ASTM E2056-04(2010)
(Practice)Standard Practice for Qualifying Spectrometers and Spectrophotometers for Use in Multivariate Analyses, Calibrated Using Surrogate Mixtures
Standard Practice for Qualifying Spectrometers and Spectrophotometers for Use in Multivariate Analyses, Calibrated Using Surrogate Mixtures
SIGNIFICANCE AND USE
This practice should be used by the developer of standard test methods that employ surrogate calibrations.
This practice assists the test method developer in setting and documenting requirements for the spectrometer/spectrophotometers that can perform the test method.
This practice assists the test method developer in setting and documenting spectral data collection and computation parameters for the test method.
This practice assists the test method developer in selecting among possible multivariate analysis procedures that could be used to establish the surrogate calibration. The practice describes statistical tests that should be performed to ensure that all multivariate analysis procedures that are allowed within the scope of the test method produce statistically indistinguishable results.
This practice describes statistical calculations that the test method developer should perform on the calibration and qualification data that should be collected as part of the ILS that establishes the test method precision. These calculations establish the level of performance that spectrometers/spectrophotometers must meet in order to perform the test method.
This practice describes how the person who calibrates a spectrometer/spectrophotometer can test the performance of said spectrometer/spectrophotometer to determine if the performance is adequate to conduct the test method.
This practice describes how the user of a spectrometer/spectrophotometer can qualify the spectrometer/spectrophotometer to conduct the test method.
SCOPE
1.1 This practice relates to the multivariate calibration of spectrometers and spectrophotometers used in determining the physical and chemical characteristics of materials. A detailed description of general multivariate analysis is given in Practice E1655. This standard refers only to those instances where surrogate mixtures can be used to establish a suitable calibration matrix. This practice specifies calibration and qualification data set requirements for interlaboratory studies (ILSs), that is, round robins, of standard test methods employing surrogate calibration techniques that do not conform exactly to Practices E1655.
Note 1—For some multivariate spectroscopic analyses, interferences and matrix effects are sufficiently small that it is possible to calibrate using mixtures that contain substantially fewer chemical components than the samples that will ultimately be analyzed. While these surrogate methods generally make use of the multivariate mathematics described in Practices E1655, they do not conform to procedures described therein, specifically with respect to the handling of outliers.
1.2 This practice specifies how the ILS data is treated to establish spectrometer/spectrophotometer performance qualification requirements to be incorporated into standard test methods.
Note 2—Spectrometer/spectrophotometer qualification procedures are intended to allow the user to determine if the performance of a specific spectrometer/spectrophotometer is adequate to conduct the analysis so as to obtain results consistent with the published test method precision.
1.2.1 The spectroscopies used in the surrogate test methods would include but not be limited to mid- and near-infrared, ultraviolet/visible, fluorescence and Raman spectroscopies.
1.2.2 The surrogate calibrations covered in this practice are: multilinear regression (MLR), principal components regression (PCR) or partial least squares (PLS) mathematics. These calibration procedures are described in detail in Practices E1655.
1.3 For surrogate test methods, this practice recommends limitations that should be placed on calibration options that are allowed in the test method. Specifically, this practice recommends that the test method developer demonstrate that all calibrations that are allowed in the test method produce statistically indistinguishable results.
1.4 For surrogate test methods that r...
General Information
Relations
Standards Content (Sample)
NOTICE: This standard has either been superseded and replaced by a new version or withdrawn.
Contact ASTM International (www.astm.org) for the latest information
Designation: E2056 − 04 (Reapproved2010)
Standard Practice for
Qualifying Spectrometers and Spectrophotometers for Use
in Multivariate Analyses, Calibrated Using Surrogate
Mixtures
This standard is issued under the fixed designation E2056; the number immediately following the designation indicates the year of
original adoption or, in the case of revision, the year of last revision. A number in parentheses indicates the year of last reapproval. A
superscript epsilon (´) indicates an editorial change since the last revision or reapproval.
