Standard Guide for Interpreting Images of Polymeric Tissue Scaffolds

SIGNIFICANCE AND USE
4.1 This document provides guidance for users who wish to obtain quantifiable data from images of tissue scaffolds manufactured from polymers that include both high water content gels and woven textiles.  
4.2 Information derived from tissue scaffold images can be used to optimize the structural characteristics of the matrix for a particular application, to develop better manufacturing procedures or to provide a measure of quality assurance and product traceability. Fig. 1 provides a summary of the key stages of image capture and analysis.
FIG. 1 Key Stages in Image Capture, Storage, and Analysis  
4.3 There is a synergy between the analysis of pores in tissue scaffolds and that of particles that is reflected in standards cited and in the analysis described in Section 9. Guide E1919 provides a compendium of standards for particle analysis that includes measurement techniques, data analytical and sampling methodologies.
SCOPE
1.1 This guide covers the factors that need to be considered in obtaining and interpreting images of tissue scaffolds including technique selection, instrument resolution and image quality, quantification and sample preparation.  
1.2 The information in this guide is intended to be applicable to porous polymer-based tissue scaffolds, including naturally derived materials such as collagen. However, some materials (both synthetic and natural) may require unique or varied sample preparation methods that are not specifically covered in this guide.  
1.3 The values stated in SI units are to be regarded as standard. No other units of measurement are included in this standard.  
1.4 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety, health, and environmental practices and determine the applicability of regulatory limitations prior to use.  
1.5 This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.

General Information

Status
Published
Publication Date
31-Jul-2020
Current Stage
Ref Project

Relations

Buy Standard

Guide
ASTM F2603-06(2020) - Standard Guide for Interpreting Images of Polymeric Tissue Scaffolds
English language
10 pages
sale 15% off
Preview
sale 15% off
Preview

Standards Content (Sample)


