Abstract

This document specifies requirements and gives recommendations for next generation sequencing (NGS) workflows. This document covers the pre-examination processes, human DNA (somatic and germline) isolation, sequencing library preparation, sequencing, sequence analysis and reporting of the examination of sequences for diagnostic purposes from isolated DNA from, e.g. formalin-fixed and paraffin embedded tissues, fresh frozen tissues, fine needle aspirates (FNA), whole blood, circulating tumour cells (CTCs), exosomes and other extracellular vesicles, circulating cell free DNA from plasma, and DNA from saliva. NOTE 1 Typical applications include, but are not limited to, NGS for oncology, pharmacogenomics and clinical genetics; approaches include panels (e.g. disease panels, exome panels, target gene panels and in silico panels), exome and whole genome sequencing, as well as certain epigenetics and certain single-cell analyses. This document is addressing in vitro diagnostic examinations including laboratory developed tests and is applicable to medical laboratories, molecular pathology laboratories and molecular genetic laboratories, in vitro diagnostic developers and manufacturers, biobanks as well as institutions and organizations performing biomedical research or sequencing services. This document is not applicable for in situ sequencing, DNA-mediated protein sequencing, forensic sequencing, sequencing of pathogens or microorganisms and microbiome analysis. NOTE 2 International, national or regional regulations or requirements or multiples of them can also apply to specific topics covered in this document.

Status
Not Published
Public Enquiry End Date
29-Nov-2026
Technical Committee
VAZ - Healthcare
Current Stage
4020 - Public enquire (PE) (Adopted Project)
Start Date
24-Sep-2026
Due Date
11-Feb-2027

Buy Documents

Draft

oSIST prEN ISO 25379-1:2026

English language (60 pages)
Preview
Preview
e-Library read for
×1 day

oSIST prEN ISO 25379-1:2026 is the Slovenian draft adoption of ISO/DIS 25379-1:2026, published by SIST and developed in ISO/TC 212. It specifies requirements and recommendations for in vitro diagnostic next generation sequencing (NGS) workflows for human DNA examination, from pre-examination through sequencing, sequence analysis and reporting for diagnostic purposes. It is written for medical laboratories, molecular pathology and molecular genetic laboratories, in vitro diagnostic developers and manufacturers, biobanks, and organizations that perform biomedical research or sequencing services.

What does oSIST prEN ISO 25379-1:2026 specify?

oSIST prEN ISO 25379-1:2026 covers NGS workflows for in vitro diagnostics and biomedical research, focused on human DNA, including somatic and germline DNA. It applies to diagnostic examinations from isolated DNA taken from sources such as FFPE tissue, fresh frozen tissue, FNA, whole blood, CTCs, exosomes and other EVs, circulating cell free DNA from plasma, and saliva.

The standard also frames typical use cases such as oncology, pharmacogenomics and clinical genetics. It covers panels, exome sequencing, whole genome sequencing, and some epigenetics and single-cell analyses.

It does not apply to in situ sequencing, DNA-mediated protein sequencing, forensic sequencing, sequencing of pathogens or microorganisms, or microbiome analysis.

The document is organized into Clauses 1 to 10, with Annex A and Annex B.

AnnexWhat it covers
ANormative requirements for an in vitro diagnostic NGS workflow for single-cell analyses
BInformative NGS workflow scheme for the examination of DNA

What are the key requirements of oSIST prEN ISO 25379-1:2026?

oSIST prEN ISO 25379-1:2026 treats the whole NGS workflow as one controlled process. The main practical message is that design, validation, reporting and routine use must all fit the intended use and the specimen type.

  • Clause 4.1 links the workflow to a quality system. Medical laboratories are directed to follow ISO 15189 or an equivalent such as ISO/IEC 17020 or ISO/IEC 17025, while IVD manufacturers should follow ISO 13485. In practice, this means the NGS workflow is not only a technical method - it has to sit inside a documented quality management system.

  • Clause 4.2 requires defined design inputs. The examination manufacturer shall document scientific validity, customer and stakeholder needs, intended use, and design input specifications. This matters because the reportable claims, specimen types, controls and performance targets all depend on those decisions.

