Standard Test Method for Characterizing the Effect of Exposure to Environmental Cycling on Thermal Performance of Insulation Products

SIGNIFICANCE AND USE
5.1 Exposing a specimen to conditions of one-directional environmental cycling can increase its moisture content until a decrease in material properties occurs (at a specific number of cycles). Such a test could be inappropriate due to the number of cycles required to cause a decrease in material properties since product performance issues often arise only after many years of exposure. The use of a preconditioning procedure is not intended to duplicate expected field performance. Rather the purpose is to increase the moisture content of test materials prior to subjecting to them to environmental cycling.  
5.2 The most important aspect of the preconditioning procedure is non-uniform moisture distribution in the specimen. The heat flow is one directional causing moisture flow towards the cold side resulting in zones of dry material on the warm side and high moisture content on the cold side. (Whether the high moisture content zone is located right at the cold surface of the specimen or at some distance from this surface depends upon temperature oscillation and ability of the cold surface to dry outwards). Because the preconditioning procedure involves thermal gradient, this preconditioning procedure results in a distribution of moisture content that may occur under field exposure conditions. However, the resulting moisture content may differ significantly from that which may be demonstrated in typical product applications.  
5.3 The preconditioning results in accumulation of moisture in the thermal insulation resulting from the simultaneous exposure to a difference in temperature and water vapor pressure. This test method is not intended to duplicate field exposure. It is intended to provide comparative ratings. As excessive accumulation of moisture in a construction system may adversely affect its performance, the designer should consider the potential for moisture accumulation and the possible effects of this moisture on the system performance.
SCOPE
1.1 This test method is applicable to preformed or field manufactured thermal insulation products, such as board stock foams, rigid fibrous and composite materials manufactured with or without protective facings. See Note 1 This test method is not applicable to high temperature, reflective or loose fill insulation.  
Note 1: If the product is manufactured with a facer, test product with facer in place.  
1.2 This test method involves two stages: preconditioning and environmental cycling. During the first stage, 25 mm (1 in.) thick specimens are used to separate two environments. Each of these environments has a constant but different temperature and humidity level. During the environmental cycling stage, specimens also divide two environments namely constant room temperature/humidity on one side and cycling temperature/ambient relative humidity on the other side.  
1.3 This test method measures the ability of the product to maintain thermal performance and critical physical attributes after being subjected to standardized exposure conditions. A comparison is made between material properties for reference specimens stored in the laboratory for the test period and specimens subjected to the two-stage test method. To eliminate the effect of moisture from the comparison, the material properties of the latter test specimens are determined after they have been dried to constant weight. The average value determined for each of the two sets of specimens is used for comparison.  
1.4 Different properties can be measured to assess the effect of environmental factors on thermal insulation. This test method requires that thermal resistance be determined based upon an average for three specimens measured after completing the test. Secondary elements of this test method include visual observations such as cracking, delamination or other surface defects, as well as the change in moisture content after each of the two stages of exposure prescribed by the...

General Information

Status
Historical
Publication Date
14-May-2015
Current Stage
Ref Project

Relations

Buy Standard

Standard
ASTM C1512-10(2015) - Standard Test Method for Characterizing the Effect of Exposure to Environmental Cycling on Thermal Performance of Insulation Products
English language
7 pages
sale 15% off
Preview
sale 15% off
Preview
Standard
REDLINE ASTM C1512-10(2015) - Standard Test Method for Characterizing the Effect of Exposure to Environmental Cycling on Thermal Performance of Insulation Products
English language
7 pages
sale 15% off
Preview
sale 15% off
Preview

Standards Content (Sample)


NOTICE: This standard has either been superseded and replaced by a new version or withdrawn.
Contact ASTM International (www.astm.org) for the latest information
Designation: C1512 − 10(Reapproved 2015)
Standard Test Method for
Characterizing the Effect of Exposure to Environmental
Cycling on Thermal Performance of Insulation Products
This standard is issued under the fixed designation C1512; the number immediately following the designation indicates the year of
original adoption or, in the case of revision, the year of last revision. A number in parentheses indicates the year of last reapproval. A
superscript epsilon (´) indicates an editorial change since the last revision or reapproval.
