ASTM E1455-03
(Practice)Standard Practice for Obtaining Colorimetric Data from a Visual Display Unit Using Tristimulus Colorimeters
Standard Practice for Obtaining Colorimetric Data from a Visual Display Unit Using Tristimulus Colorimeters
SIGNIFICANCE AND USE
This practice may be applied when tristimulus colorimeters are used to measure the colors produced on self-luminous video display devices such as CRTs and flat-panel displays, including electroluminescent (EL) panels, field emission displays (FEDs), and back-lit liquid crystal displays (LCDs). This practice is not meant to be a complete description of a procedure to measure the color coordinates of a display. Rather, it provides a method for obtaining more accurate results when certain conditions are met. It may be used by any person engaged in the measurement of color on display devices who has access to the requisite equipment.
This practice defines a class of tristimulus colorimeters that may be said to be compatible with this practice.
SCOPE
1.1 This practice is intended as an aid for improving the accuracy of colorimetric measurements made with tristimulus colorimeters on visual display units, such as cathode ray tubes (CRTs) and self-luminous flat-panel displays. It explains a useful step in the analysis of colorimetric data that takes advantage of the fact that light from such displays consists of an additive mixture of three primary colored lights. However, it is not a complete specification of how such measurements should be made.
1.2 This practice is limited to display devices and colorimetric instruments that meet linearity criteria as defined in the practice. It is not concerned with effects that might cause measurement bias such as temporal or geometric differences between the instrument being optimized and the instrument used for reference.
1.3 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use.
General Information
Relations
Standards Content (Sample)
NOTICE: This standard has either been superseded and replaced by a new version or withdrawn.
Contact ASTM International (www.astm.org) for the latest information.
Designation:E1455–03
Standard Practice for
Obtaining Colorimetric Data from a Visual Display Unit
1
Using Tristimulus Colorimeters
This standard is issued under the fixed designation E1455; the number immediately following the designation indicates the year of
original adoption or, in the case of revision, the year of last revision. A number in parentheses indicates the year of last reapproval. A
superscript epsilon (´) indicates an editorial change since the last revision or reapproval.
INTRODUCTION
This practice provides directions for correcting the results obtained with tristimulus colorimeters
when measuring the tristimulus values or chromaticity coordinates of colored displays. Tristimulus
colorimeters approximate the CIE color matching functions x¯ (l), y¯ (l), z¯ (l) to make these
measurements. The errors generated in measuring colors on a display may be minimized using this
practice.
1. Scope E1336 Test Method for Obtaining Colorimetric Data From
a Visual Display Unit by Spectroradiometry
1.1 This practice is intended as an aid for improving the
E1341 Practice for Obtaining Spectroradiometric Data from
accuracy of colorimetric measurements made with tristimulus
Radiant Sources for Colorimetry
colorimeters on visual display units, such as cathode ray tubes
2.2 ISO/CIE Standard:
(CRTs) and self-luminous flat-panel displays. It explains a
CIE Standard Colorimetric Observers, ISO/CIE
useful step in the analysis of colorimetric data that takes
10527: 1991 (E) (International Organization for Stan-
advantage of the fact that light from such displays consists of
3
dardization, Geneva, 1991)
an additive mixture of three primary colored lights. However,
it is not a complete specification of how such measurements
3. Terminology
should be made.
3.1 Definitions: Unless otherwise stated, definitions of ap-
1.2 This practice is limited to display devices and colori-
pearance terms in Terminology E284 are applicable to this
metric instruments that meet linearity criteria as defined in the
practice.
practice. It is not concerned with effects that might cause
3.2 Definitions of Terms Specific to This Standard:
measurement bias such as temporal or geometric differences
3.2.1 calibration, n— in reference to a tristimulus colorim-
between the instrument being optimized and the instrument
eter,theprocessperformedoutsideofthispracticetoadjustthe
used for reference.
tristimulus colorimeter to provide the best possible results for
1.3 This standard does not purport to address all of the
average or predefined conditions.
safety concerns, if any, associated with its use. It is the
3.2.2 optimization, n— in reference to a tristimulus colo-
responsibility of the user of this standard to establish appro-
rimeter, the process performed pursuant to this practice to
priate safety and health practices and determine the applica-
adjust the tristimulus colorimeter or to interpret its readings to
bility of regulatory limitations prior to use.
provide better results when applied to a particular display
2. Referenced Documents device.
2 3.2.3 compatible, adj— in reference to a tristimulus colo-
2.1 ASTM Standards:
rimeter, one so designed as to automate the procedure de-
E284 Terminology of Appearance
scribed in this practice.
4. Summary of Practice
1
This practice is under the jurisdiction of ASTM Committee E12 on Color and
4.1 Tristimulus colorimeters comprised of three or four
Appearance and is the direct responsibility of Subcommittee E12.06 onAppearance
detector channels are, in general, not amenable to accurate
of Displays.
Current edition approved Dec. 1, 2003. Published January 2004. Originally
approved in 1992. Last previous edition approved in 1997 as E1455 – 97. DOI:
3
10.1520/E1455-03. Currently available through the U.S. National Committee of the CIE, c/o Mr.
2
For referenced ASTM standards, visit the ASTM website, www.astm.org, or Thomas M. Lemons, TLA-Lighting Consultants, Inc., 7 Pond Street, Salem, MA
contact ASTM Customer Service at service@astm.org. For Annual Book of ASTM 01970-4819. Also included in ASTM Standards on Color and Appearance, Fifth
Standards volume information, refer to the standard’s Document Summary page on Edition, 1996.
the ASTM website.
Copyright © ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959, United States.
1
---------------------- Page: 1 ----------------------
E1455–03
4
calibration that holds for all manner of usage with different dard 10527 and the CIE in its publication No. 15.2 (1). For
illuminated devices and objects. This is because the spectral light with a color stimulus function F(l),
responsivities of their detector channels do not exactly match
830 nm
X 5 k F~l!x¯~l!dl (1)
*
the defined Commission Internationale de L’Éclairage (CIE) x¯
360 nm
(l), y¯(l), z¯ (l) functions. Factory or sub
...
Questions, Comments and Discussion
Ask us and Technical Secretary will try to provide an answer. You can facilitate discussion about the standard in here.