Sterilization of health care products — Chemical indicators — Part 5: Class 2 indicators for air removal test sheets and packs

Stérilisation des produits de santé — Indicateurs chimiques — Partie 5: Indicateurs de classe 2 pour draps et paquets prépliés servant à l'essai d'enlèvement d'air

General Information

Status
Withdrawn
Publication Date
01-Nov-2000
Withdrawal Date
01-Nov-2000
Current Stage
9599 - Withdrawal of International Standard
Completion Date
15-Mar-2007
Ref Project

Relations

Buy Standard

Standard
ISO 11140-5:2000 - Sterilization of health care products -- Chemical indicators
English language
13 pages
sale 15% off
Preview
sale 15% off
Preview

Standards Content (Sample)

INTERNATIONAL ISO
STANDARD 11140-5
First edition
2000-11-01
Sterilization of health care products —
Chemical indicators —
Part 5:
Class 2 indicators for air removal test
sheets and packs
Stérilisation des produits de santé — Indicateurs chimiques —
Partie 5: Indicateurs de classe 2 pour draps et paquets prépliés servant à
l'essai d'enlèvement d'air
Reference number
ISO 11140-5:2000(E)
© ISO 2000

---------------------- Page: 1 ----------------------
ISO 11140-5:2000(E)
PDF disclaimer
This PDF file may contain embedded typefaces. In accordance with Adobe's licensing policy, this file may be printed or viewed but shall not be
edited unless the typefaces which are embedded are licensed to and installed on the computer performing the editing. In downloading this file,
parties accept therein the responsibility of not infringing Adobe's licensing policy. The ISO Central Secretariat accepts no liability in this area.
Adobe is a trademark of Adobe Systems Incorporated.
Details of the software products used to create this PDF file can be found in the General Info relative to the file; the PDF-creation parameters
were optimized for printing. Every care has been taken to ensure that the file is suitable for use by ISO member bodies. In the unlikely event
that a problem relating to it is found, please inform the Central Secretariat at the address given below.
© ISO 2000
All rights reserved. Unless otherwise specified, no part of this publication may be reproduced or utilized in any form or by any means, elec-
tronic or mechanical, including photocopying and microfilm, without permission in writing from either ISO at the address below or ISO's mem-
ber body in the country of the requester.
ISO copyright office
Case postale 56 • CH-1211 Geneva 20
Tel. + 41 22 749 01 11
Fax + 41 22 749 09 47
E-mail copyright@iso.ch
Web www.iso.ch
Printed in Switzerland
©
ii ISO 2000 – All rights reserved

---------------------- Page: 2 ----------------------
ISO 11140-5:2000(E)
Contents Page
1 Scope . 1
2 Normative references . 1
3 Terms and definitions . 1
4 General requirements . 2
5 Air removal indicator . 2
6 Air removal indicator system . 3
7 Packaging and labelling . 3
8 Quality assurance . 4
9 Sampling conditioning . 4
Annexes
A Determination of the degree of contrast between the colour of the substrate and the indicator agent. 5
B Method of demonstrating uniform colour change on exposure to saturated steam . 6
C Method for evaluating indicator colour change on exposure to dry heat. 7
D Method for evaluating transfer of indicator to standard test pack. 8
E Standard test pack . 9
F Method of demonstrating non-uniform colour change on exposure to a standard fault condition . 10
G Steam exposure apparatus. 11
H Air porosity apparatus . 12
Bibliography. 13
©
ISO 2000 – All rights reserved iii

