Standard Test Method for Evaluation of the Environmental Stability of Calcium Phosphate Granules, Fabricated Forms, and Coatings

SCOPE
1.1 This test method covers calcium phosphate materials intended for use in surgical implant applications.
1.2 Aspects of the biological response to calcium phosphate materials in soft tissue and bone have been reported from laboratory studies and clinical use (1-10).  
1.3 The requirements of this specification apply to calcium phosphate materials such as calcium hydroxyapatite (see Specification F 1185), beta-tricalcium phosphate (see Specification F 1088), and biphasic mixtures thereof with or without intentional addition of other minor components (10 %).  
1.4 The material(s) shall be representative of that produced for sale. It shall have been produced and processed under standard manufacturing conditions.
1.5 The materials may be in the form of powders, granules, fabricated forms or coatings; and may be porous, nonporous, textured, and other implantable topographical substrate form representative of the end-use product.
1.6 The calcium phosphate material may constitute the only material in a substrate or it may be one of multiple materials so long as all other materials present do not dissolve under the test conditions described in this test method.
1.7 This test method is limited to the laboratory evaluation of the dissolution rate of a calcium phosphate material. No correlation of the results to in vivo performance is implied.
1.8 The values stated in either SI units or inch-pound units are to be regarded separately as standard. The values stated in each system may not be exact equivalents; therefore, each system shall be used independently of the other. Combining values from the two systems may result in non-conformance with the standard.
1.9 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use.

General Information

Status
Historical
Publication Date
30-Sep-2008
Current Stage
Ref Project

Relations

Buy Standard

Standard
ASTM F1926/F1926M-08 - Standard Test Method for Evaluation of the Environmental Stability of Calcium Phosphate Granules, Fabricated Forms, and Coatings
English language
4 pages
sale 15% off
Preview
sale 15% off
Preview
Standard
REDLINE ASTM F1926/F1926M-08 - Standard Test Method for Evaluation of the Environmental Stability of Calcium Phosphate Granules, Fabricated Forms, and Coatings
English language
4 pages
sale 15% off
Preview
sale 15% off
Preview

Standards Content (Sample)

NOTICE: This standard has either been superseded and replaced by a new version or withdrawn.
Please contact ASTM International (www.astm.org) for the latest information.
Designation: F1926/F1926M – 08
Standard Test Method for
Evaluation of the Environmental Stability of Calcium
1
Phosphate Granules, Fabricated Forms, and Coatings
ThisstandardisissuedunderthefixeddesignationF1926/F1926M;thenumberimmediatelyfollowingthedesignationindicatestheyear
of original adoption or, in the case of revision, the year of last revision. A number in parentheses indicates the year of last reapproval.
A superscript epsilon (´) indicates an editorial change since the last revision or reapproval.
1. Scope responsibility of the user of this standard to establish appro-
priate safety and health practices and determine the applica-
1.1 This test method covers calcium phosphate materials
bility of regulatory limitations prior to use.
intended for use in surgical implant applications.
1.2 Aspects of the biological response to calcium phosphate
2. Referenced Documents
materials in soft tissue and bone have been reported from
3
2 2.1 ASTM Standards:
laboratory studies and clinical use (1-10).
E691 Practice for Conducting an Interlaboratory Study to
1.3 The requirements of this specification apply to calcium
Determine the Precision of a Test Method
phosphate materials such as calcium hydroxyapatite (see
F1088 Specification for Beta-Tricalcium Phosphate for Sur-
Specification F1185), beta-tricalcium phosphate (see Specifi-
gical Implantation
cation F1088), and biphasic mixtures thereof with or without
F1185 SpecificationforCompositionofHydroxylapatitefor
intentional addition of other minor components (<10 %).
Surgical Implants
1.4 The material(s) shall be representative of that produced
for sale. It shall have been produced and processed under
3. Terminology
standard manufacturing conditions.
3.1 Definitions of Terms Specific to This Standard:
1.5 The materials may be in the form of powders, granules,
3.1.1 calcium phosphate—anyoneofanumberofinorganic
fabricated forms or coatings; and may be porous, nonporous,
chemicalcompoundscontainingcalciumandphosphateionsas
textured, and other implantable topographical substrate form
its principal constituents.
representative of the end-use product.
3.1.2 coating—a layer of material mechanically or chemi-
1.6 The calcium phosphate material may constitute the only
cally adhering to the surface of a substrate.
material in a substrate or it may be one of multiple materials so
longasallothermaterialspresentdonotdissolveunderthetest
4. Dissolution Media
conditions described in this test method.
4.1 Water used for preparing reagents or dissolution media
1.7 This test method is limited to the laboratory evaluation
shall be degassed carbon dioxide free deionized or distilled
of the dissolution rate of a calcium phosphate material. No
++
water and have less than 0.1 ppm of residual Ca ion.
correlation of the results to in vivo performance is implied.
4.2 Unbuffered Water Media—Deionized or distilled water
1.8 The values stated in either SI units or inch-pound units
–5 –5 –5
containing 8 3 10 M NaCl, 8 3 10 M CaCl , and 5 3 10
2
are to be regarded separately as standard. The values stated in
MK (PO ).
3 4
each system may not be exact equivalents; therefore, each
4.3 pH 5.5 MES Buffer Media—1.0 M MES, [2-(N-
system shall be used independently of the other. Combining
morphplino)ethanesulfonic acid] having a pH of 5.5 at 37 6
values from the two systems may result in non-conformance
–5 –5
0.5°Candcontaining8 310 MNaCl,8 310 MCaCl ,and
2
with the standard.
–5
53 10 MK (PO ).
3 4
1.9 This standard does not purport to address all of the
4.3.1 A buffer concentration of 1.0 M will usually provide
safety concerns, if any, associated with its use. It is the
sufficient buffer capacity to keep the solution within 60.1 pH
units of the initial value. If this is not the case, the buffer
1
This test method is under the jurisdiction ofASTM Committee F04 onMedical
capacity should be adjusted accordingly.
and Surgical Materials and Devices and is the direct responsibility of Subcommittee
F04.13 on Ceramic Materials.
Current edition approved Oct. 1, 2008. Published October 2008. Originally
3
published in 1998. Last previous edition approved in 2003 as F1926 – 03. DOI: For referenced ASTM standards, visit the ASTM website, www.astm.org, or
10.1520/F1926_F1926M-08. contact ASTM Customer Service at service@astm.org. For Annual Book of ASTM
2
The boldface numbers given in parentheses refer to a list of references at the Standards volume information, refer to the standard’s Document Summary page on
end of the text. the ASTM website.
Copyright © ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959, United States.
1

