Standard Guide for Evaluating Data Acquisition Systems Used in Cyclic Fatigue and Fracture Mechanics Testing

ABSTRACT
This guide covers how to understand and minimize the errors associated with data acquisition in fatigue and fracture mechanics testing equipment. This guide is not intended to be used instead of certified traceable calibration or verification of data acquisition systems when such certification is required. The output of the fatigue and fracture mechanics data acquisition systems described is essentially a stream of digital data. Such digital data may be considered to be divided into two types: Basic Data, which are a sequence of digital samples of an equivalent analog waveform representing the output of transducers connected to the specimen under test, and Derived Data, which are digital values obtained from the Basic Data by application of appropriate computational algorithms. In its most basic form, a mechanical testing system consists of a test frame with grips which attach to a test specimen, a method of applying forces to the specimen, and a number of transducers which measure the forces and displacements applied to the specimen. The output from these transducers may be in digital or analog form, but if they are analog, they are first amplified and filtered and then converted to digital form using analog-to-digital converters (ADCs). The resulting stream of digital data may be digitally filtered and manipulated to result in a stream of output Basic Data which is presented to the user in the form of a displayed or printed output, or as a data file in a computer. Various algorithms may be applied to the Basic Data to derive parameters representing, for example, the peaks and valleys of the forces and displacements applied to the specimen, or the stresses and strains applied to the specimen and so forth. Such parameters are the Derived Data. The whole measurement system may be divided into three sections for the purpose of verification: the mechanical test frame and its components, the electrical measurement system, and the computer processing of data. oeability, which are performed on the top coat only.
SCOPE
1.1 This guide covers how to understand and minimize the errors associated with data acquisition in fatigue and fracture mechanics testing equipment. This guide is not intended to be used instead of certified traceable calibration or verification of data acquisition systems when such certification is required. It does not cover static load verification, for which the user is referred to the current revision of Practices E 4, or static extensometer verification, for which the user is referred to the current revision of Practice E 83. The user is also referred to Practice E 467.
1.2 The output of the fatigue and fracture mechanics data acquisition systems described in this guide is essentially a stream of digital data. Such digital data may be considered to be divided into two types- Basic Data, which are a sequence of digital samples of an equivalent analog waveform representing the output of transducers connected to the specimen under test, and Derived Data, which are digital values obtained from the Basic Data by application of appropriate computational algorithms. The purpose of this guide is to provide methods that give confidence that such Basic and Derived Data describe the properties of the material adequately. It does this by setting minimum or maximum targets for key system parameters, suggesting how to measure these parameters if their actual values are not known.

General Information

Status
Historical
Publication Date
31-Oct-2010
Current Stage
Ref Project

Relations

Buy Standard

Guide
ASTM E1942-98(2010)e1 - Standard Guide for Evaluating Data Acquisition Systems Used in Cyclic Fatigue and Fracture Mechanics Testing
English language
12 pages
sale 15% off
Preview
sale 15% off
Preview

Standards Content (Sample)


