ASTM F2914-12(2024)
(Guide)Standard Guide for Identification of Shelf-Life Test Attributes for Endovascular Devices
Standard Guide for Identification of Shelf-Life Test Attributes for Endovascular Devices
SIGNIFICANCE AND USE
3.1 The purpose of this guide is to provide a procedure for determining the appropriate attributes to evaluate in a shelf-life study for an endovascular device.
SCOPE
1.1 This guide addresses the determination of appropriate device attributes for testing as part of a shelf-life study for endovascular devices. Combination and biodegradable devices (for example, drug devices, biologic devices, or drug biologics) may require additional considerations, depending on their nature.
1.2 This guide does not directly provide any test methods for conducting shelf-life testing.
1.3 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety, health, and environmental practices and determine the applicability of regulatory limitations prior to use.
1.4 This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.
General Information
Relations
Standards Content (Sample)
This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the
Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.
Designation: F2914 − 12 (Reapproved 2024)
Standard Guide for
Identification of Shelf-Life Test Attributes for Endovascular
Devices
This standard is issued under the fixed designation F2914; the number immediately following the designation indicates the year of
original adoption or, in the case of revision, the year of last revision. A number in parentheses indicates the year of last reapproval. A
superscript epsilon (´) indicates an editorial change since the last revision or reapproval.
1. Scope 4. Procedure
1.1 This guide addresses the determination of appropriate
4.1 Shelf-Life Establishment Model Introduction—The deci-
device attributes for testing as part of a shelf-life study for
sion flow chart (Fig. 1) assists study developers in selecting
endovascular devices. Combination and biodegradable devices
and justifying risk-appropriate test protocols for medical de-
(for example, drug devices, biologic devices, or drug biologics)
vices to establish shelf life. The decision flow chart is intended
may require additional considerations, depending on their
to elicit questions and an appropriate rationale for testing or not
nature.
testing a particular attribute during aging. The risk to the
patient as the device ages is one of the primary drivers. It is
1.2 This guide does not directly provide any test methods
for conducting shelf-life testing. recommended that all regulatory requirements and guidances
be considered during development of the shelf-life establish-
1.3 This standard does not purport to address all of the
ment test plan. See Fig. 1.
safety concerns, if any, associated with its use. It is the
responsibility of the user of this standard to establish appro-
4.2 Question 1: Could the device attribute change over
priate safety, health, and environmental practices and deter-
time?
mine the applicability of regulatory limitations prior to use.
4.2.1 Considerations in Evaluating Question 1—This ques-
1.4 This international standard was developed in accor-
tion must be addressed based on the device design character-
dance with internationally recognized principles on standard-
istics (and also in relation to the device being packaged,
ization established in the Decision on Principles for the
sterilized, shipped, and stored).
Development of International Standards, Guides and Recom-
4.2.1.1 Consider attributes such as the following, for ex-
mendations issued by the World Trade Organization Technical
ample:
Barriers to Trade (TBT) Committee.
(1) Material Properties/Characterization—Composition;
2. Terminology
Mechanical Properties; Corrosion Resistance
(2) Dimensional and Functional Properties—Dimensions;
2.1 Definitions:
Surface Area; Foreshortening
2.1.1 endovascular device—device used to treat vascular
(3) Deliverability and Functionality—Balloon Fatigue;
disease from within the vessel.
Balloon Rated Burst; Bond Tensile Strength
2.1.2 product—final packaged and sterilized device with all
4.2.1.2 Various sources may provide sufficient evidence to
included components.
confirm that some specific attributes do not change over time
2.1.3 shelf life—the amount of real time that a fully pack-
for the application or that the change is not a risk to the patient.
aged (and sterilized, if applicable) product can be expected to
(1) Scientific literature.
remain in storage at specified conditions and maintain its
(2) Appropriate vendor publication.
critical performance properties.
(3) In-house research.
