Standard Practice for Installing Fused Silica Open Tubular Capillary Columns in Gas Chromatographs

SIGNIFICANCE AND USE
5.1 This practice is intended to be used by all analysts using fused silica capillary chromatography. It contains the recommended steps for installation, preparation, proper installation, and continued column maintenance.
SCOPE
1.1 This practice covers the installation and maintenance of fused silica capillary columns in gas chromatographs that are already retrofitted for their use. This practice excludes information on:  
1.1.1 Injection techniques.  
1.1.2 Column selection.  
1.1.3 Data acquisition.  
1.1.4 System troubleshooting and maintenance.  
1.2 For additional information on gas chromatography, please refer to Practice E260. For specific precautions, see 7.2.2.2(1), 7.2.2.2(2), 7.2.7, and 7.2.7.2.  
1.3 The values stated in SI units are to be regarded as standard. No other units of measurement are included in this standard.  
1.4 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use. For specific safety information, see Section 6, 7.2.2.2(1), 7.2.2.2(2), 7.2.7, and 7.2.7.2.2TABLE 1 Typical Splitter Vent Flow Rates (50 to 1 split ratio) (at optimum linear velocity)    
Carrier gas  
0.25-mm ID,
cm3 /min  
0.32-mm ID,
cm3 /min  
0.53-mm ID,
cm3 /min    
helium
hydrogen  
35
70  
80
160  
125
250
Note 1—The curves were generated by plotting the height equivalent to a theoretical plate (length of column divided by the total number of theoretical plates, H.E.T.P.) against the column's average linear velocity. The lowest point on the curve indicates the carrier gas velocity in which the highest column efficiency is reached.Note 2—Gases information available from Compressed Gas Association (CGA), 4221 Walney Rd., 5th Floor, Chantilly, VA 20151-2923, http://www.cganet.com.FIG. 1 Van Deemter Profile for Hydrogen, Helium, and Nitrogen Carrier Gases    
Carrier gas: Hydrogen  
Carrier gas: Helium  
Linear velocity: 40 cm/s  
Linear velocity: 20 cm/s
Note 1—Fig. 2 shows that the resolution is similar but the analysis time is reduced by 50 % when comparing hydrogen to helium in an isothermal analysis using optimum flow velocities.Note 2—Hydrogen provides similar resolution in one-half the analysis time of helium for an isothermal analysis.Note 3—    
1. Tetrachloro-m-
8. Heptachlor epoxide  
15. Endosulfan II  
xylene  
9. γ-chlordane  
16. DDD  
2. α-BHC  
10. Endosulfan I  
17. Endrin aldehyde  
3. β-BHC  
11. α-chlordane  
18. Endosulfan sulfate  
4. γ-BHC  
12. Dieldrin  
19. DDT  
5. δ-BHC  
13. DDE  
20. Endrin ketone  
6. Heptachlor  
14. Endrin  
21. Methyoxychlor  
7. Aldrin  
Note 4—30 m, 0.25-mm ID, 0.25 μm 5 % diphenyl − 95 % dimethyl polysiloxane 0.1-μL split injection of chlorinated pesticides.    
Oven temperature: 210°C isothermal  
Injector and detector temperature: 250°C/300°C  
ECD sensitivity: 512 × 10 −11    
Split vent: 100 cm 3 /min  
FIG. 2 Hydrogen Versus Helium (Isothermal Analysis)
FIG. 3 Capping Silanol Groups with Dimethyl Dichlorosilane (DMDCS)  
Note 1—Septum bleed can obscure or co-elute with compounds of interest, thus decreasing the analytical accuracy.Note 2—    
1. 2,4,5,6-tetrachloro-  
8. Heptachlor
16. p,p-DDD  
m-xylene (IS)  
epoxide  
17. Endrin aldehyde  
2. α-BHC  
9. γ-chlordane  
18. Endosulfan sul-  
3. β-BHC  
10. Endosulfan I  
fate  
4. γ-BHC  
11. α-chlordane  
19. p,p-DDT  
5. δ-BHC  
12. Dieldrin  
20. Endrin ketone  
6. Heptachlor  
13. p,p-DDE  
21. Methyoxychlor  
7. Aldrin  
14. Endrin  
22. Decachlorobi-  
15. Endosulfan II  
phenyl (IS)
Note 3—30 m, 0.5...

