ASTM E976-10
(Guide)Standard Guide for Determining the Reproducibility of Acoustic Emission Sensor Response
Standard Guide for Determining the Reproducibility of Acoustic Emission Sensor Response
SIGNIFICANCE AND USE
Acoustic emission data is affected by several characteristics of the instrumentation. The most obvious of these is the system sensitivity. Of all the parameters and components contributing to the sensitivity, the acoustic emission sensor is the one most subject to variation. This variation can be a result of damage or aging, or there can be variations between nominally identical sensors. To detect such variations, it is desirable to have a method for measuring the response of a sensor to an acoustic wave. Specific purposes for checking sensors include: (1) checking the stability of its response with time; (2) checking the sensor for possible damage after accident or abuse; (3) comparing a number of sensors for use in a multichannel system to ensure that their responses are adequately matched; and (4) checking the response after thermal cycling or exposure to a hostile environment. It is very important that the sensor characteristics be always measured with the same sensor cable length and impedance as well as the same preamplifier or equivalent. This guide presents several procedures for measuring sensor response. Some of these procedures require a minimum of special equipment.
It is not the intent of this guide to evaluate AE system performance. Refer to Practice E750 for characterizing acoustic instrumentation and refer to Guide E2374 for AE system performance verification.
The procedures given in this guide are designed to measure the response of an acoustic emission sensor to an arbitrary but repeatable acoustic wave. These procedures in no way constitute a calibration of the sensor. The absolute calibration of a sensor requires a complete knowledge of the characteristics of the acoustic wave exciting the sensor or a previously calibrated reference sensor. In either case, such a calibration is beyond the scope of this guide.
The fundamental requirement for comparing sensor responses is a source of repeatable acoustic waves. The characteristics of the wave...
SCOPE
1.1 This guide defines simple economical procedures for testing or comparing the performance of acoustic emission sensors. These procedures allow the user to check for degradation of a sensor or to select sets of sensors with nearly identical performances. The procedures are not capable of providing an absolute calibration of the sensor nor do they assure transferability of data sets between organizations.
1.2 Units—The values stated in SI units are to be regarded as standard. No other units of measurement are included in this standard.
1.3 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use.
General Information
Relations
Buy Standard
Standards Content (Sample)
NOTICE: This standard has either been superseded and replaced by a new version or withdrawn.
Contact ASTM International (www.astm.org) for the latest information
Designation: E976 − 10
StandardGuide for
Determining the Reproducibility of Acoustic Emission
1
Sensor Response
This standard is issued under the fixed designation E976; the number immediately following the designation indicates the year of
original adoption or, in the case of revision, the year of last revision. A number in parentheses indicates the year of last reapproval. A
superscript epsilon (´) indicates an editorial change since the last revision or reapproval.
1. Scope of damage or aging, or there can be variations between
nominally identical sensors. To detect such variations, it is
1.1 This guide defines simple economical procedures for
desirable to have a method for measuring the response of a
testing or comparing the performance of acoustic emission
sensor to an acoustic wave. Specific purposes for checking
sensors. These procedures allow the user to check for degra-
sensors include: (1) checking the stability of its response with
dation of a sensor or to select sets of sensors with nearly
time; (2) checking the sensor for possible damage after
identical performances. The procedures are not capable of
accident or abuse; (3) comparing a number of sensors for use
providing an absolute calibration of the sensor nor do they
in a multichannel system to ensure that their responses are
assure transferability of data sets between organizations.
adequately matched; and (4) checking the response after
1.2 Units—The values stated in SI units are to be regarded
thermal cycling or exposure to a hostile environment. It is very
as standard. No other units of measurement are included in this
important that the sensor characteristics be always measured
standard.
with the same sensor cable length and impedance as well as the
1.3 This standard does not purport to address all of the
same preamplifier or equivalent. This guide presents several
safety concerns, if any, associated with its use. It is the procedures for measuring sensor response. Some of these
responsibility of the user of this standard to establish appro-
procedures require a minimum of special equipment.
priate safety and health practices and determine the applica-
3.2 It is not the intent of this guide to evaluate AE system
bility of regulatory limitations prior to use.
performance. Refer to Practice E750 for characterizing acous-
tic instrumentation and refer to Guide E2374 for AE system
2. Referenced Documents
performance verification.
