Medical electrical equipment -- Part 1-10: General requirements for basic safety and essential performance - Collateral Standard: Requirements for the development of physiologic closed-loop controllers

This International Standard applies to the BASIC SAFETY and ESSENTIAL PERFORMANCE of MEDICAL ELECTRICAL EQUIPMENT and MEDICAL ELECTRICAL SYSTEMS, hereafter referred to as ME EQUIPMENT and ME SYSTEMS. This collateral standard specifies requirements for the development (analysis, design, VERIFICATION and VALIDATION) of a PHYSIOLOGIC CLOSED-LOOP CONTROLLER (PCLC) as part of a PHYSIOLOGIC CLOSED-LOOP CONTROL SYSTEM (PCLCS) in ME EQUIPMENT and ME SYSTEMS to control a PHYSIOLOGIC VARIABLE. This collateral standard applies to various types of PCLC, e.g. linear and non-linear, adaptive, fuzzy, neural networks. This collateral standard does not specify: additional mechanical requirements; or additional electrical requirements. This collateral standard applies to a closed-loop controller (see Figure 1) that sets the CONTROLLER OUTPUT VARIABLE in order to adjust (i.e., change or maintain) the measured PHYSIOLOGIC VARIABLE by relating it to the REFERENCE VARIABLE. A closed-loop controller that maintains a physical or chemical VARIABLE, using feedback that is not measured from a PATIENT, is outside the scope of this standard.

Medizinische elektrische Geräte - Teil 1-10: Allgemeine Festlegungen für die Sicherheit einschließlich der wesentlichen Leistungsmerkmale - Ergänzungsnorm: Anforderungen an die Entwicklung von physiologischen geschlossenen Regelkreisen

Appareils électromédicaux -- Partie 1-10: Exigences générales pour la sécurité de base et les performances essentielles - Norme collatérale: Exigences pour le développement des régulateurs physiologiques en boucle fermée

Spécifie les exigences pour le développement (analyse, conception, vérification et validation) d'un régulateur physiologique en boucle fermée (RPBF) en tant que partie d'un système physiologique de commande en boucle fermée (SPCBF) dans les appareils em et les systèmes em pour la commande d'une variable physiologique.[
]La présente norme collatérale s'applique aux différents types de RPBF, par exemple linéaire et non-linéaire, adaptatifs, fuzzy, réseaux de neurones.[
]La présente norme collatérale s'applique à un régulateur en boucle fermée qui règle la variable de sortie du régulateur afin d'ajuster (c'est-à-dire, changer ou maintenir) la variable physiologique mesurée en effectuant une comparaison avec la variable de référence.

Medicinska električna oprema - 1-10 del: Splošne zahteve za osnovne varnostne in bistvene lastnosti - Spremljevalni standard: Zahteve za razvoj fizioloških krmilnikov s sklenjeno zanko (IEC 60601-1-10:2007)

General Information

Status
Published
Publication Date
05-Jun-2008
Current Stage
6060 - National Implementation/Publication (Adopted Project)
Start Date
09-May-2008
Due Date
14-Jul-2008
Completion Date
06-Jun-2008

Relations

Buy Standard

Standard
EN 60601-1-10:2008
English language
44 pages
sale 10% off
Preview
sale 10% off
Preview
e-Library read for
1 day

Standards Content (Sample)

SLOVENSKI STANDARD
SIST EN 60601-1-10:2008
01-julij-2008
0HGLFLQVNDHOHNWULþQDRSUHPDGHO6SORãQH]DKWHYH]DRVQRYQHYDUQRVWQHLQ
ELVWYHQHODVWQRVWL6SUHPOMHYDOQLVWDQGDUG=DKWHYH]DUD]YRMIL]LRORãNLK
NUPLOQLNRYVVNOHQMHQR]DQNR ,(&
Medical electrical equipment - Part 1-10: General requirements for basic safety and
essential performance - Collateral Standard: Requirements for the development of
physiologic closed-loop controllers (IEC 60601-1-10:2007)
Medizinische elektrische Geräte - Teil 1-10: Allgemeine Festlegungen für die Sicherheit
einschließlich der wesentlichen Leistungsmerkmale - Ergänzungsnorm: Anforderungen
an die Entwicklung von physiologischen geschlossenen Regelkreisen (IEC 60601-1-
10:2007)
Appareils électromédicaux -- Partie 1-10: Exigences générales pour la sécurité de base
et les performances essentielles - Norme collatérale: Exigences pour le développement
des régulateurs physiologiques en boucle fermée (CEI 60601-1-10:2007)
Ta slovenski standard je istoveten z: EN 60601-1-10:2008
ICS:
11.040.01 Medicinska oprema na Medical equipment in general
splošno
SIST EN 60601-1-10:2008 en,fr
2003-01.Slovenski inštitut za standardizacijo. Razmnoževanje celote ali delov tega standarda ni dovoljeno.

