Human response to vibration — Measuring instrumentation

ISO 8041:2005 specifies the performance specifications and tolerance limits for instruments designed to measure vibration values, for the purpose of assessing human response to vibration. It includes requirements for pattern evaluation, periodic verification and in-situ checks, and the specification of vibration calibrators for in-situ checks. Vibration instruments specified in ISO 8041:2005 can be single instruments, combinations of instrumentation or computer-based acquisition and analysis systems. Three levels of performance testing are defined in ISO 8041:2005: pattern evaluation, i.e. a full test of the instrument against the specifications defined in this International Standard; periodic verification, i.e. an intermediate set of tests designed to ensure that an instrument remains within the required performance specification, and in-situ checks, i.e. a minimum level of testing required to indicate that an instrument is likely to be functioning within the required performance specification.

Réponse des individus aux vibrations — Appareillage de mesure

L'ISO 8041:2005 donne les spécifications de performances et les limites de tolérance pour les instruments destinés à mesurer les valeurs de vibration afin d'évaluer la réponse des individus aux vibrations. Elle inclut les exigences relatives aux essais de conformité, à la vérification périodique et aux contrôles in-situ, ainsi que la spécification concernant les calibrateurs de vibration destinés aux contrôles in-situ. Les instruments de mesure des vibrations spécifiés dans l'ISO 8041:2005 peuvent être des instruments simples, des combinaisons d'instruments ou des systèmes d'acquisition et d'analyse informatisés. Trois niveaux de vérification des performances sont définis dans l'ISO 8041:2005: les essais de conformité, c'est-à-dire un essai complet de l'instrument par rapport aux spécifications définies dans la présente Norme internationale; la vérification périodique, c'est-à-dire un ensemble intermédiaire d'essais destiné à s'assurer qu'un instrument donné relève effectivement de la spécification de performances requise, et contrôles in-situ, c'est-à-dire un niveau minimal de vérification à appliquer pour indiquer que le fonctionnement d'un instrument donné relève vraisemblablement de la spécification de performances requise.

General Information

Status
Withdrawn
Publication Date
05-Apr-2005
Withdrawal Date
05-Apr-2005
Current Stage
9599 - Withdrawal of International Standard
Completion Date
03-May-2017
Ref Project

Relations

Buy Standard

Standard
ISO 8041:2005 - Human response to vibration -- Measuring instrumentation
English language
91 pages
sale 15% off
Preview
sale 15% off
Preview
Standard
ISO 8041:2005 - Réponse des individus aux vibrations -- Appareillage de mesure
French language
99 pages
sale 15% off
Preview
sale 15% off
Preview

Standards Content (Sample)

INTERNATIONAL ISO
STANDARD 8041
Second edition
2005-04-01


Human response to vibration —
Measuring instrumentation
Réponse des individus aux vibrations — Appareillage de mesure





Reference number
ISO 8041:2005(E)
©
ISO 2005

---------------------- Page: 1 ----------------------
ISO 8041:2005(E)
PDF disclaimer
This PDF file may contain embedded typefaces. In accordance with Adobe's licensing policy, this file may be printed or viewed but
shall not be edited unless the typefaces which are embedded are licensed to and installed on the computer performing the editing. In
downloading this file, parties accept therein the responsibility of not infringing Adobe's licensing policy. The ISO Central Secretariat
accepts no liability in this area.
Adobe is a trademark of Adobe Systems Incorporated.
Details of the software products used to create this PDF file can be found in the General Info relative to the file; the PDF-creation
parameters were optimized for printing. Every care has been taken to ensure that the file is suitable for use by ISO member bodies. In
the unlikely event that a problem relating to it is found, please inform the Central Secretariat at the address given below.


©  ISO 2005
All rights reserved. Unless otherwise specified, no part of this publication may be reproduced or utilized in any form or by any means,
electronic or mechanical, including photocopying and microfilm, without permission in writing from either ISO at the address below or
ISO's member body in the country of the requester.
ISO copyright office
Case postale 56 • CH-1211 Geneva 20
Tel. + 41 22 749 01 11
Fax + 41 22 749 09 47
E-mail copyright@iso.org
Web www.iso.org
Published in Switzerland

ii © ISO 2005 – All rights reserved

---------------------- Page: 2 ----------------------
ISO 8041:2005(E)
Contents Page
Foreword. v
1 Scope. 1
2 Normative references . 1
3 Terms, definitions and symbols . 2
3.1 Terms and definitions. 2
3.2 Symbols . 5
4 Reference environmental conditions. 7
5 Performance specifications . 7
5.1 General characteristics . 7
5.2 Display of signal magnitude . 9
5.3 Electrical output . 10
5.4 Vibration sensitivity. 10
5.5 Accuracy of indication at reference frequency under reference conditions . 10
5.6 Frequency weightings and frequency responses . 11
5.7 Amplitude linearity. 14
5.8 Instrument noise . 14
5.9 Signal-burst response . 14
5.10 Overload indication. 17
5.11 Under-range indication. 18
5.12 Time averaging. 18
5.13 Running r.m.s. acceleration. 18
5.14 Reset. 18
5.15 Timing facilities . 19
5.16 Electrical cross-talk . 19
5.17 Vibration transducer characteristics . 19
5.18 Power supply. 19
6 Mounting . 19
7 Environmental and electromagnetic criteria. 20
7.1 General. 20
7.2 Air temperature . 20
7.3 Surface temperature . 20
7.4 Electrostatic discharge. 20
7.5 Radio-frequency emissions and public-power-supply disturbances. 21
7.6 Immunity to a.c. power-frequency fields and radio-frequency fields. 21
7.7 Ingress of water and dust. 22
8 Provision for use with auxiliary devices. 22
9 Instrument marking. 22
10 Instrument documentation. 23
11 Testing and calibration. 23
12 Pattern evaluation . 24
12.1 Introduction . 24
12.2 Testing requirements. 25
12.3 Submission for testing . 25
12.4 Marking of the vibration meter and information in the instrument documentation. 25
12.5 Mandatory facilities and general requirements . 26
12.6 Initial instrument preparation . 26
© ISO 2005 – All rights reserved iii

