Standard Practice for Reporting Particle Size Characterization Data

SIGNIFICANCE AND USE
3.1 When evaluating the particle size information, if the procedures of the data processing are not available, the user of the data must make assumptions concerning the reported data in the event of analytical inconsistencies. In order for different data sets to be compared it is crucial that the parties report the analytical techniques and methods or procedures for evaluating, calculating, compiling or otherwise processing the data to be reported.  
3.2 Particle size characterization information can be reported in three levels of detail in order to satisfy user's needs.  
3.2.1 Level 1 applies when only basic information about the material is required, and shall be provided with each shipment. This level represents the minimum information that shall be reported. Level 1 information may be sufficient in such cases as identifying a certain grade of a material or when detailed knowledge of analytical methodology is not needed.  
3.2.2 Level 2 presumes the need for knowledge of methodology on the user's part and allows the user to make a more informed judgment about the information provided in Level 1.  
3.2.3 Level 3 provides detailed written procedures to allow duplication of the measurement.  
3.2.4 Information provided through Levels 2 and 3 will allow users to perform comparative material evaluations among several suppliers, set specifications or define a purchase agreement, perform inter-laboratory studies and most importantly resolve disputes among suppliers and users.  
3.3 Reported particle size measurement is a function of both the actual particle dimension and shape factor as well as the particular physical or chemical properties of the particle being measured. Caution is required when comparing data from instruments operating on different physical or chemical parameters or with different particle size measurement ranges. Sample acquisition, handling and preparation can also affect the reported particle size results.
SCOPE
1.1 This practice covers reporting particle size measurement data.  
1.2 This practice applies to particle size measurement methods, devices, detail levels, and data formats for dry powders, and wet suspensions of solids, gels, or emulsion droplets. This practice does not pertain to liquid particles. Note 1—For information on reporting liquid particle measurement data, refer to Practice E799.  
1.3 This practice does not concern particle concentration information.  
1.4 This practice uses SI (Système International) units as standard. State all numerical values in terms of SI units unless specific instrumentation software reports particle size information, including percentiles, indices, and distributions as tabulations and graphs using alternate units. In this case, present both reported and equivalent SI units in the final written report. Refer to Practice E380 for proper usage of SI units.  
1.5 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use.

General Information

Status
Historical
Publication Date
31-Mar-2014
Current Stage
Ref Project

Relations

Buy Standard

Standard
ASTM E1617-09(2014)e1 - Standard Practice for Reporting Particle Size Characterization Data
English language
7 pages
sale 15% off
Preview
sale 15% off
Preview

Standards Content (Sample)