---------------------- Page: 1 ----------------------

EUROPEAN STANDARD
EN 60601-1-10

NORME EUROPÉENNE
April 2008
EUROPÄISCHE NORM

ICS 11.040


English version


Medical electrical equipment -
Part 1-10: General requirements for basic safety
and essential performance -
Collateral Standard: Requirements for the development
of physiologic closed-loop controllers
(IEC 60601-1-10:2007)


Appareils électromédicaux -  Medizinische elektrische Geräte -
Partie 1-10: Exigences générales Teil 1-10: Allgemeine Festlegungen
pour la sécurité de base für die Sicherheit einschließlich
et les performances essentielles - der wesentlichen Leistungsmerkmale -
Norme collatérale: Exigences Ergänzungsnorm: Anforderungen
pour le développement des régulateurs an die Entwicklung von physiologischen
physiologiques en boucle fermée geschlossenen Regelkreisen
(CEI 60601-1-10:2007) (IEC 60601-1-10:2007)



This European Standard was approved by CENELEC on 2008-03-01. CENELEC members are bound to comply
with the CEN/CENELEC Internal Regulations which stipulate the conditions for giving this European Standard
the status of a national standard without any alteration.

Up-to-date lists and bibliographical references concerning such national standards may be obtained on
application to the Central Secretariat or to any CENELEC member.

This European Standard exists in three official versions (English, French, German). A version in any other
language made by translation under the responsibility of a CENELEC member into its own language and notified
to the Central Secretariat has the same status as the official versions.

CENELEC members are the national electrotechnical committees of Austria, Belgium, Bulgaria, Cyprus, the
Czech Republic, Denmark, Estonia, Finland, France, Germany, Greece, Hungary, Iceland, Ireland, Italy, Latvia,
Lithuania, Luxembourg, Malta, the Netherlands, Norway, Poland, Portugal, Romania, Slovakia, Slovenia, Spain,
Sweden, Switzerland and the United Kingdom.

CENELEC
European Committee for Electrotechnical Standardization
Comité Européen de Normalisation Electrotechnique
Europäisches Komitee für Elektrotechnische Normung

Central Secretariat: rue de Stassart 35, B - 1050 Brussels


© 2008 CENELEC - All rights of exploitation in any form and by any means reserved worldwide for CENELEC members.
Ref. No. EN 60601-1-10:2008 E

---------------------- Page: 2 ----------------------

EN 60601-1-10:2008 - 2 -

Foreword
The text of document 62A/576/FDIS, future edition 1 of IEC 60601-1-10, prepared by SC 62A, Common
aspects of electrical equipment used in medical practice, of IEC TC 62, Electrical equipment in medical
practice, and ISO SC 1, Breathing attachments and anaesthetic machines, and SC 3, Lung ventilators
and related devices, of ISO TC 121, Anaesthetic and respiratory equipment, was submitted to the
IEC-CENELEC parallel vote and was approved by CENELEC as EN 60601-1-10 on 2008-03-01
The following dates were fixed:

– latest date by which the EN has to be implemented

at national level by publication of an identical
(dop) 2008-12-01
national standard or by endorsement

