Standard Specification for Flow Table for Use in Tests of Hydraulic Cement

ABSTRACT
This specification covers requirements for the flow table and accessory apparatus used in making flow tests for consistency or mortars in tests of hydraulic cement. The flow table shall consist of an integrally cast rigid iron frame and a circular rigid table top. Other details of the flow table, its mounting and lubrication are described in the specification. The conical mold for casting the flow specimen and the caliper for measuring the diameter of the mortar after it has been spread by the operation of the table are also discussed.
SCOPE
1.1 This specification covers requirements for the flow table and accessory apparatus (Note 1) used in making flow tests for consistency of mortars in tests of hydraulic cement, such as but not limited to Test Method C1437.
Note 1: To help clarify the design of the flow table and accessory apparatus see the drawing in Fig. 1 [Fig. 2]. This drawing is for informational purposes only. See Fig. 3 of the flow table and accessory apparatus as a reference for further clarification.
FIG. 1 Flow Table and Accessory Apparatus (Partial) (IP)  
FIG. 1 Flow Table and Accessory Apparatus (Partial) (IP) (continued)  
FIG. 2 Flow Table and Accessory Apparatus (Partial) [SI]  
FIG. 2 Flow Table and Accessory Apparatus (Partial) [SI] (continued)  
FIG. 2 Flow Table and Accessory Apparatus (Partial) [SI] (continued)  
FIG. 3 Flow Table Accessory Apparatus  
1.2 The values stated in either SI units or inch-pound units are to be regarded separately as standard. The values stated in each system are not necessarily exact equivalents; therefore, to ensure conformance with the standard, each system shall be used independently of the other, and values from the two systems shall not be combined. It is permissible to use an inch-pound caliper and mold with a SI flow table or a SI caliper and mold with an inch-pound flow table. It is not permissible to mix a SI mold with an inch-pound caliper or an inch-pound mold with a SI caliper.  
1.3 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety, health, and environmental practices and determine the applicability of regulatory limitations prior to use.  
1.4 This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.

General Information

Status
Published
Publication Date
31-Mar-2023
Technical Committee
Drafting Committee
Current Stage
Ref Project

Relations

Buy Standard

Technical specification
ASTM C230/C230M-23 - Standard Specification for Flow Table for Use in Tests of Hydraulic Cement
English language
7 pages
sale 15% off
Preview
sale 15% off
Preview
Technical specification
REDLINE ASTM C230/C230M-23 - Standard Specification for Flow Table for Use in Tests of Hydraulic Cement
English language
7 pages
sale 15% off
Preview
sale 15% off
Preview

Standards Content (Sample)

This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the
Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.
Designation: C230/C230M − 23
Standard Specification for
1
Flow Table for Use in Tests of Hydraulic Cement
This standard is issued under the fixed designation C230/C230M; the number immediately following the designation indicates the year
of original adoption or, in the case of revision, the year of last revision. A number in parentheses indicates the year of last reapproval.
A superscript epsilon (´) indicates an editorial change since the last revision or reapproval.
1. Scope* C219 Terminology Relating to Hydraulic and Other Inor-
ganic Cements
1.1 This specification covers requirements for the flow table
C1437 Test Method for Flow of Hydraulic Cement Mortar
and accessory apparatus (Note 1) used in making flow tests for
consistency of mortars in tests of hydraulic cement, such as but
3. Terminology
not limited to Test Method C1437.
3.1 Definitions:
NOTE 1—To help clarify the design of the flow table and accessory
3.1.1 The terms in this specification are defined in Termi-
apparatus see the drawing in Fig. 1 [Fig. 2]. This drawing is for
nology C125 and Terminology C219.
informational purposes only. See Fig. 3 of the flow table and accessory
apparatus as a reference for further clarification.
4. Flow Table and Frame
1.2 The values stated in either SI units or inch-pound units
4.1 The flow table apparatus shall consist of an integrally
are to be regarded separately as standard. The values stated in
cast rigid iron frame and a circular rigid table top 10 6 0.1 in.
each system are not necessarily exact equivalents; therefore, to
[255 6 2.5 mm] in diameter, with a shaft attached perpendicu-
ensure conformance with the standard, each system shall be
lar to the table top by means of a screw thread. The table top
used independently of the other, and values from the two
and shaft with contact shoulder shall be mounted on a frame in
systems shall not be combined. It is permissible to use an
such a manner that it can be raised and dropped vertically
inch-pound caliper and mold with a SI flow table or a SI caliper
through the specified height of 0.500 6 0.005 in. [12.7 6
and mold with an inch-pound flow table. It is not permissible
0.13 mm] for new tables and of 0.500 6 0.015 in. [12.7 6
to mix a SI mold with an inch-pound caliper or an inch-pound
0.38 mm] for tables in use, by means of a rotated cam. The
mold with a SI caliper.
table top shall have a fine machined plane surface, free of
1.3 This standard does not purport to address all of the
blowholes and surface defects. The top shall be scribed with
safety concerns, if any, associated with its use. It is the
eight equidistant lines extending from the outside circumfer-
responsibility of the user of this standard to establish appro- 3
ence toward the center of the table, ending 2 ⁄8 in. [60 mm]
priate safety, health, and environmental practices and deter- 1
from the center. Each line shall end with a scribed arc, ⁄4 in.
mine the applicability of regulatory limitations prior to use.
[6 mm] long, whose center point is the center of the table top
1.4 This international standard was developed in accor- 3
with a radius of 2 ⁄8 in. [59.5 mm]. The scribe lines shall be
dance with internationally recognized principles on standard-
made with a 60° tool to a depth of 0.01 in. [0.25 mm]. The
ization established in the Decision on Principles for the
table top shall be of cast brass or bronze having a Rockwell
Development of International Standards, Guides and Recom-
hardness number not less than 25 HRB with an edge thickness
mendations issued by the World Trade Organization Technical
of 0.3 in. [7.5 mm], and shall have six integral radial stiffening
Barriers to Trade (TBT) Committee.
ribs. The table top and attached shaft shall weigh 9 6 0.1 lb
[4.08 6 0.05 kg] and the weight shall be symmetrical around
2. Referenced Documents
the center of the shaft.
2
2.1 ASTM Standards:
4.2 The cam and vertical shaft shall be of medium carbon
C125 Terminology Relating to Concrete and Concrete Ag-
machinery steel, hardened on the end of the shaft contacting
gregates
the cam and the tip of the cam contacting the shaft. The shaft
shall be straight and the difference between the diameter of the
1
This specification is under the jurisdiction of ASTM Committee C01 on
shaft and the diameter of the bore of the frame shall be not less
Cement and is the direct responsibility of Subcommittee C01.22 on Workability.
than 0.002 in. [0.05 mm] and not more than 0.003 in.
Current edition approved April 1, 2023. Published May 2023. Originally
[0.08 mm] for new tables and shall be maintained at 0.002 to
approved in 1949. Last previous edition approved in 2021 as C230/C230M – 21.
DOI: 10.1520/C0230_C0230M-23.
0.0
...