2
2.1 ASTM Standards:
3.3 The procedures given in this guide are designed to
E750 Practice for Characterizing Acoustic Emission Instru-
measure the response of an acoustic emission sensor to an
mentation
arbitrary but repeatable acoustic wave. These procedures in no
E2075 Practice for Verifying the Consistency of AE-Sensor
way constitute a calibration of the sensor. The absolute
Response Using an Acrylic Rod
calibration of a sensor requires a complete knowledge of the
E2374 Guide for Acoustic Emission System Performance
characteristics of the acoustic wave exciting the sensor or a
Verification
previously calibrated reference sensor. In either case, such a
3. Significance and Use calibration is beyond the scope of this guide.
3.1 Acoustic emission data is affected by several character-
3.4 The fundamental requirement for comparing sensor
istics of the instrumentation. The most obvious of these is the
responses is a source of repeatable acoustic waves. The
system sensitivity. Of all the parameters and components
characteristics of the wave do not need to be known as long as
contributing to the sensitivity, the acoustic emission sensor is
the wave can be reproduced at will. The sources and geom-
the one most subject to variation. This variation can be a result
etries given in this guide will produce primarily compressional
waves. While the sensors will respond differently to different
types of waves, changes in the response to one type of wave
1
This guide is under the jurisdiction of ASTM Committee E07 on Nondestruc-
will imply changes in the responses to other types of waves.
tive Testing and is the direct responsibility of Subcommittee E07.04 on Acoustic
Emission Method.
3.5 These procedures use a test block or rod. Such a device
CurrenteditionapprovedJune1,2010.PublishedJuly2010.Originallyapproved
provides a convenient mounting surface for the sensor and
in1984.Lastpreviouseditionapprovedin2005asE976 - 05.DOI:10.1520/E0976-
10.
when appropriately marked, can ensure that the source and the
2
For referenced ASTM standards, visit the ASTM website, www.astm.org, or
sensor are always positioned identically with respect to each
contact ASTM Customer Service at service@astm.org. For Annual Book of ASTM
other. The device or rod also provides mechanical loading of
Standards volume information, refer to the standard’s Document Summary page on
the ASTM website. the sensor
...
This document is not an ASTM standard and is intended only to provide the user of an ASTM standard an indication of what changes have been made to the previous version. Because
it may not be technically possible to adequately depict all changes accurately, ASTM recommends that users consult prior editions as appropriate. In all cases only the current version
of the standard as published by ASTM is to be considered the official document.
Designation:E976–05 Designation:E976–10
Standard Guide for
Determining the Reproducibility of Acoustic Emission
1
Sensor Response
This standard is issued under the fixed designation E976; the number immediately following the designation indicates the year of
original adoption or, in the case of revision, the year of last revision. A number in parentheses indicates the year of last reapproval. A
superscript epsilon (´) indicates an editorial change since the last revision or reapproval.
1. Scope
1.1 This guide defines simple economical procedures for testing or comparing the performance of acoustic emission sensors.
Theseproceduresallowtheusertocheckfordegradationofasensorortoselectsetsofsensorswithnearlyidenticalperformances.
The procedures are not capable of providing an absolute calibration of the sensor nor do they assure transferability of data sets
between organizations.
1.2 Units—The values stated in SI units are to be regarded as standard. No other units of measurement are included in this
standard.
1.3 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility
of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory
limitations prior to use.
2. Referenced Documents
2
2.1 ASTM Standards:
E750 Practice for Characterizing Acoustic Emission Instrumentation
E2075 Practice for Verifying the Consistency of AE-Sensor Response Using an Acrylic Rod
E2374 Guide for Acoustic Emission System Performance Verification
3. Significance and Use
3.1 Acoustic emission data is affected by several characteristics of the instrumentation.The most obvious of these is the system
sensitivity. Of all the parameters and components contributing to the sensitivity, the acoustic emission sensor is the one most
subject to variation. This variation can be a result of damage or aging, or there can be variations between nominally identical
sensors. To detect such variations, it is desirable to have a method for measuring the response of a sensor to an acoustic wave.
Specific purposes for checking sensors include: (1) checking the stability of its response with time; (2) checking the sensor for
possible damage after accident or abuse; (3) comparing a number of sensors for use in a multichannel system to ensure that their
responses are adequately matched; and (4) checking the response after thermal cycling or exposure to a hostile environment. It is
very important that the sensor characteristics be always measured with the same sensor cable length and impedance as well as the
same preamplifier or equivalent. This guide presents several procedures for measuring sensor response. Some of these procedures
require a minimum of special equipment.
3.2 It is not the intent of this guide to evaluate AE system performance. Refer to Practice E750 for characterizing acoustic
instrumentation and refer to Guide E2374 for AE system performance verification.
3.3 The procedures given in this guide are designed to measure the response of an acoustic emission sensor to an arbitrary but
repeatable acoustic wave. These procedures in no way constitute a calibration of the sensor. The absolute calibration of a sensor
requires a complete knowledge of the characteristics of the acoustic wave exciting the sensor or a previously calibrated reference
sensor. In either case, such a calibration is beyond the scope of this guide.
3.4 The fundamental requirement for comparing sensor responses is a source of repeatable acoustic waves. The characteristics
of the wave do not need to be known as long as the wave can be reproduced at will.The sources and geometries given in this guide
will produce primarily compressional waves. While the sensors will respond differently to different types of waves, changes in the
response to one type of wave will imply changes in the responses to other types of waves.
