Medical electrical equipment - Part 2-24: Particular requirements for the safety of infusion pumps and controllers (IEC 60601-2-24:1998)

Specifies the requirements for infusion pumps, infusion controllers, syringe pumps and pumps for ambulatory use. These devices are intended for use by medical staff and home patients as prescribed and medically indicated. It is the responsibility of the manufacturer to ensure that the requirements of this standard are reliably implemented.

Medizinische elektrische Geräte - Teil 2-24: Besondere Festlegungen für die Sicherheit von Infusionspumpen und Infusionsreglern (IEC 60601-2-24:1998)

Appareils électromédicaux - Partie 2-24: Règles particulières de sécurité des pompes et régulateurs de perfusion (CEI 60601-2-24:1998)

Medicinska električna oprema - 2-24. del: Posebne varnostne zahteve za infuzijske črpalke in krmilnike (IEC 60601-2-24:1998)

General Information

Status
Withdrawn
Publication Date
31-Aug-1998
Withdrawal Date
09-May-2018
Current Stage
9900 - Withdrawal (Adopted Project)
Start Date
10-May-2018
Due Date
02-Jun-2018
Completion Date
10-May-2018

Relations

Buy Standard

Standard
EN 60601-2-24:1998
English language
67 pages
sale 10% off
Preview
sale 10% off
Preview
e-Library read for
1 day

Standards Content (Sample)

SLOVENSKI STANDARD
SIST EN 60601-2-24:1998
01-september-1998
0HGLFLQVNDHOHNWULþQDRSUHPDGHO3RVHEQHYDUQRVWQH]DKWHYH]DLQIX]LMVNH
þUSDONHLQNUPLOQLNH ,(&
Medical electrical equipment - Part 2-24: Particular requirements for the safety of infusion
pumps and controllers (IEC 60601-2-24:1998)
Medizinische elektrische Geräte - Teil 2-24: Besondere Festlegungen für die Sicherheit
von Infusionspumpen und Infusionsreglern (IEC 60601-2-24:1998)
Appareils électromédicaux - Partie 2-24: Règles particulières de sécurité des pompes et
régulateurs de perfusion (CEI 60601-2-24:1998)
Ta slovenski standard je istoveten z: EN 60601-2-24:1998
ICS:
11.040.20 Transfuzijska, infuzijska in Transfusion, infusion and
injekcijska oprema injection equipment
SIST EN 60601-2-24:1998 en
2003-01.Slovenski inštitut za standardizacijo. Razmnoževanje celote ali delov tega standarda ni dovoljeno.

---------------------- Page: 1 ----------------------

SIST EN 60601-2-24:1998

---------------------- Page: 2 ----------------------

SIST EN 60601-2-24:1998

---------------------- Page: 3 ----------------------

SIST EN 60601-2-24:1998

---------------------- Page: 4 ----------------------

SIST EN 60601-2-24:1998

---------------------- Page: 5 ----------------------

SIST EN 60601-2-24:1998

---------------------- Page: 6 ----------------------

SIST EN 60601-2-24:1998
INTERNATIONAL IEC
STANDARD 60601-2-24
First edition
1998-02
Medical electrical equipment –
Part 2-24:
Particular requirements for the safety of infusion
pumps and controllers
Appareils électromédicaux –
Partie 2-24:
Règles particulières de sécurité des pompes et régulateurs
de perfusion
 IEC 1998 Copyright - all rights reserved  Droits de reproduction réservés
Aucune partie de cette publication ne peut être reproduite ni No part of this publication may be reproduced or utilized in
utilisée sous quelque forme que ce soit et par aucun any form or by any means, electronic or mechanical,
procédé, électronique ou mécanique, y compris la photo- including photocopying and microfilm, without permission in
copie et les microfilms, sans l'accord écrit de l'éditeur. writing from the publisher.
International Electrotechnical Commission 3, rue de Varembé Geneva, Switzerland
Telefax: +41 22 919 0300 e-mail: inmail@iec.ch IEC web site http: //www.iec.ch
Commission Electrotechnique Internationale PRICE CODE
XB
International Electrotechnical Commission
For price, see current catalogue

