Abstract

This document specifies requirements for single-use burettes to be used as subcomponents within sterilised infusion and transfusion sets for single use under gravity feed conditions or with pressure infusion apparatus.
NOTE 1        Annex B provides explanations about the history of the development of the standard and summarises the different arguments discussed within ISO/TC 76 during the collaboration of this document.
NOTE 2        Annex C provides recommendations regarding sustainability.
NOTE 3        Annex D provides information on attributive and variable testing.

Status
Not Published
Public Enquiry End Date
30-Dec-2025
Technical Committee
VAZ - Healthcare
Current Stage
5020 - Formal vote (FV) (Adopted Project)
Start Date
10-Sep-2026
Due Date
29-Oct-2026
Completion Date
24-Sep-2026

Buy Documents

Draft

oSIST prEN ISO 8536-5:2025

English language (15 pages)
Preview
Preview
e-Library read for
×1 day

kSIST FprEN ISO 8536-5:2026 is a CEN standard for infusion equipment for medical use, Part 5. It specifies requirements for single-use burette infusion sets used as subcomponents in sterilised infusion and transfusion sets under gravity feed conditions, and it also covers use with pressure infusion apparatus. The document is meant for manufacturers, test laboratories, quality teams, and buyers who need to check compatibility, scale accuracy, and physical performance of burette components.

What does kSIST FprEN ISO 8536-5:2026 specify?

kSIST FprEN ISO 8536-5:2026 covers the requirements for single-use burettes as components within infusion and transfusion equipment. The standard is organized around design, materials, physical performance, chemical requirements, biological requirements, and supporting annexes for testing, rationale, and sustainability.

AnnexWhat it covers
Annex APhysical tests, including the scale graduation accuracy test
Annex BRationale, guidance, and development history
Annex CSustainability aspects

In practice, the document is about the burette itself, while the surrounding infusion or transfusion set is handled by the relevant part of the ISO 8536 or ISO 1135 series.

What are the key requirements of kSIST FprEN ISO 8536-5:2026?

kSIST FprEN ISO 8536-5:2026 focuses on design, scale accuracy, material suitability, and basic physical performance. The requirements are written so that a burette can be read correctly, vented safely, and used with the intended parent set.

Design and graduated scale

  • Clause 4.1 requires a rigid or semi-rigid plastic burette that lets the user see the fluid, includes filtered air venting above the top graduation mark, and can be closed off as a separate self-vented reservoir. That matters because the burette must be readable and controllable during infusion.
  • Clause 4.1.5 links light-protective burettes to the light-protective requirements used for drip chambers in ISO 8536-15. If a product claims light protection, the manufacturer needs to check that the light-protective approach is handled consistently.
  • Clauses 4.2 to 4.4 control the graduated scale: the volume unit must be marked, the scale must be numbered and laid out according to Table 1, and any extension beyond nominal capacity must be visually distinct. This helps prevent reading errors and supports dose control.
  • Clause 4.2.7 requires scale accuracy testing in Annex A. The test confirms that the marked scale matches actual delivered volume.

Materials and performance

  • Clause 5 ties the materials to the physical, chemical, and biological requirements, and requires assessment under ISO 10993-1. In practice, material choice has to support patient safety as well as the intended use.
  • Clauses 6.1 to 6.6 cover particulate contamination, tensile strength, leakage, air-inlet device performance, filter choice, and drip chamber/drip tube performance. The practical point is to verify that the burette stays clean, mechanically sound, leak-tight, and capable of controlled drop formation.
  • Clauses 7 and 8 route chemical and biological requirements to ISO 1135-4 or ISO 8536-4, as applicable. That keeps the component aligned with the rest of the infusion or transfusion system.

Testing and usability

  • Annex A gives the physical test for scale graduation accuracy using sodium chloride solution and weighing. This is the main verification method for the graduated volume marking.
  • Clause 4.5 allows scale variation for a specific application, but recommends risk assessment under ISO 14971 and usability evaluation under IEC 62366-1. If a manufacturer changes the scale layout, it should be justified by risk and usability evidence.

