Standard Practice for General Techniques for Obtaining Infrared Spectra for Qualitative Analysis

SIGNIFICANCE AND USE
Infrared spectroscopy is the most widely used technique for identifying organic and inorganic materials. This practice describes methods for the proper application of infrared spectroscopy.
SCOPE
1.1 This practice covers the spectral range from 4000-50 cm-1 and includes techniques that are useful for qualitative analysis of liquid-, solid-, and vapor-phase samples by infrared spectrometric techniques for which the amount of sample available for analysis is not a limiting factor. These techniques are often also useful for recording spectra at frequencies higher than 4000 cm-1, in the near-infrared region.
1.2 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use. Specific precautions are given in 6.5.1.

General Information

Status
Historical
Publication Date
09-Mar-1998
Current Stage
Ref Project

Relations

Buy Standard

Standard
ASTM E1252-98(2002) - Standard Practice for General Techniques for Obtaining Infrared Spectra for Qualitative Analysis
English language
13 pages
sale 15% off
Preview
sale 15% off
Preview

Standards Content (Sample)


NOTICE: This standard has either been superseded and replaced by a new version or withdrawn.
Contact ASTM International (www.astm.org) for the latest information
Designation:E1252–98 (Reapproved2002)
Standard Practice for
General Techniques for Obtaining Infrared Spectra for
Qualitative Analysis
This standard is issued under the fixed designation E1252; the number immediately following the designation indicates the year of
original adoption or, in the case of revision, the year of last revision.Anumber in parentheses indicates the year of last reapproval.A
superscript epsilon (e) indicates an editorial change since the last revision or reapproval.
1. Scope 4. Significance and Use
1.1 This practice covers the spectral range from 4000–50 4.1 Infraredspectroscopyisthemostwidelyusedtechnique
−1
cm and includes techniques that are useful for qualitative for identifying organic and inorganic materials. This practice
analysisofliquid-,solid-,andvapor-phasesamplesbyinfrared describes methods for the proper application of infrared
spectrometric techniques for which the amount of sample spectroscopy.
available for analysis is not a limiting factor.These techniques
5. General
areoftenalsousefulforrecordingspectraatfrequencieshigher
–1
than 4000 cm , in the near-infrared region. 5.1 Infrared (IR) qualitative analysis is carried out by
functional group identification (1-3) or by the comparison of
1.2 This standard does not purport to address all of the
safety concerns, if any, associated with its use. It is the IR absorption spectra of unknown materials with those of
known reference materials, or both.These spectra are obtained
responsibility of the user of this standard to establish appro-
priate safety and health practices and determine the applica- (4-8) through transmission, reflection, and other techniques,
such as photoacoustic spectroscopy (PAS). Spectra that are to
bility of regulatory limitations prior to use. Specific precau-
tions are given in 6.5.1. be compared should be obtained by the same technique and
under the same conditions. Users of published reference
2. Referenced Documents
spectra (9-16) should be aware that not all of these spectra are
2.1 ASTM Standards: fully validated.
E131 Terminology Relating to Molecular Spectroscopy 5.1.1 Instrumentation and accessories for infrared qualita-
E168 Practices for General Techniques of Infrared Quanti- tive analysis are commercially available. The manufacturer’s
tative Analysis manual should be followed to ensure optimum performance
E334 Practices for General Techniques of Infrared Mi- and safety.
croanalysis 5.2 Transmission spectra are obtained by placing a thin
E573 Practices for Internal Reflection Spectroscopy uniform layer of the sample perpendicular to the infrared