3. Significance and Use (4) Assessment of clinically accepted device.
4.2.1.3 When using such data to justify why certain attri-
3.1 The purpose of this guide is to provide a procedure for
butes may not require shelf-life testing, consider all differences
determining the appropriate attributes to evaluate in a shelf-life
between the subject device and the source of those data to
study for an endovascular device.
ensure applicability. For example, vendor literature may not
represent the actual use of the material by the device manu-
This guide is under the jurisdiction of ASTM Committee F04 on Medical and
Surgical Materials and Devices and is the direct responsibility of Subcommittee
facturer. Additionally, further processing (for example, steril-
F04.30 on Cardiovascular Standards.
ization) may change the physical or chemical attribute(s) of the
Current edition approved Jan. 15, 2024. Published January 2024. Originally
material. Finally, consider whether there are interactions
approved in 2012. Last previous edition approved in 2018 as F2914 – 12 (2018).
DOI:10.1520/F2914-12R24. (chemical or physical) that may impact your assessment.
Copyright © ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959. United States
F2914 − 12 (2024)
FIG. 1 Device Aging Shelf-Life Establishment Flow Chart
F2914 − 12 (2024)
4.2.1.4 In order for testing to be applicable, the testing must way of stating the question is: “Will a change in the device
be conducted on articles that are representative of the final attribute resulting from aging pose a significant risk to the
device (that is, utilizing the same sterilization method and dose, patient or clinician?” Risk analysis is an appropriate technique
dimensions, material, processing conditions, and packaging). If used to answer this question. However, since risk analysis
test articles are not identical, provide appropriate justification methodologies have yet to be standardized, there is no defini-
for applicability of the testing. tive risk level that can be applied universally for all devices and
4.2.2 Justification Based upon Scientific Principles—When parameters. It will be the responsibility of individual compa-
one considers whether an attribute should be included in a nies to carefully develop the threshold for acceptable risk.
shelf-life study, the first question is whether the attribute 4.3.1 Basis for Risk Assessment—The assessment of risk
changes over time. There are several device attributes that may related to a device attribute may be conducted using clinical
be driven by physical parameters of the device that would not history (in literature or privately held) or the complaint history
change over time and therefore will not require shelf-life of a similar device used in a similar application. Additionally,
testing. The assessment should be conducted using universal a scientific/medical argument might provide adequate informa-
scientific/physical principles. In cases where the assessment is tion to assess the risk.
based on universal scientific/physical principle, appropriate 4.3.2 Risk Assessment Examples—The following examples
references should be provided. In cases where justifications of risk assessment of selected attributes are for illustrative
may be less obvious, data to support the scientific/physical purposes only; this guide cannot claim to address all
rationale shall be generated. Tables 1 and 2 list two groups of circumstances, and thus these examples should not be used to
device attributes with accompanying scientific rationale. overly influence a company’s policies. When not expected to
4.2.3 Justification Based upon Data—Scientific principles impact safety or performance, the scientific justification shall
for some device attributes/requirements are not readily evident. be documented in detail.
In such cases, one may generate data to support a rationale. It
5. Shelf-Life Establishment Report
may be advantageous to conduct testing in a manner that
allows for the data to be applicable to various size devices. In
5.1 The report shall include a complete device description,
this case, it is important to translate the device attribute (such
assumptions for device storage, and the device attributes
as system flexibility) into the underlying size independent
considered for testing in conducting a device aging shelf-life
scientific parameters (such as Young’s modulus). Testing is
establishment study. The decision to conduct testing or not for
then conducted to evaluate the stability of the core scientific
each device attribute shall be reported. The rationale for why
parameter. For each device attribute, more than one scientific
testing of a specific device attribute was determined to not be
parameter may be necessary to demonstrate stability over the
necessary (answered “no” to Question 1 or 2) shall be reported.
aging period. (For simplicity of the examples, only one test
The reported rationale shall provide sufficient detail to con-
parameter is illustrated in Table 3.) Each device attribute
vince a person with adequate engineering/scientific experience.
should be evaluated to determine what scientific parameters
References supporting rationale to not conduct testing should
may be affected by aging, and the appropriate testing to
be provided as appropriate. When testing of a specific device
mitigate each of those risks should then be conducted. The
attribute was determined to be neces
...
Questions, Comments and Discussion
Ask us and Technical Secretary will try to provide an answer. You can facilitate discussion about the standard in here.