Workplace atmospheres - Assessment of performance of instruments for measurement of airborne particle concentrations

This European Standard specifies methods for testing aerosol sampling instruments under prescribed laboratory conditions, and performance requirements that are specific to aerosol sampling instruments. These performance requirements, which include conformity with the EN 481 sampling conventions, apply only to the process of sampling the airborne particles from the air, not to the process of analysing particles collected by the process of sampling. Although analysis of samples collected in the course of testing is usually necessary in order to evaluate the sampler performance, the specified test methods ensure that analytical errors are kept very low during testing and do not contribute significantly to the end result. The determination of analytical errors and factors related to them (for example the bias, precision and limit of detection of the analytical method) is outside the scope of this standard. Where the aerosol sampling instrument requires the use of an external (rather than integral) pump, the pump is not subject to the requirements of this standard.
EN 482 contains general performance requirements for methods used for determining the concentrations of chemical agents in workplace atmospheres. These performance requirements include maximum values of overall uncertainty  (a combination of precision and bias) achievable under prescribed laboratory conditions for the methods to be used. The requirements of EN 482 apply to the combined results of sampling airborne particles and analysing collected particles. This standard specifies how the performance of aerosol measurement methods is assessed with respect to the general requirements of EN 482, through the combination of sampling and analytical errors.

Arbeitsplatzatmosphäre - Bewertung der Leistungsfähigkeit von Geräten für die Messung der Konzentration luftgetragener Partikel

Diese Europäische Norm legt Verfahren zur Prüfung von Probenahmeeinrichtungen für Aerosole unter vor-geschriebenen
Laboratoriumsbedingungen fest und gibt Leistungsanforderungen an, die für Aerosol-Probe-nahmeeinrichtungen
spezifisch sind. Diese Leistungsanforderungen, zu denen die Übereinstimmung mit den
Konventionen der EN 481 für die Probenahme gehört, gelten nur für das Verfahren der Probenahme der
luftgetragenen Partikeln aus der Luft, nicht für den Prozess der Untersuchung der Partikeln, die im Prozess der
Probenahme gesammelt wurden. Obgleich die Untersuchung der im Verlauf der Prüfung gesammelten Proben
im Allgemeinen erforderlich ist, um die Leistungsfähigkeit der Probenahmeeinrichtung zu bewerten, sichern die
festgelegten Prüfverfahren, dass Untersuchungsfehler während der Prüfung sehr niedrig gehalten werden und
nicht signifikant zum Endergebnis beitragen. Die Bestimmung analytischer Fehler und mit ihnen verbundener
Faktoren (zum Beispiel die systematische Messabweichung, die Präzision und die Nachweisgrenze des
analytischen Verfahrens) liegt außerhalb des Anwendungsbereichs dieser Norm. Wenn die Aerosol-Probenahme-einrichtung
die Verwendung einer äußeren Pumpe anstatt einer eingebauten Pumpe erfordert, unterliegt die
Pumpe nicht den Anforderungen dieser Norm.
EN 482 enthält allgemeine Leistungsanforderungen für die Verfahren, die zur Bestimmung der Konzentration
chemischer Stoffe in Arbeitsplatzatmosphären angewendet werden. Zu diesen Leistungsanforderungen gehören
Maximalwerte für die Gesamtunsicherheit (einer Kombination von Präzision und systematischer Messabweichung),
die bei den zu verwendenden Verfahren unter vorgeschriebenen Laboratoriumsbedingungen erreichbar sind.
Die Anforderungen von EN 482 gelten für die kombinierten Ergebnisse der Probenahme der luftgetragenen
Partikeln und der Untersuchung der gesammelten Teilchen

Atmosphères des lieux de travail - Evaluation des performances des intruments de mesurage des concentrations d'aérosol