---------------------- Page: 7 ----------------------

SIST EN 60601-2-24:1998
– 2 – 60601-2-24 © IEC:1998(E)
CONTENTS
Page
FOREWORD . 4
INTRODUCTION . 5
Clause
SECTION ONE – GENERAL
1 Scope and object . 6
2 Terminology and definitions . 7
3 General requirements . 10
5 Classification . 11
6 Identification, marking and documents . 11
SECTION TWO – ENVIRONMENTAL CONDITIONS
10 Environmental conditions . 13
SECTION THREE – PROTECTION AGAINST ELECTRIC SHOCK HAZARDS
14 Requirements related to classification . 14
17 Separation. 14
19 Continuous LEAKAGE CURRENTS and PATIENT AUXILIARY CURRENTS. 14
SECTION FOUR – PROTECTION AGAINST MECHANICAL HAZARDS
21 Mechanical strength. 16
SECTION FIVE – PROTECTION AGAINST HAZARDS FROM UNWANTED
OR EXCESSIVE RADIATION
36 Electromagnetic compatibility. 17
SECTION SIX – PROTECTION AGAINST HAZARDS OF IGNITION
OF FLAMMABLE ANAESTHETIC MIXTURES
SECTION SEVEN – PROTECTION AGAINST EXCESSIVE TEMPERATURES
AND OTHER SAFETY HAZARDS
44 Overflow, spillage, leakage, humidity, ingress of liquids, cleaning, sterilization,
disinfection and compatibility . 19
47 Electrostatic charges . 20
49 Interruption of the power supply . 20

---------------------- Page: 8 ----------------------

SIST EN 60601-2-24:1998
60601-2-24 © IEC:1998(E) – 3 –
Clause Page
SECTION EIGHT – ACCURACY OF OPERATING DATA
AND PROTECTION AGAINST HAZARDOUS OUTPUT
50 Accuracy of operating data. 21
51 Protection against hazardous output . 38
SECTION NINE – ABNORMAL OPERATION AND FAULT CONDITIONS:
ENVIRONMENTAL TESTS
SECTION TEN – CONSTRUCTIONAL REQUIREMENTS
54 General . 42
56 Components and general assembly. 44
Annexes
L References – Publications mentioned in this standard . 45
AA General guidance and rationale. 47

---------------------- Page: 9 ----------------------

SIST EN 60601-2-24:1998
– 4 – 60601-2-24 © IEC:1998(E)
INTERNATIONAL ELECTROTECHNICAL COMMISSION
___________

MEDICAL ELECTRICAL EQUIPMENT
Part 2-24: Particular requirements for the safety
of infusion pumps and controllers
FOREWORD
1) The IEC (International Electrotechnical Commission) is a worldwide organization for standardization comprising
all national electrotechnical committees (IEC National Committees). The object of the IEC is to promote inter-
national co-operation on all questions concerning standardization in the electrical and electronic fields. To this
end and in addition to other activities, the IEC publishes International Standards. Their preparation is entrusted
to technical committees; any IEC National Committee interested in the subject dealt with may participate in this
preparatory work. International, governmental and non-governmental organizations liaising with the IEC also
participate in this preparation. The IEC collaborates closely with the International Organization for Standardi-
zation (ISO) in accordance with conditions determined by agreement between the two organizations.
2) The formal decisions or agreements of the IEC on technical matters express, as nearly as possible, an inter-
national consensus of opinion on the relevant subjects since each technical committee has representation from
all interested National Committees.
3) The documents produced have the form of recommendations for international use and are published in the form
of standards, technical reports or guides and they are accepted by the National Committees in that sense.
4) In order to promote international unification, IEC National Committees undertake to apply IEC International
Standards transparently to the maximum extent possible in their national and regional standards. Any
divergence between the IEC Standard and the corresponding national or regional standard shall be clearly
indicated in the latter.
5) The IEC provides no marking procedure to indicate its approval and cannot be rendered responsible for any
equipment declared to be in conformity with one of its standards.
6) Attention is drawn to the possibility that some of the elements of this International Standard may be the subject
of patent rights. The IEC shall not be held responsible for identifying any or all such patent rights.
International Standard IEC 60601-2-24 has been prepared by subcommittee 62D: Electro-
medical equipment, of IEC technical committee 62: Electrical equipment in medical practice.
The text of this Particular Standard is based on the following documents:
FDIS Report on voting
62D/250/FDIS 62D/268/RVD
Full information on the voting for the approval of this Particular Standard can be found in the
report on voting indicated in the above table.
Annex L is an integral part of this standard.
Annex AA is for information only.
In this Particular Standard the following print types are used:
– requirements, compliance with which can be tested, and definitions: in roman type;
– notes, explanations, advice, introductions, general statements, exceptions and references: in
smaller type;
– test specifications: in italic type;
– TERMS DEFINED IN CLAUSE 2 OF THE GENERAL STANDARD IEC 60601-1 OR THIS PARTICULAR
STANDARD: SMALL CAPITALS.
A bilingual version of this standard may be issued at a later date.