This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the
Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.
Designation: F2603 − 06 (Reapproved 2020)
Standard Guide for
Interpreting Images of Polymeric Tissue Scaffolds
This standard is issued under the fixed designation F2603; the number immediately following the designation indicates the year of
original adoption or, in the case of revision, the year of last revision. A number in parentheses indicates the year of last reapproval. A
superscript epsilon (´) indicates an editorial change since the last revision or reapproval.
1. Scope F1854 Test Method for Stereological Evaluation of Porous
Coatings on Medical Implants
1.1 This guide covers the factors that need to be considered
F1877 Practice for Characterization of Particles
in obtaining and interpreting images of tissue scaffolds includ-
F2150 Guide for Characterization and Testing of Biomate-
ing technique selection, instrument resolution and image
rial Scaffolds Used in Regenerative Medicine and Tissue-
quality, quantification and sample preparation.
Engineered Medical Products
1.2 The information in this guide is intended to be appli-
F2450 Guide for Assessing Microstructure of Polymeric
cable to porous polymer-based tissue scaffolds, including
Scaffolds for Use in Tissue-Engineered Medical Products
naturally derived materials such as collagen. However, some
materials (both synthetic and natural) may require unique or
3. Terminology
varied sample preparation methods that are not specifically
3.1 Definitions:
covered in this guide.
3.1.1 aliasing, n—artifactual data that originates from an
1.3 The values stated in SI units are to be regarded as
insufficient sampling rate.
standard. No other units of measurement are included in this
3.1.2 biomaterial, n—a natural or synthetic material that is
standard.
suitable for introduction into living tissue especially as part of
1.4 This standard does not purport to address all of the
a medical device, such as an artificial heart valve or joint.
safety concerns, if any, associated with its use. It is the
3.1.3 blind (end) pore, n—a pore that is in contact with an
responsibility of the user of this standard to establish appro-
exposedinternalwallorsurfacethroughasingleorificesmaller
priate safety, health, and environmental practices and deter-
than the pore’s depth.
mine the applicability of regulatory limitations prior to use.
1.5 This international standard was developed in accor- 3.1.4 closed cell, n—void within a solid, lacking any con-
nectivity with an external surface. Synonym: closed pore.
dance with internationally recognized principles on standard-
ization established in the Decision on Principles for the
3.1.5 feret diameter, n—the mean value of the distance
Development of International Standards, Guides and Recom-
between pairs of parallel tangents to the periphery of a pore
mendations issued by the World Trade Organization Technical
(adapted from Practice F1877).
Barriers to Trade (TBT) Committee.
3.1.6 hydrogel, n—a water-based open network of polymer
chains that are cross-linked either chemically or through
2. Referenced Documents
crystalline junctions or by specific ionic interactions.
2.1 ASTM Standards:
3.1.7 irregular, adj—an irregular pore that cannot be de-
E1919 GuideforWorldwidePublishedStandardsRelatingto
scribed as round or spherical. A set of reference figures that
Particle and Spray Characterization (Withdrawn 2014)
define the nomenclature are given in Appendix X2. (Adapted
E2245 Test Method for Residual Strain Measurements of
from Practice F1877).
Thin, Reflecting Films Using an Optical Interferometer
3.1.8 Nyquist criterion—a criterion that states that a signal
must be sampled at a rate greater than or equal to twice its
This guide is under the jurisdiction of ASTM Committee F04 on Medical and
Surgical Materials and Devicesand is the direct responsibility of Subcommittee highest frequency component to avoid aliasing.
F04.42 on Biomaterials and Biomolecules for TEMPs.
3.1.9 permeability, n—a measure of fluid, particle, or gas
Current edition approved Aug. 1, 2020. Published August 2020. Originally
flow through an open pore structure.
approved in 2006. Last previous edition approved in 2012 as F2603 – 06 (2012).
DOI: 10.1520/F2603-06R20.
3.1.10 pixel, n—two-dimensional picture element.
For referenced ASTM standards, visit the ASTM website, www.astm.org, or
contact ASTM Customer Service at service@astm.org. For Annual Book of ASTM
3.1.11 polymer, n—a long chain molecule composed of
Standards volume information, refer to the standard’s Document Summary page on
monomers.
the ASTM website.
3.1.11.1 Discussion—A polymer may be a natural or syn-
The last approved version of this historical standard is referenced on
www.astm.org. thetic material.
Copyright © ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959. United States
F2603 − 06 (2020)
3.1.11.2 Discussion—Examples of polymers include colla-
gen and polycaprolactone.
3.1.12 pore, n—a liquid, fluid, or gas-filled externally con-
necting channel, void, or open space within an otherwise solid
or gelatinous material (for example, textile meshes composed
of many or single fibers (textile-based scaffolds), open cell
foams, (hydrogels). Synonyms: open pore, through pore.
3.1.13 porosity, n—property of a solid which contains an
inherent or induced network of channels and open spaces.
Porosity can be determined by measuring the ratio of pore
(void) volume to the apparent (total) volume of a porous
material and is commonly expressed as a percentage (Guide
F2150).
3.1.14 rectangular, adj—A pore that approximates a square
or rectangle in shape (derived from Practice F1877).
3.1.15 roundness (R), n—a measure of how closely an
object represents a circle (Practice F1877).
3.1.16 scaffold, n—a support, delivery vehicle, or matrix for
facilitating the migration, binding, or transport of cells or
bioactive molecules used to replace, repair, or regenerate
tissues. (Guide F2150).
FIG. 1 Key Stages in Image Capture, Storage, and Analysis
3.1.17 segmentation, n—a methodology for distinguishing
different regions (for example, pores and walls) within a tissue
analysis that includes measurement techniques, data analytical
scaffold image.
and sampling methodologies.
3.1.18 spherical pore, n—a pore with a generally spherical
5. Measurement Objectives
shape.
3.1.18.1 Discussion—A spherical pore appears round in a
5.1 Much of the research activity in tissue engineering is
photograph (Practice F1877).
focusedonthedevelopmentofsuitablematerialsandstructures
for optimal growth of a range of tissue types including
3.1.19 threshold, n—isolation of a range of grayscale values
cartilage, bone, and nerve. This requires a quantitative assess-
exhibited by one constituent within an image.
ment of the scaffold structure.
3.1.20 through pores, n—an inherent or induced network of
The key parameters that need to be determined are (1) the
voids or channels that permit flow of fluid from one side of the
overall level of porosity, (2) the pore size distribution, which
structure to the other.
can range from tens of nanometers to several hundred
3.1.21 tortuosity, n—a measure of the mean free path length
micrometres, and (3) the degree of interconnectivity and
of through pores relative to the sample thickness. Alternative
tortuosity of the pores.
definition: The squared ratio of the mean free path to the