General Information

Status
Historical
Publication Date
31-Dec-2012
Current Stage
Ref Project

Buy Standard

Standard
ASTM E1510-95(2013)e1 - Standard Practice for Installing Fused Silica Open Tubular Capillary Columns in Gas Chromatographs
English language
13 pages
sale 15% off
Preview
sale 15% off
Preview

Standards Content (Sample)


NOTICE: This standard has either been superseded and replaced by a new version or withdrawn.
Contact ASTM International (www.astm.org) for the latest information
´1
Designation:E1510 −95 (Reapproved 2013)
Standard Practice for
Installing Fused Silica Open Tubular Capillary Columns in
Gas Chromatographs
This standard is issued under the fixed designation E1510; the number immediately following the designation indicates the year of
original adoption or, in the case of revision, the year of last revision.Anumber in parentheses indicates the year of last reapproval.A
superscript epsilon (´) indicates an editorial change since the last revision or reapproval.
ε NOTE—Warning statements were editorially corrected in January 2013.
1. Scope 2.2 CGA Publications:
CGAP-1SafeHandlingofCompressedGasesinContainers
1.1 This practice covers the installation and maintenance of
CGA G-5.4Standard for Hydrogen Piping Systems at Con-
fused silica capillary columns in gas chromatographs that are
sumer Locations
already retrofitted for their use. This practice excludes infor-
CGA P-9The Inert Gases: Argon, Nitrogen and Helium
mation on:
CGA V-7Standard Method of Determining Cylinder Valve
1.1.1 Injection techniques.
Outlet Connections for Industrial Gas Mixtures
1.1.2 Column selection.
CGA P-12Safe Handling of Cryogenic Liquids
1.1.3 Data acquisition.
HB-3Handbook of Compressed Gases
1.1.4 System troubleshooting and maintenance.
3. Terminology
1.2 For additional information on gas chromatography,
please refer to Practice E260. For specific precautions, see
3.1 Terms and relations are defined in Practice E355.
7.2.2.2(1), 7.2.2.2(2), 7.2.7, and 7.2.7.2.
3.2 Nomenclature for open tubular or capillary columns
1.3 The values stated in SI units are to be regarded as
with a bore of 0.75 mm or less:
standard. No other units of measurement are included in this
3.3 porous layer open tubular (PLOT)—refers to columns
standard.
with particles attached on the inside wall consisting of copo-
1.4 This standard does not purport to address all of the
lymers such as styrene/divinylbenzene, molecular sieves, or
safety concerns, if any, associated with its use. It is the
adsorbents such as Al O in film thicknesses of 5 to 50 µm.
2 2
responsibility of the user of this standard to establish appro-
3.4 support coated open tubular (SCOT)—refers to fine
priate safety and health practices and determine the applica-
particles (silica or fine diatomite) coated with liquid stationary
bility of regulatory limitations prior to use. For specific safety
phase, which is then deposited on the inside column wall to
information, see Section 6, 7.2.2.2(1), 7.2.2.2(2), 7.2.7, and
improve stationary phase stability and sample capacity.
7.2.7.2.
3.5 wall coated open tubular (WCOT)—refers to columns
coated on the inside wall with a liquid stationary phase in film
2. Referenced Documents
thicknesses of 0.1 to 10.0 µm. Also referred to as FSOT or
2.1 ASTM Standards:
fused silica open tubular.
E260Practice for Packed Column Gas Chromatography
E355PracticeforGasChromatographyTermsandRelation-
4. Summary of Practice
ships
4.1 The packed gas chromatography system is described in
Practice E260 and is essentially the same as a capillary gas
chromatographysystemexceptformodificationstotheinjector
This practice is under the jurisdiction ofASTM Committee E13 on Molecular
and detector to accommodate the low flow rates and sample
Spectroscopy and Separation Science and is the direct responsibility of Subcom-
