ISO 1042:1998
(Main)Laboratory glassware — One-mark volumetric flasks
Laboratory glassware — One-mark volumetric flasks
Verrerie de laboratoire — Fioles jaugées à un trait
La présente Norme internationale spécifie les exigences d'une série de fioles jaugées à un trait, acceptable sur le plan international et satisfaisant aux besoins généraux des laboratoires.Les spécifications de la présente Norme internationale sont conformes à l'ISO 384 ainsi qu'à la Recommandation internationale N° 4 de l'OIML.
Laboratorijska steklovina - Volumetrijske steklenice z eno oznako
General Information
Relations
Standards Content (Sample)
SLOVENSKI STANDARD
01-oktober-1998
Laboratorijska steklovina - Volumetrijske steklenice z eno oznako
Laboratory glassware -- One-mark volumetric flasks
Verrerie de laboratoire -- Fioles jaugées à un trait
Ta slovenski standard je istoveten z: ISO 1042:1998
ICS:
17.060 Merjenje prostornine, mase, Measurement of volume,
gostote, viskoznosti mass, density, viscosity
71.040.20 Laboratorijska posoda in Laboratory ware and related
aparati apparatus
2003-01.Slovenski inštitut za standardizacijo. Razmnoževanje celote ali delov tega standarda ni dovoljeno.
INTERNATIONAL ISO
STANDARD 1042
Fourth edition
1998-07-01
Laboratory glassware — One-mark
volumetric flasks
Verrerie de laboratoire — Fioles jaugées à un trait
A
Reference number
Foreword
ISO (the International Organization for Standardization) is a worldwide
federation of national standards bodies (ISO member bodies). The work of
preparing International Standards is normally carried out through ISO
technical committees. Each member body interested in a subject for which
a technical committee has been established has the right to be represented
on that committee. International organizations, governmental and non-
governmental, in liaison with ISO, also take part in the work. ISO
collaborates closely with the International Electrotechnical Commission
(IEC) on all matters of electrotechnical standardization.
Draft International Standards adopted by the technical committees are
circulated to the member bodies for voting. Publication as an International
Standard requires approval by at least 75 % of the member bodies casting
a vote.
ISO 1042 was developed by Technical Committee ISO/TC 48, Laboratory
glassware and related apparatus, Subcommittee SC 1, Volumetric
instruments.
This fourth edition cancels and replaces the third edition (ISO 1042:1983)
by incorporating the following changes:
a) flasks with capacities of 1, 2, 20 and 5 000 ml have been added;
b) flasks with conical body shape have been added;
c) flasks with wider neck have been added;
d) material has been more precisely defined and a test method for
capacity introduced;
e) sizes of ground joints have been added in tables 1 and 2.
© ISO 1998
All rights reserved. Unless otherwise specified, no part of this publication may be reproduced
or utilized in any form or by any means, electronic or mechanical, including photocopying and
microfilm, without permission in writing from the publisher.
International Organization for Standardization
Case postale 56 • CH-1211 Genève 20 • Switzerland
Internet central@iso.ch
X.400 c=ch; a=400net; p=iso; o=isocs; s=central
Printed in Switzerland
ii
©
ISO ISO 1042:1998(E)
Introduction
Volumetric flasks together with analytical balances are the fundamental
tools for the preparation of volumetric standard solutions – the basis of
chemical analysis. The design of narrow-necked class A volumetric flasks
has been optimized to achieve the fewest possible acceptable errors.
With the increasing popularity of piston-operated pipettors, there is market
pressure for the manufacture of volumetric flasks with wider necks so that
pipettor tips may be inserted to remove solution directly. Wide-necked
flasks will of necessity be of lower accuracy than the corresponding
capacities of narrow-necked flasks and the insertion of any extraneous
device may introduce other errors.
It is therefore recommended that narrow-necked class A volumetric flasks
are used for the production of standard solutions and where necessary, a
suitable quantity should be poured into an intermediate vessel into which
the pipettor tip may be introduced.