6. Imaging Methods and Conditions
minimum possible path length.
6.1 There are many experimental ways of obtaining key
3.1.22 voxel, n—three-dimensional picture element.
scaffold physical parameters as described in Guide F2450.
When imaging and subsequent quantitative analysis is chosen
4. Significance and Use
as the method for determining these parameters, it is critical
4.1 This document provides guidance for users who wish to
that any image under consideration be a true representation of
obtain quantifiable data from images of tissue scaffolds manu-
the scaffold of interest. Some imaging methods require sample
factured from polymers that include both high water content
preparation.Somedonot.Whensamplepreparationisrequired
gels and woven textiles.
prior to imaging, care must be taken that the procedures do not
4.2 Information derived from tissue scaffold images can be
significantly alter the morphology of the scaffold. See Appen-
used to optimize the structural characteristics of the matrix for dix X1 for further information on sample preparation.
a particular application, to develop better manufacturing pro-
6.2 Images obtained using techniques such as light
cedures or to provide a measure of quality assurance and
microscopy, electron microscopy, and magnetic resonance
product traceability. Fig. 1 provides a summary of the key
imaging are two-dimensional (2-D) representations of a three-
stages of image capture and analysis.
dimensional (3-D) structure. These can be a planar or cross-
4.3 There is a synergy between the analysis of pores in sectional view with a relatively large depth of field or a series
tissue scaffolds and that of particles that is reflected in of physical or virtual 2-D slices, each with a small depth of
standards cited and in the analysis described in Section 9. field, that can be reassembled in a virtual environment to
Guide E1919 provides a compendium of standards for particle produce a 3-D mesostructure.
F2603 − 06 (2020)
6.3 There are limits to the extent an image (2-D or 3-D) can optical microscopy (1-3), optical coherence microscopy (4),
faithfully represent the physical artifacts that are influenced by MRI (5), and electron microscopies (6).
factors germane to the imaging method, such as spatial
6.6 The reconstruction of the mesostructure in 3-D from a
resolution and dynamic range, image contrast, and the signal-
series of 2-D images obtained from a sample that has been
to-noiseratio.Table1listssomeofthetechniquesavailablefor
physically sectioned requires considerably more effort than
producing images of porous structures, along with their con-
assembly of virtual sections produced by techniques that are
trast source, maximum demonstrated spatial resolution, and
able to focus on a plane within the sample. The virtual
typical dynamic range. Proper technique selection depends
approach is also less prone to sample distortion since it
both on the material properties of the scaffold (that is, optical
obviatestheneedforphysicalsectioningandregistrationerrors
methods cannot be used with opaque materials) the contrast
in the reassembly process. However, the techniques used to
available, and the target pore size range.
generate virtual 2-D images typically have limited penetration
6.4 The images generated by the techniques shown in Table depth.
1 cannot reproduce features smaller than the spatial resolution
6.7 Confocal microscopy (OCT), for example, has a pen-
of the method. Features that are faint, that is, those that do not
etration depth of approximately 100 µm, a value that depends
have significant contrast, or signal significantly above
on the wavelength of the light used and the amount of
background, will be resolved at length scales larger than the
scattering that occurs within the sample. Scanning acoustic
maximum resolution. Excessive contrast can also limit the
microscopy (SAM) can extend the penetration depth to ap-
penetration depth due to scattering effects. This is particularly
proximately 1 mm in polymer scaffolds albeit with a reduction
true of optical microscopies using differences in refractive
in image resolution.
index as the contrast mechanism. An appropriate level of
6.8 In general, using longer wavelength radiation to im-
contrast that can be established by experimentation is therefore
prove penetration of the radiation is accompanied by a reduc-
critical to high quality imaging.
tion in resolution.
6.5 Contrast can be enhanced by using exogenous agents,
such as florescence tags in optical microscopy and stains
7. Image Capture and Storage
containing heavy metal complexes in electron microscopies.
7.1 Image acquisition in this guide refers to the process of
Excessive contrast can be ameliorated in optical microscopies
capturing an image through digitization that is then stored for
by imbibing the structure with a fluid that has an index of
subsequent analysis. Care should be taken during this stage to
refraction similar to that of the solid making up the structure
avoid loss of fidelity by controllable factors that are not related
(this is termed “index-matching”). There are many excellent
to the methodology used to produce the image. These factors
resources describing factors influencing widefield and confocal
include the spatial sampling frequency of the detector system,
the dynamic range of analogue to digital (A/D) conversion,
segmenting (thresholding) operations (discussed in Section 9),
and both image compression and decompression.
7.2 Spatial sampling frequency and appropriate A/D con-
TABLE 1 Sources of Contrast and Techniques to Generate version are straightforward issues; the sampling frequency
Images of Tissue Scaffolds
should be at least twice the inverse spatial resolution, so as to
Generic method Contrast source Maximum resolution Physical slicing
fulfill the Nyquist criterion. Sampling at frequencies below this
(lateral/axial) required for 3D
will lead to the display of artifacts. Most image processing
imaging?
systems have anti-aliasing filters that remove frequencies
Widefield Optical Refractive index 1 µm/(10 µm) Y
greater than F /2 Hz, where F is the digital sampling rate. The
Microscopy Fluorescence
s s
Absorbance
A/D conversion should utilize a sufficient number of bits to
Confocal Optical Refractive index 0.5 µm/1 µm N
cover the dynamic range of the imaging / detector system.
Microscopy Fluorescence
Eight-bit conversion and recording is used for most common
Absorbance
Optical Coherence Refractive index 1 µm/1 µm N
imaging applications, resulting in images with 256 grayscale
Tomography (or
levels, where 0 corresponds to pure black and 255 to pure
Microscopy)
white respectively. If 8-bit conversion is used in a color (RGB)
Scanning Acoustic Acoustic 0.1 µm/0.1 µm N
Microscopy (SAM) impedance (depending on the
image there are 256 possible color combinations.
wavelength chosen)
Magnetic Nuclear spin 10 µm/10 µm N
NOTE 1—The gamut, or range of the grayscale reflects the image
Resonance
contrast.
Imaging (MRI)
X-ray Micro- Electron density 10 µm/10 µm N 7.3 It is important to record the minimum measurement
Computed
value (that is, the dimensions of a single pixel) when using
Tomography
digital capture or digitizing film-based images at all magn
...

Questions, Comments and Discussion

Ask us and Technical Secretary will try to provide an answer. You can facilitate discussion about the standard in here.