Ambient air quality - Standard method for the determination of mercury deposition

This European Standard specifies a method for the determination of the total deposition of mercury. This standard can be used within the framework of the European Council Directive on Ambient Air Quality Assessment and Management and Directive 2004/107/EC. Performance requirements with which the method should comply are specified in this European Standard. The performance characteristics of the method were determined in comparative field validation tests carried out at two European locations.
This European Standard is applicable to background sites that are in accordance with the requirements of Directive 2004/107/EC and to urban and industrial sites.
This standard allows the sampling of deposition using cylindrical deposition gauges, and analysis using Cold Vapour Atomic Absorption Spectrometry (CVAAS) or Cold Vapour Atomic Fluorescence Spectrometry (CVAFS) following existing harmonised and standardised procedures. The standard is applicable for the measurement of mercury in deposition between 1 ng/(m2•d) and 100 ng/(m2•d).
The standard is validated for the deposition range listed in Table 1.
NOTE   The range given is based upon the values measured in the field validation test. The upper and lower limits are the observed minimum and maximum values measured during the field validation tests. The actual lower limits of the working range depends on the variability of the laboratory blank, for bulk and wet-only collectors, and the precipitation. The method can be applied to higher and lower deposition rates provided that the collection characteristics are not compromised.

Außenluftbeschaffenheit - Standardisiertes Verfahren zur Bestimmung der Quecksilberdeposition

Diese Europäische Norm legt ein Verfahren für die Bestimmung der Gesamtdeposition von Queck¬silber fest. Die vorliegende Norm kann im Rahmen der Richtlinie des Europäischen Rates über die Beurteilung und die Kontrolle der Außenluftbeschaffenheit und der Richtlinie 2004/107/EG angewendet werden. In der vorliegenden Europäischen Norm sind die Leistungsanforderungen festgelegt, die dieses Verfahren erfüllen sollte. Die Verfahrenskenngrößen des Verfahrens wurden mit Vergleichsuntersuchungen zur Validierung von Feldversuchen bestimmt, die an zwei Standorten in Europa durchgeführt wurden.
Diese Europäische Norm ist anwendbar auf Hintergrundstationen, die den Anforderungen der Richtlinie 2004/107/EG entsprechen, und auf Stadtgebiete sowie Gebiete mit Industrieeinfluss.
Das Verfahren ermöglicht die Probenahme zur Bestimmung von Depositionen mit zylindrischen Auffanggefäßen und die Analyse mit der Kaltdampf Atomabsorptionsspektrometrie (CVAAS) oder Kaltdampf-Atomfluoreszenzspektro¬metrie (CVAFS) nach bestehenden, harmonisierten und standardisierten Verfahren. Das Verfahren ist anwendbar für die Messung von Quecksilberdepositionen zwischen 1 ng/(m2  d) und 100 ng/(m2  d).
Die Norm wurde für den in Tabelle 1 aufgeführten Depositionsbereich validiert.

Qualité de l'air ambiant - Méthode normalisée pour la détermination des dépôts de mercure

La présente Norme européenne spécifie une méthode pour le mesurage du dépôt total de mercure. La
présente norme peut être utilisée dans le cadre de la Directive européenne du Conseil concernant l’évaluation
et la gestion de la qualité de l’air ambiant et de la Directive 2004/107/CE. La présente Norme européenne
spécifie les exigences de performance auxquelles doit se conformer la méthode. Les caractéristiques de
performance de la méthode ont été déterminées par des essais comparatifs de validation sur le terrain
réalisés sur deux sites européens.
La présente Norme européenne s’applique aux sites de fond conformes aux exigences de la
Directive 2004/107/CE et aux sites urbains et industriels.
La présente norme spécifie l’échantillonnage de dépôts à l’aide de jauges de dépôt cylindriques, et l’analyse
par spectrométrie d’absorption atomique de vapeur froide (SAAVF) ou par spectrométrie de fluorescence
atomique de vapeur froide (SFAVF) suivant les procédures normalisées et harmonisées existantes. La norme
s’applique au mesurage du mercure dans les dépôts entre 1 ng/(m2·j) et 100 ng/(m2·j).
La norme est validée pour la plage de dépôt donnée dans le Tableau 1.

