EN 60601-2-39:1999
(Main)Medical electrical equipment - Part 2-39: Particular requirements for the safety of peritoneal dialysis equipment
Medical electrical equipment - Part 2-39: Particular requirements for the safety of peritoneal dialysis equipment
Specifies the minimum safety requirements for peritoneal dialysis equipment. Applies to equipment intended for use either by medical staff or under supervision of medical experts, including equipment operated by the patient, regardless of whether the equipment is used in a hospital or domestic environment.
Medizinische elektrische Geräte - Teil 2-39: Besondere Festlegungen für die Sicherheit von Peritoneal-Dialyse-Geräten
Appareils électromédicaux - Partie 2-39: Règles particulières de sécurité pour les appareils de dialyse péritonéale
Medical electrical equipment - Part 2-39: Particular requirements for the safety of peritoneal dialysis equipment
General Information
Relations
This May Also Interest You
IEC 60601-2-40:2024 applies to the BASIC SAFETY and ESSENTIAL PERFORMANCE of ELECTROMYOGRAPHS and EVOKED RESPONSE EQUIPMENT, hereafter referred to as ME EQUIPMENT. The following ME EQUIPMENT are excluded: - ME EQUIPMENT intended for therapeutic application; - ME EQUIPMENT intended for transcutaneous electrical nerve stimulators and electrical muscle stimulators (ME EQUIPMENT covered by IEC 60601-2-10). IEC 60601-2-40:2024 cancels and replaces the second edition published in 2016. This edition constitutes a technical revision. This edition includes the following significant technical changes with respect to the previous edition: a) added requirements for constant voltage stimulators; b) clarified requirements for VISUAL STIMULATORS.
- Standard33 pagesEnglish languagesale 10% offe-Library read for×1 day
IEC 60601-2-39:2024 applies to the BASIC SAFETY and ESSENTIAL PERFORMANCE of PERITONEAL DIALYSIS ME EQUIPMENT, hereafter referred to as PD EQUIPMENT. It applies to PD EQUIPMENT intended for use either by medical staff or under the supervision of medical experts, including PD EQUIPMENT operated by the PATIENT, regardless of whether the PD EQUIPMENT is used in a hospital or domestic environment. If a clause or subclause is specifically intended to be applicable to ME EQUIPMENT only, or to ME SYSTEMS only, the title and content of that clause or subclause will say so. If that is not the case, the clause or subclause applies both to ME EQUIPMENT and to ME SYSTEMS, as relevant. This document does not take into consideration specific safety details of the DIALYSING SOLUTION control system of PD EQUIPMENT using regeneration of DIALYSING SOLUTION or CENTRAL DELIVERY SYSTEMS for DIALYSING SOLUTION. It does, however, take into consideration the specific safety requirements of such PD EQUIPMENT concerning electrical safety and PATIENT safety. This document specifies the minimum safety requirements for PD EQUIPMENT. These PD EQUIPMENT are intended for use either by medical staff or for use by the PATIENT or other trained personnel under medical supervision. This document includes all ME EQUIPMENT that is intended to deliver a PERITONEAL DIALYSIS treatment to a PATIENT, independent of the treatment duration and location. These particular requirements do not apply to: – PRE-MANUFACTURED DIALYSING SOLUTION bags, – DIALYSING SOLUTION CIRCUITS, – DIALYSING SOLUTION CONCENTRATE, – DIALYSIS WATER supply systems (see ISO 23500-2) , – CENTRAL DELIVERY SYSTEMS for DIALYSING SOLUTION CONCENTRATES, described as systems for bulk mixing concentrate at a dialysis facility, – equipment used to perform HAEMODIALYSIS (see IEC 60601-2-16). IEC 60601-2-39:2024 cancels and replaces the third edition published in 2018. This edition constitutes a technical revision. This edition includes the following significant technical changes with respect to the previous edition: a) update of references to IEC 60601-1:2005, IEC 60601-1:2005/AMD1:2012 and IEC 60601 1:2005/AMD2:2020, of references to IEC 60601 1 2:2014 and IEC 60601 1 2:2014/AMD1:2020, of references to IEC 60601-1-8:2006, IEC 60601 1 8:2006/AMD1:2012 and IEC 60601 1 8:2006/AMD2:2020, of references to IEC 60601 1 9:2007, IEC 60601 1 9:2007/AMD1:2013 and IEC 60601 1 9:2007/AMD2:2020, of references to IEC 60601 1 10:2007, IEC 60601 1 10:2007/AMD1:2013 and IEC 60601 1 10:2007/AMD2:2020 and of references to IEC 60601 1 11:2015 and IEC 60601 1 11:2015/AMD1:2020; b) consideration of ESSENTIAL PERFORMANCE in SINGLE FAULT CONDITION regarding IEC 60601 1:2005/AMD1:2012/ISH1:2021; c) including the information given in the document 62D/1771A/INF regarding 201.11.8; d) including the information given in the document 62D/1734/INF regarding technical issues of the previous edition; e) including SECURITY (CYBERSECURITY) requirements; f) additions related to online PD SOLUTION generation (ONLINE PD); g) improvements regarding the definition of the APPLIED PART; h) improvement of the essential performance requirements clause/subclauses; i) improvements for labelling; j) other minor technical improvements; k) editorial improvements.