  • Clause 4.3 focuses development on risk and optimization. The document points to risk assessment, Design of Experiments (DoE), and guard band studies for critical input parameters. In practice, the manufacturer should identify what most affects the result, test it systematically, and define the tolerance limits before routine use.

  • Clause 4.4 requires verification and validation of the whole workflow. The manufacturer shall document verification and validation plans and confirm the intended use on the basis of the planned workflow. If products are supplied for use in the workflow, the performance requirements must still be demonstrated. This is what turns a promising assay design into one that is ready for use.

  • Clause 4.5 requires technical performance characteristics to be established. The document highlights analytical accuracy, analytical sensitivity or positive percentage agreement (PPA) when no reference method is available, analytical specificity or negative percentage agreement (NPA) when no reference method is available, precision, limits of detection (LOD) and where applicable LODP, reportable range, reference range, interferences, sequence base quality, clinically relevant performance characteristics, and the performance of each reported variant class such as SNVs, indels, CNAs and SVs. For the user, this is the evidence base for deciding what the assay can and cannot reliably report.

  • Clauses 5 and 6 define the technical workflow. The standard covers specimen collection, transport, storage, DNA isolation, DNA quality and quantity evaluation, sequencing library preparation, equipment requirements, sequencing quality control, data analysis, and QC. It explicitly includes material-specific routes such as FFPE, fresh frozen tissue, FNA, whole blood, CTCs, EVs, plasma ccfDNA, saliva and other body fluids. In practice, the lab must choose and document the right path for the specimen actually received.

  • Clauses 7 to 10 control reporting, implementation and ongoing QA. The report has to be accurate, clear and unambiguous, with defined attributes and content, and with reporting that supports clinical interpretation. Clause 9 prefers qualitative wording such as “Detected” or “Not detected” rather than “Positive” or “Negative”. Clause 10 requires ongoing monitoring, internal and external controls, documentation of performance data, and participation in proficiency testing where appropriate.

Annex A adds single-cell specific requirements, including the need to address cell capture, amplification bias, sequencing depth, data analysis and the special challenges of CTCs and single cells from tissue. Annex B gives the overall workflow view for DNA examination.

What terms does oSIST prEN ISO 25379-1:2026 define?

  • Next generation sequencing (NGS) - advanced sequencing technologies developed after DNA sequencing by the dideoxy termination method.
  • Massively parallel sequencing (MPS) - nucleic acid sequencing techniques that detect millions or billions of sequences in parallel.
  • Sequencing library - DNA, cDNA or RNA prepared for NGS within a specific size range and carrying adaptors for the sequencing platform.
  • Pre-examination processes - the steps from request and specimen collection through transport, storage and isolation of analytes.
  • Circulating cell free DNA (ccfDNA) - extracellular human DNA present in blood and plasma, including DNA in vesicles such as exosomes.
  • Laboratory developed examination - an examination made or modified and used by a health institution or medical laboratory only in its own facility for specific needs.
  • Bioinformatics pipeline - the suite of bioinformatics tools that process NGS data.

Who uses oSIST prEN ISO 25379-1:2026?

oSIST prEN ISO 25379-1:2026 is used by medical laboratories, molecular pathology laboratories and molecular genetic laboratories that design, validate and run NGS assays. It is also relevant to IVD developers and manufacturers, biobanks, and institutions or organizations providing biomedical research or sequencing services.

In practice, these users apply it to specimen handling, DNA isolation, library preparation, sequencing, data analysis, reporting, QC, validation, implementation into routine work, and ongoing performance monitoring.

Which standards are used with oSIST prEN ISO 25379-1:2026?

  • ISO 7552-2 - pre-examination for CTCs in venous whole blood, for isolated DNA.
  • ISO 18702 - pre-examination for exosomes and other extracellular vesicles in venous whole blood.
  • ISO 18704 - pre-examination for urine and other body fluids, for isolated cell-free DNA.
  • ISO 4307 - pre-examination for saliva, for isolated human DNA.
  • ISO 20166-3 - pre-examination for FFPE tissue, isolated DNA.
  • ISO 20184-3 - pre-examination for frozen tissue, isolated DNA.
  • ISO 20186-2 - pre-examination for venous whole blood, isolated genomic DNA.
  • ISO 20186-3 - pre-examination for venous whole blood, isolated circulating cell free DNA from plasma.
  • ISO 20397-1 - library preparation and nucleic acid preparation for massively parallel sequencing.
  • ISO 20397-2 - quality evaluation of sequencing data and QC requirements for sequencing data analysis.
  • ISO 13485 - quality management system requirements for IVD manufacturers.
  • ISO 15189:2022 - laboratory quality and competence requirements used throughout the workflow.
  • ISO/IEC 17020:2012 and ISO/IEC 17025:2017 - alternative competence and quality frameworks named for relevant laboratory functions.

