IEC/SC 62D - IEC_SC_62D
To prepare particular international standards and other publications to support safety and performance for medical equipment, software, and systems. Excluded are medical imaging and related equipment (see scope of SC 62B) and medical equipment using ionizing radiation in therapy (see scope of SC 62C).
IEC_SC_62D
To prepare particular international standards and other publications to support safety and performance for medical equipment, software, and systems. Excluded are medical imaging and related equipment (see scope of SC 62B) and medical equipment using ionizing radiation in therapy (see scope of SC 62C).
General Information
- Amendment12 pagesEnglish languagesale 10% offe-Library read for1 day
IEC 60601-2-46:2023 specifies safety requirements for operating tables, whether or not having electrical parts, including transporters, used for the transportation of the operating table top to or from the base or pedestal of an operating table with detachable operating table top. This particular standard does not apply to - dental patient chairs (see ISO 7494-1), - examination chairs and couches, - patient-supporting systems of diagnostic, interventional and therapeutic equipment (see IEC 60601-2-54 or IEC 60601-2-43), - operating table heating blankets (see IEC 60601-2-35), - patient transfer equipment, - delivery tables and delivery beds, - medical beds (see IEC 60601-2-52 and EN 50637), and - field tables. IEC 60601-2-46:2023 cancels and replaces the third edition published in 2016. This edition constitutes a technical revision. This edition includes the following significant technical change with respect to the previous edition: structural alignment with IEC 60601-1:2005, IEC 60601-1:2005/AMD1:2012 and IEC 60601-1:2005/AMD2:2020.
- Standard32 pagesEnglish languagesale 10% offe-Library read for1 day
- Amendment7 pagesEnglish languagesale 10% offe-Library read for1 day
- Amendment20 pagesEnglish languagesale 10% offe-Library read for1 day
IEC 80601-2-58:2024 is available as IEC 80601-2-58:2024 RLV which contains the International Standard and its Redline version, showing all changes of the technical content compared to the previous edition.IEC 80601-2-58:2024 applies to the BASIC SAFETY and ESSENTIAL PERFORMANCE of LENS REMOVAL DEVICES and VITRECTOMY DEVICES for ophthalmic surgery (as defined in 201.3.209 and 201.3.217) and associated ACCESSORIES that can be connected to this MEDICAL ELECTRICAL EQUIPMENT, hereafter referred to as ME EQUIPMENT. If a clause or subclause is specifically intended to be applicable to ME EQUIPMENT only, or to ME SYSTEMS only, the title and content of that clause or subclause will say so. If that is not the case, the clause or subclause applies both to ME EQUIPMENT and to ME SYSTEMS, as relevant. HAZARDS inherent in the intended physiological function of ME EQUIPMENT or ME SYSTEMS within the scope of this document are not covered by specific requirements in this document except in 7.2.13 of IEC 60601-1:2005 and IEC 60601 1:2005/AMD2:2020 and 8.4.1 of IEC 60601-1:2005. IEC 80601-2-58:2024 cancels and replaces the second edition published in 2014 and its Amendment 1:2016. This edition constitutes a technical revision. This edition includes the following significant technical changes with respect to the previous edition: a) the alignment of this particular standard based on the amendment of IEC 60601-1:2005, IEC 60601-1:2005/AMD1:2012 and IEC 60601-1:2005/AMD2:2020; b) updating collateral, particular and IEC 60601-1:2005, IEC 60601-1:2005/AMD1:2012 and IEC 60601-1:2005/AMD2:2020 references to align with amendments to IEC 60601-1:2005, IEC 60601-1:2005/AMD1:2012 and IEC 60601-1:2005/AMD2:2020 and other collateral standards; c) updated normative references; d) added new requirement for particulate matter from APPLIED PARTS in 201.9.5.101; e) adding the shadow light method in 201.12.1.101.7; f) clarify test conditions for EMC requirements in 202.7.1.2; g) updated Table D.4 references to include specific IEC references to the symbols and delete “Annex AA, 201.7.6.101”; h) include a new annex to address the relevant general safety and performance requirements of European regulation (EU) 2017/745 (Annex BB); i) remove all references of the LIQUEFACTION FRAGMENTATION LENS REMOVAL method.
- Standard43 pagesEnglish languagesale 10% offe-Library read for1 day
- Amendment7 pagesEnglish languagesale 10% offe-Library read for1 day
- Amendment13 pagesEnglish languagesale 10% offe-Library read for1 day
No scope available
- Amendment13 pagesEnglish languagesale 10% offe-Library read for1 day
- Amendment8 pagesEnglish languagesale 10% offe-Library read for1 day
- Amendment12 pagesEnglish languagesale 10% offe-Library read for1 day
- Amendment10 pagesEnglish languagesale 10% offe-Library read for1 day
- Amendment11 pagesEnglish languagesale 10% offe-Library read for1 day
- Amendment12 pagesEnglish languagesale 10% offe-Library read for1 day
- Amendment14 pagesEnglish languagesale 10% offe-Library read for1 day
- Amendment14 pagesEnglish languagesale 10% offe-Library read for1 day
- Amendment9 pagesEnglish languagesale 10% offe-Library read for1 day
- Amendment7 pagesEnglish languagesale 10% offe-Library read for1 day
20211012-JO-Link to 93/42/EEC and M/295 removed
- Amendment9 pagesEnglish languagesale 10% offe-Library read for1 day
IEC 60601-2-46:2010 specifies safety requirements for operating tables, whether or not having electrical parts, including transporters, used for the transportation of the table top to or from the base or pedestal of an operating table with detachable table top. This second edition cancels and replaces the first edition published in 1998 and constitutes a technical revision. This edition of IEC 60601-2-46 was revised to align structurally with the 2005 edition of IEC 60601-1.
