Standard Practice for Laser Technologies for Direct Measurement of Cross Sectional Shape of Pipeline and Conduit by Rotating Laser Diodes and CCTV Camera System

SIGNIFICANCE AND USE
4.1 Laser profiling assessment is a quality control tool for identifying and quantifying deformation, physical damage, and other pipe anomalies after installation, providing means and methods for determining the quality of workmanship and compliance with project specifications. Laser profiling capabilities include:  
4.1.1 Measurement of the structural shape, cross sectional area and defects;  
4.1.2 Collection of data needed for pipe rehabilitation or replacement design; and  
4.1.3 Post rehabilitation, replacement or new construction workmanship verification.  
4.2 A laser profile pre-acceptance and condition assessment survey provides significant information in a clear and concise manner, including but not limited to graphs and still frame digital images of pipe condition prior to acceptance, thereby providing objective data on the installed quality and percentage ovality, or degree of deformation, deflection or deviation, that is often not possible from an inspection by either a mandrel or only CCTV.
SCOPE
1.1 This practice covers the procedure for the post installation verification and acceptance of buried pipe deformation using a visible rotating laser light diode(s), a pipeline and conduit inspection analog or digital CCTV camera system and image processing software. The combination CCTV pipe inspection system, with cable distance counter or onboard distance encoder, rotating laser light diode(s) and ovality measurement software shall be used to perform a pipe measurement and ovality confirmation survey, of new or existing pipelines and conduits as directed by the responsible contracting authority. This standard practice provides minimum requirements on means and methods for laser profiling to meet the needs of engineers, contractors, owners, regulatory agencies, and financing institutions.  
1.2 This practice applies to all types of material, all types of construction, or shape.  
1.3 This practice applies to gravity flow storm sewers, drains, sanitary sewers, and combined sewers with diameters from 6 in. to 72 in. (150 mm to 1800 mm).  
1.4 The Laser Light Diode(s) shall be tested, labeled and certified to conform to US requirements for CDRH Class 2 or below (not considered to be hazardous) laser products or certified to conform to EU requirements for Class 2M or below laser products as per IEC 60825-1, or both.  
1.5 The profiling process may require physical access to lines, entry manholes and operations along roadways that may include safety hazards.  
1.6 This practice includes inspection requirements for determining pipeline and conduit ovality only and does not include all the required components of a complete inspection. The user of this practice should consider additional items outside this practice for inspection such as joint gap measurement, soil/water infiltration, crack and hole measurement, surface damage evaluation, evaluation of any pipeline repairs, and corrosion evaluation.  
1.7 This standard practice does not address limitations in accuracy due to improper lighting, dust, humidity, fog, moisture on pipe walls or horizontal/vertical offsets. Care should be taken to limit environmental factors in the pipeline that affect accuracy of the inspection.  
1.8 The values stated in inch-pound units are to be regarded as standard. The values given in parentheses are mathematical conversions to SI units that are provided for information only and are not considered standard.  
1.9 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety, health, and environmental practices and determine the applicability of regulatory limitations prior to use. There are no safety hazards specifically, however, associated with the use of the laser profiler specified (listed and labeled as specified in 1.3).  
1.10 This international standard was deve...

General Information

Status
Published
Publication Date
31-May-2022
Current Stage
Ref Project

Relations

Buy Standard

Standard
ASTM F3095-17a(2022) - Standard Practice for Laser Technologies for Direct Measurement of Cross Sectional Shape of Pipeline and Conduit by Rotating Laser Diodes and CCTV Camera System
English language
5 pages
sale 15% off
Preview
sale 15% off
Preview

Standards Content (Sample)