NOTICE: This standard has either been superseded and replaced by a new version or withdrawn.
Contact ASTM International (www.astm.org) for the latest information
´1
Designation: E1942 − 98 (Reapproved 2010)
Standard Guide for
Evaluating Data Acquisition Systems Used in Cyclic Fatigue
and Fracture Mechanics Testing
This standard is issued under the fixed designation E1942; the number immediately following the designation indicates the year of
original adoption or, in the case of revision, the year of last revision. A number in parentheses indicates the year of last reapproval. A
superscript epsilon (´) indicates an editorial change since the last revision or reapproval.
ε NOTE—3.1.4 was editorially revised in December 2011.
1. Scope E467 Practice for Verification of Constant Amplitude Dy-
namic Forces in an Axial Fatigue Testing System
1.1 This guide covers how to understand and minimize the
E1823 TerminologyRelatingtoFatigueandFractureTesting
errors associated with data acquisition in fatigue and fracture
mechanics testing equipment. This guide is not intended to be
3. Terminology
used instead of certified traceable calibration or verification of
data acquisition systems when such certification is required. It
3.1 Definitions:
does not cover static load verification, for which the user is
3.1.1 bandwidth[T ]—thefrequencyatwhichtheamplitude
referred to the current revision of Practices E4, or static
response of the channel has fallen to 1/=2 of its value at low
extensometer verification, for which the user is referred to the
frequency.
current revision of Practice E83. The user is also referred to
3.1.1.1 Discussion—This definition assumes the sensor
Practice E467.
channel response is low-pass, as in most materials testing. An
1.2 The output of the fatigue and fracture mechanics data
illustration of bandwidth is shown in Fig. 1.
acquisition systems described in this guide is essentially a
stream of digital data. Such digital data may be considered to 3.1.2 Basic Data sample—the sampled value of a sensor
be divided into two types– Basic Data, which are a sequence of waveform taken at fixed time intervals. Each sample represents
digital samples of an equivalent analog waveform representing the actual sensor value at that instant of time.
the output of transducers connected to the specimen under test,
3.1.2.1 Discussion—Fig. 2 shows examples of Basic Data
and Derived Data, which are digital values obtained from the
samples.
Basic Data by application of appropriate computational algo-
3.1.3 data rate [T ]—the date rate is ⁄td Hertz where the
rithms. The purpose of this guide is to provide methods that
time intervals between samples is t in seconds.
d
give confidence that such Basic and Derived Data describe the
3.1.3.1 Discussion—The data rate is the number of data
properties of the material adequately. It does this by setting
samples per second made available to the user, assuming the
minimum or maximum targets for key system parameters,
rate is constant.
suggesting how to measure these parameters if their actual
values are not known.
3.1.4 derived data—data obtained through processing of the
raw data.
2. Referenced Documents
3.1.4.1 Discussion—Fig. 2 illustrates examples of Derived
2.1 ASTM Standards:
Data.
E4 Practices for Force Verification of Testing Machines
3.1.5 noise level—thestandarddeviationofthedatasamples
E83 Practice for Verification and Classification of Exten-
of noise in the transducer channel, expressed in the units
someter Systems
appropriate to that channel.
3.1.6 peak—the point of maximum load in constant ampli-
This guide is under the jurisdiction of ASTM Committee E08 on Fatigue and
tude loading (see Terminology E1823).
Fracture and is the direct responsibility of SubcommitteeE08.03 on Advanced
Apparatus and Techniques.
3.1.7 phase difference [°]—the angle in degrees separating
Current edition approved Nov. 1, 2010. Published January 2011. Originally
corresponding parts of two waveforms (such as peaks), where
approved in 1998. Last previous edition approved in 2004 as E1942 - 98(2004).
DOI: 10.1520/E1942-98R10E01 one complete cycle represents 360°.
For referenced ASTM standards, visit the ASTM website, www.astm.org, or
3.1.7.1 Discussion—The phase difference of a cyclic wave-
contact ASTM Customer Service at service@astm.org. For Annual Book of ASTM
form only has meaning in reference to a second cyclic
Standards volume information, refer to the standard’s Document Summary page on
the ASTM website. waveform of the same frequency.
Copyright © ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959. United States
´1
E1942 − 98 (2010)
FIG. 1 3-dB Bandwidth of Sensor Channel FIG. 3 Sources of Error in Data Acquisition Systems
(ADCs). The resulting stream of digital data may be digitally
filtered and manipulated to result in a stream of output Basic
Data which is presented to the user in the form of a displayed
or printed output, or as a data file in a computer. Various