---------------------- Page: 3 ----------------------
ISO 8041:2005(E)
12.7 Indication at the reference frequency under reference conditions.26
12.8 Electrical cross-talk.27
12.9 Vibration transducer .27
12.10 Amplitude linearity and under-range indication .27
12.11 Frequency weightings and frequency responses.29
12.12 Instrument noise.32
12.13 Signal-burst response .32
12.14 Overload indication.32
12.15 Reset.33
12.16 Combined axis outputs.33
12.17 A.c. electrical output .33
12.18 Timing facilities .33
12.19 Power supply .34
12.20 Environmental, electrostatic and radio-frequency tests.34
12.21 Test report.38
13 Verification tests.38
13.1 Introduction.38
13.2 Testing requirements.38
13.3 Submission for a test.39
13.4 Marking of the vibration meter and information in the instrument documentation .39
13.5 Mandatory facilities and general requirements.39
13.6 Initial instrument preparation.39
13.7 Indication at the reference frequency under reference conditions.39
13.8 Electrical cross-talk.40
13.9 Amplitude linearity and under-range indication .40
13.10 Frequency weightings and frequency responses.41
13.11 Instrument noise.43
13.12 Signal-burst response .43
13.13 Overload indication.43
13.14 Reset.44
13.15 Combined axis outputs.44
13.16 Test report.44
14 In-situ checks.44
14.1 Introduction.44
14.2 Preliminary inspection.45
14.3 Vibration sensitivity (field calibration) .45
Annex A (normative) Specification for vibration field calibrator .46
Annex B (informative) Frequency weightings.49
Annex C (informative) Realization of frequency weighting filters .68
Annex D (informative) Running r.m.s. time averaging.72
Annex E (informative) Vibration transducer characteristics .74
Annex F (informative) Tests for mounting systems .76
Annex G (normative) Instrument documentation.79
Annex H (normative) Phase-response requirements for measurement of non-r.m.s. quantities .84
Bibliography.91

iv © ISO 2005 – All rights reserved

---------------------- Page: 4 ----------------------
ISO 8041:2005(E)
Foreword
ISO (the International Organization for Standardization) is a worldwide federation of national standards bodies
(ISO member bodies). The work of preparing International Standards is normally carried out through ISO
technical committees. Each member body interested in a subject for which a technical committee has been
established has the right to be represented on that committee. International organizations, governmental and
non-governmental, in liaison with ISO, also take part in the work. ISO collaborates closely with the
International Electrotechnical Commission (IEC) on all matters of electrotechnical standardization.
International Standards are drafted in accordance with the rules given in the ISO/IEC Directives, Part 2.
The main task of technical committees is to prepare International Standards. Draft International Standards
adopted by the technical committees are circulated to the member bodies for voting. Publication as an
International Standard requires approval by at least 75 % of the member bodies casting a vote.
Attention is drawn to the possibility that some of the elements of this document may be the subject of patent
rights. ISO shall not be held responsible for identifying any or all such patent rights.
ISO 8041 was prepared by Technical Committee ISO/TC 108, Mechanical vibration and shock, Subcommittee
SC 3, Use and calibration of vibration and shock measuring instruments.
This second edition cancels and replaces the first edition (ISO 8041:1990), which has been technically
revised, and incorporates its Amendment, ISO 8041:1990/Amd.1:1999, and Technical Corrigendum
ISO 8041:1990/Cor.1:1993.
The reasons for the main changes introduced in this edition are as follows:
 to improve the specifications for human response to vibration measuring instrumentation;
 to incorporate into one document the specifications introduced by the 1999 amendment to ISO 8041:1990,
which themselves were required following the introduction of new frequency weightings in
ISO 2631-1:1997;
 to recognise changes in the frequency weighting specification introduced in ISO 5349-1:2001 that allows
frequencies outside the one-third octaves from 6,3 Hz to 1250 Hz to be excluded from the weighted
acceleration calculation (this is achieved by changing the frequencies at which the tolerance is extended
to −100 % to be the lower boundary of the 6,3 Hz one-third-octave bands and the upper boundary of the
1 250 Hz one-third-octave band);
 to introduce allowances for the uncertainties of testing the conformance of the human vibration measuring
instruments;
 to introduce a hierarchy of testing requirements (pattern evaluation, periodic verification and in-situ check)
with tests defined according to the needs of this hierarchy;
 to recognise the needs for the specification and testing of new parameters such as maximum transient
vibration value (MTVV) and vibration dose value (VDV);
 to recognise the need to test multi-axis instrumentation and to test combined results from these multi-axis
inputs;
 to introduce informative tests for mounting methods.
© ISO 2005 – All rights reserved v

---------------------- Page: 5 ----------------------
INTERNATIONAL STANDARD ISO 8041:2005(E)

