ASTM F2068-03
(Specification)Standard Specification for Femoral Prostheses—Metallic Implants
Standard Specification for Femoral Prostheses—Metallic Implants
ABSTRACT
This specification covers metallic stemmed femoral prostheses used to replace the natural hip joint by means of hemi-arthroplasty or total hip surgical procedures. Prostheses for hemi-arthroplasty are intended to articulate with the natural acetabulum of the patient. Prostheses for total hip replacement are intended to articulate with prosthetic acetabular cups. Prostheses may have integral femoral heads or cones designed to accept modular heads. Femoral prostheses presented are defined as follows: type IA - single-piece (mono-block), metallic femoral total hip or hemi-arthroplasty hip prosthesis with an integral stem, neck and head, type IB - single-piece (mono-block), metallic, femoral total hip or hemi- arthroplasty hip prostheses with an integral stem, neck, and head, type IIA - modular metallic femoral hip prostheses that could include a modular (type II) head or other modular components, or both, and type IIB - Modular metallic femoral hip prosthesis that could include a modular (type II) head or other modular components, or both. Femoral prostheses shall be capable of withstanding normal static and dynamic loading in the physiological range without overload fracture, plastic deformation, or fatigue fracture. Shear strength, tensile strength, and abrasion resistance of plasma spray thermal coatings shall be tested to meet the requirements prescribed.
SCOPE
1.1 This specification covers metallic stemmed femoral prostheses used to replace the natural hip joint by means of hemi-arthroplasty or total hip surgical procedures. Prostheses for hemi-arthroplasty are intended to articulate with the natural acetabulum of the patient. Prostheses for total hip replacement are intended to articulate with prosthetic acetabular cups. Prostheses may have integral femoral heads or cones designed to accept modular heads.
1.2 Modular femoral heads, which may be affixed to cones on implants covered by this specification, are not covered by this specification. The mechanical strength, corrosion resistance, and biocompatibility of the head portions of one-piece integral implants are covered by this specification.
1.3 Femoral prostheses included within the scope of this specification are intended for fixation by press fit between the prosthesis and host bone, the use of bone cement, or through the ingrowth of host bone into a porous coating.
1.4 Custom femoral prostheses, designed explicitly for a single patient, are not covered within the scope of this specification.
1.5 Prostheses incorporating nonmetallic (for example, polymer composite) implants, nonporous bioactive ceramic coatings, or porous-polymer coatings, are specifically excluded from the scope of this specification.
1.6 The requirements for modular connections of multicomponent modular femoral hip prostheses are not covered by this specification.
1.7 The values stated in SI units are to be regarded as the standard.
General Information
Relations
Standards Content (Sample)
NOTICE: This standard has either been superseded and replaced by a new version or withdrawn.
Contact ASTM International (www.astm.org) for the latest information
Designation: F 2068 – 03
Standard Specification for
1
Femoral Prostheses—Metallic Implants
This standard is issued under the fixed designation F 2068; the number immediately following the designation indicates the year of
original adoption or, in the case of revision, the year of last revision. A number in parentheses indicates the year of last reapproval. A
superscript epsilon (e) indicates an editorial change since the last revision or reapproval.
1. Scope F 75 Specification for Cobalt-28Chromium-6Molybdenum
Alloy Castings and Casting Alloy for Surgical Implants
1.1 This specification covers metallic stemmed femoral
2
(UNS R30075)
prostheses used to replace the natural hip joint by means of
F 86 Practice for Surface Preparation and Marking of Me-
hemi-arthroplasty or total hip surgical procedures. Prostheses
2
tallic Surgical Implants
for hemi-arthroplasty are intended to articulate with the natural
F 90 Specification for Wrought Cobalt-20Chromium-
acetabulum of the patient. Prostheses for total hip replacement
15Tungsten-10Nickel Alloy for Surgical Implant Applica-
are intended to articulate with prosthetic acetabular cups.
2
tions (UNS R30605)
Prostheses may have integral femoral heads or cones designed
F 136 Specification for Wrought Titanium-6Aluminum-
to accept modular heads.