- Standard27 pagesEnglish languagesale 10% offe-Library read for1 day
IEC 80601-2-59:2017 applies to the basic safety and essential performance of screening thermographs intended to be used for the individual non-invasive febrile temperature screening of a human under controlled environmental conditions, hereafter referred to as ME equipment. This document sets laboratory characterization test limits for the screening thermograph. This edition includes the following significant technical changes with respect to the previous edition: a) updates of the normative references and the bibliography; b) expansion of the applicability to pandemic infectious diseases in general.
- Standard44 pagesEnglish languagesale 10% offe-Library read for1 day
NEXT ACTION: PUBLICATION EXPECTED BY 2019-10-31
2019-07-04 IP: BT decision D163/C033 - remove link to MDD
NEXT ACTION: TB Action by 21-22 May 2019 i.e. Next CLC TC 62 Plenary Meeting
2019-01-16 JO - In contact with CLC TC 62 to know if they will take a ballot to decouple this WI (upon BT approval) and proceed to PUB
- Amendment16 pagesEnglish languagesale 10% offe-Library read for1 day
IEC 80601-2-58:2014 applies to the basic safety and essential performance of lens removal devices and vitrectomy devices for ophthalmic surgery and associated accessories that can be connected to this medical electrical equipment, hereafter referred to as ME equipment. Hazards inherent in the intended physiological function of ME equipment or ME systems within the scope of this standard are not covered by specific requirements in this standard except in 7.2.13 and 8.4.1 of the general standard. This second edition includes changes in order to take into account the comments submitted during the approval of the first edition as a European Medical Device Directive, as well as the comments from other National Committees during the finalization of the first edition of this standard.
- Amendment14 pagesEnglish languagesale 10% offe-Library read for1 day
IEC 80601-2-49:2018 applies to basic safety and essential performance requirements of multifunction patient monitors, hereafter referred to as ME equipment or medical electrical systems. This particular standard applies to multifunction patient monitors intended for use in professional healthcare facilities as well as in the emergency medical service environment or the home healthcare environment. The scope of this document is restricted to ME equipment or medical electrical systems intended for connection to a single patient that has two or more physiological monitoring units. For purposes of this document, a pregnant mother and her fetus(es) are considered a single patient. This document does not specify requirements for individual physiological monitoring units such as ECG, invasive pressure and pulse oximetry. The particular standards related to these physiological monitoring units specify requirements from the perspective of stand-alone ME equipment. This particular standard addresses the additional requirements related to multifunction patient monitors. Multifunction patient monitors can be integrated into other ME equipment or medical electrical systems. When this is the case, other relevant standards also apply. This document does not apply to implantable parts of multifunction patient monitors. This first edition cancels and replaces the second edition of IEC 60601-2-49, published in 2011. This edition constitutes a technical revision to align with the current edition and Amendment to IEC 60601-1, new versions of collateral standards and amendments thereto. Major changes are in Clause 208 because many of the former requirements are now addressed by IEC 60601-1-8.
- Standard45 pagesEnglish languagesale 10% offe-Library read for1 day
IEC 60601-2-75:2017 applies to the basic safety and essential performance of photodynamic therapy and photodynamic diagnosis equipment. Hazards inherent in the intended physiological function of ME equipment or ME systems within the scope of this document are not covered by specific requirements in this document except in 7.2.13 and 8.4.1 of the general standard. This document applies to photodynamic therapy and photodynamic diagnosis equipment used for compensation or alleviation of disease, injury or disability. In the case of combined equipment (e.g. equipment additionally provided with a function or an applied part for the target area), such equipment also comply with any particular standard specifying safety requirements for the additional function. This particular standard does not apply to: - light therapy equipment intended for use in photothermal ablation, coagulation, and hyperthermia; - low-level laser therapy equipment not intended for use with a photosensitizer; - illumination equipment intended for use in observation, monitoring, and diagnosis, not intended for use with a photosensitizer. This document is to be used with IEC 60601-1:2005 and its amendment 1.