E932 Practice for Describing and Measuring Performance radiation path (see 9.5.1 for exception in order to eliminate
of Dispersive Infrared Spectrometers interference fringes for thin films). The sample thickness must
E1421 PracticeforDescribingandMeasuringPerformance be adequate to cause a decrease in the radiant power reaching
of Fourier Transform Infrared (FT-IR) Spectrometers: the detector at the absorption frequencies used in the analysis.
Level Zero and Level One For best results, the absorbance of the strongest bands should
E1642 PracticeforGeneralTechniquesofGasChromatog- be in the range from 1 to 2, and several bands should have
raphy Infrared (GC/IR) Analysis absorbances of 0.6 units or more. There are exceptions to this
generalization based on the polarity of the molecules being
3. Terminology
measured. For example, saturated hydrocarbons are nonpolar,
3.1 Definitions—Fordefinitionsoftermsandsymbols,refer
and their identifying bands are not strong enough unless the
−1
to Terminology E131. C-H stretch at 2920 cm is opaque and the deformation bands
are in the range from 1.5 to 2.0 absorbance units (A) at 1440
−1
to 1460 cm . Spectra with different amounts of sample in the
radiation path may be required to permit reliable analysis. If
This practice is under the jurisdiction ofASTM Committee E-13 on Molecular
spectra are to be identified by computerized curve matching,
Spectroscopy and is the direct responsibility of Subcommittee E13.03 on Infrared
Spectroscopy.
Current edition approved March 10, 1998. Published June 1998. Originally
e1 3
published as E1252–88. Last previous edition E1252–94 . The boldface numbers in parentheses refer to a list of references at the end of
Annual Book of ASTM Standards, Vol 03.06. the text.
Copyright © ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959, United States.
E1252–98 (2002)
the absorbance of the strongest band should be less than 1; Demountable spacers can be used when a longer path length is
otherwise, the effect of the instrument line shape function will required to obtain a useful spectrum.
cause errors in the relative intensities of bands in spectra
6.3 Internal Reflection Spectroscopy (IRS)—Viscous mate-
measured by dispersive spectrometers and by FT-IR spectrom- rials can be smeared on one or both sides of an internal
eters with certain apodization functions (specially triangular).
reflection element (IRE). See Practices E573 for detailed
5.2.1 Techniques for obtaining transmission spectra vary information on this technique.
with the sample state. Most samples, except free-standing thin
6.4 Disposable IR Cards —These can be used to obtain
films, require IR transparent windows or matrices containing
spectra of non-volatile liquids.Avery small drop, usually less
the sample. Table 1 gives the properties of IR window
than 10 µLof the liquid, is applied near the edge of the sample
materials commonly employed. Selection of the window ma-
application area. If the sample does not easily flow across the
terial depends on the region of the IR spectrum to be used for
substrate surface, it may be spread using an appropriate tool.
analysis, on the absence of interference with the sample, and
The sample needs to be applied in a thin layer, completely
adequate durability for the sample type.
covering an area large enough that the entire radiation beam
5.3 Spectraobtainedbyreflectionconfigurationscommonly
passes through the sample. Note that any volatile components
exhibit both reflection and absorption characteristics and are
of a mixture will be lost in this process, which may make the
affected by the refractive indices of the media and the inter-
use of a disposable card a poor choice for such systems.
faces.Spectralinterpretationshouldbebasedonreferencesrun
6.5 Solution Techniques:
underthesameexperimentalconditions.Inparticular,itshould
6.5.1 Analysis of Materials Soluble in Infrared (IR) Trans-
be realized that the spectrum of the surface of a sample
parent Solvent: The Split Solvent Technique—Many solid and
