ASTM D4327-03
(Test Method)Standard Test Method for Anions in Water by Chemically Suppressed Ion Chromatography
Standard Test Method for Anions in Water by Chemically Suppressed Ion Chromatography
SIGNIFICANCE AND USE
Ion chromatography provides for both qualitative and quantitative determination of seven common anions, F−, Cl−, NO2−, HPO4−2, Br−, NO3−, and SO4−2, in the milligram per litre range from a single analytical operation requiring only a few millilitres of sample and taking approximately 10 to 15 min for completion.
Note 2—This test method may be used to determine fluoride if its peak is in the water dip by adding one mL of eluent (at 100× the concentration in 8.3) to all 100-mL volumes of samples and standards to negate the effect of the water dip. (See 6.3, and also see 6.4.) The quantitation of unretained peaks should be avoided. Anions such as low molecular weight organic acids (formate, acetate, propionate, etc.) that are conductive coelute with fluoride and would bias fluoride quantitation in some drinking waters and most wastewaters.
Anion combinations such as Cl−/Br− and NO2−/NO3−, which may be difficult to distinguish by other analytical methods, are readily separated by ion chromatography.
SCOPE
1.1 This test method covers the sequential determination of fluoride, chloride, nitrite, ortho-phosphate, bromide, nitrate, and sulfate ions in water by chemically suppressed ion chromatography. Note 1Order of elution is dependent upon the column used; see .
1.2 This test method is applicable to drinking and wastewaters. The ranges tested for this test method for each anion were as follows (measured in mg/L):Fluoride0.26 to 8.49Chloride0.78 to 26.0Nitrite-N0.36 to 12.0Bromide0.63 to 21.0Nitrate-N0.42 to 14.0o-Phosphate0.69 to 23.1Sulfate2.85 to 95.0
1.3 It is the user's responsibility to ensure the validity of this test method for other matrices.
1.4 Concentrations as low as 0.01 mg/L were determined depending upon the anions to be quantitated, in single laboratory work. Utilizing a 50-μL sample volume loop and a sensitivity of 3 μS/cm full scale, the approximate detection limits shown in Table 1 can be achieved. If lower detection levels are required, the sensitivity may be improved by using a lower scale setting (100 μL). The analyst must assure optimum instrument performance to maintain a stable baseline at more sensitive conductivity full-scale settings.
1.5 The upper limit of this test method is dependent upon total anion concentration and may be determined experimentally as described in Annex A1. These limits may be extended by appropriate dilution or by use of a smaller injection volume.
1.6 Using alternate separator column and eluents may permit additional anions such as formate or citrate to be determined. This is not the subject of this test method.
1.7 This standard does not purport to address all of the safety problems, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use.
General Information
Relations
Standards Content (Sample)
NOTICE: This standard has either been superseded and replaced by a new version or withdrawn.
Contact ASTM International (www.astm.org) for the latest information
Designation:D4327–03
Standard Test Method for
Anions in Water by Chemically Suppressed Ion
1
Chromatography
This standard is issued under the fixed designation D4327; the number immediately following the designation indicates the year of
original adoption or, in the case of revision, the year of last revision.Anumber in parentheses indicates the year of last reapproval.A
superscript epsilon (´) indicates an editorial change since the last revision or reapproval.
1. Scope*
2
1.1 Thistestmethod coversthesequentialdeterminationof
fluoride, chloride, nitrite, ortho-phosphate, bromide, nitrate,
and sulfate ions in water by chemically suppressed ion chro-
matography.
NOTE 1—Order of elution is dependent upon the column used; see Fig.
1.