In accordance with good laboratory practice, only narrow-necked class A
volumetric flasks conforming to this International Standard should be used
for precise analytical purposes.
iii
©
INTERNATIONAL STANDARD ISO ISO 1042:1998(E)
Laboratory glassware — One-mark volumetric flasks
1 Scope
This International Standard specifies requirements for an internationally acceptable series of one-mark volumetric
flasks, suitable for general laboratory purposes.
The specifications in this International Standard are in conformity with ISO 384 and with OIML Recommendation
No. 4.
2 Normative references
The following standards contain provisions which, through references in this text, constitute provisions of this
International Standard. At the time of publication, the editions indicated were valid. All standards are subject to
revision, and parties to agreements based on this International Standard are encouraged to investigate the
possibility of applying the most recent editions of the standards indicated below. Members of IEC and ISO maintain
registers of currently valid International Standards.
ISO 383:1976, Laboratory glassware — Interchangeable conical ground joints.
ISO 719:1985, Glass — Hydrolytic resistance of glass grains at 98 °C — Method of test and classification.
ISO 4787:1984, Laboratory glassware — Volumetric glassware — Methods for use and testing of capacity.
3 Basis of adjustment
3.1 Unit of volume
The unit of volume shall be the millilitre (ml) which is equivalent to the cubic centimetre (cm ).
NOTE — The term millilitre (ml) is commonly used as a special name for the cubic centimetre (cm ), in accordance with a
decision of the twelfth Conférence Générale des Poids et Mesures. The term millilitre is acceptable, in general, for references
in International Standards to capacities of volumetric glassware and it is used, in particular, in the present text.
3.2 Reference temperature
The standard reference temperature, i.e. the temperature at which the volumetric flask is intended to contain its
nominal volume (nominal capacity), shall be 20 °C.
When the flask is required for use in a country which has adopted a standard reference temperature of 27 °C,
however, this value shall be substituted for 20 °C.
©
ISO
4 Classes of accuracy
Two classes of accuracy are specified:
Class A for the higher grade,
Class B for the lower grade.
5 Series of capacities
The series of nominal capacities of one-mark volumetric flasks is as follows (in millilitres):
1 - 2 - 5 - 10 - 20 - 25 - 50 - 100 - 200 - 250 - 500 - 1 000 - 2 000 - 5 000
All these flasks may be finished with a plain neck or may include a stopper.
NOTE — If volumetric flasks of capacities other than those listed above are required, it is recommended that they conform, as
far as possible, to the essential requirements of this International Standard.
6 Definition of capacity
The capacity of a volumetric flask is defined as the volume of water at 20 °C, expressed in millilitres, contained by
the flask at 20 °C, when filled to the graduation line.
Where, exceptionally, the reference temperature is 27 °C, this value shall be substituted for 20 °C.
Setting the meniscus shall be performed according to ISO 4787:
The meniscus is set so that the plane of the top edge of the graduation line is horizontally tangential to the lowest
point of the meniscus, the line of sight being in the same plane.
7 Accuracy
The capacity of the flasks shall not differ from the nominal capacity by more than the maximum permitted errors
shown in tables 1 and 2.
8 Construction
8.1 Material
Volumetric flasks shall be made from glass of hydrolytic class not lower than HGB3 in accordance with ISO 719 with
-6 -1
a coefficient of thermal expansion not exceeding 3,3 x 10 °C .
NOTE — Borosilicate glass 3.3 in accordance with ISO 3585 is included in this requirement.
The glass shall be as free as possible from visible defects and reasonably free from internal stress, which would
impair the performance of the flasks.
8.2 Wall thickness
Volumetric flasks shall be sufficiently robust in construction to withstand normal usage and the wall thickness shall
show no gross departures form uniformity.
SIST IS
...