1. Scope surface defects, as well as the change in moisture content after
each of the two stages of exposure prescribed by the test
1.1 This test method is applicable to preformed or field
method.
manufactured thermal insulation products, such as board stock
1.5 Characterization of the tested material is an essential
foams, rigid fibrous and composite materials manufactured
with or without protective facings. See Note 1This test method element of this test method. Material properties used for
is not applicable to high temperature, reflective or loose fill characterization will include either compressive resistance or
insulation. tensile strength values. The compressive resistance or tensile
strength is measured on two sets of specimens, one set
NOTE 1—If the product is manufactured with a facer, test product with
conditioned as defined in 1.2 and a set of reference test
facer in place.
specimenstakenfromthesamematerialbatchandstoredinthe
1.2 This test method involves two stages: preconditioning
laboratory for the whole test period. For comparison, an
and environmental cycling. During the first stage, 25 mm (1
average value is determined for each of the two sets of
in.) thick specimens are used to separate two environments.
specimens.
Each of these environments has a constant but different
1.6 The values stated in SI units are to be regarded as
temperature and humidity level. During the environmental
standard. No other units of measurement are included in this
cycling stage, specimens also divide two environments namely
standard.
constant room temperature/humidity on one side and cycling
temperature/ambient relative humidity on the other side.
1.7 This standard does not purport to address all of the
safety concerns, if any, associated with its use. It is the
1.3 This test method measures the ability of the product to
responsibility of the user of this standard to establish appro-
maintain thermal performance and critical physical attributes
priate safety and health practices and determine the applica-
after being subjected to standardized exposure conditions. A
bility of regulatory requirements prior to use.
comparison is made between material properties for reference
specimens stored in the laboratory for the test period and
2. Referenced Documents
specimens subjected to the two-stage test method.To eliminate
the effect of moisture from the comparison, the material
2.1 ASTM Standards:
properties of the latter test specimens are determined after they C165 Test Method for Measuring Compressive Properties of
have been dried to constant weight. The average value deter-
Thermal Insulations
mined for each of the two sets of specimens is used for C168 Terminology Relating to Thermal Insulation
comparison.
C177 Test Method for Steady-State Heat Flux Measure-
ments and Thermal Transmission Properties by Means of
1.4 Different properties can be measured to assess the effect
the Guarded-Hot-Plate Apparatus
of environmental factors on thermal insulation. This test
C303 Test Method for Dimensions and Density of Pre-
method requires that thermal resistance be determined based
formed Block and Board–Type Thermal Insulation
upon an average for three specimens measured after complet-
C518 Test Method for Steady-State Thermal Transmission
ing the test. Secondary elements of this test method include
Properties by Means of the Heat Flow Meter Apparatus
visual observations such as cracking, delamination or other
C870 Practice for Conditioning of Thermal Insulating Ma-
terials
ThistestmethodisunderthejurisdictionofASTMCommitteeC16onThermal
Insulation and is the direct responsibility of Subcommittee C16.33 on Insulation
Finishes and Moisture. For referenced ASTM standards, visit the ASTM website, www.astm.org, or
Current edition approved May 15, 2015. Published September 2015. Originally contact ASTM Customer Service at service@astm.org. For Annual Book of ASTM
approved in 2001. Last previous edition approved in 2010 as C1512–10. DOI: Standards volume information, refer to the standard’s Document Summary page on
10.1520/C1512-10R15. the ASTM website.
Copyright © ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959. United States
C1512 − 10 (2015)
C618 Specification for Coal Fly Ash and Raw or Calcined dryoutwards).Becausethepreconditioningprocedureinvolves
Natural Pozzolan for Use in Concrete thermal gradient, this preconditioning procedure results in a
D1621 Test Method for Compressive Properties of Rigid distribution of moisture content that may occur under field
Cellular Plastics exposure conditions. However, the resulting moisture content
D1623 Test Method for Tensile and Tensile Adhesion Prop- may differ significantly from that which may be demonstrated
erties of Rigid Cellular Plastics in typical product applications.