capacity associated with capillary columns. Refer to the gas
mittee E13.19 on Separation Science.
Current edition approved Jan. 1, 2013. Published January 2013. Originally
chromatography(GC)instrumentmanualforspecificdetailson
approved in 1993. Last previous edition approved in 2005 as E1510–95 (2005).
injector or detector pneumatics for capillary columns.
DOI: 10.1520/E1510-95R13.
Reprinted by permission of Restek Corp., 110 Benner Circle, Bellefonte, PA
4.2 Prior to performing a capillary GC analysis, the capil-
16823-8812.
lary column configuration must be determined. The stationary
For referenced ASTM standards, visit the ASTM website, www.astm.org, or
contact ASTM Customer Service at service@astm.org. For Annual Book of ASTM
Standards volume information, refer to the standard’s Document Summary page on Available from Compressed Gas Association (CGA), 4221 Walney Rd., 5th
the ASTM website. Floor, Chantilly, VA 20151-2923, http://www.cganet.com.
Copyright © ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959. United States
´1
E1510−95 (2013)
TABLE 1 Typical Splitter Vent Flow Rates (50 to 1 split ratio)
(at optimum linear velocity)
0.25-mm ID, 0.32-mm ID, 0.53-mm ID,
Carrier gas
3 3 3
cm /min cm /min cm /min
helium 35 80 125
hydrogen 70 160 250
Carrier gas: Hydrogen Carrier gas: Helium
Linear velocity: 40 cm/s Linear velocity: 20 cm/s
NOTE1—Fig.2showsthattheresolutionissimilarbuttheanalysistime
is reduced by 50% when comparing hydrogen to helium in an isothermal
analysis using optimum flow velocities.
NOTE 2—Hydrogen provides similar resolution in one-half the analysis
time of helium for an isothermal analysis.
NOTE1—Thecurvesweregeneratedbyplottingtheheightequivalentto
NOTE 3—
a theoretical plate (length of column divided by the total number of
1. Tetrachloro-m- 8. Heptachlor epoxide 15. Endosulfan II
theoretical plates, H.E.T.P.) against the column’s average linear velocity.
xylene 9. γ-chlordane 16. DDD
The lowest point on the curve indicates the carrier gas velocity in which
2. α-BHC 10. Endosulfan I 17. Endrin aldehyde
the highest column efficiency is reached.
3. β-BHC 11. α-chlordane 18. Endosulfan sulfate
NOTE 2—Gases information available from Compressed Gas Associa- 4. γ-BHC 12. Dieldrin 19. DDT
5. δ-BHC 13. DDE 20. Endrin ketone
tion (CGA), 4221 Walney Rd., 5th Floor, Chantilly, VA 20151-2923,
6. Heptachlor 14. Endrin 21. Methyoxychlor
http://www.cganet.com.
7. Aldrin
FIG. 1Van Deemter Profile for Hydrogen, Helium, and Nitrogen
Carrier Gases
NOTE 4—30 m, 0.25-mm ID, 0.25 µm 5% diphenyl−95% dimethyl
polysiloxane 0.1-µL split injection of chlorinated pesticides.
Oven temperature: 210°C isothermal
phase type, stationary phase film thickness, column inside
Injector and detector temperature: 250°C/300°C
−11
ECD sensitivity: 512 × 10
diameter, and column length must be selected. It is beyond the
Split vent: 100 cm /min
scope of this practice to provide these details. Consult a
column or instrument supplier for details on selecting the
FIG. 2Hydrogen Versus Helium (Isothermal Analysis)
appropriate capillary column configuration.
4.3 Apply caution during handling or installation to avoid
scratching or abrading the protective outer coating of the
column. Scratches or abrasions cause the fused silica capillary
column to spontaneously break or fail during usage.
5. Significance and Use
5.1 Thispracticeisintendedtobeusedbyallanalystsusing FIG. 3 Capping Silanol Groups with Dimethyl Dichlorosilane
(DMDCS)
fused silica capillary chromatography. It contains the recom-
mended steps for installation, preparation, proper installation,
and continued column maintenance.
6. Hazards
7.1.1 Cool all heated zones and replace spent oxygen and
moisture scrubbers,
6.1 Gas Handling Safety—The safe handling of compressed
7.1.2 Clean or deactivate, or both, injector and detector