---------------------- Page: 10 ----------------------

SIST EN 60601-2-24:1998
60601-2-24 © IEC:1998(E) – 5 –
INTRODUCTION
This Particular Standard deals with the safety of INFUSION PUMPS and CONTROLLERS. The
relationship between this Particular Standard, IEC 60601-1 (including amendments 1 and 2),
and the Collateral Standards is explained in 1.3.
The safe use of infusion pumps and controllers is primarily the responsibility of the OPERATOR.
It is also recognized that OPERATORS should be trained in the operation of MEDICAL ELECTRICAL
EQUIPMENT and that safe use of the EQUIPMENT can only be achieved if it is operated in
accordance with the manufacturer’s instructions for use. The minimum specified safety
requirements are considered to provide a practical degree of safety in operation. It is the
responsibility of the manufacturer to ensure that the requirements of this Particular Standard
are reliably implemented. This Particular Standard has been developed in accordance with
these principles.
Safe use can be ensured only if the associated disposable parts, especially lines and syringes
are consistent with the system. ISO 7886-2:1996, Sterile hypodermic syringes for single use –
Part 2: Syringes for use with power-driven syringe pumps should be taken into account.

---------------------- Page: 11 ----------------------

SIST EN 60601-2-24:1998
– 6 – 60601-2-24 © IEC:1998(E)
MEDICAL ELECTRICAL EQUIPMENT –
Part 2-24: Particular requirements for the safety
of infusion pumps and controllers
SECTION ONE – GENERAL
The clauses and subclauses of this section of the General Standard and of this section of the
Collateral Standard IEC 60601-1-2 apply, except as follows:
1 Scope and object
This clause of the General Standard and this clause of the Collateral Standard IEC 60601-1-2
apply, except as follows:
1.1* Scope
Addition:
This Particular Standard specifies the requirement for INFUSION PUMPS, INFUSION CONTROLLERS,
SYRINGE PUMPS and PUMPS FOR AMBULATORY USE, as defined in 2.101 to 2.110. These devices
are intended for use by medical staff and home PATIENTS as prescribed and medically
indicated. These particular requirements do not apply to devices:
1) specifically intended for diagnostic or similar use (e.g. angiography or other pumps
permanently controlled or supervised by the OPERATOR),
2) enteral infusion,
3) extracorporeal circulation of blood,
4) implantable or disposable devices,
5) EQUIPMENT specifically intended for diagnostic use within urodynamics (measurement of
pressure-volume relationship of the urinary bladder when filled through a catheter with
water);
6) EQUIPMENT specifically intended for diagnostic use within male impotence testing
(measurement of amount of liquid infused, necessary to maintain a preset pressure level
for maintaining penile erection: cavernosometry, cavernosography).
1.3 Particular standards
Addition:
This Particular Standard refers to IEC 60601-1:1988, Medical electrical equipment – Part 1:
General requirements for safety as amended by its amendment 1 (1991) and amendment 2
(1995) and to the Collateral Standard IEC 60601-1-2:1993, Medical electrical equipment –
Part 1: General requirements for safety – 2. Collateral Standard: Electromagnetic compatibility
– Requirements and tests.
For brevity, Part 1 is referred to in this Particular Standard either as the General Standard or as
the General Requirement(s) and IEC 60601-1-2 as the Collateral Standard.
The numbering of sections, clauses and subclauses of this Particular Standard corresponds to
that of the General Standard. The changes to the text of the General Standard are specified by
the use of the following words:

---------------------- Page: 12 ----------------------

SIST EN 60601-2-24:1998
60601-2-24 © IEC:1998(E) – 7 –
“Replacement” means that the clause or subclause of the General Standard is replaced
completely by the text of this Particular Standard.
“Addition” means that the text of this Particular Standard is additional to the requirements of
the General Standard.
“Amendment” means that the clause or subclause of the General Standard is amended as
indicated by the text of this Particular Standard.
Subclauses or figures which are additional to those of the General Standard are numbered
starting from 101, additional annexes are lettered AA, BB, etc., and additional items aa), bb),
etc.
The term “this Standard” is used to make reference to the General Standard, the Collateral
Standard and this Particular Standard taken together.
Where there is no corresponding section, clause or subclause in this Particular Standard, the
section, clause or subclause of the General Standard, although possibly not relevant, applies
without modification; where it is intended that any part of the General Standard, although
possibly relevant, is not to be applied, a statement to that effect is given in this Particular
Standard.
The requirements of this Particular Standard take priority over those of the General Standard.
The requirements are followed by specifications for the relevant tests.
Following the decision taken by subcommittee 62D at the meeting in Washington in 1979, a
“General guidance and rationale” section giving some explanatory notes, where appropriate,
about the more important requirements is included in annex AA.
Clauses or subclauses for which there are explanatory notes in annex AA are marked with an
asterisk (*).
It is considered that a knowledge of the reasons for these requirements will not only facilitate
the proper application of the standard but will, in due course, expedite any revision neces-
sitated by changes in clinical practice or as a result of developments in technology. However,
this annex does not form part of the requirements of this Standard.
1.5 Collateral Standards
Addition:
This Particular Standard also refers to IEC 60601-1-2, which is applicable unless otherwise
stated in a particular clause or subclause.
2 Terminology and definitions
This clause of the General Standard and of the Collateral Standard IEC 60601-1-2 apply,
except as follows:
2.1.3
ACCESSORY
Addition:
Separate programmers are regarded as accessories and therefore a component part of the
EQUIPMENT

---------------------- Page: 13 ----------------------

SIST EN 60601-2-24:1998
– 8 – 60601-2-24 © IEC:1998(E)
2.1.5
APPLIED PART
Replacement:
entirety of all parts of the EQUIPMENT including the infusion liquid pathway that is intentionally in
contact with the PATIENT being treated in NORMAL USE
2.2.18
PORTABLE EQUIPMENT
Replacement:
TRANSPORTABLE EQUIPMENT intended to be moved from one location to another while in use or
between periods of use, by one or more persons or by other means
Additional definitions:
2.101
INFUSION PUMP
EQUIPMENT intended to regulate the flow of liquids into the PATIENT under positive pressure
generated by the pump
The INFUSION PUMP may be of:
– type 1: continuous infusion flow only,
– type 2: non-continuous flow only,
BOLUS
– type 3: discrete delivery of a ,
– type 4: type 1 combined with type 3 and/or type 2 in the same EQUIPMENT,
– type 5: PROFILE PUMP.
2.102
VOLUMETRIC INFUSION PUMP
INFUSION PUMP OPERATOR
in which the delivery rate is set by the and indicated by the
EQUIPMENT in volume per unit of time, but excluding SYRINGE PUMPS
2.103
DRIP-RATE INFUSION PUMP
INFUSION PUMP in which the delivery rate is set by the OPERATOR and indicated by the
EQUIPMENT as a number of drops per unit of time
2.104
INFUSION CONTROLLER
EQUIPMENT intended to regulate the flow of liquid into the PATIENT under positive pressure
generated by gravitational force
2.105
VOLUMETRIC INFUSION CONTROLLER
INFUSION CONTROLLER in which the delivery rate is set by the OPERATOR and indicated by the
EQUIPMENT in volume per unit of time
2.106
DRIP-RATE INFUSION CONTROLLER
INFUSION CONTROLLER in which the delivery rate is set by the OPERATOR and indicated by the
EQUIPMENT as a number of drops per unit of time