E177 Practice for Use of the Terms Precision and Bias in
5.3 The preconditioning results in accumulation of moisture
ASTM Test Methods
in the thermal insulation resulting from the simultaneous
E691 Practice for Conducting an Interlaboratory Study to
exposure to a difference in temperature and water vapor
Determine the Precision of a Test Method
pressure. This test method is not intended to duplicate field
exposure. It is intended to provide comparative ratings. As
3. Terminology
excessive accumulation of moisture in a construction system
3.1 Definitions—Terms used in this test method are defined
may adversely affect its performance, the designer should
in Terminology C168 with the exceptions included as appro-
consider the potential for moisture accumulation and the
priate.
possible effects of this moisture on the system performance.
3.2 Definitions of Terms Specific to This Standard:
3.2.1 compressive resistance—thecompressiveloadperunit
6. Apparatus
of original area at the specified deformation. See Test Method
6.1 The room where the apparatus is placed shall be
C165.
maintained at a temperature and relative humidity of 24 6 3°C
3.2.2 moisture accumulation—an increase in the average
(75 6 5°F) and 50 6 10 %.
moisture content resulting from a specified exposure to condi-
6.2 Freeze-Thaw Chamber, capable of maintaining an air
tions facilitating moisture ingress into the material.
temperature of -15 6 3°C (5 6 5°F) over an extended period
3.2.3 preconditioning—a procedure which subjects test
of time. The design of the apparatus should ensure that the
specimens to standardized one directional thermal gradient.
temperature of the upper surface of the sheet metal located
3.2.4 thermal performance—comparison of thermal resis-
below the insulation specimen (measured in the center of the
tance of test specimens before and after cycling.
pan) be not higher than -4°C (25°F) when the freezer’s air
temperature reaches its lower limit. This can be achieved by
4. Summary of Test Method
placing thermal insulation between the metal pan and the
4.1 Toreducethetestingperiod,thisprocedureinvolvestwo
specimen frame and/or mixing of air in the cold chamber.
stages:
6.3 Sheet Metal Pan, placed below the specimens. This pan
4.1.1 Stage 1—Preconditioning under constant thermal gra-
performs two functions: it equalizes temperature and reduces
dient and relative humidity to accelerate ingress of moisture
diffusion of water vapor into the freeze-thaw chamber. The
into the test specimen.
distancebetweenthecoldsurfaceofthespecimenandthesheet
4.1.2 Stage 2—Exposure to constant temperature and rela-
metal should be no less than 6 mm ( ⁄4 in.) and no more than
tive humidity on one side of test specimens with cycling
12 mm ( ⁄2 in). The required space is normally maintained by
environmental conditions on the other side that include freeze-
attaching a support of the required height that is made from
thaw exposure.
6-mm ( ⁄4 in.) thick Plexiglas or other non-absorbing materials
on the inside surface of the specimen frame (see Fig. 2).
5. Significance and Use
6.4 Frame, that is placed in the door opening of the freezer
5.1 Exposing a specimen to conditions of one-directional
environmental cycling can increase its moisture content until a (see Figs. 1 and 2) or other means of specimen support. Test
frames used are made from 6 6 0.5 mm thick Plexiglas or
decrease in material properties occurs (at a specific number of
other non-absorbing material. These frames are used to mount
cycles). Such a test could be inappropriate due to the number
individual test specimens. The selection of the test frame (size
of cycles required to cause a decrease in material properties
of the test specimen) may vary based upon the thermal testing
since product performance issues often arise only after many
apparatus that is used.
years of exposure. The use of a preconditioning procedure is
not intended to duplicate expected field performance. Rather
6.5 Warm Chamber, above the test specimens that is pro-
the purpose is to increase the moisture content of test materials
vided with a heater and a temperature controller capable of
prior to subjecting to them to environmental cycling.
maintaining a temperature of 24 6 2°C (75 6 3°F) and a
5.2 The most important aspect of the preconditioning pro- humidifier capable of maintaining humidity in the warm
cedure is non-uniform moisture distribution in the specimen.
chamber of 90 6 5 %RH.
The heat flow is one directional causing moisture flow towards
6.6 Sensors, for measuring temperature of the freeze-thaw
the cold side resulting in zones of dry material on the warm
and warm chambers and relative humidity in the warm
side and high moisture content on the cold side. (Whether the
chamber.
high moisture content zone is located right at the cold surface
of the specimen or at some distance from this surface depends 6.7 Balance, capable of weighing mass of maximum 1 kg
upon temperature oscillation and ability of the cold surface to with precision of 0.01 g.
