WG 30 - TC 62/SC 62B/WG 30
TC 62/SC 62B/WG 30
General Information
IEC 61223-3-5:2019 applies to CT SCANNERS that conform to IEC 60601-2-44:2009, IEC 60601-2-44:2009/AMD1:2012 and IEC 60601-2-44:2009/AMD1:2012AMD2:2016.
IEC 60601-2-44 and this document
• defines the essential parameters which describe the performance of CT SCANNERS with regard to image quality, RADIATION OUTPUT and PATIENT positioning; the list of parameters to be tested can be found in section 4.3,
• defines the methods of testing the essential parameters, and
• evaluates compliance with the tolerances of the parameters SPECIFIED by the ACCOMPANYING DOCUMENTS.
The methods defined in IEC 60601-2-44 and this document rely on non-invasive measurements, using appropriate test equipment, performed during or after installation. Signed statements covering steps in the installation procedure can be used as part of the ACCEPTANCE TEST report.
This document applies to ACCEPTANCE TESTS and CONSTANCY TESTS on a CT SCANNER. The aim of the ACCEPTANCE TESTS is to verify compliance of the installation or MAJOR SERVICE ACTION with specifications affecting the image quality, RADIATION OUTPUT and PATIENT positioning. The CONSTANCY TESTS are performed to ensure that the functional performance of EQUIPMENT meets ESTABLISHED CRITERIA and to enable the early recognition of changes in the properties of components of the EQUIPMENT, and to verify compliance with specifications affecting the image quality, RADIATION OUTPUT and PATIENT positioning.
This document also contains requirements associated with ACCEPTANCE and CONSTANCY TESTING for the ACCOMPANYING DOCUMENTS of the CT SCANNER.
This document does not apply to
• aspects of mechanical and electrical safety, and
• aspects of mechanical, electrical and software performance, unless they are essential for performing the ACCEPTANCE TESTS and CONSTANCY TESTS, and are directly affecting image quality, RADIATION OUTPUT and PATIENT positioning.
IEC 61223-3-5:2019 cancels and replaces the first edition published in 2004, and the second edition of IEC 61223-2-6 published in 2006. This edition constitutes a technical revision.
This edition includes the following significant technical change with respect to the previous edition and to IEC 61223-2-6:
1) modification of the RADIATION protection and control;
2) modification of the acceptance testing;
3) introduction of constancy testing.
The contents of the corrigendum of June 2022 have been included in this copy.
- Standard126 pagesEnglish and French languagesale 15% off
Details the particular specification for light polyvinyl chloride sheathed cables of rated voltage of 300/500 V.[
]This consolidated version consists of the second edition (1992) and its amendment 1 (1997). Therefore, no need to order amendment in addition to this publication.
- Standard8 pagesEnglish and French languagesale 15% off
IEC 60601-2-44:2009 applies to the basic safety and essential performance of CT scanners. The scope of IEC 60601-2-44:2009 is limited to CT scanners intended to be used for both head and body characterised by an enclosure of the X-ray source(s) and imaging detector(s) in a common protective cover in the shape of a toroid. It includes safety requirements for the X-ray generators used in CT scanners, including those where high-voltage generators are integrated with an X-ray tube assembly. This third edition cancels and replaces the second edition published in 2001 and its Amendment 1 (2002). This edition constitutes a technical revision primarily related to radiation protection and control. The contents of the corrigendum of May 2010 have been included in this copy.
- Standard90 pagesEnglish and French languagesale 15% off
- Standard56 pagesEnglish and French languagesale 15% off