Biological evaluation of medical devices - Part 15: Identification and quantification of degradation products from metals and alloys (ISO/FDIS 10993-15:2019)

NEXT ACTION: TC ACTION BY 2022-10-26 : TC to send a revised annex ZA for assessment at PUB
2020-02-19- JO-CEN/TC 206 to take decision either to remove the link or send a revised annex ZA for assessment at PUB

Biologische Beurteilung von Medizinprodukten - Teil 15: Qualitativer und quantitativer Nachweis von Abbauprodukten aus Metallen und Legierungen (ISO/FDIS 10993-15:2019)

Dieses Dokument legt allgemeine Anforderungen zur Konzeption von Prüfungen für den qualitativen und quantitativen Nachweis von Abbauprodukten von fertigen Medizinprodukten aus Metall oder entsprechenden Werkstoffproben im Endzustand fest, die zur klinischen Anwendung bereit sind.
Dieses Dokument ist nur auf solche Abbauprodukte anwendbar, die durch eine chemische Veränderung des fertigen Metallproduktes in einer In vitro Prüfung erzeugt werden. Aufgrund der Natur dieser In vitro Prüfungen entsprechen die Prüfergebnisse in etwa dem In vivo Verhalten des Implantats oder des Werkstoffs. Die beschriebenen chemischen Verfahren sind ein Mittel, um Abbauprodukte für weitere Bewertungen zu erzeugen.
Dieses Dokument ist anwendbar sowohl auf Materialien, die dafür vorgesehen sind, vom Körper abgebaut zu werden, als auch auf solche, bei denen das nicht vorgesehen ist.
Dieses Dokument gilt nicht für eine Beurteilung des Abbaus durch rein mechanische Prozesse. Methoden zur Generierung dieser Art von Abbauprodukten sind, falls vorhanden, in den spezifischen Produktnormen beschrieben.
ANMERKUNG   Rein mechanischer Abbau verursacht meist partikelförmiges Material. Obwohl dieses aus dem Anwendungsbereich dieses Dokuments ausgeschlossen ist, können solche Abbauprodukte eine biologische Reaktion hervorrufen und einer biologischen Beurteilung, wie in anderen Teilen der ISO 10993 beschrieben, unterzogen werden.
In Anbetracht des breiten Spektrums an metallischen Werkstoffen, die für Medizinprodukte verwendet werden, werden keine spezifischen analytischen Verfahren zum quantitativen Nachweis der Abbauprodukte festgelegt. Der Nachweis von Spurenelementen (< 10−6 w/w), die in bestimmten Metallen oder Legierungen vorhanden sind, wird in diesem Dokument nicht beschrieben. Ebenso legt dieses Dokument keine spezifischen Anforderungen für akzeptierbare Toleranzwerte von Abbauprodukten fest.
Dieses Dokument beschreibt nicht die biologische Aktivität der Abbauprodukte. (Hierzu wird auf die anwendbaren Abschnitte von ISO 10993 1 und ISO 10993 17 verwiesen.)

Évaluation biologique des dispositifs médicaux - Partie 15: Identification et quantification des produits de dégradation issus des métaux et alliages (ISO/FDIS 10993-15:2019)

