ASTM A977/A977M-07(2013)
(Test Method)Standard Test Method for Magnetic Properties of High-Coercivity Permanent Magnet Materials Using Hysteresigraphs
Standard Test Method for Magnetic Properties of High-Coercivity Permanent Magnet Materials Using Hysteresigraphs
SIGNIFICANCE AND USE
4.1 This test method is suitable for magnet specification, acceptance, service evaluation, quality control in magnet production, research and development, and design.
4.2 When a test specimen is cut or fabricated from a larger magnet, the magnetic properties measured on it are not necessarily exactly those of the original sample, even if the material is in the same condition. In such instances, the test results must be viewed in context of part performance history.
4.3 Tests performed in general conformity to this test method and even on the same specimen, but using different test systems, may not yield identical results. The main source of discrepancies are variations between the different test systems in the geometry of the region surrounding the sample, such as, size and shape of the electromagnet pole caps (see Annex A1 and Appendix X1), air gaps at the specimen end faces, and especially the size and location of the measuring devices for H and B or for their corresponding flux values (Hall-effect probes, inductive sensing coils). Also important is the method of B calibration, for example, a volt-second calibration of the fluxmeter alone versus an overall system calibration using a physical reference sample. The method of B and H sensing should be indicated in test reports (see Section 9).
SCOPE
1.1 This test method covers how to determine the magnetic characteristics of magnetically hard materials (permanent magnets), particularly their initial magnetization, demagnetization, and recoil curves, and such quantities as the residual induction, coercive field strength, knee field, energy product, and recoil permeability. This test method is suitable for all materials processed into bulk magnets by any common fabrication technique (casting, sintering, rolling, molding, and so forth), but not for thin films or for magnets that are very small or of unusual shape. Uniformity of composition, structure, and properties throughout the magnet volume is necessary to obtain repeatable results. Particular attention is paid to the problems posed by modern materials combining very high coercivity with high saturation induction, such as the rare-earth magnets, for which older test methods (see Test Method A341/A341M) are unsuitable. An applicable international standard is IEC Publication 60404-5.
1.2 The magnetic system (circuit) in a device or machine generally comprises flux-conducting and nonmagnetic structural members with air gaps in addition to the permanent magnet. The system behavior depends on properties and geometry of all these components and on the operating temperature. This test method describes only how to measure the properties of the permanent magnet material. The basic test method incorporates the magnetic specimen in a magnetic circuit with a closed flux path. Test methods using ring samples or frames composed entirely of the magnetic material to be characterized, as commonly used for magnetically soft materials, are not applicable to permanent magnets.
1.3 This test method shall be used in conjunction with Practice A34/A34M.
1.4 The values and equations stated in customary (cgs-emu or inch-pound) or SI units are to be regarded separately as standard. Within this test method, SI units are shown in brackets except for the sections concerning calculations where there are separate sections for the respective unit systems. The values stated in each system may not be exact equivalents; therefore, each system shall be used independently of the other. Combining values from the two systems may result in nonconformance with this test method.
1.5 The names and symbols of magnetic quantities used in this test method, summarized in Table 1, are those generally accepted by the industry.
1.6 This test method is useful for magnet materials having Hci values between about 100 Oe and 35 kOe [8 kA/m and 2.8 MA/m], and Br values in the approximate range from 500 G to 20 kG [50 mT to ...
General Information
Relations
Standards Content (Sample)
NOTICE: This standard has either been superseded and replaced by a new version or withdrawn.
Contact ASTM International (www.astm.org) for the latest information
Designation: A977/A977M − 07 (Reapproved 2013)
Standard Test Method for
Magnetic Properties of High-Coercivity Permanent Magnet
Materials Using Hysteresigraphs
This standard is issued under the fixed designationA977/A977M; the number immediately following the designation indicates the year
of original adoption or, in the case of revision, the year of last revision. A number in parentheses indicates the year of last reapproval.
A superscript epsilon (´) indicates an editorial change since the last revision or reapproval.
