Evaluation and routine testing in medical imaging departments ­ - Part 2­-6: Constancy tests ­ imaging performance of computed tomography X-ray equipment

This part of IEC 61223 provides assistance in performing constancy tests on a CT scanner. It applies to those components of CT scanners which influence the image quality, patient dose and positioning. This standard
- defines the essential parameters which describe the performance of CT scanners with regard to image quality, patient dose and positioning; the list of parameters to be tested can be found in section 4.5;
- defines the methods of testing the essential parameters;
- provides criteria to be applied in the evaluation of data for compliance with the tolerances of the parameters specified by the accompanying documents and with respect to baseline values. These methods rely on non-invasive measurements, using appropriate test equipment, performed to ensure that the functional performance of equipment meets established criteria or to enable the early recognition of changes in the properties of components of the equipment. The aim is to verify compliance with specifications affecting the image quality, patient dose and patient positioning.

Essais d'évaluation et de routine dans les services d'imagerie médicale - ­ Partie 2­-6: Essais de constance ­ performance d'imagerie des équipements de tomodensitométrie à rayonnement X

La présente partie de la CEI 61223 fournit une aide à la réalisation des essais de constance des tomodensitomètres. Elle s'applique aux composants des tomodensitomètres qui influencent la qualité de l'image, la dose patient et le positionnement. La présente norme
- définit les paramètres essentiels qui décrivent les performances des tomodensitomètres concernant la qualité de l'image, la dose patient et le positionnement; la liste des paramètres à soumettre aux essais peut être trouvée au paragraphe 4.5;
- définit les méthodes d'essai des paramètres essentiels;
- fournit les critères à appliquer lors de l'évaluation des données pour établir la conformité aux tolérances des paramètres spécifiés par les documents d'accompagnement et concernant les valeurs de base. Ces méthodes reposent sur des mesurages non invasifs, utilisant des équipements d'essai appropriés, réalisés pour s'assurer que la performance fonctionnelle de l'équipement satisfait aux critères établis ou pour permettre une reconnaissance précoce des variations des propriétés des composants de l'équipements. Le but recherché est la vérification de la conformité aux spécifications concernant la qualité de l'image, la dose patient et le positionnement du patient.

General Information

Status
Replaced
Publication Date
27-Nov-2006
Current Stage
DELPUB - Deleted Publication
Completion Date
06-Dec-2019
Ref Project

Relations

Buy Standard

Standard
IEC 61223-2-6:2006 - Evaluation and routine testing in medical imaging departments ­ - Part 2­-6: Constancy tests ­ imaging performance of computed tomography X-ray equipment Released:11/28/2006 Isbn:2831888832
English and French language
67 pages
sale 15% off
Preview
sale 15% off
Preview

Standards Content (Sample)


NORME CEI
INTERNATIONALE
IEC
61223-2-6
INTERNATIONAL
Deuxième édition
STANDARD
Second edition
2006-11
Essais d’évaluation et de routine dans
les services d’imagerie médicale –
Partie 2-6:
Essais de constance –
Performance d’imagerie des équipements
de tomodensitométrie à rayonnement X

Evaluation and routine testing in
medical imaging departments –
Part 2-6:
Constancy tests –
Imaging performance of computed
tomography X-ray equipment
Numéro de référence
Reference number
CEI/IEC 61223-2-6:2006
Numérotation des publications Publication numbering

Depuis le 1er janvier 1997, les publications de la CEI As from 1 January 1997 all IEC publications are

sont numérotées à partir de 60000. Ainsi, la CEI 34-1 issued with a designation in the 60000 series. For

devient la CEI 60034-1. example, IEC 34-1 is now referred to as IEC 60034-1.

Editions consolidées Consolidated editions

Les versions consolidées de certaines publications de la The IEC is now publishing consolidated versions of its

CEI incorporant les amendements sont disponibles. Par publications. For example, edition numbers 1.0, 1.1

