Standard Practice for Use of an Alanine-EPR Dosimetry System

SIGNIFICANCE AND USE
4.1 The alanine-EPR dosimetry system provides a means for measuring absorbed dose. It is based on the measurement of specific stable free radicals in crystalline alanine generated by ionizing radiation.  
4.2 Alanine-EPR dosimetry systems are used in reference- or transfer-standard or routine dosimetry systems in radiation applications that include: sterilization of medical devices and pharmaceuticals, food irradiation, polymer modifications, medical therapy and radiation damage studies in materials (1, 13-15).
SCOPE
1.1 This practice covers dosimeter materials, instrumentation, and procedures for using the alanine-EPR dosimetry system to measure the absorbed dose in the photon or electron radiation processing of materials. The alanine system is based on electron paramagnetic resonance (EPR) spectroscopy of free radicals derived from the amino acid alanine.2  
1.2 The alanine dosimeter is classified as a type I dosimeter as it is affected by individual influence quantities in a well-defined way that can be expressed in terms of independent correction factors (see ISO/ASTM Practice 52628). The alanine dosimeter may be used in either a reference standard dosimetry system or in a routine dosimetry system.  
1.3 This document is one of a set of standards that provides recommendations for properly implementing dosimetry in radiation processing, and describes a means of achieving compliance with the requirements of ISO/ASTM 52628 “Practice for Dosimetry in Radiation Processing” for alanine dosimetry system. It should be read in conjunction with ISO/ASTM 52628.  
1.4 This practice covers the use of alanine-EPR dosimetry systems under the following conditions:  
1.4.1 The absorbed dose range is between 0.001 kGy and 150 kGy.  
1.4.2 The absorbed dose rate is up to 1 × 102 Gy s-1 for continuous radiation fields and up to 3 × 1010 Gy s-1 for pulsed radiation fields (1-4).3  
1.4.3 The radiation energy for photons and electrons is between 0.1 MeV and 30 MeV (1, 2, 5-8).  
1.4.4 The irradiation temperature is between –78 °C and +70 °C (2, 9-12).  
1.5 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety, health, and environmental practices and determine the applicability of regulatory limitations prior to use.  
1.6 This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.

General Information

Status
Published
Publication Date
31-Dec-2021
Technical Committee
Drafting Committee
Current Stage
Ref Project

Relations

Buy Standard

Standard
ASTM ISO/ASTM51607-22 - Standard Practice for Use of an Alanine-EPR Dosimetry System
English language
7 pages
sale 15% off
Preview
sale 15% off
Preview
Standard
REDLINE ASTM ISO/ASTM51607-22 - Standard Practice for Use of an Alanine-EPR Dosimetry System
English language
7 pages
sale 15% off
Preview
sale 15% off
Preview

Standards Content (Sample)