Le présent document spécifie les exigences générales pour la conception des essais d'identification et de quantification des produits de dégradation issus de dispositifs médicaux métalliques finaux ou d'échantillons des matériaux correspondants prêts pour une utilisation clinique.
Le présent document ne s'applique qu'aux produits de dégradation créés par une modification chimique du dispositif métallique final et au moyen d'un essai de dégradation in vitro. En raison de la nature des essais in vitro, les résultats d'essai donnent une approximation du comportement in vivo de l'implant ou du matériau. En raison du caractère accéléré de ces essais, la méthodologie chimique décrite est un moyen de générer des produits de dégradation en vue d'analyses ultérieures.
Le présent document s'applique aux matériaux conçus pour se dégrader dans le corps, ainsi qu'aux matériaux qui ne sont pas conçus pour se dégrader.
Le présent document n'est pas applicable à l'évaluation de la dégradation qui se produit uniquement suite à des processus mécaniques; les méthodologies de génération de ce type de produit de dégradation sont décrites, le cas échéant, dans les normes des produits considérés.
NOTE       La dégradation purement mécanique donne principalement de la matière sous forme de particules. Bien qu'ils n'entrent pas dans le domaine d'application du présent document, de tels produits de dégradation peuvent entraîner une réponse biologique et peuvent subir une évaluation biologique telle que celle décrite dans les autres parties de l'ISO 10993.
En raison de la grande variété des matériaux métalliques utilisés dans la fabrication des dispositifs médicaux, la présente partie de l'ISO 10993 n'indique aucune technique d'analyse spécifique pour la quantification des produits de dégradation. Le présent document ne traite pas de l'identification d'éléments à l'état de trace (< 10−6 w/w) contenus dans le métal ou l'alliage analysé. Il ne fournit aucune exigence spécifique relative aux niveaux admissibles de produits de dégradation.
Le présent document ne couvre pas l'activité biologique des produits de dégradation (pour cela, voir les articles correspondants de l'ISO 10993-1 et de l'ISO 10993-17).

Biološko ovrednotenje medicinskih pripomočkov - 15. del: Identifikacija in ugotavljanje količine razgradnih produktov iz kovin in zlitin (ISO/DIS 10993-15:2018)

General Information

Status
Not Published
Publication Date
15-Sep-2019
Withdrawal Date
15-Mar-2020
Current Stage
6055 - CEN Ratification completed (DOR) - Publishing
Start Date
19-Apr-2023
Completion Date
19-Apr-2023

Relations

Buy Standard

Draft
prEN ISO 10993-15:2018
English language
23 pages
sale 10% off
Preview
sale 10% off
Preview
e-Library read for
1 day

Standards Content (Sample)

SLOVENSKI STANDARD
oSIST prEN ISO 10993-15:2018
01-junij-2018
%LRORãNRRYUHGQRWHQMHPHGLFLQVNLKSULSRPRþNRYGHO,GHQWLILNDFLMDLQ
XJRWDYOMDQMHNROLþLQHUD]JUDGQLKSURGXNWRYL]NRYLQLQ]OLWLQ ,62',6


Biological evaluation of medical devices - Part 15: Identification and quantification of

degradation products from metals and alloys (ISO/DIS 10993-15:2018)

Biologische Beurteilung von Medizinprodukten - Teil 15: Qualitativer und quantitativer

Nachweis von Abbauprodukten aus Metallen und Legierungen (ISO/DIS 10993-15:2018)

Évaluation biologique des dispositifs médicaux - Partie 15: Identification et quantification

des produits de dégradation issus des métaux et alliages (ISO/DIS 10993-15:2018)
Ta slovenski standard je istoveten z: prEN ISO 10993-15
ICS:
11.100.20 %LRORãNRRYUHGQRWHQMH Biological evaluation of
PHGLFLQVNLKSULSRPRþNRY medical devices
oSIST prEN ISO 10993-15:2018 en,fr,de

2003-01.Slovenski inštitut za standardizacijo. Razmnoževanje celote ali delov tega standarda ni dovoljeno.

---------------------- Page: 1 ----------------------
oSIST prEN ISO 10993-15:2018
---------------------- Page: 2 ----------------------
oSIST prEN ISO 10993-15:2018
DRAFT INTERNATIONAL STANDARD
ISO/DIS 10993-15
ISO/TC 194 Secretariat: DIN
Voting begins on: Voting terminates on:
2018-04-24 2018-07-17
Biological evaluation of medical devices —
Part 15:
Identification and quantification of degradation products
from metals and alloys
Évaluation biologique des dispositifs médicaux —

Partie 15: Identification et quantification des produits de dégradation issus des métaux et alliages