---------------------- Page: 1 ----------------------
NOTICE: This standard has either been superseded and replaced by a new version or withdrawn.
Please contact ASTM International (www.astm
...

This document is not anASTM standard and is intended only to provide the user of anASTM standard an indication of what changes have been made to the previous version. Because
it may not be technically possible to adequately depict all changes accurately, ASTM recommends that users consult prior editions as appropriate. In all cases only the current version
of the standard as published by ASTM is to be considered the official document.
Designation:F1926–03 Designation: F 1926/F 1926M – 08
Standard Test Method for
Evaluation of the Environmental Stability of Calcium
1
Phosphate Granules, Fabricated Forms, and Coatings
This standard is issued under the fixed designation F 1926/F 1926M; the number immediately following the designation indicates the
year of original adoption or, in the case of revision, the year of last revision. A number in parentheses indicates the year of last
reapproval. A superscript epsilon (´) indicates an editorial change since the last revision or reapproval.
1. Scope
1.1 This test method covers calcium phosphate coatingsmaterials intended for use in surgical implant applications.
1.2 Aspectsofthebiologicalresponsetocalciumphosphatematerialsinsofttissueandbonehavebeenreportedfromlaboratory
2
studies and clinical use (1-10).
1.3 The requirements of this specification apply to calcium phosphate coatingsmaterials such as calcium hydroxyapatite (see
Specification F 1185), beta-tricalcium phosphate (see Specification F 1088), and biphasic mixtures thereof with or without
intentional addition of other minor components (<10 %).
1.4The coating(s) shall be representative of that produced for sale. It shall have been produced and processed under standard
manufacturing conditions.
1.5The coatings may be applied to porous, nonporous, textured, and other implantable topographical substrate forms
representative of the end-use product.
1.6The calcium phosphate coating may constitute the only coating on a substrate or be one multiple coatings.
1.7Thistestmethodislimitedtothelaboratoryevaluationofthedissolutionrateofacalciumphosphatecoatings.Nocorrelation
of the results to
1.4 The material(s) shall be representative of that produced for sale. It shall have been produced and processed under standard
manufacturing conditions.
1.5 The materials may be in the form of powders, granules, fabricated forms or coatings; and may be porous, nonporous,
textured, and other implantable topographical substrate form representative of the end-use product.
1.6 The calcium phosphate material may constitute the only material in a substrate or it may be one of multiple materials so
long as all other materials present do not dissolve under the test conditions described in this test method.
1.7 This test method is limited to the laboratory evaluation of the dissolution rate of a calcium phosphate material. No
correlation of the results to in vivo performance is implied.
1.8The values stated in both inch-pound and SI units are to be regarded separately as the standard. The values given in
parentheses are for information only.
1.8 The values stated in either SI units or inch-pound units are to be regarded separately as standard. The values stated in each
system may not be exact equivalents; therefore, each system shall be used independently of the other. Combining values from the
two systems may result in non-conformance with the standard.
1.9 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility
of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory
limitations prior to use.
2. Referenced Documents
3
2.1 ASTM Standards:
E 691 Practice for Conducting an Interlaboratory Study to Determine the Precision of a Test Method
F 1088 Specification for Beta-Tricalcium Phosphate for Surgical Implantation
F 1185 Specification for Composition of Ceramic Hydroxylapatite for Surgical ImplantationImplants
1
This test method is under the jurisdiction of ASTM Committee F04 on Medical and Surgical Materials and Devices and is the direct responsibility of Subcommittee
F04.13 on Ceramic Materials .
Current edition approved Apr. 10, 2003. Published May 2003. Originally published in 1998. Last previous edition approved in 1999 as F1926-99.
Current edition approved Oct. 1, 2008. Published October 2008. Originally published in 1998. Last previous edition approved in 2003 as F 1926 – 03.
2
The boldface numbers given in parentheses refer to a list of references at the end of the text.
3
For referencedASTM standards, visit theASTM website, www.astm.org, or contactASTM Customer Service at service@astm.org. For Annual Book of ASTM Standards
, Vol 14.02.volume information, refer to the standard’s Document Summary page on the ASTM website.
Copyright © ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959, United States.
1

---------------------- Page: 1 ----------------------
F 1926
...

Questions, Comments and Discussion

Ask us and Technical Secretary will try to provide an answer. You can facilitate discussion about the standard in here.