1. Scope 1.3 For surrogate test methods, this practice recommends
limitations that should be placed on calibration options that are
1.1 This practice relates to the multivariate calibration of
allowed in the test method. Specifically, this practice recom-
spectrometers and spectrophotometers used in determining the
mends that the test method developer demonstrate that all
physical and chemical characteristics of materials. A detailed
calibrations that are allowed in the test method produce
description of general multivariate analysis is given in Practice
statistically indistinguishable results.
E1655. This standard refers only to those instances where
surrogate mixtures can be used to establish a suitable calibra- 1.4 For surrogate test methods that reference spectrometer/
tion matrix.This practice specifies calibration and qualification spectrophotometer performance practices, such as Practices
data set requirements for interlaboratory studies (ILSs), that is, E275, E925, E932, E958, E1421, E1683,or E1944; Test
round robins, of standard test methods employing surrogate Methods E387, E388,or E579; or Guide E1866, this practice
calibration techniques that do not conform exactly to Practices recommends that instrument performance data be collected as
E1655. part of the ILS to establish the relationship between
spectrometer/spectrophotometer performance and test method
NOTE 1—For some multivariate spectroscopic analyses, interferences
precision.
andmatrixeffectsaresufficientlysmallthatitispossibletocalibrateusing
mixtures that contain substantially fewer chemical components than the
2. Referenced Documents
samples that will ultimately be analyzed. While these surrogate methods
generally make use of the multivariate mathematics described in Practices 2
2.1 ASTM Standards:
E1655, they do not conform to procedures described therein, specifically
D6277 Test Method for Determination of Benzene in Spark-
with respect to the handling of outliers.
Ignition Engine Fuels Using Mid Infrared Spectroscopy
1.2 This practice specifies how the ILS data is treated to
D6300 Practice for Determination of Precision and Bias
establish spectrometer/spectrophotometer performance qualifi-
Data for Use in Test Methods for Petroleum Products and
cation requirements to be incorporated into standard test
Lubricants
methods.
E131 Terminology Relating to Molecular Spectroscopy
NOTE 2—Spectrometer/spectrophotometer qualification procedures are
E275 Practice for Describing and Measuring Performance of
intended to allow the user to determine if the performance of a specific
Ultraviolet and Visible Spectrophotometers
spectrometer/spectrophotometer is adequate to conduct the analysis so as
E387 TestMethodforEstimatingStrayRadiantPowerRatio
to obtain results consistent with the published test method precision.
of Dispersive Spectrophotometers by the Opaque Filter
1.2.1 The spectroscopies used in the surrogate test methods
Method
would include but not be limited to mid- and near-infrared,
E388 Test Method for Wavelength Accuracy and Spectral
ultraviolet/visible, fluorescence and Raman spectroscopies.
Bandwidth of Fluorescence Spectrometers
1.2.2 The surrogate calibrations covered in this practice are:
E579 Test Method for Limit of Detection of Fluorescence of
multilinearregression(MLR),principalcomponentsregression
Quinine Sulfate in Solution
(PCR) or partial least squares (PLS) mathematics. These
E691 Practice for Conducting an Interlaboratory Study to
calibration procedures are described in detail in Practices
Determine the Precision of a Test Method
E1655.
E925 Practice for Monitoring the Calibration of Ultraviolet-
Visible Spectrophotometers whose Spectral Bandwidth
This practice is under the jurisdiction of ASTM Committee E13 on Molecular
Spectroscopy and Separation Science and is the direct responsibility of Subcom-
mittee E13.11 on Multivariate Analysis. For referenced ASTM standards, visit the ASTM website, www.astm.org, or
Current edition approved March 1, 2010. Published May 2010. Originally contact ASTM Customer Service at service@astm.org. For Annual Book of ASTM
approved in 1999. Last previous edition approved in 2004 as E2056 – 04. DOI: Standards volume information, refer to the standard’s Document Summary page on
10.1520/E2056-04R10. the ASTM website.
Copyright © ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959. United States
E2056 − 04 (2010)
does not Exceed 2 nm where appropriate, to calculate the spectral data used in the
E932 Practice for Describing and Measuring Performance of calibration and analysis.