This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the
Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.
Designation: 51607 − 22
Standard Practice for
1
Use of an Alanine-EPR Dosimetry System
This standard is issued under the fixed designation 51607; the number immediately following the designation indicates the year of
original adoption or, in the case of revision, the year of last revision. A number in parentheses indicates the year of last reapproval. A
superscript epsilon (´) indicates an editorial change since the last revision or reapproval.
1. Scope 1.5 This standard does not purport to address all of the
safety concerns, if any, associated with its use. It is the
1.1 This practice covers dosimeter materials,
responsibility of the user of this standard to establish appro-
instrumentation, and procedures for using the alanine-EPR
priate safety, health, and environmental practices and deter-
dosimetry system to measure the absorbed dose in the photon
mine the applicability of regulatory limitations prior to use.
or electron radiation processing of materials. The alanine
1.6 This international standard was developed in accor-
system is based on electron paramagnetic resonance (EPR)
dance with internationally recognized principles on standard-
spectroscopy of free radicals derived from the amino acid
2 ization established in the Decision on Principles for the
alanine.
Development of International Standards, Guides and Recom-
1.2 The alanine dosimeter is classified as a type I dosimeter
mendations issued by the World Trade Organization Technical
as it is affected by individual influence quantities in a well-
Barriers to Trade (TBT) Committee.
defined way that can be expressed in terms of independent
correction factors (see ISO/ASTM Practice 52628). The ala-
2. Referenced documents
nine dosimeter may be used in either a reference standard
4
2.1 ASTM Standards:
dosimetry system or in a routine dosimetry system.
E3083 Terminology Relating to Radiation Processing: Do-
1.3 This document is one of a set of standards that provides
simetry and Applications
4
recommendations for properly implementing dosimetry in
2.2 ISO/ASTM Standards:
radiation processing, and describes a means of achieving
51261 Practice for Calibration of Routine Dosimetry Sys-
compliance with the requirements of ISO/ASTM 52628 “Prac-
tems for Radiation Processing
tice for Dosimetry in Radiation Processing” for alanine dosim-
51707 Guide for Estimation of Measurement Uncertainty in
etry system. It should be read in conjunction with ISO/ASTM
Dosimetry for Radiation Processing
52628.
52628 Practice for Dosimetry in Radiation Processing
52701 Guide for Performance Characterization of Dosim-
1.4 This practice covers the use of alanine-EPR dosimetry
eters and Dosimetry Systems for Use in Radiation Pro-
systems under the following conditions:
cessing
1.4.1 The absorbed dose range is between 0.001 kGy and
150 kGy.
2.3 International Commission on Radiation Units and Mea-
2 -1 5
1.4.2 The absorbed dose rate is up to 1 × 10 Gy s for surements (ICRU) Reports:
10 -1
continuous radiation fields and up to 3 × 10 Gy s for pulsed
ICRU Report 80 Dosimetry Systems for Use in Radiation
3
radiation fields (1-4). Processing
1.4.3 The radiation energy for photons and electrons is
ICRU Report 85a Fundamental Quantities and Units for
between 0.1 MeV and 30 MeV (1, 2, 5-8). Ionizing Radiation
6
1.4.4 The irradiation temperature is between –78 °C and
2.4 ISO Standard:
+70 °C (2, 9-12).
12749-4 Nuclear energy – Vocabulary – Part 4: Dosimetry
for radiation processing
1
This practice is under the jurisdiction of ASTM Committee E61 on Radiation
4
Processing and is the direct responsibility of Subcommittee E61.02 on Dosimetry For referenced ASTM and ISO/ASTM standards, visit the ASTM website,
Systems. Originally developed as a joint ASTM/ISO standard in conjunction with www.astm.org, or contact ASTM Customer Service at service@astm.org. For
ISO/TC 85/WG 3. Annual Book of ASTM Standards volume information, refer to the standard’s
Current edition approved Jan. 1, 2022. Published May 2024. Originally approved Document Summary page on the ASTM website.
5
in 1994. Last previous edition approved in 2013 as ISO/ASTM 51607:2013(E). Available from International Commission on Radiation Units and Measure-
DOI: 10.1520/51607-22. ments (ICRU), 7910 Woodmont Ave., Suite 400, Bethesda, MD 20814-3095,
2
The term “electron spin resonance” (ESR) is used interchangeably with U.S.A., https://www.icru.org/.
6
electron paramagnetic resonance (EPR). Available from International Organization for Standardization (ISO), ISO
3
The boldface numbers in parentheses refer to the bibliography at the end of this Central Secretariat, Chemin de Blandonnet 8, CP 401, 1214 Vernier, Geneva,
standard. Switz
...