1.2 This test method is applicable to drinking and wastewa-
ters.Therangestestedforthistestmethodforeachanionwere
as follows (measured in mg/L):
FIG. 1 Chromatogram Showing Separation Using the AS4A
Fluoride 0.26 to 8.49
Column
Chloride 0.78 to 26.0
Nitrite-N 0.36 to 12.0
Bromide 0.63 to 21.0
sensitivity of 3 µS/cm full scale, the approximate detection
Nitrate-N 0.42 to 14.0
limits shown in Table 1 can be achieved. If lower detection
o-Phosphate 0.69 to 23.1
levels are required, the sensitivity may be improved by using a
Sulfate 2.85 to 95.0
lowerscalesetting(<3µS/cm)oralargersampleinjectionloop
1.3 Itistheuser’sresponsibilitytoensurethevalidityofthis
(>100 µL). The analyst must assure optimum instrument
test method for other matrices.
performance to maintain a stable baseline at more sensitive
1.4 Concentrations as low as 0.01 mg/L were determined
conductivity full-scale settings.
depending upon the anions to be quantitated, in single labora-
1.5 The upper limit of this test method is dependent upon
tory work. Utilizing a 50-µL sample volume loop and a
total anion concentration and may be determined experimen-
tally as described inAnnexA1. These limits may be extended
byappropriatedilutionorbyuseofasmallerinjectionvolume.
1
This test method is under the jurisdiction ofASTM Committee D19 on Water
1.6 Using alternate separator column and eluents may per-
and is the direct responsibility of Subcommittee D19.05 on Inorganic Constituents
mit additional anions such as formate or citrate to be deter-
in Water.
Current edition approved Jan. 10, 2003. Published January 2003. Originally
mined. This is not the subject of this test method.
approved in 1984. Last previous edition approved in 1997 as D4327–97. DOI:
1.7 This standard does not purport to address all of the
10.1520/D4327-03.
2 safety problems, if any, associated with its use. It is the
The following references may be consulted for additional information:
responsibility of the user of this standard to establish appro-
Small, H., Stevens, T. S., and Bauman, W. C., “Novel Ion Exchange Chromato-
graphic Method Using Conductrimetric Detection,” Analytical Chemistry, Vol 47,
priate safety and health practices and determine the applica-
1975, p. 1801.
bility of regulatory limitations prior to use.
Stevens, T. S., Turkelson, V. T., and Alve, W. R., “Determination of Anions in
BoilerBlowDownWaterwithIonChromatography,” Analytical Chemistry,Vol49,
2. Referenced Documents
1977, p. 1176.
Sawicki, E., Mulik, J. D., and Witgenstein, E., Editors, Ion Chromatographic 3
2.1 ASTM Standards:
Analysis of Environmental Pollutants, Ann Arbor Science Publishers, Ann Arbor,
D1066 Practice for Sampling Steam
MI, 1978.
D1129 Terminology Relating to Water
Mulik, J. D., and Sawicki, E., Editors, Ion Chromatographic Analysis of
Environmental Pollutants,Vol/No.2,AnnArborSciencePublishers,AnnArbor,MI,
D1193 Specification for Reagent Water
1979.
Weiss, J., Handbook of Ion Chromatography, Dionex Corp., Sunnyvale, CA,
1986.
3
Waters Innovative Methods forAnionAnalysis,WatersChromatographyDivision
For referenced ASTM standards, visit the ASTM website, www.astm.org, or
of Millipore, Method A 107 and A 116, 1990.
contact ASTM Customer Service at service@astm.org. For Annual Book of ASTM
Haddad, P. R., and Jackson, P. E., Ion Chromatography: Principles and
Standards volume information, refer to the standard’s Document Summary page on
Applications, Elsevier Scientific Publishing Co., 1990.
the ASTM website.
*A Summary of Changes section appears at the end of this standard.
Copyright © ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959, United States.
1
---------------------- Page: 1 ----------------------
D4327–03
TABLE 1 Approximate Single Laboratory Detection Limits in
that is continuously regenerated by a flow of dilute sulfuric
A,B
Reagent Water
acid. The suppressor device reduces the background conduc-
Retention MDL
tivity of the eluent to a low or negligible level by replacing the
Analyte Peak No.
Time, min mg/L
cationswiththehydrogenion,therebyconvertingtheanionsin
...
Questions, Comments and Discussion
Ask us and Technical Secretary will try to provide an answer. You can facilitate discussion about the standard in here.