The document also points to ISO 15190 for laboratory safety, ISO 8601-1 for date and time formatting in reports, and ISO 5649:2024 for concepts and specifications for laboratory-developed tests.

What does the oSIST prEN ISO 25379-1:2026 document contain?

The document contains the workflow structure for human DNA NGS examinations, from scope and terms to general requirements, technical development, reporting and quality assurance. It also includes specimen-specific pre-examination topics, library preparation steps, sequencing quality control, bioinformatics and reporting rules, and the routine-practice requirements for implementation.

It includes an informative workflow scheme in Annex B and a normative single-cell workflow in Annex A. The body of the standard also refers to tables for pre-examination documents and for performance characteristics, and to report content that includes result interpretation, limitations, report identification and review authorization.

Relations

Effective Date
19-Feb-2025

Buy Documents

Draft

oSIST prEN ISO 25379-1:2026

English language (60 pages)
Preview
Preview
e-Library read for
×1 day

Get Certified

Connect with accredited certification bodies for this standard

BSI Group

BSI (British Standards Institution) is the business standards company that helps organizations make excellence a habit.

UKAS United Kingdom Verified

TÜV Rheinland

TÜV Rheinland is a leading international provider of technical services.

DAKKS Germany Verified

TÜV SÜD

TÜV SÜD is a trusted partner of choice for safety, security and sustainability solutions.

DAKKS Germany Verified

Sponsored listings

Frequently Asked Questions

oSIST prEN ISO 25379-1:2026 is a draft published by the Slovenian Institute for Standardization (SIST). Its full title is "In vitro diagnostic Next Generation Sequencing (NGS) workflows - Part 1: Human DNA examination (ISO/DIS 25379-1:2026)". This standard covers: This document specifies requirements and gives recommendations for next generation sequencing (NGS) workflows. This document covers the pre-examination processes, human DNA (somatic and germline) isolation, sequencing library preparation, sequencing, sequence analysis and reporting of the examination of sequences for diagnostic purposes from isolated DNA from, e.g. formalin-fixed and paraffin embedded tissues, fresh frozen tissues, fine needle aspirates (FNA), whole blood, circulating tumour cells (CTCs), exosomes and other extracellular vesicles, circulating cell free DNA from plasma, and DNA from saliva. NOTE 1 Typical applications include, but are not limited to, NGS for oncology, pharmacogenomics and clinical genetics; approaches include panels (e.g. disease panels, exome panels, target gene panels and in silico panels), exome and whole genome sequencing, as well as certain epigenetics and certain single-cell analyses. This document is addressing in vitro diagnostic examinations including laboratory developed tests and is applicable to medical laboratories, molecular pathology laboratories and molecular genetic laboratories, in vitro diagnostic developers and manufacturers, biobanks as well as institutions and organizations performing biomedical research or sequencing services. This document is not applicable for in situ sequencing, DNA-mediated protein sequencing, forensic sequencing, sequencing of pathogens or microorganisms and microbiome analysis. NOTE 2 International, national or regional regulations or requirements or multiples of them can also apply to specific topics covered in this document.

This document specifies requirements and gives recommendations for next generation sequencing (NGS) workflows. This document covers the pre-examination processes, human DNA (somatic and germline) isolation, sequencing library preparation, sequencing, sequence analysis and reporting of the examination of sequences for diagnostic purposes from isolated DNA from, e.g. formalin-fixed and paraffin embedded tissues, fresh frozen tissues, fine needle aspirates (FNA), whole blood, circulating tumour cells (CTCs), exosomes and other extracellular vesicles, circulating cell free DNA from plasma, and DNA from saliva. NOTE 1 Typical applications include, but are not limited to, NGS for oncology, pharmacogenomics and clinical genetics; approaches include panels (e.g. disease panels, exome panels, target gene panels and in silico panels), exome and whole genome sequencing, as well as certain epigenetics and certain single-cell analyses. This document is addressing in vitro diagnostic examinations including laboratory developed tests and is applicable to medical laboratories, molecular pathology laboratories and molecular genetic laboratories, in vitro diagnostic developers and manufacturers, biobanks as well as institutions and organizations performing biomedical research or sequencing services. This document is not applicable for in situ sequencing, DNA-mediated protein sequencing, forensic sequencing, sequencing of pathogens or microorganisms and microbiome analysis. NOTE 2 International, national or regional regulations or requirements or multiples of them can also apply to specific topics covered in this document.