Human response to vibration — Measuring instrumentation
1 Scope
This International Standard specifies the performance specifications and tolerance limits for instruments
designed to measure vibration values, for the purpose of assessing human response to vibration. It includes
requirements for pattern evaluation, periodic verification and in-situ checks, and the specification of vibration
calibrators for in-situ checks.
Vibration instruments specified in this International Standard can be single instruments, combinations of
instrumentation or computer-based acquisition and analysis systems.
Vibration instruments specified in this International Standard are intended to measure vibrations for one or
more applications, such as
 hand-transmitted vibration (see ISO 5349-1),
 whole-body vibration (see ISO 2631-1, ISO 2631-2, ISO 2631-4), and
 low-frequency whole-body vibration in the frequency range from 0,1 Hz to 0,5 Hz (see ISO 2631-1).
Vibration instruments can be designed for measurement according to one or more of the frequency weightings
defined within each of these applications.
Three levels of performance testing are defined in this International Standard:
a) pattern evaluation, i.e. a full test of the instrument against the specifications defined in this International
Standard;
b) periodic verification, i.e. an intermediate set of tests designed to ensure that an instrument remains within
the required performance specification, and
c) in-situ checks, i.e. a minimum level of testing required to indicate that an instrument is likely to be
functioning within the required performance specification.
2 Normative references
The following referenced documents are indispensable for the application of this document. For dated
references, only the edition cited applies. For undated references, the latest edition of the referenced
document (including any amendments) applies.
ISO 2041, Vibration and shock — Vocabulary
ISO 2631-1, Mechanical vibration and shock — Evaluation of human exposure to whole-body vibration —
Part 1: General requirements
ISO 2631-2, Mechanical vibration and shock — Evaluation of human exposure to whole-body vibration —
Part 2: Vibration in buildings (1 Hz to 80 Hz)
ISO 2631-4, Mechanical vibration and shock — Evaluation of human exposure to whole-body vibration —
Part 4: Guidelines for the evaluation of the effects of vibration and rotational motion on passenger and crew
comfort in fixed-guideway transport systems
ISO 5347 (all parts), Methods for the calibration of vibration and shock pick-ups
© ISO 2005 – All rights reserved 1

---------------------- Page: 6 ----------------------
ISO 8041:2005(E)
ISO 5348, Mechanical vibration and shock — Mechanical mounting of accelerometers
ISO 5349-1:2001, Mechanical vibration — Measurement and evaluation of human exposure to
hand-transmitted vibration — Part 1: General requirements
ISO 16063 (all parts), Methods for the calibration of vibration and shock transducers
IEC 61000-4-2:2001, Electromagnetic compatibility (EMC) — Part 4-2: Testing and measurement
techniques — Electrostatic discharge immunity test
IEC 61000-4-3:2002, Electromagnetic compatibility (EMC) — Part 4-3: Testing and measurement
techniques —Radiated, radio-frequency, electromagnetic field immunity test
IEC 61000-4-6, Electromagnetic compatibility (EMC) — Part 4-6: Testing and measurement techniques —
Immunity to conducted disturbances, induced by radio-frequency fields
IEC 61000-6-2:1999, Electromagnetic compatibility (EMC) — Part 6-2: Generic standards — Immunity for
industrial environments
CISPR 22:2003, Information technology equipment — Radio disturbance characteristics — Limits and
methods of measurement
GUM, Guide to the expression of uncertainty in measurement. BIPM, IEC, IFCC, ISO, IUPAC, IUPAP, OIML,
1993
3 Terms, definitions and symbols
3.1 Terms and definitions
For the purposes of this document, the terms and definitions given in ISO 2041, together with the following,
apply.
3.1.1
vibration acceleration
component of acceleration, where the axis of measurement is specified by application standards
3.1.2
band-limiting frequency weighting
component of a frequency weighting defined by the high and low pass band-limiting filters
3.1.3
band-limited frequency range
frequency range defined by the band-limited component of a frequency weighting
3.1.4
nominal frequency range
frequency range of interest, as defined in the relevant measurement standard
3.1.5 Frequency-weighted values
3.1.5.1
time-averaged weighted acceleration value
frequency-weighted r.m.s. vibration acceleration value in a specified axis, a , in metres per second squared or
w
radians per second squared, as defined by the expression:
1
2
T

1
2

aa= ξξd (1)
()
ww


T
0
2 © ISO 2005 – All rights reserved

---------------------- Page: 7 ----------------------
ISO 8041:2005(E)
where
a (ξ ) is the translational or rotational, weighted vibration acceleration in a specified axis as a function
w
2
of the instantaneous time, ξ, in metres per second squared (m/s ) or radians per second
2
squared (rad/s ), respectively;
T is the duration of the measurement
3.1.5.2
time-averaged weighted acceleration level
frequency-weighted r.m.s. vibration acceleration level expressed in decibels, as defined by
a
w
L = 20 lg dB (2)
w
a
0
where
a is defined in 3.1.5.1;
w
−6 2
a is the reference acceleration (defined as 10 m/s in ISO 1683)
0
3.1.5.3
running r.m.s. acceleration value
frequency-weighted running r.m.s. vibration acceleration, in metres per second squared, defined by the
expression
1
t 2

1
2

at = a ξξd (3)
() ()
w,θ w

θ
t−θ

where
a (ξ ) is the frequency-weighted instantaneous vibration acceleration at time ξ , in metres per second
w
squared;
θ is the integration time of the measurement;
t is the instantaneous time
NOTE Exponential averaging may be used for the running r.m.s. method, as an approximation of the linear averaging.
The exponential averaging is defined as follows:
1
2
t