INTERNATIONAL ISO
STANDARD 1042
Fourth edition
1998-07-01
Laboratory glassware — One-mark
volumetric flasks
Verrerie de laboratoire — Fioles jaugées à un trait
A
Reference number
Foreword
ISO (the International Organization for Standardization) is a worldwide
federation of national standards bodies (ISO member bodies). The work of
preparing International Standards is normally carried out through ISO
technical committees. Each member body interested in a subject for which
a technical committee has been established has the right to be represented
on that committee. International organizations, governmental and non-
governmental, in liaison with ISO, also take part in the work. ISO
collaborates closely with the International Electrotechnical Commission
(IEC) on all matters of electrotechnical standardization.
Draft International Standards adopted by the technical committees are
circulated to the member bodies for voting. Publication as an International
Standard requires approval by at least 75 % of the member bodies casting
a vote.
ISO 1042 was developed by Technical Committee ISO/TC 48, Laboratory
glassware and related apparatus, Subcommittee SC 1, Volumetric
instruments.
This fourth edition cancels and replaces the third edition (ISO 1042:1983)
by incorporating the following changes:
a) flasks with capacities of 1, 2, 20 and 5 000 ml have been added;
b) flasks with conical body shape have been added;
c) flasks with wider neck have been added;
d) material has been more precisely defined and a test method for
capacity introduced;
e) sizes of ground joints have been added in tables 1 and 2.
© ISO 1998
All rights reserved. Unless otherwise specified, no part of this publication may be reproduced
or utilized in any form or by any means, electronic or mechanical, including photocopying and
microfilm, without permission in writing from the publisher.
International Organization for Standardization
Case postale 56 • CH-1211 Genève 20 • Switzerland
Internet central@iso.ch
X.400 c=ch; a=400net; p=iso; o=isocs; s=central
Printed in Switzerland
ii
©
ISO ISO 1042:1998(E)
Introduction
Volumetric flasks together with analytical balances are the fundamental
tools for the preparation of volumetric standard solutions – the basis of
chemical analysis. The design of narrow-necked class A volumetric flasks
has been optimized to achieve the fewest possible acceptable errors.
With the increasing popularity of piston-operated pipettors, there is market
pressure for the manufacture of volumetric flasks with wider necks so that
pipettor tips may be inserted to remove solution directly. Wide-necked
flasks will of necessity be of lower accuracy than the corresponding
capacities of narrow-necked flasks and the insertion of any extraneous
device may introduce other errors.
It is therefore recommended that narrow-necked class A volumetric flasks
are used for the production of standard solutions and where necessary, a
suitable quantity should be poured into an intermediate vessel into which
the pipettor tip may be introduced.
In accordance with good laboratory practice, only narrow-necked class A
volumetric flasks conforming to this International Standard should be used
for precise analytical purposes.
iii
©
INTERNATIONAL STANDARD ISO ISO 1042:1998(E)
Laboratory glassware — One-mark volumetric flasks
1 Scope
This International Standard specifies requirements for an internationally acceptable series of one-mark volumetric
flasks, suitable for general laboratory purposes.
The specifications in this International Standard are in conformity with ISO 384 and with OIML Recommendation
No. 4.
2 Normative references
The following standards contain provisions which, through references in this text, constitute provisions of this
International Standard. At the time of publication, the editions indicated were valid. All standards are subject to
revision, and parties to agreements based on this International Standard are encouraged to investigate the
possibility of applying the most recent editions of the standards indicated below. Members of IEC and ISO maintain
registers of currently valid International Standards.
ISO 383:1976, Laboratory glassware — Interchangeable conical ground joints.
ISO 719:1985, Glass — Hydrolytic resistance of glass grains at 98 °C — Method of test and classification.
ISO 4787:1984, Laboratory glassware — Volumetric glassware — Methods for use and testing of capacity.
3 Basis of adjustment
3.1 Unit of volume
The unit of volume shall be the millilitre (ml) which is equivalent to the cubic centimetre (cm ).
NOTE — The term millilitre (ml) is commonly used as a special name for the cubic centimetre (cm ), in accordance with a
decision of the twelfth Conférence Générale des Poids et Mesures. The term millilitre is acceptable, in general, for references
in International Standards to capacities of volumetric glassware and it is used, in particular, in the present text.