This document is not an ASTM standard and is intended only to provide the user of an ASTM standard an indication of what changes have been made to the previous version. Because
it may not be technically possible to adequately depict all changes accurately, ASTM recommends that users consult prior editions as appropriate. In all cases only the current version
of the standard as published by ASTM is to be considered the official document.
Designation: C230/C230M − 21 C230/C230M − 23
Standard Specification for
1
Flow Table for Use in Tests of Hydraulic Cement
This standard is issued under the fixed designation C230/C230M; the number immediately following the designation indicates the year
of original adoption or, in the case of revision, the year of last revision. A number in parentheses indicates the year of last reapproval.
A superscript epsilon (´) indicates an editorial change since the last revision or reapproval.
1. Scope*
1.1 This specification covers requirements for the flow table and accessory apparatus (Note 1) used in making flow tests for
consistency of mortars in tests of hydraulic cement, such as but not limited to Test Method C1437.
NOTE 1—To help clarify the design of the flow table and accessory apparatus see the drawing in Fig. 1 [Fig. 2]. This drawing is for informational purposes
only. See Fig. 3 of the flow table and accessory apparatus as a reference for further clarification.
1.2 The values stated in either SI units or inch-pound units are to be regarded separately as standard. The values stated in each
system are not necessarily exact equivalents; therefore, to ensure conformance with the standard, each system shall be used
independently of the other, and values from the two systems shall not be combined. It is permissible to use an inch-pound caliper
and mold with a SI flow table or a SI caliper and mold with an inch-pound flow table. It is not permissible to mix a SI mold with
an inch-pound caliper or an inch-pound mold with a SI caliper.
1.3 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of
the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory
limitations prior to use.
1.3 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility
of the user of this standard to establish appropriate safety, health, and environmental practices and determine the applicability of
regulatory limitations prior to use.
1.4 This international standard was developed in accordance with internationally recognized principles on standardization
established in the Decision on Principles for the Development of International Standards, Guides and Recommendations issued
by the World Trade Organization Technical Barriers to Trade (TBT) Committee.
2. Referenced Documents
2
2.1 ASTM Standards:
C125 Terminology Relating to Concrete and Concrete Aggregates
C219 Terminology Relating to Hydraulic and Other Inorganic Cements
C1437 Test Method for Flow of Hydraulic Cement Mortar
3. Terminology
3.1 Definitions:
1
This specification is under the jurisdiction of ASTM Committee C01 on Cement and is the direct responsibility of Subcommittee C01.22 on Workability.
Current edition approved Jan. 1, 2021April 1, 2023. Published January 2021May 2023. Originally approved in 1949. Last previous edition approved in 20202021 as
C230/C230M – 20.C230/C230M – 21. DOI: 10.1520/C0230_C0230M-21. 10.1520/C0230_C0230M-23.
2
For referenced ASTM standards, visit the ASTM website, www.astm.org, or contact ASTM Customer Service at service@astm.org. For Annual Book of ASTM Standards
volume information, refer to the standard’s Document Summary page on the ASTM website.
*A Summary of Changes section appears at the end of this standard
Copyright © ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959. United States
1

---------------------- Page: 1 ----------------------
C230/C230M − 23
3.1.1 The terms in this specification are defined in Terminology C125 and Terminology C219.
4. Flow Table and Frame
4.1 The flow table apparatus shall consist of an integrally cast rigid iron frame and a circular rigid table top 10 6 0.1 in. [255 6
2.5 mm] in diameter, with a shaft attached perpendicular to the table top by means of a screw thread. The table top and shaft with
contact shoulder shall be mounted on a frame in such a manner that it can be raised and dropped vertically through the specified
height of 0.500 6 0.005 in. [12.7 6 0.13 mm] for new tables and of 0.500 6 0.015 in. [12.7 6 0.38 mm] for tables in use, by
means of a rotated cam. The table top shall have a fine machined plane surface, free of blowholes and surface defects. The top
5
shall be scribed with eight equidistant lines 2 ⁄8 in. [68 mm] long, extending from the outside circumference toward the center of
3 1
the table. table, ending 2 ⁄8 in. [60 mm] from t
...