EN 14242:2023
(Main)Aluminium and aluminium alloys - Chemical analysis - Inductively coupled plasma optical emission spectrometric analysis
Aluminium and aluminium alloys - Chemical analysis - Inductively coupled plasma optical emission spectrometric analysis
This document specifies an inductively coupled plasma optical emission spectrometric method
(ICP-OES) for the analysis of aluminium and aluminium alloys.
This method is applicable to the determination of silicon, iron, copper, manganese, magnesium, chromium, nickel, zinc, titanium, gallium, vanadium, beryllium, bismuth, calcium, cadmium, cobalt, lithium, sodium, lead, antimony, tin, strontium and zirconium in aluminium and aluminium alloys.
The content of the elements to be determined should be at least 10 times higher than the corresponding detection limits.
Aluminium und Aluminiumlegierungen - Chemische Analyse - Optische Emissionsspektrometrie mit induktiv gekoppeltem Plasma
Dieses Dokument legt ein Verfahren für die optische Emissionsspektrometrie mit induktiv gekoppeltem Plasma (ICP OES, en: inductively coupled plasma optical emission spectrometric method) zur Analyse von Aluminium und Aluminiumlegierungen fest.
Dieses Verfahren ist für die Bestimmung von Silicium, Eisen, Kupfer, Mangan, Magnesium, Chrom, Nickel, Zink, Titan, Gallium, Vanadium, Beryllium, Bismut, Calcium, Cadmium, Cobalt, Lithium, Natrium, Blei, Antimon, Zinn, Strontium und Zirconium in Aluminium und Aluminiumlegierungen anwendbar.
Der Gehalt der zu bestimmenden Elemente sollte mindestens 10 mal höher sein als die entsprechende Nachweisgrenze.
Aluminium et alliages d’aluminium - Analyse chimique - Analyse par spectrométrie d’émission optique avec source à plasma induit par haute fréquence
Le présent document spécifie une méthode par spectrométrie d’émission optique avec source à plasma induit par haute fréquence (SEP) pour l’analyse de l’aluminium et des alliages d’aluminium.
Cette méthode est applicable à la détermination des teneurs en silicium, fer, cuivre, manganèse, magnésium, chrome, nickel, zinc, titane, gallium, vanadium, béryllium, bismuth, calcium, cadmium, cobalt, lithium, sodium, plomb, antimoine, étain, strontium et zirconium dans l’aluminium et les alliages d’aluminium.
Il convient que la teneur des éléments à déterminer soit au moins 10 fois supérieure aux limites de détection correspondantes.
Aluminij in aluminijeve zlitine - Kemična analiza - Analiza s spektrometrijo optične emisije z induktivno sklopljeno plazmo
Ta evropski standard podrobno opisuje korake za raztapljanje in pripravo kalibracijskih raztopin. Prednostno se uporablja za certificiranje in analizo referenc. Vsi instrumenti, vključno s programsko opremo za preskuševalne laboratorije, se razlikujejo in so predmet sprememb. Za to so določena splošna merila za umerjanje in merjenje.
To metodo je treba uporabljati s prednostnimi referenčnimi materiali, katerih snovi ima bistveno manjšo stopnjo negotovosti, kot je potrebna za ponovljivost preskusnega postopka.
General Information
Relations
Standards Content (Sample)
SLOVENSKI STANDARD
01-maj-2023
Aluminij in aluminijeve zlitine - Kemična analiza - Analiza s spektrometrijo optične
emisije z induktivno sklopljeno plazmo
Aluminium and aluminium alloys - Chemical analysis - Inductively coupled plasma optical
emission spectrometric analysis
Aluminium und Aluminiumlegierungen - Chemische Analyse - Optische
Emissionsspektrometrie mit induktiv gekoppeltem Plasma
Aluminium et alliages d’aluminium - Analyse chimique - Analyse par spectrométrie
d’émission optique avec source à plasma induit par haute fréquence
Ta slovenski standard je istoveten z: EN 14242:2023
ICS:
77.040.30 Kemijska analiza kovin Chemical analysis of metals
77.120.10 Aluminij in aluminijeve zlitine Aluminium and aluminium
alloys
2003-01.Slovenski inštitut za standardizacijo. Razmnoževanje celote ali delov tega standarda ni dovoljeno.