ICS: 11.100.20
THIS DOCUMENT IS A DRAFT CIRCULATED
This document is circulated as received from the committee secretariat.
FOR COMMENT AND APPROVAL. IT IS
THEREFORE SUBJECT TO CHANGE AND MAY
NOT BE REFERRED TO AS AN INTERNATIONAL
STANDARD UNTIL PUBLISHED AS SUCH.
IN ADDITION TO THEIR EVALUATION AS
ISO/CEN PARALLEL PROCESSING
BEING ACCEPTABLE FOR INDUSTRIAL,
TECHNOLOGICAL, COMMERCIAL AND
USER PURPOSES, DRAFT INTERNATIONAL
STANDARDS MAY ON OCCASION HAVE TO
BE CONSIDERED IN THE LIGHT OF THEIR
POTENTIAL TO BECOME STANDARDS TO
WHICH REFERENCE MAY BE MADE IN
Reference number
NATIONAL REGULATIONS.
ISO/DIS 10993-15:2018(E)
RECIPIENTS OF THIS DRAFT ARE INVITED
TO SUBMIT, WITH THEIR COMMENTS,
NOTIFICATION OF ANY RELEVANT PATENT
RIGHTS OF WHICH THEY ARE AWARE AND TO
PROVIDE SUPPORTING DOCUMENTATION. ISO 2018
---------------------- Page: 3 ----------------------
oSIST prEN ISO 10993-15:2018
ISO/DIS 10993-15:2018(E)
COPYRIGHT PROTECTED DOCUMENT
© ISO 2018

All rights reserved. Unless otherwise specified, or required in the context of its implementation, no part of this publication may

be reproduced or utilized otherwise in any form or by any means, electronic or mechanical, including photocopying, or posting

on the internet or an intranet, without prior written permission. Permission can be requested from either ISO at the address

below or ISO’s member body in the country of the requester.
ISO copyright office
CP 401 • Ch. de Blandonnet 8
CH-1214 Vernier, Geneva
Phone: +41 22 749 01 11
Fax: +41 22 749 09 47
Email: copyright@iso.org
Website: www.iso.org
Published in Switzerland
ii © ISO 2018 – All rights reserved
---------------------- Page: 4 ----------------------
oSIST prEN ISO 10993-15:2018
ISO/DIS 10993-15:2018(E)
Contents Page

Foreword ........................................................................................................................................................................................................................................iv

Introduction ..................................................................................................................................................................................................................................v

1 Scope ................................................................................................................................................................................................................................. 1

2 Normative references ...................................................................................................................................................................................... 1

3 Terms and definitions ..................................................................................................................................................................................... 2

4 Degradation test methods .......................................................................................................................................................................... 2

4.1 General ........................................................................................................................................................................................................... 2

4.2 Prerequisites ............................................................................................................................................................................................. 3

5 Reagent and sample preparation ........................................................................................................................................................ 3

5.1 Sample documentation .................................................................................................................................................................... 3

5.2 Test solution (electrolyte) ............................................................................................................................................................. 3

5.3 Preparation of test samples ......................................................................................................................................................... 4

5.3.1 Test samples......................................................................................................................................................................... 4

5.3.2 Sampling.................................................................................................................................................................................. 4

5.3.3 Sample shape ...................................................................................................................................................................... 4

5.3.4 Sample surface condition ......................................................................................................................................... 4

6 Electrochemical tests ....................................................................................................................................................................................... 4

6.1 Apparatus .................................................................................................................................................................................................... 4

6.2 Sample preparation ............................................................................................................................................................................ 5

6.3 Test conditions ........................................................................................................................................................................................ 5

6.4 Potentiodynamic measurements ......... .................................................................................................................................... 5

6.5 Potentiostatic measurements .................................................................................................................................................... 7

7 Immersion test ....................................................................................................................................................................................................... 7

7.1 Apparatus .................................................................................................................................................................................................... 7

7.2 Sample preparation ............................................................................................................................................................................ 8

7.3 Immersion test procedure ............................................................................................................................................................ 8

8 Analysis .......................................................................................................................................................................................................................... 9

9 Test report ................................................................................................................................................................................................................... 9

Annex A (normative) Electrolytes for the electrochemical tests ..........................................................................................10

Annex B (informative) Schematic diagram of the electrochemical measuring circuit ..................................11

Annex C (informative) Schematic drawing of an electrolytic cell ........................................................................................12

Annex ZA (informative) Relationship between this European Standard and the essential

requirements of Directive 93/42/EEC [OJ L 169] aimed to be covered ....................................................14