4Vanadium ELI (Extra-Low Interstitial)Alloy for Surgical
1.2 Modular femoral heads, which may be affixed to cones
2
Implant Applications (UNS R56401)
on implants covered by this specification, are not covered by
F 138 Specification for Wrought 18Chromium-14Nickel-
this specification. The mechanical strength, corrosion resis-
2.5Molybdenum Stainless Steel Bar and Wire for Surgical
tance, and biocompatibility of the head portions of one-piece
2
Implants (UNS S31673)
integral implants are covered by this specification.
F 562 Specification for Wrought Cobalt-35Nickel-
1.3 Femoral prostheses included within the scope of this
20Chromium-10Molybdenum Alloy for Surgical Implant
specification are intended for fixation by press fit between the
2
Applications (UNS R 30035)
prosthesis and host bone, the use of bone cement, or through
F 563 Specification for Wrought Cobalt-20Nickel-
the ingrowth of host bone into a porous coating.
20Chromium-3.5Molybdenum-3.5Tungsten-5Iron Alloy
1.4 Custom femoral prostheses, designed explicitly for a
2
for Surgical Implant Applications (UNS R30563)
single patient, are not covered within the scope of this
F 620 Specification for Alpha Plus Beta Titanium Alloy
specification.
2
Forgings for Surgical Implants
1.5 Prostheses incorporating nonmetallic (for example,
F 745 Specification for 18Chromium-12.5Nickel-
polymer composite) implants, nonporous bioactive ceramic
2.5Molybdenum Stainless Steel for Cast and Solution-
coatings,orporous-polymercoatings,arespecificallyexcluded
2
Annealed Surgical Implant Applications
from the scope of this specification.
F 746 Test Method for Pitting or Crevice Corrosion of
1.6 The requirements for modular connections of multicom-
2
Metallic Surgical Implant Materials
ponent modular femoral hip prostheses are not covered by this
F 748 Practice for Selecting Generic Biological Test Meth-
specification.
2
ods for Materials and Devices
1.7 The values stated in SI units are to be regarded as the
F 799 SpecificationforCobalt-28Chromium-6Molybdenum
standard.
Alloy Forgings for Surgical Implants (UNS R31537,
2
2. Referenced Documents
R31538, R31539)
F 981 Practice for Assessment of Compatibility of Bioma-
2.1 ASTM Standards:
terials for Surgical Implants with Respect to Effect of
F 67 Specification for Unalloyed Titanium for Surgical
2
Materials on Muscle and Bone
Implant Applications (UNS R50250, UNS R 50400, UNS
2
F 983 Practice for Permanent Marking of Orthopaedic Im-
R 50550, UNS R50700)
2
plant Components
F 1044 Test Method for Shear Testing of Calcium Phos-
1
2
This specification is under the jurisdiction of ASTM Committee F04 on
phate Coatings and Metallic Coatings
Medical and Surgical Materials and Devices and is the direct responsibility of
F 1108 Specification for Titanium-6Aluminum-4Vanadium
Subcommittee F04.22 on Arthroplasty.
2
Alloy Castings for Surgical Implants (UNS R56406)
Current edition approved Sept. 10, 2003. Published October 2003. Originally
e1
approved in 2000. Last previous edition approved in 2001 as F 2068 – 01 .
2
Annual Book of ASTM Standards, Vol 13.01.
Copyright © ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959, United States.
1
---------------------- Page: 1 ----------------------
F2068–03
F 1147 Test Method for Tension Testing of Calcium Phos- 3. Terminology
2
phate and Metallic Coatings
3.1 Definitions of Terms Specific to This Standard:
F 1440 Practice for Cyclic Fatigue Testing of Metallic
3.1.1 bore, n—an internal cavity, in the form of a truncated
Stemmed Hip Arthroplasty
...
Questions, Comments and Discussion
Ask us and Technical Secretary will try to provide an answer. You can facilitate discussion about the standard in here.