This document is not an ASTM standard and is intended only to provide the user of an ASTM standard an indication of what changes have been made to the previous version. Because
it may not be technically possible to adequately depict all changes accurately, ASTM recommends that users consult prior editions as appropriate. In all cases only the current version
of the standard as published by ASTM is to be considered the official document.
Designation: ISO/ASTM 51607 − 2013(E) 51607 − 22
Standard Practice for
1
Use of an Alanine-EPR Dosimetry System
This standard is issued under the fixed designation ISO/ASTM 51607; the number immediately following the designation indicates the
year of original adoption or, in the case of revision, the year of last revision. A number in parentheses indicates the year of last
reapproval. A superscript epsilon (´) indicates an editorial change since the last revision or reapproval.
1. Scope
1.1 This practice covers dosimeter materials, instrumentation, and procedures for using the alanine-EPR dosimetry system for
measuringto measure the absorbed dose in the photon andor electron radiation processing of materials. The alanine system is based
2
on electron paramagnetic resonance (EPR) spectroscopy of free radicals derived from the amino acid alanine.
1.2 The alanine dosimeter is classified as a type I dosimeter as it is affected by individual influence quantities in a well-defined
way that can be expressed in terms of independent correction factors (see ASTMISO/ASTM Practice E262852628). The alanine
dosimeter may be used in either a reference standard dosimetry system or in a routine dosimetry system.
1.3 This document is one of a set of standards that provides recommendations for properly implementing dosimetry in radiation
processing, and describes a means of achieving compliance with the requirements of ASTMISO/ASTM E262852628 “Practice for
Dosimetry in Radiation Processing” for alanine dosimetry system. It should be read in conjunction with ASTMISO/ASTM
E262852628.
1.4 This practice covers the use of alanine-EPR dosimetry systems for dose measurements under the following conditions:
5
1.4.1 The absorbed dose range is between 1 and 1.5 × 100.001 kGy and 150 kGy. Gy.
2 -1 10 -1
1.4.2 The absorbed dose rate is up to 101 × 10 Gy s for continuous radiation fields and up to 3 × 10 Gy s for pulsed radiation
3
fields (1-4).
1.4.3 The radiation energy for photons and electrons is between 0.10.1 MeV and 30 MeV (1, 2, 5-8).
1.4.4 The irradiation temperature is between –78 °C and + 70 +70 °C (2, 9-12).
1.5 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility
of the user of this standard to establish appropriate safety and healthsafety, health, and environmental practices and determine
the applicability of regulatory limitations prior to use.
1.6 This international standard was developed in accordance with internationally recognized principles on standardization
established in the Decision on Principles for the Development of International Standards, Guides and Recommendations issued
by the World Trade Organization Technical Barriers to Trade (TBT) Committee.
1
This practice is under the jurisdiction of ASTM Committee E61 on Radiation Processing and is the direct responsibility of Subcommittee E61.02 on Dosimetry Systems,
and is also under the jurisdiction of . Originally developed as a joint ASTM/ISO standard in conjunction with ISO/TC 85/WG 3.
Current edition approved April 9, 2013Jan. 1, 2022. Published June 2013May 2024. Originally published as ASTM E 1607 – 94. Last previous ASTM edition
ε1
E 1607 – 96approved in 1994. Last previous edition approved . ASTM E 1607 – 94 was adopted by ISO in 1998 with the intermediate designation ISO 15566:1998(E). The
present International Standard ISO/ASTM 51607:2013(E) replaces ISO 15566 and is a major revision of the last previous edition ISO/ASTM 51607–2004(E). in 2013 as
ISO/ASTM 51607:2013(E). DOI: 10.1520/51607-22.
2
The term “electron spin resonance” (ESR) is used interchangeably with electron paramagnetic resonance (EPR).
3
The boldface numbers in parentheses refer to the bibliography at the end of this standard.
Copyright © ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959. United States
1

---------------------- Page: 1 ----------------------
51607 − 22
2. Referenced documents
4
2.1 ASTM Standards:
E170E3083 Terminology Relating to Radiation Measurements and DosimetryProcessing: Dosimetry and Applications
E2628 Practice for Dosimetry in Radiation Processing
E2701 Guide for Performance Characterization of Dosimeters and Dosimetry Systems for Use in Radiation Processing
4
2.2 ISO/ASTM Standards:
51261 Practice for Calibration of Routine Dosimetry Systems for Radiation Processing
51707 Guide for Estimating Uncertainties Estimation of Measurement Uncertainty in Dosimetry for Radiation Processing
52628 Practice for Dosimetry in Radiatio
...

Questions, Comments and Discussion

Ask us and Technical Secretary will try to provide an answer. You can facilitate discussion about the standard in here.