3.2 Reference temperature
The standard reference temperature, i.e. the temperature at which the volumetric flask is intended to contain its
nominal volume (nominal capacity), shall be 20 °C.
When the flask is required for use in a country which has adopted a standard reference temperature of 27 °C,
however, this value shall be substituted for 20 °C.
©
ISO
4 Classes of accuracy
Two classes of accuracy are specified:
Class A for the higher grade,
Class B for the lower grade.
5 Series of capacities
The series of nominal capacities of one-mark volumetric flasks is as follows (in millilitres):
1 - 2 - 5 - 10 - 20 - 25 - 50 - 100 - 200 - 250 - 500 - 1 000 - 2 000 - 5 000
All these flasks may be finished with a plain neck or may include a stopper.
NOTE — If volumetric flasks of capacities other than those listed above are required, it is recommended that they conform, as
far as possible, to the essential requirements of this International Standard.
6 Definition of capacity
The capacity of a volumetric flask is defined as the volume of water at 20 °C, expressed in millilitres, contained by
the flask at 20 °C, when filled to the graduation line.
Where, exceptionally, the reference temperature is 27 °C, this value shall be substituted for 20 °C.
Setting the meniscus shall be performed according to ISO 4787:
The meniscus is set so that the plane of the top edge of the graduation line is horizontally tangential to the lowest
point of the meniscus, the line of sight being in the same plane.
7 Accuracy
The capacity of the flasks shall not differ from the nominal capacity by more than the maximum permitted errors
shown in tables 1 and 2.
8 Construction
8.1 Material
Volumetric flasks shall be made from glass of hydrolytic class not lower than HGB3 in accordance with ISO 719 with
-6 -1
a coefficient of thermal expansion not exceeding 3,3 x 10 °C .
NOTE — Borosilicate glass 3.3 in accordance with ISO 3585 is included in this requirement.
The glass shall be as free as possible from visible defects and reasonably free from internal stress, which would
impair the performance of the flasks.
8.2 Wall thickness
Volumetric flasks shall be sufficiently robust in construction to withstand normal usage and the wall thickness shall
show no gross departures form uniformity.
©
ISO
Figure 1 — One-mark volumetric flasks showing alternative shapes
©
ISO
Table 1 — Dimensions and maximum permitted errors for narrow-necked flasks
Essential dimensions Tolerances Recommended dimensions
3)
Nominal Internal Distance of Maximum Overall Bulb Base Wall Ground joint
2)
capacity neck graduation permitted height diameter diameter thickness
1)
diameter line errors
Class A Class B k4 k6
d h h d d s
1 2 1 2 3
ml mm mm ml ml + 5 mm mm mm
...
NORME ISO
INTERNATIONALE 1042
Quatrième édition
1998-07-01
Verrerie de laboratoire — Fioles jaugées à
un trait
Laboratory glassware — One-mark volumetric flasks
Numéro de référence
©
ISO 1998
PDF – Exonération de responsabilité
Le présent fichier PDF peut contenir des polices de caractères intégrées. Conformément aux conditions de licence d'Adobe, ce fichier peut
être imprimé ou visualisé, mais ne doit pas être modifié à moins que l'ordinateur employé à cet effet ne bénéficie d'une licence autorisant
l'utilisation de ces polices et que celles-ci y soient installées. Lors du téléchargement de ce fichier, les parties concernées acceptent de fait la
responsabilité de ne pas enfreindre les conditions de licence d'Adobe. Le Secrétariat central de l'ISO décline toute responsabilité en la
matière.
Adobe est une marque déposée d'Adobe Systems Incorporated.
Les détails relatifs aux produits logiciels utilisés pour la création du présent fichier PDF sont disponibles dans la rubrique General Info du
fichier; les paramètres de création PDF ont été optimisés pour l'impression. Toutes les mesures ont été prises pour garantir l'exploitation de
ce fichier par les comités membres de l'ISO. Dans le cas peu probable où surviendrait un problème d'utilisation, veuillez en informer le
Secrétariat central à l'adresse donnée ci-dessous.