oSIST prEN ISO 25379-1:2026 is classified under the following ICS (International Classification for Standards) categories: 11.100.10 - In vitro diagnostic test systems. The ICS classification helps identify the subject area and facilitates finding related standards.

oSIST prEN ISO 25379-1:2026 has the following relationships with other standards: It is inter standard links to SIST-TS CEN/TS 17981-1:2024. Understanding these relationships helps ensure you are using the most current and applicable version of the standard.

oSIST prEN ISO 25379-1:2026 is available in PDF format for immediate download after purchase. The document can be added to your cart and obtained through the secure checkout process. Digital delivery ensures instant access to the complete standard document.

Standards Content (Sample)

Contents

  • Introduction
  • 1 Scope
  • 2 Normative references
  • 3 Terms and definitions
  • 3.1 aliquot
  • 3.2 amplicon
  • 3.3 analyte
  • 3.4 analytical accuracy
  • 3.5 analytical sensitivity
  • 3.6 analytical specificity
  • 3.7 bioinformatics pipeline
  • 3.8 biomarker
  • 3.9 circulating cell free DNA
  • 3.10 cell free DNA
  • 3.11 clinical accuracy
  • 3.12 clinical performance
  • 3.13 clinical sensitivity
  • 3.14 clinical specificity
  • 3.15 clinical utility
  • 3.16 clinical validity
  • 3.17 closed system
  • 3.18 Design of Experiments
  • 3.19 diagnosis
  • 3.20 deoxyribonucleic acid
  • 3.21 DNA sequencing
  • 3.22 epigenetics
  • 3.23 examination
  • 3.24 examination manufacturer
  • 3.25 examination performance
  • 3.26 extracellular vesicle
  • 3.27 fine needle aspirate
  • 3.28 formalin
  • 3.29 guard band
  • 3.30 in vitro diagnostic NGS workflow
  • 3.31 intended use
  • 3.32 intermediate precision
  • 3.33 intermediate precision condition of measurement
  • 3.34 laboratory developed examination
  • 3.35 limit of detection
  • 3.36 limit of detection for platforms
  • 3.37 long-read sequencing
  • 3.38 massively parallel sequencing
  • 3.39 measurand
  • 3.40 measurement
  • 3.41 measurement procedure
  • 3.42 methylome
  • 3.43 next generation sequencing
  • 3.44 no template control
  • 3.45 panel
  • 3.46 performance characteristic
  • 3.47 performance specification
  • 3.48 pharmacogenomics
  • 3.49 post-examination processes
  • 3.50 pre-examination processes
  • 3.51 proficiency testing
  • 3.52 protein
  • 3.53 reference material
  • 3.54 reference method
  • 3.55 reference range
  • 3.56 repeatability
  • 3.57 repeatability condition of measurement
  • 3.58 reportable range
  • 3.59 reproducibility
  • 3.60 reproducibility condition of measurement
  • 3.61 ribonucleic acid
  • 3.62 sample
  • 3.63 scientific validity
  • 3.64 sequencing library
  • 3.65 sequencing library preparation
  • 3.66 short-read sequencing
  • 3.67 specimen
  • 3.68 validation
  • 3.69 venous whole blood
  • 3.70 verification
  • 3.71 workflow
  • 4 General requirements
  • 4.1 General
  • 4.2 Examination design
  • 4.3 Examination development
  • 4.4 Examination performance verification and validation

Introduction

Molecular in vitro diagnostics has enabled significant progress in medicine. Further progress is expected by new technologies analysing profiles of nucleic acids, proteins, and metabolites in human tissues and body fluids. Next Generation Sequencing (NGS) takes a prominent place in the series of molecular techniques used for diagnostics. It facilitates sequence analysis of nucleic acids that can result in precise information for diagnosis and progression of diseases.