exemple, les numéros d’édition 1.0, 1.1 et 1.2 indiquent and 1.2 refer, respectively, to the base publication,
respectivement la publication de base, la publication de the base publication incorporating amendment 1 and
base incorporant l’amendement 1, et la publication de the base publication incorporating amendments 1
base incorporant les amendements 1 et 2. and 2.
Informations supplémentaires Further information on IEC publications
sur les publications de la CEI
Le contenu technique des publications de la CEI est The technical content of IEC publications is kept
constamment revu par la CEI afin qu'il reflète l'état under constant review by the IEC, thus ensuring that
actuel de la technique. Des renseignements relatifs à the content reflects current technology. Information
cette publication, y compris sa validité, sont dispo- relating to this publication, including its validity, is
nibles dans le Catalogue des publications de la CEI available in the IEC Catalogue of publications
(voir ci-dessous) en plus des nouvelles éditions, (see below) in addition to new editions, amendments
amendements et corrigenda. Des informations sur les and corrigenda. Information on the subjects under
sujets à l’étude et l’avancement des travaux entrepris consideration and work in progress undertaken by the
par le comité d’études qui a élaboré cette publication, technical committee which has prepared this
ainsi que la liste des publications parues, sont publication, as well as the list of publications issued,
également disponibles par l’intermédiaire de: is also available from the following:
• Site web de la CEI (www.iec.ch) • IEC Web Site (www.iec.ch)
• Catalogue des publications de la CEI • Catalogue of IEC publications
Le catalogue en ligne sur le site web de la CEI The on-line catalogue on the IEC web site
(www.iec.ch/searchpub) vous permet de faire des (www.iec.ch/searchpub) enables you to search by a
recherches en utilisant de nombreux critères, variety of criteria including text searches,
comprenant des recherches textuelles, par comité technical committees and date of publication. On-
d’études ou date de publication. Des informations en line information is also available on recently
ligne sont également disponibles sur les nouvelles issued publications, withdrawn and replaced
publications, les publications remplacées ou retirées, publications, as well as corrigenda.
ainsi que sur les corrigenda.
• IEC Just Published • IEC Just Published
Ce résumé des dernières publications parues This summary of recently issued publications
(www.iec.ch/online_news/justpub) est aussi dispo- (www.iec.ch/online_news/justpub) is also available
nible par courrier électronique. Veuillez prendre by email. Please contact the Customer Service
contact avec le Service client (voir ci-dessous) Centre (see below) for further information.
pour plus d’informations.
• Service clients • Customer Service Centre
Si vous avez des questions au sujet de cette If you have any questions regarding this
publication ou avez besoin de renseignements publication or need further assistance, please
supplémentaires, prenez contact avec le Service contact the Customer Service Centre:
clients:
Email: custserv@iec.ch Email: custserv@iec.ch
Tél: +41 22 919 02 11 Tel: +41 22 919 02 11
Fax: +41 22 919 03 00 Fax: +41 22 919 03 00
.
NORME CEI
INTERNATIONALE
IEC
61223-2-6
INTERNATIONAL
Deuxième édition
STANDARD
Second edition
2006-11
Essais d’évaluation et de routine dans
les services d’imagerie médicale –
Partie 2-6:
Essais de constance –
Performance d’imagerie des équipements
de tomodensitométrie à rayonnement X

Evaluation and routine testing in
medical imaging departments –
Part 2-6:
Constancy tests –
Imaging performance of computed
tomography X-ray equipment
© IEC 2006 Droits de reproduction réservés ⎯ Copyright - all rights reserved
Aucune partie de cette publication ne peut être reproduite ni No part of this publication may be reproduced or utilized in any
utilisée sous quelque forme que ce soit et par aucun procédé, form or by any means, electronic or mechanical, including
électronique ou mécanique, y compris la photocopie et les photocopying and microfilm, without permission in writing from
microfilms, sans l'accord écrit de l'éditeur. the publisher.
International Electrotechnical Commission, 3, rue de Varembé, PO Box 131, CH-1211 Geneva 20, Switzerland
Telephone: +41 22 919 02 11 Telefax: +41 22 919 03 00 E-mail: inmail@iec.ch Web: www.iec.ch
CODE PRIX
V
PRICE CODE
Commission Electrotechnique Internationale
International Electrotechnical Commission
МеждународнаяЭлектротехническаяКомиссия
Pour prix, voir catalogue en vigueur
For price, see current catalogue

– 2– 61223-2-6 © CEI:2006
SOMMAIRE
AVANT-PROPOS.4

INTRODUCTION.8

1 Domaine d’application et objectif .10

2 Références normatives.10

3 Termes et définitions .12

4 Aspects généraux des ESSAIS DE CONSTANCE .20
4.1 Conditions préalables.20
4.2 Conditions générales à prendre en compte dans les méthodes d'essais .20
4.3 Etablissement des VALEURS DE BASE .22
4.4 Identification des équipements, des instruments et des conditions d'essais.22
4.5 Domaine des ESSAIS DE CONSTANCE .24
4.6 Equipements d'essai y compris FANTÔMES et DISPOSITIFS D'ESSAI .24
4.7 Fréquence des ESSAIS DE CONSTANCE .26
5 Méthodes d'essai des TOMODENSITOMÈTRES .26
5.1 Positionnement du SUPPORT DU PATIENT .26
5.2 Précision du positionnement du PATIENT .28
5.3 EPAISSEUR DE COUPE TOMOGRAPHIQUE.32
5.4 Dose .36
5.5 BRUIT, UNIFORMITÉ ET NOMBRES CT MOYENS.38
5.6 RÉSOLUTION SPATIALE.42