1 ξ − t
2

at = aξξexp d (4)
() ()
w,τ w 


ττ

−∞

where τ is the time constant.
3.1.5.4
maximum transient vibration value
MTVV
maximum value of the running r.m.s. vibration acceleration value when the integration time is equal to 1 s
3.1.5.5
motion sickness dose value
MSDV
1,5
integral of the squared weighted instantaneous vibration acceleration a (t) in m/s as defined by the
w
expression:
© ISO 2005 – All rights reserved 3

---------------------- Page: 8 ----------------------
ISO 8041:2005(E)
1
2
Φ

2

MSDV = a ξξd (5)
()
w


0
where Φ is the total period during which motion could occur
NOTE 1 The motion sickness dose value may be obtained from the frequency weighted r.m.s. vibration acceleration
1
2
through multiplication by Φ .
NOTE 2 For measurement instrumentation, the exposure period Φ is likely to be assumed to be equal to the
measurement period, T, unless otherwise indicated.
3.1.5.6
vibration dose value
VDV
1,75
integral of the fourth power of the weighted instantaneous vibration acceleration a (t) in m/s as defined by
w
the expression
1
4
Φ

4

VDV = a ()ξξd (6)
w


0
where Φ is the total (daily) period for which exposure occurs
NOTE 1 The vibration dose value is more sensitive to peaks than is the r.m.s. value.
NOTE 2 For measurement instrumentation, the exposure period Φ is likely to be assumed to be equal to the
measurement period, T, unless otherwise indicated.
3.1.5.7
vibration total value
combined vibration from three axes of translational vibration, as defined by the expression
22 2
a=+kakak+ a (7)
wwvxx ywy zwz
where
a , a and a are the vibration values in the three orthogonal axes x, y and z;
wx wy wz
k , k and k are multiplying constants whose values depend on the measurement application
x y z
3.1.5.8
peak vibration value
maximum modulus of the instantaneous (positive and negative) peak values of the frequency-weighted
acceleration
3.1.5.9
crest factor
parameter for a measurement period, given by the peak vibration value divided by the r.m.s. acceleration
value, with both values having the same frequency weighting
3.1.6
linear operating range
on each measurement range, the range between lower and upper boundaries over which the linearity errors
are within the applicable tolerance limits specified in this International Standard
4 © ISO 2005 – All rights reserved

---------------------- Page: 9 ----------------------
ISO 8041:2005(E)
3.1.7
overload
condition that occurs when the upper boundary of the linear operating range is exceeded
3.1.8
under-range
condition that occurs when the vibration value is below the lower boundary of the linear operating range
3.1.9
reference measurement range
level range specified for testing the characteristics of the vibration instrumentation
NOTE This range is that used for measuring the reference vibration.
3.1.10
reference vibration signal
sinusoidal vibration signal, the magnitude and frequency of which is specified in this International Standard for
testing the electromechanical performance of a human-vibration meter
NOTE Different reference vibration signals are specified according to the application of the instrumentation.
3.1.11
calibration check frequency
frequency specified for providing a check of the vibration sensitivity of the instrument
3.1.12
tone burst
one or more complete cycles of a sinusoidal signal that start and end at a zero crossing of the waveform
3.1.13
signal burst
one or more complete cycles of a periodic signal (such as saw-tooth) that start and end at a zero crossing of
the waveform
3.1.14
vibration measuring instrumentation
combination of a vibration transducer, signal processor and display, being any single instrument, or a
collection of instruments, which is capable of measuring parameters relating to human response to vibration
NOTE See Figure 1.
3.1.15
instrument documentation
instruction manual, operating procedure, or other documentation provided for the use of users of the vibration
measurement instrument
3.2 Symbols
For the purposes of this document, the following symbols and abbreviated terms are used:
a time-averaged frequency-weighted single-axis vibration acceleration
w
a (t), a (ξ ) instantaneous frequency-weighted translational or rotational single-axis acceleration at time t,
w w
or timeξ
f frequency
H overall frequency weighting function
k multiplying constants applied to the whole-body frequency-weighted acceleration value for axis i
i
n one-third-octave band number
© ISO 2005 – All rights reserved 5

---------------------- Page: 10 ----------------------
ISO 8041:2005(E)
t or ξ instantaneous time
T measurement duration
s variable of the Laplace transform
W frequency weighting x
x
Φ exposure duration
∆ϕ phase error
τ exponential averaging time constant
θ linear averaging time
MTVV maximum transient vibration value
MSDV motion sickness dose value
VDV vibration dose value

Key
1 transducer 8 frequency weighting (including band-limiting)
2 mounting system 9 band-limited output
3 vibrating surface 10 frequency-weighted output
...