1. Scope brackets except for the sections concerning calculations where
there are separate sections for the respective unit systems. The
1.1 This test method covers how to determine the magnetic
values stated in each system may not be exact equivalents;
characteristics of magnetically hard materials (permanent
therefore,eachsystemshallbeusedindependentlyoftheother.
magnets), particularly their initial magnetization,
Combining values from the two systems may result in noncon-
demagnetization, and recoil curves, and such quantities as the
formance with this test method.
residual induction, coercive field strength, knee field, energy
product, and recoil permeability. This test method is suitable
1.5 The names and symbols of magnetic quantities used in
for all materials processed into bulk magnets by any common
this test method, summarized in Table 1, are those generally
fabrication technique (casting, sintering, rolling, molding, and
accepted by the industry.
so forth), but not for thin films or for magnets that are very
1.6 This test method is useful for magnet materials having
small or of unusual shape. Uniformity of composition,
H values between about 100 Oe and 35 kOe [8 kA/m and 2.8
structure, and properties throughout the magnet volume is ci
MA/m], and B values in the approximate range from 500 G to
necessary to obtain repeatable results. Particular attention is r
20 kG [50 mT to 2 T]. High-coercivity rare-earth magnet test
paid to the problems posed by modern materials combining
specimens may require much higher magnetizing fields than
very high coercivity with high saturation induction, such as the
iron-core electromagnets can produce. Such samples must be
rare-earth magnets, for which older test methods (see Test
premagnetized externally and transferred into the measuring
Method A341/A341M) are unsuitable. An applicable interna-
yoke. Typical values of the magnetizing fields, H , required
tional standard is IEC Publication 60404-5. mag
for saturating magnet materials are shown in Table A2.1.
1.2 The magnetic system (circuit) in a device or machine
1.7 This standard does not purport to address all of the
generally comprises flux-conducting and nonmagnetic struc-
tural members with air gaps in addition to the permanent safety concerns, if any, associated with its use. It is the
responsibility of the user of this standard to establish appro-
magnet. The system behavior depends on properties and
geometry of all these components and on the operating priate safety and health practices and determine the applica-
bility of regulatory limitations prior to use.
temperature. This test method describes only how to measure
the properties of the permanent magnet material.The basic test
method incorporates the magnetic specimen in a magnetic 2. Referenced Documents
circuitwithaclosedfluxpath.Testmethodsusingringsamples
2.1 ASTM Standards:
or frames composed entirely of the magnetic material to be
A34/A34M Practice for Sampling and Procurement Testing
characterized, as commonly used for magnetically soft
of Magnetic Materials
materials, are not applicable to permanent magnets.
A340 Terminology of Symbols and Definitions Relating to
1.3 This test method shall be used in conjunction with
Magnetic Testing
Practice A34/A34M.
A341/A341M Test Method for Direct Current Magnetic
Properties of Materials Using D-C Permeameters and the
1.4 The values and equations stated in customary (cgs-emu
Ballistic Test Methods
or inch-pound) or SI units are to be regarded separately as
E177 Practice for Use of the Terms Precision and Bias in
standard. Within this test method, SI units are shown in
ASTM Test Methods
This test method is under the jurisdiction of ASTM Committee A06 on
MagneticPropertiesandisthedirectresponsibilityofSubcommitteeA06.01onTest
Methods. For referenced ASTM standards, visit the ASTM website, www.astm.org, or
CurrenteditionapprovedMay1,2013.PublishedJuly2013.Originallyapproved contact ASTM Customer Service at service@astm.org. For Annual Book of ASTM
in 1997. Last previous edition approved in 2007 as A977/A977M–07. DOI: Standards volume information, refer to the standard’s Document Summary page on
10.1520/A0977_A0977M-07R13. the ASTM website.
Copyright © ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959. United States
A977/A977M − 07 (2013)
TABLE 1 Symbols, Quantities, and Units
especially the size and location of the measuring devices for H
and B or for their corresponding flux values (Hall-effect
NOTE 1—IEC nomenclature calls B “remanence,” when B represents
r r
probes, inductive sensing coils). Also important is the method
the B at H = 0 of the outermost hysteresis loop, and it calls B “remanent
r
magnetic induction” for B at H = 0 at smaller loops.
of B calibration, for example, a volt-second calibration of the
Customary fluxmeter alone versus an overall system calibration using a
Symbol Quantity SI Unit
cgs-emu
physical reference sample. The method of B and H sensing
2 2
A Cross section of search coil [m]cm
t
should be indicated in test reports (see Section 9).