What terms does kSIST FprEN ISO 8536-5:2026 define?

kSIST FprEN ISO 8536-5:2026 uses several key technical terms in its requirements:

  • Burette - a rigid or semi-rigid chamber that can be filled, observed, and then closed off as a separate reservoir.
  • Nominal capacity - the designated volume of the burette, based on the total graduated capacity.
  • Graduated scale - the marked volume scale that must be numbered and checked for accuracy.
  • Drip chamber - the chamber where drop fall is observed before fluid continues downstream.
  • Air-inlet device - the vented entry, with filter, that lets air into the system above the burette.
  • Fluid filter / filter for blood and blood components - the filter chosen for the intended application, depending on whether the set is for infusion or transfusion use.

Who uses kSIST FprEN ISO 8536-5:2026?

kSIST FprEN ISO 8536-5:2026 is used by manufacturers of infusion and transfusion devices, component suppliers, test laboratories, and quality or regulatory teams. It is also relevant to specifiers and buyers who need to match burette components with sterilised infusion sets, pressure infusion apparatus, or transfusion equipment. In practice, these users rely on it for design checks, incoming inspection, product verification, and risk documentation.

What changed in kSIST FprEN ISO 8536-5:2026 from the previous edition?

kSIST FprEN ISO 8536-5:2026 cancels and replaces ISO 8536-5:2004. The foreword states these main changes:

  • The scope was changed to treat burettes as components within infusion and transfusion devices.
  • The scope was expanded to include additional burette applications in transfusion.
  • Clauses were added or revised for light-resistance, the graduated scale, particulate contamination, air-inlet, leakage, and drop former requirements.
  • The clauses on labelling, packaging, and disposal were removed.
  • Annexes A, B, and C were added.

Which standards are used with kSIST FprEN ISO 8536-5:2026?

kSIST FprEN ISO 8536-5:2026 relies on several referenced standards and related parts of the same family:

  • IEC 62366-1 - supports usability engineering when scale layout or scale variation is assessed.
  • ISO 1135-4 - provides transfusion equipment requirements and the basis for blood and blood component filter use, plus applicable chemical and biological requirements.
  • ISO 8536-4 - provides gravity infusion set requirements used for tensile strength, leakage, air-inlet device, and fluid filter provisions.
  • ISO 8536-8 - provides upstream leakage requirements for burettes used with pressure infusion apparatus.
  • ISO 8536-15 - provides the light-protective requirements used when a burette claims light protection.
  • ISO 10993-1 - gives the biological evaluation framework for the materials.
  • ISO 14971 - supports risk management for design and scale variations.
  • ISO 24072 - gives the aerosol bacterial retention test method for the air-inlet filter.
  • ISO/TR 8417 - is cited for background on particulate contamination and possible control measures.

What does the kSIST FprEN ISO 8536-5:2026 document contain?

kSIST FprEN ISO 8536-5:2026 contains a typical burette design figure, example graduated scales, a table for scale intervals and tolerances, and the physical test method in Annex A. The annexes also include the rationale for the requirements and sustainability guidance for design and manufacturing. The bibliography points readers to ISO 7886, ISO/TR 8417, and the European, United States, and Japanese Pharmacopoeias for related background.

Relations

Effective Date
06-Dec-2023

Buy Documents

Draft

oSIST prEN ISO 8536-5:2025

English language (15 pages)
Preview
Preview
e-Library read for
×1 day

Get Certified

Connect with accredited certification bodies for this standard

BSI Group

BSI (British Standards Institution) is the business standards company that helps organizations make excellence a habit.

UKAS United Kingdom Verified

TÜV Rheinland

TÜV Rheinland is a leading international provider of technical services.

DAKKS Germany Verified

TÜV SÜD

TÜV SÜD is a trusted partner of choice for safety, security and sustainability solutions.

DAKKS Germany Verified

Sponsored listings

Frequently Asked Questions

kSIST FprEN ISO 8536-5:2026 is a draft published by the Slovenian Institute for Standardization (SIST). Its full title is "Infusion equipment for medical use - Part 5: Burette infusion sets for single use, gravity feed (ISO/FDIS 8536-5:2026)". This standard covers: This document specifies requirements for single-use burettes to be used as subcomponents within sterilised infusion and transfusion sets for single use under gravity feed conditions or with pressure infusion apparatus. NOTE 1        Annex B provides explanations about the history of the development of the standard and summarises the different arguments discussed within ISO/TC 76 during the collaboration of this document. NOTE 2        Annex C provides recommendations regarding sustainability. NOTE 3        Annex D provides information on attributive and variable testing.