– latest date by which the national standards conflicting
(dow) 2011-03-01
with the EN have to be withdrawn
This European Standard has been prepared under a mandate given to CENELEC by the European
Commission and the European Free Trade Association and covers essential requirements of
EC Directive 93/42/EEC. See Annex ZZ.
This European Standard constitutes a collateral standard to EN 60601-1:2006, hereafter referred to as
the general standard.
In the 60601 series of publications, collateral standards specify general requirements for safety applicable
to:
– a subgroup of MEDICAL ELECTRICAL EQUIPMENT (e.g. radiological equipment); or
– a specific characteristic of all MEDICAL ELECTRICAL EQUIPMENT, not fully addressed in the general
standard (e.g. ALARM SYSTEMS).
In this collateral standard, the following print types are used:
− requirements and definitions: in roman type;
− test specifications: in italic type;
− informative material appearing outside of tables, such as notes, examples and references: in smaller type. Normative text of
tables is also in a smaller type;
− TERMS DEFINED IN CLAUSE 3 OF THE GENERAL STANDARD, IN THIS COLLATERAL STANDARD OR AS NOTED: IN
SMALL CAPITALS.
In referring to the structure of this standard, the term
– “clause” means one of the eight numbered divisions within the table of contents, inclusive of all
subdivisions (e.g. Clause 8 includes Subclauses 8.1, 8.2, etc.);
– “subclause” means a numbered subdivision of a clause (e.g. 8.1, 8.2 and 8.2.1 are all subclauses of
Clause 8).
References to clauses within this standard are preceded by the term “Clause” followed by the clause
number. References to subclauses within this standard are by number only.
In this standard, the conjunctive “or” is used as an “inclusive or” so a statement is true if any combination
of the conditions is true.

---------------------- Page: 3 ----------------------

- 3 - EN 60601-1-10:2008
The verbal forms used in this standard conform to usage described in Annex H of the ISO/IEC Directives,
Part 2. For the purposes of this standard, the auxiliary verb
− “shall” means that compliance with a requirement or a test is mandatory for compliance with this
standard;
− “should” means that compliance with a requirement or a test is recommended but is not mandatory for
compliance with this standard;
− “may” is used to describe a permissible way to achieve compliance with a requirement or test.
Clauses, subclauses and definitions for which a rationale is provided in informative Annex A are marked
with an asterisk (*).
Annexes ZA and ZZ have been added by CENELEC.
__________
Endorsement notice
The text of the International Standard IEC 60601-1-10:2007 was approved by CENELEC as a European
Standard without any modification.
In the official version, for Bibliography, the following notes have to be added for the standards indicated:
ISO 9000 NOTE Harmonized as EN ISO 9000:2005 (not modified).
ISO 14001 NOTE Harmonized as EN ISO 14001:2004 (not modified).
ISO 14021 NOTE Harmonized as EN ISO 14021:2001 (not modified).
ISO 14040 NOTE Harmonized as EN ISO 14040:2006 (not modified).
__________

---------------------- Page: 4 ----------------------

EN 60601-1-10:2008 - 4 -

Annex ZA
(normative)

Normative references to international publications
with their corresponding European publications

The following referenced documents are indispensable for the application of this document. For dated
references, only the edition cited applies. For undated references, the latest edition of the referenced
document (including any amendments) applies.

NOTE  When an international publication has been modified by common modifications, indicated by (mod), the relevant EN/HD
applies.

Publication Year Title EN/HD Year
IEC 60601-1 2005 Medical electrical equipment - EN 60601-1 2006
Part 1: General requirements for basic safety
and essential performance


IEC 60601-1-6 2006 Medical electrical equipment - EN 60601-1-6 2007
Part 1-6: General requirements for basic
safety and essential performance - Collateral
Standard: Usability


IEC 60601-1-8 2006 Medical electrical equipment - EN 60601-1-8 2007
Part 1-8: General requirements for basic
safety and essential performance - Collateral
Standard: General requirements, tests and
guidance for alarm systems in medical
electrical equipment and medical electrical
systems


IEC 62304 2006 Medical device software - Software life-cycle EN 62304 2006
processes


1) 2)
ISO 14971 - Medical devices - Application of risk EN ISO 14971 2007
management to medical devices






1)
Undated reference.
2)
Valid edition at date of issue.

---------------------- Page: 5 ----------------------

- 5 - EN 60601-1-10:2008
Annex ZZ
(informative)
Coverage of Essential Requirements of EC Directives

This European Standard has been prepared under a mandate given to CENELEC by the European
Commission and the European Free Trade Association and within its scope the standard covers all
relevant essential requirements as given in Annex I of the EC Directive 93/42/EEC.
Compliance with this standard provides one means of conformity with the specified essential
requirements of the Directive concerned.
WARNING: Other requirements and other EC Directives may be applicable to the products falling within
the scope of this standard.

