Standard Practice for Performance Evaluation and Long-Term Stability of Computed Radiography Systems

SIGNIFICANCE AND USE
4.1 This practice is intended to be used by the NDT using organization to measure baseline performance of the CR system and to monitor its performance throughout its service as an NDT imaging system. For purposes of this document, the CR System is defined as:  
4.1.1 Storage phosphor imaging plate (IP) type and manufacturer,  
4.1.2 Read out unit (scanner or reader) manufacturer and model, including applicable scanner settings (e.g., sampling resolution, PMT gain, pixel value (PV) look up table, etc.),  
4.1.3 Image acquisition and processing software, and  
4.1.4 Image display monitor.  
4.2 It is to be understood that the CR system has already been selected and purchased by the user from a manufacturer based on the inspection needs at hand. The user shall accept the CR scanner based on manufacturer’s results of Practice E2446 on the specific CR scanner as provided in a data sheet for that serialized CR scanner or other acceptance test agreed to between the user and manufacturer (not covered in this practice). This practice is not intended to be used as an “acceptance test” of the CR system, but rather to establish a performance baseline that will enable tracking while in-service.  
4.3 Although many of the properties listed in this standard have similar metrics to those found in E2446, data collection methods are not identical, and comparisons among values acquired with each standard should not be made.  
4.4 This practice defines the tests to be performed and required intervals. Also defined are the methods of tabulating results that CR users will complete following the baseline of the CR system. These tests will also be performed periodically at the stated required intervals to evaluate the CR system to determine if the system remains within acceptable operational limits as established in this practice.  
4.5 There are several factors that affect the image quality of a CR image. Factors which are dependent on the CR system performance include basic spati...
SCOPE
1.1 This practice describes the evaluation of Computed Radiology (CR) systems for industrial radiology. It is intended to ensure that the evaluation of image quality, as far as this is influenced by the CR system, meets the needs of users of this standard, and their customers, and enables process control and long-term stability of the CR system.  
1.2 This practice specifies the fundamental parameters of CR systems to be measured to determine baseline performance, and to track the long term stability of the system. These tests are for applications up to 320kV. When greater than 320kV or when a gamma source is used, these tests may still be used to characterize a system, but may need to be modified as agreed between the user and cognizant engineering organization (CEO).  
1.3 The CR system performance tests specified in this practice shall be completed upon acceptance of the system from the manufacturer and at intervals specified in this practice to monitor long term stability of the system. The intent of these tests is to monitor the system performance degradation and to identify when an action needs to be taken when the system degrades by a certain level.  
1.4 The use of gauges2 provided in this standard is mandatory for each test. In the event these tests or gauges are not sufficient, the user, in coordination with the CEO shall develop additional or modified tests, test objects, gauges, or image quality indicators to evaluate the CR system. Acceptance levels for these ALTERNATE tests shall be determined by agreement between the user and CEO.  
1.5 The values stated in either SI units or inch-pound units are to be regarded separately as standard. The values stated in each system may not be exact equivalents; therefore, each system shall be used independently of the other. Combining values from the two systems may result in non-conformance with the standard.  
1.6 This standard does not purport to address all of t...

General Information

Status
Historical
Publication Date
30-Sep-2014
Current Stage
Ref Project

Buy Standard

Standard
ASTM E2445/E2445M-14 - Standard Practice for Performance Evaluation and Long-Term Stability of Computed Radiography Systems
English language
21 pages
sale 15% off
Preview
sale 15% off
Preview
Standard
REDLINE ASTM E2445/E2445M-14 - Standard Practice for Performance Evaluation and Long-Term Stability of Computed Radiography Systems
English language
21 pages
sale 15% off
Preview
sale 15% off
Preview

Standards Content (Sample)