NORME ISO
INTERNATIONALE 8041
Deuxième édition
2005-04-01


Réponse des individus aux vibrations —
Appareillage de mesure
Human response to vibration — Measuring instrumentation




Numéro de référence
ISO 8041:2005(F)
©
ISO 2005

---------------------- Page: 1 ----------------------
ISO 8041:2005(F)
PDF – Exonération de responsabilité
Le présent fichier PDF peut contenir des polices de caractères intégrées. Conformément aux conditions de licence d'Adobe, ce fichier
peut être imprimé ou visualisé, mais ne doit pas être modifié à moins que l'ordinateur employé à cet effet ne bénéficie d'une licence
autorisant l'utilisation de ces polices et que celles-ci y soient installées. Lors du téléchargement de ce fichier, les parties concernées
acceptent de fait la responsabilité de ne pas enfreindre les conditions de licence d'Adobe. Le Secrétariat central de l'ISO décline toute
responsabilité en la matière.
Adobe est une marque déposée d'Adobe Systems Incorporated.
Les détails relatifs aux produits logiciels utilisés pour la création du présent fichier PDF sont disponibles dans la rubrique General Info
du fichier; les paramètres de création PDF ont été optimisés pour l'impression. Toutes les mesures ont été prises pour garantir
l'exploitation de ce fichier par les comités membres de l'ISO. Dans le cas peu probable où surviendrait un problème d'utilisation,
veuillez en informer le Secrétariat central à l'adresse donnée ci-dessous.


©  ISO 2005
Droits de reproduction réservés. Sauf prescription différente, aucune partie de cette publication ne peut être reproduite ni utilisée sous
quelque forme que ce soit et par aucun procédé, électronique ou mécanique, y compris la photocopie et les microfilms, sans l'accord écrit
de l'ISO à l'adresse ci-après ou du comité membre de l'ISO dans le pays du demandeur.
ISO copyright office
Case postale 56 • CH-1211 Geneva 20
Tel. + 41 22 749 01 11
Fax. + 41 22 749 09 47
E-mail copyright@iso.org
Web www.iso.org
Publié en Suisse

ii © ISO 2005 – Tous droits réservés

---------------------- Page: 2 ----------------------
ISO 8041:2005(F)
Sommaire Page
Avant-propos. v
1 Domaine d'application. 1
2 Références normatives . 1
3 Termes, définitions et symboles. 2
3.1 Termes et définitions. 2
3.2 Symboles . 8
4 Conditions ambiantes de référence. 8
5 Spécifications de performance . 9
5.1 Caractéristiques générales. 9
5.2 Affichage de la grandeur du signal. 10
5.3 Signal de sortie électrique . 12
5.4 Sensibilité aux vibrations . 12
5.5 Exactitude d'indication à la fréquence de référence dans les conditions de référence . 12
5.6 Pondérations en fréquence et réponses en fréquence. 13
5.7 Linéarité d'amplitude. 16
5.8 Bruit émis par l'instrument . 17
5.9 Réponse à un signal de train d'ondes . 17
5.10 Indication de surcharge . 20
5.11 Indication de très faible niveau . 21
5.12 Calcul de la moyenne temporelle. 21
5.13 Accélération efficace mobile . 21
5.14 Réinitialisation . 22
5.15 Dispositifs de temporisation . 22
5.16 Diaphonie électrique. 22
5.17 Caractéristiques du capteur de vibrations.22
5.18 Alimentation . 22
6 Montage . 23
7 Critères d'environnement et électromagnétiques. 23
7.1 Généralités . 23
7.2 Température de l'air. 23
7.3 Température de surface . 23
7.4 Décharge électrostatique. 24
7.5 Émissions de fréquences radioélectriques et perturbations de l'alimentation réseau. 24
7.6 Immunité aux champs de fréquences en courant alternatif et aux champs de fréquences
radioélectriques . 25
7.7 Entrée d'eau et de poussière . 25
8 Disposition pour une utilisation avec des dispositifs auxiliaires. 25
9 Marquage de l'instrument . 26
10 Documentation d'accompagnement de l'instrument . 26
11 Essais et étalonnage . 26
12 Essais de conformité. 28
12.1 Introduction . 28
12.2 Exigences d'essais . 28
12.3 Soumission aux essais . 29
12.4 Marquage de l'instrument de mesure des vibrations et informations dans la
documentation d'accompagnement de l'instrument . 29
© ISO 2005 – Tous droits réservés iii

---------------------- Page: 3 ----------------------
ISO 8041:2005(F)
12.5 Dispositifs obligatoires et exigences générales. 29
12.6 Préparation initiale de l'instrument . 29
12.7 Indication à la fréquence de référence dans les conditions de référence . 29
12.8 Diaphonie électrique. 30
12.9 Capteur de vibrations . 30
12.10 Linéarité d'amplitude et indication de très faible niveau. 30
12.11 Pondérations en fréquence et réponses en fréquence. 33
12.12 Bruit émis par l'instrument. 36
12.13 Réponse à un signal de train d'ondes . 36
12.14 Indication de surcharge. 36
12.15 Réinitialisation. 37
12.16 Résultats combinant les différentes directions. 37
12.17 Signal de sortie électrique en courant alternatif. 37
12.18 Dispositifs de temporisation. 37
12.19 Alimentation. 37
12.20 Essais d'environnement, électrostatiques et de fréquences radioélectriques . 38
12.21 Rapport d'essai . 42
13 Essais de vérification . 42
13.1 Introduction . 42
13.2 Exigences d'essais . 43
13.3 Soumission à un essai . 43
13.4 Marquage de l'instrument de mesure des vibrations et informations dans la
documentation d'accompagnement de l'instrument. 43
13.5 Dispositifs obligatoires et exigences générales. 43
13.6 Préparation initiale de l'instrument . 44
13.7 Indication à la fréquence de référence et dans les conditions de référence. 44
13.8 Diaphonie électrique. 45
13.9 Linéarité d'amplitude et indication de très faible niveau. 45
13.10 Pondérations en fréquence et réponses en fréquence. 46
13.11 Bruit émis par l'instrument. 48
13.12 Réponse à un signal de train d'ondes . 48
13.13 Indication de surcharge. 49
13.14 Réinitialisation. 49
13.15 Résultats combinant les différentes directions. 49
13.16 Rapport d'essai . 50
14 Contrôles in situ. 50
14.1 Introduction . 50
14.2 Inspection préliminaire. 50
14.3 Sensibilité aux vibrations (étalonnage de champ vibratoire). 50
Annexe A (normative) Spécification relative au calibrateur de vibration. 51
Annexe B (informative) Pondérations en fréquence. 53
Annexe C (informative) Réalisation des filtres de pondération en fréquence . 72
Annexe D (informative) Calcul de la moyenne temporelle efficace mobile . 76
Annexe E (informative) Caractéristiques du capteur de vibrations . 79
Annexe F (informative) Essais relatifs aux systèmes de montage. 82
Annexe G (normative) Documentation d'accompagnement de l'instrument . 85
Annexe H (normative) Exigences relatives à la réponse de phase pour le mesurage des
grandeurs non efficaces. 91
Bibliographie . 99