- Standard32 pagesEnglish languagesale 10% offe-Library read for1 day
NEW!IEC 60601-2-16:2018 is available as IEC 60601-2-16:2018 RLV which contains the International Standard and its Redline version, showing all changes of the technical content compared to the previous edition.IEC 60601-2-16:2018 applies to the basic safety and essential performance of haemodialysis, haemodiafiltration and haemofiltration equipment. IEC 60601-2-16:2018 does not take into consideration the dialysis fluid control system of Haemodialysis equipment using regeneration of dialysis fluid and central delivery systems. It does however take into consideration the specific safety requirements of such haemodialysis equipment concerning electrical safety and patient safety. IEC 60601-2-16:2018 specifies the minimum safety requirements for haemodialysis equipment. These devices are intended for use either by medical staff or for use by the patient or other trained personnel under the supervision of medical expertise. IEC 60601-2-16:2018 includes all electromedical equipment that is intended to deliver a haemodialysis, haemodiafiltration and haemofiltration treatment to a patient suffering from kidney failure. This fifth edition cancels and replaces the fourth edition of IEC 60601-2-16, published in 2012. This edition includes the following significant technical changes with respect to the previous edition: a) update of references to IEC 60601-1:2005 and IEC 60601-1:2005/AMD1:2012, of references and requirements to IEC 60601-1-2:2014, of references to IEC 60601-1-6:2010 and IEC 60601-1-6:2010/AMD1:2013, of references and requirements to IEC 60601-1-8:2006 and IEC 60601-1-8:2006/AMD1:2012, of references to IEC 60601-1-9:2007 and IEC 60601-1-9:2007/AMD1:2013, of references to IEC 60601-1-10:2007 and IEC 60601-1-10:2007/AMD1:2013 and of references to IEC 60601-1-11:2015; b) widening of the scope; c) editorial improvements; d) addition of requirements for anticoagulant delivery means; e) other few small technical changes.
- Standard80 pagesEnglish languagesale 10% offe-Library read for1 day
IEC 60601-2-76:2018 applies to the basic safety and essential performance of low energy ionized gas haemostasis equipment. Hazards inherent in the intended physiological function of ME Equipment or ME Systems within the scope of this document are not covered by specific requirements in this document except in 7.2.13 and 8.4.1 of the general standard. This particular standard amends and supplements IEC 60601-1:2005 and IEC 60601-1:2005/AMD1:2012.
- Standard29 pagesEnglish languagesale 10% offe-Library read for1 day
NEW!IEC 80601-2-30:2018 is available as IEC 80601-2-30:2018 RLV which contains the International Standard and its Redline version, showing all changes of the technical content compared to the previous edition.IEC 80601-2-30:2018 applies to the basic safety and essential performance of automated sphygmomanometers, hereafter referred to as ME equipment, which by means of an inflatable cuff, are used for non continuous indirect estimation of the blood pressure without arterial puncture. This document specifies requirements for the basic safety and essential performance for this ME equipment and its accessories, including the requirements for the accuracy of a determination. This document covers automatic electrically-powered ME equipment used for the intermittent, indirect estimation of the blood pressure without arterial puncture, including blood pressure monitors for the home healthcare environment. Requirements for indirect estimation of the blood pressure without arterial puncture ME equipment with an electrically-powered pressure transducer and/or displays used in conjunction with a stethoscope or other manual methods for determining blood pressure (non-automated sphygmomanometers) are specified in document ISO 81060-1. If a clause or subclause is specifically intended to be applicable to ME equipment only, or to ME systems only, the title and content of that clause or subclause will say so. If that is not the case, the clause or subclause applies both to ME equipment and to ME systems, as relevant. Hazards inherent in the intended physiological function of ME equipment or ME systems within the scope of this document are not covered by specific requirements in this document except in 201.11 and 201.105.3.3, as well as 7.2.13 and 8.4.1 of IEC 60601-1:2005. This second edition cancels and replaces the first edition published in 2009 and Amendment 1:2013. This edition constitutes a technical revision. This edition includes the following significant technical changes with respect to the previous edition: a) alignment with IEC 60601-1:2005/AMD1:2012 and IEC 60601-1-8:2006/AMD1:2012 , and with IEC 60601-1-2:2014 and IEC 60601-1-11:2015; b) referencing IEC 60601-1-10:2007 and IEC 60601-1-12; c) changing an operator-accessible cuff-sphygmomanometer connector from not compatible with the ISO 594 series to compatible with the ISO 80369 series; d) added additional requirements for public self-use sphygmomanometers; e) added a list of primary operating functions.
- Standard61 pagesEnglish languagesale 10% offe-Library read for1 day
NEW!IEC 60601-2-39:2018 is available as IEC 60601-2-39:2018 RLV which contains the International Standard and its Redline version, showing all changes of the technical content compared to the previous edition.IEC 60601-2-39:2018 applies to the basic safety and essential performance of peritoneal dialysis medical electrical equipment. It applies to peritoneal dialysis equipment intended for use either by medical staff or under the supervision of medical experts, including peritoneal dialysis equipment operated by the patient, regardless of whether the peritoneal dialysis equipment is used in a hospital or domestic environment. This third edition cancels and replaces the second edition of IEC 60601-2-39. It constitutes a technical revision.This edition includes the following significant technical changes with respect to the previous edition: - update of the references to IEC 60601-1:2005 and IEC 60601-1:2005/AMD1:2012, of references and requirements to IEC 60601-1-2:2014, of references to IEC 60601-1-6:2010 and IEC 60601-1-6:2010/AMD1:2013, of references and requirements to IEC 60601-1-8:2006 and IEC 60601-1-8:2006/AMD1:2012 and of references and requirements to IEC 60601-1-11:2015; - editorial improvements; - improvement of the essential performance requirements clause/subclauses; - new requirements for the interruption of the power supply.
