Standard Test Methods for Evaluating Design and Performance Characteristics of Motorized Treadmills

SIGNIFICANCE AND USE
4.1 The purpose of these test methods is to provide reliable and repeatable test methods for the evaluation of motorized treadmills assembled and maintained according to the manufacturer's specifications. Use of these test methods in conjunction with Specification F2115, Specification F2276, and Test Methods F2571 is intended to ensure appropriate performance and reliability of a motorized treadmill and reduce the risk of serious injury from design deficiencies.
SCOPE
1.1 These test methods specify procedures and equipment used for testing and evaluating a motorized treadmill for compliance to Specification F2115. Both design and operational parameters will be evaluated. Where possible and applicable, accepted test methods from other recognized bodies will be used and referenced. In case of a conflict between this document and Specification F2115, Specification F2115 takes precedence.  
1.2 This test method is to be used in conjunction with Specification F2276, Test Methods F2571, and Specification F2115.  
1.3 This standard takes precedence over Specification F2276 and Test Methods F2571 in areas that are specific to motorized treadmills.  
1.4 Requirements—A motorized treadmill is to be tested for all of the following parameters:  
1.4.1 Stability,  
1.4.2 Exterior design,  
1.4.3 Endurance loading,  
1.4.4 Static loading,  
1.4.5 Adjustable incline system,  
1.4.6 Controls,  
1.4.7 Motorized drive system,  
1.4.8 Folding treadmills,  
1.4.9 Additional universal design requirements,  
1.4.10 Documentation,  
1.4.11 Marking, and  
1.4.12 Warnings/Warning labels.  
1.5 This test method2 contains additional requirements to address the accessibility of the equipment for persons with disabilities.  
1.6 The values stated in SI units are to be regarded as the standard. The values in parenthesis are for information only.  
1.7 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety, health, and environmental practices and determine the applicability of regulatory limitations prior to use.  
1.8 This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.

General Information

Status
Published
Publication Date
31-May-2018
Drafting Committee
F08.30 - Fitness Products

Relations

Effective Date
01-Jun-2018
Effective Date
15-Nov-2023
Effective Date
01-May-2020
Effective Date
01-Jun-2018
Effective Date
01-Dec-2017
Effective Date
15-May-2017
Effective Date
01-Apr-2017
Effective Date
01-Oct-2016
Effective Date
01-Oct-2016
Effective Date
01-Oct-2016
Effective Date
01-Apr-2015
Effective Date
01-Apr-2015
Effective Date
01-Apr-2015
Effective Date
01-Sep-2014
Effective Date
01-Sep-2014

Overview

ASTM F2106-18 provides standardized test methods for evaluating the design and performance characteristics of motorized treadmills. Developed by ASTM International, this standard ensures that treadmills are tested for safety, performance, and reliability, following procedures recognized by fitness equipment manufacturers, independent laboratories, and regulators. The standard addresses the evaluation of treadmills as assembled and maintained according to the manufacturer's specifications and is closely aligned with related ASTM specifications to ensure comprehensive product assessment and user protection.

Key Topics

ASTM F2106-18 covers the following critical test areas for motorized treadmills:

  • Stability: Evaluation to prevent tipping over during normal use.
  • Exterior Design: Assessment of moving parts, guarding against mechanical and electrical hazards, and dimensions of footrails and handrails.
  • Endurance Loading: Testing the treadmill’s durability under repetitive stress.
  • Static Loading: Ensuring components withstand expected static loads.
  • Adjustable Incline System: Verification of incline speed, safety, and function.
  • Controls: Inspections for accessible and effective emergency stop mechanisms and user controls.
  • Motorized Drive System: Checking operational functionality and endurance.
  • Folding Treadmills: Additional stability and safety checks for treadmills with folding mechanisms.
  • Universal Design Requirements: Features for accessibility, including for individuals with disabilities.
  • Documentation, Marking, and Warning Labels: Ensuring proper instructions, marking, and safety warnings are present and clear.

Applications

The test methods provided by ASTM F2106-18 have significant practical applications across the fitness equipment industry and related sectors:

  • Manufacturers: Use these methods for product development, quality assurance, and self-certification to ensure compliance with marketplace safety expectations.
  • Testing Laboratories: Facilitate independent verification of treadmill performance and safety characteristics, including endurance and emergency stop mechanisms.
  • Fitness Centers and Gym Operators: Reference these standards for purchasing decisions, maintenance protocols, and risk mitigation.
  • Regulatory Agencies: Enable uniform assessment benchmarks for imported or domestically produced fitness equipment.
  • Consumers: Indirectly benefit from enhanced product safety, reliable performance, and clear warning labels resulting from consistent application of these requirements.
  • Accessibility Advocates: Promote inclusive fitness environments by verifying equipment accessibility as addressed in this standard.

Related Standards

ASTM F2106-18 is designed for use with the following related standards to provide comprehensive coverage of treadmill evaluation:

  • ASTM F2115 - Specification for Motorized Treadmills (primary compliance reference)
  • ASTM F2276 - Specification for Fitness Equipment (general design and safety criteria)
  • ASTM F2571 - Test Methods for Evaluating Fitness Equipment (general performance evaluation)
  • ASTM F3021/F3022 - Specifications and test methods for universal design, supporting accessibility
  • UL 1647 - Standard for Motor Operated Massage and Exercise Machines (electrical safety reference)
  • EN 957-1 - European standard for stationary training equipment safety requirements and testing

By following ASTM F2106-18 in conjunction with these referenced standards, stakeholders across the fitness industry can help ensure that motorized treadmills are safe, reliable, and accessible for a broad range of users. This ultimately contributes to the reduction of injury risk from design deficiencies, supports international standardization, and enhances consumer confidence in fitness equipment performance.

Buy Documents

Standard

ASTM F2106-18 - Standard Test Methods for Evaluating Design and Performance Characteristics of Motorized Treadmills

English language (12 pages)
sale 15% off
sale 15% off
Standard

REDLINE ASTM F2106-18 - Standard Test Methods for Evaluating Design and Performance Characteristics of Motorized Treadmills

English language (12 pages)
sale 15% off
sale 15% off

Frequently Asked Questions

ASTM F2106-18 is a standard published by ASTM International. Its full title is "Standard Test Methods for Evaluating Design and Performance Characteristics of Motorized Treadmills". This standard covers: SIGNIFICANCE AND USE 4.1 The purpose of these test methods is to provide reliable and repeatable test methods for the evaluation of motorized treadmills assembled and maintained according to the manufacturer's specifications. Use of these test methods in conjunction with Specification F2115, Specification F2276, and Test Methods F2571 is intended to ensure appropriate performance and reliability of a motorized treadmill and reduce the risk of serious injury from design deficiencies. SCOPE 1.1 These test methods specify procedures and equipment used for testing and evaluating a motorized treadmill for compliance to Specification F2115. Both design and operational parameters will be evaluated. Where possible and applicable, accepted test methods from other recognized bodies will be used and referenced. In case of a conflict between this document and Specification F2115, Specification F2115 takes precedence. 1.2 This test method is to be used in conjunction with Specification F2276, Test Methods F2571, and Specification F2115. 1.3 This standard takes precedence over Specification F2276 and Test Methods F2571 in areas that are specific to motorized treadmills. 1.4 Requirements—A motorized treadmill is to be tested for all of the following parameters: 1.4.1 Stability, 1.4.2 Exterior design, 1.4.3 Endurance loading, 1.4.4 Static loading, 1.4.5 Adjustable incline system, 1.4.6 Controls, 1.4.7 Motorized drive system, 1.4.8 Folding treadmills, 1.4.9 Additional universal design requirements, 1.4.10 Documentation, 1.4.11 Marking, and 1.4.12 Warnings/Warning labels. 1.5 This test method2 contains additional requirements to address the accessibility of the equipment for persons with disabilities. 1.6 The values stated in SI units are to be regarded as the standard. The values in parenthesis are for information only. 1.7 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety, health, and environmental practices and determine the applicability of regulatory limitations prior to use. 1.8 This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.

SIGNIFICANCE AND USE 4.1 The purpose of these test methods is to provide reliable and repeatable test methods for the evaluation of motorized treadmills assembled and maintained according to the manufacturer's specifications. Use of these test methods in conjunction with Specification F2115, Specification F2276, and Test Methods F2571 is intended to ensure appropriate performance and reliability of a motorized treadmill and reduce the risk of serious injury from design deficiencies. SCOPE 1.1 These test methods specify procedures and equipment used for testing and evaluating a motorized treadmill for compliance to Specification F2115. Both design and operational parameters will be evaluated. Where possible and applicable, accepted test methods from other recognized bodies will be used and referenced. In case of a conflict between this document and Specification F2115, Specification F2115 takes precedence. 1.2 This test method is to be used in conjunction with Specification F2276, Test Methods F2571, and Specification F2115. 1.3 This standard takes precedence over Specification F2276 and Test Methods F2571 in areas that are specific to motorized treadmills. 1.4 Requirements—A motorized treadmill is to be tested for all of the following parameters: 1.4.1 Stability, 1.4.2 Exterior design, 1.4.3 Endurance loading, 1.4.4 Static loading, 1.4.5 Adjustable incline system, 1.4.6 Controls, 1.4.7 Motorized drive system, 1.4.8 Folding treadmills, 1.4.9 Additional universal design requirements, 1.4.10 Documentation, 1.4.11 Marking, and 1.4.12 Warnings/Warning labels. 1.5 This test method2 contains additional requirements to address the accessibility of the equipment for persons with disabilities. 1.6 The values stated in SI units are to be regarded as the standard. The values in parenthesis are for information only. 1.7 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety, health, and environmental practices and determine the applicability of regulatory limitations prior to use. 1.8 This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.