© ISO 2005 – Tous droits réservés iv

---------------------- Page: 4 ----------------------
ISO 8041:2005(F)
Avant-propos
L'ISO (Organisation internationale de normalisation) est une fédération mondiale d'organismes nationaux de
normalisation (comités membres de l'ISO). L'élaboration des Normes internationales est en général confiée
aux comités techniques de l'ISO. Chaque comité membre intéressé par une étude a le droit de faire partie du
comité technique créé à cet effet. Les organisations internationales, gouvernementales et non
gouvernementales, en liaison avec l'ISO participent également aux travaux. L'ISO collabore étroitement avec
la Commission électrotechnique internationale (CEI) en ce qui concerne la normalisation électrotechnique.
Les Normes internationales sont rédigées conformément aux règles données dans les Directives ISO/CEI,
Partie 2.
La tâche principale des comités techniques est d'élaborer les Normes internationales. Les projets de Normes
internationales adoptés par les comités techniques sont soumis aux comités membres pour vote. Leur
publication comme Normes internationales requiert l'approbation de 75 % au moins des comités membres
votants.
L'attention est appelée sur le fait que certains des éléments du présent document peuvent faire l'objet de
droits de propriété intellectuelle ou de droits analogues. L'ISO ne saurait être tenue pour responsable de ne
pas avoir identifié de tels droits de propriété et averti de leur existence.
L'ISO 8041 a été élaborée par le comité technique ISO/TC 108, Vibrations et chocs mécaniques, sous-comité
SC 3, Utilisation et étalonnage des instruments de mesure des vibrations et des chocs.
Cette deuxième édition annule et remplace la première édition (ISO 8041:1990), qui a fait l'objet d'une
révision technique et intègre son amendement, ISO 8041:1990/Amd.1:1999 et son rectificatif technique,
ISO 8041:1990/Cor.1:1993.
Les raisons des principaux changements introduits dans la présente édition sont les suivants:
⎯ amélioration des spécifications relatives à l'appareillage de mesure de la réponse des individus aux
vibrations;
⎯ intégration dans un document des spécifications définies par l'amendement de 1999 à l'ISO 8041:1990,
l'existence de ces dernières se révélant nécessaire suite à l'introduction de nouvelles pondérations en
fréquence dans l'ISO 2631-1:1997;
⎯ validation des changements concernant la spécification de la pondération en fréquence établie par
l'ISO 5349-1:2001, permettant d'exclure les fréquences à l'extérieur des tiers d'octave compris entre
6,3 Hz et 1 250 Hz du calcul de l'accélération pondérée (ceci s'effectue en modifiant les fréquences pour
lesquelles la tolérance est étendue à −100 % afin qu'elles constituent la limite inférieure des bandes de
tiers d'octave de 6,3 Hz et la limite supérieure de la bande de tiers d'octave de 1 250 Hz);
⎯ prévision de tolérances eu égard aux incertitudes concernant la vérification de la conformité des
instruments de mesure de la réponse des individus aux vibrations;
⎯ établissement d'une hiérarchie propre aux exigences relatives aux essais (essais de conformité,
vérification périodique et contrôle in situ), ces derniers étant définis selon les besoins de ladite hiérarchie;
⎯ reconnaissance de la nécessité de spécifier et de vérifier les nouveaux paramètres tels que la valeur
maximale de la vibration transitoire (MTVV) et la valeur de dose vibratoire (VDV);
⎯ reconnaissance de la nécessité de soumettre à essai les appareillages de mesure des vibrations et de
vérifier les résultats combinés obtenus à partir de ces informations selon plusieurs axes;
⎯ réalisation d'essais informatifs concernant les méthodes de montage.
© ISO 2005 – Tous droits réservés v

---------------------- Page: 5 ----------------------
NORME INTERNATIONALE ISO 8041:2005(F)