C1512 − 10 (2015)
FIG. 1 Plan View of Test Equipment Setup
FIG. 2 Vertical Section at Interface Between Freezer Wall and Lid Illustrating Placement of Test Specimens in the Test Frame
7. Test Specimens specimens are tested after preconditioning and after environ-
mental cycling as described in Section 9. A second set of
7.1 Test specimens shall be square in cross-section with a
2 2 reference test specimens are stored in the laboratory for the
minimum area of 645 cm (100 in. ) and a maximum of 3716
2 2 duration of preconditioning and environmental cycling test
cm (576 in. ). The standard specimen thickness shall be 2.54
before thermal resistance and compressive resistance or tensile
cm (1 in.). Care should be taken so that the top and bottom
strength testing.
surfaces of the specimens exposed to thermal gradient are
parallel with one another and perpendicular to the sides.
8. Conditioning
7.2 All surfaces of the specimens shall be free from visible
8.1 Condition the test specimens before testing at 23 6 2°C
flaws or imperfections.
(73 64°F)and50 65 %RHrelativehumidityfornotlessthan
7.3 For comparison, two test specimen sets each consisting 40 h prior to test in accordance with Procedure A of Practice
of a minimum of three specimens are tested. One set of test C618.
C1512 − 10 (2015)
9. Procedure 9.4.3 Weigh each specimen after completion of environ-
mental cycling and calculate moisture content (% by volume).
9.1 Condition specimens to constant mass in accordance
Condition specimens to constant mass in accordance with 9.1
with Practice C870 before testing. Measure the dimensions and
and subject to testing in accordance with 9.2.
mass of each specimen in accordance with Test Method C303.
Record the initial mass of each specimen prior to subjecting to
10. Report
preconditioning procedure.
10.1 The test report shall include the following information,
9.2 Testing of Specimens Before and After Environmental
including references to applicable test methods:
Cycling:
10.1.1 The date of the report.
9.2.1 Three specimens shall be tested for thermal resistance
10.1.2 The name, address and identification of the testing
value before and after environmental cycling using Test
laboratory.
Method C518 or C177.
10.1.3 The manufacturer of the material, the date of manu-
9.2.2 Where applicable, nine specimens shall be tested for
facture and the date of receiving samples.
compressive resistance before and after environmental cycling
10.1.4 Number of samples received and the number of
using Test Method C165 or D1621.
specimens tested in respective categories.
9.2.3 Where applicable, nine specimens shall be tested for
10.1.5 The name or identification of the material tested and
tensile strength before and after environmental cycling using
description of facers (if any).
Test Method D1623.
10.1.6 The method of specimen preparation.
9.3 Preconditioning:
10.1.7 The type and size of the preconditioning set-up and
9.3.1 Test specimens are preconditioned for 28 days to
the preconditioning conditions.
increase moisture content.This is achieved under conditions of
10.1.8 The moisture content (% by volume) of each test
water vapor diffusion associated with a constant thermal
specimen after preconditioning and cycling.
gradient. The specimens are dividing two environments,
10.1.9 Average and standard deviation of these values at the
namely:
end of preconditioning stage.
9.3.1.1 Temperature of 24 6 2°C (75 6 3°F) and relative
10.1.10 The method of sealing around the test specimen.
humidity of 90 6 5 % on warm side, and
10.1.11 Averageofthetestconditionssuchasminimumand
9.3.1.2 Temperature of -15 6 3°C (5 6 5°F) and ambient
maximum temperatures in the freezing cabinet, the difference
relative humidity (uncontrolled relative humidity) on the cold
in temperature of air in the freezing cabinet and the surface of
side.
the sheet metal facing test specimens.
9.3.2 If the specimens are provided with facing, stucco
10.1.12 The moisture content, in kg/m for each test speci-
lamina or other protective finishes, these finishes should be
men and the average and standard deviation of these values at
placed on the cold side during the preconditioning exposure.
the end of the testing stage.
9.3.3 Weigh each specimen after initial preconditioning.
...