Kakovost zunanjega zraka - Standardna metoda za določevanje živega srebra v usedlinah

Ta evropski standard opredeljuje metodo za določevanje celotnega živega srebra v usedlinah. Ta standard se lahko uporablja v okviru Direktive Evropskega sveta o ocenjevanju kakovosti zunanjega zraka in Direktive 2004/107/ES. V tem evropskem standardu so določene zahteve za delovanje, ki jih mora izpolnjevati ta metoda. Značilnosti delovanja metode so bile določene s primerjalnimi validacijskimi preskusi na terenu, opravljenimi na dveh evropskih lokacijah. Ta evropski standard velja za zaledna območja, ki so v skladu z zahtevani Direktive 2004/107/ES, ter za urbana in industrijska območja. Ta standard omogoča vzorčenje usedlin s cilindričnimi merilniki usedlin in analizo z atomsko absorbcijsko spektrometrijo hladnih par (CVAAS) ali atomsko fluorescenčno spektrometrijo hladnih par (CVAFS), ki sledijo usklajenim in standardiziranim postopkom. Ta standard velja za merjenje živega srebra v usedlinah med 1 ng/(m2•d) in 100 ng/(m2•d). Standard je potrjen za razpone usedlin, navedene v preglednici 1.

General Information

Status
Published
Publication Date
01-Jun-2010
Withdrawal Date
30-Dec-2010
Current Stage
9093 - Decision to confirm - Review Enquiry
Completion Date
05-Jun-2020

Buy Standard

Standard
EN 15853:2010
English language
33 pages
sale 10% off
Preview
sale 10% off
Preview
e-Library read for
1 day

Standards Content (Sample)


2003-01.Slovenski inštitut za standardizacijo. Razmnoževanje celote ali delov tega standarda ni dovoljeno.XVHGOLQDKAußenluftbeschaffenheit - Standardisiertes Verfahren zur Bestimmung der QuecksilberdepositionQualité de l'air ambiant - Méthode normalisée pour la détermination des dépôts de mercureAmbient air quality - Standard method for the determination of mercury deposition13.040.20Kakovost okoljskega zrakaAmbient atmospheresICS:Ta slovenski standard je istoveten z:EN 15853:2010SIST EN 15853:2010en,fr,de01-november-2010SIST EN 15853:2010SLOVENSKI
STANDARD
EUROPEAN STANDARD NORME EUROPÉENNE EUROPÄISCHE NORM
EN 15853
June 2010 ICS 13.040.20 English Version
Ambient air quality - Standard method for the determination of mercury deposition
Qualité de l'air ambiant - Méthode normalisée pour la détermination des dépôts de mercure
Außenluftbeschaffenheit - Standardisiertes Verfahren zur Bestimmung der Quecksilberdeposition This European Standard was approved by CEN on 5 May 2010.
CEN members are bound to comply with the CEN/CENELEC Internal Regulations which stipulate the conditions for giving this European Standard the status of a national standard without any alteration. Up-to-date lists and bibliographical references concerning such national standards may be obtained on application to the CEN Management Centre or to any CEN member.
This European Standard exists in three official versions (English, French, German). A version in any other language made by translation under the responsibility of a CEN member into its own language and notified to the CEN Management Centre has the same status as the official versions.
CEN members are the national standards bodies of Austria, Belgium, Bulgaria, Croatia, Cyprus, Czech Republic, Denmark, Estonia, Finland, France, Germany, Greece, Hungary, Iceland, Ireland, Italy, Latvia, Lithuania, Luxembourg, Malta, Netherlands, Norway, Poland, Portugal, Romania, Slovakia, Slovenia, Spain, Sweden, Switzerland and United Kingdom.
EUROPEAN COMMITTEE FOR STANDARDIZATION
COMITÉ EUROPÉEN DE NORMALISATION EUROPÄISCHES KOMITEE FÜR NORMUNG
Management Centre:
Avenue Marnix 17,
B-1000 Brussels © 2010 CEN All rights of exploitation in any form and by any means reserved worldwide for CEN national Members. Ref. No. EN 15853:2010: ESIST EN 15853:2010

