ISO 17191:2026
(Main)Urine-absorbing products for incontinence — Measurement of airborne respirable polyacrylate superabsorbent materials — Determination of dust in collection cassettes by sodium atomic absorption spectrometry
- Abstract
This document specifies a method for the determination of polyacrylate (PA) superabsorbent materials in airborne dust by measurement of sodium (Na) by atomic absorption spectrometry (AAS). This method is applicable to the determination of collected superabsorbent powder in the range between 0,2 µg and 60 µg (limit of detection near 0,2 µg) of superabsorbent powders. It is applicable to polyacrylate superabsorbent materials, which occur in hygiene products including urine-absorbing aids for incontinent persons. The test methods are addressed to the material exclusively. They are not intended to be used, and are not applicable for use, with finished manufactured urine-absorbing aids.
- Status
- Published
- Publication Date
- 23-Jul-2026
- Technical Committee
- ISO/TC 173/SC 3 - Products for ostomy and incontinence
- Drafting Committee
- ISO/TC 173/SC 3/WG 7 - Superabsorbent polymers
- Current Stage
- 6060 - International Standard published - 60 Publication
- Start Date
- 24-Jul-2026
- Due Date
- 03-Oct-2026
- Completion Date
- 03-Oct-2026
Buy Documents
ISO 17191:2026 - Urine-absorbing products for incontinence — Measurement of airborne respirable polyacrylate superabsorbent materials — Determination of dust in collection cassettes by sodium atomic absorption spectrometry
ISO 17191:2026 is an International Standard from ISO for measuring airborne respirable polyacrylate superabsorbent materials collected in dust cassettes. It specifies a sodium atomic absorption spectrometry (AAS) method for polyacrylate (PA) superabsorbent powders used in hygiene products, including urine-absorbing aids for incontinent persons. The method is aimed at laboratories and material testers who need to quantify very low levels of this powder in collected airborne dust.
What does ISO 17191:2026 specify?
ISO 17191:2026 specifies a test method for determining PA superabsorbent material in airborne dust by measuring sodium (Na) with AAS. The document applies to the material itself, not to finished manufactured urine-absorbing aids.
The method is intended for collected superabsorbent powder in the range from 0,2 µg to 60 µg, with a limit of detection near 0,2 µg. That range matters in practice because the procedure is built for low-level contamination and dust measurement, not bulk material testing.
The document is organized into clauses on scope, normative references, terms and definitions, principle, reagents, apparatus, procedure, calculation, precision, and test report. It also includes a bibliography, one precision table, an example calculation, and figures for the cassette, drying unit, and washing stand.
What are the key requirements of ISO 17191:2026?
ISO 17191:2026 is built around contamination control, sodium extraction, and AAS measurement of very small sample amounts. In practice, the user needs clean cassettes, low-sodium reagents, and careful calibration to get usable results.
Measurement principle and sample handling
Clause 4 uses AAS to measure sodium in PA superabsorbent powders collected in air-monitoring cassettes with polytetrafluoroethylene filters and porous plastic backing pads. Sodium is released from the powder with potassium chloride (KCl) solution, then measured, and the collected PA mass is calculated after subtracting blank-cassette background. This means the sodium result is used as a proxy for the PA powder mass.
Reagents and calibration solutions
Clause 5 requires analytical-grade reagents and very low-sodium water. The standards and zeroing solution are prepared in KCl with isopropanol (IPA), and the reagents are pre-analysed for sodium before use. In practice, this keeps the calibration curve and the samples on the same chemical basis and reduces false background readings.
Apparatus and handling equipment
Clause 6 specifies an atomic absorption spectrometer with an air-acetylene flame, a sodium hollow cathode lamp, dedicated trace-level spray chamber and burner, polypropylene volumetric flasks, non-powdered plastic gloves, air-monitoring cassettes, a flatbed shaker, and a drying air filter. These items matter because trace sodium work is easily affected by contamination from ordinary lab surfaces and materials.
Cleaning and release of cassettes
Clause 7.1.1 sets out cassette cleaning before air monitoring: wet the filter with IPA, wash with 0,2 % KCl, rinse with deionized water, and dry with filtered air. A batch release check is required on 10 % of the cassettes, and if any of those contain over 0,17 µg sodium, the batch must be cleaned again. This is the document’s main control for blank quality.