NOTICE: This standard has either been superseded and replaced by a new version or withdrawn.
Contact ASTM International (www.astm.org) for the latest information
´1
Designation: E1617 − 09 (Reapproved 2014)
Standard Practice for
Reporting Particle Size Characterization Data
This standard is issued under the fixed designation E1617; the number immediately following the designation indicates the year of
original adoption or, in the case of revision, the year of last revision. A number in parentheses indicates the year of last reapproval. A
superscript epsilon (´) indicates an editorial change since the last revision or reapproval.
ε NOTE—Editorial corrections were made throughout in April 2014.
INTRODUCTION
Correlation and comparison of particle size measurement data are of general importance to
researchers, processors, suppliers, vendors, and users of particulate materials. Since there are
numerous methods and devices in use, comparisons of size measurement data are subject to perceived
inconsistencies that may be much reduced by the use of standardized reporting.
Data generated by any one size measurement method may be consistent in many respects, yet be
troublesome to correlate due to variances in data processing and reporting formats among equipment
manufacturers. Data generated from methods employing different physical principles present serious
correlation problems due to their sensing of different parameters related to particle size and some
unrelated, such as density, shape and optical properties. Standardized reporting may reduce confusion
resulting from different particle size measurement practices.
1. Scope responsibility of the user of this standard to establish appro-
priate safety and health practices and determine the applica-
1.1 Thispracticecoversreportingparticlesizemeasurement
bility of regulatory limitations prior to use.
data.
1.6 This international standard was developed in accor-
1.2 This practice applies to particle size measurement
dance with internationally recognized principles on standard-
methods, devices, detail levels, and data formats for dry
ization established in the Decision on Principles for the
powders, and wet suspensions of solids, gels, or emulsion
Development of International Standards, Guides and Recom-
droplets. This practice does not pertain to liquid particles.
mendations issued by the World Trade Organization Technical
NOTE 1—For information on reporting liquid particle measurement
Barriers to Trade (TBT) Committee.
data, refer to Practice E799.
1.3 This practice does not concern particle concentration
2. Referenced Documents
information.
2.1 ASTM Standards:
1.4 This practice uses SI (Système International) units as
E177 Practice for Use of the Terms Precision and Bias in
standard. State all numerical values in terms of SI units unless
ASTM Test Methods
specific instrumentation software reports particle size
E380 Practice for Use of the International System of Units
information, including percentiles, indices, and distributions as
(SI) (the Modernized Metric System) (Withdrawn 1997)
tabulations and graphs using alternate units. In this case,
E456 Terminology Relating to Quality and Statistics
present both reported and equivalent SI units in the final
E691 Practice for Conducting an Interlaboratory Study to
written report. Refer to Practice E380 for proper usage of SI
Determine the Precision of a Test Method
units.
E799 Practice for Determining Data Criteria and Processing
1.5 This standard does not purport to address all of the
for Liquid Drop Size Analysis
safety concerns, if any, associated with its use. It is the
1 2
This practice is under the jurisdiction ofASTM Committee E29 on Particle and For referenced ASTM standards, visit the ASTM website, www.astm.org, or
Spray Characterization and is the direct responsibility of Subcommittee E29.02 on contact ASTM Customer Service at service@astm.org. For Annual Book of ASTM
Non-Sieving Methods. Standards volume information, refer to the standard’s Document Summary page on
Current edition approved April 1, 2014. Published April 2014. Originally the ASTM website.
approved in 1994. Last previous edition approved in 2009 as E1617 – 09. DOI: The last approved version of this historical standard is referenced on
10.1520/E1617-09R14E01. www.astm.org.
Copyright © ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959. United States
´1
E1617 − 09 (2014)
3. Significance and Use 4.4.3 The report shall include a minimum reporting of size
parameter(s).Minimumreportingwillbedefinedbyagreement
3.1 When evaluating the particle size information, if the
between supplier and user of the data, such as a calculated
procedures of the data processing are not available, the user of
mean diameter, median diameter, or both.
the data must make assumptions concerning the reported data
4.4.4 The report shall include the principle of measurement,
in the event of analytical inconsistencies. In order for different
such as sieving, sedimentation, light blockage, electronic zone
data sets to be compared it is crucial that the parties report the
sensing, laser light scattering, and so forth. Often, the mea-
analytical techniques and methods or procedures for
surement principle can be defined by describing the instrumen-
evaluating, calculating, compiling or otherwise processing the
tation or apparatus used for the analysis. All references to
data to be reported.
“diameter” shall include appropriate adjectives to indicate the
3.2 Particle size characterization information can be re-
sensing basis.
ported in three levels of detail in order to satisfy user’s needs.
4.4.5 The report shall include the bases for the reported
3.2.1 Level 1 applies when only basic information about the
parameters. Examples of parameter basis are frequently
material is required, and shall be provided with each shipment.
distribution, scattering area distribution, and mass distribution.
This level represents the minimum information that shall be
4.4.6 The report shall include the range of size
reported.Level1informationmaybesufficientinsuchcasesas
measurement, in appropriate SI units.
identifying a certain grade of a material or when detailed
4.5 Particle Size Characterization Data Sheet—Level 2:
knowledge of analytical methodology is not needed.
4.5.1 The Level 2 report shall include all information listed
3.2.2 Level 2 presumes the need for knowledge of method-
under 4.4.
ology on the user’s part and allows the user to make a more
4.5.2 The report shall include the manufacturer’s identifica-
informed judgment about the information provided in Level 1.
tion of the instrument or apparatus used to perform the particle
3.2.3 Level 3 provides detailed written procedures to allow
size measurement.
duplication of the measurement.
4.5.3 The report shall include all of the parameters required
3.2.4 Information provided through Levels 2 and 3 will
allow users to perform comparative material evaluations to use the instrument or apparatus.
4.5.4 The report shall include the software version num-
among several suppliers, set specifications or define a purchase
agreement, perform inter-laboratory studies and most impor- ber(s) of any computerized instrument used to measure particle
tantly resolve disputes among suppliers and users. size.
4.5.5 The report shall include a description of the basis of
3.3 Reported particle size measurement is a function of both
calculations performed by the instrument, to the extent avail-
the actual particle dimension and shape factor as well as the
able from the instrument manufacturer. Also, the report shall
particular physical or chemical properties of the particle being
include description of any additional calculations performed
measured. Caution is required when comparing data from
outside of the instrument software package.
instruments operating on different physical or chemical param-
4.5.6 The report shall include basic statistical information,
eters or with different particle size measurement ranges.
such as standard deviation, number of degrees of freedom, and
Sample acquisition, handling and preparation can also affect
confidence interval, and shall include a description of the
the reported particle size results.
calculations used to produce this statis
...

Questions, Comments and Discussion

Ask us and Technical Secretary will try to provide an answer. You can facilitate discussion about the standard in here.