EN 14242
EUROPEAN STANDARD
NORME EUROPÉENNE
March 2023
EUROPÄISCHE NORM
ICS 77.040.30; 77.120.10 Supersedes EN 14242:2004
English Version
Aluminium and aluminium alloys - Chemical analysis -
Inductively coupled plasma optical emission spectrometric
analysis
Aluminium et alliages d'aluminium - Analyse chimique Aluminium und Aluminiumlegierungen - Chemische
- Analyse par spectrométrie d'émission optique avec Analyse - Optische Emissionsspektrometrie mit
source à plasma induit par haute fréquence induktiv gekoppeltem Plasma
This European Standard was approved by CEN on 23 January 2023.
CEN members are bound to comply with the CEN/CENELEC Internal Regulations which stipulate the conditions for giving this
European Standard the status of a national standard without any alteration. Up-to-date lists and bibliographical references
concerning such national standards may be obtained on application to the CEN-CENELEC Management Centre or to any CEN
member.
This European Standard exists in three official versions (English, French, German). A version in any other language made by
translation under the responsibility of a CEN member into its own language and notified to the CEN-CENELEC Management
Centre has the same status as the official versions.
CEN members are the national standards bodies of Austria, Belgium, Bulgaria, Croatia, Cyprus, Czech Republic, Denmark, Estonia,
Finland, France, Germany, Greece, Hungary, Iceland, Ireland, Italy, Latvia, Lithuania, Luxembourg, Malta, Netherlands, Norway,
Poland, Portugal, Republic of North Macedonia, Romania, Serbia, Slovakia, Slovenia, Spain, Sweden, Switzerland, Türkiye and
United Kingdom.
EUROPEAN COMMITTEE FOR STANDARDIZATION
COMITÉ EUROPÉEN DE NORMALISATION
EUROPÄISCHES KOMITEE FÜR NORMUNG
CEN-CENELEC Management Centre: Rue de la Science 23, B-1040 Brussels
© 2023 CEN All rights of exploitation in any form and by any means reserved Ref. No. EN 14242:2023 E
worldwide for CEN national Members.
Contents Page
European foreword . 3
1 Scope . 4
2 Normative references . 4
3 Terms and definitions . 4
4 Principle . 4
5 Reagents . 5
6 Apparatus .10
7 Sampling .11
7.1 General .11
7.2 Test sample .11
8 Procedure.11
8.1 Test portion .11
8.2 Dissolution procedure I with sodium hydroxide solution .11
8.3 Dissolution procedure II with nitric acid and hydrofluoric acid .12
8.4 Dissolution procedure III with a mixture of hydrochloric acid and nitric acid .12
8.5 Dissolution procedure IV with hydrochloric acid .13
8.6 Calibration solutions and drift correction solution .13
8.6.1 General .13
8.6.2 Preparation of the calibration solutions .14
8.7 Measurements .14
8.7.1 Adjustment of the apparatus .14
8.7.2 Measurement of the calibration solutions .14
8.7.3 Measurement of the test solutions .15
8.8 Calibration curves .15
9 Correction of short-term fluctuations and drift .15
9.1 General .15
9.2 Short-term fluctuations .15
9.3 Drift .15
10 Investigation of interferences .15
11 Expression of the results .16
11.1 Correction .16
11.2 Result .16
12 Test report .16
Annex A (informative) Analytical wavelengths .17
Annex B (informative) Plasma optical emission spectrometer — Suggested performance
criteria to be checked .19
Bibliography .21
European foreword
This document (EN 14242:2023) has been prepared by Technical Committee CEN/TC 132 “Aluminium
and aluminium alloys”, the secretariat of which is held by AFNOR.
This European Standard shall be given the status of a national standard, either by publication of an
identical text or by endorsement, at the latest by September 2023, and conflicting national standards shall
be withdrawn at the latest by September 2023.
Attention is drawn to the possibility that some of the elements of this document may be the subject of
patent rights. CEN shall not be held responsible for identifying any or all such patent rights.
This document supersedes EN 14242:2004.
The main changes compared to the previous edition are listed below:
— modification of the title and Scope;
— new subclause 5.15.6;
— several editorial modifications.
Any feedback and questions on this document should be directed to the users’ national standards body.
A complete listing of these bodies can be found on the CEN website.
According to the CEN-CENELEC Internal Regulations, the national standards organisations of the
following countries are bound to implement this European Standard: Austria, Belgium, Bulgaria, Croatia,
Cyprus, Czech Republic, Denmark, Estonia, Finland, France, Germany, Greece, Hungary, Iceland, Ireland,
Italy, Latvia, Lithuania, Luxembourg, Malta, Netherlands, Norway, Poland, Portugal, Republic of North
Macedonia, Romania, Serbia, Slovakia, Slovenia, Spain, Sweden, Switzerland, Türkiye and the United
Kingdom.