NOTICE: This standard has either been superseded and replaced by a new version or withdrawn.
Contact ASTM International (www.astm.org) for the latest information
Designation: E2445/E2445M − 14
Standard Practice for
Performance Evaluation and Long-Term Stability of
1
Computed Radiography Systems
ThisstandardisissuedunderthefixeddesignationE2445/E2445M;thenumberimmediatelyfollowingthedesignationindicatestheyear
of original adoption or, in the case of revision, the year of last revision.Anumber in parentheses indicates the year of last reapproval.
A superscript epsilon (´) indicates an editorial change since the last revision or reapproval.
1. Scope 1.5 The values stated in either SI units or inch-pound units
are to be regarded separately as standard. The values stated in
1.1 This practice describes the evaluation of Computed
each system may not be exact equivalents; therefore, each
Radiology (CR) systems for industrial radiology. It is intended
system shall be used independently of the other. Combining
to ensure that the evaluation of image quality, as far as this is
values from the two systems may result in non-conformance
influenced by the CR system, meets the needs of users of this
with the standard.
standard, and their customers, and enables process control and
1.6 This standard does not purport to address all of the
long-term stability of the CR system.
safety concerns, if any, associated with its use. It is the
1.2 This practice specifies the fundamental parameters of
responsibility of the user of this standard to establish appro-
CRsystemstobemeasuredtodeterminebaselineperformance,
priate safety and health practices and determine the applica-
and to track the long term stability of the system. These tests
bility of regulatory limitations prior to use.
are for applications up to 320kV. When greater than 320kV or
when a gamma source is used, these tests may still be used to
2. Referenced Documents
characterize a system, but may need to be modified as agreed
3
2.1 ASTM Standards:
between the user and cognizant engineering organization
E746Practice for Determining Relative Image Quality Re-
(CEO).
sponse of Industrial Radiographic Imaging Systems
1.3 The CR system performance tests specified in this
E1316Terminology for Nondestructive Examinations
practice shall be completed upon acceptance of the system
E1647Practice for Determining Contrast Sensitivity in Ra-
fromthemanufacturerandatintervalsspecifiedinthispractice
diology
tomonitorlongtermstabilityofthesystem.Theintentofthese
E2002PracticeforDeterminingTotalImageUnsharpnessin
tests is to monitor the system performance degradation and to
Radiology
identify when an action needs to be taken when the system
E2007Guide for Computed Radiography
degrades by a certain level.
E2033Practice for Computed Radiology (Photostimulable
2
1.4 The use of gauges provided in this standard is manda-
Luminescence Method)
tory for each test. In the event these tests or gauges are not
E2446Practice for Classification of Computed Radiology
sufficient,theuser,incoordinationwiththeCEOshalldevelop
Systems
additional or modified tests, test objects, gauges, or image
qualityindicatorstoevaluatetheCRsystem.Acceptancelevels
3. Terminology
for theseALTERNATE tests shall be determined by agreement
3.1 Definitions—The definition of terms relating to gamma-
between the user and CEO.
and X-radiology, which appear in Terminology E1316, Guide
E2007,andPracticeE2033shallapplytothetermsusedinthis
1
This practice is under the jurisdiction of ASTM Committee E07 on Nonde-
practice.
structive Testing and is the direct responsibility of Subcommittee E07.01 on
Radiology (X and Gamma) Method. 3.2 Definitions of Terms Specific to This Standard:
Current edition approved Oct. 1, 2014. Published October 2014. Originally
3.2.1 aliasing—artifacts that appear in an image when the
approved in 2005. Last previous edition approved in 2010 as E2445/E2445M-
spatial frequency of the input is higher than the output is
05(2010). DOI:10.1520/E2445_E2445M-14.
2
capable of reproducing.
ThesolesourceofsupplyoftheapparatusshowninAppendixX2knowntothe
committee at this time is Rockwell Collins’ARINC, 1300 Thomas Drive, Panama
City Beach, FL 32408, Phone: 405-605-7095, ARINC part number A0295224002
(USAF design) or A0295224003 (NAVAIR design). The NAVAIR design includes
3
two additional test targets that are not used in this test standard. If you are aware of For referenced ASTM standards, visit the ASTM website, www.astm.org, or
alternative suppliers, please provide this information to ASTM International contact ASTM Customer Service at service@astm.org. For Annual Book of ASTM
Headquarters.Your comments will receive careful consideration at a meeting of the Standards volume information, refer to the standard’s Document Summary page on
1
responsible technical committee, which you may attend. the ASTM website.
Copyright © ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959. United States
1

----
...