ASTM F2106-18 is classified under the following ICS (International Classification for Standards) categories: 97.220.30 - Indoor sports equipment. The ICS classification helps identify the subject area and facilitates finding related standards.

ASTM F2106-18 has the following relationships with other standards: It is inter standard links to ASTM F2106-12, ASTM F3021-23, ASTM F1749-15(2020), ASTM F2115-18, ASTM F2277-17a, ASTM F2277-17, ASTM F3021-17, ASTM F3022-16, ASTM F3021-16, ASTM F3022-16e1, ASTM F3021-15, ASTM F1749-15, ASTM F3022-15, ASTM F3022-14, ASTM F3021-14. Understanding these relationships helps ensure you are using the most current and applicable version of the standard.

ASTM F2106-18 is available in PDF format for immediate download after purchase. The document can be added to your cart and obtained through the secure checkout process. Digital delivery ensures instant access to the complete standard document.

Standards Content (Sample)


This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the
Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.
Designation: F2106 − 18 An American National Standard
Standard Test Methods for
Evaluating Design and Performance Characteristics of
Motorized Treadmills
This standard is issued under the fixed designation F2106; the number immediately following the designation indicates the year of
original adoption or, in the case of revision, the year of last revision.Anumber in parentheses indicates the year of last reapproval.A
superscript epsilon (´) indicates an editorial change since the last revision or reapproval.
INTRODUCTION
The goal of these test methods is to provide reliable and repeatable methods for the evaluation of
motorized treadmills. Users of the equipment must recognize that conformance to a standard will not
necessarily prevent injuries. Like other physical activities, exercise involving a treadmill involves a
risk of injury, particularly if the equipment is not maintained or used properly.
The equipment user must recognize, however, that a standard alone will not necessarily prevent
injuries.Likeotherphysicalactivities,exerciseinvolvingfitnessequipmentinvolvestheriskofinjury,
particularly if the equipment is used improperly or not properly maintained. In addition, users with
physical limitations should seek medical advice or instruction from the fitness facility, or both, prior
to using this equipment. Certain physical conditions or limitations may preclude some persons from
using this equipment as intended by the manufacturer, and using this equipment may increase the risk
of injury.
1. Scope 1.4.5 Adjustable incline system,
1.4.6 Controls,
1.1 These test methods specify procedures and equipment
1.4.7 Motorized drive system,
used for testing and evaluating a motorized treadmill for
1.4.8 Folding treadmills,
compliance to Specification F2115. Both design and opera-
1.4.9 Additional universal design requirements,
tional parameters will be evaluated. Where possible and
1.4.10 Documentation,
applicable,acceptedtestmethodsfromotherrecognizedbodies
1.4.11 Marking, and
will be used and referenced. In case of a conflict between this
1.4.12 Warnings/Warning labels.
document and Specification F2115, Specification F2115 takes
precedence. 1.5 This test method contains additional requirements to
address the accessibility of the equipment for persons with
1.2 This test method is to be used in conjunction with
disabilities.
Specification F2276, Test Methods F2571, and Specification
F2115. 1.6 The values stated in SI units are to be regarded as the
standard. The values in parenthesis are for information only.
1.3 This standard takes precedence over Specification
1.7 This standard does not purport to address all of the
F2276 and Test Methods F2571 in areas that are specific to
safety concerns, if any, associated with its use. It is the
motorized treadmills.
responsibility of the user of this standard to establish appro-
1.4 Requirements—Amotorized treadmill is to be tested for
priate safety, health, and environmental practices and deter-
all of the following parameters:
mine the applicability of regulatory limitations prior to use.
1.4.1 Stability,
1.8 This international standard was developed in accor-
1.4.2 Exterior design,
dance with internationally recognized principles on standard-
1.4.3 Endurance loading,
ization established in the Decision on Principles for the
1.4.4 Static loading,
Development of International Standards, Guides and Recom-
mendations issued by the World Trade Organization Technical
1 Barriers to Trade (TBT) Committee.
These test methods are under the jurisdiction of ASTM Committee F08 on
Sports Equipment, Playing Surfaces, and Facilities and are the direct responsibility
of Subcommittee F08.30 on Fitness Products.
CurrenteditionapprovedJune1,2018.PublishedJuly2018.Originallyapproved This work was funded, in part, by the Rehabilitation Engineering Research
in 2001. Last previous edition approved in 2012 as F2106–12. DOI: 10.1520/ CenteronRecTechthroughtheNationalInstituteonDisability,IndependentLiving,
F2106-18. and Rehabilitation Research grant #90RE5009-01-00.
Copyright © ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959. United States
F2106 − 18
2. Referenced Documents 6.1.1 Apparatus and Set Up—Place treadmill on a nonskid
3 10° surface in the orientation that is least stable. A method of
2.1 ASTM Standards:
applying a steady state force equal to 1.0 × maximum user
F1749Specification for Fitness Equipment and Fitness Fa-
weight in the vertical direction must be provided. Possible
cility Safety Signage and Labels
methods of providing force include, but are not limited to,
F2115Specification for Motorized Treadmills
pneumatic cylinder(s) or dead weights.
F2276Specification for Fitness Equipment
6.1.2 Calibration—Using an angle measuring instrument
F2277TestMethodsforEvaluatingDesignandPerformance
accuratetowithin0.1°,verifythenonskidsurfaceis10 60.5°.
Characteristics of Selectorized Strength Equipment
Calibrate the load measurement apparatus to confirm accuracy
F2571TestMethodsforEvaluatingDesignandPerformance
to within 620 N (4.5 lb) over entire user weight range.
Characteristics of Fitness Equipment
F3021Specification for Universal Design of Fitness Equip-
6.1.3 Procedure—Test the treadmill as follows:
ment for Inclusive Use by Persons with Functional Limi-
6.1.3.1 Using the aforementioned load apparatus, apply a
tations and Impairments
vertical load equal to 1.0 × maximum specified user weight 6
F3022Test Method for Evaluating the Universal Design of
5% in a non-impact manner at the point on the foot rail which
Fitness Equipment for Inclusive Use by Persons with
creates the most instability. Verify that the treadmill does not
Functional Limitations and Impairments
tip over.
2.2 UL Standard:
6.1.3.2 Repeat 6.1.3.1 with the treadmill oriented in any
UL 1647Motor Operated Massage and Exercise Machines
other directions of potential instability.
6.1.3.3 For folding treadmills, the treadmill shall, after
3. Terminology
completing6.1.3.1and6.1.3.2,befoldedtoitsstorageposition
3.1 Definitions—For definitions applicable to this standard,
per manufacturer’s instructions and placed on the 10° inclined
see Specifications F2115 and F3021.
surface in all orientations that could cause instability. No
additional load is to be applied.
4. Significance and Use
6.1.4 Pass/Fail Criteria—In none of the above test condi-
4.1 The purpose of these test methods is to provide reliable
tions shall the treadmill tip over.
and repeatable test methods for the evaluation of motorized
6.1.5 Precision and Bias—No information is presented
treadmills assembled and maintained according to the manu-
about either the precision or bias of this test for measuring
facturer’s specifications. Use of these test methods in conjunc-
stability since the test result is non-quantitative.
tion with Specification F2115, Specification F2276, and Test
Methods F2571 is intended to ensure appropriate performance
6.2 Exterior Design:
and reliability of a motorized treadmill and reduce the risk of
6.2.1 Rotating Parts in the Accessible Area and Rear Roller
serious injury from design deficiencies.
Area—The purpose of this test is to evaluate the risk of injury
due to a moving mechanical part. Methodology entails inser-
5. Sample Preparation
tion of an articulate probe that simulates an accepted case
5.1 Assemble and adjust the treadmill on a horizontal
finger into all questionable areas.
surface according to the manufacturer’s instructions. Once
6.2.1.1 Apparatus and Set Up—This test requires a probe as
fully assembled, verify according to the manufacturer’s in-
specified inTest Methods F2277, Fig. 3.Verify that all guards
structions that the moving surface has been adjusted to the
are properly positioned and secured and the moving surface is
proper tension and alignment. Unless otherwise stated, the
centered per instructions in the owner’s manual. Apply power
treadmill must pass the following tests without adjustment
to the treadmill so that its incline can be elevated and the
from this initial condition. Apply power to the treadmill and
moving surface can be run.
verify that the unit functions properly. If the unit is equipped
6.2.1.2 Calibration—Verify the probe conforms to the di-
with an adjustable incline system, operate it through its full
mensions of Test Methods F2277, Fig. 3.
range.
6.2.1.3 Procedure—With no power applied, insert the probe
5.2 The individual test methods will describe any variations
at any pull-in nip points or areas accessible to any mechanical
or modifications that are required to the test sample.
hazards on the treadmill. Points of insertion include, but shall
not be limited to the rear roller guards, foot rail to moving
6. Test Methods and Procedures
surface interface, motor compartment/moving surface gap and
6.1 Stability—The treadmill shall be tested by a series of
any openings in the motor compartment or shroud. For all
load applications in the orientation that is most obviously
insertions, the probe is to be rotated and bent in all possible
unstable.
configurationsandapplicationforceshallnotexceed4.4N(1.0
lb). Elevate the treadmill to its maximum incline position.
Under the assumption there shouldn’t be any hazard, run the
For referenced ASTM standards, visit the ASTM website, www.astm.org, or
contact ASTM Customer Service at service@astm.org. For Annual Book of ASTM
moving surface at minimum speed and verify the finger will
Standards volume information, refer to the standard’s Document Summary page on
not get trapped anywhere. If any other incline positions reveal
the ASTM website.
apotentialpull-inornippoints,thetestshallberepeatedatthat