B Magnetic induction at BH [T] G
d max
B Magnetic induction at low point of [T] G
rec
recoil loop
5. Measuring Methods and Apparatus
B Magnetic induction at remanence [T] G
r
5.1 Measuring Flux and Induction (Flux Density):
d Diameter of pole piece [m] cm
l
d Diameter of homogeneous field [m] cm
5.1.1 In the preferred B-measuring method, the total flux is
H Magnetic field strength at BH [A/m] Oe
d max
measured with a sensing coil (search coil) that surrounds the
H Magnetic field strength at low [A/m] Oe
p
point of test specimen and is wound as closely as possible to the
recoil loop
specimen surface. Its winding length should be no more than a
l Distance between pole faces [m] cm
third of the specimen length, preferably less than one fifth, and
l Length of test sample [m] cm
r
N Number of turns of test coil must be centered on the specimen. The leads shall be twisted
e Voltage induced in test coil V V
tightly.Asthefluxchangesinresponsetosweepingtheapplied
d Total air gap between test [m] cm
field, H, the total flux is measured by taking the time integral
sample and
pole faces
of the voltage induced in this coil. This measurement is taken
µ A constant with value µ =4π
0 0
with a fluxmeter. Modern hysteresigraphs use electronic inte-
-7
grating fluxmeters that allow convenient continuous integra-
H/m
µ Recoil permability
tion and direct graphic recording of magnetization curves. If
rec
the signal is large enough, high-speed voltage sampling at the
coil and digital integration is also possible.
2.2 Magnetic Materials Procedure Association Standard:
5.1.2 The magnetic induction B is determined by dividing
MMPANo. 0100–00 Standard Specifications for Permanent
the total flux by the area-turns product NA of the B-sensing
Magnet Materials
coil. For permanent magnets in general, and especially for
2.3 International Electrotechnical Commission Document:
high-coercivitymaterials,anair-fluxcorrectionisrequired(see
Publication 60404-5 Magnetic Materials– Part 5: Permanent
5.1.5).
Magnet (Magnetically Hard) Materials – Methods of
5.1.3 Thetotalerrorofmeasuring Bshallbenotgreaterthan
Measurement of Magnetic Properties
62%.
5.1.4 The change of magnetic induction, ∆ B = B – B,in
2 1
3. Terminology
thetimeintervalbetweenthetimes t and t isgivenasfollows:
1 2
3.1 Basic magnetic units are defined in Terminology A340
t
∆ B 5 ~10 /AN! edt ~customary units! (1)
*
and MMPA No. 0100–00. Additional definitions with symbols
t
and units are given in Table 1 and Figs. 1-3 of this test method.
t
∆ B 5 1/AN edt SI units (2)
~ ! * ~ !
t
4. Significance and Use
where:
4.1 This test method is suitable for magnet specification,
B = magnetic induction, G [T];
2 2
acceptance, service evaluation, quality control in magnet
A = cross-sectional area of the test specimen, cm [m ];
production, research and development, and design.
N = number of turns on the B-sensing coil;
e = voltage induced in the coil, V;
4.2 When a test specimen is cut or fabricated from a larger
t = time, s; and
magnet, the magnetic properties measured on it are not
t
* edt = voltage integral = flux, V-s [Weber].
t
necessarily exactly those of the original sample, even if the 1
material is in the same condition. In such instances, the test
5.1.5 The change in the magnetic induction shall be cor-
results must be viewed in context of part performance history.
rectedtotakeintoaccounttheairfluxoutsidethetestspecimen
that is linked by the sensing coil. The corrected change, B ,
corr
4.3 Tests performed in general conformity to this test
is given as follows:
methodandevenonthesamespecimen,butusingdifferenttest
t
systems, may not yield identical results. The main source of
∆ B 5 10 /AN edt 2 ∆ H A 2 A /A customary units
~ ! * ~ ! ~ !
corr t
t
discrepancies are variations between the different test systems 1
(3)
in the geometry of the region surrounding the sample, such as,
t
size and shape of the electromagnet pole caps (see Annex A1 2
∆B 5 1/AN edt 2 µ ∆H A 2 A /A SI units (4)
~ ! * ~ ! ~ !
corr 0 t
t
and Appendix X1), air gaps at the specimen end faces, and
where:
Available from Magnetic Materials ProducersAssociation, 8 S. MichiganAve., A = average cross-sectional area of the sensing coil,
2 2
Suite 1000, Chicago, IL 60603.
cm [m ];
Available from International Electrotechnical Commission (IEC), 3 rue de
∆ H = change in field from t until t , Oe [A/m]; and
1 2
Varembé, P.O. Box 131, CH-1211, Geneva 20, Switzerland.