gases and cryogenic liquids for use in chromatography is the
responsibility of every laboratory. The Compressed Gas sleeves (if necessary),
7.1.3 Replace critical injector and detector seals,
Association, a member group of specialty and bulk gas
suppliers, publishes the following guidelines to assist the 7.1.4 Replace septum,
laboratory chemist to establish a safe work environment: CGA 7.1.5 Set make-up and detector gas flow rates,
P-1, CGA G-5.4, CGA P-9, CGA V-7, CGA P-12, and HB-3.
7.1.6 Carefully inspect the column for damage or breakage,
7.1.7 Cutapproximately10cmfromeachendofthecolumn
7. Installation Procedure for Fused Silica Capillary
using a ceramic scoring wafer or sapphire scribe,
Columns
7.1.8 Install nut and appropriately sized ferrule on both
7.1 Abriefoutlineofthestepsnecessaryforinstallingfused column ends,
silica capillary columns in capillary dedicated gas chromato- 7.1.9 Cut an additional 10 cm from each end of the column
graphs is as follows: to remove ferrule shards,
´1
E1510−95 (2013)
7.1.21 Reinject a non-retained substance (usually methane)
to set the proper linear velocity,
7.1.22 Run test mixtures to confirm proper installation and
column performance, and
7.1.23 Calibrate instrument and inject samples.
7.2 The following section provides in-depth information on
instrument preparation procedures for installing and operating
fused silica capillary columns in capillary dedicated gas
chromatographs:
7.2.1 Gas Purification—The carrier gas must contain less
than 1 ppm of oxygen, moisture, or any other trace contami-
nants. Otherwise, oxygen and moisture degrade column
NOTE 1—Septum bleed can obscure or co-elute with compounds of
performance, decrease column lifetime, and increase back-
interest, thus decreasing the analytical accuracy.
ground stationary phase bleed. Contaminants such as trace
NOTE 2—
hydrocarbons cause ghost peaks to appear during temperature
1. 2,4,5,6-tetrachloro- 8. Heptachlor 16. p,p-DDD
programming and degrade the validity of the analytical data.
m-xylene (IS) epoxide 17. Endrin aldehyde
Make-up gas should also be contaminant-free or baseline
2. α-BHC 9. γ-chlordane 18. Endosulfan sul-
3. β-BHC 10. Endosulfan I fate
fluctuations and excessive detector noise may occur. Detector
4. γ-BHC 11. α-chlordane 19. p,p-DDT
gases such as hydrogen and compressed air should be free of
5. δ-BHC 12. Dieldrin 20. Endrin ketone
water and hydrocarbon or excessive baseline noise may occur.
6. Heptachlor 13. p,p-DDE 21. Methyoxychlor
7. Aldrin 14. Endrin 22. Decachlorobi-
7.2.1.1 Install purifiers as closely as possible to the GC’s
15. Endosulfan II phenyl (IS)
bulkhead fitting, rather than system-wide. If purifiers are
installed system-wide, a leaky fitting downstream of the
NOTE 3—30 m, 0.53-mm ID, 0.50 µm 5% diphenyl−95% dimethyl
polysiloxane 0.1 µL direct injection of 50 pg pesticide standard. purifier could allow oxygen and moisture to enter the gas
stream and degrade column performance.
Oven temperature: 150 to 275°C at 4°C/min,
hold15min
7.2.1.2 Only high–purity gases should be used for capillary
Injector temperature: 250°C Detector temperature: 300°C
chromatography.All regulators should be equipped with stain-
Carrier gas: Helium
Linear velocity: 40 cm/s (Flow rate: 10 cm /min) less steel diaphragms. Regulators equipped with rubber or
−11
ECD sensitivity: 8 × 10 AFS
elastomeric diaphragms should not be used because oxygen,
moisture, and elastomeric contaminants migrate through the
FIG. 4ECD Septum Bleed
diaphragm and enter the flow.
7.2.1.3 Both indicating and non-indicating traps are avail-
able from most capillary column suppliers. Indicating purifiers
7.1.10 Mount the capillary column in the oven using a
are recommended since they allow analysts to visually assess
bracket to protect the column from becoming scratched or
