Standard Test Methods for Specific Gravity of Halogenated Organic Solvents and Their Admixtures

SCOPE
1.1 These test methods cover the determination of the specific gravity of halogenated organic solvents and solvent admixtures. They define suitable apparatus and procedures and furnish details underlying the interpretation of test data and the selection of numerical limits for agreement among interested persons and agencies.  
1.2 Two methods are covered as follows:  
1.2.1 Method A , specific gravity by means of a hydrometer.  
1.2.2 Method B , specific gravity by means of a pycnometer.  Note 1-In referee problems, Method B may be used.
1.2.3 Method C , specific gravity by means of an electronic densitometer.  
1.3 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use.

General Information

Status
Historical
Publication Date
09-Jan-2000
Current Stage
Ref Project

Relations

Buy Standard

Standard
ASTM D2111-95(2000) - Standard Test Methods for Specific Gravity of Halogenated Organic Solvents and Their Admixtures
English language
4 pages
sale 15% off
Preview
sale 15% off
Preview

Standards Content (Sample)

Designation: D 2111 – 95 (Reapproved 2000)
Standard Test Methods for
Specific Gravity of Halogenated Organic
1
Solvents and Their Admixtures
This standard is issued under the fixed designation D 2111; the number immediately following the designation indicates the year of
original adoption or, in the case of revision, the year of last revision. A number in parentheses indicates the year of last reapproval. A
superscript epsilon (e) indicates an editorial change since the last revision or reapproval.
This standard has been approved for use by agencies of the Department of Defense.
Specific gravity x/y°C, example 20/4°C (2)
1. Scope
When using an electronic densitometer to determine specific gravity,
1.1 These test methods cover the determination of the
the temperature of the material to be tested and the water reference will
specific gravity of halogenated organic solvents and solvent
be the same. Examples 25/25°C, 20/20°C.
admixtures. They define suitable apparatus and procedures and
3.1.2 density—the mass of a given material per unit volume.
furnish details underlying the interpretation of test data and the
3.1.2.1 Discussion—Density for chlorinated solvents is nor-
selection of numerical limits for agreement among interested
mally stated in grams per cubic centimetre. Pounds per gallon
persons and agencies.
is also commonly used.
1.2 Two methods are covered as follows:
1.2.1 Method A, specific gravity by means of a hydrometer.
4. Significance and Use
1.2.2 Method B, specific gravity by means of a pycnometer.
4.1 The density or specific gravity of a pure chlorinated
NOTE 1—In referee problems, Method B may be used.
solvent at a given temperature is constant. Density or specific
gravity can be used in identification of materials, the assay of
1.2.3 Method C, specific gravity by means of an electronic
binary mixtures, and as an indication of purity of a given
densitometer.
solvent.
1.3 This standard does not purport to address all of the
safety concerns, if any, associated with its use. It is the
5. Test Temperatures
responsibility of the user of this standard to establish appro-
5.1 ASTM specifications normally state the temperatures
priate safety and health practices and determine the applica-
for specific gravity of halogenated organic solvents at 25/25°C.
bility of regulatory limitations prior to use.
20/20°C and 60/60°F are other commonly used temperatures.
2. Referenced Documents
METHOD A—SPECIFIC GRAVITY BY MEANS
2.1 ASTM Standards:
OF A HYDROMETER
2
E 1 Specification for ASTM Thermometers
2
E 100 Specification for ASTM Hydrometers
6. Apparatus
6.1 Hydrometer—The hydrometers to be used shall be those
3. Terminology
specified in Specification E 100, as follows:
3.1 Definitions:
Nominal Specific Gravity Range ASTM Hydrometer No.
3.1.1 specific gravity—the ratio of the mass in air of a given
0.900 to 0.950 107H
volume of the material at a stated temperature to the mass in air 0.950 to 1.000 108H
1.000 to 1.050 125H
of an equal volume of distilled water at a stated temperature.
1.050 to 1.100 126H
3.1.1.1 Discussion—When the temperature of the material
1.100 to 1.150 127H
and of the water are the same, the specific gravity of the 1.150 to 1.200 128H
1.200 to 1.250 129H
material is expressed as follows:
1.250 to 1.300 130H
Specific gravity x/x°C, example 25/25°C (1) 1.300 to 1.350 131H
1.350 to 1.400 132H
When the temperature of the material and of the water are not the
1.400 to 1.450 133H
same, the specific gravity of the material is expressed as follows: 1.450 to 1.500 134H
1.500 to 1.550 135H
1.550 to 1.600 136H
1.600 to 1.650 137H
1
These test methods are under the jurisdiction of ASTM Committee D26 on
Halogenated Organic Solvents and Fire Extinguishing Agents and are the direct
6.2 Hydrometer Cylinder—The vessel in which the sample
responsibility of Subcommittee D26.04 on Test Methods.
for the gravity test is confined shall be made of clear glass and
Current edition approved March 15, 1995. Published May 1995. Originally
shall be cylindrical in shape. For convenience in pouring, it
published as D 2111 – 62. Last previous edition D 2111 – 92.
2
Annual Book of ASTM Standards, Vol 14.03. may have a lip on the rim. The inside diameter shall be at least
Copyright © ASTM, 100 Barr Harbor Drive, West Conshohocken, PA 19428-2959, United States.
1

---------------------- Page: 1 ----------------------
D 2111
25.0 mm (1.0 in.) greater than the outside diameter of the freshly boiled distilled water that has been cooled to 22 to
hydrometer used in it. The height of the cylinder shall be such 24°C. Place it in the water bath maintained at 25.0 6 0.5°C
that the length of the column of sample it contains is greater by
until the pycnometer and its contents are at a constant volume
at least 25.0 mm than the portion of the hydrometer that is at 25°C.
immersed beneath the surface of the sample after a state of
9.2 After immersion in the bath for at least 30 min, adjust
equilibrium has been reached.
the level of liquid to the proper point on t
...