- Standard31 pagesEnglish languagesale 10% offe-Library read for1 day
Establishes particular requirements for the safety of electromyographs and evoked response equipment as defined in clause 2 of this standard.
- Standard33 pagesEnglish languagesale 10% offe-Library read for1 day
IEC 60601-2-41:2021 applies to the BASIC SAFETY and ESSENTIAL PERFORMANCE of SURGICAL LUMINAIRES and LUMINAIRES FOR DIAGNOSIS, hereafter referred to as ME EQUIPMENT. If a clause or subclause is specifically intended to be applicable to ME EQUIPMENT only, or to ME SYSTEMS only, the title and content of that clause or subclause will say so. If that is not the case, the clause or subclause applies both to ME EQUIPMENT and to ME SYSTEMS, as relevant. This particular standard does not apply to – headlights; – endoscopes, laparoscopes and their light sources, which are covered by IEC 60601-2-18; – luminaires used in dentistry, which are covered by ISO 9680; – luminaires for general purposes, which are covered by IEC 60598-2-1 and IEC 60598-2-4; – luminaires dedicated to therapeutic purposes; – special purpose lights with different conditions of use such as light sources intended solely for decontamination of air and surfaces, UV lights for dermatological diagnosis, slit lamps for ophthalmology, lights for surgical microscopes and lights for surgical navigation systems; – lights connected to surgical instruments, such as luminous retractors; – luminaires for emergency lighting, which are covered by IEC 60598-2-22. SURGICAL LUMINAIRES and LUMINAIRES FOR DIAGNOSIS are medical devices and not general lighting equipment. IEC 60601-2-41:2021 cancels and replaces the second edition published in 2009 and Amendment 1:2013. This edition constitutes a technical revision. This edition includes the following significant technical changes with respect to the previous edition: a) revised the statement of essential performance; b) added exposure limits, test conditions, calculation methods and safety warnings related to photobiological hazards; c) removed the terms “MINOR SURGICAL LUMINAIRES” and “MAJOR SURGICAL LUMINAIRES”; d) added definitions of MAXIMUM ILLUMINANCE DISTANCE and REFERENCE DISTANCE and allowed MANUFACTURERS to measure some performance characteristics at the REFERENCE DISTANCE that they specify; e) replaced the region of acceptable chromaticity in (x,y) colour space with a requirement for Du,v; f) added a requirement for acceptable drift of the lighthead when attached to the suspension system; g) added a requirement for fluid ingress protection; h) revised Table 201.101 of IEC 60601-2-41:2009 and IEC 60601-2-41:2009/AMD1:2013 and moved it to Annex BB; i) specified a new device for measuring SHADOW DILUTION in a simulated cavity; j) specified test conditions for luminaires equipped with distance sensors.
- Standard63 pagesEnglish languagesale 10% offe-Library read for1 day
IEC 80601-2-77:2019 applies to the BASIC SAFETY and ESSENTIAL PERFORMANCE of ROBOTICALLY ASSISTED SURGICAL EQUIPMENT (RASE) and ROBOTICALLY ASSISTED SURGICAL SYSTEMS (RASS), referred to as ME EQUIPMENT and ME SYSTEMS together with their INTERACTION CONDITIONS and INTERFACE CONDITIONS.
- Standard55 pagesEnglish languagesale 10% offe-Library read for1 day
IEC 60601-2-50:2020 is available as IEC 60601-2-50:2020 RLV which contains the International Standard and its Redline version, showing all changes of the technical content compared to the previous edition.IEC 60601-2-50:2020 applies to the BASIC SAFETY and ESSENTIAL PERFORMANCE of INFANT PHOTOTHERAPY EQUIPMENT, as defined in 201.3.203, also referred to as ME EQUIPMENT. If a clause or subclause is specifically intended to be applicable to ME EQUIPMENT only, or to ME SYSTEMS only, the title and content of that clause or subclause will say so. If that is not the case, the clause or subclause applies both to ME EQUIPMENT and to ME SYSTEMS, as relevant. HAZARDS inherent in the intended physiological function of ME EQUIPMENT or ME SYSTEMS within the scope of this document are not covered by specific requirements in this document, except in 7.2.13 and 8.4.1 of the general standard. IEC 60601-2-50:2020 specifies safety requirements for INFANT PHOTOTHERAPY EQUIPMENT, but alternate methods of compliance with a specific clause by demonstrating equivalent safety will not be judged as non-compliant if the MANUFACTURER has demonstrated in his RISK MANAGEMENT FILE that the RISK presented by the HAZARD has been found to be of an acceptable level when weighed against the benefit of treatment from the device. IEC 60601-2-50:2020 does not apply to: - devices supplying heat via BLANKETS, PADS or MATTRESSES in medical use; for information, see IEC 60601-2-35; - INFANT INCUBATORS; for information, see IEC 60601-2-19; - INFANT TRANSPORT INCUBATORS; for information, see IEC 60601-2-20; - INFANT RADIANT WARMERS; for information, see IEC 60601-2-21. IEC 60601-2-50:2020 cancels and replaces the second edition published in 2009 and Amendment 1:2016. This edition constitutes a technical revision. IEC 60601-2-50:2020 includes the following significant technical change with respect to the previous edition: re-dating of normative references.