Cette norme spécifie les méthodes de test des instruments d'échantillonnage des concentrations d'aérosol soumis à des conditions de laboratoire imposées, ainsi que les exigences requises pour ces instruments d'échantillonnage d'aérosols. Ces exigences qui comprennent notamment la conformité aux conventions d'échantillonnage EN 481, s'appliquent uniquement au processus d'échantillonnage des particules d'aérosol contenues dans l'air, et non pas au processus d'analyse des particules recueillies dans le cadre du processus d'échantillonnage. L'analyse des échantillons recueillis au cours des essais est généralement nécessaire pour évaluer les performances de l'échantillonneur ; toutefois, les méthodes d'essai spécifiées permettent de conserver un très faible taux d'erreurs analytiques au cours des essais, qui n'ont pas beaucoup d'incidence sur le résultat final. La détermination des erreurs analytiques (par exemple le biais, la fidélité et la limite de détection de la méthode d'analyse) n'entre pas dans le domaine d'application de la présente norme européenne. Certains échantillonneurs ont une pompe intégrée, d'autres non. Lorsque l'instrument nécessite l'utilisation d'une pompe externe (plutôt qu'intégrée), cette pompe n'est pas soumise aux exigences de la présente norme.
L'EN 482 indique les exigences requises pour les méthodes utilisées afin de déterminer les concentrations d'agents chimiques dans les atmosphères des lieux de travail. Ces exigences comprennent notamment la détermination des valeurs maximales de l'incertitude globale (qui combine précision et biais) pouvant être obtenues dans le cadre des conditions de laboratoire prescrites pour les méthodes à utiliser. Les exigences de l'EN 482 s'appliquent aux résultats combinés de l'échantillonnage des particules d'aérosol et de l'analyse des particules recueillies.

Zrak na delovnem mestu - Ocenjevanje lastnosti merilnikov za merjenje koncentracij lebdečih delcev

General Information

Status
Withdrawn
Publication Date
11-Dec-2001
Withdrawal Date
17-Jun-2014
Current Stage
9960 - Withdrawal effective - Withdrawal
Completion Date
18-Jun-2014

Relations

Buy Standard

Standard
EN 13205:2002
English language
46 pages
sale 10% off
Preview
sale 10% off
Preview
e-Library read for
1 day

Standards Content (Sample)


SLOVENSKI STANDARD
01-maj-2002
=UDNQDGHORYQHPPHVWX2FHQMHYDQMHODVWQRVWLPHULOQLNRY]DPHUMHQMH
NRQFHQWUDFLMOHEGHþLKGHOFHY
Workplace atmospheres - Assessment of performance of instruments for measurement
of airborne particle concentrations
Arbeitsplatzatmosphäre - Bewertung der Leistungsfähigkeit von Geräten für die Messung
der Konzentration luftgetragener Partikel
Atmospheres des lieux de travail - Evaluation des performances des intruments de
mesurage des concentrations d'aérosol
Ta slovenski standard je istoveten z: EN 13205:2001
ICS:
13.040.30 Kakovost zraka na delovnem Workplace atmospheres
mestu
2003-01.Slovenski inštitut za standardizacijo. Razmnoževanje celote ali delov tega standarda ni dovoljeno.

EUROPEAN STANDARD
EN 13205
NORME EUROPÉENNE
EUROPÄISCHE NORM
December 2001
ICS 13.040.30
English version
Workplace atmospheres - Assessment of performance of
instruments for measurement of airborne particle concentrations
Atmosphères des lieux de travail - Evaluation des Arbeitsplatzatmosphäre - Bewertung der Leistungsfähigkeit
performances des intruments de mesurage des von Geräten für die Messung der Konzentration
concentrations d'aérosols luftgetragener Partikel
This European Standard was approved by CEN on 16 November 2001.
CEN members are bound to comply with the CEN/CENELEC Internal Regulations which stipulate the conditions for giving this European
Standard the status of a national standard without any alteration. Up-to-date lists and bibliographical references concerning such national
standards may be obtained on application to the Management Centre or to any CEN member.
This European Standard exists in three official versions (English, French, German). A version in any other language made by translation
under the responsibility of a CEN member into its own language and notified to the Management Centre has the same status as the official
versions.
CEN members are the national standards bodies of Austria, Belgium, Czech Republic, Denmark, Finland, France, Germany, Greece,
Iceland, Ireland, Italy, Luxembourg, Netherlands, Norway, Portugal, Spain, Sweden, Switzerland and United Kingdom.
EUROPEAN COMMITTEE FOR STANDARDIZATION
COMITÉ EUROPÉEN DE NORMALISATION
EUROPÄISCHES KOMITEE FÜR NORMUNG
Management Centre: rue de Stassart, 36  B-1050 Brussels
© 2001 CEN All rights of exploitation in any form and by any means reserved Ref. No. EN 13205:2001 E
worldwide for CEN national Members.

