirements for IMPLANTABLE CARDIOVERTER DEFIBRILLATORS and the
101 functions of ACTIVE IMPLANTABLE MEDICAL DEVICES intended to treat tachyarrhythmia, to provide basic
102 assurance of safety for both patients and users.
103 An external defibrillator is a MEDICAL DEVICE used, in the emergency setting, to deliver a high-energy
104 shock to the heart, by means of ELECTRODEs applied to the external chest wall, in patients suffering
105 ventricular fibrillation (a rapid, disorganized and potentially lethal heart rhythm abnormality), to
106 restore normal heart action. External defibrillators can also be used, in emergency or elective settings,
107 to terminate other ventricular or atrial tachyarrhythmias by delivery of a high-energy shock,
108 synchronized to the intrinsic cardiac rhythm, a procedure known as CARDIOVERSION. In patients known
109 to be at RISK of such arrhythmias, due to the occurrence of previous episodes or the presence of specific
110 predisposing cardiac conditions, an IMPLANTABLE CARDIOVERTER DEFIBRILLATOR might be implanted to
111 perform similar functions. The implantable device, which is much smaller than an external defibrillator,
112 is contained within a sealed, encapsulating enclosure. It generates high voltage PULSES from an enclosed,
113 miniature, electrical battery. The PULSES are transmitted to the heart by means of implanted, insulated
114 conductors with ELECTRODEs (LEADS). The IMPLANTABLE CARDIOVERTER DEFIBRILLATOR can also incorporate
115 other sensing and pacing functions, such as rate support for bradycardia and ANTITACHYCARDIA PACING
116 (ATP) to terminate certain tachyarrhythmias without the need of a high-energy shock. The defibrillator
117 can be adjusted non-invasively by means of an electronic device, known as a programmer.
118 In recent years, other active implantable cardiovascular devices have emerged, most notably devices
119 that perform the function of improving cardiac output by optimizing ventricular synchrony, in addition
120 to performing ICD functions.
121 Although these devices can deliver an additional therapy with respect to ICD devices, most of their
122 requirements are similar so that, in most cases, the concepts that apply to ICDs also apply to CRT-D
123 devices, and the appropriate way to test a CRT-D device is similar to the way ICDs are tested.
124 This document is relevant to all parts of ACTIVE IMPLANTABLE MEDICAL DEVICES intended to treat
125 tachyarrhythmia other than pacing functions to control bradyarrhythmia or provide cardiac
126 resynchronization. Typical examples are IMPLANTABLE PULSE GENERATORs, LEADS, ADAPTORS, ACCESSORIES,
127 programmers and the related software (bradyarrhythmia and cardiac resynchronization pacing
128 functions are dealt with in ISO 14708-2).
129 The requirements of this document supplement or modify those of ISO 14708-1, Implants for surgery —
130 Active implantable medical devices — Part 1: General requirements for safety, marking and for
131 information to be provided by the manufacturer, hereinafter referred to as ISO 14708-1. The
132 requirements of this document take priority over those of ISO 14708-1.
133 Figures or tables that are additional to those of ISO 14708-1 are numbered starting from 101.
134 Annex D describes a coding system that may be used to designate tachyarrhythmia therapy modes. All
135 annexes are informative.
6 © ISO 2018 – All rights reserved