- Standard33 pagesEnglish languagesale 10% offe-Library read for1 day
IEC 60601-2-35:2020 applies to the BASIC SAFETY and ESSENTIAL PERFORMANCE of HEATING DEVICES using BLANKETS, PADS or MATTRESSES in medical use, also referred to as ME EQUIPMENT. HEATING DEVICES intended to prewarm a bed are included in the scope of this document. If a clause or subclause is specifically intended to be applicable to ME EQUIPMENT only, or to ME SYSTEMS only, the title and content of that clause or subclause will say so. If that is not the case, the clause or subclause applies both to ME EQUIPMENT and to ME SYSTEMS, as relevant. If a clause or subclause is specifically intended to be applicable to a specifically defined type of ME EQUIPMENT, as is the case with FORCED AIR DEVICES, then the title and content of that clause or subclause will say so. If that is not the case, the clause or subclause applies both to ME EQUIPMENT and to ME SYSTEMS, as relevant. HAZARDS inherent in the intended physiological function of ME EQUIPMENT or ME SYSTEMS within the scope of this document are not covered by specific requirements in this document, except in 7.2.13 and 8.4.1 of the general standard. IEC 60601-2-35:2020 does not apply to: - HEATING DEVICES intended for physiotherapy; - INFANT RADIANT WARMERS; for information, see IEC 60601-2-21; - INFANT INCUBATORS; for information, see IEC 60601-2-19; - INFANT TRANSPORT INCUBATORS, for information, see IEC 60601-2-20; - cooling devices. IEC 60601-2-35:2020 cancels and replaces IEC 80601-2-35 published in 2009 and Amendment 1:2016. This edition constitutes a technical revision. IEC 60601-2-35:2020 includes the following significant technical change with respect to the previous edition: re-dating of normative references.
- Standard73 pagesEnglish languagesale 10% offe-Library read for1 day
- Standard73 pagesEnglish languagesale 10% offe-Library read for1 day
IEC 60601-2-21:2020 is available as IEC 60601-2-21:2020 RLV which contains the International Standard and its Redline version, showing all changes of the technical content compared to the previous edition.IEC 60601-2-21:2020 applies to the BASIC SAFETY and ESSENTIAL PERFORMANCE of INFANT RADIANT WARMERS as defined in 201.3.204, also referred to as ME EQUIPMENT. If a clause or subclause is specifically intended to be applicable to ME EQUIPMENT only, or to ME SYSTEMS only, the title and content of that clause or subclause will say so. If that is not the case, the clause or subclause applies both to ME EQUIPMENT and to ME SYSTEMS, as relevant. HAZARDS inherent in the intended physiological function of ME EQUIPMENT or ME SYSTEMS within the scope of this document are not covered by specific requirements in this document, except in 7.2.13 and 8.4.1 of the general standard. IEC 60601-2-21:2020 specifies the safety requirements for INFANT RADIANT WARMERS, but alternate methods of compliance with a specific clause, by demonstrating equivalent safety, will not be judged as non-compliant, if the MANUFACTURER has demonstrated in his RISK MANAGEMENT FILE that the RISK presented by the HAZARD has been found to be of an acceptable level when weighed against the benefit of treatment from the device. This particular standard does not apply to: - devices supplying heat via BLANKETS, PADS or MATTRESSES in medical use; for information, see IEC 60601-2-35; - INFANT INCUBATORS; for information, see IEC 60601-2-19; - INFANT TRANSPORT INCUBATORS, for information, see IEC 60601-2-20; - INFANT PHOTOTHERAPY EQUIPMENT, for information, see IEC 60601-2-50. SKIN TEMPERATURE SENSORS which are applied to operate a BABY CONTROLLED RADIANT WARMER including the displayed value are not considered to be a CLINICAL THERMOMETER in the sense of the particular standard ISO 80601-2-56. IEC 60601-2-21:2020 cancels and replaces the second edition published in 2009 and Amendment 1:2016. This edition constitutes a technical revision. IEC 60601-2-21:2020 includes the following significant technical change with respect to the previous edition: alignment with IEC 60601-1:2005 and IEC 60601-1:2005/AMD1:2012.