This document specifies requirements for single-use burettes to be used as subcomponents within sterilised infusion and transfusion sets for single use under gravity feed conditions or with pressure infusion apparatus. NOTE 1        Annex B provides explanations about the history of the development of the standard and summarises the different arguments discussed within ISO/TC 76 during the collaboration of this document. NOTE 2        Annex C provides recommendations regarding sustainability. NOTE 3        Annex D provides information on attributive and variable testing.

kSIST FprEN ISO 8536-5:2026 is classified under the following ICS (International Classification for Standards) categories: 11.040.20 - Transfusion, infusion and injection equipment. The ICS classification helps identify the subject area and facilitates finding related standards.

kSIST FprEN ISO 8536-5:2026 has the following relationships with other standards: It is inter standard links to SIST EN ISO 8536-5:2013. Understanding these relationships helps ensure you are using the most current and applicable version of the standard.

kSIST FprEN ISO 8536-5:2026 is available in PDF format for immediate download after purchase. The document can be added to your cart and obtained through the secure checkout process. Digital delivery ensures instant access to the complete standard document.

Standards Content (Sample)

Contents

  • 1 Scope
  • 2 Normative references
  • 3 Terms and definitions
  • 4 General requirements
  • 4.1 Design
  • 4.2 Graduated scale
  • 4.3 Numbering of scales
  • 4.4 Overall length of scale to nominal capacity line
  • 4.5 Variation of scale
  • 5 Materials
  • 6 Physical requirements
  • 6.1 Particulate contamination
  • 6.2 Tensile strength
  • 6.3 Leakage
  • 6.4 Air-inlet device
  • 6.5 Fluid filter / filter for blood and blood components
  • 6.6 Drip chamber and drip tube
  • 7 Chemical requirements
  • 8 Biological requirements
  • Annex A
  • Annex B
  • 1. Particulate contamination
  • 2. Leakage
  • 3. Air-inlet
  • 4. Drop former
  • Annex C
  • Bibliography

1 Scope

This document specifies requirements for single-use burettes to be used as subcomponents within sterilised infusion and transfusion sets for single use under gravity feed conditions or with pressure infusion apparatus.

2 Normative references

The following documents are referred to in the text in such a way that some or all of their content constitutes requirements of this document. For dated references, only the edition cited applies. For undated references, the latest edition of the referenced document (including any amendments) applies. IEC 62366-1, Medical devices — Part 1: Application of usability engineering to medical devices ISO 1135-4, Transfusion equipment for medical use — Part 4: Transfusion sets for single use, gravity feed ISO 8536-4, Infusion equipment for medical use — Part 4: Infusion sets for single use, gravity feed ISO 8536-8, Infusion equipment for medical use — Part 8: Infusion sets for single use with pressure infusion apparatus ISO 8536-15, Infusion equipment for medical use — Part 15: Light-protective infusion sets for single use ISO 10993-1, Biological evaluation of medical devices — Part 1: Evaluation and testing within a risk management process ISO 14971, Medical devices — Application of risk management to medical devices ISO 24072, Aerosol bacterial retention test method for air-inlet filter on administration devices

3 Terms and definitions

No terms and definitions are listed in this document.

ISO and IEC maintain terminology databases for use in standardization at the following addresses:

  • ISO Online browsing platform: available at https:// www .iso .org/ obp
  • IEC Electropedia: available at https:// www .electropedia .org/

oSIST prEN ISO 8536-5:2025 ISO/DIS 8536-5:2025(en)

4 General requirements

4.1 Design

4.1.1 A typical design along with the nomenclature to be used for burettes is given in Figure 1. Key a 1 injection site 2 air inlet with air filter and closure 3 graduated burette a 4 shut-off valve 5 drip tube 6 drip chamber b c 7 fluid filter / filter for blood and blood components a Optional.

b The fluid filter may be positioned at other sites, for example preferably near the patient access. Generally a fluid filter has a nominal pore size of 15 µm.

Indicates alternative locations of the filter for blood and blood components. Other designs are acceptable, if the same safety aspects are ensured.