This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the
Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.
Designation: F3095 − 17a (Reapproved 2022)
Standard Practice for
Laser Technologies for Direct Measurement of Cross
Sectional Shape of Pipeline and Conduit by Rotating Laser
Diodes and CCTV Camera System
This standard is issued under the fixed designation F3095; the number immediately following the designation indicates the year of
original adoption or, in the case of revision, the year of last revision. A number in parentheses indicates the year of last reapproval. A
superscript epsilon (´) indicates an editorial change since the last revision or reapproval.
1. Scope* measurement, soil/water infiltration, crack and hole
measurement, surface damage evaluation, evaluation of any
1.1 This practice covers the procedure for the post installa-
pipeline repairs, and corrosion evaluation.
tion verification and acceptance of buried pipe deformation
using a visible rotating laser light diode(s), a pipeline and 1.7 This standard practice does not address limitations in
conduit inspection analog or digital CCTV camera system and accuracy due to improper lighting, dust, humidity, fog, mois-
image processing software. The combination CCTV pipe ture on pipe walls or horizontal/vertical offsets. Care should be
inspection system, with cable distance counter or onboard taken to limit environmental factors in the pipeline that affect
distance encoder, rotating laser light diode(s) and ovality accuracy of the inspection.
measurement software shall be used to perform a pipe mea-
1.8 The values stated in inch-pound units are to be regarded
surement and ovality confirmation survey, of new or existing
as standard. The values given in parentheses are mathematical
pipelines and conduits as directed by the responsible contract-
conversions to SI units that are provided for information only
ing authority. This standard practice provides minimum re-
and are not considered standard.
quirements on means and methods for laser profiling to meet
1.9 This standard does not purport to address all of the
the needs of engineers, contractors, owners, regulatory
safety concerns, if any, associated with its use. It is the
agencies, and financing institutions.
responsibility of the user of this standard to establish appro-
1.2 This practice applies to all types of material, all types of
priate safety, health, and environmental practices and deter-
construction, or shape.
mine the applicability of regulatory limitations prior to use.
There are no safety hazards specifically, however, associated
1.3 This practice applies to gravity flow storm sewers,
with the use of the laser profiler specified (listed and labeled as
drains, sanitary sewers, and combined sewers with diameters
specified in 1.3).
from 6 in. to 72 in. (150 mm to 1800 mm).
1.10 This international standard was developed in accor-
1.4 The Laser Light Diode(s) shall be tested, labeled and
dance with internationally recognized principles on standard-
certified to conform to US requirements for CDRH Class 2 or
ization established in the Decision on Principles for the
below (not considered to be hazardous) laser products or
Development of International Standards, Guides and Recom-
certified to conform to EU requirements for Class 2M or below
mendations issued by the World Trade Organization Technical
laser products as per IEC 60825-1, or both.
Barriers to Trade (TBT) Committee.
1.5 The profiling process may require physical access to
2. Referenced Documents
lines, entry manholes and operations along roadways that may
include safety hazards.
2.1 ASTM Standards:
E177 Practice for Use of the Terms Precision and Bias in
1.6 This practice includes inspection requirements for de-
ASTM Test Methods
termining pipeline and conduit ovality only and does not
E691 Practice for Conducting an Interlaboratory Study to
include all the required components of a complete inspection.
Determine the Precision of a Test Method
The user of this practice should consider additional items
F1216 Practice for Rehabilitation of Existing Pipelines and
outside this practice for inspection such as joint gap
Conduits by the Inversion and Curing of a Resin-
Impregnated Tube
This practice is under the jurisdiction ofASTM Committee F36 on Technology
and Underground Utilities and is the direct responsibility of Subcommittee F36.20
on Inspection and Renewal of Water and Wastewater Infrastructure. For referenced ASTM standards, visit the ASTM website, www.astm.org, or
Current edition approved June 1, 2022. Published June 2022. Originally contact ASTM Customer Service at service@astm.org. For Annual Book of ASTM
approved in 2014. Last previous edition approved in 2017 as F3095-17a. DOI: Standards volume information, refer to the standard’s Document Summary page on
10.1520/F3095-17AR22. the ASTM website.
*A Summary of Changes section appears at the end of this standard
Copyright © ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959. United States
F3095 − 17a (2022)
F2019 Practice for Rehabilitation of Existing Pipelines and methods for determining the quality of workmanship and