- Standard40 pagesEnglish languagesale 10% offe-Library read for1 day
IEC 60601-2-19:2020 is available as IEC 60601-2-19:2020 RLV which contains the International Standard and its Redline version, showing all changes of the technical content compared to the previous edition.IEC 60601-2-19:2020 applies to the BASIC SAFETY and ESSENTIAL PERFORMANCE of INFANT INCUBATORS, as defined in 201.3.209, also referred to as ME EQUIPMENT. If a clause or subclause is specifically intended to be applicable to ME EQUIPMENT only, or to ME SYSTEMS only, the title and content of that clause or subclause will say so. If that is not the case, the clause or subclause applies both to ME EQUIPMENT and to ME SYSTEMS, as relevant. HAZARDS inherent in the intended physiological function of ME EQUIPMENT or ME SYSTEMS within the scope of this document are not covered by specific requirements in this document, except in 7.2.13 and 8.4.1 of the general standard. IEC 60601-2-19:2020 specifies safety requirements for INFANT INCUBATORS, but alternate methods of compliance with a specific clause, by demonstrating equivalent safety, will not be judged as non-compliant, if the MANUFACTURER has demonstrated in his RISK MANAGEMENT FILE that the RISK presented by the HAZARD has been found to be of an acceptable level when weighed against the benefit of treatment from the device. IEC 60601-2-19:2020 does not apply to: - INFANT TRANSPORT INCUBATORS; for information, see IEC 60601-2-20; - INFANT RADIANT WARMERS; for information, see IEC 60601-2-21; - devices supplying heat via BLANKETS, PADS or MATTRESSES in medical use; for information, see IEC 60601-2-35; - INFANT PHOTOTHERAPY EQUIPMENT; for information see IEC 60601-2-50. SKIN TEMPERATURE SENSORS which are applied to operate a BABY CONTROLLED INCUBATOR including the displayed value are not considered to be a CLINICAL THERMOMETER in the sense of the particular standard ISO 80601-2-56. IEC 60601-2-19:2020 cancels and replaces the second edition published in 2009 and Amendment 1:2016. This edition constitutes a technical revision. IEC 60601-2-19:2020 includes the following significant technical change with respect to the previous edition: re-dating of normative references.
- Standard42 pagesEnglish languagesale 10% offe-Library read for1 day
NEW!IEC 60601-2-2:2017 is available as IEC 60601-2-2:2017 RLV which contains the International Standard and its Redline version, showing all changes of the technical content compared to the previous edition.IEC 60601-2-2:2017 applies to the basic safety and essential performance of HF surgical equipment and HF surgical accessories. HF surgical equipment having a rated output power not exceeding 50 W (for example for micro-coagulation, or for use in dentistry or ophthalmology) is exempt from certain of the requirements of this particular standard. These exemptions are indicated in the relevant requirements. The object of this particular standard is to establish particular basic safety and essential performance requirements for HF surgical equipment and HF surgical accessories. This sixth edition cancels and replaces the fifth edition published in 2009. This edition constitutes a technical revision. This edition includes the following significant technical changes with respect to the previous edition: - refinement and additions to the defined terms; - additional separation of the requirements for HF surgical equipment and HF surgical accessories; - a new requirement for adult neutral electrodes to be contact quality monitoring neutral electrodes; - new requirements for devices that have or use a high current mode.
- Standard93 pagesEnglish languagesale 10% offe-Library read for1 day
IEC 60601-2-20:2020 is available as IEC 60601-2-20:2020 RLV which contains the International Standard and its Redline version, showing all changes of the technical content compared to the previous edition.IEC 60601-2-20:2020 applies to the BASIC SAFETY and ESSENTIAL PERFORMANCE of INFANT TRANSPORT INCUBATOR equipment, as defined in 201.3.208, also referred to as ME EQUIPMENT. If a clause or subclause is specifically intended to be applicable to ME EQUIPMENT only, or to ME SYSTEMS only, the title and content of that clause or subclause will say so. If that is not the case, the clause or subclause applies both to ME EQUIPMENT and to ME SYSTEMS, as relevant. HAZARDS inherent in the intended physiological function of ME EQUIPMENT or ME SYSTEMS within the scope of this document are not covered by specific requirements in this document, except in 7.2.13 and 8.4.1 of the general standard. IEC 60601-2-20:2020 specifies safety requirements for INFANT TRANSPORT INCUBATORS, but alternate methods of compliance with a specific clause, by demonstrating equivalent safety, will not be judged as non-compliant, if the MANUFACTURER has demonstrated in his RISK MANAGEMENT FILE that the RISK presented by the HAZARD has been found to be of an acceptable level when weighed against the benefit of treatment from the device. This particular standard does not apply to: - devices supplying heat via BLANKETS, PADS or MATTRESSES in medical use; for information, see IEC 60601-2-35; - INFANT INCUBATORS which are not INFANT TRANSPORT INCUBATOR; for information see IEC 60601-2-19; - INFANT RADIANT WARMERS; for information, see IEC 60601-2-21; - INFANT PHOTOTHERAPY; for information, see IEC 60601-2-50. IEC 60601-2-20:2020 cancels and replaces the second edition published in 2009 and Amendment 1:2016. This edition constitutes a technical revision. IEC 60601-2-20:2020 includes the following significant technical change with respect to the previous edition: re-dating of normative references.