algorithms may be applied to the Basic Data to derive
parameters representing, for example, the peaks and valleys of
the forces and displacements applied to the specimen, or the
stresses and strains applied to the specimen and so forth. Such
parameters are the Derived Data.
4.1.1 The whole measurement system may be divided into
three sections for the purpose of verification: the mechanical
test frame and its components, the electrical measurement
system, and the computer processing of data. This guide is
specifically concerned only with the electrical measurement
system commencing at the output of the transducers. Before
FIG. 2 Basic and Derived Data
themechanicalsystemisinvestigatedfordynamicerrorsbythe
methods given in Practice E467, this guide can be used to
ascertain that the electrical measurement system has adequate
3.1.8 sampling rate [T ]—the rate at which the analog-to-
performance for the measurements required for Practice E467.
digital converter samples a waveform. This rate may not be
If the requirements of Practice E467 for the mechanical system
visible to the user of the data acquisition system.
and the recommendations of this guide are met, then the user
3.1.8.1 Discussion—A distinction is made here between
has confidence that the Basic Data produced by the testing
sampling rate and data rate, because in some data acquisition
system are adequate for processing by subsequent computer
systems, the analog waveform may be sampled at a much
algorithms to produce further Derived Data.
higher rate than the rate at which data are made available to the
4.1.2 At each stage of the flow of data in the electrical
user.(Suchatechniqueiscommonlyknownas over-sampling).
measurement system, errors can be introduced. These should
3.1.9 word size—the number of significant bits in a single
be considered in the sequence in which these are dealt with in
data sample.
this guide. The sequence includes:
3.1.9.1 Discussion—The word size is one parameter which
determines the system resolution. Usually it will be determined 4.2 Errors Due to Bandwidth Limitations in the Signal
by the analog-digital converter used, and typically may be 12 Conditioning—Where there is analog signal conditioning prior
to analog-to-digital conversion, there will usually be restric-
or 16 bits. If the word size is w, then the smallest step change
w
in the data that can be seen is 1 part in 2 , that is the tions on the analog bandwidth in order to minimize noise and,
–w
in some cases, to eliminate products of demodulation. After
quantization step is d=2 .
digital conversion, additional digital filtering may be applied to
3.1.10 valley—The point of minimum load in constant
reduce noise components. These bandwidth restrictions result
amplitude loading (see Terminology E1823).
in cyclic signals at higher frequencies having an apparent
4. Description of a Basic Data Acquisition System amplitude which is lower than the true value, and if the
waveform is not sinusoidal, also having waveform distortion.
4.1 In its most basic form, a mechanical testing system
The bandwidth restrictions also cause phase shifts which result
consists of a test frame with grips which attach to a test
in phase measurement errors when comparing phase in two
specimen, a method of applying forces to the specimen, and a
channels with different bandwidths.
number of transducers which measure the forces and displace-
ments applied to the specimen (see Fig. 3). The output from 4.3 Errors Due to Incorrect Data Rate—Errors can result
these transducers may be in digital or analog form, but if they from an insufficient data rate, where the intervals between data
are analog, they are first amplified and filtered and then samples are too large and intervening events are not recorded
converted to digital form using analog-to-digital converters in the Basic Data. These result also in errors in the Derived
´1
E1942 − 98 (2010)
Data, for example, when the peak value of a waveform is 5.5 Minimum Data Rate—For measurement of the peak
missed during sampling. Data skew, where the Basic Data are value of sinusoidal or square waveforms, the minimum data
not acquired at the same instant in time, can produce similar rate is 50 points/cycle, or 50f points/s. For measurement of the
errors to phase shifts between channels. peak value of triangular waveforms, the minimum data rate is
400 points/cycle, or 400f points/s. If the data acquisition
4.4 Errors Due to Noise and Drift—Noise added to the
system produces the peak value as an output, then the internal
signal being measured causes measurement uncertainty. Short–
Basic Data rate used should equal or exceed the appropriate
term noise causes variability or random error, and includes
minimumdatarate(dependingonwaveformtype).Thisshould
analog noise at the transducer output due to electrical or
be verified even if the external rate at which samples are
mechanical pick up, and analog noise added in the amplifier,
presented is less than this minimum value. For a discussion of