recordedbyreflectionwilloftendifferfromthespectrumofthe
liquid samples are soluble in solvents that are transparent in
bulkmaterialasrecordedbytransmissionspectroscopy.Thisis
parts of the infrared spectral region. A list of solvents com-
because the chemistry of the surface often differs from that of
monly used in obtaining solution spectra is given in Table 2.
the bulk, due to factors such as surface oxidation, migration of
The selection of solvents depends on several factors. The
species from the bulk to the surface, and possible surface
sample under examination must have adequate solubility, it
contaminants. Some surface measurements are extremely sen-
must not react with the solvent, and the solvent must have
sitive to small amounts of materials present on a surface,
appropriate transmission regions that enable a useful spectrum
whereas transmission spectroscopy is relatively insensitive to
to be obtained. Combinations of solvents and window materi-
these minor components.
als can often be selected that will allow a set of qualitative
5.3.1 Reflection spectra are obtained in four configurations:
solution-phasespectratobeobtainedovertheentireIRregion.
5.3.1.1 Specular reflectance (7.5),
One example of this “split solvent” technique utilizes carbon
5.3.1.2 Diffuse reflectance (7.6),
tetrachloride (CCl ) and carbon disulfide (CS ) as solvents.
4 2
5.3.1.3 Reflection-absorption (7.7),
NOTE 1—Warning:Both CCl and CS are toxic; keep in a well
4 2
5.3.1.4 Internal reflection (7.9). Refer to Practices E573.
ventilated hood. Use of these solvents is prohibited in many laboratories.
This technique is also called Attenuated Total Reflection
Inaddition,CS is extremelyflammable;keepawayfromignitionsources,
(ATR), and
even a steam bath. Moreover, CS is reactive (sometimes violently) with
5.3.1.5 Grazing angle reflectance.
primaryandsecondaryaliphaticaminesandmustnotbeusedasasolvent
5.4 Photoacoustic IR spectra (11.2).
for these compounds. Similarly, CCl reacts with aluminum metal.
Depending on conditions such as temperature and particle size, the
5.5 Emission spectroscopy (11.4).
reaction has been lethally violent.
TEST METHODS AND TECHNIQUES
6.5.1.1 Absorption by CCl is negligible in the region
−1 −1
4000-1330 cm and by CS in the region 1330-400 cm in
6. Analysis of Liquids
cells of about 0.1 mm thickness. (Other solvents can be used.)
6.1 Fixed Cells—A wide range of liquid samples of low to
Solutions are prepared, usually in the 5–10% weight/volume
moderate viscosity may be introduced into a sealed fixed-path
range, and are shaken to ensure uniformity. The solutions are
length cell. These are commercially available in a variety of
transferred by clean pipettes or by syringes that have been
materials and path lengths.Typical path lengths are 0.01 to 0.2
cleaned with solvent and dried to avoid cross-contamination
mm. See 5.2 for considerations in selection of cell materials
with a previous sample. If the spectrum of a 10% solution
and path lengths.
contains many bands that are too deep and broad for accurate
6.2 Capillary Films—Some liquids are too viscous to force
frequency measurement, thinner cells or a more dilute solution
into or out of a sealed cell. Examination of viscous liquids is
must be used.
accomplished by placing one or more drops in the center of a
NOTE 2—New syringes should be cleaned before use. Glass is the
flat window. Another flat window is then placed on top of the
preferred material. If plastic is used as containers, lids, syringes, pipettes,
liquid. Pressure is applied in order to form a bubble-free
and so forth, analytical blanks are necessary as a check against contami-
capillary film covering an area large enough that the entire
nation.
radiation beam passes through the film. The film thickness is
regulated by the amount of pressure applied and the viscosity
of the liquid. A capillary film prepared in this manner has a
path length of about 0.01 mm. Volatile and highly fluid
The 3M disposable IR Card is manufactured by 3M Co., Disposable Products