Réponse des individus aux vibrations — Appareillage de
mesure
1 Domaine d'application
La présente Norme internationale donne les spécifications de performances et les limites de tolérance pour
les instruments destinés à mesurer les valeurs de vibration afin d'évaluer la réponse des individus aux
vibrations. Elle inclut les exigences relatives aux essais de conformité, à la vérification périodique et aux
contrôles in situ, ainsi que la spécification concernant les calibrateurs de vibration destinés aux contrôles
in situ.
Les instruments de mesure des vibrations spécifiés dans la présente Norme internationale peuvent être des
instruments simples, des combinaisons d'instruments ou des systèmes d'acquisition et d'analyse informatisés.
Les instruments de mesure des vibrations spécifiés dans la présente Norme internationale sont destinés à
mesurer les vibrations pour une ou plusieurs applications, telles que
⎯ les vibrations transmises par la main (voir l'ISO 5349-1),
⎯ les vibrations globales du corps (voir l'ISO 2631-1, l'ISO 2631-2 et l'ISO 2631-4), et
⎯ les vibrations globales du corps en basses fréquences dans la gamme de fréquences comprises entre
0,1 Hz et 0,5 Hz (voir l'ISO 2631-1).
Les instruments de mesure des vibrations peuvent être conçus pour effectuer un mesurage selon une ou
plusieurs pondérations en fréquence définies pour chacune de ces applications.
Trois niveaux de vérification des performances sont définis dans la présente Norme internationale:
a) essais de conformité, c'est-à-dire un essai complet de l'instrument par rapport aux spécifications définies
dans la présente Norme internationale;
b) vérification périodique, c'est-à-dire un ensemble intermédiaire d'essais destinés à s'assurer qu'un
instrument donné relève effectivement de la spécification de performances requise, et
c) contrôles in situ, c'est-à-dire un niveau minimal de vérification à appliquer pour indiquer que le
fonctionnement d'un instrument donné relève vraisemblablement de la spécification de performances
requise.
2 Références normatives
Les documents de référence suivants sont indispensables pour l'application du présent document. Pour les
références datées, seule l'édition citée s'applique. Pour les références non datées, la dernière édition du
document de référence (y compris les éventuels amendements) s'applique.
ISO 2041, Vibrations et chocs — Vocabulaire
ISO 2631-1, Vibrations et chocs mécaniques — Évaluation de l'exposition des individus à des vibrations
globales du corps — Partie 1: Spécifications générales
© ISO 2005 – Tous droits réservés 1

---------------------- Page: 6 ----------------------
ISO 8041:2005(F)
ISO 2631-2, Vibrations et chocs mécaniques — Évaluation de l'exposition des individus à des vibrations
globales du corps — Partie 2: Vibrations dans les bâtiments (1 Hz à 80 Hz)
ISO 2631-4, Vibrations et chocs mécaniques — Évaluation de l'exposition des individus à des vibrations
globales du corps — Partie 4: Lignes directrices pour l'évaluation des effets des vibrations et du mouvement
de rotation sur le confort des passagers et du personnel dans les systèmes de transport guidé
ISO 5347 (toutes les parties), Méthodes pour l'étalonnage de capteurs de vibrations et de chocs
ISO 5348, Vibrations et chocs mécaniques — Fixation mécanique des accéléromètres
ISO 5349-1:2001, Vibrations mécaniques — Mesurage et évaluation de l'exposition des individus aux
vibrations transmises par la main — Partie 1: Exigences générales
ISO 16063 (toutes les parties), Méthodes pour l'étalonnage des transducteurs de vibrations et de chocs
CEI 61000-4-2:2001, Compatibilité électromagnétique (CEM) — Partie 4-2: Techniques d'essai et de
mesure — Essai d'immunité aux décharges électrostatiques
CEI 61000-4-3:2002, Compatibilité électromagnétique (CEM) — Partie 4-3: Techniques d'essai et de
mesure — Essai d'immunité aux champs électromagnétiques rayonnés aux fréquences radioélectriques
CEI 61000-4-6, Compatibilité électromagnétique (CEM) — Partie 4-6: Techniques d'essai et de mesure —
Immunité aux perturbations conduites, induites par les champs radioélectriques
CEI 61000-6-2:1999, Compatibilité électromagnétique (CEM) — Partie 6-2: Normes génériques — Immunité
pour les environnements industriels
CISPR 22:2003, Appareils de traitement de l'information — Caractéristiques des perturbations
radioélectriques — Limites et méthodes de mesure
GUM, Guide pour l'expression de l'incertitude de mesure. BIPM, CEI, FICC, ISO, OIML, UICPA, UIPPA, 1995
3 Termes, définitions et symboles
3.1 Termes et définitions
Pour les besoins du présent document, les termes et définitions donnés dans l'ISO 2041 ainsi que les
suivants s'appliquent.
3.1.1
accélération de vibration
élément d'accélération dont l'axe de mesurage est spécifié par les normes d'application
3.1.2
pondération en fréquence à limite de bande
élément d'une pondération en fréquence défini par les filtres de limite de bandes passe-haut et passe-bas
3.1.3
gamme de fréquences à limite de bande
gamme de fréquences définie par l'élément limiteur de bande d'une pondération en fréquence
3.1.4
gamme de fréquences nominale
gamme de fréquences représentative, telle que définie dans l'étalon approprié
2 © ISO 2005 – Tous droits réservés

---------------------- Page: 7 ----------------------
ISO 8041:2005(F)
3.1.5 Valeurs pondérées en fréquence
3.1.5.1
moyenne temporelle de valeur d'accélération pondérée
valeur d'accélération de vibration efficace pondérée en fréquence sur un axe spécifié, a en mètres par
w
seconde carrée ou en radians par seconde carrée, telle que définie par l'expression
1
T 2
⎛⎞
1
2
⎜⎟
aa= ξξd (1)
()
ww