A977/A977M − 07 (2013)
FIG. 1 Normal and Intrinsic Hysteresis Loops and Initial Magnetization Curves for Permanent Magnet Materials Illustrating Two Ex-
tremes of Virgin Sample Behavior
-7
5.2.2 The change of intrinsic induction in the test specimen
µ = magnetic constant [4π 10 H/m].
can be determined by integrating the voltage induced in a
5.2 Determining Intrinsic Induction :
device comprising two sensing coils, both subject to the same
5.2.1 For high-coercivity magnets, it is more convenient to
applied field H, where the test specimen is contained in only
sense directly an electrical signal proportional to the intrinsic
one of the coils (Coil 1). If each individual coil has the same
induction, derive the average B by dividing this flux by the
i
area-turns product, and if the windings are connected electri-
area-turns product of the surrounding B coil, and to plot B
i
cally in opposition, the signal induced by the flux linking Coil
versus H. B then is obtained by mathematical or electronic
2 (not containing the specimen) will compensate for the output
addition of H to B.
A977/A977M − 07 (2013)
5.2.3 The two-sensing-coil device shall lie totally within the
homogeneous field defined by Eq A1.1 and Eq A1.2. Test
specimens of lower-coercivity magnets having a range of
cross-sectional areas and shapes can then be measured with the
samecoildevice.Anarrangementofside-by-sidecoilsofequal
size is useful. Serious errors, however, are incurred when
measuring B this way on high-B or high/coercivity magnets,
i r
or both, at applied fields of about 10 kOe or more. The errors
are most severe for test specimens of short pole-to-pole length.
Local pole-piece saturation causes strong field inhomogene-
ities. The specimen then must fill the cross section of Coil 1,
andCoil2mustbeathinandflatcoil,oracoaxialannularcoil,
either centered on the specimen or in close proximity to its
surface (see 5.3).
5.2.4 The total error of measuring B shall be not greater
i
than 62%.
5.3 Measuring the Magnetic Field Strength:
5.3.1 For correct magnetization curves, one should know
FIG. 2 Normal and Intrinsic Demagnetization Curves with Sym-
bols for Special Points of Interest and Definition of Salient Prop- the magnetic field strength, H, inside the test specimen,
erties. Illustration of Maximum Energy Product, Coercive Fields,
averaged over the specimen volume if H is not uniform. But
and Definition of Knee Field
this inner field cannot be measured. At the surface of the test
specimen, H is equal to the local field strength just inside the
specimen in those locations (and only there) where the H
vectorisparalleltothesidesurfaceofthespecimen.Therefore,
a magnetic field strength sensor of small dimensions relative to
the specimen is placed near the specimen surface and sym-
metrical with respect to the end faces, covering the shortest
possible center portion of the specimen length. It shall be so
oriented that it correctly measures the tangential field compo-
nent.
5.3.2 To determine the magnetic field strength, a flat surface
coil, a tightly fitted annular coil, a magnetic potentiometer, or
a Hall probe is used together with suitable instruments. The
dimensions of the magnetic field sensor and its location shall
be such that it is within an area of limited diameter around the
test specimen (see Annex A1).
5.3.3 The provisions of 5.3.2 are adequate for measure-
ments on magnets having low-to-moderate intrinsic coercivity,
such as Alnico and bonded ferrites. For high-coercivity, dense
ferrites and especially for most rare earth-transition metal
FIG. 3 Normal and Intrinsic Demagnetization Curves with Sym-
materials, it is essential for accurate measurement to use thin
bols for Special Points of Interest and Definition of Salient Prop-
erties. Illustration of Recoil Loop. Recoil Permeability is Defined
flat or radially thin annular H-sensing coils of short length
as µ = ∆B/∆H
rec
(<1/5 to 1/3 of the specimen length), centered on the specimen
and placed as close as possible to the specimen surface.
5.3.4 The same considerations apply to the H-flux compen-
of Coil 1 except for B within the test specimen. The change of
i
sation coil used in B measurements (see 5.2.3.) When pole
i
intrinsic induction in the specimen then is given as follows:
saturation can occur, Coil 2 also shall be a thin conforming flat
t
8 surface coil for rectangular specimen shapes or a thin annular
∆ B 5 10 / AN edt customary units (5)
~ ! * ~ !
i
t
coil closely surrounding a cyclindrical specimen, and the
t
specimen essentially shall fill the open cross-sectional area of
∆B 5 ~1/AN! edt ~SI units! (6)
*
i
t
t
...