Available from Underwriters Laboratories (UL), Corporate Progress, 333
Pfingsten Rd., Northbrook, IL 60062. incline position.
F2106 − 18
6.2.1.4 Pass/Fail Criteria—The probe shall not become (1) Precision—The precision of this test method has not
entrapped in any mechanical hazard. Entrapment is defined to been determined.
have occurred if the force to pull out the probe is greater than (2) Bias—Thebiasofthistestmethodincludesquantitative
4.4 N (1.0 lb). estimates of the uncertainties of the measuring devices, the
calibrationsoftestingequipment,andtheskilloftheoperators.
6.2.1.5 Precision and Bias—No information is presented
At this time, the statements on bias should be limited to
about either the precision or bias of this test for evaluating
documented performance of particular laboratories.
hazards of rotating parts since the test result is non-
6.2.4 Moving Surface—The purpose of this test is to verify
quantitative.
the dimensional and marking compliance of the moving
6.2.2 Guarding from Electrical Hazards—The purpose of
surface.
this test is to verify that all electrical elements are adequately
6.2.4.1 Apparatus and Set Up—No set up required, dimen-
guarded to prevent electric shock from un-insulated live parts
sional inspection only.
and film-coated wire. Methodology entails insertion of an
6.2.4.2 Calibration—Verify that the distance measuring
articulate probe that simulates a worst case finger into all
equipment is calibrated and accurate to within 61 mm (0.040
questionable areas.
in.).
6.2.2.1 Apparatus and Set Up—This test requires an articu-
6.2.4.3 Procedure—ConsultthemovingsurfacetableSpeci-
lating probe as specified in UL1647. Verify that all guards are
fication F2115. Refer to Fig.3 in Specification F2115 and
properly positioned and secured and the moving surface is
measurethelengthandwidthofthemovingsurface.Verifythe
centered per instructions in the owner’s manual. Treadmill to
presenceofmovementindicatorsonthemovingsurface.Verify
be tested in all incline positions and all normal usage positions
that they meet the minimum dimensions specified in Specifi-
that may present a hazard.
cationF2115.Operatethetreadmillandverifythataportionof
6.2.2.2 Calibration—Verifyarticulateprobesconformtothe
the movement indicator is visible at all times as the moving
dimensions of UL 1647.
surface rotates.
6.2.2.3 Procedure—With no power applied, insert probe at
6.2.4.4 Pass/Fail Criteria—The dimensions of the moving
any points where contact with electrical elements is possible.
surface shall conform to dimensional requirements of Table 1
Points of insertion include, but shall not be limited to, the
of Specification F2115.
motor compartment/moving surface gap and any openings in
6.2.4.5 Precision and Bias—No information is presented
the motor compartment or shroud. For all insertions, the probe
abouteithertheprecisionorbiasofthistestsincethetestresult
is to be rotated and bent in all possible configurations and
is non-quantitative.
application force shall not exceed 4.4 N (1.0 lb). Elevate the
(1) Precision—The precision of this test method has not
treadmilltoallotherinclinepositionsthatmaypresentahazard
been determined.
and repeat the insertion test. The components not involved in
(2) Bias—Thebiasofthistestmethodincludesquantitative
the insertion area may now be removed to clearly see whether
estimates of the uncertainties of the measuring devices, the
the probe can contact any electrical hazard.
calibrationsoftestingequipment,andtheskilloftheoperators.
6.2.2.4 Pass/Fail Criteria—The probe shall not contact any
At this time, the statements on bias should be limited to
electrical hazard.
documented performance of particular laboratories.
6.2.2.5 Precision and Bias—No information is presented
6.2.5 Moving Surface Slip Resistance—The purpose of this
about either the precision or bias of this test for evaluating
test is to evaluate the slip resistance of the moving surface of
accessibility of electrical components since the test result is
the treadmill.
non-quantitative.
6.2.5.1 Apparatus and Set Up—Thetreadmillshallbesetup
6.2.3 Foot Rails—The purpose of this test is to verify the
in accordance with Test Methods F2571. The moving surface
dimensional compliance of the foot rails.
shall be restrained from moving.
6.2.3.1 Apparatus and Set Up—No set up required, dimen-
6.2.5.2 Calibration—Per Test Methods F2571, no calibra-
sional inspection only.
tion is required.
6.2.3.2 Calibration—Verify that the distance measuring
6.2.5.3 Procedure—Secure the moving surface so that it
equipment is calibrated and accurate to within 61 mm (0.04
cannot move. Conduct the test in accordance with Test Meth-
in.).
ods F2571.
6.2.3.3 Procedure—Verify that the foot rails extend from
6.2.5.4 Pass/Fail Criteria—Moving surfaces that do not
either the base of the masts or treadmill uprights or the end of
resist foot slippage shall fail the test.
the motor cover to the rear end caps of the treadmill. Measure
6.2.5.5 Precision and Bias—No information is presented
the width of the moving surface. Measure the width of the foot
about either the precision or bias of this test for evaluating
rail surface.
moving surface slip resistance since the test result is non-
6.2.3.4 Pass/Fail Criteria—The dimensions of the foot rails quantitative.
shall conform to dimensional requirements of Specification
6.2.6 Hand Grips and Grippable Surfaces—The purpose of
F2115.
this test is to verify the presence and dimensional compliance
of the handrails.
6.2.3.5 Precision and Bias—No information is presented
abouteithertheprecisionorbiasofthistestsincethetestresult 6.2.6.1 Apparatus and Set Up—No set up required, dimen-
is non-quantitative. sional inspection only.
F2106 − 18
6.2.6.2 Calibration—Verify that the distance measuring conventional belt and deck), the moving surface may be
equipment is calibrated and accurate to within 61 mm (0.040 repositioned periodically during the test.
in.).
6.3.1.4 Pass/Fail Criteria—Upon completion of the test,
6.2.6.3 Procedure—Verify how many front/side handrails
remove the test unit from the fixture and confirm normal
arepresent.Measurethelengthofallsegmentsofthehandrail.
operation of all functions. The treadmill must meet life
MeasurethedistancefromLineAonthemovingsurfacetothe
requirements per classification of use per subsection 4.3 of
top of the grip surface. Evaluate through the full range of
Specification F2115. The unit must not show structural cracks
inclination. If so equipped, measure the horizontal distance
or other indications of impending failure.
between the side handrails.
6.3.1.5 Precision and Bias—No information is presented
6.2.6.4 Pass/Fail Criteria—Thedimensionsofthehandrails
about either the precision or bias of this test for evaluating
shall conform to dimensional requirements of Specification
frame endurance since the test result is non-quantitative.
F2115.
6.3.2 Treadmill Switch and Switch Actuation Mechanism
6.2.6.5 Precision and Bias—No information is presented
Endurance—Stop, Pause or End functions (see Note 1).
abouteithertheprecisionorbiasofthistestsincethetestresult
is non-quantitative.
NOTE 1—If several means of stopping exist—for example, Stop switch
and Pause switch—each means must be separately tested only if the
(1) Precision—The precision of this test method has not
actuator types are different.
been determined.
(2) Bias—Thebiasofthistestmethodincludesquantitative
6.3.2.1 Apparatus and Set Up—Provide a mechanism to
estimates of the uncertainties of the measuring devices, the
repeatedlyactivatetheswitchofinterestataratenottoexceed
calibrationsoftestingequipment,andtheskilloftheoperators.
2Hz.andtoaccumulateacountofactuations.Activationforce
At this time, the statements on bias should be limited to
for testing shall be 1.5× 6 10% the minimum actuating force
documented performance of particular laboratories.
for the particular switch. For push switches, the activating
6.3 Endurance Loading—These tests are intended to con- mechanism can be a simulated finger mounted to an appropri-
firm the endurance of the main components of the treadmill.
atepneumaticcylinderoperatingparalleltothelineofactionof
the switch, a test system designed specifically for switch
6.3.1 Mechanical Frame Endurance—The stationary mov-
ing surface, frame, and structural components of the treadmill testing (having an air cylinder plunger with a rounded simu-
are subjected to a repeated load equal to 1.5 times the lated finger 0.675 diameter, 45 durometer rubber, SF-45), or
maximum user weight as specified by the manufacturer. other appropriate actuating means. For non-returning switches,
a mechanism must be provided to pull out the switch between
6.3.1.1 Apparatus and Set Up—A pneumatic cylinder or
actuations, with a force equal to 1.5× 6 10% the minimum
other appropriate load application device shall be fixtured to
activating force. For lanyard-type, pull-off mechanisms, the
apply load to the moving surface at the centerline of the
treadmill 33% of the length of the usable moving surface switch can be tested as outlined above without the lanyard
rearward from the front of the usable moving surface.A30 by attached. If this method is employed, a separate lanyard
30 cm (11.8 by 11.8 in.) load application area shall be used for
strength test must be conducted by applying a load equal to 5
the test. Timing controls must be supplied to regulate the load
times the maximum activation force of the switch on the
application frequency. Provide appropriate instrumentation to
lanyard. A means of automatically sensing activation (for
assure that the load is consistent throughout the test duration.
example, sensing a “beep” from the control panel internal
6.3.1.2 Calibration—Verify that the load and frequency
electronics) will allow monitoring each activation response.
measuring equipment is properly calibrated and that the load
6.3.2.2 Calibration—Verifythattheloadapplicationsystem
accuracy is within 65% of the applied load and that the
applies 1.5 times 6 10% of the minimum activation force for
frequency measurement is accurate to within 60.2 Hz.
the switch to be tested. Using appropriate instrumentation, set
6.3.1.3 Procedure—Elevate the treadmill to the incline that
the timing of the activator to 0.1 to 0.2 s “on” time and the
will provide the most stress to the frame and incline system
repeat rate as desired but not over 2 Hz. Verify activation
components and position the treadmill under the load applica-
counter operation for at least 100 cycles. If a lanyard pull is
tion device. No power shall be connected to the treadmill.
required, the pull force must guarantee activation.
Block the treadmill into position so that the vibrations created
6.3.2.3 Procedure—Actuatetheswitchbeingtestedwiththe