Dispersive Infrared Spectrometers
4.4 The test method should specify the exact mathematics
E958 Practice for Measuring Practical Spectral Bandwidth
that are to be used to develop the multivariate calibration.
of Ultraviolet-Visible Spectrophotometers
Allowable spectral preprocessing methods should be defined.
E1421 Practice for Describing and Measuring Performance
The specific mathematics (MLR, PCR or PLS) should be
of Fourier Transform Mid-Infrared (FT-MIR) Spectrom-
specified,andtheacceptablerangeforthenumbersofvariables
eters: Level Zero and Level One Tests
should be given.
E1655 Practices for Infrared Multivariate Quantitative
4.5 When the ILS is conducted to establish the precision of
Analysis
the surrogate test method, the calibration data for all of the
E1683 Practice for Testing the Performance of Scanning
participating laboratories should be collected and used to
Raman Spectrometers
calculate a pooled standard error of calibration for the test
E1866 Guide for Establishing Spectrophotometer Perfor-
method. The pooled standard error of calibration and its
mance Tests
associated degrees of freedom should be reported in the test
E1944 Practice for Describing and Measuring Performance
of Laboratory Fourier Transform Near-Infrared (FT-NIR) method.
Spectrometers: Level Zero and Level One Tests
4.5.1 When a user is calibrating a spectrometer/
spectrophotometer, the standard error of calibration is calcu-
3. Terminology
lated and compared to the pooled standard error of calibration
from the ILS to determine if the performance of the calibrated
3.1 Definitions:
3.1.1 For definitions of terms and symbols relating to spectrometer/spectrophotometer is adequate to produce analy-
ses of the precision specified in the test method.
infrared, ultraviolet/visible and Raman spectroscopy, refer to
Terminology E131.
4.5.2 If a user is purchasing a precalibrated spectrometer/
3.1.2 For definitions of terms and symbols relating to
spectrophotometer, the instrument vendor should supply the
multivariate analysis, refer to Practices E1655.
standard error of calibration and its statistical comparison to
the pooled standard error of calibration.
3.2 Definitions of Terms Specific to This Standard:
3.2.1 spectrometer/spectrophotometer qualification, n—the
4.6 During the ILS, each participating laboratory analyzes a
procedures by which a user demonstrates that the performance
set of qualification samples and reports both the compositions
of a specific spectrometer/spectrophotometer is adequate to
of the qualification set and the estimates made using the
conduct a multivariate analysis so as to obtain precision
multivariate analysis. A pooled error of qualification is calcu-
consistent with that specified in the test method.
lated and reported as part of the test method along with its
corresponding degrees of freedom.
3.2.2 surrogate calibration, n—a multivariate calibration
that is developed using a calibration set which consists of
4.6.1 Before a user may use the spectrometer/
mixtureswithpre-specifiedandreproduciblecompositionsthat spectrophotometer, it must be qualified to perform the surro-
contain substantially fewer chemical components than the
gate test method. The qualification set is analyzed, and a
samples that will ultimately be analyzed. standard error of qualification is calculated. The standard error
of qualification is statistically compared with the pooled
3.2.3 surrogate test method, n—a standard test method that
standard error of qualification to determine if the performance
is based on a surrogate calibration.
ofthecalibratedspectrometer/spectrophotometerisadequateto
4. Summary of Practice produce analyses of the precision specified in the test method.
4.6.2 Spectrometer/spectrophotometer qualification is re-
4.1 Asurrogatetestmethodmustspecifythecompositionof
quiredregardlessofwhetherthecalibrationisperformedbythe
two sets of samples. One set is used to calibrate the
vendor or the user.
spectrometers/spectrophotometers. The second set of samples
4.6.3 Spectrometer/spectrophotometer qualification should
is used to qualify the spectrometer/spectrophotometer to per-
berepeatedaftermajormaintenancehasbeenperformedonthe
form the analysis. The compositions of both sets are expressed
spectrometer/spectrophotometer so as to determine whether
in terms of weight or volume fraction depending on whether
recalibration is required.
thesamplesarepreparedgravimetricallyorvolumetrically.The
compositions of both sets should be specified in the surrogate
test method. If the surrogate test method is being used to 5. Significance and Use
estimate a physical property, then the test method should
5.1 This practice should be used by the developer of
indicate what value of the property is to be assigned to each of
standard test methods that employ surrogate calibrations.
the calibration and qualification samples.