- Standard59 pagesEnglish languagesale 10% offe-Library read for×1 day
- Amendment12 pagesEnglish languagesale 10% offe-Library read for×1 day
IEC 60601-2-41:2021 applies to the BASIC SAFETY and ESSENTIAL PERFORMANCE of SURGICAL LUMINAIRES and LUMINAIRES FOR DIAGNOSIS, hereafter referred to as ME EQUIPMENT. If a clause or subclause is specifically intended to be applicable to ME EQUIPMENT only, or to ME SYSTEMS only, the title and content of that clause or subclause will say so. If that is not the case, the clause or subclause applies both to ME EQUIPMENT and to ME SYSTEMS, as relevant. This particular standard does not apply to – headlights; – endoscopes, laparoscopes and their light sources, which are covered by IEC 60601-2-18; – luminaires used in dentistry, which are covered by ISO 9680; – luminaires for general purposes, which are covered by IEC 60598-2-1 and IEC 60598-2-4; – luminaires dedicated to therapeutic purposes; – special purpose lights with different conditions of use such as light sources intended solely for decontamination of air and surfaces, UV lights for dermatological diagnosis, slit lamps for ophthalmology, lights for surgical microscopes and lights for surgical navigation systems; – lights connected to surgical instruments, such as luminous retractors; – luminaires for emergency lighting, which are covered by IEC 60598-2-22. SURGICAL LUMINAIRES and LUMINAIRES FOR DIAGNOSIS are medical devices and not general lighting equipment. IEC 60601-2-41:2021 cancels and replaces the second edition published in 2009 and Amendment 1:2013. This edition constitutes a technical revision. This edition includes the following significant technical changes with respect to the previous edition: a) revised the statement of essential performance; b) added exposure limits, test conditions, calculation methods and safety warnings related to photobiological hazards; c) removed the terms “MINOR SURGICAL LUMINAIRES” and “MAJOR SURGICAL LUMINAIRES”; d) added definitions of MAXIMUM ILLUMINANCE DISTANCE and REFERENCE DISTANCE and allowed MANUFACTURERS to measure some performance characteristics at the REFERENCE DISTANCE that they specify; e) replaced the region of acceptable chromaticity in (x,y) colour space with a requirement for Du,v; f) added a requirement for acceptable drift of the lighthead when attached to the suspension system; g) added a requirement for fluid ingress protection; h) revised Table 201.101 of IEC 60601-2-41:2009 and IEC 60601-2-41:2009/AMD1:2013 and moved it to Annex BB; i) specified a new device for measuring SHADOW DILUTION in a simulated cavity; j) specified test conditions for luminaires equipped with distance sensors.
- Standard63 pagesEnglish languagesale 10% offe-Library read for×1 day
IEC 80601-2-26:2019 applies to the BASIC SAFETY and ESSENTIAL PERFORMANCE of ELECTROENCEPHALOGRAPHS as defined in 201.3.204, hereafter also referred to as ME EQUIPMENT or ME SYSTEM. This document is applicable to ELECTROENCEPHALOGRAPHS intended for use in professional healthcare facilities, the EMERGENCY MEDICAL SERVICES ENVIRONMENT or the HOME HEALTHCARE ENVIRONMENT. This document does not cover requirements for other equipment used in electroencephalography such as: – phono-photic stimulators; – EEG data storage and retrieval; – ME EQUIPMENT particularly intended for monitoring during electro-convulsive therapy. If a clause or subclause is specifically intended to be applicable to ME EQUIPMENT only, or to ME SYSTEMS only, the title or content of that clause or subclause will say so. If that is not the case, the clause or subclause applies both to ME EQUIPMENT and to ME SYSTEMS, as follows. The clause or subclause applies to ME EQUIPMENT, as default. For ME EQUIPMENT with the corresponding safety measure or function not completely integrated into the ME EQUIPMENT but instead implemented in an ME SYSTEM, the ME EQUIPMENT MANUFACTURER specifies in the ACCOMPANYING DOCUMENTS which functionality and safety requirements are provided by the ME SYSTEM to comply with this document. The ME SYSTEM is verified accordingly. HAZARDS inherent in the intended physiological function of ME EQUIPMENT or ME SYSTEMS within the scope of this document are not covered by specific requirements in this document. IEC 80601-2-26:2019 cancels and replaces the third edition of IEC 60601-2-26 published in 2012. This edition constitutes a technical revision to align with Amendment 1:2012 of IEC 60601-1:2005, new versions of collateral standards and amendments thereto.