---------------------- Page: 3 ----------------------
ISO 11140-5:2000(E)
Foreword
ISO (the International Organization for Standardization) is a worldwide federation of national standards bodies (ISO
member bodies). The work of preparing International Standards is normally carried out through ISO technical com-
mittees. Each member body interested in a subject for which a technical committee has been established has the
right to be represented on that committee. International organizations, governmental and non-governmental, in liai-
son with ISO, also take part in the work. ISO collaborates closely with the International Electrotechnical Commission
(IEC) on all matters of electrotechnical standardization.
International Standards are drafted in accordance with the rules given in the ISO/IEC Directives, Part 3.
Draft International Standards adopted by the technical committees are circulated to the member bodies for voting.
Publication as an International Standard requires approval by at least 75 % of the member bodies casting a vote.
Attention is drawn to the possibility that some of the elements of this part of ISO 11140 may be the subject of patent
rights. ISO shall not be held responsible for identifying any or all such patent rights.
International Standard ISO 11140-5 was prepared by Technical Committee ISO/TC 198, Sterilization of health care
products.
ISO 11140 consists of the following parts, under the general title Sterilization of health care products — Chemical
indicators:
— Part 1: General requirements
— Part 2: Test equipment and methods
— Part 3: Class 2 indicators for steam penetration test sheets
— Part 4: Class 2 indicators for steam penetration test packs
— Part 5: Class 2 indicators for air removal test sheets and packs
Annexes A, B, C, D, E, F, G and H form a normative part of this part of ISO 11140.
©
iv ISO 2000 – All rights reserved

---------------------- Page: 4 ----------------------
ISO 11140-5:2000(E)
Introduction
The air removal test is used to evaluate the efficacy of air removal during the pre-vacuum phase of a pre-vacuum
sterilization cycle. Retention of air due to an inefficient air removal stage or the presence of an air leak during the air
removal stage are circumstances which can lead to failure of the test. This part of ISO 11140 describes the require-
ments for Class 2 indicators and systems for air removal test.
For a description of the classes of chemical indicators, see ISO 11140-1.
Because a range of different tests in different countries have historically been named the Bowie-Dick test, no
reference to this term is used in ISO 11140-3, ISO 11140-4 or ISO 11140-5.
The difference between the steam penetration test (ISO 11140-3 and ISO 11140-4) and the air removal test
(ISO 11140-5) is described in the chemical indicator guidance document (ISO 15882).
©
ISO 2000 – All rights reserved v

---------------------- Page: 5 ----------------------
INTERNATIONAL STANDARD ISO 11140-5:2000(E)
Sterilization of health care products — Chemical indicators —
Part 5:
Class 2 indicators for air removal test sheets and packs
1 Scope
This part of ISO 11140 specifies the requirements for an indicator and alternative test system used to evaluate the
effectiveness of air removal during the pre-vacuum phase of pre-vacuum steam sterilization cycles.
Additionally, this part of ISO 11140 includes test methods and equipment used to meet these performance require-
ments.
2 Normative references
The following normative documents contain provisions which, through reference in this text, constitute provisions of
this part of ISO 11140. For dated references, subsequent amendments to, or revisions of, any of these publications
do not apply. However, parties to agreements based on this part of ISO 11140 are encouraged to investigate the
possibility of applying the most recent editions of the normative documents indicated below. For undated references,
the latest edition of the normative document referred to applies. Members of ISO and IEC maintain registers of cur-
rently valid International Standards.
ISO 5-4:1995, Photography — Density measurements — Part 4: Geometric conditions for reflection density.
ISO 11140-1:1995, Sterilization of health care products — Chemical indicators — Part 1: General requirements.
IEC 60584-2:1982, Thermocouples — Part 2: Tolerances.
IEC 60751:1983, Industrial platinum resistance thermometer sensors.
3 Terms and definitions
For the purposes of this part of ISO 11140, the terms and definitions given in ISO 11140-1 and the following apply.
3.1
air removal indicator
indicator to be used in the standard test pack to determine the efficacy of the air removal phase in the steam sterili-
zation process
NOTE See annex E for description of the standard test pack.
3.2
air removal indicator system
specific test load containing an indicator to determine the efficacy of the air removal phase in the steam sterilization
process
NOTE The system may be user-assembled or pre-assembled. The test load may be disposable, for limited reuse, or reusable.
©
ISO 2000 – All rights reserved 1