- Standard47 pagesEnglish languagesale 10% offe-Library read for1 day
IEC 80601-2-26:2019 applies to the BASIC SAFETY and ESSENTIAL PERFORMANCE of ELECTROENCEPHALOGRAPHS as defined in 201.3.204, hereafter also referred to as ME EQUIPMENT or ME SYSTEM. This document is applicable to ELECTROENCEPHALOGRAPHS intended for use in professional healthcare facilities, the EMERGENCY MEDICAL SERVICES ENVIRONMENT or the HOME HEALTHCARE ENVIRONMENT. This document does not cover requirements for other equipment used in electroencephalography such as: – phono-photic stimulators; – EEG data storage and retrieval; – ME EQUIPMENT particularly intended for monitoring during electro-convulsive therapy. If a clause or subclause is specifically intended to be applicable to ME EQUIPMENT only, or to ME SYSTEMS only, the title or content of that clause or subclause will say so. If that is not the case, the clause or subclause applies both to ME EQUIPMENT and to ME SYSTEMS, as follows. The clause or subclause applies to ME EQUIPMENT, as default. For ME EQUIPMENT with the corresponding safety measure or function not completely integrated into the ME EQUIPMENT but instead implemented in an ME SYSTEM, the ME EQUIPMENT MANUFACTURER specifies in the ACCOMPANYING DOCUMENTS which functionality and safety requirements are provided by the ME SYSTEM to comply with this document. The ME SYSTEM is verified accordingly. HAZARDS inherent in the intended physiological function of ME EQUIPMENT or ME SYSTEMS within the scope of this document are not covered by specific requirements in this document. IEC 80601-2-26:2019 cancels and replaces the third edition of IEC 60601-2-26 published in 2012. This edition constitutes a technical revision to align with Amendment 1:2012 of IEC 60601-1:2005, new versions of collateral standards and amendments thereto.
- Standard39 pagesEnglish languagesale 10% offe-Library read for1 day
IEC 80601-2-78:2019 applies to the general requirements for BASIC SAFETY and ESSENTIAL PERFORMANCE of MEDICAL ROBOTS that physically interact with a PATIENT with an IMPAIRMENT to support or perform REHABILITATION, ASSESSMENT, COMPENSATION or ALLEVIATION related to the PATIENT’S MOVEMENT FUNCTIONS, as intended by the MANUFACTURER. IEC 80601-2-78:2019 does not apply to • external limb prosthetic devices (use ISO 22523), • electric wheelchairs (use ISO 7176 (all parts)), • diagnostic imaging equipment (e.g. MRI, use IEC 60601-2-33), and • personal care ROBOTS (use ISO 13482).
- Standard79 pagesEnglish languagesale 10% offe-Library read for1 day
IEC 60601-2-31:2020 is available as IEC 60601-2-31:2020 RLV which contains the International Standard and its Redline version, showing all changes of the technical content compared to the previous edition. IEC 60601-2-31:2020 applies to the BASIC SAFETY and ESSENTIAL PERFORMANCE of EXTERNAL PACEMAKERS powered by an INTERNAL ELECTRICAL POWER SOURCE, hereafter referred to as ME EQUIPMENT. This document applies to PATIENT CABLES as defined in 201.3.209, but does not apply to LEADS as defined in 201.3.206. HAZARDS inherent in the intended physiological function of ME EQUIPMENT within the scope of this document are not covered by specific requirements in this document except in 7.2.13 and 8.4.1 of the general standard. This document does not apply to the implantable parts of ACTIVE IMPLANTABLE MEDICAL DEVICES covered by ISO 14708-1. This document does not apply to EXTERNAL PACEMAKERS which can be connected directly or indirectly to a SUPPLY MAINS. This document does not apply to transthoracic and oesophageal pacing ME EQUIPMENT and antitachycardia ME EQUIPMENT. IEC 60601-2-31:2020 cancels and replaces the second edition published in 2008 and Amendment 1:2011. This edition constitutes a technical revision. This edition includes the following significant technical changes with respect to the previous edition: a) The requirement for testing for energy reduction has been removed; b) The test for exposure to external defibrillation has been completely revised; c) The exclusion for testing ESD immunity only with respect to air discharges has been removed; d) Alignment with the latest edition of ISO 14708-2 for pacemakers, as well as the associated EMC standard ISO 14117; e) Additional rationale for all changes.
- Standard57 pagesEnglish languagesale 10% offe-Library read for1 day
IEC 60601-2-83:2019 is applicable to the BASIC SAFETY and ESSENTIAL PERFORMANCE of HOME LIGHT THERAPY EQUIPMENT, intended for use in the HOME HEALTHCARE ENVIRONMENT. HOME LIGHT THERAPY EQUIPMENT is typically used by a LAY OPERATOR. The scope of this document includes all light sources except laser.
- Standard34 pagesEnglish languagesale 10% offe-Library read for1 day
IEC 80601-2-60:2019 applies to the BASIC SAFETY and ESSENTIAL PERFORMANCE OF DENTAL UNITS, DENTAL PATIENT CHAIRS, DENTAL HANDPIECES AND DENTAL OPERATING LIGHTS, hereafter referred to as DENTAL EQUIPMENT. If a clause or subclause is specifically intended to be applicable to ME EQUIPMENT only, or to ME SYSTEMS only, the title and content of that clause or subclause will say so. If that is not the case, the clause or subclause applies both to ME EQUIPMENT and to ME SYSTEMS, as relevant. HAZARDS inherent in the intended physiological function of ME EQUIPMENT or ME SYSTEMS within the scope of this document are not covered by specific requirements in this document except in 7.2.13 and 8.4.1 of the general standard. IEC 80601-2-60:2019 cancels and replaces the first edition published in 2012. This edition constitutes a technical revision. This edition includes the following significant technical changes with respect to the previous edition: a) alignment with IEC 60601-1:2005 and IEC 60601-1:2005/AMD1:2012.