Figure 1 — Example of a burette

oSIST prEN ISO 8536-5:2025 ISO/DIS 8536-5:2025(en) 4.1.2 The burette shall consist of a tube of rigid or semi-rigid plastics material and shall permit observation of fluid in the chamber.

4.1.3 The burette shall be provided with filtered air-venting capability located in a position above the top graduation mark.

4.1.4 The burette shall be capable of receiving fluid from the main container (or from the injection site / LAV as applicable) and of being closed off and serving as a separate self-vented reservoir. 4.1.5 For burettes that claim light protection, the requirements for drip chambers specified in ISO 8536-15 shall apply to the graduated burette which is denoted as item 3 in Figure 1. 4.1.6 The nominal volume of the burette shall be designated by the total graduated capacity.

4.2 Graduated scale

4.2.1 The burette shall have either only one scale or more than one identical scale, which shall be graduated and numbered at least at the intervals given in Table 1. The unit of volume shall be marked on the burette. NOTE The scale interval can be less (finer) than the scale interval given in Table 1. 4.2.2 The total graduated capacity may be equal to, or greater than, the nominal capacity. If the scale is extended beyond the nominal capacity, the extended portion shall be differentiated from the rest of the scale. Examples of means of differentiation are the following:

a) encircling the scale number of the nominal capacity line; b) using smaller scale numbers for the extra graduation lines; c) using shorter graduation lines for the extra graduation lines; d) using a broken line for the optional vertical line of the extra scale length. 4.2.3 Graduation lines shall be of uniform thickness. They shall lie in planes at right angles to the axis of the burette.

4.2.4 Graduation lines shall be evenly spaced along the longitudinal axis between the zero graduation line and the line for the total graduated capacity.

4.2.5 When the burette is held vertically, the ends of all graduation lines of similar length shall be vertically beneath each other.

The lengths of the short graduation lines on each scale are recommended to be approximately half the length of the long lines. If different graduation line configurations are used, this could be submitted to usability evaluation according to IEC 62366-1.

Examples of scales and the numbering of graduation lines are shown in Figure 2. 4.2.6 The zero-position mark on the burette shall be located in a position which compensates for the volume displaced by any shut-off device, the position of the outlet relative to the bottom of the burette, or any other feature of the bottom cap which can affect the reading.

oSIST prEN ISO 8536-5:2025 ISO/DIS 8536-5:2025(en) 4.2.7 The accuracy of the scale shall be tested in accordance with Annex A and shall meet the accuracy requirements of Table 1.

Table 1 — Volume and scale intervals for burettes Tolerance on any graduated capacity Minimum Increment Less than half overall length between Equal to or Nominal capac- Less than and more of scale to nom- Scale interval graduation greater than ity of burette 20 % nominal than or equal inal capacity lines to be half nominal capacity to 20 % nomimark numbered capacity nal capacity V ml mm ml ml ±(1,5 % of V +1 % of expelled ±5 % of ex- V < 50 67 2,0 10 volume) pelled volume ±(1,5 % of V +1 % of expelled ±5 % of ex- V ≥ 50 75 5,0 10 volume) pelled volume NOTE Expelled volume means all the liquid ejected to the point where the meniscus of the water level is coincident with the zero mark of the scale.

NOTE The vertical lines are optional.

Figure 2 — Typical graduated scales for use in burettes

oSIST prEN ISO 8536-5:2025 ISO/DIS 8536-5:2025(en)

4.3 Numbering of scales

4.3.1 Graduation lines shall be numbered at least at the volume of increments given in Table 1. In addition, the line denoting the nominal capacity or the lines denominal capacity and the total graduation capacity, if these differ, shall be numbered.

Examples of scale numbering are shown in Figure 2.

4.3.2 When the burette is held vertically with the scale to the front, the numbers shall appear vertical on the scale and be approximately centred on the graduation lines to which they relate. The numbers shall be close to, but shall not touch, the ends of the graduation lines to which they relate.

4.4 Overall length of scale to nominal capacity line

The overall length of the scale shall be as given in Table 1.

4.5 Variation of scale

If necessary by a specific application, the scale may vary and this requirement does not preclude the provision of additional graduation marks within the scale or as extensions to the scale. Any variation of the scale or graduation is recommended to be assessed for risk according to ISO 14971 and for usability according to IEC 62366-1.

...