The NGS technique, however, has a very complex workflow that contains many steps. The target nucleic acids can originate from different sources (e.g. tissues, blood, and body fluids). The profiles of the isolated DNA or methylated DNA can change during specimen collection, transport, storage and processing (e.g. formalin fixation), making the outcome from diagnostics or research unreliable or even impossible because the subsequent analytical assay will not determine the situation in the patient, but an artificial profile generated during the pre-examination process. The available material can be small, the cells in a tissue can be dispersed heterogeneously (e.g. ratio of tumour to normal), the target nucleic acids can be circulating in blood or body fluids free of cells or in circulating cells (e.g. circulating tumour cells (CTCs)). For a successful and reliable sequencing result, a suitable strategy needs to be chosen for every case depending on the available material and disease conditions. Therefore, the NGS workflow can differ from case to case, and the NGS workflow steps need to be carefully considered and chosen to get a sound and reliable result to determine the best available treatment for the patient. In addition, sequencing platforms can differ in their technique (e.g. detection of a change in a current or fluorescence) and approach (e.g. panels, short-read sequencing, longread sequencing) for sequence assessment. The bioinformatics analysis can differ in approach and ability to detect non-conformities and unreliable sequencing results. To enable such capabilities, NGS metadata needs to be collected during all workflow steps from the patient to the reporting. In addition, controls and added controls need to be analysed properly. In this way, non-conformities or detected unreliabilities can be reported to the patient and the treating physician. The reporting of diagnostic NGS results can differ in clarity and depth, which can lead to different interpretations. Standardization of the entire NGS workflow from specimen collection to the reporting of the results to the patient and the treating physician is needed for the development of reliable NGS examinations. This document draws upon previous work to standardize the steps for NGS examinations from tissues, blood and body fluids in what is referred to as the pre-examination phase (sample collection), the examination phase (library preparation, sequencing), and the post-examination phase (analysis and reporting). In this document, the following verbal forms are used:

  • “shall” indicates a requirement;
  • “should” indicates a recommendation;
  • “may” indicates a permission;
  • “can” indicates a possibility or a capability.

oSIST prEN ISO 25379-1:2026

oSIST prEN ISO 25379-1:2026 DRAFT International Standard ISO/DIS 25379-1:2026(en) In vitro diagnostic Next Generation Sequencing (NGS) workflows — Part 1:

Human DNA examination

1 Scope

This document specifies requirements and gives recommendations for next generation sequencing (NGS) workflows for in vitro diagnostics and biomedical research. This document covers the pre-examination processes including human DNA (somatic and germline) isolation, sequencing library preparation, sequencing, sequence analysis and reporting of the examined sequences for diagnostic purposes from isolated DNA from, e.g. formalin-fixed and paraffin embedded tissues, fresh frozen tissues, fine needle aspirates (FNA), whole blood, circulating tumour cells (CTCs), exosomes and other extracellular vesicles, circulating cell free DNA from plasma, and DNA from saliva.

NOTE 1 Typical applications include, but are not limited to, NGS for oncology, pharmacogenomics and clinical genetics; approaches include panels (e.g. disease panels, exome panels, target gene panels and in silico panels), exome and whole genome sequencing, as well as certain epigenetic and certain single-cell analyses. This document addresses in vitro diagnostic examinations including laboratory developed examinations and is applicable to medical laboratories, molecular pathology laboratories and molecular genetic laboratories, in vitro diagnostic developers and manufacturers, biobanks and institutions and organizations performing biomedical research or sequencing services.

This document is not applicable for in situ sequencing, DNA-mediated protein sequencing, forensic sequencing, sequencing of pathogens or microorganisms and microbiome analysis. NOTE 2 International, national or regional regulations or requirements or multiples of them can also apply to specific topics covered in this document.