---------------------- Page: 6 ----------------------

IEC 60601-1-10
Edition 1.0 2007-11
INTERNATIONAL
STANDARD
NORME
INTERNATIONALE
Medical electrical equipment –
Part 1-10: General requirements for basic safety and essential performance –
Collateral Standard: Requirements for the development of physiologic
closed-loop controllers

Appareils électromédicaux –
Partie 1-10: Exigences générales pour la sécurité de base et les performances
essentielles – Norme collatérale: Exigences pour le développement des
régulateurs physiologiques en boucle fermée

INTERNATIONAL
ELECTROTECHNICAL
COMMISSION
COMMISSION
ELECTROTECHNIQUE
PRICE CODE
INTERNATIONALE
W
CODE PRIX
ICS 11.040 ISBN 2-8318-9448-4

---------------------- Page: 7 ----------------------

– 2 – 60601-1-10 © IEC:2007
CONTENTS
FOREWORD.4
INTRODUCTION.7

1 Scope, object and related standards.8
1.1 * Scope .8
1.2 Object .8
1.3 Related standards .8
1.3.1 IEC 60601-1 .8
1.3.2 Particular standards .9
2 Normative references .9
3 Terms and definitions .9
4 * General requirements .14
5 ME EQUIPMENT identification, marking and documents .14
5.1 * Instructions for use .14
5.2 Technical description.15
6 Accuracy of controls and instruments and protection against hazardous outputs .15
6.1 * USABILITY .15
6.2 ALARM SYSTEMS .15
6.3 * PCLCS VARIABLE logging .15
6.4 * DISTRIBUTED PCLCS .16
7 * PROGRAMMABLE ELECTRICAL MEDICAL SYSTEMS (PEMS) .16
8 Requirements for PHYSIOLOGIC CLOSED-LOOP CONTROLLER (PCLC) development.16
8.1 * General.16
8.2 Attributes/activities of the PCLC development PROCESS .17
8.2.1 RECORDS and PROCESS scaling .17
8.2.2 Equipment specifications .17
8.2.3 * Disturbance management.20
8.2.4 * PCLC VERIFICATION.21
8.2.5 * PCLCS VALIDATION .21

Annex A (informative) General guidance and rationale.22
Annex B (informative) Description of dynamic performance of a PCLCS .32
Annex C (informative) Guide to marking and labelling requirements for ME EQUIPMENT
and ME SYSTEMS.36

Bibliography.37

Index of defined terms used in this collateral standard.38

Figure 1 – Functional diagram indicating typical components of a PHYSIOLOGIC CLOSED-
LOOP CONTROL SYSTEM (PCLCS) utilizing a PCLC .10
Figure B.1 – Example of PCLCS dynamic performance with no STEADY-STATE DEVIATION.33
Figure B.2 – Example of PCLCS dynamic performance with STEADY-STATE DEVIATION.34
Figure B.3 – Example of PCLCS dynamic performance transient COMMAND VARIABLE.35

---------------------- Page: 8 ----------------------

60601-1-10 © IEC:2007 – 3 –

Table A.1 – Examples of ME EQUIPMENT or ME SYSTEMS that incorporate a PCLCS .22
Table C.2 – ACCOMPANYING DOCUMENTS, instructions for use.36
Table C.3 – ACCOMPANYING DOCUMENTS, technical description .36

---------------------- Page: 9 ----------------------

– 4 – 60601-1-10 © IEC:2007
INTERNATIONAL ELECTROTECHNICAL COMMISSION
____________

MEDICAL ELECTRICAL EQUIPMENT –

Part 1-10: General requirements for basic safety
and essential performance –
Collateral Standard:
Requirements for the development of
physiologic closed-loop controllers


