Standard Practice for Use of a Radiochromic Optical Waveguide Dosimetry System

SIGNIFICANCE AND USE
4.1 The radiochromic optical waveguide dosimetry system provides a means of measuring absorbed dose in materials. Under the influence of ionizing radiation such as photons, chemical reactions take place in the radiochromic optical waveguide creating and/or modifying optical absorbance bands in the visible region of the spectrum. Optical response is determined at selected wavelengths using the equations in 3.1.6. Examples of appropriate wavelengths for the analysis for specific dosimetry systems are provided by their manufacturers and in Refs (1) through (5).
4.2 In the application of a specific dosimetry system, absorbed dose is determined by use of a calibration curve traceable to national or international standards.
4.3 The absorbed dose determined is usually specified in water. Absorbed dose in other materials may be determined by applying the conversion factors discussed in ISO/ASTM Guide 51261.
NOTE 1—For a comprehensive discussion of various dosimetry methods applicable to the radiation types and energies discussed in this practice, see ICRU Reports 14, 17, and 34.
4.4 These dosimetry systems commonly are applied in the industrial radiation processing of a variety of products, for example, the sterilization of medical devices and radiation processing of foods (4-6).
SCOPE
This practice covers the procedures for handling, testing, and using a radiochromic optical waveguide dosimetry system to measure absorbed dose in materials irradiated by photons in terms of absorbed dose in water.
1.2 This practice applies to radiochromic optical waveguide dosimeters that can be used within part or all of the specified ranges as follows:
1.2.1 The absorbed dose range is from 1 to 10 000 Gy for photons.
1.2.2 The absorbed dose rate is from 0.001 to 1000 Gy/s.
1.2.3 The radiation energy range for photons is from 0.1 to 10 MeV.
1.2.4 The irradiation temperature range is from –78 to +60°C.
1.3 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use.

General Information

Status
Historical
Publication Date
20-Mar-2012
Technical Committee
Drafting Committee
Current Stage
Ref Project

Buy Standard

Standard
ASTM ISO/ASTM51310-04(2012) - Standard Practice for Use of a Radiochromic Optical Waveguide Dosimetry System
English language
5 pages
sale 15% off
Preview
sale 15% off
Preview

Standards Content (Sample)