---------------------- Page: 14 ----------------------

SIST EN 60601-2-24:1998
60601-2-24 © IEC:1998(E) – 9 –
2.107
SPECIAL USE EQUIPMENT
EQUIPMENT in which the delivery rate is set by the OPERATOR and indicated by the EQUIPMENT in
units other than those defined in 2.101 to 2.106
2.108
SYRINGE PUMP
EQUIPMENT intended for controlled infusion of liquids into the PATIENT by means of one or more
single action syringe(s) or similar container(s) (e.g. where the cartridge is emptied by pushing
on its plunger)) and in which the delivery rate is set by the OPERATOR and indicated by the
EQUIPMENT in volume per unit of time
2.109
INFUSION PUMP FOR AMBULATORY USE
EQUIPMENT intended for the controlled infusion of liquids into the PATIENT and intended to be
carried continuously by the PATIENT
2.110
PROFILE PUMP
EQUIPMENT intended for controlled infusion of liquids into the PATIENT by means of a
programmed sequence of delivery rates
2.111
REGION OF CONTROL
that part of the EQUIPMENT within which flow regulation, flow shut-off or air detection occurs,
within the body of the EQUIPMENT or remotely
2.112
ADMINISTRATION SET
device(s) that convey(s) liquid from the supply via the EQUIPMENT to the PATIENT
2.113
PATIENT LINE
that part of the ADMINISTRATION SET between the EQUIPMENT and the PATIENT
2.114
SUPPLY LINE
that part of the ADMINISTRATION SET between the liquid supply and the EQUIPMENT
2.115
OCCLUSION ALARM THRESHOLD (PRESSURE)
value of the physical quantity at which the occlusion alarm is activated
2.116
KEEP OPEN RATE (KOR)
low predetermined rate(s) to which the EQUIPMENT reverts under specified conditions with the
object of keeping the PATIENT LINE open
NOTE – The abbreviation KVO (Keep-Vein-Open Rate) is commonly used as a synonym of KOR.
2.117
FREE FLOW
flow in an ADMINISTRATION SET which is not controlled by the EQUIPMENT, for example, due to the
unintended effects of gravity by the removal of the ADMINISTRATION SET from the EQUIPMENT
2.118
ADMINISTRATION SET CHANGE INTERVAL
time recommended by the manufacturer of the EQUIPMENT for using the ADMINISTRATION SET

---------------------- Page: 15 ----------------------

SIST EN 60601-2-24:1998
– 10 – 60601-2-24 © IEC:1998(E)
2.119
BOLUS
discrete quantity of liquid which is delivered in a short time
2.120
INTERMEDIATE RATE defined as follows:
– for volumetric infusion pumps and volumetric infusion controllers, set the rate to 25 ml/h;
– for drip-rate infusion pumps and drip-rate infusion controllers, set the rate to
20 drops/minute;
– for syringe pumps, set the rate to 5 ml/h;
– for special use equipment and infusion pumps for ambulatory use, set the rate specified by
the manufacturer as typical for the equipment.
2.121
MINIMUM RATE
lowest rate selectable by the OPERATOR, but not less than 1 ml/h
NOTE – For INFUSION PUMPS FOR AMBULATORY USE it is the lowest selectable rate.
2.122
MAXIMUM INFUSION PRESSURE
maximum pressure which can be generated by the EQUIPMENT under conditions of total
occlusion at the end of the PATIENT LINE
2.123
PATIENT END
that end of the PATIENT LINE where connection to the PATIENT takes place
3 General requirements
This clause of the General Standard applies, except as follows:
3.6* Addition
SINGLE FAULT CONDITIONS occurring in those protective systems specified in 51.5 and 51.102
shall become obvious to the OPERATOR within the ADMINISTRATION SET CHANGE INTERVAL. SINGLE
FAULT CONDITIONS occurring in the protective system specified in clause 51.103 shall cause the
cessation of delivery and the generation of an alarm within a time interval less than the volume
ADMINISTRATION SET
of the between the air detector and the venous cannula connected to it
divided by the maximum flow rate of the pump.
NOTE – Acceptable methods of complying with this requirement are, for example:
1) a safety system check initiated and controlled by the EQUIPMENT, first at the beginning of the
ADMINISTRATION SET CHANGE INTERVAL, and then repeated continuously as warranted;
2) one or more protective systems checks initiated by the OPERATOR and controlled by the EQUIPMENT within
the ADMINISTRATION SET CHANGE INTERVAL, with the OPERATOR initiating checks before or during the
infusion;
3) a safety system check carried out by the OPERATOR at least once within the ADMINISTRATION SET CHANGE
INTERVAL (see 6.8.2 a) 24)).
The following are not regarded as SINGLE FAULT CONDITIONS, but are regarded as NORMAL USE
CONDITIONS:
– leakage from the ADMINISTRATION SET and/or the liquid supply;
– depletion of the INTERNAL ELECTRICAL POWER SOURCE;
– mispositioning and/or incorrect filling of a drip chamber;

