Standard Test Method for Platelet Leukocyte Count&mdash;An <emph type="bdit">In-Vitro</emph > Measure for Hemocompatibility Assessment of Cardiovascular Materials

SIGNIFICANCE AND USE
4.1 The purpose of this test method is to determine if medical materials exposed to human whole blood will adversely affect the platelet and leukocyte counts in whole blood. A large number of platelets and leukocytes as part of thrombi adhering to the material will be reflected by a decrease in their counts in blood. Thrombogenic materials should not be used for cardiovascular medical devices, unless the purpose of the device is to promote thrombosis.
SCOPE
1.1 This test method assists in the evaluation of cardiovascular device materials for their ability to induce thrombus formation. Thrombus formation is assessed by means of a reduction in human platelets and leukocytes when consumed by thrombus after activation on the material surface. This assay may be part of the hemocompatibility evaluation for devices and materials contacting human blood, as per ANSI/AAMI/ISO 10993–4. See also Test Method F2382.  
1.2 All safety policies and practices shall be observed during the performance of this test method. All human blood and any materials that had contact with human blood shall be bagged in a biohazard bag, properly labeled as the contents, and disposed of by appropriate means.  
1.3 The human blood should be handled at Biosafety Level 2 as recommended in the Centers for Disease Control/National Institutes of Health Manual Biosafety in Microbiological Laboratories. The human blood donor must have tested negative for Hepatitis B (HBV) and Human Immunodeficiency (HIV) viruses. The blood should be treated like any patient blood in using universal precautions.  
1.4 The values stated in SI units are to be regarded as standard. No other units of measurement are included in this standard.  
1.5 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use. Some specific hazards statements are given in Section 7 on Hazards.

General Information

Status
Historical
Publication Date
14-Feb-2013
Current Stage
Ref Project

Relations

Buy Standard

Standard
ASTM F2888-13 - Standard Test Method for Platelet Leukocyte Count&mdash;An <emph type="bdit">In-Vitro</emph > Measure for Hemocompatibility Assessment of Cardiovascular Materials
English language
4 pages
sale 15% off
Preview
sale 15% off
Preview

Standards Content (Sample)

NOTICE: This standard has either been superseded and replaced by a new version or withdrawn.
Contact ASTM International (www.astm.org) for the latest information
Designation: F2888 − 13
Standard Test Method for
Platelet Leukocyte Count—An In-Vitro Measure for
1
Hemocompatibility Assessment of Cardiovascular Materials
This standard is issued under the fixed designation F2888; the number immediately following the designation indicates the year of
original adoption or, in the case of revision, the year of last revision.Anumber in parentheses indicates the year of last reapproval.A
superscript epsilon (´) indicates an editorial change since the last revision or reapproval.
1. Scope 2. Referenced Documents
2
1.1 This test method assists in the evaluation of cardiovas- 2.1 ASTM Standards:
cular device materials for their ability to induce thrombus F2382TestMethodforAssessmentofIntravascularMedical
formation. Thrombus formation is assessed by means of a Device Materials on Partial Thromboplastin Time (PTT)
reduction in human platelets and leukocytes when consumed 2.2 Other Standards:
bythrombusafteractivationonthematerialsurface.Thisassay ANSI/AAMI/ISO 10993–4Biological Evaluation of Medi-
may be part of the hemocompatibility evaluation for devices cal Devices—Part 4: Selection of Tests for Interactions
3
and materials contacting human blood, as per ANSI/AAMI/ with Blood
ISO 10993–4. See also Test Method F2382. Centers for Disease Control/National Institutes of Health
Manual Biosafety in Microbiological Laboratories, 1993
1.2 All safety policies and practices shall be observed
during the performance of this test method. All human blood
3. Summary of Test Method
and any materials that had contact with human blood shall be
3.1 This test method identifies materials which are capable
bagged in a biohazard bag, properly labeled as the contents,
of activating blood platelets and leukocytes on their surface
and disposed of by appropriate means.
when exposed to freshly drawn human blood and causing the
1.3 The human blood should be handled at Biosafety Level
formation of thrombi on the material surface. A significant
2 as recommended in the Centers for Disease Control/National
decrease in the number of platelets and leukocytes when
Institutes of Health Manual Biosafety in Microbiological
counted by a blood analyzer is an indication of these cells
Laboratories. The human blood donor must have tested nega-
beingentrappedinthrombi.Thematerialsareexposedtoblood
tive for Hepatitis B (HBV) and Human Immunodeficiency
immediately after the blood is drawn with anticoagulant.
(HIV) viruses. The blood should be treated like any patient
Another anticoagulant is added at the appropriate time (one
blood in using universal precautions.
hour) to stop the reaction. Different blood analyzers may be
used.
1.4 The values stated in SI units are to be regarded as
standard. No other units of measurement are included in this
4. Significance and Use
standard.
4.1 The purpose of this test method is to determine if
1.5 This standard does not purport to address all of the
medical materials exposed to human whole blood will ad-
safety concerns, if any, associated with its use. It is the
verselyaffecttheplateletandleukocytecountsinwholeblood.
responsibility of the user of this standard to establish appro-
A large number of platelets and leukocytes as part of thrombi
priate safety and health practices and determine the applica-
adhering to the material will be reflected by a decrease in their
bility of regulatory limitations prior to use. Some specific
counts in blood. Thrombogenic materials should not be used
hazards statements are given in Section 7 on Hazards.
2
For referenced ASTM standards, visit the ASTM website, www.astm.org, or
1
ThistestmethodisunderthejurisdictionofASTMCommitteeF04onMedical contact ASTM Customer Service at service@astm.org. For Annual Book of ASTM
andSurgicalMaterialsandDevicesandisthedirectresponsibilityofSubcommittee Standards volume information, refer to the standard’s Document Summary page on
F04.16 on Biocompatibility Test Methods. the ASTM website.
3
Current edition approved Feb. 15, 2013. Published March 2013. DOI: 10.1520/ Available fromAmerican National Standards Institute (ANSI), 25 W. 43rd St.,
F2888–13. 4th Floor, New York, NY 10036, http://www.ansi.org.
Copyright © ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959. United States
1

---------------------- Page: 1 ----------------------
F2888 − 13
for cardiovascular medical devices, unless the purpose of the
C = percentage (%) of negative control.
device is to promote thrombosis.
When applicable, a comparison article should be used in the
formula in place of the negative control.
5. Interferences
5.1 There is potential for interference if the materials of the
10. Preparation of Apparatus
test tubes used are thrombogenic (for example, glass tubes).
10.1 Initialize the hematology analyzer and allow it to
Therefore, polyethylene or
...

Questions, Comments and Discussion

Ask us and Technical Secretary will try to provide an answer. You can facilitate discussion about the standard in here.