Sampling equipment that can be used for precipitation sampling . 17Annex B (informative)
Sampling procedure . 20Annex C (informative)
Analytical procedure . 22Annex D (informative)
Summary of field trial validation tests . 27Annex E (informative)
Relationship between this European Standard and the Essential Requirements of EU Directives . 32Bibliography . 33 SIST EN 15853:2010

This European Standard specifies a method for the determination of the total deposition of mercury. This standard can be used within the framework of the European Council Directive on Ambient Air Quality Assessment and Management and Directive 2004/107/EC. Performance requirements with which the method should comply are specified in this European Standard. The performance characteristics of the method were determined in comparative field validation tests carried out at two European locations. This European Standard is applicable to background sites that are in accordance with the requirements of Directive 2004/107/EC and to urban and industrial sites.
This standard allows the sampling of deposition using cylindrical deposition gauges, and analysis using Cold Vapour Atomic Absorption Spectrometry (CVAAS) or Cold Vapour Atomic Fluorescence Spectrometry (CVAFS) following existing harmonised and standardised procedures. The standard is applicable for the measurement of mercury in deposition between 1 ng/(m2·d) and 100 ng/(m2·d). The standard is validated for the deposition range listed in Table 1. Table 1 — Working range of this standard method Working range ng/(m2.d) Lower limit Upper limit 1 100
NOTE The range given is based upon the values measured in the field validation test. The upper and lower limits are the observed minimum and maximum values measured during the field validation tests. The actual lower limits of the working range depends on the variability of the laboratory blank, for bulk and wet-only collectors, and the precipitation. The method can be applied to higher and lower deposition rates provided that the collection characteristics are not compromised. 2 Normative references The following referenced documents are indispensable for the application of this document. For dated references, only the edition cited applies. For undated references, the latest edition of the referenced document (including any amendments) applies.
ENV 13005, Guide to the expression of uncertainty in measurement CR 14377, Air quality — Approach to uncertainty estimation for ambient air reference measurement methods EN ISO 17852, Water quality — Determination of mercury — Method using atomic fluorescence spectrometry (ISO 17852:2006) EN ISO 20988, Air quality — Guidelines for estimating measurement uncertainty (ISO 20988:2007) 3 Terms, definitions and abbreviations For the purposes of this document, the following terms, definitions and abbreviations apply. SIST EN 15853:2010