Annex ZB (informative) Relationship between this European Standard and the essential

requirements of Directive 90/385/EEC [OJ L 189] aimed to be covered ................................................15

Annex ZC (informative) Relationship between this European Standard and the general

health and safety requirements of Regulation (EU) 2017/745 on medical devices

aimed to be covered .......................................................................................................................................................................................17

Bibliography .............................................................................................................................................................................................................................18

© ISO 2018 – All rights reserved iii
---------------------- Page: 5 ----------------------
oSIST prEN ISO 10993-15:2018
ISO/DIS 10993-15:2018(E)
Foreword

ISO (the International Organization for Standardization) is a worldwide federation of national standards

bodies (ISO member bodies). The work of preparing International Standards is normally carried out

through ISO technical committees. Each member body interested in a subject for which a technical

committee has been established has the right to be represented on that committee. International

organizations, governmental and non-governmental, in liaison with ISO, also take part in the work.

ISO collaborates closely with the International Electrotechnical Commission (IEC) on all matters of

electrotechnical standardization.

The procedures used to develop this document and those intended for its further maintenance are

described in the ISO/IEC Directives, Part 1. In particular the different approval criteria needed for the

different types of ISO documents should be noted. This document was drafted in accordance with the

editorial rules of the ISO/IEC Directives, Part 2 (see www .iso .org/ directives).

Attention is drawn to the possibility that some of the elements of this document may be the subject of

patent rights. ISO shall not be held responsible for identifying any or all such patent rights. Details of

any patent rights identified during the development of the document will be in the Introduction and/or

on the ISO list of patent declarations received (see www .iso .org/ patents).

Any trade name used in this document is information given for the convenience of users and does not

constitute an endorsement.

For an explanation on the meaning of ISO specific terms and expressions related to conformity assessment,

as well as information about ISO's adherence to the World Trade Organization (WTO) principles in the

Technical Barriers to Trade (TBT) see the following URL: www .iso .org/ iso/ foreword .html.

This document was prepared by Technical Committee ISO/TC 194, Biological and clinical evaluation of

medical devices.

This second edition cancels and replaces the first edition (ISO 10993-15:2000), which has been

technically revised.
The main changes compared to the previous edition are as follows:

a) document now considers materials designed to degrade in the body as well as materials that are

not intended to degrade;

b) information on test methods amended to consider nanomaterials and relevant material specific

standards;
c) test solution (electrolyte) more specified;
d) sample shape more specified;
e) immersion test procedure expanded;
f) Annex C changed to normative and now Annex A.
A list of all parts in the ISO 10993- series can be found on the ISO website.
iv © ISO 2018 – All rights reserved
---------------------- Page: 6 ----------------------
oSIST prEN ISO 10993-15:2018
ISO/DIS 10993-15:2018(E)
Introduction

One of the potential health hazards resulting from medical devices may be due to the interactions

of their electrochemically induced degradation products with the biological system. Therefore, the

evaluation of potential degradation products from metallic materials by methods suitable for testing

the electrochemical behavior of these materials is a necessary step in the biological performance

testing of materials.

The body environment typically contains cations of sodium, potassium, calcium, and magnesium, and

anions of chloride, bicarbonate, phosphate, and organic acids generally in concentrations between 2 x 10

–3 mol and 150 x 10 –3 mol. A range of organic molecules such as proteins, enzymes, and lipoproteins is

also present, but their concentrations may vary to a great extent. Earlier studies assumed that organic

molecules did not exert a significant influence on the degradation of metallic implants, but newer

investigations indicate that implant–protein interactions should be taken into account. Depending on

a particular product or application, altering the pH of the testing environment may also need to be

considered.

In such biological environments, metallic materials may undergo a certain degradation, and the

different degradation products may interact with the biological system in different ways. Therefore,

the identification and quantification of these degradation products is an important step in evaluating

the biological performance of medical devices.
© ISO 2018 – All rights reserved v
---------------------- Page: 7 ----------------------
oSIST prEN ISO 10993-15:2018
---------------------- Page: 8 ----------------------
oSIST prEN ISO 10993-15:2018
DRAFT INTERNATIONAL STANDARD ISO/DIS 10993-15:2018(E)
Biological evaluation of medical devices —
Part 15:
Identification and quantification of degradation products
from metals and alloys
1 Scope

This document provides guidance on general requirements for the design of tests for identifying and

quantifying degradation products from final metallic medical devices or corresponding material

samples finished as ready for clinical use.