This document is not an ASTM standard and is intended only to provide the user of an ASTM standard an indication of what changes have been made to the previous version. Because
it may not be technically possible to adequately depict all changes accurately, ASTM recommends that users consult prior editions as appropriate. In all cases only the current version
of the standard as published by ASTM is to be considered the official document.
Designation: E2445 − 05 (Reapproved 2010) E2445/E2445M − 14
Standard Practice for
Qualification Performance Evaluation and Long-Term
1
Stability of Computed RadiologyRadiography Systems
This standard is issued under the fixed designation E2445;E2445/E2445M; the number immediately following the designation indicates
the year of original adoption or, in the case of revision, the year of last revision. A number in parentheses indicates the year of last
reapproval. A superscript epsilon (´) indicates an editorial change since the last revision or reapproval.
1. Scope
1.1 This practice specifies the fundamental parameters of computed radiography systems to assure satisfactory and repeatable
results for nondestructive testing.
1.1 This practice describes the evaluation of Computed Radiology (CR) systems for industrial radiography.radiology. It is
intended to ensure that the evaluation of image quality, as far as this is influenced by the scanner/IPCR system, meets the needs
of users and enables the test of long-term stability. of this standard, and their customers, and enables process control and long-term
stability of the CR system.
1.2 This practice specifies the fundamental parameters of CR systems to be measured to determine baseline performance, and
to track the long term stability of the system. These tests are for applications up to 320kV. When greater than 320kV or when a
gamma source is used, these tests may still be used to characterize a system, but may need to be modified as agreed between the
user and cognizant engineering organization (CEO).
1.3 The CR system performance tests specified in this practice shall be completed upon acceptance of the system from the
manufacturer and at intervals specified in this practice to monitor long term stability of the system. The intent of these tests is to
monitor the system performance degradation and to identify when an action needs to be taken when the system degrades by a
certain level.
1.4 Each of the tests described may be performed with individual gages specified. The user shall decide which tests shall be used
2
for system control using individual test objects or the CR test phantomThe use of gauges (providedAppendix X1). The computed
radiological tests, specified as “user tests” in this practice, may be utilized at appropriate intervals determined by the user, based
on the application of the examination operations. The in this standard is mandatory for each test. In the event these tests or gauges
are not sufficient, the user, in coordination with the CEO shall develop additional or modified tests, test objects, gauges, or image
quality indicators to evaluate the CR system. Acceptance levels for these ALTERNATE tests shall be appropriate for the materials
and range of use of the system. Fading, uniformity, and erasure tests shall also be part of the control system. All other tests for
qualification and capability are to be performed and certified by the CR equipment manufacturer. determined by agreement
between the user and CEO.
1.5 The values stated in either SI units or inch-pound units are to be regarded as the standard. Values in inch-pound units are
for information purposes.separately as standard. The values stated in each system may not be exact equivalents; therefore, each
system shall be used independently of the other. Combining values from the two systems may result in non-conformance with the
standard.
1.6 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility
of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory
limitations prior to use.
1
This practice is under the jurisdiction of ASTM Committee E07 on Nondestructive Testing and is the direct responsibility of Subcommittee E07.01 on Radiology (X and
Gamma) Method.
Current edition approved June 1, 2010Oct. 1, 2014. Published November 2010October 2014. Originally approved in 2005. Last previous edition approved in 20052010
as E2445 - 05.E2445/E2445M-05(2010). DOI:10.1520/E2445–05R10. DOI:10.1520/E2445_E2445M-14.
2
The sole source of supply of the apparatus shown in Appendix X2 known to the committee at this time is Nuclear Associates, A Division of Cardinal Health, 120 Andrews
Road, Hicksville, NY 11801, Phone: 1-888-466-8257, Catalog Number: 07-605-2435. Rockwell Collins’ARINC, 1300 Thomas Drive, Panama City Beach, FL 32408, Phone:
405-605-7095, ARINC part number A0295224002 (USAF design) or A0295224003 (NAVAIR design). The NAV
...

Questions, Comments and Discussion

Ask us and Technical Secretary will try to provide an answer. You can facilitate discussion about the standard in here.