by the test do not cause it to move on the floor. Apply power
load application system and confirm that the force applied
to the load application system. Adjust the applied load to 1.5
causes the switch to function. The function of the switch shall
times maximum user weight 65% at a frequency of 0.5 to 2
be confirmed on every actuation of the switch via a circuit that
Hz.Assure that the moving surface rebounds completely prior
actuates a counter. A manual check must be made of switch
to the next cycle. Verify that the counter is recording the
function before and after the full test sequence, and during the
number of cycles. Inspect the treadmill and test apparatus
automated sequence, some activation-confirming feedback
approximatelyevery250000cyclesorasneededtoensurethat
must be observable for each activation. As an alternate to this
the test is functioning properly. If the design of the treadmill is
procedure, if the switch is installed according to the switch
such that keeping the moving surface stationary for the
manufacturer’s recommendations, the switch manufacturer’s
durationofthetestwouldabnormallystressthemovingsurface
(that is, a treadmill that employs moving slats rather than a testing may be used.
F2106 − 18
6.3.2.4 Pass/Fail Criteria—The switch(s) tested must be andthatdonotactuallyprovidetheenergytoadjusttheincline,
functional as described in procedure at the end of the life test are excluded from this test.
requirements for the classification for use as follows: Con-
6.5.1 Apparatus and Set Up—A combination of vertical
sumer Treadmills—1560 actuations without damage or non-
displacement and time measuring instruments which yields a
functionality. Institutional Treadmills—46 800 actuations
combinedaccuracyof2.5mm/s(0.1in./s)willbenecessaryfor
without damage or non-functionality.
thevelocitymeasurements.Thevelocitywillbedeterminedby
dividing the displacement by the time required for movement.
6.3.2.5 Precision and Bias—No information is presented
about either the precision or bias of this test for evaluating As an alternate, an instrument that directly measures vertical
velocity having the accuracy stated above may be used. The
actuator endurance since the test result is non-quantitative.
instrumentation is to be fixtured to measure the relative
6.4 Static Loading of Structures—The purpose of this test is
velocity of any pinch or shear points that occur during
to evaluate the static structural integrity of the deck, foot rails,
actuation of the incline system. A method of applying maxi-
any frame structure and user supports (that is, handlebars and
mum specified user weight that is independent of angle of
side handrails) as applicable to user safety.
inclination must be provided.
6.4.1 Moving Surface and Foot Rails:
6.5.2 Calibration—Verify that the instrumentation is prop-
6.4.1.1 Apparatus and Set Up—Supply a means of provid-
erly calibrated and yields an accuracy within 2.5 mm/s (0.1
ing a steady state load (pneumatic cylinder, weights, etc.) on
in./s) of actual.
the deck and foot rails, capable of adjustment to apply vertical
6.5.3 Procedure—With the treadmill in its minimum incline
load to the test member at minimum, mid-range and maximum
positionandnoloadapplied,beginrecordingdata.Operatethe
incline for the moving surface. Load is to be applied on a 300
incline control to elevate the treadmill to maximum elevation
by 300 mm (11.8 by 11.8 in.) square area located on the
as quickly as possible. Once the treadmill has reached full
centerline of the moving surface 33% of the usable surface
inclination,operatetheinclinecontroltoreturnthetreadmillto
length back from the front of the usable moving surface. For
its minimum inclination as quickly as possible.
the foot rails, the vertical load is to be applied at the
6.5.3.1 Apply maximum user weight to a point 33% of the
longitudinal centerline of the foot rail.
length of the usable moving surface rearward from the front of
6.4.1.2 Calibration—Verify load application system is cali-
the usable moving surface. Repeat the procedure described in
brated and is accurate to within 65% of the applied load.
6.5.3.
6.4.1.3 Procedure—With the treadmill fixed to prevent
6.5.4 Pass/Fail Criteria—The maximum measured speed
movementandmovingsurfaceheldstationary,applyaforceof
shall not exceed 25 mm/s (l.0 in./s).
3× the maximum user weight for consumer treadmills or 4×
6.5.5 Precision and Bias—No information is presented
maximumuserweightforcommercialtreadmillstothemoving
abouteithertheprecisionorbiasofthistestsincethetestresult
surfaceasdescribedin6.3.1.1.Thetreadmillistobesupported
is non-quantitative.
only at points specifically intended for support during use.
(1) Precision—The precision of this test method has not
Apply the load for 5 to 15 s. Remove the load. Then apply
been determined.
vertical force of 2× the maximum user weight for consumer
(2) Bias—Thebiasofthistestmethodincludesquantitative
treadmills or 3× maximum user weight for commercial tread-
estimates of the uncertainties of the measuring devices, the
mills to the longitudinal center of one of the foot rails over an
calibrationsoftestingequipment,andtheskilloftheoperators.
areaof180mm(7.1in.)inlengthby180mm(7.1in.)inwidth,
At this time, the statements on bias should be limited to
or the width of the foot rails, whichever is less. Maintain load
documented performance of particular laboratories.
for 5 to 15 s, then remove load.
6.6 Controls—This section is to verify that the proper stop
6.4.1.4 Pass/Fail Criteria—Examine all frame members,
controls exist and that the accuracy of the speed readout is
welds,jointsandthedeckforcracks,separationsorfailure.No
adequate.
visible evidence of cracks, separations, or other structural
damage is allowable. Cosmetic damage is not considered a
6.6.1 Verification of Emergency Stop—This test is a visual
failure. inspection of the unit to verify the presence of an emergency
stop.
6.4.1.5 Precision and Bias—No information is presented
abouteithertheprecisionorbiasofthistestsincethetestresult 6.6.1.1 Apparatus and Set Up—The sample shall be set up
is non-quantitative. as described in Section 5.
(1) Precision—The precision of this test method has not
6.6.1.2 Calibration—Nocalibrationrequired.Visualinspec-
been determined.
tion only.
(2) Bias—Thebiasofthistestmethodincludesquantitative
6.6.1.3 Procedure—Verify that stop mechanisms per Speci-
estimates of the uncertainties of the measuring devices, the
fication F2115 exist.
calibrationsoftestingequipment,andtheskilloftheoperators.
6.6.1.4 Pass/Fail Criteria—Emergency stop mechanisms
At this time, the statements on bias should be limited to
shall conform to the requirements of Specification F2115.
documented performance of particular laboratories.
6.6.1.5 Precision and Bias—No information is presented
6.5 Adjustable Incline System Function—Note that tread- about either the precision or bias of this test for evaluating
mills that use a mechanical incline position system such as a emergency stop actuation mechanisms since the test result is
ratchet, pin, or other means to provide limited fixed positions, non-quantitative.
F2106 − 18
6.6.2 Verification of Stop Mechanism Function—This test is (2) Bias—Thebiasofthistestmethodincludesquantitative
avisualandfunctionalinspectionoftheunittoensurethatstop estimates of the uncertainties of the measuring devices, the
mechanisms function and are user accessible. calibrationsoftestingequipment,andtheskilloftheoperators.
6.6.2.1 Apparatus and Set Up—Place the treadmill on a At this time, the statements on bias should be limited to
documented performance of particular laboratories.
level surface. Supply power to the treadmill.
6.6.5 Push Button Stop Switch Size—This test is a dimen-
6.6.2.2 Calibration—No calibration required. Visual and
functional inspection only. sional inspection of the sample to ensure Emergency stop push
button size.
6.6.2.3 Procedure—Verify that stop mechanisms per Speci-
6.6.5.1 Apparatus and Set Up—The sample shall be set up
fication F2115 exist. Operate all stop mechanisms to verify
accessibility. Verify the accessibility with a 5th percentile as described in Section 5.
female running at 8 km/h (5 mph). Perform function tests at 6.6.5.2 Calibration—Verify that the distance measuring
minimum and maximum incline at 8 km/h (5.0 mph) or equipment is calibrated and accurate to within 1 mm (0.040
maximum speed, whichever is less, with a 100 kg (220.5 lb) in.).
user. 6.6.5.3 Procedure—Measure the surface area of the push
6.6.2.4 Pass/Fail Criteria—Stop devices must be readily button stop switch.
accessible and cause the moving surface to decelerate to rest
6.6.5.4 Pass/Fail Criteria—Push button stop switch size
and stop the motion of the incline system. shall conform to the requirements of Specification F2115.
6.6.2.5 Precision and Bias—No information is presented 6.6.5.5 Precision and Bias—No information is presented
about either the precision or bias of this test for evaluating
abouteithertheprecisionorbiasofthistestsincethetestresult
accessibility and function of stop mechanisms since the test is non-quantitative.
result is non-quantitative.
(1) Precision—The precision of this test method has not
6.6.3 Emergency Stop Appearance—This test is a visual been determined.
inspection of the sample to ensure the compliance of the (2) Bias—Thebiasofthistestmethodincludesquantitative
emergency stop color, size and shape. estimates of the uncertainties of the measuring devices, the
calibrationsoftestingequipment,andtheskilloftheoperators.
6.6.3.1 Apparatus and Set Up—The sample shall be set up
At this time, the statements on bias should be limited to
as described in Section 5.
documented performance of particular laboratories.
6.6.3.2 Calibration—No calibration required. Visual test
6.6.6 Push Button Stop Switch Color—This test is a visual
only.
inspection of the sample to ensure the compliance of the push
6.6.3.3 Procedure—Verify that the emergency stop color,
button stop switch color.
shape, and size differ from other available controls.
6.6.6.1 Apparatus and Set Up—The sample shall be set up
6.6.3.4 Pass/Fail Criteria—Emergency stop color, shape,
as described in Section 5.
and size shall conform to the requirements of Specification
6.6.6.2 Calibration—No calibration required. Visual test
F2115.
only.
6.6.3.5 Precision and Bias—No information is presented
6.6.6.3 Procedure—Verify that the push button stop switch
about either the precision or bias of this test for evaluating the
color is red.
emergency stop color, shape and size since the test result is
6.6.6.4 Pass/Fail Criteria—ush button stop switch shall
non-quantitative.
conform to the requirements of Specific
...