5.1.1 This practice assists the test method developer in
4.2 The surrogate test method should specify the minimum
setting and documenting requirements for the spectrometer/
spectrometer/spectrophotometer requirements for instruments
spectrophotometers that can perform the test method.
that can be used to perform the test method.
5.1.2 This practice assists the test method developer in
4.3 The spectrometer/spectrophotometer test method should setting and documenting spectral data collection and compu-
specify the exact conditions that are to be used to collect and, tation parameters for the test method.
E2056 − 04 (2010)
5.1.3 This practice assists the test method developer in 6.2.1 Thesetsofsurrogatesamplesthatareusedtocalibrate
selecting among possible multivariate analysis procedures that the spectrometers/spectrophotometers should satisfy the re-
could be used to establish the surrogate calibration. The quirements of Practices E1655.If k is the number of variables
practice describes statistical tests that should be performed to (MLR wavelengths or frequencies, PCR principal components
ensurethatallmultivariateanalysisproceduresthatareallowed or PLS latent variables) used in the model, then the minimum
within the scope of the test method produce statistically number of calibration samples should be the greater of 24 or
indistinguishable results. 6k. If the calibration set is derived from an experimental
5.1.4 This practice describes statistical calculations that the design, and if the spectra have been shown to be linear
functions of the component concentrations, then fewer calibra-
test method developer should perform on the calibration and
qualification data that should be collected as part of the ILS tion samples can be used, but in all cases the minimum number
of calibration samples should be the greater of 24 or 4k. The
that establishes the test method precision. These calculations
establish the level of performance that spectrometers/ experimental design must independently vary all components
over the desired analysis range.
spectrophotometers must meet in order to perform the test
method.
6.2.2 When calibrating for a single component, the calibra-
tion set should uniformly span the range over which the
5.2 This practice describes how the person who calibrates a
analysis of that component is to be conducted. Additional
spectrometer/spectrophotometer can test the performance of
components that are present in the calibration set to simulate
said spectrometer/spectrophotometer to determine if the per-
interferences should be independently and uniformly varied
formance is adequate to conduct the test method.
over a range at least as large as is likely to be encountered
5.3 This practice describes how the user of a spectrometer/
during actual application of the test method.
spectrophotometer can qualify the spectrometer/
6.2.3 When calibrating for a property that depends on more
spectrophotometer to conduct the test method.
than one chemical component, the calibration set should
uniformly span the range over which the property analysis is to
6. Surrogate Calibrations
be conducted, and all components that contribute to the
6.1 Practices E1655 assumes that the calibration set used to property should be varied independently.
develop a multivariate model contains samples of the same 6.2.4 The test method should specify the compositions of
type as those that are to eventually be analyzed using the
the calibration samples, including components and target
model. Practices E1655 requires use of outlier statistics to concentrations. The purity of materials to be used in preparing
ensure that samples being analyzed are sufficiently similar to
the calibration samples should also be specified in the test
the calibration samples to produce meaningful results. For method.
some spectroscopic analyses, however, it is possible to cali-
6.3 Qualification Sets:
brate using gravimetrically or volumetrically prepared mix-
6.3.1 The sets of surrogate samples that are used to qualify
tures that contain significantly fewer components than the
the spectrometers/spectrophotometers should satisfy the vali-
samples that will ultimately be analyzed. For these surrogate
dation requirements of Practices E1655.If k is the number of
test methods, the outlier statistics described in Practices E1655
variables (MLR wavelengths or frequencies, PCR principal
arenotappropriatesinceallsamplesareexpectedtobeoutliers
components or PLS latent variables) used in the model, then
relative to the simplified calibrations. Thus, surrogate test
the minimum number of qualification samples should be the
methods cannot fulfill the requirements of Practices E1655.
greater of 20 or 5k. If t
...
Questions, Comments and Discussion
Ask us and Technical Secretary will try to provide an answer. You can facilitate discussion about the standard in here.