© ISO 1998
Droits de reproduction réservés. Sauf prescription différente, aucune partie de cette publication ne peut être reproduite ni utilisée sous quelque
forme que ce soit et par aucun procédé, électronique ou mécanique, y compris la photocopie et les microfilms, sans l'accord écrit de l’ISO à
l’adresse ci-après ou du comité membre de l’ISO dans le pays du demandeur.
ISO copyright office
Case postale 56 � CH-1211 Geneva 20
Tel. + 41 22 749 01 11
Fax. + 41 22 734 10 79
E-mail copyright@iso.ch
Web www.iso.ch
Version française parue en 2000
ImpriméenSuisse
ii © ISO 1998 – Tous droits réservés
Avant-propos
L'ISO (Organisation internationale de normalisation) est une fédération mondiale d'organismes nationaux de
normalisation (comités membres de l'ISO). L'élaboration des Normes internationales est en général confiée aux
comités techniques de l'ISO. Chaque comité membre intéressé par une étude a le droit de faire partie du comité
technique créé à cet effet. Les organisations internationales, gouvernementales et non gouvernementales, en
liaison avec l'ISO participent également aux travaux. L'ISO collabore étroitement avec la Commission
électrotechnique internationale (CEI) en ce qui concerne la normalisation électrotechnique.
Les Normes internationales sont rédigées conformément aux règles données dans les Directives ISO/CEI, Partie 3.
Les projets de Normes internationales adoptés par les comités techniques sont soumis aux comités membres pour
vote. Leur publication comme Normes internationales requiert l'approbation de 75 % au moins des comités
membres votants.
L’attention est appelée sur le fait que certains des éléments de la présente Norme internationale peuvent faire
l’objet de droits de propriété intellectuelle ou de droits analogues. L’ISO ne saurait être tenue pour responsable de
ne pas avoir identifié de tels droits de propriété et averti de leur existence.
La Norme internationale ISO 1042 a été élaborée par le comité technique ISO/TC 48, Verrerie de laboratoire et
appareils connexes, sous-comité SC 1, Instruments volumétriques.
Cette quatrième édition annule et remplace la troisième édition (ISO 1042:1983), par rapport à laquelle les
modifications suivantes ont été apportées:
a) des fioles jaugées de capacités 1 ml, 2 ml, 20 ml et 5 000 ml ont été ajoutées;
b) des fioles jaugées dont le corps est de forme conique ont été ajoutées;
c) des fioles jaugées à col large ont été ajoutées;
d) le matériau est défini de manière plus précise et une méthode d'essai pour la mesure de la capacité a été
introduite;
e) les dimensions des joints rodés ont été ajoutées dans les Tableaux 1 et 2.
© ISO 1998 – Tous droits réservés iii
Introduction
Les fioles jaugées constituent avec les balances analytiques les outils fondamentaux requis pour la préparation de
solutions volumétriques étalons, base de l'analyse chimique. La conception des fioles jaugées à col étroit de la
classe A a été optimisée afin d'obtenir un risque d'erreur acceptable minimal.
Devant le succès grandissant des pipettes à piston, le marché connaît une augmentation de la demande relative à
la fabrication de fioles jaugées à col large permettant d'insérer le bout des pipettes de sorte à aspirer directement
la solution. À capacité égale, les fioles jaugées à col large seront inévitablement moins précises que les fioles
jaugées à col étroit, et l'insertion de tout instrument quelconque peut entraîner de nouvelles erreurs.
Il est par conséquent recommandé d'utiliser les fioles jaugées à col étroit de la classe A pour la fabrication des
solutions étalons et, si nécessaire, d'en verser la quantité convenable dans un récipient intermédiaire dans lequel il
est possible d'insérer le bout de la pipette.
Conformément aux bonnes pratiques de laboratoire, il convient d'utiliser uniquement des fioles jaugées à col étroit
de la classe A conformes à la présente Norme internationale pour des applications analytiques de précision.
iv © ISO 1998 – Tous droits réservés
NORME INTERNATIONALE ISO 1042:1998(F)
Verrerie de laboratoire — Fioles jaugées à un trait
1 Domaine d'application
La présente Norme internationale spécifie les exigences d'une série de fioles jaugées à un trait, acceptable sur le
plan international et satisfaisant aux besoins généraux des laboratoires.