Sample analysis and instrument checks
Clause 7.1.2 and 7.2 require the cassette to be prepared with IPA and KCl, shaken for extraction, and then analysed by aspirating directly from the cassette. The calibration uses sodium standards and a zeroing solution, and the instrument is checked after every five samples; if a standard changes by more than 5 %, the tube is treated as plugged and the setup is cleaned and recalibrated if needed. This keeps the readings stable when particles obstruct the aspiration path.
Calculation and reporting
Clause 8 requires a second-order regression calibration, not a linear one, because the AAS response to sodium is non-linear. The measured sodium concentration is converted to apparent PA mass using a conversion factor that depends on the material’s degree of neutralization, then corrected by the mean blank and compared with the limit of detection. Results at or below the limit of detection are reported as ND.
What terms does ISO 17191:2026 define?
ISO 17191:2026 does not provide a formal term list, but it uses several key terms in the procedure.
- Polyacrylate (PA) superabsorbent materials - the absorbent polymer powders that the method measures.
- Atomic absorption spectrometry (AAS) - the analytical technique used to measure sodium in the extraction solution.
- Potassium chloride (KCl) - the salt solution used to release sodium from the collected powder.
- Isopropanol (IPA) - the solvent used to wet the filter and support cleaning and extraction.
- Air-monitoring cassettes - the cassettes that collect airborne dust for later analysis.
- Limit of detection (LOD) - the threshold below which the method reports none detected.
- Blank cassette - a cleaned cassette used to measure background sodium and correct sample results.
Who uses ISO 17191:2026?
ISO 17191:2026 is used by analytical laboratories, quality-control teams, and material suppliers working with PA superabsorbent powders. It is relevant to makers of hygiene products and urine-absorbing aids, as well as to labs that need to verify airborne dust from the material itself. It also supports product development and incoming inspection where very low sodium-based dust levels must be measured consistently.
What changed in ISO 17191:2026 from the previous edition?
ISO 17191:2026 is the second edition and it cancels and replaces ISO 17191:2004. The stated main change is clarification of the test procedures.
Which standards are used with ISO 17191:2026?
ISO 17191:2026 cites ISO 3696:1987 for water used in analytical laboratory work. That reference matters because the method requires low-sodium water of Grade 1 quality.
The introduction also places the document alongside the ISO 17190 series, which covers test methods for polymer-based absorbent materials. Those related standards help readers understand that ISO 17191 focuses on dust measurement for PA superabsorbent material, while the ISO 17190 series covers broader material characterization.
What does the ISO 17191:2026 document contain?
ISO 17191:2026 contains the principle of the method, detailed reagent and apparatus requirements, sample preparation steps, sodium extraction and AAS measurement instructions, and calculation formulas for apparent mass, blank correction, and detection limit. It also includes precision data from a recovery trial in Table 1 and an example showing how to turn sodium readings into reported PA mass or ND results.
The figures show the air-monitoring cassette, the cassette-drying unit, and the washing stand used in the procedure. The test report clause lists the information to record so the measurement can be traced and repeated.
Relations
- Effective Date
- 28-Oct-2023
Buy Documents
ISO 17191:2026 - Urine-absorbing products for incontinence — Measurement of airborne respirable polyacrylate superabsorbent materials — Determination of dust in collection cassettes by sodium atomic absorption spectrometry
Get Certified
Connect with accredited certification bodies for this standard

BSI Group
BSI (British Standards Institution) is the business standards company that helps organizations make excellence a habit.

TÜV Rheinland
TÜV Rheinland is a leading international provider of technical services.

TÜV SÜD
TÜV SÜD is a trusted partner of choice for safety, security and sustainability solutions.