This document is not an ASTM standard and is intended only to provide the user of an ASTM standard an indication of what changes have been made to the previous version. Because
it may not be technically possible to adequately depict all changes accurately, ASTM recommends that users consult prior editions as appropriate. In all cases only the current version
of the standard as published by ASTM is to be considered the official document.
Designation: F2106 − 12 F2106 − 18 An American National Standard
Standard Test Methods for
Evaluating Design and Performance Characteristics of
Motorized Treadmills
This standard is issued under the fixed designation F2106; the number immediately following the designation indicates the year of
original adoption or, in the case of revision, the year of last revision. A number in parentheses indicates the year of last reapproval. A
superscript epsilon (´) indicates an editorial change since the last revision or reapproval.
INTRODUCTION
The goal of these test methods is to provide reliable and repeatable methods for the evaluation of
motorized treadmills. Users of the equipment must recognize that conformance to a standard will not
necessarily prevent injuries. Like other physical activities, exercise involving a treadmill involves a
risk of injury, particularly if the equipment is not maintained or used properly.
The equipment user must recognize, however, that a standard alone will not necessarily prevent
injuries. Like other physical activities, exercise involving fitness equipment involves the risk of injury,
particularly if the equipment is used improperly or not properly maintained. In addition, users with
physical limitations should seek medical advice or instruction from the fitness facility, or both, prior
to using this equipment. Certain physical conditions or limitations may preclude some persons from
using this equipment as intended by the manufacturer, and using this equipment may increase the risk
of injury.
1. Scope
1.1 These test methods specify procedures and equipment used for testing and evaluating a motorized treadmill for compliance
to Specification F2115. Both design and operational parameters will be evaluated. Where possible and applicable, accepted test
methods from other recognized bodies will be used and referenced. In case of a conflict between this document and Specification
F2115, Specification F2115 takes precedence.
1.2 This test method is to be used in conjunction with Specification F2276, Test Methods F2571, and Specification F2115.
1.3 This standard takes precedence over Specification F2276 and Test Methods F2571 in areas that are specific to motorized
treadmills.
1.4 Requirements—A motorized treadmill is to be tested for all of the following parameters:
1.4.1 Stability,
1.4.2 Exterior design,
1.4.3 Endurance,Endurance loading,
1.4.4 Static loading,
1.4.5 Overheating,
1.4.5 Adjustable incline system function,system,
1.4.6 User interface parameters,Controls,
1.4.7 Motorized drive system operation,system,
1.4.8 Folding treadmills,
1.4.9 Additional universal design requirements,
1.4.10 Documentation,
1.4.11 Warning label compliance, Marking, and
1.4.12 Documentation.Warnings/Warning labels.
ThisThese test method ismethods are under the jurisdiction of ASTM Committee F08 on Sports Equipment, Playing Surfaces, and Facilities and isare the direct
responsibility of Subcommittee F08.30 on Fitness Products.
Current edition approved March 1, 2012June 1, 2018. Published April 2012July 2018. Originally approved in 2001. Last previous edition approved in 20102012 as
F2106 – 03 (2010).F2106 – 12. DOI: 10.1520/F2106-12.10.1520/F2106-18.
Copyright © ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959. United States
F2106 − 18
1.5 This test method contains additional requirements to address the accessibility of the equipment for persons with disabilities.
1.6 The values stated in SI units are to be regarded as the standard. The values in parenthesis are for information only.
1.7 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility
of the user of this standard to establish appropriate safety, health, and environmental practices and determine the applicability of
regulatory limitations prior to use.
1.8 This international standard was developed in accordance with internationally recognized principles on standardization
established in the Decision on Principles for the Development of International Standards, Guides and Recommendations issued
by the World Trade Organization Technical Barriers to Trade (TBT) Committee.
2. Referenced Documents
2.1 ASTM Standards:
F1749 Specification for Fitness Equipment and Fitness Facility Safety Signage and Labels
F2115 Specification for Motorized Treadmills
F2276 Specification for Fitness Equipment
F2277 Test Methods for Evaluating Design and Performance Characteristics of Selectorized Strength Equipment
F2571 Test Methods for Evaluating Design and Performance Characteristics of Fitness Equipment
F3021 Specification for Universal Design of Fitness Equipment for Inclusive Use by Persons with Functional Limitations and
Impairments
F3022 Test Method for Evaluating the Universal Design of Fitness Equipment for Inclusive Use by Persons with Functional
Limitations and Impairments
2.2 UL Standards:Standard:
UL 1647 Motor Operated Massage and Exercise Machines
2.3 European Standard:
EN 957-1 Stationary Training Equipment—Part 1: General Safety Requirements and Test Methods
3. Terminology
3.1 Definitions—For definitions applicable to this standard, see SpecificationSpecifications F2115 and F3021.
4. Significance and Use
4.1 The purpose of these test methods is to provide reliable and repeatable test methods for the evaluation of motorized
treadmills assembled and maintained according to the manufacturer’s specifications. Use of these test methods in conjunction with
Specification F2115, Specification F2276, and Test Methods F2571 is intended to insureensure appropriate performance and
reliability of a motorized treadmill and reduce the risk of serious injury from design deficiencies.
5. Certification
5.1 5.1 These test methods permit self-certification. It is recommended that each manufacturer employ an independent
laboratory to evaluate and validate that their designs and test procedures conform and comply with these test methods and
Specification F2115, Specification F2276, and Test Methods F2571.
6. Units of Measure
6.1 The values stated in SI units are to be regarded as the standard. The values in parenthesis are for information only.
5. Sample Preparation
5.1 Assemble and adjust the treadmill on a horizontal surface according to the manufacturer’s instructions. On treadmills that
are Once fully assembled, verify according to the manufacturer’s instructions that the moving surface has been adjusted to the
proper tension and alignment. Unless otherwise stated, the treadmill must pass the following tests without adjustment from this
initial condition. Apply power to the treadmill and verify that the unit functions properly. If the unit is equipped with an adjustable
incline system, operate it through its full range.
5.2 The individual test methods will describe any variations or modifications that are required to the test sample.
This work was funded, in part, by the Rehabilitation Engineering Research Center on RecTech through the National Institute on Disability, Independent Living, and
Rehabilitation Research grant #90RE5009-01-00.
For referenced ASTM standards, visit the ASTM website, www.astm.org, or contact ASTM Customer Service at service@astm.org. For Annual Book of ASTM Standards
volume information, refer to the standard’s Document Summary page on the ASTM website.
Available from Underwriters Laboratories (UL), Corporate Progress, 333 Pfingsten Rd., Northbrook, IL 60062.
F2106 − 18
8. Report
8.1 Record of Tests—Maintain complete test records and test summary reports for all testing, whether performed by the
manufacturer or an independent laboratory. The records can be stored on paper, electronically, or on photographs, or a combination
thereof. A copy of the test summary must be kept by the laboratory that performed the test for a minimum of five years from the
date of the test and by the manufacturer for a minimum of five years past the end of production of the model tested. The summary
shall include the signature of the technician(s) performing the tests and a management representative of the laboratory performing
the test. The test summary shall include the following information:
8.1.1 Manufacturer’s name and location,
8.1.2 Information provided by the manufacturer to accurately identify the configuration of, and specific unit provided to, the
testing agency,
8.1.3 Dates over which the tests were conducted, and
8.1.4 Name and location of the testing laboratory, if different from the manufacturer.
8.1.5 Summary and results of each test performed including method and apparatus used. This shall include what the desired
requirement was and whether the test sample met that parameter or failed. If the test requires a specific number of cycles to be
met, then the report must include the number of cycles actually conducted. If the treadmill fails to meet a parameter, then that
failure must be noted in clear and accurate terms to enable a reader of the report to understand at a later date what transpired.
6. Test Methods and Procedures
6.1 Stability—The treadmill shall be tested by a series of load applications in the orientation that is most obviously unstable.
6.1.1 Apparatus and Set Up—Place treadmill on a nonskid 10° surface in the orientation that is least stable. A method of
applying a steady state force equal to 1.0 × maximum user weight in the vertical direction must be provided. Possible methods
of providing force include, but are not limited to, pneumatic cylinder(s) or dead weights.
6.1.2 Calibration—Using an angle measuring instrument accurate to within 0.1°, verify the nonskid surface is 10 6 0.5°.
Calibrate the load measurement apparatus to confirm accuracy to within 620 N (4.5 lb) over entire user weight range.
6.1.3 Procedure—Test the treadmill as follows:
6.1.3.1 Using the aforementioned load apparatus, apply a vertical load equal to 1.0 × maximum specified user weight 6 5 %
in a non-impact manner at the point on the foot rail which creates the most instability. Verify that the treadmill does not tip over.
6.1.3.2 Repeat 9.1.3.16.1.3.1 with the treadmill oriented in any other directions of potential instability.
6.1.3.3 For folding treadmills, the treadmill shall, after completing 9.1.3.16.1.3.1 and 9.1.3.26.1.3.2, be folded to its storage
position per manufacturer’s instructions and placed on the 10° inclined surface in all orientations that could cause instability. No
additional load is to be applied.
6.1.4 Pass/Fail Criteria—In none of the above test conditions shall the treadmill tip over.
6.1.5 Precision and Bias—No information is presented about either the precision or bias of this test 9.1for measuring stability
since the test result is non-quantitative.
6.2 Exterior Design:
6.2.1 Rotating Parts in the Accessible Area and Rear Roller Area—The purpose of this test is to evaluate the risk of injury due
