Audiometers - Part 4: Equipment for extended high-frequency audiometry

Specifies requirements for audiometric equipment designed for use in pure tone audiometry in the frequency range from 8 000 Hz to 16 000 Hz, in addition to those that are applicable and specified in EN 60645-1

Audiometer - Teil 4: Geräte für die Audiometrie in einem erweiterten Hochtonbereich

Audiomètres - Partie 4: Equipement pour l'audiométrie étendue au domaine des fréquences élevées

Donne des spécifications concernant les équipments audiométriques conçus pour une utilisation en audiométrie tonale dans un domaine de fréquences compris entre 8 000 Hz et 16 000 Hz, en plus des prescriptions qui sont spécifiées dans la EN 60645-1 et qui sont également applicables.

Avdiometri - 4. del: Oprema za razširjeno visokofrekvenčno avdiometrijo (IEC 60645-4:1994)

General Information

Status
Withdrawn
Publication Date
30-Jun-1999
Withdrawal Date
15-Oct-2017
Technical Committee
Current Stage
9900 - Withdrawal (Adopted Project)
Start Date
16-Oct-2017
Due Date
08-Nov-2017
Completion Date
16-Oct-2017

Relations

Buy Standard

Standard
EN 60645-4:1999
English language
13 pages
sale 10% off
Preview
sale 10% off
Preview
e-Library read for
1 day

Standards Content (Sample)

SLOVENSKI STANDARD
SIST EN 60645-4:1999
01-julij-1999
$YGLRPHWULGHO2SUHPD]DUD]ãLUMHQRYLVRNRIUHNYHQþQRDYGLRPHWULMR ,(&

Audiometers - Part 4: Equipment for extended high-frequency audiometry
Audiometer - Teil 4: Geräte für die Audiometrie in einem erweiterten Hochtonbereich
Audiomètres - Partie 4: Equipement pour l'audiométrie étendue au domaine des
fréquences élevées
Ta slovenski standard je istoveten z: EN 60645-4:1995
ICS:
17.140.50 Elektroakustika Electroacoustics
SIST EN 60645-4:1999 en
2003-01.Slovenski inštitut za standardizacijo. Razmnoževanje celote ali delov tega standarda ni dovoljeno.

---------------------- Page: 1 ----------------------

SIST EN 60645-4:1999

---------------------- Page: 2 ----------------------

SIST EN 60645-4:1999

---------------------- Page: 3 ----------------------

SIST EN 60645-4:1999

---------------------- Page: 4 ----------------------

SIST EN 60645-4:1999

---------------------- Page: 5 ----------------------

SIST EN 60645-4:1999

---------------------- Page: 6 ----------------------

SIST EN 60645-4:1999
NORME
CEI
INTERNATIONALE IEC
60645-4
INTERNATIONAL
Première édition
STAN DARD
First edition
1994-09
Audiomètres
Partie 4:
Equipement pour l'audiométrie étendue au
domaine des fréquences élevées
Audiometers
Part 4:
Equipment for extended high-frequency
audiometry
© IEC 1994 Droits
de reproduction réservés — Copyright - all rights reserved
Aucune partie de cette publication ne peut être reproduite ni No part of this publication may be reproduced or utilized in
utilisée sous quelque forme que ce soit et par aucun any form or by any means, electronic or mechanical,
procédé, électronique ou mécanique, y compris la photo- including photocopying and microfilm, without permission in
copie et les microfilms, sans l'accord écrit de l'éditeur. writing from the publisher.

International Electrotechnical Commission
3, rue de Varembé Geneva, Switzerland

Telefax: +41 22 919 0300 e-mail: inmail@iec.ch IEC web site http: //www.iec.ch
CODE PRIX
Commission Electrotechnique Internationale
K
PRICE CODE
International Electrotechnical Commission
IEC
MermvHaponHan 3newrpoTexHH4ecKaa
HOMHCCHA
Pour prix, voir catalogue en vigueur
• • For price, see current catalogue

---------------------- Page: 7 ----------------------

SIST EN 60645-4:1999
645-4 © IEC:1994 — 3 —
CONTENTS
Page
FOREWORD 5
INTRODUCTION 7
Clause
1 Scope 9
2 Normative references 9
3 Explanation of terms 9
11
4 Facilities on extended high-frequency audiometers
11
5 General requirements
13
6 Test signals
13 6.1 Test signal frequencies
13
6.2 Frequency accuracy
13
6.3 Test signal level range for earphones
13
6.4 Masking sounds
15
7 Signal level control
Marking 15
7.1
15
7.2 Accuracy
17
7.3 Tone switching
Masking level 17
7.4
8 Unwanted sound 17
9 Transducers 17
Earphones
9.1 17
17
9.2 Loudspeaker
19 9.3 Bone vibrator
10 Audiogram format 19
19
10.1 Extended high-frequency audiogram
19
10.2 Combined audiogram