---------------------- Page: 8 ----------------------
oSIST prEN ISO 14708-6:2018
ISO/DIS 14708-6:2018(E)
ISO 14708-6:201x
136 Implants for surgery — Active implantable medical devices —
137 Part 6: Particular requirements for active implantable medical
138 devices intended to treat tachyarrhythmia (including implantable
139 defibrillators)
140 1 Scope
141 This document specifies requirements that are applicable to IMPLANTABLE CARDIOVERTER DEFIBRILLATORS
142 AND CRT-DS and the functions of ACTIVE IMPLANTABLE MEDICAL DEVICES intended to treat tachyarrhythmia.
143 The tests that are specified in ISO 14708 are type tests and are to be carried out on samples of a device to
144 show compliance.
145 This document was designed for tachyarrhythmia PULSE generators used with either endocardial or
146 epicardial LEADs. At the time of this edition, the authors recognized the emergence of technologies that
147 do not use ENDOCARDIAL or EPICARDIAL LEADS for which adaptations of this part will be required. Such
148 adaptations are left to the discretion of MANUFACTURERS incorporating these technologies.
149 This document is also applicable to some non-implantable parts and accessories of the devices (see
150 Note 1).
151 The characteristics of the IMPLANTABLE PULSE GENERATOR or LEAD shall be determined by either the
152 appropriate method detailed in this document or by any other method demonstrated to have accuracy
153 equal to, or better than, the method specified. In the case of dispute, the method detailed in this
154 document shall apply.
155 Any aspect of an ACTIVE IMPLANTABLE MEDICAL DEVICE intended to treat bradyarrhythmias or cardiac
156 resynchronization is covered by ISO 14708-2.
157 NOTE 1  The device that is commonly referred to as an ACTIVE IMPLANTABLE MEDICAL DEVICE might in fact be a single
158 device, a combination of devices, or a combination of a device or devices and one or more accessories. Not all of
159 these parts are required to be either partially or totally implantable, but there is a need to specify some
160 requirements of non-implantable parts and accessories if they could affect the safety or performance of the
161 implantable device.
162 NOTE 2  The terminology used in this European Standard is intended to be consistent with the terminology of
163 Directive 90/385/EEC.
164 NOTE 3  In this document, terms printed in small capital letters are used as defined in Clause 3. Where a defined
165 term is used as a qualifier in another term, it is not printed in small capital letters unless the concept thus qualified
166 is also defined.
167 NOTE 4  Particular requirements for congestive heart failure devices are under consideration. These types of
168 devices are not covered by this document.
169 2 Normative references
170 The following documents, in whole or in part, are normatively referenced in this document and are
171 indispensable for its application. For dated references, only the edition cited applies. For undated
172 references, the latest edition of the referenced document (including any amendments) applies.
173 ISO 5841-3:2013, Implants for surgery — Cardiac pacemakers — Part 3: Low-profile connectors (IS-1) for
174 implantable pacemakers
© ISO 2018 – All rights reserved 7

---------------------- Page: 9 ----------------------
oSIST prEN ISO 14708-6:2018
ISO/DIS 14708-6:2018(E)
ISO 14708-6:201x
175 ISO 8601:2004, Data elements and interchange formats — Information interchange — Representation of
176 dates and times
177 ISO 11318:2002, Cardiac defibrillators — Connector assembly DF-1 for implantable defibrillators —
178 Dimensions and test requirements
179 ISO 14117:2018, Active implantable medical devices — Electromagnetic compatibility — EMC test
180 protocols for implantable cardiac pacemakers and implantable cardioverter defibrillators
181 ISO 14708-1:2014, Implants for surgery — Active implantable medical devices — Part 1: General
182 requirements for safety, marking and for information to be provided by the manufacturer
183 ISO 14708-2:2018, Implants for surgery — Active implantable medical devices — Part 2: Cardiac
184 pacemakers
185 ISO 15223-1, Medical devices — Symbols to be used with medical device labels, labelling and information
186 to be supplied — Part 1: General requirements
187 IEC 60878, Graphical symbols for electrical equipment in medical practice
188 3 Terms and definitions
189 For the purposes of this document, the terms and definitions given in ISO 14708-1 and the following
190 apply.
191  ISO Online browsing platform: available at http://www.iso.org/obp
192  IEC Electropedia: available at http://www.electropedia.org/
193 3.101
194 adaptor
195 special connector used between an otherwise incompatible IMPLANTABLE PULSE GENERATOR and a LEAD
196 [SOURCE: ISO 14708-2:2018, 3.102]
197 3.102
198 implantable cardioverter defibrillator
199 ICD
200 ACTIVE IMPLANTABLE MEDICAL DEVICE comprising an IMPLANTABLE PULSE GENERATOR and LEAD(S) that is
201 intended to detect and correct tachycardias and fibrillation by application of
202 CARDIOVERSION/DEFIBRILLATION PULSE(s) to the heart
203 3.103
204 implantable pulse generator
205 IPG
206 part of the ACTIVE IMPLANTABLE MEDICAL DEVICE, including the power supply and electronic circuit that
207 produces an electrical output
208 NOTE 1 to entry: For purposes of this document, the term IMPLANTABLE PULSE GENERATOR describes any ACTIVE
209 IMPLANTABLE MEDICAL DEVICE that incorporates functions intended to treat tachyarrhythmias.
210 [SOURCE: ISO 14708-2:2018, 3.104, modified – “ACTIVE IMPLANTABLE MEDICAL DEVICE” substituted for
211 “PACEMAKER”, NOTE 1 to entry added]
8 © ISO 2018 – All rights reserved