to a moving mechanical part. Methodology entails insertion of an articulate probe that simulates an accepted case finger into all
questionable areas.
6.2.1.1 Apparatus and Set Up—This test requires a probe as specified in Fig. 2, “TestTest Methods F2277Finger,” of EN 957-1.
, Fig. 3. Verify that all guards are properly positioned and secured and the moving surface is centered per instructions in the
owner’s manual. Apply power to the treadmill so that its incline can be elevated and the moving surface can be run.
6.2.1.2 Calibration—Verify the probe conforms to the dimensions of Fig. 2, “TestTest Methods F2277Finger,” of EN 957-1. ,
Fig. 3.
6.2.1.3 Procedure—With no power applied, insert the probe at any pull-in nip points or areas accessible to any mechanical
hazards on the treadmill. Points of insertion include, but shall not be limited to the rear roller guards, foot rail to moving surface
interface, motor compartment/moving surface gap and any openings in the motor compartment or shroud. For all insertions, the
probe is to be rotated and bent in all possible configurations and application force shall not exceed 4.4 N (1(1.0 lb). Elevate the
treadmill to its maximum incline position. Under the assumption there shouldn’t be any hazard, run the moving surface at
minimum speed and verify the finger will not get trapped anywhere. If any other incline positions reveal a potential pull-in or nip
points, the test shall be repeated at that incline position.
6.2.1.4 Pass/Fail Criteria—The probe shall not become entrapped in any mechanical hazard. Entrapment is defined to have
occurred if the force to pull out the probe is greater than 4.4 N (1(1.0 lb).
6.2.1.5 Precision and Bias—No information is presented about either the precision or bias of this test 9.2.1for evaluating
hazards of rotating parts since the test result is non-quantitative.
6.2.2 Guarding from Electrical Hazards—The purpose of this test is to verify that all electrical elements are adequately guarded
to prevent electric shock from un-insulated live parts and film-coated wire. Methodology entails insertion of an articulate probe
that simulates a worst case finger into all questionable areas.
F2106 − 18
6.2.2.1 Apparatus and Set Up—This test requires an articulating probe as specified in UL 1647. Verify that all guards are
properly positioned and secured and the moving surface is centered per instructions in the owner’s manual. Treadmill to be tested
in all incline positions and all normal usage positions that may present a hazard.
6.2.2.2 Calibration—Verify articulate probes conform to the dimensions of UL 1647.
6.2.2.3 Procedure—With no power applied, insert probe at any points where contact with electrical elements is possible. Points
of insertion include, but shall not be limited to, the motor compartment/moving surface gap and any openings in the motor
compartment or shroud. For all insertions, the probe is to be rotated and bent in all possible configurations and application force
shall not exceed 4.4 N (1(1.0 lb). Elevate the treadmill to all other incline positions that may present a hazard and repeat the
insertion test. The components not involved in the insertion area may now be removed to clearly see whether the probe can contact
any electrical hazard.
6.2.2.4 Pass/Fail Criteria—The probe shall not contact any electrical hazard.
6.2.2.5 Precision and Bias—No information is presented about either the precision or bias of this test 9.2.2for evaluating
accessibility of electrical components since the test result is non-quantitative.
6.2.3 Foot Rails—The purpose of this test is to verify the dimensional compliance of the foot rails.
6.2.3.1 Apparatus and Set Up—No set up required, dimensional inspection only.
6.2.3.2 Calibration—Verify that the distance measuring equipment is calibrated and accurate to within 61 mm (0.040(0.04 in.).
6.2.3.3 Procedure—Measure the length of the foot rails. Measure the distance from the forward edge usable moving surface to
the beginning of the foot rail. Locate the center of the usable moving surface. Measure the distance from this point to the end of
the foot rail. Verify that the foot rails extend from either the base of the masts or treadmill uprights or the end of the motor cover
to the rear end caps of the treadmill. Measure the width of the moving surface. Measure the width of the foot rail surface.
6.2.3.4 Pass/Fail Criteria—The dimensions of the foot rails shall conform to dimensional requirements of subsection 4.3.3 of
Specification F2115.
6.2.3.5 Precision and Bias—No information is presented about either the precision or bias of this test 9.2.3 for measuring foot
rail dimensions since the test result is non-quantitative.
(1) Precision—The precision of this test method has not been determined.
(2) Bias—The bias of this test method includes quantitative estimates of the uncertainties of the measuring devices, the
calibrations of testing equipment, and the skill of the operators. At this time, the statements on bias should be limited to
documented performance of particular laboratories.
6.2.4 Moving Surface—The purpose of this test is to verify the dimensional and marking compliance of the moving surface.
6.2.4.1 Apparatus and Set Up—No set up required, dimensional inspection only.
6.2.4.2 Calibration—Verify that the distance measuring equipment is calibrated and accurate to within 61 mm (0.040 in.).
6.2.4.3 Procedure—Consult the moving surface table Specification F2115. Refer to Fig. 3 in Specification F2115 and measure
the length and width of the moving surface. Verify the presence of movement indicators on the moving surface. Verify that they
meet the minimum dimensions specified in Specification F2115. Operate the treadmill and verify that a portion of the movement
indicator is visible at all times as the moving surface rotates.
6.2.4.4 Pass/Fail Criteria—The dimensions of the moving surface shall conform to dimensional requirements of subsection
4.3.4.3Table 1 of Specification F2115.
6.2.4.5 Precision and Bias—No information is presented about either the precision or bias of this test 9.2.4 for measuring
moving surface dimensions since the test result is non-quantitative.
(1) Precision—The precision of this test method has not been determined.
(2) Bias—The bias of this test method includes quantitative estimates of the uncertainties of the measuring devices, the
calibrations of testing equipment, and the skill of the operators. At this time, the statements on bias should be limited to
documented performance of particular laboratories.
6.2.5 Moving Surface Slip Resistance—The purpose of this test is to evaluate the slip resistance of the moving surface of the
treadmill.
6.2.5.1 Apparatus and Set Up—The treadmill shall be set up in accordance with Test Methods F2571. The moving surface shall
be restrained from moving.
6.2.5.2 Calibration—Per Test Methods F2571, no calibration is required.
6.2.5.3 Procedure—Secure the moving surface so that it cannot move. Conduct the test in accordance with Test Methods F2571.
6.2.5.4 Pass/Fail Criteria—Moving surfaces that do not resist foot slippage shall fail the test.
6.2.5.5 Precision and Bias—No information is presented about either the precision or bias of this test 9.2.5for evaluating
moving surface slip resistance since the test result is nonquantitative.non-quantitative.
6.2.6 Handrails—Hand Grips and Grippable Surfaces—The purpose of this test is to verify the presence and dimensional
compliance of the handrails.
6.2.6.1 Apparatus and Set Up—No set up required, dimensional inspection only.
6.2.6.2 Calibration—Verify that the distance measuring equipment is calibrated and accurate to within 61 mm (0.040 in.).
F2106 − 18
6.2.6.3 Procedure—Verify how many front/side handrails are present. Measure the length of all segments of the handrail.
Measure the distance from Line A on the moving surface to the top of the grip surface. Evaluate through the full range of
inclination. If so equipped, measure the horizontal distance between the side handrails.
6.2.6.4 Pass/Fail Criteria—The dimensions of the handrails shall conform to dimensional requirements of subsection 4.3.5 of
Specification F2115.
6.2.6.5 Precision and Bias—No information is presented about either the precision or bias of this test 9.2.6 for measuring
handrail dimensions since the test result is non-quantitative.
(1) Precision—The precision of this test method has not been determined.
(2) Bias—The bias of this test method includes quantitative estimates of the uncertainties of the measuring devices, the
calibrations of testing equipment, and the skill of the operators. At this time, the statements on bias should be limited to
documented performance of particular laboratories.
6.3 Endurance—Endurance Loading—These tests are intended to confirm the endurance of the main components of the
treadmill.
6.3.1 Mechanical Frame Endurance—The stationary moving surface, frame, and structural components of the treadmill are
subjected to a repeated load equal to 1.5 times the maximum user weight as specified by the manufacturer.
6.3.1.1 Apparatus and Set Up—A pneumatic cylinder or other appropriate load application device shall be fixtured to apply load
to the moving surface at the centerline of the treadmill 33 % of the length of the usable moving surface rearward from the front
of the usable moving surface. A 30 by 30 cm (12(11.8 by 1211.8 in.) load application area shall be used for the test. Timing controls
must be supplied to regulate the load application frequency. Provide appropriate instrumentation to assure that the load is consistent
throughout the test duration.
6.3.1.2 Calibration—Verify that the load and frequency measuring equipment is properly calibrated and that the load accuracy
is within 65 % of the applied load and that the frequency measurement is accurate to within 60.2 Hz.
6.3.1.3 Procedure—Elevate the treadmill to the incline that will provide the most stress to the frame and incline system
components and position the treadmill under the load application device. No power shall be connected to the treadmill. Block the
treadmill into position so that the vibrations created by the test do not cause it to move on the floor. Apply power to the load
application system. Adjust the applied load to 1.5 times maximum user weight 65 % at a frequency of 0.5 to 2 Hz. Assure that
the moving surface rebounds completely prior to the next cycle. Verify that the counter is recording the number of cycles. Inspect
the treadmill and test apparatus approximately every 250 000 cycles or as needed to insureensure that the test is functioning
properly. If the design of the treadmill is such that keeping the moving surface stationary for the duration of the test would
abnormally stress the moving surface (that is, a treadmill that employs moving slats rather than a conventional belt and deck), the
moving surface may be repositioned periodically during the test.
6.3.1.4 Pass/Fail Criteria—Upon completion of the test, remove the test unit from the fixture and confirm normal operation of