---------------------- Page: 8 ----------------------

SIST EN 60645-4:1999
645-4 © IEC:1994 — 5 —
INTERNATIONAL ELECTROTECHNICAL COMMISSION
AUDIOMETERS
Part 4: Equipment for extended high-frequency audiometry
FOREWORD
1) The IEC (International Electrotechnical Commission) is a worldwide organization for standardization
comprising all national electrotechnical committees (IEC National Committees). The object of the IEC is to
promote international cooperation on all questions concerning standardization in the electrical and
electronic fields. To this end and in addition to other activities, the IEC publishes International Standards.
Their preparation is entrusted to technical committees; any IEC National Committee interested in the
subject dealt with may participate in this preparatory work. International, governmental and
non-governmental organizations liaising with the IEC also participate in this preparation. The IEC
collaborates closely with the International Organization for Standardization (ISO) in accordance with
conditions determined by agreement between the two organizations.
2)
The formal decisions or agreements of the IEC on technical matters, prepared by technical committees on
which all the National Committees having a special interest therein are represented, express, as nearly as
possible, an international consensus of opinion on the subjects dealt with.
3)
They have the form of recommendations for international use published in the form of standards, technical
reports or guides and they are accepted by the National Committees in that sense.
4)
In order to promote international unification, IEC National Committees undertake to apply IEC International
Standards transparently to the maximum extent possible in their national and regional standards. Any
divergence between the IEC Standard and the corresponding national or regional standard shall be clearly
indicated in the latter.
This part of International Standard IEC 645 has been prepared by IEC technical committee 29:
Electroacoustics.
It forms the first edition of IEC 645-4.
The text of this standard is based on the following documents:
DIS Report on voting
29(CO)215
29(CO)220
Full information on the voting for the approval of this standard can be found in the report
on voting indicated in the above table.

---------------------- Page: 9 ----------------------

SIST EN 60645-4:1999
645-4 © IEC:1994 — 7 —
INTRODUCTION
This International Standard describes equipment which is designed for the measurement
of hearing in the frequerrcy range from 8 000 Hz to 16 000 Hz. Part 1 of IEC 645 specifies
the requirements for pure tone audiometers with frequencies within the conventional range
for audiometry from 125 Hz to 8 000 Hz. Many of the requirements are common for both
frequency ranges and therefore this standard includes only such requirements that are
specific to the high-frequency range.

---------------------- Page: 10 ----------------------

SIST EN 60645-4:1999
645-4 © IEC:1994 - 9 -
AUDIOMETERS
Part 4: Equipment for extended high-frequency audiometry
1 Scope
This International Standard specifies requirements for audiometric equipment designed
for use in pure tone audiometry in the frequency range from 8 000 Hz to 16 000 Hz, in
addition to those that are applicable and specified in IEC 645-1.
The purpose of the standard is to ensure that extended high-frequency audiometry
performed on a given human ear with different types of equipment which comply with this
standard shall give substantially similar results.
2 Normative references
The following normative documents contain provisions which, through reference in this
text, constitute provisions of this part of IEC 645. At the time of publication, the editions
indicated were valid. All normative documents are subject to revision, and pa rties to
agreements based on this part of IEC 645 are encouraged to investigate the possibility of
applying the most recent editions of the normative documents indicated below. Members
of IEC and ISO maintain registers of currently valid International Standards.
IEC 645-1: 1992,
Audiometers - Part 1: Pure tone audiometers
ISO 389-7: Acoustics - Reference zero for the calibration of audiometric equipment -
Part 7: Reference threshold of hearing under free-field and diffuse-field listening
conditions (Revision of ISO 226:1987)*
ISO 8253-1: 1991, Acoustics - Audiometric test methods - Part 1: Basic pure tone air
and bone conduction threshold audiometry
ISO 266: 1975, Acoustics - Preferred frequencies for measurements
ISO 8253-2: 1992,
Acoustics - Audiometric test methods - Part 2: Sound Field Audio-
metry with pure tone and narrow-band test signals
3 Explanation of terms
For the purpose of this standard the following terms apply:
3.1 extended high-frequency: An audiometric test frequency in the range from 8 000 Hz
to 16 000 Hz.
NOTE – The frequency 8 000 Hz is considered both as the highest frequency in the conventional
...

Questions, Comments and Discussion