2 Normative references

The following documents are referred to in the text in such a way that some or all of their content constitutes requirements of this document. For dated references, only the edition cited applies. For undated references, the latest edition of the referenced document (including any amendments) applies. ISO 7552-2, Molecular in vitro diagnostic examinations — Specifications for pre-examination processes for circulating tumor cells (CTCs) in venous whole blood — Part 2: Isolated DNA ISO 18702, Molecular in vitro diagnostic examinations — Specifications for pre-examination processes for exosomes and other extracellular vesicles in venous whole blood — DNA, RNA and proteins ISO 18704, Molecular in vitro diagnostic examinations — Requirements and recommendations for preexamination processes for urine and other body fluids — Isolated cell-free DNA ISO 4307, Molecular in vitro diagnostic examinations — Specifications for pre-examination processes for saliva - Isolated human DNA ISO 13485, Medical devices — Quality management systems — Requirements for regulatory purposes ISO 15189:2022, Medical laboratories — Requirements for quality and competence

oSIST prEN ISO 25379-1:2026 ISO/DIS 25379-1:2026(en) ISO/IEC 17020:2012, Conformity assessment — Requirements for the operation of various types of bodies performing inspection ISO/IEC 17025:2017, General requirements for the competence of testing and calibration laboratories ISO 20166-3, Molecular in vitro diagnostic examinations — Specifications for pre-examination processes for formalin-fixed and paraffin-embedded (FFPE) tissue — Part 3: Isolated DNA ISO 20184-3, Molecular in vitro diagnostic examinations — Specifications for pre-examination processes for frozen tissue — Part 3: Isolated DNA ISO 20186-2, Molecular in vitro diagnostic examinations — Specifications for pre-examination processes for venous whole blood — Part 2: Isolated genomic DNA ISO 20186-3, Molecular in vitro diagnostic examinations — Specifications for pre-examination processes for venous whole blood — Part 3: Isolated circulating cell free DNA from plasma ISO 20397-1:2022, Biotechnology — Massively parallel sequencing — Part 1: Nucleic acid and library preparation ISO 20397-2, Biotechnology — Massively parallel sequencing — Part 2: Quality evaluation of sequencing data

3 Terms and definitions

For the purposes of this document, the terms and definitions given in ISO 15189 and the following apply. ISO and IEC maintain terminology databases for use in standardization at the following addresses:

  • ISO Online browsing platform: available at https:// www .iso .org/ obp
  • IEC Electropedia: available at https:// www .electropedia .org/

3.1 aliquot

portion of a larger amount of homogenous material, assumed to be taken with negligible sampling error Note 1 to entry: The term is usually applied to fluids. Tissues are heterogeneous and therefore cannot be aliquoted. [1][2][3] Note 2 to entry: The definition is derived from .

[SOURCE: ISO 20184-3:2021, 3.1, modified ─ Note 2 to entry was added.]

3.2 amplicon

specific DNA fragment produced by a DNA-amplification technology, such as the polymerase chain reaction (PCR) [SOURCE: ISO 13495:2013, 3.3.1]

3.3 analyte

component represented in the name of a measurable quantity Note 1 to entry: The term “analyte”, or the name of a substance or compound, is not to be confused with the term “measurand”, because analytes are not quantities.

EXAMPLE Amount of substance of glucose in plasma is a measurand (3.39), while glucose is the analyte. [SOURCE: ISO 17511:2021, 3.1, modified — Note 1 and example were added]

oSIST prEN ISO 25379-1:2026 ISO/DIS 25379-1:2026(en)

3.4 analytical accuracy

closeness of the agreement between the result of an examination (3.23) and a true property value Note 1 to entry: For NGS-based examinations, accuracy represents the degree of concordance (or agreement) of results between a sequence obtained from the examination and the same sequence determined by a valid comparator method, or between a reference sample run on an NGS-based examination and the high confidence sequence of the reference. [SOURCE: ISO/IEC Guide 99:2007, 2.13, modified – Original notes were deleted, Note 1 to entry was added; "analytical accuracy" replaces all old terms, "a measured quantity value" was replaced by "the result of an examination" and "quantity value of a measurand" was replaced by "property value".]

3.5 analytical sensitivity

sensitivity of a measurement procedure quotient of the change in a measurement (3.40) indication and the corresponding change in a value of a quantity being measured Note 1 to entry: The sensitivity of a measurement procedure (3.41) can depend on the value of the quantity being measured.

Note 2 to entry: The change considered in the value of the quantity being measured only becomes useful when large compared with the resolution.