Annexe A (informative) Guide des actions à engager .46
Annexe B (informative) Méthodes d'essai alternatives relatives à la RÉSOLUTION
SPATIALE.50
Annexe C (informative) EPAISSEUR DE COUPE TOMOGRAPHIQUE pour un balayage
hélicoïdal .56
Annexe D (informative) Méthode visuelle concernant la RÉSOLUTION À BAS CONTRASTE .58

Bibliographie.60

Index des termes .64

Figure 1 – Système de coordonnées utilisé pour les tomodensitomètres .20

Tableau B.1 − Comparaison des procédures d'essai de la RÉSOLUTION SPATIALE .50

61223-2-6 © IEC:2006 – 3 –
CONTENTS
FOREWORD.5

INTRODUCTION.9

1 Scope and object.11

2 Normative references .11

3 Terms and definitions .13

4 General aspects of CONSTANCY TESTS .21
4.1 Preconditions .21
4.2 General conditions affecting test procedures .21
4.3 Establishment of BASELINE VALUES .23
4.4 Identification of equipment, instrumentation and test conditions .23
4.5 Scope of the CONSTANCY TESTS.25
4.6 Test equipment, including PHANTOMS and TEST DEVICES .25
4.7 Frequency of CONSTANCY TESTS .27
5 Test methods for CT SCANNERS.27
5.1 Positioning of the PATIENT SUPPORT.27
5.2 PATIENT positioning accuracy .29
5.3 TOMOGRAPHIC SECTION THICKNESS .33
5.4 Dose .37
5.5 NOISE, UNIFORMITY AND MEAN CT NUMBERS .39
5.6 SPATIAL RESOLUTION.43

Annex A (informative) Guidance on action to be taken .47
Annex B (informative) Alternate test methods for SPATIAL RESOLUTION .51
Annex C (informative) TOMOGRAPHIC SECTION THICKNESS for helical scanning .57
Annex D (informative) Visual method for LOW CONTRAST RESOLUTION .59

Bibliography.61

Index of terms.65

Figure 1 – Coordinate system used for CT SCANNERS.21

Table B.1 − Comparison of SPATIAL RESOLUTION test procedures .51

– 4– 61223-2-6 © CEI:2006
COMMISSION ÉLECTROTECHNIQUE INTERNATIONALE

___________
ESSAIS D'ÉVALUATION ET DE ROUTINE

DANS LES SERVICES D'IMAGERIE MÉDICALE –

Partie 2-6: Essais de constance – Performance d'imagerie des

équipements de tomodensitométrie à rayonnement X

AVANT-PROPOS
1) La Commission Électrotechnique Internationale (CEI) est une organisation mondiale de normalisation
composée de l'ensemble des comités électrotechniques nationaux (Comités nationaux de la CEI). La CEI a
pour objet de favoriser la coopération internationale pour toutes les questions de normalisation dans les
domaines de l'électricité et de l'électronique. A cet effet, la CEI – entre autres activités – publie des Normes
internationales, des Spécifications techniques, des Rapports techniques, des Spécifications accessibles au
public (PAS) et des Guides (ci-après dénommés "Publication(s) de la CEI"). Leur élaboration est confiée à des
comités d'études, aux travaux desquels tout Comité national intéressé par le sujet traité peut participer. Les
organisations internationales, gouvernementales et non gouvernementales, en liaison avec la CEI, participent
également aux travaux. La CEI collabore étroitement avec l'Organisation Internationale de Normalisation (ISO),
selon des conditions fixées par accord entre les deux organisations.
2) Les décisions ou accords officiels de la CEI concernant les questions techniques représentent, dans la mesure
du possible, un accord international sur les sujets étudiés, étant donné que les Comités nationaux de la CEI
intéressés sont représentés dans chaque comité d’études.
3) Les Publications de la CEI se présentent sous la forme de recommandations internationales et sont agréées
comme telles par les Comités nationaux de la CEI. Tous les efforts ra
...

Questions, Comments and Discussion

Ask us and Technical Secretary will try to provide an answer. You can facilitate discussion about the standard in here.