1 Scope
This document specifies an inductively coupled plasma optical emission spectrometric method
(ICP-OES) for the analysis of aluminium and aluminium alloys.
This method is applicable to the determination of silicon, iron, copper, manganese, magnesium,
chromium, nickel, zinc, titanium, gallium, vanadium, beryllium, bismuth, calcium, cadmium, cobalt,
lithium, sodium, lead, antimony, tin, strontium and zirconium in aluminium and aluminium alloys.
The content of the elements to be determined should be at least 10 times higher than the corresponding
detection limits.
2 Normative references
The following documents are referred to in the text in such a way that some or all of their content
constitutes requirements of this document. For dated references, only the edition cited applies. For
undated references, the latest edition of the referenced document (including any amendments) applies.
EN 12258-2:2004, Aluminium and aluminium alloys - Terms and definitions - Part 2: Chemical analysis
EN 14361, Aluminium and aluminium alloys - Chemical analysis - Sampling from metal melts
EN ISO 648, Laboratory glassware - Single-volume pipettes (ISO 648)
EN ISO 1042, Laboratory glassware - One-mark volumetric flasks (ISO 1042)
EN ISO 3696, Water for analytical laboratory use - Specification and test methods (ISO 3696)
3 Terms and definitions
For the purposes of this document, the terms and definitions given in EN 12258-2:2004 apply.
ISO and IEC maintain terminology databases for use in standardization at the following addresses:
— IEC Electropedia: available at https://www.electropedia.org/
— ISO Online browsing platform: available at https://www.iso.org/obp
4 Principle
A test portion is dissolved with:
— a sodium hydroxide solution followed by acidification with a mixture of nitric acid and hydrochloric
acid; or
— nitric acid and hydrofluoric acid; or
— a mixture of hydrochloric acid and nitric acid; or
— hydrochloric acid and hydrogen peroxide
according to the alloy type and the element to be determined.
After suitable dilution and, if necessary, addition of an internal reference element, the solution is
nebulized into an inductively coupled plasma optical emission spectrometer and the intensity of the
emitted light (including, where appropriate, that of the internal reference element) is measured. The
emission signals on the selected analytical lines (see Annex A) are then compared with those of the
calibration solutions.
NOTE 1 The ranges of application and the accuracy of the method or any alternative steps are validated by the
laboratory. Approximate ranges of application are given in Annex A.
NOTE 2 All instrumentation, including software used in the laboratories, are different and subject to change.
Therefore, only general criteria for calibration and measurement are specified.
5 Reagents
During the analysis, unless otherwise specified, use only reagents of recognized analytical grade and only
grade 2 water as specified in EN ISO 3696 or equivalent quality.
The same reagents should be used for preparation of calibration solutions and of sample solutions.
5.1 Aluminium, purity ≥ 99,999 % by mass.
5.2 Sodium carbonate (Na CO )
2 3
5.3 Potassium carbonate (K CO )
2 3
5.4 Sodium nitrite (NaNO )
5.5 Potassium disulphate (K S O )
2 2 7
5.6 Nitric acid, ρ = 1,40 g/ml approximately.
5.7 Nitric acid solution, 1 + 1
Carefully add 500 ml of nitric acid (5.6) to 400 ml water, allow to cool, dilute to 1 l with water and mix.
5.8 Nitric acid solution, 4 mol/l
Carefully add 27,7 ml of nitric acid (5.6) to 50 ml water, allow to cool, dilute to 100 ml with water and
mix.
5.9 Hydrochloric acid, ρ = 1,19 g/ml, approximately
5.10 Hydrochloric acid solution, 1 + 1
Carefully add 500 ml of hydrochloric acid (5.9) to 400 ml water, allow to cool, dilute to 1 l with water and
mix.
5.11 Hydrofluoric acid, ρ = 1,14 g/ml, approximately.
5.12 Sulphuric acid, ρ = 1,84 g/ml approximately.
5.13 Hydrogen peroxide, 30 % (by mass) solution.
5.14 Sodium hydroxide solution, 400 g/l
Transfer 400,0 g of sodium hydroxide (NaOH) into a plastic beaker with a lid and carefully add 500 ml of
water. Transfer the solution into a 1 000 ml volumetric plastic flask.
Dilute to the mark with water and mix.
5.15 Standard solutions
The standard solutions shall be traceable to international units mass or amount of substances i.e.
kilogram or mol. They shou
...