ASTM F2503-20
(Practice)Standard Practice for Marking Medical Devices and Other Items for Safety in the Magnetic Resonance Environment
Standard Practice for Marking Medical Devices and Other Items for Safety in the Magnetic Resonance Environment
SIGNIFICANCE AND USE
4.1 Interactions of medical devices and other items with the MR environment has resulted in serious injuries and death of patients and other individuals. Additionally, hazards stemming from equipment malfunction are of concern. Section 4.2 lists possible direct and indirect causes of hazards in the MR environment.
4.2 Potential direct and indirect causes of hazards:
4.2.1 Direct causes:
4.2.1.1 mechanical causes, including magnetically induced displacement force, torque, and vibration
4.2.1.2 electromagnetic causes, including induction (heating, stimulation) and discharge (spark gap)
4.2.1.3 acoustic causes
4.2.2 Indirect causes:
4.2.2.1 malfunction of items, for example of vital components such as valves, monitors and pumps
4.3 This practice provides a uniform system for marking to indicate the conditions for which it has been determined that a medical device or other item may be safely placed and used in the MR environment. It provides simple visual icons and terms which are intended to reduce injuries and other mishaps that occur when items that pose hazards in the MR environment are brought into the MR environment.
SCOPE
1.1 This international standard applies to the practice of marking of items that might be used in the magnetic resonance (MR) environment.
1.2 The purpose of this practice is to mark items that might be brought into the MR environment and to recommend information that should be included in the marking.
1.3 The standard specifies the permanent marking of items, which are used in an MR environment, by means of terms and icons.
1.4 MR image artifacts are not considered to be a performance issue and so are not addressed in this international standard practice (see X1.5).
1.5 The values stated in SI units are to be regarded as standard. No other units of measurement are included in this standard.
1.6 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety, health, and environmental practices and determine the applicability of regulatory limitations prior to use.
1.7 This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.
General Information
Buy Standard
Standards Content (Sample)
NOTICE: This standard has either been superseded and replaced by a new version or withdrawn.
Contact ASTM International (www.astm.org) for the latest information
Designation: F2503 − 20
Standard Practice for
Marking Medical Devices and Other Items for Safety in the
1
Magnetic Resonance Environment
This standard is issued under the fixed designation F2503; the number immediately following the designation indicates the year of
original adoption or, in the case of revision, the year of last revision. A number in parentheses indicates the year of last reapproval. A
superscript epsilon (´) indicates an editorial change since the last revision or reapproval.
2
1. Scope 2.2 ASTM Standards:
F2052 Test Method for Measurement of Magnetically In-
1.1 This international standard applies to the practice of
duced Displacement Force on Medical Devices in the
marking of items that might be used in the magnetic resonance
Magnetic Resonance Environment
(MR) environment.
F2119 Test Method for Evaluation of MR Image Artifacts
1.2 The purpose of this practice is to mark items that might
from Passive Implants
be brought into the MR environment and to recommend
F2182 Test Method for Measurement of Radio Frequency
information that should be included in the marking.
Induced Heating On or Near Passive Implants During
Magnetic Resonance Imaging
1.3 The standard specifies the permanent marking of items,
which are used in an MR environment, by means of terms and F2213 Test Method for Measurement of Magnetically In-
duced Torque on Medical Devices in the Magnetic Reso-
icons.
nance Environment
1.4 MR image artifacts are not considered to be a perfor-
2.3 Other Standards:
mance issue and so are not addressed in this international
IEC 60601-2-33 Medical Electrical Equipment—Part 2-33:
standard practice (see X1.5).
Particular Requirements for the Safety of Magnetic Reso-
1.5 The values stated in SI units are to be regarded as 3
nance Equipment for Medical Diagnosis
standard. No other units of measurement are included in this
ISO 14971 Medical Devices—Application of Risk Manage-
standard. 4
ment to Medical Devices
1.6 This standard does not purport to address all of the
ISO/IEC Guide 51 Safety Aspects—Guidelines for their
4
safety concerns, if any, associated with its use. It is the
Inclusion in Standards
responsibility of the user of this standard to establish appro-
ISO TS 10974 Assessment of the Safety of Magnetic Reso-
priate safety, health, and environmental practices and deter-
nance Imaging for Patients with an Active Implantable
4
mine the applicability of regulatory limitations prior to use.
Medical Device
1.7 This international standard was developed in accor-
3. Terminology
dance with internationally recognized principles on standard-
ization established in the Decision on Principles for the
3.1 Definitions:
Development of International Standards, Guides and Recom-
3.1.1 harmful interaction—unintended direct or indirect in-
mendations issued by the World Trade Organization Technical
teraction of items with MR equipment, especially with the
Barriers to Trade (TBT) Committee.
static magnetic field, the gradient fields and the RF fields of the
MR equipment, that can pose hazards to patients or other
2. Referenced Documents
persons.
2.1 The following referenced documents are indispensable
3.1.1.1 Discussion—In this context, the affected image qual-
for the application of this document. For dated references, only
ity or image artifacts are not considered to be a harmful
the edition cited applies. For undated references, the latest
interaction.
edition of the referenced document (including any amend-
ments) applies.
2
For referenced ASTM standards, visit the ASTM website, www.astm.org, or
contact ASTM Customer Service at service@astm.org. For Annual Book of ASTM
1
This practice is under the jurisdiction of ASTM Committee F04 on Medical and Standards volume information, refer to the standard’s Document Summary page on
Surgical Materials and Devices and is the direct responsibility of Subcommittee the ASTM website.