Conduits by the Pulled in Place Installation of Glass compliance with project specifications. Laser profiling capa-
Reinforced Plastic Cured-in-Place (GRP-CIPP) Using the bilities include:
UV-Light Curing Method 4.1.1 Measurement of the structural shape, cross sectional
area and defects;
2.2 Other Documents:
4.1.2 Collection of data needed for pipe rehabilitation or
IEC 60825-1 Safety of Laser Products—Part 1: Equipment
3 replacement design; and
Classification and Requirements, Jan 2011
4.1.3 Post rehabilitation, replacement or new construction
CDRH Regulations CFR 21, Section I, Subchapter J, Parts
4 workmanship verification.
1002 to 1040.11
4.2 Alaser profile pre-acceptance and condition assessment
survey provides significant information in a clear and concise
3. Terminology
manner, including but not limited to graphs and still frame
3.1 Definitions of Terms Specific to This Standard:
digital images of pipe condition prior to acceptance, thereby
3.1.1 authority, n—party reasonable for the generation and
providingobjectivedataontheinstalledqualityandpercentage
verification of performance to job specification(s) and contract
ovality, or degree of deformation, deflection or deviation, that
requirements.
is often not possible from an inspection by either a mandrel or
3.1.2 barrel distortion, n—distortion of an image produced
only CCTV.
by an optical system that causes straight lines at image margins
to bulge outwards.
5. Contract Responsibilities
3.1.3 CCTV, n—aclosedcircuitpipelineandconduitinspec-
5.1 Apartfromtheprovisionsgenerallyincludedinatesting
tion television system including a camera, camera transporter,
and certification contract, the laser profiling survey contract
integrated lighting, central control system, video monitor and
shall define and assign responsibilities for the following items:
recording device.
5.2 Access to the survey site to be provided to the extent
3.1.4 laser, n—a solid state device that produces a mono-
that the contracting authority can provide such access.
chromatic and coherent beam of visible light in an intense,
5.3 Theutilityownershallensurethatalllinestobeprofiled
narrow beam.
are free of debris, obstructions, and cleaned within 24 h prior
3.1.5 laser light diode, n—a mobile, certified “eye safe”
to the profiling inspection. Standing or flowing water or debris
laser light source and internal optics capable of projecting a
shall not exceed 10 % of the nominal pipe diameter, or six (6)
narrow beam of laser light onto an internal pipe wall in pipes
in. (150 mm) in depth, whichever is the lesser.
from 6 to 72 in. (150 to 1800 mm) in diameter regardless of
material, design, or shape.
6. Equipment
3.1.6 laser profile, n—the spatial intensity profile of a laser
6.1 The laser profiling equipment, including laser diode(s)
beam at a particular plane transverse to the beam propagation
and CCTV inspection system (Fig. 1) shall be configured and
path.
calibratedbythelaserprofilingequipmentmanufacturer,asper
the same specifications of the equipment “Certificate of Accu-
3.1.7 laser profiling survey, n—asurveycomposedoftaking
racy” in accordance with Section 9.
measurements of the cross sectional shape of the pipe at
various stations along its alignment, processing the data using
6.2 Only calibration and laser distance software algorithms,
an appropriate software and producing a condition assessment
as specified by the software manufacturer, shall be used as per
report using laser profiling technology.
the same specifications of the equipment “Certificate of Accu-
racy.”
3.1.8 ovality, n—percentage of shape deflection in circular
and noncircular pipes as calculated per Practice F1216 or
Practice F2019 as defined in Annex A1.
3.1.9 profiling software, n—the software that analyzes the
collected data from a laser profiling effort into deformed cross
sectional profiles along the pipe of conduit alignment.
4. Significance and Use
4.1 Laser profiling assessment is a quality control tool for
identifying and quantifying deformation, physical damage, and
other pipe anomalies after installation, providing means and
Available from International Electrotechnical Commission (IEC), 3 rue de
Varembé, P.O. Box 131, CH-1211 Geneva 20, Switzerland, http://www.iec.ch.
Available from Center for Devices and Radiological Health (CDRH), Food and
Drug Administration, 10903 New Hampshire Avenue, WO66-4621, Silver Spring,
MD 20993, http://www.fda.gov. FIG. 1 CCTV—Rotating Laser Profile Camera in Pipe
F3095 − 17a (2022)
6.3 The profiling survey software shall be a version that rotations) images shall be taken every inch or two full laser
meets or exceeds the contents of this practice. ring (complete rotations) images per centimeter along the
longitudinal distance of the pipeline.Any filtering or averaging
6.4 The processing computer shall be equal to or exceed
of readings and data points shall be reported and approved by
those specified by the software manufacturer.
the engineer. In areas where the pipe exceeds performance
7. Software limits, a full ring (stationary scan) shall be
...

Questions, Comments and Discussion

Ask us and Technical Secretary will try to provide an answer. You can facilitate discussion about the standard in here.