1
This guide is under the jurisdiction ofASTM Committee E07 on Nondestructive Testing and is the direct responsibility of Subcommittee E07.04 onAcoustic Emission.
Current edition approved Dec. 1, 2005. Published December 2005. Originally approved in 1984. Last previous edition approved in 2000 as E976-00. DOI:
10.1520/E0976-05.on Acoustic Emission Method.
Current edition approved June 1, 2010. Published July 2010. Originally approved in 1984. Last previous edition approved in 2005 as E976 - 05. DOI: 10.1520/E0976-10.
2
For referencedASTM standards, visit theASTM website, www.astm.org, or contactASTM Customer Ser
...
This document is not an ASTM standard and is intended only to provide the user of an ASTM standard an indication of what changes have been made to the previous version. Because
it may not be technically possible to adequately depict all changes accurately, ASTM recommends that users consult prior editions as appropriate. In all cases only the current version
of the standard as published by ASTM is to be considered the official document.
Designation:E976–05 Designation:E976–10
Standard Guide for
Determining the Reproducibility of Acoustic Emission
1
Sensor Response
This standard is issued under the fixed designation E976; the number immediately following the designation indicates the year of
original adoption or, in the case of revision, the year of last revision. A number in parentheses indicates the year of last reapproval. A
superscript epsilon (´) indicates an editorial change since the last revision or reapproval.
1. Scope
1.1 This guide defines simple economical procedures for testing or comparing the performance of acoustic emission sensors.
Theseproceduresallowtheusertocheckfordegradationofasensorortoselectsetsofsensorswithnearlyidenticalperformances.
The procedures are not capable of providing an absolute calibration of the sensor nor do they assure transferability of data sets
between organizations.
1.2 Units—The values stated in SI units are to be regarded as standard. No other units of measurement are included in this
standard.
1.3 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility
of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory
limitations prior to use.
2. Referenced Documents
2
2.1 ASTM Standards:
E750 Practice for Characterizing Acoustic Emission Instrumentation
E2075 Practice for Verifying the Consistency of AE-Sensor Response Using an Acrylic Rod
E2374 Guide for Acoustic Emission System Performance Verification
3. Significance and Use
3.1 Acoustic emission data is affected by several characteristics of the instrumentation.The most obvious of these is the system
sensitivity. Of all the parameters and components contributing to the sensitivity, the acoustic emission sensor is the one most
subject to variation. This variation can be a result of damage or aging, or there can be variations between nominally identical
sensors. To detect such variations, it is desirable to have a method for measuring the response of a sensor to an acoustic wave.
Specific purposes for checking sensors include: (1) checking the stability of its response with time; (2) checking the sensor for
possible damage after accident or abuse; (3) comparing a number of sensors for use in a multichannel system to ensure that their
responses are adequately matched; and (4) checking the response after thermal cycling or exposure to a hostile environment. It is
very important that the sensor characteristics be always measured with the same sensor cable length and impedance as well as the
same preamplifier or equivalent. This guide presents several procedures for measuring sensor response. Some of these procedures
require a minimum of special equipment.
3.2 It is not the intent of this guide to evaluate AE system performance. Refer to Practice E750 for characterizing acoustic
instrumentation and refer to Guide E2374 for AE system performance verification.
3.3 The procedures given in this guide are designed to measure the response of an acoustic emission sensor to an arbitrary but
repeatable acoustic wave. These procedures in no way constitute a calibration of the sensor. The absolute calibration of a sensor
requires a complete knowledge of the characteristics of the acoustic wave exciting the sensor or a previously calibrated reference
sensor. In either case, such a calibration is beyond the scope of this guide.
3.4 The fundamental requirement for comparing sensor responses is a source of repeatable acoustic waves. The characteristics
of the wave do not need to be known as long as the wave can be reproduced at will.The sources and geometries given in this guide
will produce primarily compressional waves. While the sensors will respond differently to different types of waves, changes in the
response to one type of wave will imply changes in the responses to other types of waves.
1
This guide is under the jurisdiction ofASTM Committee E07 on Nondestructive Testing and is the direct responsibility of Subcommittee E07.04 onAcoustic Emission.
Current edition approved Dec. 1, 2005. Published December 2005. Originally approved in 1984. Last previous edition approved in 2000 as E976-00. DOI:
10.1520/E0976-05.on Acoustic Emission Method.
Current edition approved June 1, 2010. Published July 2010. Originally approved in 1984. Last previous edition approved in 2005 as E976 - 05. DOI: 10.1520/E0976-10.
2
For referencedASTM standards, visit theASTM website, www.astm.org, or contactASTM Customer Ser
...
Questions, Comments and Discussion
Ask us and Technical Secretary will try to provide an answer. You can facilitate discussion about the standard in here.