This document is applicable only to those degradation products generated by chemical alteration of the

final metallic device in an in vitro accelerated degradation test. Because of the accelerated nature of

these tests, the test results may not reflect the implant or material behavior in the body. The described

chemical methodologies are a means to generate degradation products for further assessments.

This document considers both materials designed to degrade in the body as well as materials that are

not intended to degrade.

This document is not applicable to degradation products induced by applied mechanical stress.

Mechanically induced degradation, such as wear, can be covered in the appropriate product-specific

standard. Where product-group standards provide applicable product-specific methodologies for the

identification and quantification of degradation products, those standards should be considered.

Because of the wide range of metallic materials used in medical devices, no specific analytical

techniques are identified for quantifying the degradation products. The identification of trace elements

(< 10 w/w) contained in the specific metal or alloy is not addressed in this part of ISO 10993, nor are

specific requirements for acceptable levels of degradation products provided in this part of ISO 10993.

This document does not address the biological activity of the degradation products; see instead the

applicable clauses of ISO 10993-1 and ISO 10993-17.
2 Normative references

The following documents are referred to in the text in such a way that some or all of their content

constitutes requirements of this document. For dated references, only the edition cited applies. For

undated references, the latest edition of the referenced document (including any amendments) applies.

ISO 3585, Borosilicate glass 3.3 — Properties
ISO 3696, Water for analytical laboratory use — Specification and test methods
ISO 8044, Corrosion of metals and alloys — Basic terms and definitions

ISO 10993-1, Biological evaluation of medical devices — Part 1: Evaluation and testing within a risk

management process

ISO 10993-9, Biological evaluation of medical devices — Part 9: Framework for identification and

quantification of potential degradation products

ISO 10993-12, Biological evaluation of medical devices — Part 12: Sample preparation and reference

materials
© ISO 2018 – All rights reserved 1
---------------------- Page: 9 ----------------------
oSIST prEN ISO 10993-15:2018
ISO/DIS 10993-15:2018(E)

ISO 10993-13, Biological evaluation of medical devices — Part 13: Identification and quantification of

degradation products from polymeric medical devices

ISO 10993-14, Biological evaluation of medical devices — Part 14: Identification and quantification of

degradation products from ceramics

ISO 10993-16, Biological evaluation of medical devices — Part 16: Toxicokinetic study design for

degradation products and leachables
3 Terms and definitions

For the purposes of this document, the terms and definitions given in ISO 8044, ISO 10993-1,

ISO 10993-9, ISO 10993-12 and the following apply.

ISO and IEC maintain terminological databases for use in standardization at the following addresses:

IEC Electropedia: available at http:// www .electropedia .org/
ISO Online browsing platform: available at https:// www .iso .org/ obp
3.1
alloy

material composed of a metallic element with one or more addition(s) of other metallic and/or non-

metallic elements
3.2
electrolyte
solution containing ions with the capacity to conduct electric current
3.3
open-circuit potential

potential of an electrode measured with respect to a reference electrode or another electrode when no

current flows to or from it
3.4
passive limit potential
electrode potential of the positive limit of the passive range
Note 1 to entry: See Figure 1.
3.5
breakdown potential

critical electrode potential above which localized or transpassive corrosion is found to occur

Note 1 to entry: See Figure 1.
4 Degradation test methods
4.1 General

To identify and quantify degradation products from metals and alloys in medical devices, a combination

of two procedures is described. The choice of test procedure shall be justified according to the function

of the medical device.

The first procedure described is a combination of a potentiodynamic test and a potentiostatic test. The

second procedure described is an immersion test.
2 © ISO 2018 – All rights reserved
---------------------- Page: 10 ----------------------
oSIST prEN ISO 10993-15:2018
ISO/DIS 10993-15:2018(E)

The potentiodynamic test is used to determine the general electrochemical behavior of the material

under consideration and to determine certain specific points (E and E ) on the potential/current

a p
density curve.

