Standard Practice for Internal Temperature Measurements in Low-Conductivity Materials

SCOPE
1.1 This practice covers methods for instrumenting low-conductivity specimens for testing in an environment subject to rapid thermal changes such as produced by rocket motors, atmospheric re-entry, electric-arc plasma heaters, etc. Specifically, practices for bare-wire thermocouple instrumentation applicable to sheath-type thermocouples are discussed.  
1.2 The values stated in inch-pound units are to be regarded as the standard. The metric equivalents of inch-pound units may be approximate.  
1.3 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use.

General Information

Status
Historical
Publication Date
09-Oct-1996
Current Stage
Ref Project

Relations

Buy Standard

Standard
ASTM E377-96 - Standard Practice for Internal Temperature Measurements in Low-Conductivity Materials
English language
4 pages
sale 15% off
Preview
sale 15% off
Preview

Standards Content (Sample)


NOTICE: This standard has either been superseded and replaced by a new version or discontinued.
Contact ASTM International (www.astm.org) for the latest information.
Designation: E 377 – 96
Standard Practice for
Internal Temperature Measurements in Low-Conductivity
Materials
This standard is issued under the fixed designation E 377; the number immediately following the designation indicates the year of
original adoption or, in the case of revision, the year of last revision. A number in parentheses indicates the year of last reapproval. A
superscript epsilon (e) indicates an editorial change since the last revision or reapproval.
1. Scope and lead wires, (3) proper selection of thermocouple type and
size, corresponding to both the temperature range to be
1.1 This practice covers methods for instrumenting low-
measured and the chemical compatibility with the environ-
conductivity specimens for testing in an environment subject to
ment, and (4) proper use of instrumentation for readout of
rapid thermal changes such as produced by rocket motors,
thermocouple emf.
atmospheric re-entry, electric-arc plasma heaters, etc. Specifi-
cally, practices for bare-wire thermocouple instrumentation
NOTE 1—Reader is referred to ASTM MNL 12, and STP 492, as well
applicable to sheath-type thermocouples are discussed. as Kinzie, P.A., Thermocouple Temperature Measurement, for needed
information.
1.2 The values stated in inch-pound units are to be regarded
as the standard. The metric equivalents of inch-pound units
3.2 The most important sources of error beyond the above
may be approximate.
basic areas are the following:
1.3 This standard does not purport to address all of the
3.2.1 The thermal disturbance produced in the low-
safety concerns, if any, associated with its use. It is the
conductivity material at the vicinity of the thermocouple sensor
responsibility of the user of this standard to establish appro-
hot junction due to the sensor size, configuration, and instal-
priate safety and health practices and determine the applica-
lation method.
bility of regulatory limitations prior to use.
3.2.2 Electrical shorting of lead wires due to the electrical
conductivity of the virgin or charred ablation material, and
2. Significance and Use
3.2.3 Thermocouple sensor hot junction location accuracy.
2.1 Internal temperature measurements are made on both
in-flight vehicles and on-ground test specimens; and, because 4. Thermal Disturbance at Vicinity of Thermocouple
Sensor Hot Junction
of the importance of the temperature measurements to the
design of various missile and spacecraft heat shields, it is
4.1 General—Ideally, to measure the true internal tempera-
essential that care be taken to minimize the sources of error in
ture of a solid body, it would be desirable not to have any
obtaining these measurements.
foreign substance present that would create a disturbance
2.2 Over the past several years, the problems of using
affecting the natural flow of heat in the body. Since it is
thermocouples to obtain accurate temperature measurements in
physically impossible to exclude the temperature sensor from
low-conductivity specimens have been studied by various
the internal confines of the body, it is necessary that the thermal
people to isolate the sources of error and to establish improved
disturbance introduced by the sensor be minimized for accurate
temperature measurement techniques. The major sources of
temperature measurements.
error are listed in this document and recommended solutions to
4.2 Thermocouple Junction Bead Diameter:
the problems are given.
4.2.1 General—Excessively large junction beads result in
lower than true temperature measurements in low-conductivity
3. General
materials (conductivity of material less than conductivity of
3.1 Before proceeding to the major sources of error, it is
thermocouple wire) because of the heat sink effect of the bead.
assumed that the reader is familiar with basic methods of
4.2.2 Recommendations—To minimize this effect, the junc-
forming and using thermocouples , that is (1) electric welding
tion bead diameter should be no larger than 1.5 wire diameters
to form junctions, (2) maintaining cleanliness of junction area
for butt-welded junctions and 2 wire diameters for other types
of welds.
4.3 Thermocouple Wire in Isothermal Surface of Hot Junc-
This practice is under the jurisdiction of ASTM Committee E-21 on Space
tion:
Simulation and Applications of Space Technology and is the direct responsibility of
Subcommittee E21.08 on Thermal Protection.
4.3.1 General—Because many materials have low thermal
Current edition approved Oct. 10, 1996. Published December 1996. Originally
conductivity compared with those of thermocouple assemblies,
published as E 377 – 68. Last previous edition E 377 – 68 (1992).
2 it has been found that certain methods of installing sensors can
ANSI MC96.1-1975. Temperature Measurement Thermocouples (Sponsor
produce significant errors in internal temperature measurement
ISA).
Copyright © ASTM, 100 Barr Harbor Drive, West Conshohocken, PA 19428-2959, United States.
E 377
NOTE 1—If a number of thermocouples in depth are required, drill holes at varying locations on the circumference.
NOTE 2—Eliminate air pockets in junction plane by filling hole with same or similar compound as that used to make test specimen.
NOTE 3—This is a schematic representation and is not intended to be an engineering drawing.
FIG. 1 Summary of Recommended Practices for Mounting Thermocouples—Schematic Representation for “One-Piece” Cylindrical
Specimen
(1-4). Errors of several hundred degrees are possible unless specimen (thus eliminating air pockets) with the same or
heat conduction away from the sensor hot junction, by the similar compound (such as an epoxy plastic) as that used to
sensor materials, is minimized. Test results show that a make the specimen.
thermocouple having a sufficient length of bare wire in the
5. Electrical Shorting by Conductive Char Layers
isothermal surface that includes the junction will minimize
5.1 General—The char layer formed by most organic ma-
these errors.
4.3.2 Recommendations—It is therefore recommended that terials becomes highly conductive (electrically) as pyrolysis
progresses. Care should be taken to avoid the possibility of
the configuration of the thermocouple sensor be such that the
leads perpendicular to the heat flow have a length equivalent to electrical shorting of thermocouple lead wires not protected by
proper insulation methods. Studies (1) have shown that short-
at least 25 wire diameters on both sides of the junction in the
same isothermal surface that includes the hot junction. ing can result in temperature errors of as much as 110 C (200
F) in thermocouples which do not emplo
...