This document is not an ASTM standard and is intended only to provide the user of an ASTM standard an indication of what changes have been made to the previous version. Because
it may not be technically possible to adequately depict all changes accurately, ASTM recommends that users consult prior editions as appropriate. In all cases only the current version
of the standard as published by ASTM is to be considered the official document.
Designation: C1512 − 10 C1512 − 10 (Reapproved 2015)
Standard Test Method for
Characterizing the Effect of Exposure to Environmental
Cycling on Thermal Performance of Insulation Products
This standard is issued under the fixed designation C1512; the number immediately following the designation indicates the year of
original adoption or, in the case of revision, the year of last revision. A number in parentheses indicates the year of last reapproval. A
superscript epsilon (´) indicates an editorial change since the last revision or reapproval.
1. Scope
1.1 This test method is applicable to preformed or field manufactured thermal insulation products, such as board stock foams,
rigid fibrous and composite materials manufactured with or without protective facings. See Note 1 This test method is not
applicable to high temperature, reflective or loose fill insulation.
NOTE 1—If the product is manufactured with a facer, test product with facer in place.
1.2 This test method involves two stages: preconditioning and environmental cycling. During the first stage, 25 mm (1 in.) thick
specimens are used to separate two environments. Each of these environments has a constant but different temperature and
humidity level. During the environmental cycling stage, specimens also divide two environments namely constant room
temperature/humidity on one side and cycling temperature/ambient relative humidity on the other side.
1.3 This test method measures the ability of the product to maintain thermal performance and critical physical attributes after
being subjected to standardized exposure conditions. A comparison is made between material properties for reference specimens
stored in the laboratory for the test period and specimens subjected to the two-stage test method. To eliminate the effect of moisture
from the comparison, the material properties of the latter test specimens are determined after they have been dried to constant
weight. The average value determined for each of the two sets of specimens is used for comparison.
1.4 Different properties can be measured to assess the effect of environmental factors on thermal insulation. This test method
requires that thermal resistance be determined based upon an average for three specimens measured after completing the test.
Secondary elements of this test method include visual observations such as cracking, delamination or other surface defects, as well
as the change in moisture content after each of the two stages of exposure prescribed by the test method.
1.5 Characterization of the tested material is an essential element of this test method. Material properties used for
characterization will include either compressive resistance or tensile strength values. The compressive resistance or tensile strength
is measured on two sets of specimens, one set conditioned as defined in 1.2 and a set of reference test specimens taken from the
same material batch and stored in the laboratory for the whole test period. For comparison, an average value is determined for each
of the two sets of specimens.
1.6 The values stated in SI units are to be regarded as standard. The values given in parentheses are for information only. No
other units of measurement are included in this standard.
1.7 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility
of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory
requirements prior to use.
2. Referenced Documents
2.1 ASTM Standards:
C165 Test Method for Measuring Compressive Properties of Thermal Insulations
C168 Terminology Relating to Thermal Insulation
This test method is under the jurisdiction of ASTM Committee C16 on Thermal Insulation and is the direct responsibility of Subcommittee C16.33 on Insulation Finishes
and Moisture.
Current edition approved May 1, 2010May 15, 2015. Published July 2010September 2015. Originally approved in 2001. Last previous edition approved in 20072010 as
C1512–07.–10. DOI: 10.1520/C1512-10.10.1520/C1512-10R15.
For referenced ASTM standards, visit the ASTM website, www.astm.org, or contact ASTM Customer Service at service@astm.org. For Annual Book of ASTM Standards
volume information, refer to the standard’s Document Summary page on the ASTM website.
Copyright © ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959. United States
C1512 − 10 (2015)
C177 Test Method for Steady-State Heat Flux Measurements and Thermal Transmission Properties by Means of the
Guarded-Hot-Plate Apparatus
C303 Test Method for Dimensions and Density of Preformed Block and Board–Type Thermal Insulation
C518 Test Method for Steady-State Thermal Transmission Properties by Means of the Heat Flow Meter Apparatus
C870 Practice for Conditioning of Thermal Insulating Materials
C618 Specification for Coal Fly Ash and Raw or Calcined Natural Pozzolan for Use in Concrete
D1621 Test Method for Compressive Properties of Rigid Cellular Plastics
D1623 Test Method for Tensile and Tensile Adhesion Properties of Rigid Cellular Plastics
E177 Practice for Use of the Terms Precision and Bias in ASTM Test Methods
E691 Practice for Conducting an Interlaboratory Study to Determine the Precision of a Test Method
3. Terminology
3.1 Definitions—Terms used in this test method are defined in Terminology C168 with the exceptions included as appropriate.