all functions. The treadmill must meet life requirements per classification of use per subsection 4.44.3 of Specification F2115. The
unit must not show structural cracks or other indications of impending failure.
6.3.1.5 Precision and Bias—No information is presented about either the precision or bias of this test 9.3.1for evaluating frame
endurance since the test result is non-quantitative.
6.3.2 Treadmill Switch and Switch Actuation Mechanism Endurance—Stop, Pause or End functions (see Note 1).
NOTE 1—If several means of stopping exist—for example, Stop switch and Pause switch—each means must be separately tested only if the actuator
types are different.
6.3.2.1 Apparatus and Set Up—Provide a mechanism to repeatedly activate the switch of interest at a rate not to exceed 2 Hz.
and to accumulate a count of actuations. Activation force for testing shall be 1.5 times 1.5× 6 10 % the minimum actuating force
for the particular switch. For push switches, the activating mechanism can be a simulated finger mounted to an appropriate
pneumatic cylinder operating parallel to the line of action of the switch, a test system designed specifically for switch testing
(having an air cylinder plunger with a rounded simulated finger 0.675 diameter, 45 durometer rubber, SF-45), or other appropriate
actuating means. For non-returning switches, a mechanism must be provided to pull out the switch between actuations, with a force
equal to 1.5 times 1.5× 6 10 % the minimum activating force. For lanyard-type, pull-off mechanisms, the switch can be tested as
outlined above without the lanyard attached. If this method is employed, a separate lanyard strength test must be conducted by
applying a load equal to 5 times the maximum activation force of the switch on the lanyard. A means of automatically sensing
activation (for example, sensing a “beep” from the control panel internal electronics) will allow monitoring each activation
response.
6.3.2.2 Calibration—Verify that the load application system applies 1.5 times 6 10 % of the minimum activation force for the
switch to be tested. Using appropriate instrumentation, set the timing of the activator to 0.1 to 0.2 s “on” time and the repeat rate
as desired but not over 2 Hz. Verify activation counter operation for at least 100 cycles. If a lanyard pull is required, the pull force
must guarantee activation.
6.3.2.3 Procedure—Actuate the switch being tested with the load application system and confirm that the force applied causes
the switch to function. The function of the switch shall be confirmed on every actuation of the switch via a circuit that actuates
a counter. A manual check must be made of switch function before and after the full test sequence, and during the automated
F2106 − 18
sequence, some activation-confirming feedback must be observable for each activation. As an alternate to this procedure, if the
switch is installed according to the switch manufacturer’s recommendations, the switch manufacturer’s testing may be used.
6.3.2.4 Pass/Fail Criteria—The switch(s) tested must be functional as described in procedure at the end of the life test
requirements for the classification for use as follows: Consumer Treadmills—1560 actuations without damage or non-functionality.
Institutional Treadmills—46 800 actuations without damage or non-functionality.
6.3.2.5 Precision and Bias—No information is presented about either the precision or bias of this test 9.3.2for evaluating
actuator endurance since the test result is non-quantitative.
6.4 Static Loading of Structures—The purpose of this test is to evaluate the static structural integrity of the deck, foot rails, any
frame structure and user supports (that is, handlebars and side handrails) as applicable to user safety.
6.4.1 Moving Surface and Foot Rails:
6.4.1.1 Apparatus and Set Up—Supply a means of providing a steady state load (pneumatic cylinder, weights, etc.) on the deck
and foot rails, capable of adjustment to apply vertical load to the test member at minimum, mid-range and maximum incline for
the moving surface. Load is to be applied on a 300 by 300 mm (12(11.8 by 1211.8 in.) square area located on the centerline of
the moving surface 33 % of the usable surface length back from the front of the usable moving surface. For the foot rails, the
vertical load is to be applied at the longitudinal centerline of the foot rail.
6.4.1.2 Calibration—Verify load application system is calibrated and is accurate to within 65 % of the applied load.
6.4.1.3 Procedure—With the treadmill fixed to prevent movement and moving surface held stationary, apply a force of three (3)
times 3× the maximum user weight for consumer treadmills or four (4) times 4× maximum user weight for commercial treadmills
to the moving surface as described in 9.3.1.16.3.1.1. The treadmill is to be supported only at points specifically intended for support
during use. Apply the load for 5 to 15 s. Remove the load. Then apply vertical force of 2 times 2× the maximum user weight for
consumer treadmills or 3 times 3× maximum user weight for commercial treadmills to the longitudinal center of one of the foot
rails over an area of 180 mm (12(7.1 in.) in length by 180 mm (12(7.1 in.) in width, or the width of the foot rails, whichever is
less. Maintain load for 5 to 15 s, then remove load.
6.4.1.4 Pass/Fail Criteria—Examine all frame members, welds, joints and the deck for cracks, separations or failure. No visible
evidence of cracks, separations, or other structural damage is allowable. Cosmetic damage is not considered a failure.
6.4.1.5 Precision and Bias—No information is presented about either the precision or bias of this test 9.4.1 for evaluating
structural integrity of moving surface and foot rails since the test result is nonquantitative.non-quantitative.
(1) Precision—The precision of this test method has not been determined.
(2) Bias—The bias of this test method includes quantitative estimates of the uncertainties of the measuring devices, the
calibrations of testing equipment, and the skill of the operators. At this time, the statements on bias should be limited to
documented performance of particular laboratories.
9.4.2 Folding Treadmills—The purpose of this test is to confirm that folding treadmills are sufficiently stable such that the
handrail/console does not fold down unexpectedly during use. Further, this test is to visually confirm the presence and function
of the locking mechanism(s) to secure the deck and moving surface in the folded-up storage position for treadmills where it is
folded up for storage or to secure the handrails in the use position.
9.4.2.1 Apparatus and Set Up—Supply a means of providing a steady state horizontal load equal to 180 N (40 lb). Possible
methods of providing this force include, but are not limited to, pneumatic cylinder(s) or weights with a pulley system.
9.4.2.2 Calibration—Verify and or calibrate the load application apparatus to confirm accuracy to within 65 % of the applied
load.
9.4.2.3 Procedure—Apply a load of 668 N (150 lb) on a 300 by 300-mm area ⁄3 of the length of the usable surface forward
of the rear roller. Apply the 180 N (40 lb) load horizontally to the topmost handrail in the direction that could cause folding. Verify
that the handlebars do not fold down when this load is applied. For treadmills where the deck and moving surface fold up for
storage verify that the deck and moving surface do not reach the balance point before the latching mechanism engages. Visually
verify that a locking mechanism is present to secure the deck and moving surface in the storage position. For treadmills where the
handrails and upright fold up for use verify that the handrails and uprights do not reach their balance point before the locking
mechanism engages. Visually verify the presence of a locking mechanism to secure the handrails and uprights in the use position.
9.4.2.4 Pass/Fail Criteria—For the first test, the treadmill shall not fold under the applied load. In the second test, the balance
point must not be achieved until the locking mechanism is activated.
9.4.2.5 Precision and Bias—No information is presented about either the precision or bias of test 9.4.2 for evaluating the
resistance to unintentional folding of folding treadmills since the test result is non-quantitative.
9.5 Overheating—For an empirically determined worst case thermal combination of moving surface speed and incline setting,
exterior surface temperatures shall be checked.
9.5.1 Apparatus and Set Up—Use a load application device that will provide a load to the motor and drive system equivalent
to a maximum weight user operating at maximum speed. Use appropriate temperature measuring instruments, such as
thermocouples or other equivalent devices, to measure the temperature of the surfaces and components of the treadmill. The test
should be conducted at an ambient temperature of 25 6 5°C (77 6 9°F).
F2106 − 18
9.5.2 Calibration—Verify that the applied load results in an input power which generates a worst case thermal condition for a
maximum user weight user (normally this is at low operating speeds). Verify that the thermocouples and infrared gun are properly
calibrated and accurate to within 61°C (2°F). Ambient temperature should be showing on all temperature readouts prior to running
the treadmill.
9.5.3 Procedure—Operate treadmill until the temperatures stabilize at a combination of load and speed that produces the worst
case thermal conditions as determined in 9.5.2. At the end of this period, measure and record temperature for all locations.
9.5.4 Pass/Fail Criteria—All temperatures must be below 60°C (140°F) for metallic surfaces and 85°C (185°F) for non-metallic
surfaces. If the test is conducted at an ambient temperature of other than 25°C (77°F), the temperatures are to be corrected to that
temperature.
9.5.5 Precision and Bias—No information is presented about either the precision or bias of test 9.5 for overheating since the
test result is non-quantitative.
6.5 Adjustable Incline System Function—Note that treadmills that use a mechanical incline position system such as a ratchet,
pin, or other means to provide limited fixed positions, and that do not actually provide the energy to adjust the incline, are excluded
from this test.
6.5.1 Apparatus and Set Up—A combination of vertical displacement and time measuring instruments which yields a combined
accuracy of 2.5 mm/s (0.1 in./s) will be necessary for the velocity measurements. The velocity will be determined by dividing the
displacement by the time required for movement. As an alternate, an instrument that directly measures vertical velocity having the
accuracy stated above may be used. The instrumentation is to be fixtured to measure the relative velocity of any pinch or shear
points that occur during actuation of the incline system. A method of applying maximum specified user weight that is independent
of angle of inclination must be provided.
6.5.2 Calibration—Verify that the instrumentation is properly calibrated and yields an accuracy within 2.5 mm/s (0.1 in./s) of
actual.
6.5.3 Procedure—With the treadmill in its minimum incline position and no load applied, begin recording data. Operate the