Sponsored listings
Frequently Asked Questions
ISO 17191:2026 is a standard published by the International Organization for Standardization (ISO). Its full title is "Urine-absorbing products for incontinence — Measurement of airborne respirable polyacrylate superabsorbent materials — Determination of dust in collection cassettes by sodium atomic absorption spectrometry". This standard covers: This document specifies a method for the determination of polyacrylate (PA) superabsorbent materials in airborne dust by measurement of sodium (Na) by atomic absorption spectrometry (AAS). This method is applicable to the determination of collected superabsorbent powder in the range between 0,2 µg and 60 µg (limit of detection near 0,2 µg) of superabsorbent powders. It is applicable to polyacrylate superabsorbent materials, which occur in hygiene products including urine-absorbing aids for incontinent persons. The test methods are addressed to the material exclusively. They are not intended to be used, and are not applicable for use, with finished manufactured urine-absorbing aids.
This document specifies a method for the determination of polyacrylate (PA) superabsorbent materials in airborne dust by measurement of sodium (Na) by atomic absorption spectrometry (AAS). This method is applicable to the determination of collected superabsorbent powder in the range between 0,2 µg and 60 µg (limit of detection near 0,2 µg) of superabsorbent powders. It is applicable to polyacrylate superabsorbent materials, which occur in hygiene products including urine-absorbing aids for incontinent persons. The test methods are addressed to the material exclusively. They are not intended to be used, and are not applicable for use, with finished manufactured urine-absorbing aids.
ISO 17191:2026 is classified under the following ICS (International Classification for Standards) categories: 11.180.20 - Aids for incontinence and ostomy. The ICS classification helps identify the subject area and facilitates finding related standards.
ISO 17191:2026 has the following relationships with other standards: It is inter standard links to ISO 17191:2004. Understanding these relationships helps ensure you are using the most current and applicable version of the standard.
ISO 17191:2026 is available in PDF format for immediate download after purchase. The document can be added to your cart and obtained through the secure checkout process. Digital delivery ensures instant access to the complete standard document.
Standards Content (Sample)
Contents
- Introduction
- 1 Scope
- 2 Normative references
- 3 Terms and definitions
- 4 Principle
- 5 Reagents
- 6 Apparatus
- 1 HEPA filter
- 3 Tygon tubing
- 7 Procedure
- 7.1 Sample preparation and precautionary measures
- 1 Tygon tubing
- 7.2 Sodium determination
- 8 Calculation
- 8.4 Subtract the mean blank value m from each sample result to obtain the corrected level of
- 9 Precision
- 10 Test report
- Bibliography
Introduction
This document was originally developed by the European Nonwovens Association (EDANA) as one of an eleven-part series of test methods. It specifies a test method for determining the amount of airborne dust emanating from polyacrylate superabsorbent material. The other test methods in the original EDANA [1] series have been transformed into International Standards as the ISO 17190 series . The documents in the [1] ISO 17190 series series differ from this document in that they cover test methods for the characterization of polymer-based absorbent materials.
This test method has been in practical use for several years, and has proven to be reliable with respect to common criteria of quality of test methods (validity, repeatability, etc.).
International Standard ISO 17191:2026(en) Urine-absorbing products for incontinence — Measurement of airborne respirable polyacrylate superabsorbent materials - Determination of dust in collection cassettes by sodium atomic absorption spectrometry
1 Scope
This document specifies a method for the determination of polyacrylate (PA) superabsorbent materials in airborne dust by measurement of sodium (Na) by atomic absorption spectrometry (AAS). This method is applicable to the determination of collected superabsorbent powder in the range between 0,2 µg and 60 µg (limit of detection near 0,2 µg) of superabsorbent powders. It is applicable to polyacrylate superabsorbent materials, which occur in hygiene products including urineabsorbing aids for incontinent persons. The test methods are addressed to the material exclusively. They are not intended to be used, and are not applicable for use, with finished manufactured urine-absorbing aids.
2 Normative references
The following documents are referred to in the text in such a way that some or all of their content constitutes requirements of this document. For dated references, only the edition cited applies. For undated references, the latest edition of the referenced document (including any amendments) applies. ISO 3696:1987, Water for analytical laboratory use — Specification and test methods
3 Terms and definitions
No terms and definitions are listed in this document.
ISO and IEC maintain terminology databases for use in standardization at the following addresses:
- IEC Electropedia: available at http:// www .electropedia .org/
- ISO Online browsing platform: available at http:// www .iso .org/ obp
4 Principle
Atomic absorption analysis (AAS) is used to determine the sodium content in polyacrylate (PA) superabsorbent powders collected in air-monitoring cassettes.