together with digital noise, or quantization, due to the finite
data rate, see A1.3.1.
digital word length of the ADC system.
4.4.1 Long-term effects, such as drifts in the transducer 5.6 Actual Data Rate—The actual data rate must equal or
output or its analog signal conditioning due to temperature or
exceed the minimum data rate. If the actual data rate is not
aging effects, are indistinguishable from slow changes in the known, then it must be ascertained using a method such as that
forces and displacements seen by the specimen, and cause a in A1.3.2.
more systematic error.
5.7 Maximum Permitted Noise Level—The noise level is the
4.4.2 Further details of these sources of error are given in
standard deviation of the noise in the transducer channel,
Annex A1.
expressed in the units appropriate to the channel. The maxi-
mum permitted noise level is 0.2 % of the expected peak value
5. System Requirements
of the waveform being measured. For example, if the expected
5.1 How This Section is Organized—This section gives the
peak value in a load channel is 100 kN, then the standard
steps that must be taken to ensure the errors are controlled. deviation of the noise in that channel must not exceed 0.2 kN.
There are several sources of error in the electrical system, and
5.8 Actual Noise Level—The actual noise level must be
these may add both randomly and deterministically. To give
equal to or less than the maximum permitted noise level. If the
reasonable assurance that these errors have a minor effect on
actual noise level is not known, then it must be ascertained
overall accuracy of a system with 1 % accuracy, recommenda-
using a method such as that in A1.4.6. Guidance on how to
tions are given in this guide, which result in a 0.2 % error
investigate sources of noise is given in A1.4.7.
bound for each individual source of error. However, AnnexA1
5.8.1 If the actual noise level exceeds the maximum permit-
also shows how the error varies with each parameter, so that
ted noise level, it can usually be reduced by reducing
the user may choose to use larger or smaller error bounds with
bandwidth,butthiswillrequirebeginningagainat5.3toverify
appropriate adjustments to bandwidth, data rate, and so forth.
that the bandwidth reduction is permissible.
5.1.1 In this section, which is intended to be used in the
5.9 Maximum Permissible Phase Difference and Maximum
order written, a minimum value or a maximum value is
Permissible Data Skew—These terms are discussed in A1.5.1
recommended for each parameter. If the actual value of each
and A1.5.2. No value is recommended for the maximum
parameterisknown,thenthesystemrequirementisthatineach
permissible phase difference and data skew between channels,
case either:
since this is very dependent on the testing application. If
Maximum value ≥ actual value
typical phase shifts between displacement and force due to the
or
material under test are 10 to 20°, then an acceptable value for
Minimum value ≤ actual value.
themaximumphasedifferencemightbe1°.However,iftypical
However, if the actual value is not known, then help is given
phase shifts are 2 to 3°, the acceptable value for the maximum
as to how to determine it.
phase difference might be only 0.1°.
5.2 Frequency and Waveshape—The first step is to deter-
5.10 Actual Phase Shift and Data Skew—Methods for esti-
mine the highest cyclic frequency, f Hz, at which testing will
mating the combined effect of phase shift and data skew in a
occur, and the waveshape to be employed (for example,
data acquisition system are given in A1.5.3.
sinusoidal, triangular, square).
5.3 Minimum Bandwidth—If the waveform is sinusoidal or
6. Report
square, then the minimum bandwidth is 10f Hz to measure the
6.1 The purpose of the report is to record that due consid-
peak value. If the waveform is triangular, then the minimum
bandwidth is 100f Hz. For example, for a 10–Hz sinusoidal eration was given to essential performance parameters of the
dataacquisitionsystemwhenperformingaparticularfatigueor
waveform,theminimumbandwidthis100Hz.Foradiscussion
of minimum bandwidth, see A1.2.1 and A1.2.2. fracture mechanics test. Since the report should ideally be an
attachment to each set of such test results, it should be
5.4 Actual Bandwidth—The actual bandwidth must be equal
sufficient but succinct. The report should contain the following
to or greater than the minimum bandwidth. If this condition
information, preferably in a tabular format.
cannot be met, then the errors will increase as shown in A1.2.1
andA1.2.2.Iftheactualbandwidthisnotknown,thenitcanbe 6.2 Measurement Equipment Description—This should in-
ascertained using one of the suggested methods in A1.2.3,or clude the manufacturer’s name, model number, and serial
otherwise. number for the test hardware used.
´1
E1942 −
...

Questions, Comments and Discussion

Ask us and Technical Secretary will try to provide an answer. You can facilitate discussion about the standard in here.