materials may be lost from films prepared in this manner. Division.
E1252–98 (2002)
TABLE 1 Properties of Window Materials (in order of long-wavelength limit)
A
Cutoff Range Useful Transmission Range
Chemical Water Refractive at
Window Material Remarks
−1 −1
Composition Solubility Index (;µm)
(µm) (cm ) (µm) (cm )
B
Glass SiO + ;2.5 ;4000 0.35–2 28 570–5000 insoluble 1.5–1.9 HF, alkali
B
Quartz (fused) SiO ;3.5 ;2857 0.2–4 50 000–2500 insoluble 1.43 4.5 HF
SIlicon Nitrate Si N 0.3–4.5 33 000–2200
3 4
Silicon Carbide SiC 0.6–5 16 600–2000
C
Calcite CaCO 0.2–5 50 000–2000 1.65, 1.5 0.589 Reacts with acids
Sapphire Al O ;5.5 ;1818 0.2–5.5 50 000–1818 insoluble 1.77 0.55 Good strength, no cleavage
2 3
ALON 9AI O .5AIN 0.2–5.5 50 000–1700 1.8 0.6
2 3
Spinel MgAI O 0.2–6 50 000–1600 1.68 0.6
2 4
B
Strontium Titanate SrTiO 0.39–6 25 000–1700 insoluble 2.4 HF
B
Titanium Dioxide TiO 0.42–6 24 000–1700 insoluble 2.6–2.9 H SO and Alkali
2 2 4
B
Lithium Fluoride LiF ;6.0 ;1667 0.2–7 50 000–1429 slightly 1.39 1.39 Acid
B
Zirconia ZrO 0.36–7 27 000–1500 insoluable 2.15 HF and H SO
2 2 4
D
Silicon Si 1.5–7 and 6600–1430 insoluble 3.4 11.0 Reacts with HF, alkali
10–FIR
Yttria Y 0.25–8 40 000–1250 1.9 0.6
Yttria (La-doped) 0.09La O - 0.25–8 40 000–1250 1.8 0.6
2 3
0.91Y O
2 3
E B
IRTRAN I MgF 2–8 5 000–1 250 slightly 1.3 6.7 HNO
2 3
B
Magnesium Oxide MgO 0.4–8 25 000–1300 insoluble 1.6 5 Acid and NH salts
B
Fluorite CaF ;8.0 ;1250 0.2–10 50 000–1000 insoluble 1.40 8.0 Amine salt and NH salts
2 4
Strontium Fluoride SrF 0.13–11 77 000–909 slightly 1.4
E
IRTRAN III CaF 0.2–11 50 000–909 insoluble 1.34 5.0 Polycrystalline, no cleavage
Gallium Phosphide GaP 0.5–11 20 000–910
GaP
Lead Fluoride PbF 0.3–12 3450–833 1.7 1
F B
Servofrax As S 1–12 10 000–833 insoluble 2.59 0.67 Alkali , softens at 195°C
2 3
slightly (hot)
Barium Fluoride BaF ;11 ;909 0.2–13 50 000–769 insoluble 1.45 5.1
AMTIR GeAsSe Glass 0.9–14 11 000–725 insoluble 2.5 10 Hard, brittle, attacked by alkali, good
ATR material
E
IRTRAN II ZnS 1–14 10 000–714 insoluble 2.24 5.5 Insoluble in most solvents
Indium Phosphide InP 1–14 10 000–725
Potassium Floride KF 0.16–15 62 500–666 soluble 1.3 0.3 Extremely deliquescent: not
recommended for routine use
G
Rock salt NaCl ;16 ;625 0.2–16 50 000–625 soluble 1.52 4.7 Soluble in glycerine
Cadmium Sulfide CdS 0.5–16 20 000–625
Arsenic Triselenide As Se 0.8–17 12 500–600 slightly 2.8 Soluble in bases
2 3
Gallium Arsenide GaAs 1–17 10 000–600 insoluble 3.14 Slightly soluable in acids and bases
Germanium Ge 2–20 5 000–500 insoluble 4.0 13.0
G
Sylvite KCl 0.3–21 33 333–476 soluble 1.49 0.5 Soluble in glycerine
E
IRTRAN IV ZnSe 1–21 10 000–476 insoluble 2.5 1.0 Polycrystalline
Sodium Bromide NaBr 0.2–23 50 000–435 Soluble 1.7 0.35
Sodium Iodide NaI 0.25–25 40 000–400 Soluble 1.7 0.5
H
Silver Chloride AgCl ;22 ;455 0.6–25 16 6667–400 insoluble 2.0 3.8 Soft, darkens in light reacts with
metals
Potassium Bromide KBr ;25 ;400 0.2–27 50 000–370 soluble 1.53 8.6 Soluble in alcohol; fogs
B
Cadmium Telluride CdTe ;28 ;360 0.5–28 20 000–360 insoluble 2.67 10 Acids, HNO
Thallium Chloride 0.4–30 25 000–330 slightly 2.2 0.75 Toxic
TICI
KRS-6 Tl CIBr 0.4–32 25 000–310 slightly 2.0–2.3 0.6–24 Toxic
H
Silver Bromide AgBr ;35 ;286 2–35 5 000–286 insoluble Soft, darkens in light , reacts with
metals
B
KRS-5 Tl2Brl ;40 ;250 0.7–38 14 286–263 slightly 2.38 4.0 Toxic, soft, soluble in alcohol, HNO
Cesium Bromide CsBr ;35 ;286 0.3–40 33 333–250 soluble 1.66 8.0 Soft, fogs, soluble alcohols
Potassium Iodide Kl 0.15–45 66 600–220
Thallium Bromide TIBr 0.45–45 22 000–220 slightly 2.3 0.6–25 Toxic
Cesium Iodide CsI ;52 ;192 0.3–50 33 330–220 soluble 1.74 8.0
Low-density (CH CH )n 20–220 500–45 insoluble 1.52 Very soft, organic liquids penetrate
2 2
polyethylene into polymer at ambient
temperature
I
Type 61 (CH CH )n 2–220 5 000–45 insoluble 1.52 Softens at 90°C
2 2
I J
Type 62 (CF CF )n 2–220 5 000–45 insoluble 1.52 Useful to 200°C for short durations
2 2
J B
Dia
...