ASTM D4052-96
(Test Method)Standard Test Method for Density and Relative Density of Liquids by Digital Density Meter
Standard Test Method for Density and Relative Density of Liquids by Digital Density Meter
SCOPE
1.1 This test method covers the determination of the density or relative density of petroleum distillates and viscous oils that can be handled in a normal fashion as liquids at test temperatures between 15 and 35°C. Its application is restricted to liquids with vapor pressures below 600 mm Hg (80 kPa) and viscosities below about 15000 cSt (mm 2/s) at the temperature of test.
1.2 This test method should not be applied to samples so dark in color that the absence of air bubbles in the sample cell cannot be established with certainty. For the determination of density in crude oil samples use Test Method D 5002.
1.3 The accepted units of measure for density are grams per millilitre or kilograms per cubic metre.
1.4 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use. For specific hazard statements, see 7.4, 7.5, and 9.1.
General Information
Relations
Standards Content (Sample)
NOTICE: This standard has either been superseded and replaced by a new version or
withdrawn. Contact ASTM International (www.astm.org) for the latest information.
Designation: D 4052 – 96 An American National Standard
Designation: 365/84(86)
AMERICAN SOCIETY FOR TESTING AND MATERIALS
100 Barr Harbor Dr., West Conshohocken, PA 19428
Reprinted from the Annual Book of ASTM Standards. Copyright ASTM
Standard Test Method for
Density and Relative Density of Liquids by Digital Density
1
Meter
This standard is issued under the fixed designation D 4052; the number immediately following the designation indicates the year of
original adoption or, in the case of revision, the year of last revision. A number in parentheses indicates the year of last reapproval. A
superscript epsilon (e) indicates an editorial change since the last revision or reapproval.
This method was adopted as a joint ASTM-IP standard in 1984.
1. Scope D 5002 Test Method for Density and Relative Density of
5
Crude Oils by Digital Density Analyzer
1.1 This test method covers the determination of the density
or relative density of petroleum distillates and viscous oils that
3. Terminology
can be handled in a normal fashion as liquids at test tempera-
3.1 Definitions:
tures between 15 and 35°C. Its application is restricted to
3.1.1 density—mass per unit volume at a specified tempera-
liquids with vapor pressures below 600 mm Hg (80 kPa) and
2
ture.
viscosities below about 15 000 cSt (mm /s) at the temperature
3.1.2 relative density—the ratio of the density of a material
of test.
at a stated temperature to the density of water at a stated
1.2 This test method should not be applied to samples so
temperature.
dark in color that the absence of air bubbles in the sample cell
cannot be established with certainty. For the determination of
4. Summary of Test Method
density in crude oil samples use Test Method D 5002.
4.1 A small volume (approximately 0.7 mL) of liquid
1.3 The accepted units of measure for density are grams per
sample is introduced into an oscillating sample tube and the
millilitre or kilograms per cubic metre.
change in oscillating frequency caused by the change in the
1.4 This standard does not purport to address all of the
mass of the tube is used in conjunction with calibration data to
safety concerns, if any, associated with its use. It is the
determine the density of the sample.
responsibility of the user of this standard to establish appro-
priate safety and health practices and determine the applica-
5. Significance and Use
bility of regulatory limitations prior to use. For specific hazard
5.1 Density is a fundamental physical property that can be
statements, see Note 1 and Note 2.
used in conjunction with other properties to characterize both
the light and heavy fractions of petroleum and petroleum
2. Referenced Documents
products.
2.1 ASTM Standards:
2 5.2 Determination of the density or relative density of
D 1193 Specification for Reagent Water
petroleum and its products is necessary for the conversion of
3
D 1250 Guide for Petroleum Measurement Tables
measured volumes to volumes at the standard temperature of
D 4057 Practice for Manual Sampling of Petroleum and
15°C.
4
Petroleum Products
D 4177 Practice for Automatic Sampling of Petroleum and
6. Apparatus
4
Petroleum Products
6.1 Digital Density Analyzer—A digital analyzer consisting
D 4377 Test Method for Water in Crude Oils by Potentio-
of a U-shaped, oscillating sample tube and a system for
4
metric Karl Fischer Titration
electronic excitation, frequency counting, and display. The
analyzer must accommodate the accurate measurement of the
1 sample temperature during measurement or must control the
This test method is under the jurisdiction of ASTM Committee D-2 on
Petroleum Products and Lubricantsand is the direct responsibility of Subcommittee sample temperature as described in 6.2. The instrument shall be
D02.04 on Hydrocarbon Analysis.
capable of meeting the precision requirements described in this
Current edition approved Apr. 10, 1996. Published June 1996. Originally
test method.
published as D 4052 – 81. Last previous edition D 4052 – 95.
2
Annual Book of ASTM Standards, Vol 11.01.
3
Annual Book of ASTM Standards, Vol 05.01.
4 5
Annual Book of ASTM Standards, Vol 05.02. Annual Book of ASTM Standards, Vol 05.03.
1
---------------------- Page: 1 ----------------------
D 4052
6.2 Circulating Constant-Temperature Bath, (optional) ca- 8.3.1 Mix the sample if required to homogenize. The mixing
pable of maintaining the temperature of the circulating liquid may be accomplished as described in Practice D 4177 (Section
constant to 60.05°C in the desired range. Temperature control 11) or Test Method D 4377 (A.1). Mixing at room temperature
can be maintained as part of the density analyzer instrument in an open container can result in the loss of volatile material,
package. so mixing in closed, pressurized containers or at sub-ambient
6.3 Syringes, at least 2 mL in volume with a tip or
...
Questions, Comments and Discussion
Ask us and Technical Secretary will try to provide an answer. You can facilitate discussion about the standard in here.