The immersion test is used to chemically degrade the test material to generate degradation products to

be analyzed.

If there is the possibility of the loss of a coating from a metallic substrate due to degradation, the

potential degradation products from the substrate material shall be considered, as well as the

coating itself. In addition, if a metallic substrate coated with a non-metallic material is to be tested,

the requirements of ISO 10993-13 and/or ISO 10993-14 shall be considered in order to determine the

potential degradation products of the coating.

The identified and quantified degradation products form the basis for evaluation of biological response

and, if appropriate, toxicokinetic studies in accordance with ISO 10993-16.

For those medical devices composed of or containing nanoscale materials, and for those instances where

metallic degradation products are within the nanoscale size range (approximately 1 nm to 100 nm), the

user is referred to ISO 10993-22 when creating their risk assessment documents.

If the medical device is made using a metal or metal alloy designed to be absorbed by the body, the

user is directed to relevant material specific standards (see bibliography) for methods and specific

considerations (e.g. electrolyte, atmosphere, etc.) appropriate for this class of materials.

4.2 Prerequisites

The rates of electrochemical degradation reactions are sensitive to small variations in test conditions,

instrumentation, sample conditions, and preparation. Therefore, electrochemical degradation testing

shall be carried out in an appropriately equipped laboratory by experienced and qualified personnel.

This includes proper maintenance and calibration of the test equipment. The methods and operating

conditions of the equipment shall also be validated.

Fulfillment of electrochemical test conditions for stability, warm-up time, etc., can be demonstrated by

[1]
conformance to.
5 Reagent and sample preparation
5.1 Sample documentation
The general composition of the material(s) under test shall be documented.
5.2 Test solution (electrolyte)

The test solution (electrolyte) to be used shall be appropriate for the intended use of the medical device.

All chemicals shall be of analytical grade and dissolved in water of grade 2 in accordance with ISO 3696.

The first choice for the electrolyte shall be an aqueous solution of 0,9 % sodium chloride.

Dependent on the composition and corrosion mechanism of the metal or alloy being tested, other

electrolytes may be used, such as artificial saliva or artificial plasma. Examples of electrolyte

compositions are given in Annex A, but other more material and physiologically relevant electrolyte

solutions and test conditions may be utilized. Possible effect of implant–protein interactions should be

taken into account.

NOTE Formulations for artificial sweat, gastrointestinal fluids, and lung fluids have been used [see

Bibliography].

In the test report, the choice of electrolyte shall be justified. If other than an aqueous solution of 0,9 %

sodium chloride is used, the pH of the electrolyte shall be specified.
© ISO 2018 – All rights reserved 3
---------------------- Page: 11 ----------------------
oSIST prEN ISO 10993-15:2018
ISO/DIS 10993-15:2018(E)
5.3 Preparation of test samples
5.3.1 Test samples

The sensitivity of chemical degradation testing is related to variation in material composition, to

material processing, and to surface-finishing procedures. The sampling procedure, sample shape,

and surface preparation are critical. In addition, confined spaces can result in crevice corrosion and

defects in coatings can cause pit corrosion, which shall be taken into consideration. The samples shall

be representative of the final devices.
5.3.2 Sampling

For each chemical test, at least two test samples shall be prepared as specified in ISO 10993-12. If

substantial deviations in the test results are found, the reasons for the deviation shall be determined,

and more samples shall be tested.

If the metallic sample has anisotropic properties due to manufacturing conditions, tests involving

single-surface exposure should include samples cut parallel to both the transverse and longitudinal

manufacturing directions.
5.3.3 Sample shape

Standard samples, either circular- or rectangular-section bars or flat coupons, or one single free

surface, may be used for degradation testing if they are prepared in a manner comparable to the final

medical device. Samples of actual device components may be of any shape and condition; however, the

testing shall be carried out under well-controlled conditions which shall be reported.

The surface area of the sample exposed to the electrolyte shall be determined to +/- 10% of the total

geometrical area to assure an accurate and repeatable determination of the degradation rates.