3.2 Definitions of Terms Specific to This Standard:
3.2.1 compressive resistance—the compressive load per unit of original area at the specified deformation. See Test Method
C165.
3.2.2 moisture accumulation—an increase in the average moisture content resulting from a specified exposure to conditions
facilitating moisture ingress into the material.
3.2.3 preconditioning—a procedure which subjects test specimens to standardized one directional thermal gradient.
3.2.4 thermal performance—comparison of thermal resistance of test specimens before and after cycling.
4. Summary of Test Method
4.1 To reduce the testing period, this procedure involves two stages:
4.1.1 Stage 1—Preconditioning under constant thermal gradient and relative humidity to accelerate ingress of moisture into the
test specimen.
4.1.2 Stage 2—Exposure to constant temperature and relative humidity on one side of test specimens with cycling
environmental conditions on the other side that include freeze-thaw exposure.
5. Significance and Use
5.1 Exposing a specimen to conditions of one-directional environmental cycling can increase its moisture content until a
decrease in material properties occurs (at a specific number of cycles). Such a test could be inappropriate due to the number of
cycles required to cause a decrease in material properties since product performance issues often arise only after many years of
exposure. The use of a preconditioning procedure is not intended to duplicate expected field performance. Rather the purpose is
to increase the moisture content of test materials prior to subjecting to them to environmental cycling.
5.2 The most important aspect of the preconditioning procedure is non-uniform moisture distribution in the specimen. The heat
flow is one directional causing moisture flow towards the cold side resulting in zones of dry material on the warm side and high
moisture content on the cold side. (Whether the high moisture content zone is located right at the cold surface of the specimen or
at some distance from this surface depends upon temperature oscillation and ability of the cold surface to dry outwards). Because
the preconditioning procedure involves thermal gradient, this preconditioning procedure results in a distribution of moisture
content that may occur under field exposure conditions. However, the resulting moisture content may differ significantly from that
which may be demonstrated in typical product applications.
5.3 The preconditioning results in accumulation of moisture in the thermal insulation resulting from the simultaneous exposure
to a difference in temperature and water vapor pressure. This test method is not intended to duplicate field exposure. It is intended
to provide comparative ratings. As excessive accumulation of moisture in a construction system may adversely affect its
performance, the designer should consider the potential for moisture accumulation and the possible effects of this moisture on the
system performance.
6. Apparatus
6.1 The room where the apparatus is placed shall be maintained at a temperature and relative humidity of 24 6 3°C (75 6 5°F)
and 50 6 10 %.
6.2 Freeze-Thaw Chamber, capable of maintaining an air temperature of -15 6 3°C (5 6 5°F) over an extended period of time.
The design of the apparatus should ensure that the temperature of the upper surface of the sheet metal located below the insulation
specimen (measured in the center of the pan) be not higher than -4°C (25°F) when the freezer’s air temperature reaches its lower
limit. This can be achieved by placing thermal insulation between the metal pan and the specimen frame and/or mixing of air in
the cold chamber.
C1512 − 10 (2015)
6.3 Sheet Metal Pan, placed below the specimens. This pan performs two functions: it equalizes temperature and reduces
diffusion of water vapor into the freeze-thaw chamber. The distance between the cold surface of the specimen and the sheet metal
1 1
should be no less than 6 mm ( ⁄4 in.) and no more than 12 mm ( ⁄2 in). The required space is normally maintained by attaching
a support of the required height that is made from 6-mm ( ⁄4 in.) thick Plexiglas or other non-absorbing materials on the inside
surface of the specimen frame (see Fig. 2).
6.4 Frame, that is placed in the door opening of the freezer (see Figs. 1 and 2) or other means of specimen support. Test frames
used are made from 6 6 0.5 mm thick Plexiglas or other non-absorbing material. These frames are used to mount individual test
specimens. The selection of the test frame (size of the test specimen) may vary based upon the thermal testing apparatus that is
used.