- Standard45 pagesEnglish languagesale 10% offe-Library read for1 day
- Amendment7 pagesEnglish languagesale 10% offe-Library read for1 day
- Amendment5 pagesEnglish languagesale 10% offe-Library read for1 day
- Amendment5 pagesEnglish languagesale 10% offe-Library read for1 day
IEC 60601-2-25:2011 applies to the basic safety and essential performance of electrocardiographs intended by themselves or as a part of a medical electrical system, for the production of electrocardiogram reports for diagnostic purposes. The equipment intended for use under extreme or uncontrolled environmental conditions outside the hospital environment or physician's office, such as in ambulances and air transport, shall comply with this particular standard. Additional standards may apply to Medical electrical equipment for those environments of use. Not included within the scope of this particular standard are: a) the part of Medical electrical equipment that provides vectorcardiographic loops; b) ambulatory electrocardiographic medical electrical equipment covered by IEC 60601-2-47 where not intended for obtaining electrocardiogram reports for diagnostic purposes; c) cardiac monitors covered by IEC 60601-2-27 where not intended for obtaining electrocardiographic reports for diagnostic purposes. This second edition cancels and replaces the first edition of IEC 60601-2-25, published in 1993 and the first edition of IEC 60601-2-51, published in 2003. Updating the particular standards to refer to the third edition of the general standard provided the opportunity to merge the first editions of IEC 60601-2-25 and IEC 60601-2-51 into one standard. Reformatting and technical changes were both made. This second edition of IEC 60601-2-25 constitutes a technical revision of both those standards.
- Standard100 pagesEnglish languagesale 10% offe-Library read for1 day
IEC 60601-2-23:2011 applies to the basic safety and essential performance of transcutaneous partial pressure monitoring equipment. It applies to transcutaneous monitors used with adults, children and neonates, and it includes the use of these devices in foetal monitoring during birth. IEC 60601-2-23:2011 does not apply to haemoglobin saturation oximeters or to devices applied to surfaces of the body other than the skin (for example conjunctiva, mucosa). This third edition cancels and replaces the second edition published in 1999 and constitutes a technical revision. This edition of IEC 60601-2-23 was revised to align structurally with the 2005 edition of IEC 60601-1.
- Standard51 pagesEnglish languagesale 10% offe-Library read for1 day
IEC 60601-2-18:2009 establishes particular basic safety and essential performance requirements for endoscopic equipment. The minimum safety requirements specified in this particular standard are considered to provide for a practical degree of safety in the operation of endoscopic equipment. IEC 60601-2-18:2009 amends and supplements IEC 60601-1:2005. This third edition cancels and replaces the second edition, published in 1996, and its Amendment 1 (2000). This edition constitutes a technical revision and has been aligned or harmonized with IEC 60601-1:2005. The main changes with respect to the previous edition include: - alignment of requirements with IEC 60601-1:2005; - inclusion of essential performance requirements; - the inclusion of energized endoscopes and energized endotherapy devices used through second and subsequent punctures within the scope of the standard; - reference to IEC 60601-2-2 for the dielectric strength testing of HF energized endotherapy devices, rather than defining different tests.
- Standard51 pagesEnglish languagesale 10% offe-Library read for1 day
IEC 60601-2-5:2009 applies to the basic safety and essential performance of ultrasonic physiotherapy equipment employing a single plane unfocused circular transducer per treatment head, producing static beams perpendicular to the face of the treatment head. This standard can also be applied to ultrasonic physiotherapy equipment used for compensation or alleviation of disease, injury or disability. This third edition cancels and replaces the second edition published in 2000. This edition constitutes a technical revision. The numbering was revised to agree with IEC 60601-1:2005 (third edition). Beyond this, essential performance characteristics are defined in 201.4.3.101, guidance on maintenance is added in 201.7.9.2.1, a new requirement regarding dielectric withstand was added in 201.8.8.3. The clause on transducer surface temperature rise, 201.11, has been modified to allow for simulated use conditions. Measurements of ultrasound-related parameters are now referenced to IEC 61689:2007 (second edition). The most important change in the ultrasound-related parameters is the definition of effective radiating area, 201.3.207. This change will also affect the value of the effective intensity and its uncertainty.
- Standard37 pagesEnglish languagesale 10% offe-Library read for1 day
IEC 80601-2-71:2015(E) applies to the BASIC SAFETY and ESSENTIAL PERFORMANCE of FUNCTIONAL NIRS EQUIPMENT intended to be used by themselves, or as a part of an ME SYSTEM, for the production of FUNCTIONAL NIRS EQUIPMENT output for adjunctive diagnostic purposes, hereinafter referred to as ME EQUIPMENT.
- Standard40 pagesEnglish languagesale 10% offe-Library read for1 day