Note 3 to entry: The analytical sensitivity of a measuring system is the slope of the calibration curve. Note 4 to entry: Analytical sensitivity is often confused with positive percentage agreement (PPA), because similar calculations are used for both the analytical sensitivity and the PPA. However, the term PPA instead of analytical sensitivity is only used, if a reference method is not available or not used. The PPA represents merely an estimation of the analytical sensitivity. PPA is applicable only to measurements of quantities that have the nature of a count. [SOURCE: ISO/IEC Guide 99:2007, 4.12, modified — terms changed from "sensitivity of a measuring system /sensitivity" changed to "analytical sensitivity/sensitivity of a measurement procedure"; "an indication of a measuring system" replaced by "a measurement indication", "measuring system" replaced by "measurement procedure" in Note 1; new Note 3 and Note 4 were added.]

3.6 analytical specificity

selectivity of a measurement procedure capability of a measuring system, using a specified measurement procedure (3.41), to provide measurement (3.40) results for one or more measurands (3.39) which do not depend on each other nor on any other quantity in the system undergoing measurement Note 1 to entry: Lack of analytical specificity is called analytical interference (see ISO 18113-1:2022, 3.2.2). Note 2 to entry: Specificity of a measurement procedure (3.41) should not be confused with diagnostic specificity (3.14) (see ISO 18113-1:2022, 3.2.18).

Note 3 to entry: ISO/IEC Guide 99:2007 uses the term selectivity for this concept instead of specificity. Note 4 to entry: Adapted from ISO/IEC Guide 99:2007, 4.13.

Note 5 to entry: Selectivity as used in metrology is a concept close to specificity as it is sometimes used in chemistry. “Specificity” of measurements has some risk of ambiguity with “clinical/diagnostic specificity”. Note 6 to entry: Analytical specificity is often confused with negative percentage agreement (NPA), because the same calculations are used for both the analytical specificity and the NPA. However, the term NPA instead of analytical specificity is only used, if a reference method (3.54) is not available or is not used. The NPA represents merely an estimation of the analytical specificity.

[SOURCE: ISO 18113-1:2022, 3.2.5, modified — Note 2 was deleted and Note 5 and 67 were added.]

oSIST prEN ISO 25379-1:2026 ISO/DIS 25379-1:2026(en)

3.7 bioinformatics pipeline

suite of different bioinformatics tools that process the next generation sequencing (3.43) data

3.8 biomarker

molecular marker detectable and/or quantifiable molecule or group of molecules used to indicate a biological condition, state, identity or characteristic or an organism [SOURCE: ISO 16577:2022, 3.4.28]

3.9 circulating cell free DNA

ccfDNA extracellular human DNA present in blood and plasma Note 1 to entry: ccfDNA includes DNA present in vesicles such as exosomes. [SOURCE: EN ISO 20186-3:2019, 3.5]

3.10 cell free DNA

cfDNA extracellular human DNA present in body liquids such as urine and plasma Note 1 to entry: cfDNA can include DNA present in vesicles such as exosomes. [SOURCE: ISO/DIS 18704, 3.8]

3.11 clinical accuracy

diagnostic accuracy extent of agreement between the outcome of the examination (3.23) and the target clinical condition

3.12 clinical performance

ability of an examination (3.23) to yield results that are correlated with a particular clinical condition, physiological or pathological state in accordance with the target population and intended use(r) Note 1 to entry: Although sometimes referred to as diagnostic performance or clinical validity (3.16); clinical performance is the harmonized term endorsed by the Global Harmonization Task Force (GHTF) and its successor, the International Medical Devices Regulators Forum (IMDRF).

Note 2 to entry: Evaluation of clinical performance often relies on the outcome of other types of clinical examinations (3.23) to define “true positive or true negative” results.

3.13 clinical sensitivity

diagnostic sensitivity ability of an in vitro diagnostic examination (3.23) procedure to have positive results associated with a particular disease or condition Note 1 to entry: Also defined as percent positivity in samples where the target marker is known to be present. Note 2 to entry: Diagnostic sensitivity is expressed as a percentage (number fraction multiplied by 100), calculated as 100 × the number of true positive values (TP) divided by the sum of the number of true positive values (TP) plus the number of false negative values (FP), or 100 × TP/(TP + FN). This calculation is based on a study design where only one sample is taken from each subject.