outdoor air in the troposphere, excluding workplace air
3.1.2 atmospheric deposition mass flow rate (unit of mass per unit of area per unit of time) describing the process of the transfer of any species from the atmosphere to an environmental compartment, e.g. air to water, air to soil
NOTE
Atmospheric deposition is the sum of the depositions of sedimenting particles, non-sedimenting particles and gases. 3.1.3 bulk collector equipment to collect bulk deposition
NOTE
Bulk collectors sample deposition at all times. A bulk sampler can consist of a funnel-bottle combination or a wide mouthed jar. 3.1.4 bulk deposition sum of the deposition of sedimenting wet and dry particles
NOTE Bulk deposition is a part of the atmospheric deposition. 3.1.5
combined standard measurement uncertainty
standard measurement uncertainty that is obtained using the individual standard measurement uncertainties associated with the input quantities in a measurement model NOTE
In case of correlations of input quantities in a measurement model, covariances should also be taken into account when calculating the combined standard measurement uncertainty [ISO/IEC Guide 99 (VIM)]. 3.1.6 coverage factor
number larger than one by which a combined standard measurement uncertainty is multiplied to obtain an expanded measurement uncertainty NOTE
A coverage factor is usually symbolized k [ISO/IEC Guide 99 (VIM)]. 3.1.7 detection limit (DL), instrumental lowest amount of an analyte that is detectable using an instrument as determined by repeated measurements of a reagent blank 3.1.8 detection limit (DL), method lowest amount of an analyte detectable after the whole measurement process as determined by repeated measurements of different field blanks 3.1.9 dry deposition sum of the deposition of sedimenting dry particles, non sedimenting particles and gases
NOTE
Dry deposition includes the following processes: atmospheric turbulent diffusion, adsorption, absorption, impaction and gravitational settling. The dry deposition process is affected by the type of underlying surface and surface conditions. Dry deposition is a part of the atmospheric deposition. SIST EN 15853:2010

expanded uncertainty
product of a combined standard measurement uncertainty and a factor larger than the number one
NOTE 1
The factor depends upon the type of probability distribution of the output quantity in a measurement model and on the selected coverage probability.
NOTE 2
The term "factor" in this definition refers to a coverage factor.
NOTE 3
Expanded measurement uncertainty is termed "overall uncertainty" in paragraph 5 of Recommendation INC-1 (1980) (see the GUM) and simply "uncertainty" in IEC documents [ISO/IEC Guide 99 (VIM)].
NOTE 4 For the purpose of this document the expanded uncertainty is the combined standard uncertainty multiplied by a coverage factor k = 2 resulting in an interval with a level of confidence of 95 %.
3.1.11 field blank artificial sample (e.g. de-ionised water) taken through the same procedure as the precipitation sample, except that this has not been exposed to precipitation NOTE
It is transported to the sampling site, mounted in the sampling unit, returned to the laboratory and worked up in the same way as the deposition sample. 3.1.12 laboratory blank artificial sample (e.g. de-ionised water) taken through the same procedure in the laboratory as the precipitation sample, worked up in the same way as the deposition sample 3.1.13 measurement uncertainty non-negative parameter characterizing the dispersion of the quantity values being attributed to a measurand, based on the information used
NOTE 1 Measurement uncertainty includes components arising from systematic effects, such as components associated with corrections and the assigned quantity values of measurement standards, as well as the definitional uncertainty. Sometimes estimated systematic effects are not corrected for but, instead, associated measurement uncertainty components are incorporated. NOTE 2 The parameter may be, for example, a standard deviation called standard measurement uncertainty (or a specified multiple of it), or the half-width of an interval, having a stated coverage probability. [ISO/IEC Guide 99:2007 (VIM)]
3.1.14 precipitation rain, snow, sleet, snow pellets ("graupel") and hail 3.1.15 reagent blank artificial sample (e.g. de-ionised water) processed through the analytical measurement steps 3.1.16 wet deposition sum of depositions of sedimenting wet particles and droplets
NOTE Wet particles and droplets in the atmosphere undergo the process of scavenging of any gases and/or particles. Wet deposition is a part of the atmospheric deposition.
3.1.17 wet-only collector equipment to collect wet deposition, typically a funnel-bottle combination SIST EN 15853:2010

Wet-only collectors collect only during wet precipitation events 3.2 Abbreviations CRM
Certified Reference Material CVAAS Cold Vapour Atomic Absorption Spectrometry CVAFS Cold Vapour Atomic Fluorescence Spectrometry EMEP Co-operative Programme for Monitoring and Evaluation of the Long-range Transmission of Air
pollutants in Europe (European Monitoring and Evaluation Programme) WMO/GAW World Meteorologi
...

Questions, Comments and Discussion

Ask us and Technical Secretary will try to provide an answer. You can facilitate discussion about the standard in here.