NOTICE: This standard has either been superseded and replaced by a new version or withdrawn.
Contact ASTM International (www.astm.org) for the latest information
ISO/ASTM 51310:2004 (Reapproved 2012)(E)
Standard Practice for
Use of a Radiochromic Optical Waveguide Dosimetry
System
This standard is issued under the fixed designation ISO/ASTM 51310; the number immediately following the designation indicates the
year of original adoption or, in the case of revision, the year of last revision.
1. Scope E925 Practice for Monitoring the Calibration of Ultraviolet-
Visible Spectrophotometers whose Spectral Bandwidth
1.1 This practice covers the procedures for handling,
does not Exceed 2 nm
testing, and using a radiochromic optical waveguide dosimetry
E958 Practice for Estimation of the Spectral Bandwidth of
system to measure absorbed dose in materials irradiated by
Ultraviolet-Visible Spectrophotometers
photons in terms of absorbed dose in water.
E1026 Practice for Using the Fricke Dosimetry System
1.2 This practice applies to radiochromic optical waveguide
2.2 ISO/ASTM Standards:
dosimeters that can be used within part or all of the specified
51261 Guide for Selection and Calibration of Dosimetry
ranges as follows:
Systems for Radiation Processing
1.2.1 The absorbed dose range is from 1 to 10 000 Gy for
51400 Practice for Characterization and Performance of a
photons.
High-Dose Radiation Dosimetry Calibration Laboratory
1.2.2 The absorbed dose rate is from 0.001 to 1000 Gy/s.
51707 Guide for Estimating Uncertainties in Dosimetry for
1.2.3 The radiation energy range for photons is from 0.1 to
Radiation Processing
10 MeV.
2.3 International Commission on Radiation Units and Mea-
1.2.4 The irradiation temperature range is from –78 to
surements (ICRU) Reports:
+60°C.
ICRU Report 14 Radiation Dosimetry: X-Rays and Gamma
1.3 This standard does not purport to address all of the
RayswithMaximumPhotonEnergiesBetween0.6and50
safety concerns, if any, associated with its use. It is the
MeV
responsibility of the user of this standard to establish appro-
ICRUReport17 RadiationDosimetry:X–RaysGeneratedat
priate safety and health practices and determine the applica-
Potentials of 5 to 150 kV
bility of regulatory limitations prior to use.
ICRU Report 34 The Dosimetry of Pulsed Radiation
ICRU Report 60 Fundamental Quantities and Units for
2. Referenced documents
Ionizing Radiation
2.1 ASTM Standards:
3. Terminology
E170 Terminology Relating to Radiation Measurements and
Dosimetry
3.1 Definitions:
E275 PracticeforDescribingandMeasuringPerformanceof
3.1.1 analysis wavelength—wavelength used in a spectro-
Ultraviolet and Visible Spectrophotometers
photometric instrument for the measurement of optical absor-
E668 Practice for Application of Thermoluminescence-
bance or reflectance.
Dosimetry (TLD) Systems for Determining Absorbed
3.1.2 calibration curve—graphical representation of the do-
DoseinRadiation-HardnessTestingofElectronicDevices
simetry system’s response function.
3.1.3 dosimeter batch—quantity of dosimeters made from a
specific mass of material with uniform composition, fabricated
This guide is under the jurisdiction of ASTM Committee E61 on Radiation
in a single production run under controlled, consistent condi-
Processing and is the direct responsibility of Subcommittee E61.02 on Dosimetry
Systems, and is also under the jurisdiction of ISO/TC 85/WG 3. tions and having a unique identification code.
CurrenteditionapprovedMarch21,2012.PublishedNovember2012.Originally
3.1.4 dosimetry system—system used for determining ab-
ε1
published as ASTM E 1310–89. Last previous ASTM edition E 1310–98 . ASTM
sorbed dose, consisting of dosimeters, measurement instru-
E 1310–94 was adopted by ISO in 1998 with the intermediate designation ISO
15559:1998(E). The present International Standard ISO/ASTM
ments and their associated reference standards, and procedures
51310:2004(2012)(E) replaces ISO 15559 and is a reapproval of the last previous
for the system’s use.
edition ISO/ASTM 51310:2004(E).
For referenced ASTM and ISO/ASTM standards, visit the ASTM website,
www.astm.org, or contact ASTM Customer Service at service@astm.org. For
Annual Book of ASTM Standards volume information, refer to the standard’s Available from the International Commission on Radiation Units and
Document Summary page on the ASTM website. Measurements, 7910 Woodmont Ave., Suite 800, Bethesda, MD 20814, U.S.A.
© ISO/ASTM International 2022 – All rights reserved
ISO/ASTM 51310:2004 (2012)(E)
3.1.5 measurement quality assurance plan—documented 4.2 In the application of a specific dosimetry system,
program for the measurement process that ensures on a absorbed dose is determined by use of a calibration curve
continuing basis that the overall uncertainty meets the require- traceable to national or international standards.
mentsofthespecificapplication.Thisplanrequirestraceability
4.3 The absorbed dose determined is usually specified in