FOREWORD
1) The International Electrotechnical Commission (IEC) is a worldwide organization for standardization comprising
all national electrotechnical committees (IEC National Committees). The object of IEC is to promote
international co-operation on all questions concerning standardization in the electrical and electronic fields. To
this end and in addition to other activities, IEC publishes International Standards, Technical Specifications,
Technical Reports, Publicly Available Specifications (PAS) and Guides (hereafter referred to as “IEC
Publication(s)”). Their preparation is entrusted to technical committees; any IEC National Committee interested
in the subject dealt with may participate in this preparatory work. International, governmental and non-
governmental organizations liaising with the IEC also participate in this preparation. IEC collaborates closely
with the International Organization for Standardization (ISO) in accordance with conditions determined by
agreement between the two organizations.
2) The formal decisions or agreements of IEC on technical matters express, as nearly as possible, an international
consensus of opinion on the relevant subjects since each technical committee has representation from all
interested IEC National Committees.
3) IEC Publications have the form of recommendations for international use and are accepted by IEC National
Committees in that sense. While all reasonable efforts are made to ensure that the technical content of IEC
Publications is accurate, IEC cannot be held responsible for the way in which they are used or for any
misinterpretation by any end user.
4) In order to promote international uniformity, IEC National Committees undertake to apply IEC Publications
transparently to the maximum extent possible in their national and regional publications. Any divergence
between any IEC Publication and the corresponding national or regional publication shall be clearly indicated in
the latter.
5) IEC provides no marking procedure to indicate its approval and cannot be rendered responsible for any
equipment declared to be in conformity with an IEC Publication.
6) All users should ensure that they have the latest edition of this publication.
7) No liability shall attach to IEC or its directors, employees, servants or agents including individual experts and
members of its technical committees and IEC National Committees for any personal injury, property damage or
other damage of any nature whatsoever, whether direct or indirect, or for costs (including legal fees) and
expenses arising out of the publication, use of, or reliance upon, this IEC Publication or any other IEC
Publications.
8) Attention is drawn to the Normative references cited in this publication. Use of the referenced publications is
indispensable for the correct application of this publication.
9) Attention is drawn to the possibility that some of the elements of this IEC Publication may be the subject of
patent rights. IEC shall not be held responsible for identifying any or all such patent rights.
International standard IEC 60601-1-10 has been prepared by IEC subcommittee 62A:
Common aspects of electrical equipment used in medical practice, of IEC technical committee
62: Electrical equipment in medical practice, and ISO subcommittees SC1: Breathing
attachments and anaesthetic machines, and SC3: Lung ventilators and related devices of ISO
technical committee 121: Anaesthetic and respiratory equipment.
It is published as double logo standard.
This first edition constitutes a collateral standard to IEC 60601-1: Medical electrical
equipment – Part 1: General requirements for safety and essential performance hereafter
referred to as the general standard.

---------------------- Page: 10 ----------------------

60601-1-10 © IEC:2007 – 5 –
The text of this collateral standard is based on the following documents:
FDIS Report on voting
62A/576/FDIS 62A/585/RVD

Full information on the voting for the approval of this collateral standard can be found in the
report on voting indicated in the above table. In ISO, the standard has been approved by
18 P-members out of 19 having cast a vote.
This publication has been drafted in accordance with the ISO/IEC Directives, Part 2.
In the 60601 series of publications, collateral standards specify general requirements for
safety applicable to:
– a subgroup of MEDICAL ELECTRICAL EQUIPMENT (e.g. radiological equipment); or
– a specific characteristic of all MEDICAL ELECTRICAL EQUIPMENT, not fully addressed in the
general standard (e.g. ALARM SYSTEMS).
In this collateral standard, the following print types are used:
– requirements and definitions: roman type.
– test specifications: italic type.
– informative material appearing outside of tables, such as notes, examples and references: in smaller type.
Normative text of tables is also in a smaller type.
– TERMS DEFINED IN CLAUSE 3 OF THE GENERAL STANDARD, IN THIS COLLATERAL STANDARD OR AS
NOTED: SMALL CAPITALS.
In referring to the structure of this standard, the term
– “clause” means one of the eight numbered divisions within the table of contents, inclusive
of all subdivisions (e.g. Clause 8 includes Subclauses 8.1, 8.2, etc.);
– “subclause” means a numbered subdivision of a clause (e.g. 8.1, 8.2 and 8.2.1 are all
subclauses of Clause 8).
References to clauses within this standard are preceded by the term “Clause” followed by the
clause number. References to subclauses within this standard are by number only.
In this standard, the conjunctive “or” is used as an “inclusive or” so a statement is true if any
combination of the conditions is true.
The verbal forms used in this standard conform to usage described in Annex H of the ISO/IEC
Directives, Part 2. For the purposes of this standard, the auxiliary verb:
– “shall” means that compliance with a requirement or a test is mandatory for compliance
with this standard;
– “should” means that compliance with a requirement or a test is recommended but is not
mandatory for compliance with this standard;
– “may” is used to describe a permissible way to achieve compliance with a requirement or
test.
Clauses, subclauses and definitions for which a rationale is provided in informative Annex A
are marked with an asterisk (*).
A list of all parts of the IEC 60601 series, published under the general title: Medical electrical
equipment, can be found on the IEC website.