Note 3 to entry: The target condition is defined by criteria independent of the examination procedure under consideration.

oSIST prEN ISO 25379-1:2026 ISO/DIS 25379-1:2026(en) Note 4 to entry: Clinical sensitivity is often confused with positive percentage agreement (PPA), because the same calculations are used for both the clinical sensitivity and the PPA. However, the term PPA instead of clinical sensitivity is only used, if a reference method is not available or is not used. The PPA represents merely an estimation of the clinical sensitivity.

[SOURCE: ISO 18113-1:2022, 3.2.17, modified — The term "clinical sensitivity" was added; second sentence in Note 1 was deleted; Note 4 was added.]

3.14 clinical specificity

diagnostic specificity ability of an in vitro diagnostic examination (3.23) procedure to have negative results associated with an absence of particular disease or condition Note 1 to entry: Also defined as percent negativity in samples where the target marker is known to be absent. For information regarding description of the diagnostic performance characteristics of an IVD medical device, see Reference [4].

Note 2 to entry: Clinical specificity is expressed as a percentage (number fraction multiplied by 100), calculated as 100 × the number of true negative values (TN) divided by the sum of the number of true negative plus the number of false positive values (FP), or 100 × TN/(TN + FP). This calculation is based on a study design where only one sample is taken from each subject.

Note 3 to entry: The target condition is defined by criteria independent of the examination (3.23) procedure under consideration.

Note 4 to entry: Clinical specificity is often confused with negative percentage agreement (NPA), because the same calculations are used for both the clinical specificity and the NPA. However, the term NPA instead of clinical specificity is only used, if a reference method (3.54) is not available or is not used. The NPA represents merely an estimation of the clinical specificity.

[SOURCE: ISO 18113-1:2022, 3.2.18, modified — The term "clinical specificity" was added; Note 4 was added.]

3.15 clinical utility

ability of a screening or diagnostic examination (3.23) to prevent or ameliorate adverse health outcomes such as mortality, morbidity, or disability through the adoption of efficacious treatments conditioned on examination results [5][6] Note 1 to entry: Clinical utility can be part of scientific validity and clinical performance (3.12) . [6] [SOURCE: ]

3.16 clinical validity

predictive value of an examination (3.23) for a given clinical outcome Note 1 to entry: Clinical validity is primarily determined by the clinical sensitivity (3.13) and clinical specificity (3.14) with which an examination identifies individuals with a defined clinical condition within a given population. The clinical validity (3.16) of a genetic examination is the likelihood that, e.g. cancer will develop in someone with a positive examination result.

3.17 closed system

non-modifiable system (by the user) provided by the vendor including all necessary components for the examination (3.23) Note 1 to entry: Necessary components are hardware, software, procedures and reagents.

oSIST prEN ISO 25379-1:2026 ISO/DIS 25379-1:2026(en)

3.18 Design of Experiments

DoE <NGS> structured, statistical approach for planning sequencing runs and experimental conditions in order to systematically evaluate the effects of key variables (e.g. library preparation methods, input nucleic acid quantity, sequencing depth, and indexing strategies) on sequencing quality, coverage, and downstream data interpretation

3.19 diagnosis

identification of a disease from its signs and symptoms, where the diagnostic process can involve examinations (3.23) and tests for classification of an individual's condition into separate and distinct categories or subclasses that allow medical decisions about treatment and prognosis to be made

3.20 deoxyribonucleic acid

DNA polymer of deoxyribonucleotides occurring in a double-stranded (dsDNA) or single-stranded (ssDNA) form

3.21 DNA sequencing

determining the order of nucleotide bases (e.g. adenine, guanine, cytosine and thymine) in a molecule of DNA (3.20) Note 1 to entry: Sequence is generally described from the 5’ end. [SOURCE: ISO 17822:2020, 3.19, modified — Added e.g.]

3.22 epigenetics

study of changes of genome usage without alterations in the underlying DNA sequence [7] [SOURCE: ]

3.23 examination

analytical test set of operations having the objective of determining the numerical value or characteristics of a property Note 1 to entry: An examination can be the total of a number of activities, observations or measurements required to determine a value or characteristic.

...