---------------------- Page: 6 ----------------------
ISO 11140-5:2000(E)
3.3
equilibration time
period which elapses between the attainment of the sterilization temperature in the sterilizer chamber and the attain-
ment of the sterilization temperature in all parts of the load
4 General requirements
Unless otherwise specified in this part of ISO 11140, the general requirements given in ISO 11140-1 apply.
5 Air removal indicator
5.1 Format
5.1.1 The indicator agent shall be uniformly distributed on its substrate to cover not less than 30 % of the test area
of the substrate.
NOTE The pattern of indicator agent distribution should allow easy judgement of the uniformity of the colour change.
−1 −2
5.1.2 The indicator shall have an air porosity greater than 0,026 ml� s � mm . Compliance shall be tested in
accordance with annex H.
5.1.3 The substrate of the indicator shall have a uniform ground colour which provides a difference in colour density
of not less than 0,3 between ground colour and either the changed or the unchanged indicator, as specified by the
manufacturer, when the difference in colour density is determined using a reflective densitometer.
Compliance shall be tested in accordance with annex A.
5.1.4 The indicator dimensions shall be (200� 20) mm� (275� 25) mm.
5.2 Performance
5.2.1 The indicator shall show a uniform colour change (as specified by the manufacturer) after exposure to satu-
� �
rated steam at 134 C for 3,5 min� 5 s and/or exposure to saturated steam at 121 C for 15 min� 5s, or such
other combinations of time and temperature as the manufacturer shall specify for the intended use of product. In all
+1,5

cases the permitted tolerance on the test temperature shall be C and the time given shall be the time within
0
which the colour change shall occur.
Compliance shall be tested in accordance with annex B.
5.2.2 When placed in the centre of the standard test pack as given in annex E, the indicator shall show a non-uni-
� �
+1

form colour change when the temperature at the centre of the standard test pack is 2 C lower than the tem-
0

perature of the chamber drain at the beginning of the final 1 min at 134 C, or at the beginning of the final 5 min of a

15 min cycle at 121 C of the exposure phase of the steam exposure apparatus. This is the standard fault condition
generated by inadequate air removal from the chamber. Any other combination of time and temperature stated by the
manufacturer shall exhibit a similar response at the beginning of the final 30 % of the exposure time.
Compliance shall be tested in accordance with annex F.

5.2.3 After exposure to dry heat at (140� 2) C for 30 min, the indicator shall show either no change or a change
that is markedly different from the change occurring after exposure to a steam sterilization process.
Compliance shall be tested in accordance with annex C.
5.2.4 Transfer of the indicator agent to the material of the standard test pack in which it is intended to be used, shall
not compromise the result of the test.
©
2 ISO 2000 – All rights reserved

---------------------- Page: 7 ----------------------
ISO 11140-5:2000(E)
Compliance shall be tested in accordance with annex D.
NOTE Although some transfer may be possible without adversely affecting the performance of the indicator or test pack, there
are currently no test methods available to verify the acceptable limits of indicator agent transfer.
5.2.5 The indicator shall comply with the requirements of this part of ISO 11140 for the duration of the shelf life
specified by the manufacturer.
NOTE A documented accelerated ageing procedure may be used in demonstrating compliance.
6 Air removal indicator system
6.1 Format
6.1.1 The indicator agent shall be uniformly distributed on its substrate to cover not less than 30 % of the test area
of the substrate.
6.1.2 The substrate of the indicator shall have a uniform ground colour which provides a difference in colour density
of not less than 0,3 between ground colour and either the changed or the unchanged indicator agent, as specified by
the manufacturer, when the difference in colour density is determined using a reflective densitometer.
Compliance shall be tested in accordance with annex A.
6.2 Performance
6.2.1 The indicator shall show a uniform colour change complying with 6.1.2 both after exposure to saturated steam
� � � �
+1,5 +1,5
� �
at 134 C for 3,5 min� 5s and/or exposure to saturated steam at 121 C for 15 min� 5s, or such
0 0
other combinations of time and temperature as the manufacturer shall specify for the intended use of product. In all
+1,5�
cases the permitted tolerance on the test temperature shall be C and the time given shall be the time within
0
which the colour change shall occur.
Compliance shall be tested in accordance with annex B.
6.2.2 After exposure to conditions used to produce a standard fault condition as given in 5.2.2 in the standard test
pack as given in annex E, the indicator shall show a non-uniform colour change.
Compliance shall be tested in accordance with annex F.

6.2.3 After exposure to dry heat at (140� 2) C for 30 min, the chemical indicator system shall show either no
change or a change that is markedly different from the change occurring after exposure to a steam sterilization pro-
cess. Compliance shall be tested in accordance with annex C.
6.2.4 Transfer of the indicator agent to the chemical indicator system shall not compromise the result of the test.
6.2.5 The indicator shall comply with the require
...