- Standard39 pagesEnglish languagesale 10% offe-Library read for×1 day
NEW!IEC 60601-2-39:2018 is available as IEC 60601-2-39:2018 RLV which contains the International Standard and its Redline version, showing all changes of the technical content compared to the previous edition.IEC 60601-2-39:2018 applies to the basic safety and essential performance of peritoneal dialysis medical electrical equipment. It applies to peritoneal dialysis equipment intended for use either by medical staff or under the supervision of medical experts, including peritoneal dialysis equipment operated by the patient, regardless of whether the peritoneal dialysis equipment is used in a hospital or domestic environment. This third edition cancels and replaces the second edition of IEC 60601-2-39. It constitutes a technical revision.This edition includes the following significant technical changes with respect to the previous edition: - update of the references to IEC 60601-1:2005 and IEC 60601-1:2005/AMD1:2012, of references and requirements to IEC 60601-1-2:2014, of references to IEC 60601-1-6:2010 and IEC 60601-1-6:2010/AMD1:2013, of references and requirements to IEC 60601-1-8:2006 and IEC 60601-1-8:2006/AMD1:2012 and of references and requirements to IEC 60601-1-11:2015; - editorial improvements; - improvement of the essential performance requirements clause/subclauses; - new requirements for the interruption of the power supply.
- Standard31 pagesEnglish languagesale 10% offe-Library read for×1 day
Establishes particular requirements for the safety of electromyographs and evoked response equipment as defined in clause 2 of this standard.
- Standard33 pagesEnglish languagesale 10% offe-Library read for×1 day
IEC 60601-2-25:2011 applies to the basic safety and essential performance of electrocardiographs intended by themselves or as a part of a medical electrical system, for the production of electrocardiogram reports for diagnostic purposes. The equipment intended for use under extreme or uncontrolled environmental conditions outside the hospital environment or physician's office, such as in ambulances and air transport, shall comply with this particular standard. Additional standards may apply to Medical electrical equipment for those environments of use. Not included within the scope of this particular standard are: a) the part of Medical electrical equipment that provides vectorcardiographic loops; b) ambulatory electrocardiographic medical electrical equipment covered by IEC 60601-2-47 where not intended for obtaining electrocardiogram reports for diagnostic purposes; c) cardiac monitors covered by IEC 60601-2-27 where not intended for obtaining electrocardiographic reports for diagnostic purposes. This second edition cancels and replaces the first edition of IEC 60601-2-25, published in 1993 and the first edition of IEC 60601-2-51, published in 2003. Updating the particular standards to refer to the third edition of the general standard provided the opportunity to merge the first editions of IEC 60601-2-25 and IEC 60601-2-51 into one standard. Reformatting and technical changes were both made. This second edition of IEC 60601-2-25 constitutes a technical revision of both those standards.
- Standard100 pagesEnglish languagesale 10% offe-Library read for×1 day
- Draft18 pagesEnglish languagesale 10% offe-Library read for×1 day
IEC 60601-2-40:2024 applies to the BASIC SAFETY and ESSENTIAL PERFORMANCE of ELECTROMYOGRAPHS and EVOKED RESPONSE EQUIPMENT, hereafter referred to as ME EQUIPMENT. The following ME EQUIPMENT are excluded: - ME EQUIPMENT intended for therapeutic application; - ME EQUIPMENT intended for transcutaneous electrical nerve stimulators and electrical muscle stimulators (ME EQUIPMENT covered by IEC 60601-2-10). IEC 60601-2-40:2024 cancels and replaces the second edition published in 2016. This edition constitutes a technical revision. This edition includes the following significant technical changes with respect to the previous edition: a) added requirements for constant voltage stimulators; b) clarified requirements for VISUAL STIMULATORS.
- Standard33 pagesEnglish languagesale 10% offe-Library read for×1 day
Questions, Comments and Discussion
Ask us and Technical Secretary will try to provide an answer. You can facilitate discussion about the standard in here.