whether the purifier has exceeded its useful life. Also, a
abraded and to prevent it from touching the oven wall,
moisture trap should be installed prior to the oxygen trap. If
7.1.11 Connect the column to the inlet at the appropriate
hydrocarbon contamination is suspected, a hydrocarbon trap
distance as indicated in the instrument manual,
should be installed between the moisture and oxygen trap.
7.1.12 Set the approximate column flow rate by adjusting
Sincemostindicatingtrapsaremadefromglass,careshouldbe
the head pressure (see column manufacturer’s literature),
taken not to apply lateral torque on the fittings, or they will
7.1.13 Set split vent, septa purge, and any other applicable
snap. To prevent spontaneous breakage of the trap, the line
inlet gases according to the instrument specifications,
leading to and from the purifier should be coiled to relieve
7.1.14 Confirmflowbyimmersingcolumnoutletinavialof
strain and isolate instrument vibrations.
acetone or methylene chloride,
7.1.15 Connectthecolumntothedetectorattheappropriate 7.2.2 Carrier Gas Selection—Afastcarriergasthatexhibits
a flat van Deemter profile is essential to obtain optimum
distance as indicated in the instrument manual,
7.1.16 Check for leaks at the inlet or outlet using a thermal capillary column performance. Because capillary columns
average30minlength(comparedto2mforpackedcolumns),
conductivity leak detector (do not use soaps or liquid-based
leak detectors), a carrier gas that minimizes the effect of dead time is
important. In addition, capillary columns are usually head
7.1.17 Set injector and detector temperatures and turn on
detector when temperatures have equilibrated (Warning—Do pressure controlled (not flow controlled like most packed
columns), which cause the carrier gas flow rate to decrease by
not exceed the phase’s maximum operating temperature),
7.1.18 Inject a non-retained substance (usually methane) to 40%whenthecolumnisprogrammedfromambientto300°C.
Therefore, a carrier gas that retains high efficiency over a wide
set the proper dead time (linear velocity),
range of flow rates is essential towards obtaining good resolu-
7.1.19 Check system integrity by making sure dead volume
tion throughout a temperature–programmed chromatographic
peak does not tail,
analysis.
7.1.20 Condition the column at the maximum operating
temperature for 2 h (consult column manufacturer’s literature) 7.2.2.1 The optimum average linear gas velocity for hydro-
to stabilize the baseline, gen (u : 40 cm/s) is greater than all the others, and hydrogen
opt
´1
E1510−95 (2013)
exhibits the flattest van Deemter profile. Helium is the next Proper safety precautions should be utilized to prevent an
best choice (u : 20 cm/s). Note that head pressures at explosion in the oven chamber. Some gas chromatographs are
opt
optimum flow rates are similar for hydrogen and helium designed with spring–loaded doors, perforated or corrugated
because hydrogen has half the viscosity but double the linear metal oven chambers, and back pressure/flow controlled
velocityashelium.Becauseofthelowoptimumlinearvelocity pneumatics,whichminimizethehazardswhenusinghydrogen
(u : 10 cm/s) and steep van Deemter profile, nitrogen gives carrier gas. Additional precautions used by analysts include:
opt
inferior performance with capillary columns and is usually not (a)Frequently check for carrier gas leaks using a sensi-
recommended. tive electronic leak detector,
7.2.2.2 Temperature programming usually provides similar (b)Use electronic sensors that shut down the carrier gas
analysis times between hydrogen and helium since the elution flow should an explosive atmosphere be
...

Questions, Comments and Discussion

Ask us and Technical Secretary will try to provide an answer. You can facilitate discussion about the standard in here.