Standard Specification for Polydioxanone Polymer Resins for Surgical Implants

SCOPE
1.1 This specification covers virgin polydioxanone homopolymer resins intended for use in surgical implants.  
1.2 Polydioxanone is commonly abbreviated as PDO, and is alternatively referred to as poly(para dioxanone) or poly(ρ-dioxanone) with the acronym PPD. Additionally, it may be referred to as PDS as it is the polymer of composition of PDS suture (Ethicon, Inc.), representing an early and widely used application of polydioxanone polymer.  
1.3 This specification covers virgin polydioxanone resins able to be fully solvated at 30 °C by fluorinated solvents such as hexafluoroisopropanol (HFIP) or hexafluoroacetone (HFA).  
1.4 Homopolymers of this composition are known to be semi-crystalline. Within this specification, semi-crystallinity within the resin is defined by the presence of a DSC (differential scanning calorimetry) crystalline endotherm peak upon annealing between 105 and 115 °C. While the presence of a crystalline endotherm indicates semi-crystallinity, the percentage and morphology of the crystalline phase are highly dependent on processing, and in particular on the thermal history of the material. Therefore, the thermal properties and percent crystallinity of the virgin polymer resin (with exception of melting temperature) are not necessarily indicative of final product quality.  
1.5 This specification addresses material characteristics of the virgin polydioxanone-based resins intended for use in surgical implants and does not apply to packaged and sterilized finished implants fabricated from these materials, nor does it address the characteristics of polydioxanone resins with compounded materials such as dyes, polymeric or ceramic compounds, or any other additives.  
1.6 As with any material, some characteristics may be altered by processing techniques (such as molding, extrusion, machining, assembly, sterilization, and so forth) required for the production of a specific part or device. Therefore, properties of fabricated forms of this resin should be evaluated independently using appropriate test methods to ensure safety and efficacy.  
1.7 Biocompatibility testing is not a requirement since this specification is not intended to cover fabricated devices. While biocompatibility testing of resin may provide an early indication of potential safety, biocompatibility analysis of the final finished device is required to determine safety and suitability for any implant device. Refer to Supplementary Requirement S1 of this standard and Guide F2902 for relevant biocompatibility information.  
1.8 The values stated in SI units are to be regarded as standard. No other units of measurement are included in this standard.  
1.9 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety, health, and environmental practices and determine the applicability of regulatory limitations prior to use.  
1.10 This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.

General Information

Status
Published
Publication Date
30-Nov-2021
Current Stage
Ref Project

Buy Standard

Technical specification
ASTM F3384-21 - Standard Specification for Polydioxanone Polymer Resins for Surgical Implants
English language
9 pages
sale 15% off
Preview
sale 15% off
Preview

Standards Content (Sample)

This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the
Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.
Designation:F3384 −21
Standard Specification for
1
Polydioxanone Polymer Resins for Surgical Implants
This standard is issued under the fixed designation F3384; the number immediately following the designation indicates the year of
original adoption or, in the case of revision, the year of last revision. A number in parentheses indicates the year of last reapproval. A
superscript epsilon (´) indicates an editorial change since the last revision or reapproval.
1. Scope ties of fabricated forms of this resin should be evaluated
independently using appropriate test methods to ensure safety
1.1 Thisspecificationcoversvirginpolydioxanonehomopo-
and efficacy.
lymer resins intended for use in surgical implants.
1.7 Biocompatibility testing is not a requirement since this
1.2 Polydioxanone is commonly abbreviated as PDO, and is
specification is not intended to cover fabricated devices. While
alternatively referred to as poly(para dioxanone) or poly(ρ-
biocompatibility testing of resin may provide an early indica-
dioxanone) with the acronym PPD. Additionally, it may be
tion of potential safety, biocompatibility analysis of the final
referred to as PDS as it is the polymer of composition of PDS
finished device is required to determine safety and suitability
suture (Ethicon, Inc.), representing an early and widely used
for any implant device. Refer to Supplementary Requirement
application of polydioxanone polymer.
S1 of this standard and Guide F2902 for relevant biocompat-
1.3 This specification covers virgin polydioxanone resins ibility information.
able to be fully solvated at 30 °C by fluorinated solvents such
1.8 The values stated in SI units are to be regarded as
as hexafluoroisopropanol (HFIP) or hexafluoroacetone (HFA).
standard. No other units of measurement are included in this
standard.
1.4 Homopolymers of this composition are known to be
semi-crystalline. Within this specification, semi-crystallinity
1.9 This standard does not purport to address all of the
within the resin is defined by the presence of a DSC (differ-
safety concerns, if any, associated with its use. It is the
ential scanning calorimetry) crystalline endotherm peak upon
responsibility of the user of this standard to establish appro-
annealing between 105 and 115 °C. While the presence of a
priate safety, health, and environmental practices and deter-
crystalline endotherm indicates semi-crystallinity, the percent-
mine the applicability of regulatory limitations prior to use.
age and morphology of the crystalline phase are highly
1.10 This international standard was developed in accor-
dependent on processing, and in particular on the thermal
dance with internationally recognized principles on standard-
history of the material. Therefore, the thermal properties and
ization established in the Decision on Principles for the
percentcrystallinityofthevirginpolymerresin(withexception
Development of International Standards, Guides and Recom-
of melting temperature) are not necessarily indicative of final
mendations issued by the World Trade Organization Technical
product quality.
Barriers to Trade (TBT) Committee.
1.5 This specification addresses material characteristics of
2. Referenced Documents
the virgin polydioxanone-based resins intended for use in
2
2.1 ASTM Standards:
surgicalimplantsanddoesnotapplytopackagedandsterilized
D1505 Test Method for Density of Plastics by the Density-
finished implants fabricated from these materials, nor does it
Gradient Technique
address the characteristics of polydioxanone resins with com-
D2857 Practice for Dilute Solution Viscosity of Polymers
pounded materials such as dyes, polymeric or ceramic
D3418 Test Method for Transition Temperatures and En-
compounds, or any other additives.
thalpies of Fusion and Crystallization of Polymers by
1.6 As with any material, some characteristics may be
Differential Scanning Calorimetry
altered by processing techniques (such as molding, extrusion,
D4603 Test Method for Determining Inherent Viscosity of
machining, assembly, sterilization, and so forth) required for
Poly(Ethylene Terephthalate) (PET) by Glass Capillary
the production of a specific part or device. Therefore, proper-
Viscometer
D5296 Test Method for Molecular Weight Averages and
1
This specification is under the jurisdiction of ASTM Committee F04 on
2
Medical and Surgical Materials and Devices and is the direct responsibility of For referenced ASTM standards, visit the ASTM website, www.astm.org, or
Subcommittee F04.11 on Polymeric Materials. contact ASTM Customer Service at service@astm.org. For Annual Book of ASTM
Current edition approved Dec. 1, 2021. Published December 2021. DOI: Standards volume informatio
...

Questions, Comments and Discussion

Ask us and Technical Secretary will try to provide an answer. You can facilitate discussion about the standard in here.