---------------------- Page: 11 ----------------------

– 6 – 60601-1-10 © IEC:2007

The committee has decided that the contents of this publication will remain unchanged until
the maintenance result date indicated on the IEC web site under "http://webstore.iec.ch" in
the data related to the specific publication. At this date, the publication will be
• reconfirmed;
• withdrawn;
• replaced by a revised edition, or
• amended

---------------------- Page: 12 ----------------------

60601-1-10 © IEC:2007 – 7 –
INTRODUCTION
The use of PHYSIOLOGIC CLOSED-LOOP CONTROLLERS in ME EQUIPMENT and ME SYSTEMS are
expected to provide a successful strategy to improve PATIENT safety and reduce healthcare
)
1
costs [9][10][11][12][13] . New RISKS that are not directly addressed by previous standards
are emerging in the development of this equipment. MANUFACTURERS employ a variety of
methods to validate the safety and integrity of control systems with varying degrees of
success. Classical methods of software VALIDATION for PHYSIOLOGIC CLOSED-LOOP
CONTROLLERS can be insufficient to ensure performance with acceptable RISKS under all
clinical and physiologic conditions.
—————————
)
1
Figures in square brackets refer to the Bibliography.

---------------------- Page: 13 ----------------------

– 8 – 60601-1-10 © IEC:2007
MEDICAL ELECTRICAL EQUIPMENT –

Part 1-10: General requirements for basic safety
and essential performance –
Collateral Standard:
Requirements for the development of
physiologic closed-loop controllers



1 Scope, object and related standards
1.1 * Scope
This International Standard applies to the BASIC SAFETY and ESSENTIAL PERFORMANCE of
MEDICAL ELECTRICAL EQUIPMENT and MEDICAL ELECTRICAL SYSTEMS, hereafter referred to as
ME EQUIPMENT and ME SYSTEMS.
This collateral standard specifies requirements for the development (analysis, design,
VERIFICATION and VALIDATION) of a PHYSIOLOGIC CLOSED-LOOP CONTROLLER (PCLC) as part of a
PHYSIOLOGIC CLOSED-LOOP CONTROL SYSTEM (PCLCS) in ME EQUIPMENT and ME SYSTEMS to
control a PHYSIOLOGIC VARIABLE.
NOTE A PHYSIOLOGIC VARIABLE can be a body chemistry (e.g. electrolytes, blood glucose), a physical property
(e.g. PATIENT temperature, electrophysiologic, hemodynamic), or a pharmaceutical concentration.
This collateral standard applies to various types of PCLC, e.g. linear and non-linear, adaptive,
fuzzy, neural networks.
This collateral standard does not specify:
− additional mechanical requirements; or
− additional electrical requirements.
This collateral standard applies to a closed-loop controller (see Figure 1) that sets the
CONTROLLER OUTPUT VARIABLE in order to adjust (i.e., change or maintain) the measured
PHYSIOLOGIC VARIABLE by relating it to the REFERENCE VARIABLE.
A closed-loop controller that maintains a physical or chemical VARIABLE, using feedback that is
not measured from a PATIENT, is outside the scope of this standard.
1.2 Object
The object of this collateral standard is to specify general requirements that are in addition to
those of the general standard and to serve as the basis for particular standards.
1.3 Related standards
1.3.1 IEC 60601-1
For ME EQUIPMENT and ME SYSTEMS, this collateral standard complements IEC 60601-1.
When referring to IEC 60601-1 or to this collateral standard, either individually or in
combination, the following conventions are used:
− "the general standard" designates IEC 60601-1 alone;
− "this collateral standard" designates IEC 60601-1-10 alone;