---------------------- Page: 16 ----------------------

SIST EN 60601-2-24:1998
60601-2-24 © IEC:1998(E) – 11 –
– air in the SUPPLY LINE or the REGION OF CONTROL;
– pulling on the PATIENT LINE (see ISO 8536-4).
5 Classification
This clause of the General Standard applies, except as follows:
5.2 Amendment:
Delete TYPE B APPLIED PART;
5.6 Amendment:
CONTINUOUS OPERATION
Delete all except for .
6 Identification, marking and documents
This clause of the General Standard and this clause of the Collateral Standard IEC 60601-1-2
apply, except as follows:
6.1 Marking on the outside of EQUIPMENT or EQUIPMENT parts
Addition:
aa)If detachable liquid reservoirs or PATIENT LINE(S) of specific sizes or brands, or containing
specific concentrations of drugs need to be used to maintain safe NORMAL USE of the
EQUIPMENT then relevant markings shall be fixed or indicated in a prominent place on the
EQUIPMENT which either identify those conditions or provide location of such information.
Compliance is checked by inspection.
6.1 q) Physiological effects
Replacement:
The body of the EQUIPMENT shall be marked with the following:
1) symbol No. 14 of appendix D of the General Standard or a statement to refer the OPERATOR
to the ACCOMPANYING DOCUMENTS;
2) an arrow or other appropriate symbol indicating the correct direction of flow if the
ADMINISTRATION SET can be incorrectly loaded;
3) EQUIPMENT as defined in 2.103 and 2.106 shall additionally be marked as follows:
"Caution: this equipment controls the drip rate not the volume delivered."
Additional items:
6.1.201 of the Collateral Standard, IEC 60601-1-2
Addition:
Compliance is checked by inspection.
6.8 ACCOMPANYING DOCUMENTS

---------------------- Page: 17 ----------------------

SIST EN 60601-2-24:1998
– 12 – 60601-2-24 © IEC:1998(E)
6.8.2 Instructions for use
a) Addition:
The instructions for use shall also include the following:
1) a list of the recommended ADMINISTRATION SET(S) to be used;
2) a warning of the consequences of the use of unsuitable ADMINISTRATION SET(S);
ACCESSORIES
3) a list of particular recommended by the manufacturer for use with the
EQUIPMENT;
4) permitted EQUIPMENT orientation and methods and precautions concerning its mounting, for
example, stability on a pole;
5) instructions regarding loading, priming, changing and reloading the ADMINISTRATION SET(S),
and the ADMINISTRATION SET CHANGE INTERVAL to maintain the specified performance;
6) instructions regarding the use of clamps on an ADMINISTRATION SET, the avoidance of FREE
FLOW conditions and the procedure to be followed when changing liquid containers;
7) where gravity is relevant to performance, the acceptable height range of the liquid
container above the PATIENT's heart;
8) the means provided to protect the PATIENT from air infusion;
9) a statement of the MAXIMUM INFUSION PRESSURE generated and the OCCLUSION ALARM
THRESHOLD (PRESSURE)(S) of the EQUIPMENT;
10) a statement of the maximum time for activation of the occlusion alarm when the
EQUIPMENT is operating at the MINIMUM RATE and the INTERMEDIATE RATE and at the
minimum and maximum selectable OCCLUSION ALARM THRESHOLD (PRESSURE)(S);
11) a statement of the BOLUS volume generated as a result of the EQUIPMENT operating at the
INTERMEDIATE RATE and reaching the minimum and maximum OCCLUSION ALARM THRESHOLD
(PRESSURE) (see also 51.5 b));
12) a statement of the means provided (if any) to manage the BOLUS before occlusion release;
13) a statement to indicate to the OPERATOR if the EQUIPMENT cannot be used as PORTABLE
EQUIPMENT;
14) precautions required with drop detectors, for example with respect to placement, cleanli-
ness, liquid level, ambient light;
15) recommendations on any specific method of cleaning and maintaining the EQUIPMENT;
16) the typical operating time when the EQUIPMENT is operating from the INTERNAL ELECTRICAL
POWER SOURCE at the INTERMEDIATE RATE;
17) a statement of KEEP OPEN RATE(S), and when initiated;
18) a list of alarms and their operating conditions;
19)* a warning that under certain circumstances the specified accuracy may not be maintained.
NOTE – The manufacturer must specify the parameters in which the device cannot maintain the specified accuracy;
e.g. minimum/maximum viscosity of liquids, reaction time of the safety system, scope of the risk analysis, etc.
20)*reference to a guide on the SAFETY HAZARDS associated with the interconnection of other
infusion systems or ACCESSORIES to the PATIENT LINE;
21) the rate obtained when the prime/purge or BOLUS control is operated, and a statement of
any alarm disabled;
22) a warning statement on the possible SAFETY HAZARDS associated with external radio-
frequency interference (RFI) or electromagnetic radiation which may affect the safe
operation of the EQUIPMENT. This statement should include examples of typical EQUIPMENT
which may generate such radiation;
23) the selectable rate range and the increments of selection;
24) guidance on tests to permit the OPERATOR to check the correct functioning of alarm(s) and
the operational safety of the EQUIPMENT;