to, and consistency with, nationally or internationally recog-
water.Absorbed dose in other materials may be determined by
nized standards.
applyingtheconversionfactorsdiscussedinISO/ASTMGuide
3.1.6 net response, ∆R— radiation–induced change in the 51261.
relationship of measured absorbance at a specific wavelength
NOTE1—Foracomprehensivediscussionofvariousdosimetrymethods
determined by subtracting the pre–irradiation response, R ,
applicable to the radiation types and energies discussed in this practice,
from the post–irradiation response, R:
see ICRU Reports 14, 17, and 34.
∆R5 R 2 R (1)
4.4 These dosimetry systems commonly are applied in the
industrial radiation processing of a variety of products, for
with:
example, the sterilization of medical devices and radiation
A
λ processing of foods (4-6).
R 5
A
λref
Aλ
R (2) 5. Apparatus
0 5 0
F G
A
λref
5.1 The following shall be used to determine absorbed dose
and where:
with radiochromic optical waveguide dosimetry systems:
A = optical absorbance at the analysis wavelength,λ, and
λ
5.1.1 Dosimeters—A batch or portion of a batch of ra-
A = optical absorbance at a reference wavelength, λ .
λref ref
diochromic optical waveguide dosimeters.
3.1.7 optical waveguide—device that contains an optical
5.1.2 Spectrophotometer or Photometer—An instrument,
path at a high index of refraction relative to the material either a spectrophotometer equipped with a special dosimeter
enclosing the optical path.
holder and associated coupling optics (see Ref 7 for an
example), or a modified photometer (see Fig. 1 for a block
3.1.8 radiochromic optical waveguide—specially prepared
diagram of an instrument that uses a reference wavelength),
optical waveguide containing ingredients that undergo an
having documentation covering analysis wavelengths, accu-
ionizing radiation–induced change in photometric absorbance.
racy of wavelength selection, absorbance determination, spec-
This change in absorbance can be related to absorbed dose in
tral bandwidth, and stray light rejection.
water (1, 2).
5.1.3 Holder, to position the dosimeter reproducibly in the
3.1.9 reference wavelength, λ —wavelength selected for
ref
measuring light beam.
comparison with the analysis wavelength. This wavelength is
chosen to minimize effects associated with optical coupling
6. Performance check of instrumentation
and other geometric variations in the dosimeter.
6.1 Checkanddocumenttheperformanceofthephotometer
3.1.10 response function—mathematical representation of
or spectrophotometer (seeASTM Practices E275, E925, E958,
therelationshipbetweendosimeterresponseandabsorbeddose
and E1026). Use reference standards traceable to national or
for a given dosimetry system.
international standards, unless the photometer’s or spectropho-
3.2 Definitions or other terms used in this standard that tometer’s design precludes such use.
pertain to radiation measurement and dosimetry may be found 6.1.1 When using a photometer, check and document the
inASTM Terminology E170. Definitions in E170 are compat- accuracy of the absorbance scale at intervals not to exceed one
iblewithICRU60;thatdocument,therefore,maybeusedasan month during periods of use, or whenever there are indications
alternative reference.
of poor performance.
4. Significance and use
4.1 The radiochromic optical waveguide dosimetry system
provides a means of measuring absorbed dose in materials.
Under the influence of ionizing radiation such as photons,
chemical reactions take place in the radiochromic optical
waveguidecreatingand/ormodifyingopticalabsorbancebands
in the visible region of the spectrum. Optical response is
determined at selected wavelengths using the equations in
3.1.6.Examplesofappropriatewavelengthsfortheanalysisfor
specificdosimetrysystemsareprovidedbytheirmanufacturers
and in Refs (1-5).
The boldface numbers in parentheses refer to the bibliography at the end of this
practice. FIG. 1Block Diagram of the Instrument Described in Section 5
© ISO/ASTM International 2018 – All rights reserved
ISO/ASTM 51310:2004 (2012)(E)
6.1.2 Whenusingaspectrophotometer,checkanddocument 8.1.3 Visually inspect the dosimeters for imperfections (for
the precision and bias of the wavelength scale and absorbance example, loss of end fittings). Discard any dosimeters that
scale at or near the selected analysis wavelength(s) at intervals show imperfections.
not to exceed one month during periods of use, or whenever 8.1.4 Identify the dosimeters with an appropriate code that
there are indications of poor performance. can be related to the manufacturer, type, and batch.
6.1.3 Document the comparison of information obtained in 8.1.5 Store the dosimeters in accordance with the manufac-
turer’s written recommendations.
6.1.1 or 6.1.2 with the original instrument specification to
verify adequate performance.
8.2 Irradiation Procedure:
8.2.1 Determine the pre-irradiation response, R , for each
7. Calibration of the dosimetry system
dosimeter at the selected analysis wavelength(s). This may be