If representative samples are used, consideration shall be made regarding whether the differences

between the representative sample and the final medical device or component could affect the results

of the test. Testing of representative samples instead of the final medical device shall be supported by a

description of any differences between the representative sample and the final device. The report shall

contain a detailed rationale for why each difference is not expected to alter the biocompatibility of the

final device.
5.3.4 Sample surface condition

Since the surface condition of a material may affect its electrochemical behavior, the surface condition of

the test sample shall be identical to the final medical device and shall be described in the test report. For

comparing test results of different materials, the surface condition of the test samples shall be the same.

6 Electrochemical tests
6.1 Apparatus

Test cells of borosilicate glass, in appropriate sizes, in accordance with ISO 3585, with a means of

controlling the bath temperature within ± 1 °C.
-9 –1

Scanning potentiostat with a potential range ± 2 V and a current output range from 10 A to 10 A.

Potential-measuring instrument with a high input impedance (>10 Ω) and a sensitivity and

accuracy to detect a change of 1 mV over a potential range between ± 2 V.

Current-measuring instrument capable of measuring a current to ± 1 % of the absolute value over a

-9 –1
current range between 10 A and 10 A.
4 © ISO 2018 – All rights reserved
---------------------- Page: 12 ----------------------
oSIST prEN ISO 10993-15:2018
ISO/DIS 10993-15:2018(E)
Working electrode (test sample).

Counter-electrode(s) such as platinum (grid, plate, or wire) or vitreous carbon with an area at least 10

times that of the working electrode.
Reference electrode.
pH-meter with a sensitivity of ± 0,1.

A schematic diagram of the electrochemical measurement circuit which may be used as a system with

variable potential is given in Annex B.
A schematic drawing of an electrolytic cell is given in Annex C.
6.2 Sample preparation

Mount the test sample in a watertight electrode holder so that only the test surface is in contact with

the electrolyte. Take care to avoid the creation of conditions where crevice corrosion can occur due to

the formation of a crevice between the mounting and the sample. Before testing, clean the specimen

ultrasonically for 10 min to 15 min in ethanol, carefully rinse with water of grade 2 in accordance with

ISO 3696, and immediately transfer into the test cell.
6.3 Test conditions

Fill the test cell with the test solution (electrolyte). If the electrochemical behavior is temperature

sensitive in the range of 10 °C to 50 °C, maintain the electrolyte cell at (37 ± 1) °C. Reduce the oxygen level

3 –1

in the electrolyte by bubbling oxygen-free nitrogen or argon at a rate of approximately 100 cm ▪min

for not less than 30 min prior to the start of the test. The electrolyte shall be agitated either by the

bubbling gas or mechanical means to avoid concentration gradients. If gas agitation is used, take care

not to have any gas bubbles adhering to the active test surface.

Magnetic stirrers often interfere with electrochemical test cells. If they are used, their effect on the test

cell shall be determined as part of the validation of test equipment (see 4.2).
6.4 Potentiodynamic measurements

Measure the open-circuit potential not less than 2 h after the immersion of the working electrode. This

potential shall be the starting potential for potentiodynamic measurements. The sweep rate shall be

1,0 mV▪s , except in tests where the sweep rate has little effect, where the test may be accelerated by

increasing the sweep rate to 10 mV▪s . Record the potential/current density curve up to a maximum

of 2 000 mV or a maximum current density of 1,0 mA▪cm , whichever comes first, to evaluate the

transpassive range of the sample (see Figure 1). To ensure consistency, reverse the scan and continue

back at least to the open-circuit potential. Then repeat the test back to 2 000 mV or 1,0 mA▪cm . If

the curves are not reproducible, then continue cycling 5 to 10 times. If consistent potential/current

density curves are not achieved after 5 to 10 cycles, investigate possible causes such as test set- up,

electrode function, innate material properties, etc. The log current density/potential curves should also

be recorded (see Figure 2). Record the breakdown potential (E ) from the last cycle taken (see Figure 1).

Noble metals may b
...

Questions, Comments and Discussion

Ask us and Technical Secretary will try to provide an answer. You can facilitate discussion about the standard in here.