Contents
page
Foreword.4
Introduction .5
1 Scope .6
2 Normative references.6
3 Terms and definitions .7
4 Requirements.9
4.1 Summary of requirements.9
4.2 Accuracy.9
5 Test methods .10
5.1 Choice of laboratory tests to be used.10
5.2 Overview of test methods.11
6 Types of evaluation .11
7 Instructions for use .12
8 Marking, quality control .12
8.1 Marking.12
8.2 Quality control .13
Annex A (normative) Laboratory testing of samplers with respect to sampling conventions .14
A.1 Principle.14
A.2 Test method.14
A.2.1 General.14
A.2.1 Test conditions .14
A.2.2 Test variables.14
A.2.4 Experimental requirements .16
A.3 Calculation methods .18
A.3.2 Symbols and abbreviated terms .18
A.3.3 Determination of the sampling efficiency .19
A.3.4 Calculation of the sampled aerosol concentration. .19
A.3.5 Calculation of the ideal sampled aerosol concentration .19
A.3.6 Calculation of the sampler bias.19
A.3.7 Application of a sampler correction factor.20
A.3.8 Calculation of the uncertainty in the estimated sampler bias.20
A.3.9 Calculation of the sampler accuracy.20
A.4 Test report .21
A.4.1 Testing laboratory details and sponsoring organisation .21
A.4.2 Description of the tested sampler .21
A.4.3 Critical review of sampling process (see clause 5).21
A.4.4 Laboratory methods used .21
A.4.5 Details of experimental design.22
A.4.6 Presentation of experimental results.22
A.4.7 Data analysis.22
A.4.8 Sampler performance.22
Annex B (normative) Laboratory comparison of instruments.25
B.1 Principle .25
B.2 Test method.25
B.2.2 Test conditions .25
B.2.3 Test variables .26
B.2.4 Experimental requirements .28
B.3 Calculation methods .29
B.3.1 Symbols and abbreviated terms.29
B.3.2 Distribution of ratios .29
B.3.3 Correction factor.29
B.3.4 Accuracy.29
B.3.5 Temperature stability.30
B.3.6 Time stability .30
B.4 Test Report.30
B.4.1 Testing laboratory details and sponsoring organisation .30
B.4.2 Description of the candidate instrument and reference sampler .30
B.4.3 Critical review of sampling process (clause 5).30
B.4.4 Test facilities .30
B.4.5 Details of experimental design.31
B.4.6 Data analysis.31
B.4.7 Candidate instrument performance.31
B.4.8 Summary and information for the user.31
Annex C (informative) Recommended procedure for field comparison of instruments.32
C.1 Principle.32
C.2 Comparison procedure.32
C.2.2 Comparison of two types of personal instrument.33
C.2.3 Comparison of two types of static instrument .33
C.2.4 Periodic validation .33
C.3 Calculation methods .33
C.3.1 Symbols and abbreviated terms .33
C.3.2 Estimation of the correction function.33
C.3.3 Exclusion of outliers .34
C.3.4 Residual uncertainty after transformation by the correction function.34
C.3.5 Equivalence.34
C.4 Documentation .34
C.4.1 General.34
C.4.2 Description of the candidate instrument and reference sampler .
...

Questions, Comments and Discussion