incline control to elevate the treadmill to maximum elevation as quickly as possible. Once the treadmill has reached full inclination,
operate the incline control to return the treadmill to its minimum inclination as quickly as possible.
6.5.3.1 Apply maximum user weight to a point 33 % of the length of the usable moving surface rearward from the front of the
usable moving surface. Repeat step the procedure described in 9.6.36.5.3.
6.5.4 Pass/Fail Criteria—The maximum measured speed shall not exceed 25 mm/s (l(l.0 in./s).
6.5.5 Precision and Bias—No information is presented about either the precision or bias of this test 9.6 for measuring velocity
of the incline system since the test result is non-quantitative.
(1) Precision—The precision of this test method has not been determined.
(2) Bias—The bias of this test method includes quantitative estimates of the uncertainties of the measuring devices, the
calibrations of testing equipment, and the skill of the operators. At this time, the statements on bias should be limited to
documented performance of particular laboratories.
6.6 User Interface Parameters—Controls—This section is to verify that the proper stop controls exist and that the accuracy of
the speed readout is adequate.
6.6.1 Verification of Emergency Stop—This test is a visual inspection of the unit to verify the presence of an emergency stop.
6.6.1.1 Apparatus and Set Up—The sample shall be set up as described in Section 5.
6.6.1.2 Calibration—No calibration required. Visual inspection only.
6.6.1.3 Procedure—Verify that stop mechanisms per Specification F2115 exist.
6.6.1.4 Pass/Fail Criteria—Emergency stop mechanisms shall conform to the requirements of Specification F2115.
6.6.1.5 Precision and Bias—No information is presented about either the precision or bias of this test for evaluating emergency
stop actuation mechanisms since the test result is non-quantitative.
6.6.2 Verification of Stop Mechanism—Mechanism Function—This test is a visual and functional inspection of the unit to
insureensure that stop mechanisms existfunction and are user accessible.
6.6.2.1 Apparatus and Set Up—Place the treadmill on a level surface. Supply power to the treadmill.
6.6.2.2 Calibration—No calibration required. Visual and functional inspection only.
6.6.2.3 Procedure—Verify that stop mechanisms per subsection 4.8.2 of Specification F2115 exist. Operate all stop mechanisms
to verify accessibility. Perform the testVerify the accessibility with a 5th percentile female running at 8 km/h (5 mph). Perform
function tests at minimum and maximum incline at 8 kph (5km/h (5.0 mph) or maximum speed, whichever is less, with a 100 kg
(220(220.5 lb) user.
6.6.2.4 Pass/Fail Criteria—Stop devices must be readily accessible and cause the moving surface to decelerate to rest and stop
the motion of the incline system.
6.6.2.5 Precision and Bias—No information is presented about either the precision or bias of this test 9.7.1for evaluating
accessibility and function of stop mechanisms since the test result is non-quantitativenon-quantitative.
6.6.3 Emergency Stop Appearance—This test is a visual inspection of the sample to ensure the compliance of the emergency
stop color, size and shape.
6.6.3.1 Apparatus and Set Up—The sample shall be set up as described in Section 5.
F2106 − 18
6.6.3.2 Calibration—No calibration required. Visual test only.
6.6.3.3 Procedure—Verify that the emergency stop color, shape, and size differ from other available controls.
6.6.3.4 Pass/Fail Criteria—Emergency stop color, shape, and size shall conform to the requirements of Specification F2115.
6.6.3.5 Precision and Bias—No information is presented about either the precision or bias of this test for evaluating the
emergency stop color, shape and size since the test result is non-quantitative.
6.6.4 Emergency Stop Location—This test is a visual and dimensional inspection of the sample to ensure the emergency stop
location.
6.6.4.1 Apparatus and Set Up—The sample shall be set up as described in Section 5.
6.6.4.2 Calibration—Verify that the distance measuring equipment is calibrated and accurate to within 1 mm (0.040 in.).
6.6.4.3 Procedure—Inspect the treadmill to verify that there is an emergency stop centrally located on the unit or that there is
an emergency stop on each side of the console. For centrally located emergency stops, measure the distance of the centerline of
the emergency stop to the centerline longitudinal centerline of the unit.
6.6.4.4 Pass/Fail Criteria—Emergency stop location shall conform to the requirements of Specification F2115.
6.6.4.5 Precision and Bias—No information is presented about either the precision or bias of this test since the test result is
non-quantitative.
(1) Precision—The precision of this test method has not been determined.
(2) Bias—The bias of this test method includes quantitative estimates of the uncertainties of the measuring devices, the
calibrations of testing equipment, and the skill of the operators. At this time, the statements on bias should be limited to
documented performance of particular laboratories.
6.6.5 Push Button Stop Switch Size—This test is a dimensional inspection of the sample to ensure Emergency stop push button
size.
6.6.5.1 Apparatus and Set Up—The sample shall be set up as described in Section 5.
6.6.5.2 Calibration—Verify that the distance measuring equipment is calibrated and accurate to within 1 mm (0.040 in.).
6.6.5.3 Procedure—Measure the surface area of the push button stop switch.
6.6.5.4 Pass/Fail Criteria—Push button stop switch size shall conform to the requirements of Specification F2115.
6.6.5.5 Precision and Bias—No information is presented about either the precision or bias of this test since the test result is
non-quantitative.
(1) Precision—The precision of this test method has not been determined.
(2) Bias—The bias of this test method includes quantitative estimates of the uncertainties of the measuring devices, the
calibrations of testing equipment, and the skill of the operators. At this time, the statements on bias should be limited to
documented performance of particular laboratories.
6.6.6 Push Button Stop Switch Color—This test is a visual inspection of the sample to ensure the compliance of the push button
stop switch color.
6.6.6.1 Apparatus and Set Up—The sample shall be set up as described in Section 5.
6.6.6.2 Calibration—No calibration required. Visual test only.
6.6.6.3 Procedure—Verify that the push button stop switch color is red.
6.6.6.4 Pass/Fail Criteria—ush button stop switch shall conform to the requirements of Specification F2115.
6.6.6.5 Precision and Bias—No information is presented about either the precision or bias of this test for evaluating the push
button stop switch color since the test result is non-quantitative.
6.6.7 Pull-Cord Emergency Stop Location—This test is a dimensional inspection of the sample to ensure the pull-cord
emergency stop location.
6.6.7.1 Apparatus and Set Up—The sample shall be set up as described in Section 5.
6.6.7.2 Calibration—Verify that the distance measuring equipment is calibrated and accurate to within 1 mm (0.040 in.).
6.6.7.3 Procedure—Measure the distance of the pull-cord emergency stop centerline to the centerline longitudinal centerline of
the unit.
6.6.7.4 Pass/Fail Criteria—Emergency stop location shall conform to the requirements of Specification F2115.
6.6.7.5 Precision and Bias—No information is presented about either the precision or bias of this test since the test result is
non-quantitative.
(1) Precision—The precision of this test method has not been determined.
(2) Bias—The bias of this test method includes quantitative estimates of the uncertainties of the measuring devices, the
calibrations of testing equipment, and the skill of the operators. At this time, the statements on bias should be limited to
documented performance of particular laboratories.
6.6.8 Pull-Cord Activation—This test is a functional inspection of the sample to ensure the compliance of the pull-cord
activation.
6.6.8.1 Apparatus and Set Up—The sample shall be set up as described in Section 5.
6.6.8.2 Calibration—Verify that the force meter is calibrated per operator instruction.
F2106 − 18
6.6.8.3 Procedure—While pulling in a horizontal direction, measure force to disengage the pull cord from the product. Repeat
the process to measure the force to disengage pull clip from clothing. Compare the forces. Repeat the function test described in
6.6.2.
6.6.8.4 Pass/Fail Criteria—Verify that the moving surface decelerates to rest and that the incline function ceases when the pull
cord is activated with a force equal to 50% of the force required to disengage the cord clip from exercise clothing.
6.6.8.5 Precision and Bias—No information is presented about either the precision or bias of this test since the test result is
non-quantitative.
(1) Precision—The precision of this test method has not been determined.
(2) Bias—The bias of this test method includes quantitative estimates of the uncertainties of the measuring devices, the
calibrations of testing equipment, and the skill of the operators. At this time, the statements on bias should be limited to
documented performance of particular laboratories.
6.6.9 Pull-Cord Length—This test is a dimensional inspection of the sample to ensure the compliance of the pull-cord length.
6.6.9.1 Apparatus and Set Up—The sample shall be set up as described in Section 5.
6.6.9.2 Calibration—Verify that the distance measuring equipment is calibrated and accurate to within 1 mm (0.040 in.).
6.6.9.3 Procedure—Measure the pull in its usable position.
6.6.9.4 Pass/Fail Criteria—Pull length shall conform to the requirements of Specification F2115.
6.6.9.5 Precision and Bias—No information is presented about either the precision or bias of this test for evaluating the pull
length since the test result is non-quantitative.
6.6.10 Pull-Cord Color—This test is a visual inspection of the sample to ensure the compliance of the pull-cord color.
6.6.10.1 Apparatus and Set Up—The sample shall be set up as described in Section 5.
6.6.10.2 Calibration—No calibration required. Visual test only.
6.6.10.3 Procedure—Verify that the pull-cord actuator color is red.
6.6.10.4 Pass/Fail Criteria—Push button stop switch shall conform to the requirements of Specification F2115.
6.6.10.5 Precision and Bias—No information is presented about either the precision or bias of this test since the test result is
non-quantitative.
6.6.11 Speed Indication Accuracy:
6.6.11.1 Apparatus and Set Up—Install a speed measurement device on moving surface.
6.6.11.2 Calibration—Verify sensor is properly calibrated and is accurate to 60.2 kph (65 %).km/h (60.1 mph).
6.6.11.3 Procedure—Operate the treadmill at 33 %, 66 % and 100 % of maximum speed. Measure the moving surface speed
with the speed sensor.
6.6.11.4 Pass/Fail Criteria—The speed sensor reading for each
...

Questions, Comments and Discussion

Ask us and Technical Secretary will try to provide an answer. You can facilitate discussion about the standard in here.

Loading comments...