---------------------- Page: 18 ----------------------

SIST EN 60601-2-24:1998
60601-2-24 © IEC:1998(E) – 13 –
25) data as evaluated by the test methods of 50.101 to 50.108 at the rates indicated in table
102, including an explanation for the OPERATOR of the data presentation;
26) the time for which the electronic memory is retained following switch-off;
27) for SPECIAL USE EQUIPMENT, the conversion factor(s) for volume divided by unit of time;
28)* the maximum volume that may be infused under SINGLE FAULT CONDITIONS;
29) guidance on the safe operation of the EQUIPMENT if it is connected operationally to a
remote control device;
30) information concerning type(s) of battery to be used and where available;
31) a statement of the meaning of claimed IP-classification.
6.8.201 of IEC 60601-1-2
Addition:
6.8.3 Technical description
Addition:
The technical description shall also include the following:
aa) the sensitivity of the air detector, if included to comply with 51.9, over the specified range
of rates for a single bubble;
bb) the units of measurement used for calibration of the EQUIPMENT;
cc) a description of any battery charging system;
dd) a functional description of the means provided to protect the PATIENT from EQUIPMENT error
resulting in overinfusion and, where applicable, in underinfusion;
ee) the manufacturer shall disclose the ADMINISTRATION SET(S) used for all the tests in this
standard.
Compliance is checked by inspection of the ACCOMPANYING DOCUMENTS.
SECTION TWO – ENVIRONMENTAL CONDITIONS
The clauses and subclauses of this section of the General Standard apply, except as follows:
10 Environmental conditions
This clause of the General Standard applies, except as follows:
Replacement:
10.2.1 a) An ambient temperature between +5 °C and +40 °C.
10.2.1 b) A relative humidity between 20 % and 90 %.

---------------------- Page: 19 ----------------------

SIST EN 60601-2-24:1998
– 14 – 60601-2-24 © IEC:1998(E)
SECTION THREE – PROTECTION AGAINST ELECTRIC SHOCK HAZARDS
The clauses and subclauses of the General Standard apply, except as follows:
14 Requirements related to classification
This clause of the General Standard applies, except as follows:
Replacement:
14.6 b) EQUIPMENT shall be of Type BF or CF.
14.6 d) EQUIPMENT intended for DIRECT CARDIAC APPLICATION having one or more APPLIED
PARTs of TYPE CF may have one or more additional APPLIED PARTS of TYPE BF which may be
applied simultaneously if the requirements of 6.1 l) and 19.3 for such EQUIPMENT have been
met.
17 Separation
This clause of the General Standard applies, except as follows:
Item c) is not applicable.
19 Continuous LEAKAGE CURRENTS and PATIENT AUXILIARY CURRENTS
This clause of the General Standard applies, except as follows:
19.4 Tests:
d) Measuring arrangement
Addition:
3) Measurement of
...

Questions, Comments and Discussion

Ask us and Technical Secretary will try to provide an answer. You can facilitate discussion about the standard in here.