Les spécifications de la présente Norme internationale sont conformes à l'ISO 384 ainsi qu'à la Recommandation
o
internationale N 4del'OIML.
2 Références normatives
Les documents normatifs suivants contiennent des dispositions qui, par suite de la référence qui y est faite,
constituent des dispositions valables pour la présente Norme internationale. Pour les références datées, les
amendements ultérieurs ou les révisions de ces publications ne s’appliquent pas. Toutefois, les parties prenantes
aux accords fondés sur la présente Norme internationale sont invitées à rechercher la possibilité d'appliquer les
éditions les plus récentes des documents normatifs indiqués ci-après. Pour les références non datées, la dernière
édition du document normatif en référence s’applique. Les membres de l'ISO et de la CEI possèdent le registre des
Normes internationales en vigueur.
ISO 383:1976, Verrerie de laboratoire — Assemblages coniques rodés interchangeables.
ISO 719:1985, Verre — Résistance hydrolytique du verre en grains à 98 °C — Méthode d'essai et classification.
ISO 4787:1984, Verrerie de laboratoire — Verrerie volumétrique — Méthodes d'utilisation et de vérification de la
capacité.
3 Données fondamentales pour l'étalonnage
3.1 Unité de volume
L'unité de volume doit être le millilitre (ml), qui est équivalent à un centimètre cube (cm ).
NOTE Le terme millilitre (ml) est couramment utilisé comme nom spécial donné au centimètre cube (cm ), conformément à
la décision de la Douzième Conférence Générale des Poids et Mesures. Le terme millilitre est généralement admis pour
désigner les capacités de la verrerie volumétrique dans les Normes internationales et il est, en particulier, utilisé dans le présent
texte.
3.2 Température de référence
La température normale de référence, c'est-à-dire la température à laquelle la fiole jaugée est censée contenir son
volume nominal (capacité nominale), doit être de 20 °C.
Quand la fiole doit être utilisée dans un pays ayant adopté une température normale de référence de 27 °C, cette
valeur doit être substituée à 20 °C.
© ISO 1998 – Tous droits réservés 1
4 Classes de précision
Deux classes de précision sont spécifiées:
� classe A pour la catégorie supérieure,
� classe B pour la catégorie inférieure.
5 Série des capacités
La série des capacités nominales des fioles jaugées à un trait est la suivante (en millilitres):
1 - 2- 5- 10 - 20- 25 - 50- 100- 200 - 250 - 500 - 1000 - 2000 - 5000
Toutes ces fioles peuvent se terminer par un col ordinaire ou être pourvues d'un bouchon.
NOTE S'il est nécessaire de réaliser des fioles jaugées de capacités autres que celles indiquées ci-dessus, il est
recommandé de se conformer, dans la mesure du possible, aux exigences essentielles de la présente Norme internationale.
6 Définition de la capacité
La capacité d'une fiole jaugée est définie par le volume d'eau à 20 °C, exprimé en millilitres, contenu dans la fiole à
20 °C, quand elle est remplie jusqu'au trait repère.
Quand, exceptionnellement, la température de référence est de 27 °C, cette valeur doit être substituée à 20 °C.
La position du ménisque doit être déterminée conformément à l'ISO 4787:
Le ménisque est ajusté de telle façon que le plan horizontal passant par le bord supérieur du trait repère soit
tangent au ménisque en son point le plus bas, la ligne de visée étant dans le même plan.
7 Précision
La capacité de la fiole ne doit pas différer de la capacité nominale de plus des erreurs maximales tolérées,
indiquées dans les Tableaux 1 et 2.
8 Construction
8.1 Matériau
Les fioles jaugées doivent être faites en verre appartenant à une classe de résistance hydrolytique qui ne doit pas
être inférieure à la ca
...
Questions, Comments and Discussion