done for each dosimeter or by use of an average R determined
7.1 Prior to use, the dosimetry system (consisting of a
by reading several dosimeters and documenting the
specific batch of dosimeters and specific measurement instru-
uncertainty, provided this practice meets the precision require-
ments) shall be calibrated in accordance with the user’s
ments for the application.
documented procedure that specifies details of the calibration
8.2.2 Where necessary, package the dosimeters in a UV-
process and quality assurance requirements. This calibration
opaque material.
process shall be repeated at regular intervals to ensure that the
8.2.3 Mark the packaged dosimeters appropriately for iden-
accuracy of the absorbed dose measurement is maintained
tification.
within required limits. Calibration methods are described in
8.2.4 Irradiate the dosimeters.
ISO/ASTM Guide 51261.
NOTE 3—The dosimeters may be irradiated in the product undergoing
7.2 Calibration of Dosimeters—Irradiationisacriticalcom-
processingorinamediumofsimilarcomposition,orwater,ofappropriate
ponent of the calibration of the dosimetry system. Calibration
dimensions so as to approximate electron equilibrium conditions. Such
shall be performed in one of three ways by irradiating the
equilibrium conditions may not exist within dosimeters placed throughout
dosimeters at:
the product under actual processing conditions. This particularly is the
7.2.1 an accredited calibration laboratory that provides an case near interfaces of different materials. Irradiation under nonequilib-
rium conditions, such as on the surface of a product package, is often used
absorbed dose (or an absorbed-dose rate) having measurement
to monitor the absorbed dose delivered to the product and may be related
traceability to nationally or internationally recognized
totheabsorbeddosewithintheproductbycorrectionfactorsundercertain
standards, or
conditions.
7.2.2 an in-house calibration facility that provides an ab-
8.3 Analysis Procedure:
sorbed dose (or an absorbed-dose rate) having measurement
8.3.1 Avoid any exposure to stray ultraviolet radiation that
traceability to nationally or internationally recognized
may induce coloration of the dosimeter (see 8.1.1).
standards, or
8.3.2 Determine the post-irradiation response, R,atthe
7.2.3 a production or research irradiation facility together
selected analysis wavelength(s) used for calibration of the
with reference or transfer standard dosimeters that have mea-
dosimetry system.
surement traceability to nationally or internationally recog-
8.3.3 Calculate the net response, ∆ R, as follows:
nized standards.
∆R 5 R 2 R (3)
7.3 When the optical waveguide dosimeter is used as a
8.3.4 Determine the absorbed dose from the calibration
transfer standard dosimeter, the calibration irradiation may be
curve or response function.
performed only as stated in 7.2.1,orin 7.2.2 at a facility that
meets the requirements in ISO/ASTM Practice 51400.
9. Characterization of each batch of dosimeters
7.4 Measurement Instrument Calibration and Performance
9.1 Reproducibility of Net Response:
Verification—For the calibration of the instruments, and for the
9.1.1 Determine the reproducibility of net response for each
verification of instrument performance between calibrations,
batch of dosimeters by analyzing the data from the sets of
see ISO/ASTM Guide 51261 and/or instrument-specific oper-
dosimeters irradiated during the calibration process at each
ating manuals.
dose value.
9.1.2 Use the sample standard deviation (S ) determined
8. Procedure
n-1
during calibration to calculate the coefficient of variation (CV)
8.1 Examination and Storage Procedure:
for each dose value as follows:
8.1.1 Exposure to ultraviolet (UV) radiation may cause the
S
n21
dosimeter to change color. Perform tests to ensure that the
CV 5 100 3 (4)
F G
∆R
handling and reading environment does not cause measurable
color development. If needed, place UVfilters over fluorescent
9.1.3 Document these coefficients of variation and note any
lights or windows to reduce color development.
that are unusually large.
NOTE 2—Dosimeters may be stored in UV–opaque material to further
NOTE 4—In general, if the value of the coefficient of variation is greater
avoid the effects noted in 8.1.1.
than62 %, then a re-determination of the data should be considered or, in
the extreme, the batch should be rejected.
8.1.2 Handle the dosimeter along the sides, never at the
ends. Handling should be kept to a minimum. 9.2 Effect of Absorbed Dose Rate:
© ISO/ASTM International 2018 – All rights reserved
ISO/ASTM 51310:2004 (2012)(E)
9.2.1 The shape (slope) of the calibration curve associated 11. Minimum documentation
with some radiochromic optical waveguide dosimeters may be
11.1 Record the dosimeter manufacturer, type, batch
affected by the absorbed dose rate
...

Questions, Comments and Discussion

Ask us and Technical Secretary will try to provide an answer. You can facilitate discussion about the standard in here.