IEC 61010-2-012:2019 is available as IEC 61010-2-012:2019 RLV which contains the International Standard and its Redline version, showing all changes of the technical content compared to the previous edition.
IEC 61010-2-012:2019 specifies safety requirements for electrical equipment and its accessories within the categories a) through c), wherever it is intended to be used, whenever that equipment incorporates one or more of the following characteristics:
– A refrigerating system that is acted on or impacted by an integral heating function such that the combined heating and refrigerating system generates additional and/or more severe hazards than those for the two systems if treated separately.
– The materials being treated in the intended application introduce significant heat into the refrigerating system, so that the refrigerating system in the application yields additional and/or more severe hazards than those for the refrigerating system if operated at the maximum rated ambient temperature alone.
– An irradiation function for the materials being treated presenting additional hazards.
– A function to expose the materials being treated to excessive humidity, carbon dioxide, salt mist, or other substances which can result in additional hazards.
– A function of mechanical movement presenting additional hazards.
– Provision for an operator to walk in to the operating area to load or unload the materials being treated.
It has the status of a group safety publication in accordance with IEC Guide 104.
This second edition cancels and replaces the first edition published in 2016. This edition constitutes a technical revision.
This edition includes the following significant technical changes with respect to the previous edition:
a) alignment with changes introduced by Amendment 1 of IEC 61010-1:2010;
b) changes related to the use of small capitals for defined terms only;
c) clarifications for cooling tests in 4.4.2.10;
d) requirements for overtemperature protection in 10.101, including deletion of the second part of the sentence in item b), and the deletion of item c);
e) changes pertaining to the accurate employment of terms "temperature", "operating temperature", "working temperature", "application temperature", "room temperature" and "ambient temperature" in 3.5.104, 3.5.105, 4.3.1, 4.3.2, 5.4.2, 8.2.1, 8.2.2, 11.7.2.101.2, 11.7.2.101.3, 13.2.102, 14.102, 15.101, 15.102, 15.103, Introduction and many other locations. For the purpose of clarification, the definition of 3.5.114, controlled temperature, is added.

  • Standard
    177 pages
    English and French language
    sale 15% off

IEC 61010-2-011:2019 is available as IEC 61010-2-011:2019 RLV which contains the International Standard and its Redline version, showing all changes of the technical content compared to the previous edition.
IEC 61010-2-011:2019 specifies particular safety requirements for the following types a) to c) of electrical equipment and their accessories, wherever they are intended to be used, whenever that equipment incorporates refrigerating systems as an integral part of, or separate from, the equipment and the equipment is in direct control of the refrigerating system. This document details all the requirements when up to 150 g of flammable refrigerant are used per stage of a refrigerating system. Additional requirements beyond the current scope of this document apply if a refrigerant charge of flammable refrigerant exceeds this amount.
a) Electrical test and measurement equipment
This is equipment which by electromagnetic means tests, measures, indicates or records one or more electrical or physical quantities, also non-measuring equipment such as signal generators, measurement standards, power supplies for laboratory use, transducers, transmitters, etc.
b) Electrical industrial process-control equipment
This is equipment which controls one or more output quantities to specific values, with each value determined by manual setting, by local or remote programming, or by one or more input variables.
c) Electrical laboratory equipment
This is equipment which measures, indicates, monitors, inspects or analyses materials, or is used to prepare materials, and includes in vitro diagnostic (IVD) equipment. This second edition cancels and replaces the first edition published in 2016. This edition constitutes a technical revision. This edition includes the following significant technical changes with respect to the previous edition:
a) alignment with changes introduced by Amendment 1 of IEC 61010-1:2010;
b) introduction of new defined terms or modified terms to align with Part 2-012 and other source documents. Editorial changes to use small capitals only for defined terms. Note the difference of defined term in 4.3.2.101 and abnormal operation in 11.7.104.3 and 11.7.104.5;
c) clarifications for cooling tests in 4.4.2.10;
d) changes pertaining to the accurate employment of the following terms: temperature, operating temperature, application temperature, controlled temperature, room ambient and ambient temperature;
e) use of defined term to replace cooling system;
f) move text of 4.4.2.101 to 4.3.2.101, since the purpose of as defined, is to simulate failure of the ambient conditions of 1.4.1 but not of the of the equipment;
g) use of the term equipment to replace unit, apparatus, appliance, where applicable;
h) in 5.1.2 dd) for high and low sides for each stage are required only under ;
i) use of defined term to replace normal operation;
j) use of defined term to replace user.
It has the status of a group safety publication in accordance with IEC Guide 104.

  • Standard
    103 pages
    English and French language
    sale 15% off

IEC 61010-2-010:2019 is available as IEC 61010-2-010:2019 RLV which contains the International Standard and its Redline version, showing all changes of the technical content compared to the previous edition.
IEC 61010-2-010:2019 specifies particular safety requirements for the following types a) to c) of electrical equipment and their accessories, wherever they are intended to be used, whenever the heating of materials is one of the functions of the equipment.
a) Electrical test and measurement equipment
This is equipment which by electromagnetic means tests, measures, indicates or records
one or more electrical or physical quantities, also non-measuring equipment such as
signal generators, measurement standards, power supplies for laboratory use,
transducers, transmitters, etc.
b) Electrical industrial process-control equipment
This is equipment which controls one or more output quantities to specific values, with
each value determined by manual setting, by local or remote programming, or by one or
more input variables.
c) Electrical laboratory equipment
This is equipment which measures, indicates, monitors, inspects or analyses materials, or
is used to prepare materials, and includes in vitro diagnostic (IVD) equipment. This fourth edition cancels and replaces the third edition published in 2014. This edition constitutes a technical revision. This edition includes the following significant technical changes with respect to the previous edition:
a) alignment with changes introduced by Amendment 1 of IEC 61010-1:2010;
b) alignment with IEC 61010-2-011 and IEC 61010-2-012:
- new matching Introduction clarifying which standard(s) to use;
- new 5.4.101 instructions for flammable liquid;
- subclause 9.5 on flammable liquids replaced with text from IEC 61010-2-012;
c) subclause 5.2.101 deleted;
d) requirements in 10.101 b) and c) clarified.
It has the status of a group safety publication in accordance with IEC Guide 104.