---------------------- Page: 10 ----------------------
oSIST prEN ISO 14708-6:2018
ISO/DIS 14708-6:2018(E)
ISO 14708-6:201x
212 3.104
213 sensitivity
214 sensing threshold
215 minimum signal required to consistently control the function of the IMPLANTABLE PULSE GENERATOR
216 [SOURCE: ISO 14708-2:2018, 3.108]
217 3.106
218 electrode
219 electrically conducting part (usually the termination of a LEAD), which is designed to form an interface
220 with body tissue or body fluid
221 [SOURCE: ISO 14708-2:2018, 3.109]
222 3.107
223 endocardial lead
224 LEAD with an ELECTRODE designed to make a contact with the endocardium, or inner surface of the heart
225 [SOURCE: ISO 14708-2:2018, 3.112]
226 3.108
227 epicardial lead
228 LEAD with an ELECTRODE designed to make a contact with the epicardium, or outer surface of the heart
229 [SOURCE: ISO 14708-2:2018, 3.113]
230 3.109
231 transvenous
232 approach to the heart through the venous system
233 [SOURCE: ISO 14708-2:2018, 3.114]
234 3.110
235 insertion diameter
236 lead
237 minimum bore of a rigid cylindrical tube into which the LEAD (not including the connector) can be
238 inserted
239 [SOURCE: ISO 14708-2:2018, 3.115]
240 3.112
241 lead pacing impedance
242 Z
p
243 impedance that is formed by the ratio of a voltage PULSE to the resulting current
244 [SOURCE: ISO 14708-2:2018, 3.117].
245 3.114
246 model designation
247 name and/or a combination of letters and numbers used by a MANUFACTURER to distinguish, by function
248 or type, one device from another
249 [SOURCE: ISO 14708-2:2018, 3.119]
© ISO 2018 – All rights reserved 9

---------------------- Page: 11 ----------------------
oSIST prEN ISO 14708-6:2018
ISO/DIS 14708-6:2018(E)
ISO 14708-6:201x
250 3.115
251 serial number
252 unique combination of letters and/or numbers, selected by the MANUFACTURER, intended to distinguish a
253 device from other devices with the same MODEL DESIGNATION
254 [SOURCE: ISO 14708-2:2018, 3.120]
255 3.117
256 pulse
257 electrical output of an IMPLANTABLE PULSE GENERATOR other than CD PULSE intended to stimulate the
258 myocardium
259 [SOURCE: ISO 14708-2:2018, 3.122, modified – added “other than CD PULSE”.]
260 3.118
261 pulse amplitude
262 amplitude of the PULSE
263 [SOURCE: ISO 14708-2:2018, 3.123]
264 3.119
265 pulse duration
266 duration of the PULSE
267 [SOURCE: ISO 14708-2:2018, 3.124]
268 3.120
269 pulse interval
270 interval between equivalent points of two consecutive PULSEs
271 [SOURCE: ISO 14708-2:2018, 3.125]
272 3.122
273 automatic sensitivity control
274 automatic adjustment of the SENSITIVITY in response to available physiological signals
275 3.123
276 beginning of service
277 BOS
278 time at which an individual IMPLANTABLE PULSE GENERATOR is first released by the MANUFACTURER as fit for
279 being placed on the market
280 [SOURCE: ISO 14708-2:2018, 3.138]
281 3.124
282 end of service
283 EOS
284 time at which the PROLONGED SERVICE PERIOD has elapsed and no further pacing function is specified nor
285 can be expected
286 [SOURCE: ISO 14708-2: 2018, 3.139]
10 © ISO 2018 – All rights reserved

---------------------- Page: 12 ----------------------
oSIST prEN ISO 14708-6:2018
ISO/DIS 14708-6:2018(E)
ISO 14708-6:201x
287 3.125
288 prolonged service period
289 PSP
290 period beyond the RECOMMENDED REPLACEMENT TIME during which the IMPLANTABLE PULSE GENERATOR
291 continues to function as specified by the MANU
...

Questions, Comments and Discussion

Ask us and Technical Secretary will try to provide an answer. You can facilitate discussion about the standard in here.