⎜⎟T
0
⎝⎠

a (ξ) est l'accélération en translation ou en rotation de vibration pondérée, sur un axe spécifié, en
w
2
fonction du temps instantané, ξ, en mètres par seconde carrée (m/s ) ou en radians par
2
seconde carrée (rad/s ), respectivement;
T est la durée du mesurage
3.1.5.2
moyenne temporelle du niveau d'accélération pondérée
niveau d'accélération de vibration efficace pondéré en fréquence, exprimé en décibels, tel que défini par
a
w
L = 20 lg dB (2)
w
a
0

a est définie en 3.1.5.1;
w
−6 2
a est l'accélération de référence (définie comme 10 m/s dans l'ISO 1683)
0
3.1.5.3
valeur efficace mobile de l'accélération
la valeur efficace mobile de l'accélération de vibration pondérée en fréquence, exprimée en mètres par
seconde carrée, est définie par l'expression suivante
1
2
t
⎛⎞
1
2
⎜⎟
at = a ξξd (3)
() ()
w,θ w

⎜⎟
θ
⎝⎠t−θ

a (ξ) est l'accélération de vibration instantanée pondérée en fréquence au temps ξ, exprimée en
w
mètre par seconde carrée;
θ est le temps d'intégration du mesurage;
t est le temps instantané
NOTE Le calcul de la moyenne exponentielle peut être utilisé pour la méthode de la valeur efficace mobile, comme
valeur approximative du calcul de la moyenne linéaire. Il se définit comme suit:
1
t 2
⎛⎞
1 ξ − t
⎛⎞
2
⎜⎟
at = aξξexp d (4)
() ()
w,τ w ⎜⎟

⎜⎟
ττ
⎝⎠
⎝⎠−∞
où τ est la constante de temps.
© ISO 2005 – Tous droits réservés 3

---------------------- Page: 8 ----------------------
ISO 8041:2005(F)
3.1.5.4
valeur maximale de la vibration transitoire
MTVV
valeur maximale de la valeur efficace mobile de l'accélération de vibration lorsque le temps d'intégration est
égal à 1 s
3.1.5.5
valeur de la dose relative au mal des transports
MSDV
1,5
intégrale de l'accélération de vibration instantanée pondérée mise au carré a (t) en m/s , telle que définie
w
par l'expression
1
2
Φ
⎛⎞
2
⎜⎟
MSDV = a ξξd (5)
()
w

⎜⎟
⎝⎠0
où Φ est la période totale au cours de laquelle un déplacement peut se produire
NOTE 1 La valeur de la dose relative au mal des transports peut être obtenue à partir de la valeur efficace de
1
2
l'accélération de vibration pondérée en fréquence multipliée par Φ .
NOTE 2 Pour les appareillages de mesure, la période d'exposition Φ est supposée être égale à la période de
mesurage, T sauf indication contraire.
3.1.5.6
valeur de dose vibratoire
VDV
1,75
intégrale de la racine quatrième de l'accélération de vibration instantanée pondérée a (t) en m/s , telle que
w
définie par l'expression
1
4
Φ
⎛⎞
4
⎜⎟
VDV = a ξξd (6)
()
w

⎜⎟
⎝⎠0
où Φ est la période (quotidienne) totale d'exposition effective
NOTE 1 La valeur de dose vibratoire est plus sensible aux crêtes que la valeur efficace.
NOTE 2 Pour les appareillages de mesure, la période d'exposition Φ est supposée être égale à la période de
mesurage, T sauf indication contraire.
3.1.5.7
valeur de vibration totale
vibration combinée à partir de trois axes de vibration en translation, telle que définie par l'expression
22 2
ak=+a ka+ka (7)
wwvxx ywy zwz

a , a et a sont les valeurs de vibration sur les trois axes orthogonaux x, y et z;
wx wy wz
k , k et k sont des constantes de multiplication dont les valeurs dépendent de l'application de
x y z
mesurage
4 © ISO 2005 – Tous droits réservés

---------------------- Page: 9 ----------------------
ISO 8041:2005(F)
3.1.5.8
valeur de vibration de crête
module maximal des valeurs de crête instantanées (positive et négative) de l'accélération pondérée en
fréquence
3.1.5.9
facteur de crête
paramètre relatif à une période de mesurage, donné par la valeur de vibration de crête divisée par la valeur
efficace d'accélération, les deux valeurs ayant la même pondération en fréquence
3.1.6
plage de fonctionnement linéaire
pour chaque gamme de mesure, gamme comprise entre les limites inférieure et supérieure à l'intérieur
desquelles les erreurs de linéarité se situent dans les limites de tolérance applicables spécifiées dans la
présente Norme internationale
3.1.7
surcharge
condition se produisant en cas de dépassement de la limite supérieure de la plage de fonctionnement linéaire
3.1.8
sous estimation
condition se produisant lorsque la valeur de vibration se situe en dessous de la limite inférieure de la plage de
fonctionnement linéaire
3.1.9
gamme de mesure de référence
gamme de niveau spécifiée pour la vérification des caractéristiques des instruments de mesure des vibrations
NOTE Cette gamme est celle utilisée pour mesurer les vibrations de référence.
3.1.10
signal de vibration de référence
signal de vibration sinusoïdal, dont l'amplitude et la fréquence sont spécifiées dans la présente Norme
internationale en vue de la vérification des performances électromécaniq
...

Questions, Comments and Discussion

Ask us and Technical Secretary will try to provide an answer. You can facilitate discussion about the standard in here.