ASTM D6698-07
(Test Method)Standard Test Method for On-Line Measurement of Turbidity Below 5 NTU in Water
Standard Test Method for On-Line Measurement of Turbidity Below 5 NTU in Water
SCOPE
1.1 This test method is applicable to the on-line measurement of turbidity under 5 nephelometric turbidity units (NTU) in water.
1.2 It is the user's responsibility to ensure the validity of this test method for waters of untested matrices.
1.3 In this test method calibration standards are defined in NTU values, but other assigned turbidity units are assumed to be equivalent.
1.4 This method assigns traceable reporting units to the type of respective technology that was used to perform the measurement. Units are numerically equivalent with respect to the calibration standard. For example, a 1 NTU formazin standard is also equal to a 1FNU standard, a 1 FNRU standard and so forth.
This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use.
General Information
Relations
Standards Content (Sample)
NOTICE: This standard has either been superseded and replaced by a new version or withdrawn.
Contact ASTM International (www.astm.org) for the latest information
Designation: D6698 − 07
StandardTest Method for
1
On-Line Measurement of Turbidity Below 5 NTU in Water
This standard is issued under the fixed designation D6698; the number immediately following the designation indicates the year of
original adoption or, in the case of revision, the year of last revision. A number in parentheses indicates the year of last reapproval. A
superscript epsilon (´) indicates an editorial change since the last revision or reapproval.
1. Scope D3864 Guide for Continual On-Line Monitoring Systems
for Water Analysis
1.1 This test method is applicable to the on-line measure-
2.2 Other Standards
ment of turbidity under 5 nephelometric turbidity units (NTU)
EPA 180.1 Methods for Chemical Analysis of Water and
in water.
4
Wastes, Turbidity
1.2 Itistheuser’sresponsibilitytoensurethevalidityofthis
4
GLI Method 2
test method for waters of untested matrices.
Hach Method 8195 Determination of Turbidity by Nephelo-
5
1.3 In this test method calibration standards are defined in
metry EMMC Format
6
NTU values, but other assigned turbidity units are assumed to
ISO 7027 Determination of Turbidity
7
be equivalent.
Standard Method 2130B
1.4 Thismethodassignstraceablereportingunitstothetype 2.3 Other Documents
8
of respective technology that was used to perform the mea- US Patent 4,283,143
8
surement. Units are numerically equivalent with respect to the US Patent 4,291,980
5
calibration standard. For example, a 1 NTU formazin standard US Patent 5,777,011
is also equal to a 1FNU standard, a 1 FNRU standard and so
3. Terminology
forth.
1.5 This standard does not purport to address all of the 3.1 Definitions—For definitions of terms used in this
method refer to Terminology D1129.
safety concerns, if any, associated with its use. It is the
responsibility of the user of this standard to establish appro-
3.2 Definitions of Terms Specific to This Standard:
priate safety and health practices and determine the applica-
3.2.1 calibration turbidity standard—a turbidity standard
bility of regulatory limitations prior to use.
that is traceable and equivalent to the reference turbidity
standard to within statistical errors, including commercially
2. Referenced Documents
prepared 4000 NTU Formazin, stabilized formazin, and sty-
2
2.1 ASTM Standards:
renedivinylbenzene (SDVB). These standards may be used to
D1129 Terminology Relating to Water calibrate the instrument.
D1193 Specification for Reagent Water
NOTE 1—Calibration standards may be instrument specific.
D1889 Test Method for Turbidity of Water (Withdrawn
3 3.2.2 calibration verification standards—defined standards
2007)
used to verify the accuracy of a calibration in the measurement
D2777 Practice for Determination of Precision and Bias of
range of interest. These standards may not be used to perform
Applicable Test Methods of Committee D19 on Water
calibrations, only calibration verifications. Included standards
D3370 Practices for Sampling Water from Closed Conduits
are opto-mechanical light scatter devices, gel-like standards, or
any other type of stable liquid standard.
1
This test method is under the jurisdiction of ASTM Committee D19 on Water
and is the direct responsibility of Subcommittee D19.03 on Sampling Water and
4
Water-Formed Deposits, Analysis of Water for Power Generation and Process Use, Available from United States Environmental Protection Association (EPA),
On-Line Water Analysis, and Surveillance of Water. Ariel Rios Bldg., 1200 Pennsylvania Ave., NW, Washington, DC 20460, http://
Current edition approved June 15, 2007. Published July 2007. Originally www.epa.gov.
5
approved in 2001. Last previous edition approved in 2001 as D6698 – 01. DOI: Hach Company, P.O. Box 389, Loveland, CO, 80539-0389, www.hach.com.
6
10.1520/D6698-07. Available from International Organization for Standardization (ISO), 1 rue de
2
For referenced ASTM standards, visit the ASTM website, www.astm.org, or Varembé, Case postale 56, CH-1211, Geneva 20, Switzerland, http://www.iso.ch.
7
contact ASTM Customer Service at service@astm.org. For Annual Book of ASTM Standard Methods for the Examination of Water and Wastewater, 21st Edition,
Standards volume information, refer to the standard’s Document Summary page on American Public Health Association, Washington, DC, 2005, www.standardmeth-
the ASTM website. ods.org.
3 8
The last approved version of this historical standard is referenced on Available from AMCO Clear, P.O. Box 245, Powell, OH, 43065, www.amco-
www.astm.org. clear.com.
Copyright © ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959. United States
1
---------------------- Page: 1 ----------------------
D6698 − 07
NOTE 2—Calibration verification standards may be instrument specific.
5. Significance and Use
3.2.3 in-situ nephelometer—a turbidimeter that determines
5.1 Turbid
...