6.5 Warm Chamber, above the test specimens that is provided with a heater and a temperature controller capable of maintaining
a temperature of 24 6 2°C (75 6 3°F) and a humidifier capable of maintaining humidity in the warm chamber of 90 6 5 %RH.
6.6 Sensors, for measuring temperature of the freeze-thaw and warm chambers and relative humidity in the warm chamber.
6.7 Balance, capable of weighing mass of maximum 1 kg with precision of 0.01 g.
7. Test Specimens
2 2 2
7.1 Test specimens shall be square in cross-section with a minimum area of 645 cm (100 in. ) and a maximum of 3716 cm
(576 in. ). The standard specimen thickness shall be 2.54 cm (1 in.). Care should be taken so that the top and bottom surfaces of
the specimens exposed to thermal gradient are parallel with one another and perpendicular to the sides.
7.2 All surfaces of the specimens shall be free from visible flaws or imperfections.
7.3 For comparison, two test specimen sets each consisting of a minimum of three specimens are tested. One set of test
specimens are tested after preconditioning and after environmental cycling as described in Section 9. A second set of reference test
specimens are stored in the laboratory for the duration of preconditioning and environmental cycling test before thermal resistance
and compressive resistance or tensile strength testing.
8. Conditioning
8.1 Condition the test specimens before testing at 23 6 2°C (73 6 4°F) and 50 6 5 %RH relative humidity for not less than
40 h prior to test in accordance with Procedure A of Practice C618.
9. Procedure
9.1 Condition specimens to constant mass in accordance with Practice C870 before testing. Measure the dimensions and mass
of each specimen in accordance with Test Method C303. Record the initial mass of each specimen prior to subjecting to
preconditioning procedure.
9.2 Testing of Specimens Before and After Environmental Cycling:
9.2.1 Three specimens shall be tested for thermal resistance value before and after environmental cycling using Test Method
C518 or C177.
9.2.2 Where applicable, nine specimens shall be tested for compressive resistance before and after environmental cycling using
Test Method C165 or D1621.
FIG. 1 Plan View of Test Equipment Setup
C1512 − 10 (2015)
FIG. 2 Vertical Section at Interface Between Freezer Wall and Lid Illustrating Placement of Test Specimens in the Test Frame
9.2.3 Where applicable, nine specimens shall be tested for tensile strength before and after environmental cycling using Test
Method D1623.
9.3 Preconditioning:
9.3.1 Test specimens are preconditioned for 28 days to increase moisture content. This is achieved under conditions of water
vapor diffusion associated with a constant thermal gradient. The specimens are dividing two environments, namely:
9.3.1.1 Temperature of 24 6 2°C (75 6 3°F) and relative humidity of 90 6 5 % on warm side, and
9.3.1.2 Temperature of -15 6 3°C (5 6 5°F) and ambient relative humidity (uncontrolled relative humidity) on the cold side.
9.3.2 If the specimens are provided with facing, stucco lamina or other protective finishes, these finishes should be placed on
the cold side during the preconditioning exposure.
9.3.3 Weigh each specimen after initial preconditioning. Moisture content (% by volume) of the specimen is calculated after
completing the preconditioning exposure. Normally the specimens are returned to the same equipment but conditions on the cold
side are changed and cycling under environmental conditions which include freeze-thaw cycling on the cold side proceeds.
9.4 Environmental Cycling Conditions:
9.4.1 Place test specimens in the test frame (Fig. 2) and seal the edges of the test specimens to prevent passage of air around
the edges.
9.4.2 Test specimens shall be placed for 20 days (40 cycles) separating two environments:
9.4.2.1 Warm chamber where temperature and relative humidity are maintained at 24 6 2°C (75 6 3°F) and 90 6 5 %RH; and
9.4.2.2 Environmental cycling chamber where conditions require temperature cycling between two levels: -15 6 3°C (5 6 5°F)
and 15 6 3°C (59 6 5°F). The total cycling period is twelve hours, divided equally into cold and warm exposures. The warm
exposure (at least 4 h at temperature higher than 5°C) is ended with the transition period of no longer than 2 h. D
...

Questions, Comments and Discussion

Ask us and Technical Secretary will try to provide an answer. You can facilitate discussion about the standard in here.