PA dust samples are collected in polystyrene acrylonitrile air-monitoring cassettes with polytetrafluoroethylene filters and porous plastic backing pads. The sodium is determined after being released from the PA superabsorbent powder by exchange with potassium in a potassium chloride solution. The amount of dissolved sodium is then determined using AAS, and the mass of the PA superabsorbent powder collected is calculated after subtracting the background sodium levels present in blank cassettes.
To minimize sodium contamination, cassettes are prewashed before collecting the airborne superabsorbent powder. For the analysis, an aliquot of the potassium chloride solution is drawn from the cassette directly into the atomic absorption spray chamber.
ISO 17191:2026(en) This test is designed to determine low levels of sodium and requires that very clean handling conditions be observed. The use of deionized/distilled water containing very low sodium content is critical for successful analysis.
5 Reagents
Use only reagents of recognized analytical grade, unless otherwise specified. Store all reagents and solutions in plastic containers with the exception of KCl (5.1) and IPA (5.3). Store these in the glass containers in which they are shipped.
Pre-analyse reagents and solutions specified in 5.1, 5.2 and 5.3 for sodium (Na) using AAS, before using to clean cassettes or for Na analysis.
5.1 Potassium chloride (KCl), KCl content 99,999 % minimum; Na concentration less than 10 μg/g. 5.2 Water, deionized or distilled, in accordance with ISO 3696:1987, Grade 1, containing < 0,1 mg/l of Na. 5.3 Isopropanol (IPA), analytical reagent grade.
1) 5.4 Sodium standard reference solution, r(Na) = 1 000 mg/l [1 000 ppm ]. 5.5 Potassium chloride solution, r(KCl) = 2 000 mg/l.
Dissolve 2,0 g of potassium chloride (5.1) in 1 l of deionized water (5.2) in a polypropylene flask. The solution shall contain < 0,1 mg/l of Na.
5.6 Sodium standard solutions, for AAS as follows: r(Na) = 0,1 mg/l, 0,5 mg/l, 1,0 mg/l, 2,0 mg/l. Using polypropylene volumetric flasks, dilute the 1 000 mg/l sodium standard reference solution (5.4) with the 2 000 mg/l potassium chloride solution (5.5) as required, and add isopropanol (5.3) to obtain a solutio of 10 % by volume.
5.7 Atomic absorption zeroing solution.
Add isopropanol (5.3) to 2 000 mg/l potassium chloride solution (5.5) to obtain a solution of exactly 10,0 % isopropanol by volume.
6 Apparatus
6.1 Atomic absorption spectrometer (AAS).
Use a single-beam spectrometer with air-acetylene flame, or equivalent, capable of photometric accuracy of ± 0,000 5 AU.
6.2 Spraychamber and burner.
Use a unit that is dedicated for trace-level sodium determination. 6.3 Hollow cathode lamp for sodium.
6.4 Pipette, capable of measuring 0,5 ml.
6.5 Pipette, capable of measuring 4,5 ml.
1) 1 mg/l = 1 ppm. The use of ppm is deprecated.
ISO 17191:2026(en) 6.6 Plastic gloves, non-powdered.
6.7 Volumetric flasks, polypropylene, 1 000 ml and 100 ml capacity. 2) 6.8 Air-monitoring cassettes , see Figure 1.
Key 1 cassette outlet 2 support pad 3 filter 4 cassette ring (middle) 5 cassette ring (bottom) a Air flow.
Figure 1 — Air-monitoring cassette 6.9 Flatbed laboratory shaker, capable of holding and shaking air-monitoring cassettes (6.8). 2) Commercially available air-monitoring cassettes with the following specifications have been shown to be suitable:
three-piece clear polystyrene cassette for respirable dust cyclone sampling, as specified in NIOSH methods 7500 & 0600 http:// www .cdc .gov/ niosh/ docs/ 2003 -154/ pdfs/ 7500 .pdf, 37 mm diameter, with porous high density polyethylene or PTFE support pad [e.g. cassette SKC catalogue part No. 225-3; support pad = SKC catalogue part No. 225-2902; filter
...