3
F04.15 on Material Test Methods. Available from International Electrotechnical Commission (IEC), 3, rue de
Current edition approved Feb. 1, 2020. Published February 2020. Originally Varembé, P.O. Box 131, CH-1211 Geneva 20, Switzerland, http://www.iec.ch.
4
approved in 2005. Last previous edition approved in 2013 as F2503–13. DOI: Available from American National Standards Institute (ANSI), 25 W. 43rd St.,
10.1520/F2503-20. 4th Floor, New York, NY 10036, http://www.ansi.org.
Copyright © ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959. United States
1
---------------------- Page: 1 ----------------------
F2503 − 20
3.1.2 hazard—potential source of harm. ISO/IEC Guide 51 via additional language the conditions in which an item can be
used safely within the MR environment.
3.1.3 item—object that might be brought into the MR
3.1.13 MR Safe—an item that poses no known hazards
environme
...
This document is not an ASTM standard and is intended only to provide the user of an ASTM standard an indication of what changes have been made to the previous version. Because
it may not be technically possible to adequately depict all changes accurately, ASTM recommends that users consult prior editions as appropriate. In all cases only the current version
of the standard as published by ASTM is to be considered the official document.
Designation: F2503 − 13 F2503 − 20
Standard Practice for
Marking Medical Devices and Other Items for Safety in the
1
Magnetic Resonance Environment
This standard is issued under the fixed designation F2503; the number immediately following the designation indicates the year of
original adoption or, in the case of revision, the year of last revision. A number in parentheses indicates the year of last reapproval. A
superscript epsilon (´) indicates an editorial change since the last revision or reapproval.
1. Scope
1.1 This international standard applies to the practice of marking of items that might be used in the magnetic resonance (MR)
environment.
1.2 The purpose of this practice is to mark items that might be brought into the MR environment and to recommend information
that should be included in the marking.
1.3 The standard specifies the permanent marking of items, which are used in an MR environment, by means of terms and icons.
1.4 MR image artifacts are not considered to be a performance issue and so are not addressed in this international standard
practice (see X1.5).
1.5 The values stated in SI units are to be regarded as standard. No other units of measurement are included in this standard.
1.6 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility
of the user of this standard to establish appropriate safety safety, health, and healthenvironmental practices and determine the
applicability of regulatory limitations prior to use.
1.7 This international standard was developed in accordance with internationally recognized principles on standardization
established in the Decision on Principles for the Development of International Standards, Guides and Recommendations issued
by the World Trade Organization Technical Barriers to Trade (TBT) Committee.
2. Referenced Documents
2.1 The following referenced documents are indispensable for the application of this document. For dated references, only the
edition cited applies. For undated references, the latest edition of the referenced document (including any amendments) applies.
2
2.2 ASTM Standards:
F2052 Test Method for Measurement of Magnetically Induced Displacement Force on Medical Devices in the Magnetic
Resonance Environment
F2119 Test Method for Evaluation of MR Image Artifacts from Passive Implants
F2182 Test Method for Measurement of Radio Frequency Induced Heating On or Near Passive Implants During Magnetic
Resonance Imaging
F2213 Test Method for Measurement of Magnetically Induced Torque on Medical Devices in the Magnetic Resonance
Environment
2.3 Other Standards:
IEC 60601-2-33 Medical Electrical Equipment—Part 2-33: Particular Requirements for the Safety of Magnetic Resonance
3
Equipment for Medical Diagnosis
4
ISO 14971 Medical Devices—Application of Risk Management to Medical Devices
4
ISO/IEC Guide 51 Safety Aspects—Guidelines for their Inclusion in Standards
1
This practice is under the jurisdiction of ASTM Committee F04 on Medical and Surgical Materials and Devices and is the direct responsibility of Subcommittee F04.15
on Material Test Methods.
Current edition approved June 1, 2013Feb. 1, 2020. Published July 2013February 2020. Originally approved in 2005. Last previous edition approved in 20082013 as
F2503–08.–13. DOI: 10.1520/F2503-13.10.1520/F2503-20.
2
For referenced ASTM standards, visit the ASTM website, www.astm.org, or contact ASTM Customer Service at service@astm.org. For Annual Book of ASTM Standards
volume information, refer to the standard’s Document Summary page on the ASTM website.
3
Available from International Electrotechnical Commission (IEC), 3, rue de Varembé, P.O. Box 131, CH-1211 Geneva 20, Switzerland, http://www.iec.ch.
4
Available from American National Standards Institute (ANSI), 25 W. 43rd St., 4th Floor, New York, NY 10036, http://www.ansi.org.
Copyright © ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959. United States
1
---------------------- Page: 1 ----------------------
F2503 − 20
ISO TS 10974 Assessment of the Safety of Magnetic Resonance Imaging for Patients with an Active Implantable Medical
4
Device
3. Terminology
3.1 Definitions:
3.1.1 harmful interaction—unintended direct or indirect interaction of items with MR equipment, especially with the static
magnetic field, the gradient fields and the RF fields of the MR equipment, that can pose hazards to patients or other persons.
NOTE 1—In this context, the affected image quality or image artifacts are not considere
...
Questions, Comments and Discussion
Ask us and Technical Secretary will try to provide an answer. You can facilitate discussion about the standard in here.