---------------------- Page: 14 ----------------------

60601-1-10 © IEC:2007 – 9 –
− "this standard" designates the combination of the general standard and this collateral
standard.
1.3.2 Particular standards
A requirement in a particular standard takes priority over the corresponding requirement in
this collateral standard.
2 Normative references
The following referenced documents are indispensable for the application of this document.
For dated references, only the edition cited applies. For undated references, the latest edition
of the referenced document (including any amendments) applies.
IEC 60601-1:2005, Medical electrical equipment – Part 1: General requirements for basic
safety and essential performance
IEC 60601-1-6:2006, Medical electrical equipment – Part 1-6: General requirements for basic
safety and essential performance – Collateral Standard: Usability
IEC 60601-1-8:2006, Medical electrical equipment – Part 1-8: General requirements for basic
safety and essential performance – Collateral Standard: General requirements, tests and
guidance for alarm systems in medical electrical equipment and medical electrical systems
IEC 62304:2006, Medical device software – Software life cycle processes
ISO 14971, Medical devices – Application of risk management to medical devices
3 Terms and definitions
For the purposes of this document, the terms and definitions given in IEC 60601-1:2005,
IEC 60601-1-6:2006, IEC 60601-1-8:2006 and the following apply.
NOTE An index of defined term used in this collateral standard is found beginning on page 38.
3.1
ACTUATOR
A
part of a PCLCS that performs a specified output function (see, for example, Figure 1, A)
EXAMPLE 1 A heater delivers thermal energy.
EXAMPLE 2 An infusion pump delivers a fluid or drug.
EXAMPLE 3 An anaesthetic agent vaporizer delivers a vapour concentration.
EXAMPLE 4 A ventilator delivers an inspiratory volume.
3.2
COMMAND OVERSHOOT
y
co
for a step response, the maximum positive deviation of the PHYSIOLOGIC VARIABLE (y), from the
COMMAND VARIABLE (c)
NOTE See also Annex B.

---------------------- Page: 15 ----------------------

– 10 – 60601-1-10 © IEC:2007
v
p
PCLC
c m
w ex y
+
C D E A P
-
f
F
IEC  2068/07

Elements VARIABLES
PCLC PHYSIOLOGIC CLOSED-LOOP CONTROLLER
A ACTUATOR
m MANIPULATED VARIABLE
C COMMAND TRANSFER ELEMENT w REFERENCE VARIABLE
D COMPARING ELEMENT e ERROR VARIABLE
E CONTROL TRANSFER ELEMENT x CONTROLLER OUTPUT VARIABLE
F MEASURING TRANSFER ELEMENT f FEEDBACK VARIABLE
P PATIENT TRANSFER ELEMENT y PHYSIOLOGIC VARIABLE
v PATIENT DISTURBANCE VARIABLE
p

c COMMAND VARIABLE
NOTE DISTURBANCE VARIABLES (v), not shown, can act on any element or VARIABLE.
Figure 1 – Functional diagram indicating typical components of a PHYSIOLOGIC CLOSED-
LOOP CONTROL SYSTEM (PCLCS) utilizing a PCLC
3.3
* COMMAND TRANSFER ELEMENT
C
part of a PCLCS that provides an output having a deterministic relationship to the COMMAND
VARIABLE (c) (see, for example, Figure 1, C)
3.4
* COMMAND VARIABLE
c
VARIABLE which, after signal conversion or other processing by the COMMAND TRANSFER
ELEMENT (C), gives the REFERENCE VARIABLE (w) (see, for example, Figure 1, c)
3.5
* COMPARING ELEMENT
D
element with two inputs and one output, the output VARIABLE being the difference between the
input VARIABLES (see, for example, Figure 1, D)
[IEC 60050-351, definition 351-28-03, modified]
NOTE The difference can be simple subtraction, classification within a value range, or a complex relationship
such as results from a neural network calculation.
3.6
CONTROL TRANSFER ELEMENT
E
part of a PCLC that provides an output having a deterministic relationship to the FEEDBACK
VARIABLE (f) (see, for example, Figure 1, E)

---------------------- Page: 16 ----------------------

60601-1-10 © IEC:
...

Questions, Comments and Discussion

Ask us and Technical Secretary will try to provide an answer. You can facilitate discussion about the standard in here.