IEC 60645-6:2022 is available as IEC 60645-6:2022 RLV which contains the International Standard and its Redline version, showing all changes of the technical content compared to the previous edition.
IEC 606045-6:2022 applies to instruments designed primarily for the measurement of otoacoustic emissions in the human external auditory meatus evoked by acoustic probe stimuli. This document defines the characteristics to be specified by the manufacturer, specifies minimum mandatory functions for two types of instruments and provides performance specifications applicable to both instrument types. This document describes methods to be used to demonstrate conformance with the specifications in this document and guidance on methods for periodic calibration. The purpose of this document is to ensure that measurements made under comparable test conditions with different instruments complying with this document will be consistent. Instruments can provide a measurement function not specifically within the scope of this document and still comply with the relevant requirements of this document for the functions that are within the scope. This document is not intended to restrict development or incorporation of new features, nor to discourage innovative approaches. IEC 606045-6:2022 cancels and replaces the first edition published in 2009. This edition constitutes a technical revision.
This edition includes the following significant technical changes with respect to the previous edition:
a) the nominal test frequency used in DPOAE is now defined as the higher of the two frequencies, f2;
b) the permitted deviation of the stimulus signal for TEOAE has been specified;
c) the frequency range for DPOAE stimulus signals has been redefined,
d) the stimulus level requirements for TEOAE have been redefined;
e) the stimulus level requirements for DPOAE have been redefined;
f) the harmonic distortion requirements for DPOAE have been redefined;
g) a minimum measurement range for DPOAE has been added.

  • Standard
    34 pages
    English and French language
    sale 15% off

IEC 60645-3:2020 specifies a means of describing the physical characteristics, in terms of electrical waveforms, of audiometric reference and test signals of short duration and methods for their measurement.
The object of this document is to ensure that audiometric stimuli of short duration are specified and measured in the same way and that the calibration of equipment using such signals is carried out using defined methods.
This document does not describe the method of use of short-duration test signals.
IEC 60645-3:2020 cancels and replaces the second edition published in 2007. This edition constitutes a technical revision.
This edition includes the following significant technical changes with respect to the previous edition:
a) new figures of reference signals;
b) changes in definitions.

  • Standard
    28 pages
    English and French language
    sale 15% off

IEC 60645-1:2017 is also available as IEC 60645-1:2017 RLV which contains the International Standard and its Redline version, showing all changes of the technical content compared to the previous edition.
IEC 60645-1:2017 specifies general requirements for audiometers designed for use in determining hearing threshold levels, relative to standard reference threshold levels established by means of psychoacoustic test methods, and those designed to perform psychoacoustic tests using speech material.
The object of this standard is to ensure:
a) that tests of hearing in the frequency range 125 Hz to 16 kHz on a given human ear, performed with different pure-tone audiometers which comply with this standard, give substantially the same results;
b) that the results obtained represent a valid comparison between the hearing of the ear tested and the reference threshold of hearing;
c) that a means of presenting speech material to a subject in a standardized manner is provided. This will ensure that tests of hearing using a specific speech signal and a specific manner of signal presentation, when performed with different audiometers which comply with this standard, give substantially the same results;
d) that audiometers are classified according to the range of test signals they present, according to the mode of operation or according to their presumed primary application.
This fourth edition cancels and replaces the third edition, published in 2012, and the first edition IEC 60645-2, published in 1993. This edition constitutes a technical revision.
This edition now includes the requirements for both pure-tone (prior edition of IEC 60645-1) and speech audiometers (prior edition of IEC 60645-2) into a single document. The technical requirements in this edition remain similar to the intent of the prior two documents, but now eliminate technical and editorial contradictions caused by two separate standards with different review cycles applying to an audiometer.

  • Standard
    86 pages
    English and French language
    sale 15% off

IEC 60645-1:2012 specifies general requirements for audiometers and particular requirements for pure-tone audiometers designed for use in determining hearing threshold levels, relative to standard reference threshold levels established by means of psychoacoustic test methods. The object of this standard is to ensure:
a) That tests of hearing in the frequency range 125 Hz to 16 000 Hz on a given human ear, performed with different audiometers which comply with this standard shall give substantially the same results;
b) That the results obtained represent a valid comparison between the hearing of the ear tested and the reference threshold of hearing;
c) That audiometers are classified according to the range of test signals they generate, according to the mode of operation or according to the complexity of the range of auditory functions they test.
This third edition cancels and replaces the second edition, published in 2001, and IEC 60645-4 published in 1994. It constitutes an editorial revision.

  • Standard
    68 pages
    English and French language
    sale 15% off

IEC 60645-7:2009 applies to instruments designed for the measurement of auditory evoked potentials from the inner ear, the auditory nerve and the brainstem, evoked by acoustic and/or vibratory stimuli of short duration. IEC 60645-7:2009 defines the characteristics to be specified by the manufacturer, specifies performance requirements for two types of instrument, screening and diagnostic, and specifies the functions to be provided on these types. The purpose of IEC 60645-7:2009 is to ensure that measurements made under comparable test conditions with different instruments complying with this standard will be consistent. IEC 60645-7:2009 is not intended to restrict development or incorporation of new features, nor to discourage innovative approaches. The application of electric stimuli for special purposes is beyond the scope of this standard.

  • Standard
    25 pages
    English and French language
    sale 15% off

IEC 60645-6:2009 applies to instruments designed primarily for the measurement of otoacoustic emissions in the human external acoustic meatus evoked by acoustic probe pulses or tones. This standard defines the characteristics to be specified by the manufacturer, lays down performance specifications for two types of instruments and specifies the functions to be provided on these types. IEC 60645-6:2009 describes methods of test to be used for approval testing and guidance on methods for undertaking routine calibration. The purpose of IEC 60645-6:2009 is to ensure that measurements made under comparable test conditions with different instruments complying with the standard will be consistent. Instruments which provide a measurement function not specifically within the scope of the standard shall still comply with any relevant requirements. This standard is not intended to restrict development or incorporation of new features, nor to discourage innovative approaches.

  • Standard
    32 pages
    English and French language
    sale 15% off

This part of IEC 60645 specifies a means of describing the physical characteristics of audiometric reference and test signals of short duration and methods for their measurement. The object of this standard is to ensure that audiometric stimuli of short duration are specified and measured in the same way and that the calibration of equipment using such signals is carried out using defined methods. This standard does not describe the method of use of short duration test signals. This second edition cancels and replaces the first edition published in 1994. This edition constitutes a technical revision. Specific changes in this edition concern new figures of reference signals and changes in definitions.

  • Standard
    25 pages
    English and French language
    sale 15% off

Applies to instruments designed primarily for the measurement of acoustic impedance/admittance in the human external acoustic meatus using a stated probe tone. It is recognized that other probe signals may also be used. The standard defines the characteristics to be specified by the manufacturer, lays down performance specifications for three types of instruments and specifies the facilities to be provided on these types. This standard describes methods of test to be used for approval testing and guidance on methods for undertaking routine calibration. The purpose of this standard is to ensure that measurements made under comparable test conditions with different instruments complying with the standard will be consistent. The standard is not intended to restrict development or incorporation of new features, nor to discourage innovative approaches. This first edition of IEC 60645-5 cancels and replaces the first edition of IEC 61027, published in 1991, and constitutes a technical revision.

  • Standard
    57 pages
    English and French language
    sale 15% off